Sunteți pe pagina 1din 5

Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE DS gantry floor anchored? gantry cable access? standard cable lengths voltage/amps/MDC cat.

No. UPS recommended ceiling height minimum ceiling height recommended room size minimum room size delivery opening YES REAR A-Gen=21' OCGen=33' Gen-Gtry=16' OC-Gtry=16' 208V/30A/E4502EE E4502JS 9'-0" 8'-2 1/2" 12'-10" x 12'-2" 9'-8" X 9'-8" 28" X 79"

Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE 2000D gantry floor anchored? gantry cable access? YES REAR A-Gen=21' OCGen=16' Gen-Gtry=16' Aws-Gtry=16' 208V/30A/E4502EE E4502JK 9'-0" 8'-2 1/2" 14'-6" X 14'-0" 11'-6" x 11'-6" 30" X 77"

standard cable lengths voltage/amps/MDC cat. No. UPS recommended ceiling height minimum ceiling height recommended room size minimum room size delivery opening

Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE DMR gantry floor anchored? gantry cable access? standard cable lengths voltage/amps UPS recommended ceiling height minimum ceiling height recommended room size minimum room size delivery opening YES REAR A-Gen=21' OC-Gen=16' Gen-Gtry=16' 208V/30A E4502JK 9'-0" 8'-2 1/2" 11'-6" X 10'-0" 10'-0" x 10'-0" 30" X 77"

Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE INSTRUMENTARIUM (DIAMOND) gantry floor SEISMIC Areas anchored? gantry cable REAR access? voltage/amps recommended ceiling height minimum ceiling height recommended room size minimum room size delivery opening 208V/20A 9'-0" 8'-2 1/2" 16'-4" X 11'-0" 10'-0" x 10'-0" 36" X 82"

Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE INSTRUMENTARIUM (PERFORMA) gantry floor SEISMIC Areas anchored? gantry cable REAR access? voltage/amps recommended ceiling height minimum ceiling height recommended room size minimum room size delivery opening 208V/20A 9'-0" 8'-2 1/2" 16'-4" X 11'-0" 10'-0" x 10'-0" 36" X 82"

S-ar putea să vă placă și