Sunteți pe pagina 1din 2

CAPNOSTAT 5 MAINSTREAM CO2 SENSOR

21st Century CO2 Technology for mainstream CO2 monitoring


in critically ill, intubated patients.
The CAPNOSTAT 5 Sensor is the ideal capnography
solution for your patient monitoring system. It
provides technologically advanced measurement of
End-Tidal Carbon Dioxide (ETCO2), respiration rate,
and a clear, accurate capnogram at all respiratory
rates up to 150 breaths per minute.
The CAPNOSTAT 5 Sensor is:
Compact
Flexible
Affordable
Reliable
The CAPNOSTAT 5 can be combined with other
components and software to provide respiratory
monitoring capabilities that are comprehensive, yet
remarkably flexible.

800.243.3444

OPTION

ITS ALL INSIDE!


As the recognized global leader in capnography,
Respironics has been providing innovative and costeffective products to OEM customers for over
18 years. Now, our engineers have developed a new
miniature PC board that puts all measurement and
analysis functions right inside the sensor head. This
keeps costs and space requirements down by eliminating the need for a PC board inside your monitor.
The data output is a simple RS232 signal, so its a
snap to integrate the new CAPNOSTAT 5 Sensor
into your patient monitoring system. We also use
the very latest Digital Signal Processor (DSP) for
reliable and flexible analog-to-digital conversion.

WWW.OEM.RESPIRONICS.COM

203.697.6488

CAPNOSTAT 5 MAINSTREAM CO2 SENSOR SPECIFICATIONS


Transducer Type
Principle of Operation
Initialization Time
CO2 Measurement Range
Rise Time
CO2 Resolution
CO2 Accuracy

GO MAINSTREAM
The CAPNOSTAT 5 Sensor is designed for
mainstream measurements using sophisticated infrared absorption spectroscopy
to measure ETCO2. With mainstream,
measurements are taken directly at the
patients airway, so response is faster and
there is less chance of erroneous, artifact
data. The analyzer portion of the sensor
cannot be contaminated with patient
secretions, and there are no pumping
or pneumatic components to replace.
Maintenance requirements and overall
cost-of-ownership are minimized. Our
exclusive family of airway adapters are
easy to use and durable enough even for
long-term ventilator patients. Yet, they are
inexpensive and disposable.

CO2 Stability
CO2 Noise
Respiration Rate Range
Respiration Rate Accuracy
Compensations
(Supplied)
Calibration
Airway Adapters
Voltage Requirements
Power Rating
Interconnection

Note: LoFlo and


CAPNOSTAT are
interchangeable with
the host monitor. Either
device may be connected
and used via the RS232
connector.

The CAPNOSTAT 5 Sensor can be integrated with other components, software


packages and accessories, making it
easy for you to configure your patient
monitoring systems with state-of-the-art
respiratory management and analysis
technology now and into the future.

Temperature and Humidity


Water Resistance
Shock Impact

Data Interface
Data Output
Regulatory

COMPLETE OEM SOLUTIONS


The CAPNOSTAT 5 is just one of many
solutions we offer to our OEM customers.
Respironics comprehensive engineering
team customizes innovative products to
meet customer requirements and provides
clinical and marketing support to help
meet the growing needs of your business.

Mainstream CO2 Sensor


Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
Capnogram, displayed in less than 15 seconds at an ambient temperature
of 25C, full specifications within 2 minutes
0 to 150 mm Hg, 0 to 19.7%, 0 to 20 kPa (at 760 mm Hg)
Barometric Pressure supplied by host
Less than 60 ms - Adult Reusable or Single-Patient-Use Airway Adapter
Less than 60 ms - Infant Reusable or Single-Patient-Use Airway Adapter
0.1 mm Hg
0 to 69 mm Hg
0.25 mm Hg
70 to 150 mm Hg
0 40 mm Hg
2 mm Hg
41 70 mm Hg
5% of reading
8% of reading
71 100 mm Hg
10% of reading
101 150 mm Hg
Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will be maintained over a 120-hour period
RMS noise of the sensor shall be less than or equal to 0.25 mm Hg at 7.5% CO2
0 to 150 Breaths Per Minute (BPM)
1 breath
Barometric Pressure 400 mm Hg to 850 mm Hg
Operator selectable O2, N2O, HE and Agent compensation.
No routine user calibration required. An airway adapter zero is required when changing
to a different style of airway adapter.
Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant
Adapter taper meets ISO 5356-1
+5.0 VDC, 1.125 W Typical, 1.5 W Maximum.
Rated input: 1.1 Watts typical. Steady State
Up to 1.5 Watts maximum on power up (warm up)
Standard
Lemo Redel 8-pin plastic, Material PSU
Sensor Plug: PAB.M0.8GL.AC39GZ
Bend Relief: GMA.1B.030.DJ
Host Receptacle: PKB.M0.8GL.LJ
Pin out:
1
VA
5.0V
2
Shield
Shield
3
DGND
Digital return
4
VSRC
5.0V
5
TxD
Serial data from CAPNOSTAT
6
RxD
Serial data from Host
7
AGND
Analog return
8
SYNC
Waveform Synchronization
Operating:
0 to 45C, 10 to 90% RH, non-condensing
Storage:
-40 to 70C, <90% RH, non-condensing
IPX4 Splash-proof (sensor head only)
EN60068-2-6 Sinusoidal Vibration
EN60068-2-27 Shock
EN60068-2-64 Random Vibration
Able to withstand repeated 6-foot drops onto tiled floor while operating
RS232 compatible, bi-directional, 19200 Baud, Standard N-8-1.
CO2 gas concentration (mm Hg), End-tidal CO2, Inspired CO2, Respiratory Rate
Gas and barometric pressure compensated when supplied by host
Designed to meet IEC 60601-1-2, EN55011 CISPIR 11 Class B (Radiated and
Conductive Emissions), IEC 61000-4-2 Electrostatic Discharge Immunity, IEC 61000-4-3
Radiated Immunity, Designed to comply with 93/42/EEC (MDD CE Marking), FDA
Standards, ASTM F1456-01 Minimum Performance and Safety Requirements for
Capnometers and ISO21647. Medical Electrical Equipment performance requirements
for the basic safety and essential performance of respiratory gas monitors.

Specifications are subject to change without notice.

Customer Service: 1-800-243-3444, option 3 or 203-697-6488


www.oem.respironics.com
Respironics, CAPNOSTAT and LoFlo are registered trademarks and 21st Century CO2
Technology is a trademark of Respironics, Inc. and its affiliates.
CAUTION: U.S. federal law restricts this device to sale by or on the order of a physician.
2007 Respironics, Inc. and its affiliates. All rights reserved.

Geyer KB 7/10/07 MCI 4101194