QA
Procedure
s
For LINAC
Tolerance
Parameter
Frequency
Procedure
Daily
Dosimetry
Xray output constancy
Electron output constancy
Mechanical
Localizing lasers
Distance indicator
Safety
Door interlock
Audiovisual monitor
Dosimetry
Xray output constancy
Monthly
Electron Constancy
Backup monitor constancy
Xray central axis PDD, TAR
NonIMRT
s
IMRT
3%
3%
2mm
2mm
1.5mm
2mm
2% IMRT
2% SRS
dose rate
2%
2%
2%
Rate
2mm at
depth
2%
constancy
Electron beam flatness
3%
3%
Functional
Functional
interlocks
Mechanical checks
Light/radiation field
2mm or 1%
coincidence
Gantry/collimator angle
on side
1 degree
indicators
1mm
1mm
Functional
Functional
constancy
Electron central axis
constancy
Xray and electron symmetry
Safety interlocks
Emergency off switches
Wedge, electron cone
SRS SBRT
Wedge position
Tray position
Applicator position
Field size indicators
Cross-hair centering
2mm
2mm
2mm
2mm
2mm
2mm/1
diameter
2mm/1
indicators
Wedge and block inspection
Jaw symmetry
Field light intensity
Localizing lasers
degree
+/- 2mm
degree
Functional
2mm
Functional
+/- 1mm
1mm/0.5
degree
Less then
1mm
Annual
Dosimetry
Xray/electron output
calibration constancy
2%
(>5mu)
5% (2-4mu)
5% (2-
2% (>5mu)
4mu)
2%
(<5mu)
2%
2%
electron cones
Central axis parameter
2%
constancy PDD,TAR
Off-axis factor constancy
Transmission factor
2%
2%
2%
constancy
Monitor chamber linearity
Xray output constancy
1%
2%
2%
Functional
2mm
diameter
2mm
diameter
2mm
2mm
diameter
2mm
1mm
diameter
diameter
diameter
Isocenter
Coincidence of radiation and
2mm
mechanical isocenter
Tabletop Sag
Vertical travel of table
Respiratory Gating
Beam energy constancy
Temporal accuracy of
diameter
2mm
2mm
2%
100 ms
phase/amplitude
Calibration of surrogate
100 ms
phase/amplitude
Interlock testing
Functioning
QA
Procedure
s for OBI
Planar kv and MV imaging
Non-
SRS/SBR
SRS/SB
RT
Daily
Collision interlock
Function
Functiona
Positioning/repositioning
Imaging and treatment
al
2mm
2mm
l
1mm
1mm
2mm
1mm
2mm
Baseline
2mm
Baseline
5mm
Baseline
5mm
Baseline
coordinate coincidence
Monthly
Imaging and treatment
coincidence (4 cardinal
angles)
Scaling
Spatial Resolution, Contrast,
and uniformity
Annual
Full range of travel SDD
Imaging dose, beam
quality/energy
CBCT (kV and MV)
Daily
Collision interlock
Function
Functiona
Positioning/reposiotioning
Imaging and treatment
al
2mm
2mm
l
1mm
1mm
2mm
Baseline
1mm
Baseline
Baseline
Baseline
coordinate coincidence
Monthly
Geometric distortion
Spatial resolution, contrast,
HU constancy, and Uniformity.
Annual
Imaging Dose
Tolerance Parameters
0 5 HU
X or y direction 1mm
0 5 HU
X and Y direction 1mm
Within 0 5 HU
0 5 HU
Manufacturer Specs
Manufacturer Specs
Manufacturer Specs
Monthly
Annual
As needed
or software upgrades.
Daily/Morning QA for HDR Afterloader
DATE
TEST CRITERIA
Viewing and intercom systems functional
Emergency container and stopwatch present
TCS self-test successful
All lights on TCS console function
Date and time in TCS correct (to within 1 hr)
Decayed source activity in TCS correct
Survey meter battery OK, meter ON
Door interlocks function, prevent treatment
Source exposure indicators at TCS and door work
Source positioning accuracy (+/-1mm)
Interrupt button functions
Emergency off button functions
Timer ends treatment correctly
Timer accuracy (vs. stopwatch)
Printer Operational
Acceptable
time
(YES/NO)
(sec)
20
60
stopwatch (sec)
yes
yes
Programmed
position (cm)
130
125
121
Interlock
Differenc
Offset (YES/NO)
e (mm)
0
0
0
YES/NO
YES/NO
Functions
i) Door Interlock
O.K.
ii) Missing
Applicator
O.K.
Interlock O.K.
vi) Emergency Off and Interrupt
button O.K.
HDR Accessories
ii) Transfer tubes checked .OK
ii) Patient monitoring system
O.K.
iii) Audio and visual system O.K.
1. Khan FM. The Physics of Radiation Therapy. Lippincott Williams & Wilkins; 2010.
2. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: quality assurance of medical
accelerators. Med Phys. 2009;36(9):4197-4212. http://dx.doi.org/10.1118/1.3190392