User Manual
Table of Contents
Table of Contents
List of Figures
ix
List of Tables
xi
Preface
xiii
v
Preview copyGenerated 7/9/2010
User Manual
Table of Contents
25
41
vi
DRAFT
User Manual
Table of Contents
47
59
63
67
vii
Preview copyGenerated 7/9/2010
User Manual
Table of Contents
Appendix ASpecifications
73
81
List of Addenda
83
Index
85
viii
DRAFT
User Manual
List of Figures
List of Figures
Figure 1-1: Selenia Dimensions .................................................................................................................... 1
Figure 1-2: Label Locations........................................................................................................................... 8
Figure 2-1: System Power Controls ............................................................................................................... 9
Figure 2-2: Acquisition Workstation Controls and Displays ........................................................................ 10
Figure 2-3: Tubestand Controls and Indicators............................................................................................ 12
Figure 2-4: C-arm Controls ......................................................................................................................... 13
Figure 2-5: Compression Device................................................................................................................. 13
Figure 2-6: Compression Display................................................................................................................ 13
Figure 2-7: Tubehead Display..................................................................................................................... 14
Figure 2-8: Dual Function Footswitches...................................................................................................... 14
Figure 2-9: The Startup Screen.................................................................................................................... 15
Figure 2-10: How to Log In......................................................................................................................... 16
Figure 2-11: C-arm Controls (left side shown) ............................................................................................. 17
Figure 2-12: Power Buttons ........................................................................................................................ 23
Figure 3-1: An Example Select Function to Perform Screen......................................................................... 25
Figure 3-2: An Example Gain Calibration Screen ........................................................................................ 26
Figure 3-3: How to Select a Patient............................................................................................................. 27
Figure 3-4: How to Add a New Patient ....................................................................................................... 28
Figure 3-5: The Filter Tab in the Patient Filter Screen.................................................................................. 29
Figure 3-6: An Example Generator Tab in the Procedure Screen................................................................. 31
Figure 3-7: The Add View Screen ............................................................................................................... 33
Figure 3-8: The Add Procedure Dialog Box ................................................................................................ 34
Figure 3-9: The Edit View Screen................................................................................................................ 34
Figure 3-10: The Print Screen ..................................................................................................................... 37
Figure 4-1: The Preview Screen .................................................................................................................. 41
Figure 4-2: The Tools Tab in the Procedure Screen..................................................................................... 42
Figure 4-3: Marked Images in a Procedure.................................................................................................. 42
Figure 4-4: Image Review Tools.................................................................................................................. 43
Figure 4-5: Image Review Tabs................................................................................................................... 44
Figure 4-6: Icons Available on the Notices Tab........................................................................................... 44
Figure 4-7: The Exposure Index .................................................................................................................. 44
Figure 5-1: C-arm Accessories .................................................................................................................... 47
Figure 5-2: How to Install the Conventional Face Shield............................................................................. 48
Figure 5-3: How to Align the Retractable Face Shield on the C-arm............................................................ 49
Figure 5-4: Installation................................................................................................................................ 50
Figure 5-5: Operation ................................................................................................................................. 50
Figure 5-6: How to Install a Compression Paddle ....................................................................................... 53
Figure 5-7: How to Remove the Compression Paddle ................................................................................. 53
Figure 5-8: The FAST Compression Mode Slide .......................................................................................... 54
Figure 5-9: Installation of the Magnification Stand ...................................................................................... 55
Figure 5-10: How to Attach the Localization Crosshair Device ................................................................... 56
Figure 5-11: How to Install and Remove the Magnification Crosshair Device ............................................. 57
Figure 6-1: Screening Example, Conventional Procedure............................................................................ 60
Figure 8-1: The Admin Screen .................................................................................................................... 68
Figure A-1: Tubestand Dimensions ............................................................................................................. 73
Figure A-2: Acquisition Workstation Dimensions........................................................................................ 74
ix
Preview copyGenerated 7/9/2010
User Manual
List of Figures
x
DRAFT
User Manual
List of Tables
List of Tables
Table 2-1: C-arm Functional Tests ............................................................................................................... 17
Table 3-1: Taskbar Menus ........................................................................................................................... 39
Table 8-1: Admin Screen Functions ............................................................................................................. 69
Table 8-2: Radiologic Technologist ManagerService Tools Functions....................................................... 71
Table A-1: mA Setting as a Function of kV................................................................................................... 78
Table B-1: System Messages ........................................................................................................................ 82
xi
Preview copyGenerated 7/9/2010
User Manual
List of Tables
xii
DRAFT
User Manual
Preface
Intended Use
Preface
1.0
Intended Use
The Hologic Selenia Dimensions Digital Breast Tomosynthesis System generates digital
mammographic images that can be used for screening and diagnosis of breast cancer. The
Selenia Dimensions system is intended for use in the same clinical applications as Full Field
Digital Mammography systems for screening mammograms. Specifically, the Selenia
Dimensions system can be used to acquire two-dimensional full field digital mammograms
and three-dimensional tomosynthesis mammograms. The screening examination will consist
of a two-dimensional image set or a two-dimensional and tomosynthesis image set. The
Selenia Dimensions system may also be used for additional diagnostic workup of the breast.
2.0
System Capabilities
The system provides the user interfaces for the performance of screening and diagnostic
mammograms:
3.0
4.0
The service personnel to install the system, set the site system configurations and
calibrations, and find faults
Users
xiii
Preview copyGenerated 7/9/2010
User Manual
Preface
Training Requirements
5.0
Training Requirements
Hologic does not accept the responsibility for injury or damage from wrong system
operation.
Make sure that you receive training on the Selenia Dimensions before use on patients.
Hologic training programs address MQSA training regulations for any Technologist or
Physician.
Refer to this manual for directions on how to use Selenia Dimensions.
6.0
7.0
8.0
xiv
DRAFT
WARNING!
Warning:
Caution:
Note
User Manual
Preface
Terms and Definitions
9.0
AEC
Annotations
Collimator
Combo Procedure
Conventional Mammography
Diagnostic Workstation
DICOM
Gantry
Grid
HIS
HTC
Image Receptor
MQSA
Notice
PACS
Pend
Projection Images
RIS
ROI
Region of Interest
SID
Tomosynthesis
UPS
xv
Preview copyGenerated 7/9/2010
User Manual
Preface
International Symbols
Protective Earth
terminal
Off
On
Off
On
WEEE
Dangerous Voltage
Manufacturer
Date of Manufacture
xvi
DRAFT
User Manual
Chapter 1General Information
System Description
System Description
Legend for Figure 1-1
1.
Gantry
2.
C-arm
3.
Acquisition Workstation
2.0
Safety Information
Read and understand this manual before you use the system. Keep the manual available
during the patient procedures.
Always follow all the instructions in this manual. Hologic does not accept the responsibility
for injury or damage from wrong system operation. Hologic can arrange for training at your
facility.
The Selenia Dimensions has protective devices, but the Technologist must understand how
to safely use the system. The Technologist must remember the health hazards of x-rays.
2.1
General Safety
The Selenia Dimensions system is classified as CLASS I, TYPE B APPLIED PART, IPX0,
permanently connected equipment, continuous operation with short term loading per IEC
60601-1. There are no special provisions to protect the system from flammable anesthetics
or ingress of liquids.
WARNING!
1
Preview copyGenerated 7/9/2010
User Manual
Chapter 1General Information
Safety Information
WARNING!
WARNING!
Warning:
Warning:
Warning:
Warning:
Caution:
Caution:
Only use the approved accessories with this equipment. The failure to
follow this caution can cause errors and possible data loss.
Note
Hologic does not provide the Gantry power cable for some countries. If
the power cable is not provided, the installed cable must meet the
following requirements and all local codes that apply: 3 conductor, 8
AWG (10 mm2) copper not more than 25 feet (7.62 meters) in length.
2
DRAFT
User Manual
Chapter 1General Information
Safety Information
2.2
Patient Safety
WARNING!
WARNING!
WARNING!
Warning:
Warning:
Keep the hands of the patient away from all buttons and
switches at all times.
Warning:
3
Preview copyGenerated 7/9/2010
User Manual
Chapter 1General Information
Safety Information
2.3
4
DRAFT
Warning:
You increase the patient dose to high levels if you increase the
AEC exposure adjustment setting. You increase the image noise
or decrease image quality if you decrease the AEC exposure
adjustment setting.
Warning:
Put both footswitches away from the patient and C-arm area to
prevent any accidental footswitch use. When the patient has a
wheelchair, put the footswitches away from the area.
Warning:
Radiation Safety
WARNING!
Warning:
Warning:
WARNING!
Warning:
Warning:
You must keep your complete body behind the radiation shield
for the time of the exposure for maximum protection from x-ray
exposure.
User Manual
Chapter 1General Information
Safety Information
2.4
2.5
2.6
Data Loss
Warning:
Do not move the C-arm while the system retrieves the image.
Caution:
Caution:
Do not put any magnetic media near or on devices that create any
magnetic fields, because stored data can be lost.
Equipment Damage
Caution:
Caution:
Caution:
Caution:
Use the least possible amount of cleaning fluids. The fluids must not
flow or run.
Caution:
5
Preview copyGenerated 7/9/2010
User Manual
Chapter 1General Information
Safety Information
2.7
Interlocks
The Selenia Dimensions has safety interlocks:
6
DRAFT
The C-arm vertical drive and rotation is disabled when 45 Newtons (10 pounds) or
greater of compression force is displayed.
If the x-ray button is released before the end of the exposure, the exposure stops and an
alarm message appears.
When in Tomo mode, the system does not allow the Grid in the x-ray field
(Tomosynthesis option).
Mirror and Filter interlocks prevent the x-ray exposure when the Light Field Mirror or the
Filter is not aligned.
User Manual
Chapter 1General Information
Compliance
3.0
Compliance
This section describes the mammography system compliance requirements and the
responsibilities of the manufacturer.
3.1
Compliance Requirements
The manufacturer has the responsibility for the safety, reliability, and performance of this
equipment with the following provisions:
3.2
The network and communication equipment must be installed to meet IEC Standards.
The complete system (network and communications equipment and Selenia Dimensions
Mammography System) must be in compliance with IEC 60601-1 and IEC 60601-1-1.
Compliance Statements
The manufacturer states this device is made to meet the following requirements:
7
Preview copyGenerated 7/9/2010
User Manual
Chapter 1General Information
Label Locations
4.0
Label Locations
8
DRAFT
User Manual
Chapter 2System Controls and Indicators
System Power Controls
2.
Emergency Off Switch (two on the Gantry, one on the Acquisition Workstation)
3.
4.
5.
9
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Acquisition Workstation Controls and Display
2.0
Trackball
2.
Scroll Wheel
3.
Compression Release
4.
5.
Fingerprint Reader
6.
7.
Touchscreen Display
8.
9.
CD/DVD Drive
10
DRAFT
User Manual
Chapter 2System Controls and Indicators
Acquisition Workstation Controls and Display
2.1
Keyboard
Use the keyboard in the front drawer of the Acquisition Workstation for data entry.
2.2
2.3
2.4
Preview Display
See the images on the Preview Display.
11
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Tubestand Controls and Indicators
3.0
12
DRAFT
User Manual
Chapter 2System Controls and Indicators
Tubestand Controls and Indicators
3.1
C-arm Controls
The C-arm Controls provide the
Collimator and C-arm functions.
See Section 5.0, page 17.
3.2
2.
3.
AEC Buttons
4.
5.
6.
Paddle Clamp
POSITAEC
ION
Compression Thickness
13
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Tubestand Controls and Indicators
3.3
Tubehead Display
The Tubehead Display shows:
SID
Filter Type
Collimator Setting
Paddle Position
cm Rh
SID: 70 9 C
x2
Coll: 24
3.4
Put both footswitches away from the patient and C-arm area to
prevent any accidental footswitch use. When the patient has a
wheelchair, put the footswitches away from the area.
C-AR
COMPR
ES
SION
C-arm Down
2.
C-arm Up
3.
Compression Down
4.
Compression Up
14
DRAFT
User Manual
Chapter 2System Controls and Indicators
How to Turn On the Selenia Dimensions
4.0
Preparation
1. Reset all three Emergency Off switches.
2. Make sure that both system circuit breakers are in the On position.
3. Remove any obstructions to the C-arm movement and to the view of the Operator.
4.2
Startup
1. Press the UPS button at the rear of the Acquisition Workstation if the UPS was shut
down.
2. Press the computer power button at the rear of the Acquisition Workstation (see
Figure 2-1, page 9, number 4).
Note
The Startup screen includes a Shutdown button that turns off the whole
system, and a Reboot button that restarts the system.
15
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
How to Turn On the Selenia Dimensions
4.3
Log In
16
DRAFT
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
5.0
Compression Release
2.
(Future use)
3.
4.
(Future use)
5.
Collimator Override
6.
7.
8.
Counterclockwise C-arm
Rotation
9.
Compression Up
The following Functional Tests make sure that the control operates correctly.
Functional Test
Press a Compression Down button:
17
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
Functional Test
Press a Compression Up button:
Compression Release
C-arm Up
18
DRAFT
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
Functional Test
Press the C-arm Down button:
19
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
20
DRAFT
Functional Test
Press the Counterclockwise C-arm Rotation to start
counterclockwise C-arm rotation.
Right Panel
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
Functional Test
Push the C-arm Rotation switch away from you to move the
C-arm toward you.
Pull the C-arm Rotation switch toward you to move the
C-arm away.
The C-arm movement stops when the switch is released.
21
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
Perform the Functional Tests
Functional Test
Collimator Override
Press the light field lamp button to see the x-ray field for
approximately 30 seconds.
The light field lamp automatically illuminates with the start
of the Compression Down movement.
Motor Enable
C-arm Zero
Shifting Paddle
System
AEC
POSITION
22
DRAFT
User Manual
Chapter 2System Controls and Indicators
How to Turn Off the System
6.0
Functional Test
There are three Emergency Off switches, one on each side of
the Gantry and one on the Acquisition Workstation.
Press any of the Emergency Off switches to turn Off the
Gantry.
Turn the Emergency Off switch by one-quarter turn to reset
the switch.
7.0
23
Preview copyGenerated 7/9/2010
User Manual
Chapter 2System Controls and Indicators
How to Remove All Power from the Acquisition Workstation
24
DRAFT
User Manual
Chapter 3The User Interface
Select the Function to Perform
The Select Patient screen appears if you are not scheduled to perform
any Quality Control tasks.
If you select the Skip button, the Select Patient screen appears.
25
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
How to Perform the Quality Control Tasks
2.0
26
DRAFT
Note
When the Start button is not enabled for a Quality Control task, select
the Mark Completed button.
Note
You can perform required Quality Control tasks at another time. Select
the Admin button (on the Select Patient screen). Select the Quality
Control button on the Admin screen to display the list.
Note
User Manual
Chapter 3The User Interface
How to Select a Patient
3.0
The Current User tab displays the procedures for the current Operator.
3.1
Delete a patient from the worklist (Delete)see Section 3.4, page 28.
Search for a patient in the Modality Worklist (Query)see Section 3.7, page 30.
27
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
How to Select a Patient
3.2
3.3
3.4
28
DRAFT
Note
Note
User Manual
Chapter 3The User Interface
How to Select a Patient
3.5
The Filter tab accesses the parameters that select the patients who appear on the selected
tab page.
The Columns tab adds or removes columns on the selected tab page.
Select a line in the Results list to activate the Open button. When you select the Open
button, the Procedure screen for that patient appears.
Note
3.6
Your logon permissions can allow you to add, change or delete the tabs
on the Select a Patient screen from the Patient Filter screen.
The Save button changes the name of the selected tab.
The Save As button creates a new tab.
The Delete button deletes the selected tab.
29
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
How to Select a Patient
3.7
3.8
Bar Code ReaderThe field in which the bar code reader scans is configurable. Scan
the configured field bar code. The scheduled procedure displays and the patient is
added to the local database. By default, the user can scan on the Patient ID, Accession
Number, or Requested Procedure ID.
KeyboardUse one or more fields to query the Modality Worklist Provider. All fields to
query are configurable. The default fields are as follows: the Patient name, the Patient ID,
Accession Number, Requested Procedure ID, Scheduled Procedure Date. The scheduled
procedure displays and the patient is added to the local database.
3.9
30
DRAFT
User Manual
Chapter 3The User Interface
The Procedure Screen
4.0
4.1
4.1.2
Standard
Enhanced
Manual
AEC: Auto-Time
AEC: Auto-kV
AEC: Auto-Filter
The user selects the kV, mAs, Focal Spot, and Filter.
The user selects the kV, Focal Spot, and Filter. The system
selects the mAs.
The user selects the Focal Spot. The system selects the kV,
mAs, and Filter (Rhodium).
The user selects the Focal Spot. The system selects the kV,
mAs, and Filter.
Use the Automatic Exposure Control modes (AEC) to let the system control the
exposure techniques.
31
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
The Procedure Screen
4.1.3
4.2
4.3
You can Accept the image. The locked image transmits to output devices with all
attributes and marks.
You can Reject the image. The Preview closes. You can repeat the rejected view, or
select another view.
You can Pend the image. The image saves for future review.
32
DRAFT
User Manual
Chapter 3The User Interface
The Procedure Screen
4.4
View Modifiers
ID = Implant Displaced
RL = Rolled Lateral
RM = Rolled Medial
RI = Rolled Inferior
RS = Rolled Superior
NP = Nipple in Profile
AC = Anterior Compression
IMF = Infra-Mammary Fold
AX = Axillary Tissue
Note
2. Select the tab, then select the view. You can select a maximum of 3 View Modifiers from
the right panel of the screen.
3. Select the Add button. A thumbnail image for each view that you select appears in the
bottom of the window.
4. To remove a selected view, select the view then select the TRASH icon.
5. To remove all selected views, select the Clear button.
33
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
The Procedure Screen
4.5
4.6
User Manual
Chapter 3The User Interface
How to Access Image Review Features
4.7
Close Procedure
Discontinued
Return To procedure
If there are images marked as Pend, you must respond to the confirmation prompt to close
the procedure In-Progress.
If MPPS is activated, messages are sent to the output devices when you select Complete and
Discontinue. You can also click and hold the tab above the thumbnail images to resend a
message about the procedure status during the procedure. A Procedure Action dialog box
appears with buttons to resend a status or to return to the procedure.
5.0
6.0
6.1
6.2
35
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
How to Use the On-Demand Outputs
7.0
7.1
How to Archive
1. Select the Archive button.
2. Select a storage device from a drop-down menu.
3. Select the Start button to copy all selected images from the opened case study to the
selected device.
Note
36
DRAFT
Use the Manage Queue utility in the task bar to review the image
status.
User Manual
Chapter 3The User Interface
How to Use the On-Demand Outputs
7.2
How to Print
7.2.1
37
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
How to Use the Paddle Shift Feature
7.2.2
7.3
How to Export
1. Select the Export button.
2. Select a device from a drop-down menu.
3. Select the Start button to copy all accepted images from the open procedure to the
selected device.
8.0
2. From the paddle shift section of the screen, select the button for the new paddle
position. The paddle moves to the new position.
38
DRAFT
User Manual
Chapter 3The User Interface
About the Taskbar
9.0
Menu
Information Icon
Select the Information icon to display a menu.
This section of the taskbar flashes a yellow color if an alarm exists.
Select Acknowledge All to remove the flashing indication. Select the
Manage Alarms option to display and close any open alarms.
Current User Name
Select the user name to display a menu.
Log Out returns you to the Sign-in screen.
My Settings displays the Edit Operator screen to review or edit
Operator information, change your password, or use the
Fingerprint Capture feature.
Print prints the displayed patient list to a connected printer.
Output Device Icons
Select an output icon (query retrieve, workstation/archive, printer,
or CD-ROM) to display a menu. Each menu selection takes you to
the screen for that selection.
Manage Queues displays the status of jobs in the queue and job
details for the selected output, and allows you to filter the queue
display.
Notice Icon
Select the Notice icon to display a menu.
Select Acknowledge All to remove the flashing indication.
The Manage Notices icon allows you to select the display options
for the message like date or title.
System Status Icons
Select the Tubehead icon to display a menu. When a green
checkmark appears next to the tubehead icon, the detector and
generator are ready for use.
Clear All Faults deletes all error messages.
Zero Tubehead puts the tubehead at zero degrees of rotation for the
next exposure.
System Diagnostics accesses Subsystem settings.
System Defaults opens the Gantry Defaults screen to set the
Compression and Generator default values.
About displays information about the Acquisition Workstation.
39
Preview copyGenerated 7/9/2010
User Manual
Chapter 3The User Interface
About the Taskbar
40
DRAFT
User Manual
Chapter 4The Images
Introduction
Introduction
After you make an exposure, the acquired image displays on the Preview screen. Review the
image and add the annotations, then Accept, Reject, or Pend the image. A thumbnail image
appears in the Case Study area of the screen.
If you select the Pend button, a question mark ? appears on the thumbnail image.
1.1
1.2
41
Preview copyGenerated 7/9/2010
User Manual
Chapter 4The Images
How to Review the Images
2.0
42
DRAFT
User Manual
Chapter 4The Images
How to Review the Images
2.1
1.
2.
The Window Level tool with the Trackball changes the brightness and contrast.
3.
The Window Level Fine Adjustment tool enters the specified contrast and brightness values.
4.
5.
The SNR/CNR button calculates the signal-to-noise ratio and contrast-to-noise ratio in the ACR
Phantom image.
6.
The AEC button displays the AEC Sensor areas used for the exposure calculation. The sensor areas
display on the Preview Screen.
7.
8.
9.
The Fit-to-Viewport button fits the image within the image tile.
10. The True Size button displays the image in the actual size of the breast.
11. The View actual pixels button displays the image in full resolution.
12. The Multi-Up Display button selects the number of tiles to display.
13. The Image Tile Advance button sets the active Multi-Up tile.
14. The Mirror button reverses (mirrors) the image.
15. The Tag for Print button tags the projection or reconstruction images of a tomosynthesis image to
print later (Tomosynthesis option).
43
Preview copyGenerated 7/9/2010
User Manual
Chapter 4The Images
How to Review the Images
2.2
Notices: Make annotations about the image. Use the Draw or Elliptical button
(Numbers 1 and 2) in Figure 4-6 to mark the area of interest. Use the Arrow
button (Number 3) to send the annotation as a Notice.
2.2.2
44
DRAFT
User Manual
Chapter 4The Images
Send the Images to the Output Devices
2.2.3
2.3
2.3.2
Note
3.0
45
Preview copyGenerated 7/9/2010
User Manual
Chapter 4The Images
Send the Images to the Output Devices
46
DRAFT
User Manual
Chapter 5How to Use the Accessories
Introduction
Introduction
The Selenia Dimensions can perform screening or diagnostic applications with specified
accessories. This chapter describes how to use all possible system accessories. Your
accessories depend on your system configuration.
2.0
47
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
The Patient Face Shields
3.0
3.1
Warning:
Warning:
48
DRAFT
User Manual
Chapter 5How to Use the Accessories
The Patient Face Shields
3.2
Figure 5-3: How to Align the Retractable Face Shield on the C-arm
To remove the Retractable Face Shield:
1. Press and hold the Unlock Lever (number 1, Figure 5-4) in the Up position.
2. Lift the Face Shield from the slots and remove from the C-arm.
49
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
The Patient Face Shields
3.3
50
DRAFT
User Manual
Chapter 5How to Use the Accessories
Compression Paddles
4.0
Compression Paddles
The system can identify each paddle and automatically adjust the collimator.
4.1
18 x 24 cm Frameless
Screening Paddle
4.2
24 x 29 cm Frameless
Screening Paddle
Spot Contact
Frameless Paddle
10 cm Contact
Frameless Paddle
15 cm Contact
Frameless Paddle
51
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
Compression Paddles
4.3
Localization Paddles
10 cm Rectangular Opening
Localization Paddle
10 cm Perforated
Localization Paddle
4.4
52
DRAFT
15 cm Rectangular Opening
Localization Paddle
15 cm Perforated
Localization Paddle
Magnification Paddles
10 cm Magnification
Paddle
10 cm Magnification
Localization Perforated
Paddle
15 cm Magnification
Paddle
User Manual
Chapter 5How to Use the Accessories
Compression Paddles
4.5
4.6
4.7
Paddle Shift
The system allows most paddles to move to the left or right of the center position. The feature
helps small-breast examinations with lateral views. When a lateral view is selected, the
system automatically moves the collimator for the selected paddle position.
53
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
Compression Paddles
4.8
The compression is more uniform from the chest wall to the nipple.
When the FAST Compression mode is selected, the paddle automatically tilts when
the compression is applied. The paddle starts at the flat position until some
compression force is applied. The paddle then tilts until its maximum angle is
reached.
The FAST Compression mode does not require excessive compression, but you must
use enough compression to prevent the movement of the breast. You should use a
consistent amount of compression, especially for related left and right views.
The FAST Compression mode may not be best for breasts that are equal or
symmetrical in thickness from the chest wall to the anterior area of the breast.
4.8.2
54
DRAFT
User Manual
Chapter 5How to Use the Accessories
Magnification Stand
5.0
Magnification Stand
The Selenia Dimensions Magnification Stand has a breast platform and an abdominal shield.
When the Magnification Stand is installed, the HTC grid automatically retracts and the x-ray
exposure techniques are set to the Magnification default values. When the platform is
installed, only use the Magnification paddles (see Section 4.4, page 52).
5.1
There are two sets of hooks for the Magnification StandOne set is for
1.8x, and the other set is for 1.5x. See Figure 5-1, page 47, numbers 2
and 3.
5. Put the hooks of the Magnification Stand into the C-arm slots. Slide the Magnification
Stand down, until the thin black lines on the Magnification Stand and the black line of
the C-arm meet. See Figure 5-9, number 2.
6. Release the levers. The locking pins slide into holes and lock the device.
To remove the Magnification Stand:
1. Remove the Magnification paddle.
2. Hold the handles of the Magnification Stand and compress the release levers.
3. Lift and remove the device from the C-arm.
55
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
Crosshair Devices
6.0
Crosshair Devices
6.1
6.1.2
6.2
56
DRAFT
User Manual
Chapter 5How to Use the Accessories
Crosshair Devices
6.3
Figure 5-11: How to Install and Remove the Magnification Crosshair Device
6.3.1
6.3.2
6.4
57
Preview copyGenerated 7/9/2010
User Manual
Chapter 5How to Use the Accessories
Crosshair Devices
58
DRAFT
User Manual
Chapter 6Clinical Procedures
Standard Workflow
1.0
Warning:
Warning:
Keep the hands of the patient away from all buttons and
switches at all times.
Warning:
Put both footswitches away from the patient and C-arm area to
prevent any accidental footswitch use. When the patient has a
wheelchair, put the footswitches away from the area.
Standard Workflow
Preparation:
1. Select a patient from the worklist, or manually add a new patient.
2. Identify the required procedures.
3. Select the output device set if a different or additional device is needed.
4. Install the paddle.
5. Select the first view.
At the Gantry:
1. Set C-arm height and rotation angle.
2. Make sure the light field illuminates the correct area.
3. Position the patient and compress the breast.
At the Acquisition Workstation:
1. Set the exposure technique.
2. Acquire the image.
3. Release the patient.
4. Preview the image. Look at the Exposure Index to make sure that the exposure is within
acceptable range.
5. You can use the Window/Level tool or other Preview options during image preview.
6. Accept, Reject, or Pend the image.
7. Perform the Acquisition cycle as required for the requested procedures.
8. If necessary, add an additional view or procedure.
9. Make sure that the patient is safely away from the system after you complete the
examination.
10. Close the procedure.
59
Preview copyGenerated 7/9/2010
User Manual
Chapter 6Clinical Procedures
Example Screening Procedure
2.0
2.1
2.2
Use the light field lamp as necessary to see the x-ray field.
Apply the compression slowly. As necessary, stop and make the adjustments to
patient position.
60
DRAFT
User Manual
Chapter 6Clinical Procedures
Example Screening Procedure
2.3
WARNING!
3. Press and hold the x-ray buttons for the complete exposure. During the exposure, a
System Message, a tone, and the x-ray indicator on the control panel indicate an
exposure is in progress.
4. Release the compression device. If the automatic release feature is set, the compression
device automatically lifts after the exposure.
2.4
You can Accept the image. Wait for the image to appear as a thumbnail image on
the Procedure screen. The image transmits to the output device.
If there are image problems, you can Reject the image and you must enter the
reason. The system automatically adds another icon for the same view. Repeat the
exposure.
You can Pend the image. The image saves for future review.
The selection to send the image to the Output Devices at the close of the patient or when
accepted is service-configurable.
2.5
2.6
61
Preview copyGenerated 7/9/2010
User Manual
Chapter 6Clinical Procedures
Example Screening Procedure
62
DRAFT
User Manual
Chapter 7Maintenance and Cleaning
General Information
General Information
Call the Hologic Technical Support telephone number for the current list of recommended
cleaning solutions.
Before each examination, clean and use a disinfectant on any part of the system which
touches a patient. Give the attention to the paddles and the Image Receptor Device.
Caution:
Do not use any heat source (like a heating pad) on the image receptor.
Be careful with the plastic compression paddles. Inspect the paddles. Replace the paddle
when you see damage.
1.1
If more than soap and water is required, Hologic recommends any one of the following:
10% chlorine bleach and water with one part commercially available chlorine bleach
(normally 5.25% chlorine and 94.75% water) and nine parts water
After you apply any of the above solutions, use a pad and apply a diluted dishwashing liquid
to clean any parts which touch the patient.
Caution:
Warning:
63
Preview copyGenerated 7/9/2010
User Manual
Chapter 7Maintenance and Cleaning
General Information
1.2
Do not use a strong detergent, abrasive cleaner, high alcohol concentration, or methanol
at any concentration. If skin preparation contains a high alcohol concentration, allow
time to dry before compression.
Never allow any liquids to enter the internal parts of the equipment. Do not apply
cleaning sprays or liquids to the equipment. Always use a clean cloth and apply the
spray or liquid. If liquid enters the system, disconnect the electrical supply and schedule
inspection by service personnel before the system is returned to use.
Caution:
64
DRAFT
Always follow the instructions from the germicide manufacturer. The instructions
include the methods of concentration, storage, application, contact time, wash
requirements, protective clothing, shelf life, and for removal. The instructions make sure
best and safe product use.
User Manual
Chapter 7Maintenance and Cleaning
Acquisition Workstation
2.0
Acquisition Workstation
2.1
Always use a clean, soft, lint-free cloth to clean the display area. Microfiber cloths,
available at most camera stores, are recommended.
There are many commercially available products to clean LCD displays. Any of the products
free of the ingredients described above and used according to the directions of the
manufacturer can be used.
2.2
2.3
2.4
Apply the adhesive side of cellophane tape, then remove the tape.
Apply a product with ammonia base to a cloth, and clean the Fingerprint Reader
window.
65
Preview copyGenerated 7/9/2010
User Manual
Chapter 7Maintenance and Cleaning
Acquisition Workstation
66
DRAFT
User Manual
Chapter 8System Administration Interface
How to Use the Admin Screen
67
Preview copyGenerated 7/9/2010
User Manual
Chapter 8System Administration Interface
How to Use the Admin Screen
Refer to Table 8-1, page 69 for descriptions of the Admin screen functions.
68
DRAFT
User Manual
Chapter 8System Administration Interface
How to Use the Admin Screen
System
Connectivity
Screen Function
Manage Operators
My Settings
Procedure Editor
Procedure Order
Quality Control
QC Report
Test Patterns
Reject and Repeat
Report
System Tools
Gantry Defaults
Subsystem
Log Viewer
Preferences
About
Query Retrieve
Import
Manage Output
Groups
Incoming Log
Description
Add, delete or change Operator information.
Change the information for the current Operator.
Add or Edit the procedures, or change the view order for each user.
Change the procedure list order.
Select a Quality Control task to perform or mark completed.
Create a QC Report.
Select and send the test patterns to output devices.
Create a Reject and Repeat Report.
The Interface for Service for the configuration of and identification of
problems in the Acquisition Workstation.
Set the Gantry default values.
Displays the status of all subsystems.
Review the system log files.
Set the system preferences.
Describes the system.
Query the configured devices.
Import the data from a DICOM source.
Add, delete, or edit output groups.
Shows log entries for images that do not import during manual import
or DICOM store.
You must have permissions to access all features. The permission level controls the functions you can change.
69
Preview copyGenerated 7/9/2010
User Manual
Chapter 8System Administration Interface
How to Use the System Tools
2.0
2.1
70
DRAFT
User Manual
Chapter 8System Administration Interface
How to Use the System Tools
Screen Functions
About
FAQ
Glossary
Platform
Shortcuts
Connectivity
Film & Image
Information
Licensing
User Interface
Internationalization
QC
Troubleshooting AWS
Computer
Log
Backups
AWS
Description
The introduction to the service tool.
List of common questions.
List of terms and descriptions.
List of directories, software version numbers, and
system software statistics.
List of Windows shortcuts.
List of Installed Devices.
Create an Image Report. Create a QC Report.
List of Installed Licenses.
Change the options in the Software application.
Select the local language and culture.
Set Quality Control Settings.
Allows for download of images.
System Management and Network Information.
Change the event record options.
Control the backups for the system.
71
Preview copyGenerated 7/9/2010
User Manual
Chapter 8System Administration Interface
How to Use the System Tools
72
DRAFT
User Manual
Appendix ASpecifications
Product Measurements
Appendix ASpecifications
1.0
Product Measurements
1.1
Width
66 cm (26 inches)
Depth
Weight
73
Preview copyGenerated 7/9/2010
User Manual
Appendix ASpecifications
Product Measurements
1.2
Acquisition Workstation
207 cm
(81.5 in.)
92.7 cm
(36.5 in.)
58.5 cm
(23 in.)
74
DRAFT
Height
Width
Depth
Weight
User Manual
Appendix ASpecifications
Operation and Storage Environment
2.0
2.2
20 C to 30 C
Storage Environment
2.2.1
Temperature Range
10 C to 40 C
2.2.2
X-ray Detector
Temperature Range
10 C to 30 C indefinitely
10 C to 35C for a maximum of 12 hours
2.2.3
3.0
Gantry
Acquisition Workstation
Temperature Range
10 C to 40 C
Windows XPE
Computer Memory
4 GB RAM minimum
700 GB minimum
CDRW/CDR/DVD+/-RW/DAT
Display
2 or 3 Megapixels as required
Touchscreen Display
Network Interface
100/1000Base-T Ethernet
Remote Diagnostics
Internet
Application specific
75
Preview copyGenerated 7/9/2010
User Manual
Appendix ASpecifications
Electrical Input
4.0
Electrical Input
4.1
4.2
5.0
Tubestand
Mains Voltage
Mains Impedance
Mains Frequency
50/60 Hz 5%
<5A
65 A for 3 seconds
Acquisition Workstation
Mains Voltage
Mains Frequency
50/60 Hz 5%
Power Consumption
C-arm
Rotation Range
Conventional Mammography:
+195 +3/0.5 to 0 0.5 to 155 +0.5/3
Tomosynthesis option:
+180 0.5 to 0 0.5 to 140 0.5
accurate to 0.5
Rotation Acceleration
18/s2 +18/-9%
Rotation Deceleration
18/s2 +18/-9%
18/s 25%
Note
The angular velocity is the average of the velocity of the tube arm
rotating clockwise between 0 and 90 or rotating counterclockwise
between 90 and 0. The angular velocity does not include the time to
accelerate from zero velocity and decelerate to zero velocity.
76
DRAFT
User Manual
Appendix ASpecifications
Tubestand Technical Information
5.2
Compression
Manual Compression Force
Motorized Compression
Functions in three operating modes, Pre-compression, FullRange, Dual Compression, user selectable through
software
Pre-Compression Force
Compression Controls
Compression Release
Automatic Compression Release
Compression Paddles
5.3
X-ray Tube
Focal Spot
Tube Voltage
20 kV to 49 kV
Anode Material
Tungsten
X-ray Window
Beryllium 0.63 mm
77
Preview copyGenerated 7/9/2010
User Manual
Appendix ASpecifications
Tubestand Technical Information
5.4
Filtration
5.4.1
kV/mA Range
Table A-1: mA Setting as a Function of kV
kV
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
78
DRAFT
LG FS mA
SM FS mA
100
110
110
120
130
130
140
150
160
160
170
180
190
200
200
200
190
180
180
180
170
170
160
160
150
150
150
140
140
140
30
30
30
30
30
40
40
40
40
40
50
50
50
50
50
50
50
50
50
50
User Manual
Appendix ASpecifications
Imaging System Technical Information
5.5
X-ray Collimation
Collimation Fields
5.6
5.7
Within 2% of SID
X-ray Generator
Type
Rating
Electrical Power Capacity
kV Range
kV accuracy
mAs Range
mAs Accuracy
mA Range
6.0
7.0 cm x 8.5 cm
10 cm x 10 cm
15 cm x 15 cm
18 cm x 24 cm
18 cm x 29 cm (Tomosynthesis option)
24 cm x 29 cm
Image Receptor
Fluid Ingress
Deflection
Active Imaging Area
DQE (Conventional Mammography)
DQE (Tomosynthesis option)
Dynamic Range and Linearity
Uniformity
79
Preview copyGenerated 7/9/2010
User Manual
Appendix ASpecifications
Imaging System Technical Information
80
DRAFT
User Manual
Appendix BThe System Messages and Alert Messages
Error Recovery and Troubleshooting
2.0
Fault Levels
There are five fault levels: Warning, Minor, Major, Critical, and Alerts.
2.1.1
Warning Faults
The Warning faults are not displayed to the user. These faults are recorded in the log files.
Warning fault designs:
2.1.2
Minor Faults
Minor fault designs:
2.1.3
Major Faults
Major fault designs:
2.1.4
Critical Faults
Critical fault designs:
2.1.5
Alert Messages
The alert messages are routine messages that can prevent an exposure. An Alert
message remains active until the required action is complete or the condition does
not exist.
81
Preview copyGenerated 7/9/2010
User Manual
Appendix BThe System Messages and Alert Messages
Types of Messages and Alert messages
2.2
System Messages
When the following system messages appear, perform the action indicated to clear the
message and allow the next exposure.
Table B-1: System Messages
Icon
82
DRAFT
Message
User Action
Paddle is moving
User Manual
List of Addenda
List of Addenda
Write the title of all Addenda which are supplied for this manual, with their part number and
revision.
Add all the Addenda to the manual after this record page.
Title
Part Number
Revision
83
Preview copyGenerated 7/9/2010
User Manual
List of Addenda
84
DRAFT
User Manual
Index
A
Index
A
accept images 61
accept rejected images 61
accessories
compression paddles 51
crosshair devices 56
face shield 48
install on C-arm 47
Magnification Stand 55
acquire images 32, 61
acquisition modes 31
Acquisition Workstation
controls and displays 10
maintenance 65
Touchscreen 1011
add
new patient 28
procedure 34
view 33
admin screen 67
Admin button 30
functions 69
AEC buttons 13
AEC Sensor position 13
alerts 81
archive 36
B
bar code reader 1011
C
calibrations, perform 26
C-arm
controls and indicators 1213
displays 13
functional tests 17
rotation 20
slots for accessories 47
cine tab 44
circuit breaker
Acquisition Workstation 9
Gantry 9
cleaning solutions and methods 64
clinical procedures 59
close
patient 35
procedure 35
procedure with MPPS 35
collimator
override 22
programmed to paddle 51
programmed to paddle position 22, 53
comments tab 44
compliance
requirements 7
statements 7
compression
controls and displays 13
paddles 51
release 10, 61
compression force, range 13
compression paddles, routine screening 51
computer power button 9
contact paddles 51
controls
Acquisition Workstation 10
C-arm 14
collimator override 22
compression 14, 17, 22
compression brake 17
handwheels 13
indicators 9
light field lamp 22
manual compression release 18
system 9
conventional face shield, install 48
cybersecurity statement xiv
D
data loss 5
definitions xv
delete patient 28
display
Acquisition Workstation 10
clean 65
compression force 13
preview image 11
thickness 13
types of images selection 45
window level 43
85
Preview copyGenerated 7/9/2010
User Manual
Index
E
E
edit
patient information 28
user information 39
view 34
Emergency Off Switch 910, 12
functional test 23
export 38
exposure
parameters 31
sequence 61
techniques, set 31
exposure index 44
F
face shield 48
faults 81
filter, information options 29
fingerprint reader 10
clean 65
footswitches 14
function, select to perform 26
functional tests 17
C-arm controls locations 17
Emergency Off Switch 23
G
gantry controls and indicators 12
generator tab, set techniques 31
glossary xv
H
Hologic technical support 63
I
images
accept 61
acquisition mode 31
output options 45
preview 61
reject 61
review 42
review tools 35
review tools tab 43
store 61
unacceptable 61
Implant Displaced views 33
Implant Present button 32
86
DRAFT
indicators 9
install
compression paddles 53
conventional face shield 48
tomo face shield 49
intended use xiii
interlocks 6
international symbols xvi
K
keyboard 1011
clean 65
L
label locations 8
laser film printer, isolation requirements 3
light field lamp 17, 22
use 60
localization crosshair device
align 57
install and remove 56
use 56
localization paddles 52
Log In 16
Log Out 30
M
magnification crosshair device, install and remove 57
magnification paddles 52
Magnification Stand 55
install and remove 55
maintenance
Acquisition Workstation 65
general 63
manage
alarms 39
notices 39
output groups 35
messages and alerts 81
MPPS, messages 35
My Settings icon 39
N
Notices tab 44
O
On/Off button 9
on-demand outputs 36
User Manual
Index
P
P
paddles
compression 5152
install 53
remove 53
shift 53
shift to new position 38
patient
add 28
delete 28
edit information 28
face shield 48
filter 29
open 27
position 60
safety 3
pended image, accept or reject 61
perform functional tests 17
permissions, by user group 70
power button and controls 9
prerequisites for system use xiii
preview image screen 11, 41, 61
procedure screen 31
procedures
add 34
close 35
Q
QC tasks, perform 26
quality control requirements xiv
query worklist 30
R
radiation safety 4
reject images 61
remove
compression paddles 53
conventional face shield 48
localization crosshair device 56
magnification crosshair device 57
Magnification Stand 55
Tomo face shield 49
view 33
requirements
quality control xiv
training xiv
review
images 42
rejected image 61
ROI tab 44
S
safety
data loss 5
equipment damage 5
general information 1
patient 3
radiation 4
screening
acquire the image 61
position patient 60
screen example 60
screens
add new patient 28
add view 33
admin 67
filter patient information 29
preview image 41
query 30
Select Function to Perform 25
select patient 27
scroll wheel 10
select
exposure parameters 31
output sets 35
patient 27
send images to outputs 45
shift paddle 38, 53
Shifting Paddle system 22
specifications 73
spot compression paddles 51
statement, cybersecurity xiv
system
administration 67
capabilities xiii
description 1
messages 82
power controls 9
ready 61
status icons 39
87
Preview copyGenerated 7/9/2010
User Manual
Index
T
T
taskbar 39
terms and definitions xv
tomo face shield
install 49
use 50
tools, image review 43
Touchscreen 11
clean 65
training requirements xiv
tubehead, display 14
tubestand controls and indicators 12
turn off system 23
turn on system 15
Log In 16
preparation 15
startup 15
U
uninterrupted power supply (UPS), power button 9
user interface 25
users menu 39
V
view
add 33
edit 34
W
warnings, cautions and notes xiv5
defined xiv
window level 43
workflow, standard 59
worklist, query 30
X
x-ray
acquire image 32, 61
collimated fields 22
indicators 32, 61
x-ray switch 10
88