Sunteți pe pagina 1din 994

Phillips

Setup Manuals
Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Setup Philips C-arm BV Libra / Endura / Pulsera


(native DICOM ie newer approach)
with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual).................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 8
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual)(1)................ 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................26
n Kodak DryView 8900 LASER IMAGER......................................................................... 28
n Kodak DryView 8150 LASER IMAGER......................................................................... 30
n Kodak Color Medical Imager 1000 ................................................................................. 32

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Kodak DryView 8300

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8300, IP=accordingly, port nbr=104, Printer Type=Kodak 8300
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=8INx10IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=2, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=0,10
n to print select Target=DryView, Print page format=2x3-6 , click Ok&Export

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=2
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips C-Arm Kodak 190 camera


- BV family:
Libra, Endura, Pulsera

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 190 rel. 1.7.29

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=Kodak, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=190, IP=accordingly, port nbr=5040, Printer Type=Kodak 190
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnif=CUBIC, Smoothing=NORMAL, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=CS435\CN0
n to print select Target=Kodak, Print page format=3x4-12 , click Ok&Export

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips C-Arm
Fiber Cable
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=9410, IP=accordingly, port nbr=1024, Printer Type=Kodak 8700
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n on Device Setup, SCP PRINTERS, Registered Users:


n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87

Philips C-Arm
Fiber Cable
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8500 Plus

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 9410 v1.6, 8500 v1.34

Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into 9410 Local Printers

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=9410, IP=accordingly, port nbr=1024, Printer Type=Kodak 8500
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=11INx14IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8500, Model=8500, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8500, AE Title= 9410, Port= 1024, Max Associations= 5

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n on Device Setup, SCP PRINTERS, Registered Users:


n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8500, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 9410 v1.6, 8100 v2.4.1

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=9410, IP=accordingly, port nbr=1024, Printer Type=Kodak 8100
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low


n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips C-Arm COPCIL board
- BV family: Copper Cable
Libra, Endura, Pulsera
Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=9410, IP=accordingly, port nbr=1024, Printer Type=Kodak 8300
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=8INx10IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=2, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=0,10
n to print select Target=DryView, Print page format=2x3-6 , click Ok&Export

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low


n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips C-Arm
- BV family: Kodak 160 camera
Crossed
Libra, Endura, Pulsera
Network Cable

Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, KELI 160 v7.2

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=Kodak, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=160, IP=accordingly, port nbr=5040, Printer Type=Kodak 160
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnif=CUBIC, Smoothing=NORMAL, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=CS435\CN0
n to print select Target=Kodak, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips C-Arm
Kodak 2180 camera
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 2180 rel. 2.17H

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=Kodak, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=2180, IP=accordingly, port nbr=5040, Printer Type=Kodak 2180
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnif=CUBIC, Smoothing=NORMAL, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=CS435\CN0
n to print select Target=Kodak, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips C-Arm
Kodak 1120 camera
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=Kodak, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=1120, IP=accordingly, port nbr=5040, Printer Type=Kodak 1120
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnif=CUBIC, Smoothing=NORMAL, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=CS435\CN0
n to print select Target=Kodak, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Interpolation: NoMag=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips C-Arm
Crossed
- BV family:
Network
Libra, Endura, Pulsera
Cable

Kodak MIM 200 Kodak DryView 8200

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM 200 v6.1, DV8200 v3.5
Note: This setup is basically also valid for 25 PS computer

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8200, IP=accordingly, port nbr=5040, Printer Type=Kodak 8200
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Supported Item: Remote Image Deletion=On, Rotation=On


n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips C-Arm
Fiber Cable
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8700, IP=accordingly, port nbr=5040, Printer Type=Kodak 8700
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Media Info: Media=35x43cm, 4096x5223, 5133, Blue


n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87

Philips C-Arm
Fiber Cable
- BV family:
Libra, Endura, Pulsera
Fiber Cable

Kodak MIM 200 Kodak DryView 8500 Plus

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 8500 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into MIM200 Destination

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8500, IP=accordingly, port nbr=5040, Printer Type=Kodak 8500
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=11INx14IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8500, Set:
n General: Log.Name=8500
n General: Database Name=8500, Model=8500, Memory=32

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Media Info: Media=11x14in, 3388x4283, 4193, Blue


n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=11INx14IN (Always=N)
n Negotiate Color Print=Off

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 8100 v2.4.1
Note: This setup is basically also valid for 25 PS computer

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8100, IP=accordingly, port nbr=5040, Printer Type=Kodak 8100
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips C-Arm COPCIL board
- BV family: Copper Cable
Libra, Endura, Pulsera
Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B
Note: This setup is basically also valid for 25 PS computer

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8300, IP=accordingly, port nbr=5040, Printer Type=Kodak 8300
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=8INx10IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=2, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=0,10
n to print select Target=DryView, Print page format=2x3-6 , click Ok&Export

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Kodak DryView 8900

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=8900, IP=accordingly, port nbr=5040, Printer Type=Kodak 8900
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Image Size=On


n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=N), Contrast=9 (Always=N)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=Y)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=bv

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Kodak DryView 8150

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=DryView, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=DV8150, IP=accordingly, port nbr=5040, Printer Type=Kodak 8100
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=14INx17IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=11, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=Pmsv3k.w87,9
n to print select Target=DryView, Print page format=3x4-12 , click Ok&Export

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Formatting: Req.Size Mix Row=On, Trim=On,


n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name= Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=bv

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips C-Arm
- BV family:
Libra, Endura, Pulsera

Kodak CMI 1000

Requirements: BV Libra, Endura, Pulsera rel.1.2 with DFI sw v1.4


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

BV Setup:
n by using the BV Service Tool
n select Program Manual/DICOM Target Configuration/Printer Targets
n select Printer Target=no name
n set: Name=CMI1000, Type=Print, Order in list=eg 1, Enable=Yes
n AE title=CMI1000, IP=accordingly, port nbr=5040, Printer Type=Kodak 8300
n select next
n check that Priority=MED, Destination=PROCESSOR, Bin=1,
n Orientation=PORTRAIT, Film Size=8INx10IN, Medium Type=BLUE FILM
n Border Density=BLACK, Border Density Value=0, Copies=1
n Magnification=CUBIC, Smoothing=2, Min Density=20, Max Density=300
n Polarity=NORMAL, Trim=NO, Configuration Information=0,10
n to print select Target=CMI1000, Print page format=2x3-6, click Ok&Export

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips C-arm BV Libra / Endura / Pulsera with LASER IMAGER Products

n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)


n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Setup Philips C-arm Libra/Endura/Pulsera/BV300


via Merge DICOM-converter (ie older approach)
with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual).................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 8
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual)(1)................ 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................26
n Kodak DryView 8900 LASER IMAGER......................................................................... 28
n Kodak DryView 8150 LASER IMAGER......................................................................... 30
n Kodak Color Medical Imager 1000 ................................................................................. 32

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board

Video Cable
Image Video OUT
Merge Video Board
Kodak DryView 8300

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 3. 8300,
n AE Title=8300, IP Number (ie IP address)=accordingly, Port=104
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=8300.fmt,
- AE_TITLE=8300, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8300] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 8300, PORT_NUMBER=104

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/8300.prn under [AETITLE_OVERRIDE] check:


- Medium_Type=CURRENT, Film_Destination=PROCESSOR,
- Film_Size=8INX10IN, Magnification_Type=CUBIC,
- Smoothing_Type=0, Border_Density=BLACK,
- Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,10,
- Polarity=NORMAL

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable Kodak 190 camera
Serial 232 Link OUT Merge Comm Board

Video Cable
Image Video OUT
Merge Video Board

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 190 rel. 1.7.29

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=K190
n specify print format from list 14. MLP190,
n AE Title=190, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_K190] check:
- LOGGING=YES, NAME=K190, FORMAT_FILE=MLP190.fmt,
- AE_TITLE=190, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [190] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 190, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/MLP190.prn under [AETITLE_OVERRIDE] check:


- Film_Size=14INX17IN, Magnification_Type=CUBIC,
- Smoothing_Type=NORMAL, Border_Density=BLACK,
- Min_Density=20, Max_Density=300, Trim=NO
- Configuration_ Information=CS490, Polarity=NORMAL

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network Kodak DryView 8700 Plus


Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
yView
- BV300 3M 87
00 Dr

Fiber Cable
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Fiber Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak 9410 PACS Link

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=9410, IP Number (ie IP address)=accordingly, Port=1024
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=9410, PROTOCOL= DICOM BASIC PRINT,

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [9410] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 9410, PORT_NUMBER=1024
n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network Kodak DryView 8500 Plus


Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
yView
- BV300 3M 87
00 Dr

Fiber Cable
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Fiber Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak 9410 PACS Link

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 9410 v1.6, 8500 v1.34

Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into 9410 Local Printers

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=9410, IP Number (ie IP address)=accordingly, Port=1024
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=9410, PROTOCOL= DICOM BASIC PRINT,

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [9410] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 9410, PORT_NUMBER=1024
n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8500, Model=8500, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8500, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8500, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network
Merge
RS422
Philips C-Arm Keypad
- BV family: Control 9-pin 9-pole Plug
Merge ExamWorks COPCIL Board
Libra, Endura, Pulsera
Copper Cable
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Data 37-pin
Video Cable
Image Video OUT
Merge Video Board Kodak DryView 8100
Kodak 9410 PACS Link

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 9410 v1.6, 8100 v2.4.1

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=9410, IP Number (ie IP address)=accordingly, Port=1024
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=9410, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [9410] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 9410, PORT_NUMBER=1024

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]


check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
Control 9-pin
- BV family:
Merge ExamWorks COPCIL Board
Libra, Endura, Pulsera
Copper Cable
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Data 37-pin
Video Cable
Image Video OUT
Merge Video Board Kodak DryView 8300
Kodak 9410 PACS Link

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 3. 8300,
n AE Title=9410, IP Number (ie IP address)=accordingly, Port=1024
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=8300.fmt,
- AE_TITLE=9410, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8300] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 9410, PORT_NUMBER=1024
n into usr/home/merge/xpi/config/8300.prn under [AETITLE_OVERRIDE] check:

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- Medium_Type=CURRENT, Film_Destination=PROCESSOR,
- Film_Size=8INX10IN, Magnification_Type=CUBIC,
- Smoothing_Type=0, Border_Density=BLACK,
- Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,10,
- Polarity=NORMAL

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
Kodak 160 camera
RS232 Cable Crossed
Serial 232 Link OUT Merge Comm Board
Network Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, KELI 160 v7.2

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=K160
n specify print format from list 13. MIM,
n AE Title=160, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_K160] check:
- LOGGING=YES, NAME=K160, FORMAT_FILE=MIM.fmt,
- AE_TITLE=160, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [160] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/MIM.prn under [AETITLE_OVERRIDE] check:


- Magnification_Type=CUBIC, Smoothing_Type=NORMAL,
- Border_Density=BLACK, Min_Density=20, Max_Density=300,
- Trim=NO, Configuration_ Information=CS490, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
Kodak 2180 camera
RS232 Cable Crossed
Serial 232 Link OUT Merge Comm Board
Network Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 2180 rel. 2.17H

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=K2180
n specify print format from list 13. MIM,
n AE Title=2180, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_K2180] check:
- LOGGING=YES, NAME=K2180, FORMAT_FILE=MIM.fmt,
- AE_TITLE=2180, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [2180] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/MIM.prn under [AETITLE_OVERRIDE] check:


- Magnification_Type=CUBIC, Smoothing_Type=NORMAL,
- Border_Density=BLACK, Min_Density=20, Max_Density=300,
- Trim=NO, Configuration_ Information=CS490, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off
2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
Kodak 1120 camera
RS232 Cable Crossed
Serial 232 Link OUT Merge Comm Board
Network Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=K1120
n specify print format from list 13. MIM,
n AE Title=1120, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_K1120] check:
- LOGGING=YES, NAME=K1120, FORMAT_FILE=MIM.fmt,
- AE_TITLE=1120, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [1120] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/MIM.prn under [AETITLE_OVERRIDE] check:


- Magnification_Type=CUBIC, Smoothing_Type=NORMAL,
- Border_Density=BLACK, Min_Density=20, Max_Density=300,
- Trim=NO, Configuration_ Information=CS490, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable Crossed
Serial 232 Link OUT Merge Comm Board
Network Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200 Kodak DryView 8200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM 200 v6.1, DV8200 v3.5

Note: This setup is basically also valid for 25 PS computer

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=8200, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=8200, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8200] check:

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040
n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)


n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network Kodak DryView 8700 Plus


Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
yView
- BV300 3M 87
00 Dr

Fiber Cable
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Fiber Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=8700, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- AE_TITLE=8700, PROTOCOL= DICOM BASIC PRINT,


- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8700] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040
n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network Kodak DryView 8500 Plus


Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
yView
- BV300 3M 87
00 Dr

Fiber Cable
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Fiber Cable

Video Cable
Image Video OUT
Merge Video Board
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 8500 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into MIM200 Destination

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=8500, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- AE_TITLE=8500, PROTOCOL= DICOM BASIC PRINT,


- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8500] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040
n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8500, Set:
n General: Log.Name=8500
n General: Database Name=8500, Model=8500, Memory=32
n Media Info: Media=11x14in, 3388x4283, 4193, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=11INx14IN (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n Negotiate Color Print=Off

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network
Merge
RS422
Philips C-Arm Keypad
- BV family: Control 9-pin 9-pole Plug
Merge ExamWorks COPCIL Board
Libra, Endura, Pulsera
Copper Cable
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Data 37-pin
Video Cable
Image Video OUT
Merge Video Board Kodak DryView 8100
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=8100, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=8100, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8100] check:
- SERVICE_LIST=Grayscale_Print_Service_List

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040


n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]
check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
Control 9-pin
- BV family:
Merge ExamWorks COPCIL Board
Libra, Endura, Pulsera
Copper Cable
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board
Data 37-pin
Video Cable
Image Video OUT
Merge Video Board Kodak DryView 8300
Kodak MIM 200

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 3. 8300,
n AE Title=8300, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=8300.fmt,
- AE_TITLE=8300, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8300] check:
- SERVICE_LIST=Grayscale_Print_Service_List

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

- HOST_NAME=IPaddress of MIM200, PORT_NUMBER=5040


n into usr/home/merge/xpi/config/8300.prn under [AETITLE_OVERRIDE] check:
- Medium_Type=CURRENT, Film_Destination=PROCESSOR,
- Film_Size=8INX10IN, Magnification_Type=CUBIC,
- Smoothing_Type=0, Border_Density=BLACK,
- Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,10,
- Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board

Video Cable
Image Video OUT
Merge Video Board
Kodak DryView 8900

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=8900, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=8900, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8900] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 8900, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]


check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=bv

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board

Video Cable
Image Video OUT
Merge Video Board
Kodak DryView 8150

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=DryView
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=DV8150, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_DryView] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=DV8150, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [DV8150] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of 8150, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]


check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name= Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=bv

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

Philips C-arm Libra / Endura / Pulsera / BV300 via DICOM with


Kodak Color Medical Imager 1000

DICOM Network
Merge
Philips C-Arm Keypad
- BV family:
Merge ExamWorks
Libra, Endura, Pulsera
- BV300
25-to-25 pin
RS232 Cable
Serial 232 Link OUT Merge Comm Board

Kodak CMI 1000


Video Cable
Image Video OUT
Merge Video Board

Requirements: BV Libra, Endura, Pulsera, BV300 rel.3, Merge ExamWorks v1.4


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Merge ExamWorks Setup:


n by using the Merge keyboard and the Merge touch screen to give inputs
n at Merge touchscreen select Utils, Exit to Service, enter the password
n press Ctrl-Alt-Backspace to enter Service Mode, press Enter
n enter the bvfamily login and the password
n cd xpi/bin
n su
n enter the superuser password (to get superuser rights)
n config.pl
n select 7. Perform Output Configuration only
n select 2. Configure with Storage Commitment=No
n select 2. Will at least one DICOM Storage target use data compression=No
n enter n to add new target
n select 2. DICOM Basic Print
n name=CMI1000
n specify print format from list 2. 9410_PACS_LINK,
n AE Title=CMI1000, IP Number (ie IP address)=accordingly, Port=5040
n Select 1. Default to this target being selected on startup=Yes
n Press Enter to return to targets menu, press q to quit the targets menu
n select 1. Accept changes in Output Configuration=Yes
n Press Enter to return to main menu, press q to terminate configuration, exit, exit
n
n into usr/home/merge/xpi/config/targets.ini under [TARGET_nbr_of_CMI1000] check:
- LOGGING=YES, NAME=DryView, FORMAT_FILE=9410_PACS_LINK.fmt,
- AE_TITLE=CMI1000, PROTOCOL= DICOM BASIC PRINT,
- DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [CMI1000] check:
- SERVICE_LIST=Grayscale_Print_Service_List
- HOST_NAME=IPaddress of CMI1000, PORT_NUMBER=5040

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips C-arm Libra/Endura/Pulsera/BV300 via Merge-box with LASER IMAGER Products

n into usr/home/merge/xpi/config/9410_PACS_LINK.prn under [AETITLE_OVERRIDE]


check:
- Magnification_Type=CUBIC, Smoothing_Type=11,
- Border_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips CT Aura / Secura with LASER IMAGER Products

Setup Philips CT Aura / Secura


with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual........................................................ 4
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 5
n Kodak DryView 8700 LASER IMAGER / Standard ....................................................... 6
n Kodak Digital Science 969 HQ LASER IMAGER...................................................... 7
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 8
n Kodak Digital Science 959 XL LASER IMAGER ...................................................... 9
n Kodak DryView 8100 LASER IMAGER......................................................................... 10
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) (1) ............................... 12
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) (4) ............................... 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 24

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips CT Aura / Secura with LASER IMAGER Products

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 27


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 28
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) (1) .............................. 30
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 32
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 34
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600) (1) ............................................................................................................ 36
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 38
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 40
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 42
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 44
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) (4,5)............................. 46
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5) ................................................................. 48
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4,5) ................................................................. 50
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600) (4,5) .......................................................................................................... 52
n Kodak DryView 8900 LASER IMAGER......................................................................... 54
n Kodak DryView 8150 LASER IMAGER......................................................................... 56
n Kodak Color Medical Imager 1000 ................................................................................. 58

5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips CT
Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
anag
putM
i-In
Mult
8800
3M

iew
00 DryV
3M 87

Fiber Cable
8700 Dual

8800 MIM

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 8800 rel. 4.2, 8700 rel. 1.15

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual
Philips CT
Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
anag
putM
i-In
Mult
8800
3M

iew
00 DryV
3M 87

Fiber Cable
8700 Dual
iew
00 DryV
3M 87

Fiber Cable

8800 MIM
8700 Dual

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 8800 rel. 5.1, 8700 rel. 1.22
CT Aura / Secura Setup (so the 2 cameras are treated independently):
n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface type=PCI/SBus, Printer type=3M 8700, Name=dv8700nbr1
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n into Advanced Printer Properties: Shared Interface=On, Dest. Nbr=0, Proceed, Exit
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface type=PCI/SBus, Printer type=3M 8700, Name=dv8700nbr2
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n into Advanced Printer Properties: Shared Interface=On, Dest. Nbr=1
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle
Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
8800/8700 Dual Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak DryView 8700 LASER IMAGER / Plus

Philips CT
Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

iew
00 DryV
3M 87

8700 Plus

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 8700 Plus rel. 1.15

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnif.=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard
Philips CT
Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
Image Data OUT

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 8700 Standard rel. 1.15

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnification=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak Digital Science 969 HQ LASER IMAGER

Philips CT RS422 37-to-9 pin


Philips 37-pin
Aura / Secura Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M H Q L a ser Im a g er

M969 HQ

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, M969 HQ rel. 3.0

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M HQ 969, Name=hq969
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correction=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak Digital Science 969 HQT LASER IMAGER

Philips CT M969 HQT


Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT 3M H Q L a ser Im a g er

Philips 37-pin
Data Cable
Image Data OUT

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, M969 HQT rel. 1.0
Note: in case of image quality problems update HQT to rel. 2.0

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M HQ 969, Name=hqt969
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:

n on TDB/C board: Jumpers=RS422, Switch=Center


n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=No, Scale Format=Nos, Scale Magnif.=No
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak Digital Science 959 XL LASER IMAGER

Philips CT RS422 37-to-9 pin


Philips 37-pin
Aura / Secura Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a se r Im a g er

M959 XL

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, M959 XL rel. 3.8

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 959 XL, Name=xl959
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup:
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu: 959 Default
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak DryView 8100 LASER IMAGER

DryView 8100

Philips CT RS422 9-pole


Aura / Secura Philips 37-pin RS422 37-to-9 pin plug
Control Cable Host Adaptor Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 8100 v1.1

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8100
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=180, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ)

Philips CT
Aura / Secura Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 DryV
iew Fiber Cable

Fiber Cable 9410 PACS Link

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into 9410 Local Printers and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD Camera Config.File

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips CT Aura / Secura with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=ev, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips CT RS422 37-to-9 pin


Philips 37-pin
Aura / Secura Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 9410 v1.6, 8100 v2.4.1

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8100
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=ev, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips CT Aura / Secura with LASER IMAGER Products

n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)


n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Aura / Secura Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM 200 v6.1, KELI 160 v7.2

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELI-160 (or MLP-190), Name=k160
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ctas,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips CT Aura / Secura with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Aura / Secura Control Cable
Kodak 2180 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM 200 v6.1, KELP 2180 v2.17H

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELP 2180, Name=k2180
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips CT Aura / Secura with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Size Scale=On


n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Aura / Secura Control Cable
Kodak 1120 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELP 1120, Name=k1120
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips CT Aura / Secura with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Size Scale=On


n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ )

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable 00 DryV
iew
3M 87

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into MIM 200 Destinations and in case of 8500 and HQ
by installing 8500 and HQ into EasyVision/EasyRAD Camera Config.File

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=MOV
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips CT Aura / Secura with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips CT Philips 37-pin Octopus Adaptor Cable COPCIL Board
Control Adaptor Cable Control 9-pin
Aura / Secura Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (ie EasyVision) 4.2, MIM200 v6.1, 8100 v2.4.1

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer= 3M 8100 (or 8700), Name=dv8100
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips CT Aura / Secura with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300

Philips CT
Aura / Secura

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8300, port=104, AETitle=8300
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=4K or 16K,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=8300
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=1
n Film Layout : Format=1 on 1, Mode=Portrait
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips CT Kodak 190 camera


Aura / Secura

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 190 rel. 1.7.29

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 190, port=5040, AETitle=190
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak MLP-190, Name=k190,
n AE title=190, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ)

DICOM Network

Kodak DryView 8700 Plus

Philips CT
Fiber Cable
Aura / Secura 00 DryV
iew
3M 87

Fiber Cable

Kodak 9410 PACS Link

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into 9410 Local Printers and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD DICOM Camera Config.File
(if HQ DICOM Camera Config.File not yet available it will also work with 8700)

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8700, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips CT Aura / Secura with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT
Aura / Secura

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, 9410 v1.6, 8100 v2.4.1

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8100, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips CT Aura / Secura with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Kodak DryView 8300

Control 9 pin
Philips CT COPCIL board
Aura / Secura Copper Cable

Data 37 pin

Kodak 9410 PACS Link

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=9410
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips CT Aura / Secura with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

Kodak DryView 8610


Control 9 pin
Philips CT COPCIL board
Aura / Secura Copper Cable

Data 37 pin

Kodak 9410 PACS Link

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 5.1
Requirements: 9410 v1.6, 8610 v2.4 (if 8600 v2.5)
Note: This setup is basically also valid for 8600 camera

EasyVision Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8610, Name=dv8610, AE title=9410
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print) (Off if 8600)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips CT Aura / Secura with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8610, Model=8600/8610, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=12 (8 if 8600), Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=1, Installed TFT/ULUT=don’t care
n Max.Density=300, Contrast=5, Border Density=0
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8610, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8610, Copies=1, Contrast Table=5, Max.Density=300
n Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Gamma Table=Linear in Transmittance, Contrast Group=1, Kernel Group=1
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, VRTX32=1.08, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P, 2:1P, 4:1P, 6:1P
n Gamma Table=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips CT
Kodak 160 camera
Aura / Secura

Crossed Network Cable

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM 200 v6.1, KELI 160 v7.2

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=160
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELI-160 (or MLP-190), Name=k160
n AE title=160, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips CT Aura / Secura with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips CT
Kodak 2180 camera
Aura / Secura

Fiber Cable

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) v4.3
Requirements: MIM200 v6.1, 2180 rel. 2.17H

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=2180
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELP2180 (or MLP-190), Name=k2180,
n AE title=2180, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips CT Aura / Secura with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips CT
Kodak 1120 camera
Aura / Secura

Fiber Cable

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) v4.3
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=1120
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELP1120 (or MLP-190), Name=k1120,
n AE title=1120, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips CT Aura / Secura with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Kodak DryView 8200

Philips CT
Aura / Secura

Crossed Network Cable

Kodak MIM 200

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM200 v6.1, 8200 v3.5
Note: This setup is basically also valid for 25 PS computer

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8200
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM,Printer=3M 8200 (or 8700),Name=k8200,AE title=8200
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips CT Aura / Secura with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:


n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ )

DICOM Network

iew
00 DryV
3M 87

Philips CT
Fiber Cable
Aura / Secura

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into MIM200 Destination and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD DICOM Camera Config.File
(if HQ DICOM Camera Config.File not yet available it will also work with 8700)

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8700
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8700, AE title=8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips CT Aura / Secura with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT
Aura / Secura

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2, MIM200 v6.1, 8100 v2.4.1
Note: This setup is basically also valid for 25 PS computer

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8100
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8100 (or 8700),
n Name=dv8100, AE title=8100
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips CT Aura / Secura with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:


n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 46


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips CT COPCIL board
Aura / Secura Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (i.e., EasyVision) 4.2
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B
Note: This setup is basically also valid for 25 PS computer

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8300
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=8300
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 47


Philips CT Aura / Secura with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.811
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=u1_109b.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 48


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips CT COPCIL board
Aura / Secura Copper Cable

Data 37 pin

Kodak MIM 200


Note: Filming Station of CT Aura / Secura is an EasyVision workstation
Requirements: CT Station (ie EasyVision rel. 5.1)
Requirements: MIM200 v6.1, 8610 v2.4 (if 8600 v2.5)
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8600 camera

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8610
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8610, Name=dv8610, AE title=8610
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 49


Philips CT Aura / Secura with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8610 (or 8600), Set:
n General: Log.Name=8610 (or 8600)
n General: Database=8610 (or 8600), Model=8610 (or 8600), Memory=32
n Media Info: Media=8x10in, 5025x6200, 6020, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12 (8 if 8600), Header/Line,Transfer=2000,Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=25.8065
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.5
n M952 Usage: Min H/V Mag=0.26, Max H/V Mag=16, Min Dmax 952=2
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= upmsv3kp.w8u (Always=N), Contrast (dry)=5 (Always=N)
n Modality Group=1, Conv.Group=1, Gamma Table=Linear in Transmittance
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=1 (Always=Y), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None (Always=N)
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off
8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 50


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips CT
Aura / Secura

Kodak DryView 8900

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (ie EasyVision) 5.2 Level 2
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8900, port=5040, AETitle=8900
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface= DICOM, Printer= Kodak 8900 standard
n (or 3M / Kodak 8900 and in this case 2mm Border=No, or 3M 8200),
n Name=dv8900, AE title=8900
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any

Author: gian.valesano@kodak.com, Edition: May ´05 51


Philips CT Aura / Secura with LASER IMAGER Products

n Media Types: Media=35x35cm,9100x9087, 8831, Any


n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=0 (Always=Y), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ct

Author: gian.valesano@kodak.com, Edition: May ´05 52


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips CT
Aura / Secura

Kodak DryView 8150

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (ie EasyVision) 5.2 Level 2
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8150, port=5040, AETitle=DV8150
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Kodak 8150 DICOM (or 3M 8200),
n Name=dv8150, AE title=DV8150
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any

Author: gian.valesano@kodak.com, Edition: May ´05 53


Philips CT Aura / Secura with LASER IMAGER Products

n Media Types: Media=35x35cm, 4438x4414,4286, Any


n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=0 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ct

Author: gian.valesano@kodak.com, Edition: May ´05 54


Philips CT Aura / Secura with LASER IMAGER Products

Philips CT Aura / Secura via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips CT
Aura / Secura

Kodak CMI 1000

Note: Filming Station of CT Aura / Secura is an EasyVision workstation


Requirements: CT Station (ie EasyVision) 5.2 Level 2
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3))

CT Aura / Secura Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of CMI-1000, port=5040, AETitle=CMI1000
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Print SOP Classes set:
n Basic Greyscale=On, Basic Color=On
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface= DICOM,
n Printer= Kodak CMI-1000 DICOM (or Codonics-NP1660MD-DICOM),
n Name=CMI1000, AE title=CMI1000
n into Config. Media List:
n if Kodak CMI-1000 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, A4 Color RGB=On, all others Off
n if Letter inside CMI-1000 Paper Tray, A Color RGB=On, others=Off
n if Codonics-NP1660 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, set A4 P Col=On, others=Off
n if Letter inside CMI-1000 Paper Tray, set A P Col=On, others=Off
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray

Author: gian.valesano@kodak.com, Edition: May ´05 55


Philips CT Aura / Secura with LASER IMAGER Products

(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):


at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=N), TFT Set (From list)=Pmsv3k.w87 (Always=N)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=0 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 56


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

Setup for Philips CT Mx8000 / Brilliance


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

Philips CT Mx8000 / Brilliance via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips CT Mx8000

Kodak DryView 8300


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8300 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8300, AETitle=8300
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=104, Max PDU=16384
n into Expert Mode set: Port=104, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8300
n select Options set Film Format=6 (2x3), Film Size=8x10 or 14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

Philips CT Mx8000 / Brilliance via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips CT Mx8000 Kodak 190 camera

Requirements: Mx8000 CT v3.52 (based on Unix platform)


Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 190 rel. 1.7.29

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=190 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=190, AETitle=190
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=190
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips CT Mx8000
Fiber Cable

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into 9410 Local Printers

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8700 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8700, AETitle=9410
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=1024, Max PDU=16384
n into Expert Mode set: Port=1024, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8700
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None


n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=13
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT Mx8000

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100
Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 9410 v1.6, 8100 v2.4.1

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8100 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8100, AETitle=9410
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=1024, Max PDU=16384
n into Expert Mode set: Port=1024, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8100
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

on Device Setup, SCP PRINTERS, Current Providers:


n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=13
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

Philips CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips CT Mx8000 COPCIL board
Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8300 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8300, AETitle=8300
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=1024, Max PDU=16384
n into Expert Mode set: Port=1024, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8300
n select Options set Film Format=6 (2x3), Film Size=8x10 or 14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n on Device Setup, SCP PRINTERS, Current Providers:


n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips CT Mx8000
Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, KELI 160 v7.2

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=160 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=160, AETitle=160
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=160
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n General: Database Name=KELI 160, Model=KELI 160


n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips CT Mx8000
Kodak 2180 camera

Fiber Cable

Kodak MIM 200


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, 2180 rel. 2.17H

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=2180 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=2180, AETitle=2180
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=2180
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly


n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips CT Mx8000
Kodak 1120 camera

Fiber Cable

Kodak MIM 200


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=1120 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=1120, AETitle=1120
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=1120
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly


n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Kodak DryView 8200

Philips CT Mx8000

Crossed Network Cable

Kodak MIM 200


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM 200 v6.1, DV8200 v3.5

Note: This setup is basically also valid for 25 PS computer

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8200 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8200, AETitle=8200
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8200
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:


n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips CT Mx8000
Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into MIM200 Destination

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8700 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8700, AETitle=8700
n set Capability file=Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8700
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Node Name=MIM, IPAddr=accordingly


n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT Mx8000

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8100 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8100, AETitle=8100
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8100
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9


n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips CT Mx8000 COPCIL board
Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8300 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8300, AETitle=8300
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8300
n select Options set Film Format=6 (2x3), Film Size=8x10 or 14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:


n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips CT Mx8000

Kodak DryView 8900


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8900 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8900, AETitle=8900
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8900
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Media Types: Media=35x35cm,9100x9087, 8831, Any


n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=13 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ct

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips CT Mx8000

Kodak DryView 8150


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

CT Setup:
n login as Configuration
n select Network/Add Node=Other
n set Name=DV8150 and click OK
n set IP addr.=accordingly, Add Device=Printer, Name=DV8150, AETitle=DV8150
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer=DV8150
n select Options set Film Format=20 (4x5), Film Size=14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Media Types: Media=35x35cm, 4438x4414,4286, Any


n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=13 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ct

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

CT Mx8000 / Brilliance via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips CT
- Mx8000
- Brilliance

Kodak CMI 1000


Requirements: Mx8000 CT v3.52 (based on Unix platform)
Requirements: Mx8000 CT v4.1 (based on Windows platform)
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

CT Mx8000 Brilliance Setup:


n login as Configuration
n select Network/Add Node=Other
n set Name= CMI1000 and click OK
n set IP addr.=accordingly, Add Device=Printer,
n Name=CMI1000, AETitle=CMI1000
n set Capability file= Kodak190_DICOM
n into Expert Mode set: Enable_AutoDisconnect=On, Port=5040, Max PDU=16384
n into Expert Mode set: Port=5040, ARTIM=5, click OK, OK, Exit, OK (reboot)
n login as MD
n select images to be printed, select File/Film, select Tasks/Master Film
n into Master Film:
n select File/Printer= CMI1000
n select Options set Film Format=6 (2x3), Film Size=8x10 or 14x17 (Inch)
n select Options set Film Orientation=Portrait, Print Priority=High, Nbr.Copies=1
n select Options set Medium Type=Blue Film, Film Destination=Processor
n select Options set Magnif.Type=Cubic, Smoothing=Printer Default
n select Options set Border Density=Black, Empty Image Density=Black
n select Options set Trim=No, Polarity=Normal
n select File/Print

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips CT Mx8000 / Brilliance with LASER IMAGER Products

n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=13 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips CT Tomoscan M/EG with LASER IMAGER Products

Setup Philips CT Tomoscan M/EG


with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 4
n Kodak DryView 8700 LASER IMAGER / Standard ....................................................... 5
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual.............................................................. 6
n Kodak DryView 8500 LASER IMAGER / Plus............................................................... 7
n Kodak DryView 8500 LASER IMAGER / Standard ....................................................... 8
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual.............................................................. 9
n Kodak DryView 8300 LASER IMAGER / Standalone.................................................... 10
n Kodak Digital Science 969 HQ LASER IMAGER...................................................... 11
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 12
n Kodak Digital Science 959 XL LASER IMAGER ...................................................... 13
n Kodak DryView 8100 LASER IMAGER......................................................................... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 15
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 21
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) . 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 27

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips CT Tomoscan M/EG with LASER IMAGER Products

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 29


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 30
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 31
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 33
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 35
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 37
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 39
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 41
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 43
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 45
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER(4,5)................................................................... 47
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER(4,5)................................................................... 49
n Kodak DryView 8900 LASER IMAGER......................................................................... 51
n Kodak DryView 8150 LASER IMAGER......................................................................... 53
n Kodak Color Medical Imager 1000 ................................................................................. 55

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual
25-to-25 Null RS232 25-to-9 pin
Philips 25-pin
Philips CT Tomoscan M/EG Modem Cable Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

er
nag
ut Ma
-Inp
ulti
0M
880
3M

iew
0 DryV
3M 870

Fiber Cable
8700 Dual

8800 MIM

Requirements: Tomoscan M/EG rel. 3.0, 8800 rel. 4.2, 8700 rel. 1.15

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8700 LASER IMAGER / Plus
Philips 25-pin RS232 25-to-9 pin
Philips CT Tomoscan M/EG 25-to-25 Null
Control Cable Host Adaptor Cable
Modem Cable
Host Control 232OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

iew
0 DryV
3M 870

8700 Plus

Requirements: Tomoscan M/EG rel. 3.0, 8700 Plus rel. 1.15

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8700 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard
Philips 25-pin 25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG
Control Cable Modem Cable Host Adaptor Cable
Host Control 232 OUT
iew
00 DryV
3M 87

Philips 37-pin
I mage Data OUT Data Cable

Requirements: Tomoscan M/EG rel. 3.0, 8700 Standard rel. 1.15

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

8700 Standard Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual
25-to-25 Null RS232 25-to-9 pin
Philips 25-pin
Philips CT Tomoscan M/EG Modem Cable Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

er
nag
ut Ma
-Inp
ulti
0M
880
3M

iew
0 DryV
3M 870

Fiber Cable
8500 Dual

8800 MIM

Requirements: Tomoscan M/EG rel. 3.0, 8800 rel. 5.1, 8500 rel. 1.22

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8800/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8500 LASER IMAGER / Plus
Philips 25-pin RS232 25-to-9 pin
Philips CT Tomoscan M/EG 25-to-25 Null
Control Cable Host Adaptor Cable
Modem Cable
Host Control 232OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

iew
0 DryV
3M 870

8500 Plus

Requirements: Tomoscan M/EG rel. 3.0, 8500 Plus rel. 1.22

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8500 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard
Philips 25-pin 25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG
Control Cable Modem Cable Host Adaptor Cable
Host Control 232 OUT
iew
00 DryV
3M 87

Philips 37-pin
I mage Data OUT Data Cable

Requirements: Tomoscan M/EG rel. 3.0, 8500 Standard rel. 1.22

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8500 Standard Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips CT Tomoscan M/EG Philips 25-pin RS232 25-to-9 pin


25-to-25 Null
Control Cable Host Adaptor Cable
Modem Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
Manag
put
ulti-In

Fiber Cable
00M
3M88

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: Tomoscan M/EG rel. 3.0, 8800 rel. 5.1, 8300 rel. 4.2

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

8300 Setup:
n on 8300 Operator Menu set: Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None, Modality Group=0

8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu: Pixel Corr.=Yes, Scale Format=Yes, Scale Mag.=Yes
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 25-pin 25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG
Control Cable Modem Cable Host Adaptor Cable
Host Control 232 OUT

Philips 37-pin Data Cable


I mage Data OUT

Requirements: Tomoscan M/EG rel. 3.0, 8300 rel. 4.6

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

8300 Setup:
n on 8300 Operator Menu set:
n Density=16, Contrast=10, Sharp/Smooth=Auto
n Film Layout : Format=6 on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Delimiter=Comma, Protocol=3M
n Digital Interface: Parity=None, Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak Digital Science 969 HQ LASER IMAGER
25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG Philips 25-pin
Modem Cable Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M H Q L a ser Im a g e r

M969 HQ

Requirements: Tomoscan M/EG rel. 3.0, M969 HQ rel. 3.0

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

M969 HQ Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 25-pin 25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG
Control Cable Modem Cable Host Adaptor Cable 3M H Q L a ser Im a g er

Host Control 232 OUT

Philips 37-pin
I mage Data OUT Data Cable

Requirements: Tomoscan M/EG rel. 3.0, M969 HQT rel. 1.0


Note: in case of image quality problems update HQT to rel. 2.0

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

M969 HQT Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak Digital Science 959 XL LASER IMAGER
25-to-25 Null RS232 25-to-9 pin
Philips CT Tomoscan M/EG Philips 25-pin
Modem Cable Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a se r I m a g er

M959 XL

Requirements: Tomoscan M/EG rel. 3.0, M959 XL rel. 3.8

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

Setup:
n on Philips Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center
n on XL Input Module install LUT 322A0

M959 XL Setup:
n on COMM Menu: 959 Default
n on EIB Menu: Parity=Disable, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG with


Kodak DryView 8100 LASER IMAGER

Philips 25-pin 25-to-25 Null RS232 9-pole


Philips CT Tomoscan M/EG RS232 25-to-9 pin plug
Control Cable Modem Cable Host Adaptor Cable
Host Control 232 OUT

Philips 37-pin
Image Data OUT Data Cable

DryView 8100

Requirements: Tomoscan M/EG rel. 3.0, 8100 v1.1

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Disable

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips CT Tomoscan M/EG Philips 25-pin 25-to-25 Null RS232 25-to-9 pin
Control Cable Modem Cable Host Adaptor Cable UKEIB
Host Control 232 OUT Host
Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC 25-9 DPC-UKEIB Cable

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 Dr
yView Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Tomoscan M/EG rel. 3.0, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Setup:
n on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=60, Parity=None, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Kpad=ctm, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA


n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 25-pin
25-to-25 Null
Control Cable RS232 25-to-9 pin
Philips CT Tomoscan M/EG Modem Cable
Host Adaptor Cable
Host UKEIB
Host Control 232 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Tomoscan M/EG rel. 3.0, 9410 v1.6, 8100 v2.4.1

Setup:
n on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Note: if 25-to-25 Null Modem Cable is not available,


cross pin 2 and 3 on 25-side of 25-to-9 RS232 Host Adaptor Cable

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=60, Parity=None, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=ctm, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips CT Tomoscan M/EG with LASER IMAGER Products

n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

RS23-to-RS422 Kpad / Host Control


Philips 25-pin
25-to-15 pin Octopus Adaptor Cable
Control Cable
Converter Box
Crossed Kodak 160 camera
Host Control 232 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Image Data OUT Data Cable

Kodak MIM 200

Requirements: Tomoscan M/EG rel. 3.0


Requirements: MIM 200 v6.1, KELI 160 v7.2

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Tomoscan M/EG CT:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ctm,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=512, Lines=512, Parity=None, Timer=59, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips CT Tomoscan M/EG with LASER IMAGER Products

n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

RS23-to-RS422 Kpad / Host Control


Philips CT Tomoscan M/EG Philips 25-pin
25-to-15 pin Octopus Adaptor Cable
Control Cable
Converter Box
Host Control 232 OUT Kodak 2180 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Image Data OUT Data Cable

Kodak MIM 200

Requirements: Tomoscan M/EG rel. 3.0


Requirements: MIM 200 v6.1, KELP 2180 v2.17H

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Tomoscan M/EG CT:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ctm,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=512, Lines=512, Parity=None, Timer=59, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Pixel Pitch=12.6582


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

RS23-to-RS422 Kpad / Host Control


Philips CT Tomoscan M/EG Philips 25-pin
25-to-15 pin Octopus Adaptor Cable
Control Cable
Converter Box
Host Control 232 OUT Kodak 1120 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Image Data OUT Data Cable

Kodak MIM 200

Requirements: Tomoscan M/EG rel. 3.0


Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Tomoscan M/EG CT:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ctm,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=512, Lines=512, Parity=None, Timer=59, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Pixel Pitch=12.5


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
25cm=10inch Extension
Philips CT (not mandatory) Cable +
Tomoscan M/EG Kpad / Host Control
Philips 25-pin
RS23-to-RS422 Octopus Adaptor Cable iew
Control Cable 3M 87
00 DryV

25-to-15 pin
Host Control Fiber Cable
Converter Box
232 OUT
Philips 37-pin RS422 37-to-50 pin Fiber Cable
Image Data OUT Data Cable Digital Adaptor Cable

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Tomoscan M/EG rel. 3.0, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Tomoscan M/EG CT:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ct,
n Mod.Name=ct, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=512, Lines=512, Parity=None, Timer=59, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
25cm=10inch Extension
Philips CT (not mandatory) Cable +
RS422
Tomoscan M/EG Kpad / Host Control
Philips 25-pin 9-pole Plug
RS23-to-RS422 Octopus Adaptor Cable
Control Cable Control 9-pin
25-to-15 pin COPCIL Board
Host Control
Converter Box Copper Cable
232 OUT
Philips 37-pin RS422 37-to-50 pin
Image Data OUT Data Cable Digital Adaptor Cable

Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Tomoscan M/EG rel. 3.0, MIM200 v6.1, 8100 v2.4.1

Setup: on Tomoscan M/EG install Camera Config.File 3M-LASER 952(EPROM)

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Tomoscan M/EG CT:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ct,
n Mod.Name=ct, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=512, Lines=512, Parity=None, Timer=59, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Pixel Pitch=12.8205


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300

Philips CT Tomoscan M/EG

Requirements: Philips Tomoscan M/EG rel. 3.0


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

CT M/EG Setup:
- log as sa in Service Menu and go to Database Modification/Printer Device/Apply
- set Name=Kodak 8300, AE=8300, IP=accordingly, Port=104
- set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
- select in Film Formats: 8x10 1,1; 2,1; 2,2; 3,2
- press Add, press Set Current
- log as oper in Main Menu and go to Configure/Print and select 8300

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0 ie Auto
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips CT Tomoscan M/EG with LASER IMAGER Products

Philips CT Tomoscan M/EG via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips CT Tomoscan M/EG Kodak 190 camera

Requirements: Philips CT Tomoscan M/EG v4.0, 190 rel. 1.7.29

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 190, AE=190, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 190

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with|


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
DICOM Network

8700 Plus

Philips CT Tomoscan M/EG


Fiber Cable
iew
00 DryV
3M 87

Fiber Cable

9410 PACS Link

Requirements: Philips CT Tomoscan M/EG rel. 4.0, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Philips CT Tomoscan M/EG Setup:
n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8700, AE=9410, IP=accordingly, Port=1024
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8700
9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips CT Tomoscan M/EG with LASER IMAGER Products

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT Tomoscan M/EG

Data 37-pin

9410 PACS Link


DryView 8100

Requirements: Philips CT Tomoscan M/EG rel. 4.0, 9410 v1.6, 8100 v2.4.1

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8100, AE=9410, IP=accordingly, Port=1024
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8100

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=13
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips CT Tomoscan M/EG with LASER IMAGER Products

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300
Control 9 pin
Philips CT Tomoscan M/EG COPCIL board
Copper Cable

Data 37 pin

9410 PACS Link

Requirements: Tomoscan M/EG v4.0, 9410 v1.6, 8300 v 4.9.1, 8300 LUT v3.0.3B

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8300, AE=9410, IP=accordingly, Port=1024
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 8x10 1,1; 2,1; 2,2; 3,2
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8300

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips CT Tomoscan M/EG with LASER IMAGER Products

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e.,Auto
n on Operator Menu set: Film Layout: 1:1 Portrait, 2:1P, 4:1P, 6:1P
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips CT Tomoscan M/EG


Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM 200 v6.1, KELI 160 v7.2

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n pick up e.g., 190, set Name=Kodak 160, AE=160, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 160
n to print select an image and go to Data Transfer, Manual Print, Image, Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips CT Tomoscan M/EG


Kodak 2180 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips CT Tomoscan M/EG rel. 4.0


Requirements: MIM200 v6.1, 2180 rel. 2.17H

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 2180, AE=2180, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 2180

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips CT Tomoscan M/EG


Kodak 1120 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 1120, AE=1120, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 1120

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips CT Tomoscan M/EG with LASER IMAGER Products

CT Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips CT Tomoscan M/EG


Crossed
Network Cable

Kodak MIM 200


Kodak DryView 8200

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8200, AE=8200, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8200
n to print select an image and go to Data Transfer, Manual Print, Image, Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips CT Tomoscan M/EG


Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM200 v6.1, 8700 v1.34
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8700, AE=8700, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8700

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 46


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with|


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips CT Tomoscan M/EG

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8100, AE=8100, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8100

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 47


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=13 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 48


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips CT COPCIL board
Tomoscan M/EG Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

Philips CT Tomoscan M/EG Setup:


n log as sa in Service Menu and go to Database Modification/Printer Device/Apply
n set Name=Kodak 8300, AE=8300, IP=accordingly, Port=5040
n set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
n select in Film Formats: 8x10 1,1; 2,1; 2,2; 3,2
n press Add, press Set Current
n log as oper in Main Menu and go to Configure/Print and select 8300

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 49


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.811
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e.,Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 50


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips CT
Tomoscan M/EG

Kodak DryView 8900

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Philips CT Tomoscan M/EG Setup:


- log as sa in Service Menu and go to Database Modification/Printer Device/Apply
- set Name=Kodak 8900, AE= 8900, IP=accordingly, Port=5040
- set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
- select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
- press Add, press Set Current
- log as oper in Main Menu and go to Configure/Print and select 8900

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4

Author: gian.valesano@kodak.com, Edition: May ´05 51


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Min Vert Sep=4, Trim Width Enabled=4


n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=13 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ctm

Author: gian.valesano@kodak.com, Edition: May ´05 52


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips CT
Tomoscan M/EG

Kodak DryView 8150

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Philips CT Tomoscan M/EG Setup:


- log as sa in Service Menu and go to Database Modification/Printer Device/Apply
- set Name=Kodak 8150, AE=DV8150, IP=accordingly, Port=5040
- set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
- select in Film Formats: 14x17 1,1; 2,1; 2,2; 3,2; 3,3; 4,3; 5,3; 4,4; 5,4
- press Add, press Set Current
- log as oper in Main Menu and go to Configure/Print and select 8150

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2

Author: gian.valesano@kodak.com, Edition: May ´05 53


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10


n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=13 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ct

Author: gian.valesano@kodak.com, Edition: May ´05 54


Philips CT Tomoscan M/EG with LASER IMAGER Products

Tomoscan M/EG via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips CT
Tomoscan M/EG

Kodak CMI 1000

Requirements: Philips CT Tomoscan M/EG v4.0


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Philips CT Tomoscan M/EG Setup:


- log as sa in Service Menu and go to Database Modification/Printer Device/Apply
- set Name=Kodak CMI1000, AE= CMI1000, IP=accordingly, Port=5040
- set Medium=Blue Film, Orient.=Portrait, Interpolation=Cubic, Density=300
- select in Film Formats: 8x10 1,1; 2,1; 2,2; 3,2
- press Add, press Set Current
- log as oper in Main Menu and go to Configure/Print and select CMI1000

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=13 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 55


Philips CT Tomoscan M/EG with LASER IMAGER Products

n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No


n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 56


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Specific Setup for Philips DR Digital Diagnost


with LASER IMAGERS

OEM Data
Manufacturer Philips
Modality DR
Model Digital Diagnost
sw version v1.4
Implement. Class UID 1.3.46.670589.26.1.4
Implement. Version Name DigiDiagnost1.4

Kodak Data
DICOM-interface MIM internal / external or 9410 as below reported
DICOM-intf sw version as below reported
Printer model All cameras as below reported
Printer sw version as below reported

What’s new:
The setup of a Kodak printer for a given OEM Modality was split in 2 separate documents:

- the generic Setup Catalog for Kodak Cameras


- the specific Setup Manual for a given OEM Modality

************* the generic Setup Catalog for Kodak Cameras:


the Setup Catalog for Kodak Cameras contains that part of a printer installation which is Generic, ie common to
all DICOM connections independently from the Specific modality connected.

Therefore, as first step, setup the printer as per the correspondent page of the Setup Catalog for Kodak Cameras
(ie go to the Setup Catalog doc and do that first, then come back
here and finish install as per this Specific Setup Manual doc).

************ the specific Setup Manual for a given OEM Modality:


the Setup Manual for a given OEM Modality contains that part of a printer installation which is Specific to that
particular OEM Modality to be connected, ie the OEM console Settings and the printer Image Quality
parameters.

Therefore, as second and final step, setup the printer as per the indications reported here into the Setup Manual
for that particular OEM Modality.
Note that the Setup Manual has 2 sections ie the DryView section and the Wet section.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

What’s inside this document:


n OEM Modality Setup (ie how to configure,into the OEM console,the Kodak printer) 2
n DryView correspondance with Kodak Setup Catalog for DryView cameras................ 3
n Image Quality setup into MIM (or 9410) for DryView cameras ............................... 4
n More info on OEM Modality when connected to DryView cameras .............................. 5
n Wet correspondance with Kodak Setup Catalog for Wet cameras .................................. 6
n Image Quality setup into MIM (or 190) for Wet cameras ............................................... 7
n More info on OEM Modality eg Page-Mode and IQ parameters sent out by OEM ........ 8
n Miscellaneous: Deviations from Kodak Setup Catalog / Known Problems / Notes........ 9

OEM Modality Setup


ie how to configure into the OEM console the Kodak printer:

Note: MIM setup is basically also valid for 9410 by using AE=9410, Port=1024
Note: for each DryView camera select into the Modality the correspondent Prototype,
in particular for 8900 choose the Kodak 8900 standard
(or 3M / Kodak 8900 and in this case 2mm Border=No, or 3M 8200),
in particular for CMI1000 choose the Kodak CMI1000 or Codonics 1660MD.

Digital Diagnost Setup:


n into HCU/Configuration select Printers/Add and set:
n Name=DryView, Model=8150, Manufacturer=Kodak,
n Description=DryView, Protocol=DICOM,
n from list of Supported Prototypes choose Kodak-DryView-8150-DICOM (or 8200)
n press Set-button (the one with an arrow bringing left column into right column)
n Prototype= Kodak-DryView-8150-DICOM (or 8200)
n AET=DV8150, Hostname=DV8150, IP=accordingly, Port=5040
n Remove Media not used, eg leave 14x17 blue film P/L 8 bit Monochrome2
n Click OK/Apply/Save, click Return, click Restart=Confirm

Author: gian.valesano@kodak.com, Edition: May ´05 2


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

DryView section:
Setup for Kodak DryView cameras ie
8900, 8700, 8610, 8600, 8500, 8300, 8200, 8150, 8100, 1000

Correspondance with Kodak Setup Catalog for DryView cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

MIM (inside printer) 8900 8150


into Kodak catalog 19 20
see Setup as per page
MIM sw version 5.6.b5 5.7.b8
Printer sw version Rev.E ie 2.11 1.0.2.b04

MIM external 8700, 8500 8100 8300 8610 (8600) 8200 1000
into Kodak catalog 15 16 17 18 14 21
see Setup as per page
MIM sw version 6.1 6.1 6.1 6.1 6.1 2.0.b3
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5) 4.9.1 1.0.b8

9410 external 8700, 8500 8100 8300 8610 (8600)


into Kodak catalog 7 8 9 10
see Setup as per page
9410 sw version 1.6 1.6 1.6 1.6
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5)

Author: gian.valesano@kodak.com, Edition: May ´05 3


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 9410) for DryView cameras:

MIM (or 9410) 8900, 8150, 8700, 8300 8610, 8600


Parameters 8500, 8200, 8100, (see Note2 below,
1000 confirm Y/N)
(see Note1 below)
TFT/ULUT Pmsv3k.w87 u1_109s.w8u (gsdfaa0p.w8u)
(8610 Modality/Contrast Group) 6
(Always Apply=Y/N) (Y) (Y)
Contrast 9 10 3
(Always Apply=Y/N) (Y) (Y) (Y)
Max.Density 300 300 350
(Always Apply=Y/N) (Y) (Y) (Y)
Smoothing 0 1 1
(Always Apply=Y/N) (Y) (Y) (Y)
Presentation ie PLUT=Y/N N N N
(not available on 9410 )
8610 Convolution / Kernel Group Not available Not available 1
8610 Gamma Table Not available Not available Transmittance

Note1: in particular for 8100, 8200, 8150 use TFT=Pmsb.w87.

Note2: OEM Modality tested also with 8300 standalone Yes


ie without MIM or 9410 in front? (Yes/No, Mod.Gr.=0, D=16, C=10, smooth=1
parameters)

Author: gian.valesano@kodak.com, Edition: May ´05 4


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to DryView cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg TFT=Wrkstn2a.w87, Dmax=3.00, C=2.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for Value sent out Value sent out
Parameters Modality 8900, 8150, 8700, for 8300 for 8610, 8600
(Yes/No) 8500, 8200, 8100,
1000
Config. Information Yes LUT=0,9 LUT=0,10 LUT=0,3
(LUT=TFT,Contrast)
Max. Density Yes 300 300 300
Magnification Yes NONE NONE NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 5


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Wet section:
Setup for Kodak Wet cameras ie 160, 2180, 1120, 190

Correspondance with Kodak Setup Catalog for Wet cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

160 2180 1120 190


into Kodak catalog 11 12 13 6
see Setup as per page
MIM sw version 6.1 6.1 6.1 Not present
Printer sw version 7.2 2.17H 2.40 / 032 1.7.29

Note: in regards to 969 parameters use the same as per 8700 setup.

Author: gian.valesano@kodak.com, Edition: May ´05 6


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 190) for Wet cameras:

MIM (or 190) Parameters 160, 2180, 1120, 190


(see Note below)
Curve Shape 3
(Always Apply=Y/N) (Y)
Contrast 0
(Always Apply=Y/N) (Y)
Max.Density 300
(Always Apply=Y/N) (Y)
PLUT=Y/N N
(not available on 190 )

Note: reported setup was tested eg for 160 and tentatively is valid also for other cameras ie
2180, 1120, 190.

Author: gian.valesano@kodak.com, Edition: May ´05 7


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to Wet cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg CurveShape=0,Dmax=3.00,Contrast=0.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for
Parameters Modality 160, 2180, 1120,
(Yes/No) 190
Config. Information Yes CS435\CN0
(CurveShape\Contrast)
Max. Density Yes 300
Magnification Yes NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 8


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Miscellaneous:

Deviations in respect to the printer setup reported into Kodak Catalog:

e.g. disable N-Event ie NeventReports=No (Always=Y)

Known Problems or Limitations:

Notes:

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Setup Philips DR Digital Diagnost


via Merge DICOM-converter (ie older approach)
with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 4
n Kodak DryView 8700 LASER IMAGER / Standard ....................................................... 5
n Kodak Digital Science 969 HQ LASER IMAGER...................................................... 6
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 7
n Kodak DryView 8100 LASER IMAGER......................................................................... 8
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 9
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 11
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 13
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 15
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) . 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 21

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Connection via DICOM:

n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 23


n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 24
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 26
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 28
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 30
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 32
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 34
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 36
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5) ................................................................. 38
n Kodak DryView 8900 LASER IMAGER......................................................................... 40
n Kodak DryView 8150 LASER IMAGER......................................................................... 42
n Kodak Color Medical Imager 1000 ................................................................................. 44

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
anag
tM
Inpu
ulti-
0M
880
3M

iew
00 DryV
3M 87

Fiber Cable
8700 Dual

8800 MIM

Requirements: Digital Diagnost rel. 1.1, 8800 rel. 5.1, 8700 rel. 1.22

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700/8800 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700/8800

Diagnost Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Frame=No,Pxl Corr.=No,Scale Format=No,Scale Magnif.=No,Slides=Super
n Override 952 Beta=No (interpolation curve=11 sent from modality)
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

iew
00 DryV
3M 87

8700 Plus

Requirements: Digital Diagnost rel.1.1, 8700 rel. 1.22

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Framing=No, Pixel Correct.=No, Scale Format=No, Scale Magnif.=No
n Override 952 Beta=No (interpolation curve=11 sent from modality)
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
Image Data OUT

Requirements: Digital Diagnost rel.1.1, 8700 rel. 1.22

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Framing=No, Pixel Correct.=No, Scale Format=No, Scale Magnif.=No
n Override 952 Beta=No (interpolation curve=11 sent from modality)
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak Digital Science 969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M H Q L aser Im ager

M969 HQ

Requirements: Digital Diagnost rel. 1.1, M969 HQ rel. 5.1

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 969 is defined
- ItemName=SiteInformation set Preferred Printer=3M 969

Diagnost Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Frame=No,Pxl Corr.=No,Scale Format=No,Scale Magnif.=No,Slides=Super
n Override 952 Beta=No (interpolation curve=11 sent from modality)
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422 OUT 3 M H Q L aser Im a ger

Philips 37-pin
Data Cable
Image Data OUT

Requirements: Digital Diagnost rel.1.1, M969 HQT rel. 1.0

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 969 is defined
- ItemName=SiteInformation set Preferred Printer=3M 969

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu:
n Framing=No, Pixel Correct.=No, Scale Format=No, Scale Magnif.=No
n Override 952 Beta=No (interpolation curve=11 sent from modality)
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak DryView 8100 LASER IMAGER

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

DryView 8100

Requirements: Digital Diagnost rel.1.1, 8100 rel. 1.1

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips Digital Diagnost Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 DryV
iew Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Digital Diagnost v1.1, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
and if HQ by installing HQ into Digital Diagnost Camera Config.File

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=digdiag, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips Digital Diagnost Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Digital Diagnost rel.1.1, 9410 v1.6, 8100 v2.4.1

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=60, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=digdiag, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On


n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips Digital Diagnost Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Digital Diagnost v1.1


Requirements: MIM 200 v6.1, 160 v7.2

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak 160 (or 2180) is defined
- ItemName=SiteInformation set Preferred Printer=Kodak 160 (or 2180)

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Direct Connect, Config.Add, Digital, Unqualified:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dd,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=4.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips Digital Diagnost Control Adaptor Cable
Control Cable
Kodak 2180 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Digital Diagnost v1.1


Requirements: MIM 200 v6.1, 2180 v2.17H

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak 2180 is defined
- ItemName=SiteInformation set Preferred Printer=Kodak 2180

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Direct Connect, Config.Add, Digital, Unqualified:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dd,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=4.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips Digital Diagnost Control Adaptor Cable
Control Cable
Kodak 1120 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Digital Diagnost v1.1


Requirements: MIM 200 v6.1, 1120 v2.40/032

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak 1120 (or 2180) is defined
- ItemName=SiteInformation set Preferred Printer=Kodak 1120 (or 2180)

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Direct Connect, Config.Add, Digital, Unqualified:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dd,
n Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=4.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips Digital Diagnost Control Adaptor Cable
Control Cable 00 DryV
iew
3M 87

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Digital Diagnost v1.1, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination
and if HQ by installing HQ into Digital Diagnost Camera Config.File

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Unqualified:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dd,
n Mod.Name=dd, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips Digital Diagnost Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Digital Diagnost v1.1, MIM200 v6.1, 8100 v2.4.1

Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 8700 is defined
- ItemName=SiteInformation set Preferred Printer=3M 8700

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Unqualified:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dd,
n Mod.Name=dd, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Philips Digital Diagnost via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER
DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Kodak 190 camera
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board

Requirements: Digital Diagnost v1.1, Merge DPI v2.4, 190 rel. 1.7.29
Philips Digital Diagnost Setup:
n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI
Merge DPI Setup:
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 190=NER_ANY
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=190, ADDRESS=NER_ANY, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_ANY] set:
- PORT_NUMBER=5040, HOST_NAME=NER_ANY
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL
190 Setup: On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK 9410ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable Fiber Cable 3M 87
00 Dr
yView

Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
9410 PACS Link
DryView 8700 Plus

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Philips Digital Diagnost Setup:
n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI
Merge DPI Setup:
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=MERGE_XPI1, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
9410 PACS Link DryView 8100

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: 9410 v1.6, 8100 v2.4.1

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=MERGE_XPI1, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=Off, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 160 camera
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Digital Diagnost v1.1, Merge DPI v2.4


Requirements: MIM 200 v6.1, KELI 160 v7.2

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_160
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_160, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_160] set:
- PORT_NUMBER=5040, HOST_NAME=NER_160
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=4 i.e.,575 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 2180 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: MIM200 v6.1, 2180 v2.17H

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_2180
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_2180, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_2180] set:
- PORT_NUMBER=5040, HOST_NAME=NER_2180
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=4 i.e.,575 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 1120 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Digital Diagnost v1.1, Merge DPI v2.4


Requirements: MIM200 v6.1, 1120 v2.40/032

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_1120
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_1120, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_1120] set:
- PORT_NUMBER=5040, HOST_NAME=NER_1120
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=4 i.e.,575 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8200

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4, MIM200 v6.1, 8200 v3.5
Note: This setup is basically also valid for 25 PS computer

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8200
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_8200, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8200] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8200
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n on Configure, Input, Network, Add:


n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off
8200 Setup on CONFIGURATION Menu:
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable Fiber Cable 3M 870
0 DryV
iew

Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8700 Plus

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM200=NER_8700
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_8700, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8700] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8700
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

MIM200 Setup:
n on Configure, Network
n Node=MIM, IPAddr=accordingly, Subnet=accordingly, Gateway=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On, Landscape=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Digital Diagnost Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200 Kodak DryView 8100

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4, MIM200 v6.1, 8100 v2.4.1
Note: This setup is basically also valid for 25 PS computer

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM200=NER_8100
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_8100, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8100] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8100
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:


n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board

Kodak DryView 8900

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8900= 8900
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8900, ADDRESS=8900, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8900] set:
- PORT_NUMBER=5040, HOST_NAME=8900
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,9 , Polarity=NORMAL

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Subnet Mask=accordingly, Gateway Addr=accordingly


n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=dd

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8150

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8150= 8150
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8150, ADDRESS=8150, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8150] set:
- PORT_NUMBER=5040, HOST_NAME=8150
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,9 , Polarity=NORMAL

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

n Subnet Mask=accordingly, Gateway Addr=accordingly


n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=dd

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

Digital Diagnost via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Kodak CMI 1000


Data Cable
Image Data OUT Merge Data Board

Requirements: Philips Digital Diagnost v1.1, Merge DPI rel. 2.4


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Philips Digital Diagnost Setup:


n on Digital Diagnost into System/Service/Configuration:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify DPI is defined
- ItemName=SiteInformation set Preferred Printer=DPI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of CMI1000= CMI1000
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME= CMI1000, ADDRESS= CMI1000,
- PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [CMI1000] set:
- PORT_NUMBER=5040, HOST_NAME= CMI1000
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information= LUT=0,9 , Polarity=NORMAL

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips DR Digital Diagnost via Merge-box with LASER IMAGER Products

(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):


at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Setup Philips DSA models Allura


H/V2000, H/V3000, H/V4000, H/V5000
with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 4
n Kodak DryView 8700 LASER IMAGER / Standard ....................................................... 5
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual.............................................................. 6
n Kodak DryView 8500 LASER IMAGER / Plus............................................................... 7
n Kodak DryView 8500 LASER IMAGER / Standard ....................................................... 8
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual.............................................................. 9
n Kodak DryView 8300 LASER IMAGER / Standalone.................................................... 10
n Kodak Digital Science 969 HQ LASER IMAGER...................................................... 11
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 12
n Kodak Digital Science 959 XL LASER IMAGER ...................................................... 13
n Kodak DryView 8100 LASER IMAGER......................................................................... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 15
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 21
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) . 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 27

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 29


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 30
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 31
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 33
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 35
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 37
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 39
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 41
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 43
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 45
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5) ................................................................. 47
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4,5) ................................................................. 49
n Kodak DryView 8900 LASER IMAGER......................................................................... 51
n Kodak DryView 8150 LASER IMAGER......................................................................... 53
n Kodak Color Medical Imager 1000 ................................................................................. 55

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
Mana
nput
ulti-I
0M
880
3M

yView
00 Dr
3M 87

Fiber Cable
8700 Dual

8800 MIM

Requirements: V3000 rel. 4.3.3, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8700 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8700 Plus

Requirements: Philips V3000 rel. 4.3.3, 8700 Plus rel. 1.15

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=First Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: Philips V3000 rel. 4.3.3, 8700 Standard rel. 1.15

Setup: on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:

n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=First Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
Mana
nput
ulti-I
0M
880
3M

yView
00 Dr
3M 87

Fiber Cable
8500 Dual

8800 MIM

Requirements: V3000 rel. 4.3.3, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8500 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8500 Plus

Requirements: Philips V3000 rel. 4.3.3, 8500 Plus rel. 1.22

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=First Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: V3000 rel. 4.3.3, 8500 Standard rel. 1.22

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=First Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual
Philips 37-pin RS422 37-to-9 pin
Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
nag
ut Ma
lti-Inp
00 Mu

Fiber Cable 3M88

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: V3000 rel. 4.3.3, 8800 rel. 5.1, 8300 rel. 4.2
Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout: Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0
8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips V3000
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


I mage Data OUT

Requirements: Philips V3000 rel. 4.3.3, 8300 rel. 4.2

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout: Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Odd
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak Digital Science 969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M H Q L a ser Im ag er

M969 HQ

Requirements: Philips V3000 rel. 4.3.3, M969 HQ rel. 3.0

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=15, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT 3 M H Q L a s er Im a g e r

Philips 37-pin
Data Cable
I mage Data OUT

Requirements: Philips V3000 rel. 4.3.3, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=First Fit
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak Digital Science 959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a s e r I m a g er

M959 XL

Requirements: Philips V3000 rel. 4.3.3, M959 XL rel. 3.8

Setup:
n on Philips V3000 rel. 4.3.3 install Camera Config.File 3M M952
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu: 959 Default
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak DryView 8100 LASER IMAGER

DryView 8100

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips V3000 Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

Requirements: V3000 rel. 4.3.3, 8100 v1.1

Setup: on V3000 rel. 4.3.3 install Camera Config.File 3M M952

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=10, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=15, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips V3000 Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 DryV
iew Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: V3000 rel. 4.3.3 , 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=v3000, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=15, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips V3000 Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: V3000 rel. 4.3.3, 9410 v1.6, 8100 v2.4.1

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
on Device Setup, HOST KEYPADS:
n Kpad=v3000, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=15, D=14

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips V3000 Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Integris V3000 rel. 7.2.1


Requirements: MIM 200 v6.1, KELI 160 v7.2

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=Kodak Ektascan 100, HCU=Laser
n cycle power V3000

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Integris V/H DF:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=v3000,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=960, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips V3000 Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 2180 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Integris V3000 rel. 7.2.1


Requirements: MIM 200 v6.1, KELP 2180 v2.17H

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=Kodak Ektascan 100, HCU=Laser
n cycle power V3000

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Integris V/H DF:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=v3000,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=960, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips V3000 Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 1120 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Integris V3000 rel. 7.2.1


Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=Kodak Ektascan 100, HCU=Laser
n cycle power V3000

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Integris V/H DF:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=v3000,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=960, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips V3000 Control Adaptor Cable
Control Cable 00 DryV
iew
3M 87

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Integris V3000 rel. 7.2.1, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM 200 Destinations
Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Integris V/H DF:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=v3000,
n Mod.Name=v3000, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=960, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips V3000 Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Integris V3000 rel. 7.2.1, MIM200 v6.1, 8100 v2.4.1

Note: same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Integris V/H DF:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=v3000,
n Mod.Name=v3000, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=960, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8300

Requirements: Philips V3000 rel. 4.3.3, Merge DPI rel. 2.2


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT rel. 3.0.3B

Note: same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8300=IMN_8300
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=8300, ADDRESS=IMN_8300, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_8300] set:
- PORT_NUMBER=104, HOST_NAME=IMN_8300
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=0
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
8300 Setup:
n on Operator Menu set: D=16, C=10, Sharp/Smooth=0 i.e.,Auto
n on Operator Menu set: Film Layout: 1 on 1=Landscape, 2:1L, 4:1L
n Note: due to V3000 matrix 1024x1280, 4:1 allowed only with Film Layout: 4:1L
n on 8300 Service Menu into Configure User set: Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Kodak 190 camera
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board

Requirements: Philips V3000 rel. 4.3.3, Merge DPI v2.4, 190 rel. 1.7.29

Philips V3000 Setup: on V3000 install Camera Config.File 3M M952

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 190=NER_ANY
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=190, ADDRESS=NER_ANY, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_ANY] set:
- PORT_NUMBER=5040, HOST_NAME=NER_ANY
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

V3000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable Fiber Cable 3M 87
00 Dr
yView

Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
9410 PACS Link
DryView 8700 Plus

Requirements: Philips V3000 rel. 4.3.3, Merge DPI rel. 2.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000
V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000
Merge DPI Setup:
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,15 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=15, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

V3000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips 37-pin RS422 37-to-25 pin Merge DPI
Philips V3000 Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
9410 PACS Link DryView 8100

Requirements: Philips V3000 rel. 4.3.3, Merge DPI rel. 2.2, 9410 v1.6, 8100 v2.4.1

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,15 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=15, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

V3000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER
DICOM Network

Control 9-pin
COPCIL Board
Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board Kodak DryView 8300
Kodak 9410 PACS Link

Requirements: Philips V3000 rel. 4.3.3, Merge DPI rel. 2.2


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=0
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type =831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e.,Auto
n on Operator Menu set: Film Layout: 1 on 1=Landscape, 2:1L, 4:1L
n Note: due to V3000 matrix 1024x1280, 4:1 allowed only with Film Layout: 4:1L
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 160 camera
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: MIM 200 v6.1, KELI 160 v7.2

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=ANY
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=ANY, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [ANY] set:
- PORT_NUMBER=5040, HOST_NAME=ANY
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
DICOM Network

Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 2180 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips V3000 rel. 4.3.3, Merge DPI v2.4


Requirements: MIM200 v6.1, 2180 v2.17H

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_2180
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_2180, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_2180] set:
- PORT_NUMBER=5040, HOST_NAME=NER_2180
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
DICOM Network

Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 1120 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: MIM200 v6.1, 1120 v2.40/032

Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_1120
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_1120, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_1120] set:
- PORT_NUMBER=5040, HOST_NAME=NER_1120
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e.,435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8200

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: MIM200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer


Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8200
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8200,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8200] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8200
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=LUT=0,9, Polarity=NORMAL
MIM200 Setup:
n on Configure, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable Fiber Cable 3M 87
00 Dr
yView

Control 422OUT Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8700 Plus

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note:This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination
Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000
V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000
Merge DPI Setup:
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8700
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8700,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8700] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8700
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=LUT=0,9, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 46


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On, Landscape=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 47


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips 37-pin RS422 37-to-25 pin Merge DPI
Philips V3000 Copper Cable
Control Cable Merge Adaptor Cable
Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200 Kodak DryView 8100

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer


Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8100
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8100,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8100] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8100
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 48


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e.,external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 49


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER
DICOM Network

Control 9-pin
COPCIL Board
Philips V3000 Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board Kodak DryView 8300
Kodak MIM 200

Requirements: Philips V3000 rel. 4.3.3, Merge DPI rel. 2.2


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer


Note: Same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
n on service terminal press shift-s during booting to enter the Service Mode
n on Service software select 2 (Servicesoftware) / Run / Config / Hardware
n set: Remote control HCU=3M M952, HCU=Laser
n cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8300
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8300,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8300] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8300
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=2
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 50


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off
8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e.,Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 51


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8900

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Note: same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
- on service terminal press shift-s during booting to enter the Service Mode
- on Service software select 2 (Servicesoftware) / Run / Config / Hardware
- set: Remote control HCU=3M M952, HCU=Laser
- cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8900= 8900
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8900,ADDRESS= 8900,PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8900] set:
- PORT_NUMBER=5040, HOST_NAME= 8900
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8900 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 52


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=11 (Always=N), Image Polarity=Normal (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=v3000

Author: gian.valesano@kodak.com, Edition: May ´05 53


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips V3000
Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8150

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Note: same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
- on service terminal press shift-s during booting to enter the Service Mode
- on Service software select 2 (Servicesoftware) / Run / Config / Hardware
- set: Remote control HCU=3M M952, HCU=Laser
- cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8150= 8150
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8150,ADDRESS= 8150,PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8150] set:
- PORT_NUMBER=5040, HOST_NAME= 8150
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8150 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 54


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=v3000

Author: gian.valesano@kodak.com, Edition: May ´05 55


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products

Philips V3000 via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Philips 37-pin RS422 37-to-25 pin Merge DPI


Digital Diagnost Control Cable Merge Adaptor Cable
Merge Comm Board

Philips 37-pin Kodak CMI 1000


Data Cable
Image Data OUT Merge Data Board

Requirements: Philips V3000 v4.3.3, Merge DPI v2.4


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Note: same setup valid for Integris H/V2000, H/V3000, H/V4000, H/V5000

V3000 Setup:
- on service terminal press shift-s during booting to enter the Service Mode
- on Service software select 2 (Servicesoftware) / Run / Config / Hardware
- set: Remote control HCU=3M M952, HCU=Laser
- cycle power V3000

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of CMI1000=CMI1000
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=CMI1000,ADDRESS=CMI1000,
- PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [CMI1000] set:
- PORT_NUMBER=5040, HOST_NAME= CMI1000
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=8INX10IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 56


Philips DSA models Allura H/V2000, H/V3000, H/V4000, H/V5000 with
LASER IMAGER Products
CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 57


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Specific Setup for Philips DSA Allura Xper


with LASER IMAGERS

OEM Data
Manufacturer Philips
Modality DSA
Model Allura Xper
sw version FD20
Implement. Class UID 1.3.46.670589.7.28.2.0.1
Implement. Version Name AlluraXper20ACP

Kodak Data
DICOM-interface MIM internal / external or 9410 as below reported
DICOM-intf sw version as below reported
Printer model All cameras as below reported
Printer sw version as below reported

What’s new:
The setup of a Kodak printer for a given OEM Modality was split in 2 separate documents:

- the generic Setup Catalog for Kodak Cameras


- the specific Setup Manual for a given OEM Modality

************* the generic Setup Catalog for Kodak Cameras:


the Setup Catalog for Kodak Cameras contains that part of a printer installation which is Generic, ie common to
all DICOM connections independently from the Specific modality connected.

Therefore, as first step, setup the printer as per the correspondent page of the Setup Catalog for Kodak Cameras
(ie go to the Setup Catalog doc and do that first, then come back
here and finish install as per this Specific Setup Manual doc).

************ the specific Setup Manual for a given OEM Modality:


the Setup Manual for a given OEM Modality contains that part of a printer installation which is Specific to that
particular OEM Modality to be connected, ie the OEM console Settings and the printer Image Quality
parameters.

Therefore, as second and final step, setup the printer as per the indications reported here into the Setup Manual
for that particular OEM Modality.
Note that the Setup Manual has 2 sections ie the DryView section and the Wet section.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

What’s inside this document:


n OEM Modality Setup (ie how to configure,into the OEM console,the Kodak printer) 2
n DryView correspondance with Kodak Setup Catalog for DryView cameras................ 3
n Image Quality setup into MIM (or 9410) for DryView cameras ............................... 4
n More info on OEM Modality when connected to DryView cameras .............................. 5
n Wet correspondance with Kodak Setup Catalog for Wet cameras .................................. 6
n Image Quality setup into MIM (or 190) for Wet cameras ............................................... 7
n More info on OEM Modality eg Page-Mode and IQ parameters sent out by OEM ........ 8
n Miscellaneous: Deviations from Kodak Setup Catalog / Known Problems / Notes........ 9

OEM Modality Setup


ie how to configure into the OEM console the Kodak printer:

Note: MIM setup is basically also valid for 9410 by using AE=9410, Port=1024
Note: for each DryView camera select into the Modality the correspondent Prototype,
in particular for 8900 choose the Kodak 8900 standard,
in particular for CMI1000 choose the Kodak CMI1000 or Codonics 1660MD.

Allura Xper Setup:


n press Windows Startmenu key
n enter into Service Functions/Yes/Service Login/Yes/click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Configuration/Networking/ACP/DICOM Printer
n select Available Templates=Kodak-DryView-8150 (or 8200)-DICOM, click Add
n set Name=DryView, AE=DV8150, Port=5040, IP HostName=DV8150, IP=accordingly,
n into Print Medium Types, move into the Unconfigured list:
n all filmsizes with C (ie Clear Base) both 8 and 12 (ie bit)
n all filmsizes with B (ie Blue Base) and 12 (ie bit)
n all filmsizes with B (ie Blue Base) and 8 (ie bit) that will not be used eg 14x14B_8
n into Advanced Settings leave defaults ArtimTimeout=60, Autom.Assoc.Timeout=0
n click Apply/OK/Yes/OK/OK/OK/Up/Logoff/Yes/OK/OK/OK/Next/OK/Exit
n press Windows Startmenu key and ShutDown/Restart the computer

Author: gian.valesano@kodak.com, Edition: May ´05 2


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

DryView section:
Setup for Kodak DryView cameras ie
8900, 8700, 8610, 8600, 8500, 8300, 8200, 8150, 8100, 1000

Correspondance with Kodak Setup Catalog for DryView cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

MIM (inside printer) 8900 8150


into Kodak catalog 19 20
see Setup as per page
MIM sw version 5.6.b5 5.7.b8
Printer sw version Rev.E ie 2.11 1.0.2.b04

MIM external 8700, 8500 8100 8300 8610 (8600) 8200 1000
into Kodak catalog 15 16 17 18 14 21
see Setup as per page
MIM sw version 6.1 6.1 6.1 6.1 6.1 2.0.b3
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5) 4.9.1 1.0.b8

9410 external 8700, 8500 8100 8300 8610 (8600)


into Kodak catalog 7 8 9 10
see Setup as per page
9410 sw version 1.6 1.6 1.6 1.6
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5)

Author: gian.valesano@kodak.com, Edition: May ´05 3


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 9410) for DryView cameras:

MIM (or 9410) 8900, 8150, 8700, 8300 8610, 8600


Parameters 8500, 8200, 8100, (see Note2 below,
1000 confirm Y/N)
(see Note1 below)
TFT/ULUT Pmsv3k.w87 u1_109s.w8u (gsdfaa0p.w8u)
(8610 Modality/Contrast Group) 6
(Always Apply=Y/N) (Y) (Y)
Contrast 9 10 3
(Always Apply=Y/N) (Y) (Y) (Y)
Max.Density 300 300 350
(Always Apply=Y/N) (Y) (Y) (Y)
Smoothing 0 1 1
(Always Apply=Y/N) (Y) (Y) (Y)
Presentation ie PLUT=Y/N N N N
(not available on 9410 )
8610 Convolution / Kernel Group Not available Not available 1
8610 Gamma Table Not available Not available Transmittance

Note1: in particular for 8100, 8200, 8150 use TFT=Pmsb.w87.

Note2: OEM Modality tested also with 8300 standalone Yes


ie without MIM or 9410 in front? (Yes/No, Mod.Gr.=0, D=16, C=10, smooth=1
parameters)

Author: gian.valesano@kodak.com, Edition: May ´05 4


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to DryView cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg TFT=Wrkstn2a.w87, Dmax=3.00, C=2.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for Value sent out Value sent out
Parameters Modality 8900, 8150, 8700, for 8300 for 8610, 8600
(Yes/No) 8500, 8200, 8100,
1000
Config. Information Yes LUT=0,9 LUT=0,10 LUT=0,3
(LUT=TFT,Contrast)
Max. Density Yes 300 300 300
Magnification Yes NONE NONE NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 5


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Wet section:
Setup for Kodak Wet cameras ie 160, 2180, 1120, 190

Correspondance with Kodak Setup Catalog for Wet cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

160 2180 1120 190


into Kodak catalog 11 12 13 6
see Setup as per page
MIM sw version 6.1 6.1 6.1 Not present
Printer sw version 7.2 2.17H 2.40 / 032 1.7.29

Note: in regards to 969 parameters use the same as per 8700 setup.

Author: gian.valesano@kodak.com, Edition: May ´05 6


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 190) for Wet cameras:

MIM (or 190) Parameters 160, 2180, 1120, 190


(see Note below)
Curve Shape 3
(Always Apply=Y/N) (Y)
Contrast 0
(Always Apply=Y/N) (Y)
Max.Density 300
(Always Apply=Y/N) (Y)
PLUT=Y/N N
(not available on 190 )

Note: reported setup was tested eg for 160 and tentatively is valid also for other cameras ie
2180, 1120, 190.

Author: gian.valesano@kodak.com, Edition: May ´05 7


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to Wet cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg CurveShape=0,Dmax=3.00,Contrast=0.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for
Parameters Modality 160, 2180, 1120,
(Yes/No) 190
Config. Information Yes CS435\CN0
(CurveShape\Contrast)
Max. Density Yes 300
Magnification Yes NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 8


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Miscellaneous:

Deviations in respect to the printer setup reported into Kodak Catalog:

e.g. disable N-Event ie NeventReports=No (Always=Y)

Known Problems or Limitations:

Notes:

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips DSA model DSI with LASER IMAGER Products

Setup for Philips DSA model DSI with LASER IMAGERS

(note: models EasyDiagnost, MultiDiagnost, OmniDiagnost, Eleva could


have either the setup of Philips DSI or the setup of Philips EasyVision)

Connection POINT-TO-POINT:
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 4
n Kodak DryView 8700 LASER IMAGER / Standard ....................................................... 5
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual.............................................................. 6
n Kodak DryView 8500 LASER IMAER / Plus ................................................................. 7
n Kodak DryView 8500 LASER IMAGER / Standard ....................................................... 8
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual.............................................................. 9
n Kodak DryView 8300 LASER IMAGER / Standalone.................................................... 10
n Kodak Digital Science 969 HQ LASER IMAGER...................................................... 11
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 12
n Kodak Digital Science 959 XL LASER IMAGER ...................................................... 13
n Kodak DryView 8100 LASER IMAGER......................................................................... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 15
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 21
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) . 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 27

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips DSA model DSI with LASER IMAGER Products

Connection via DICOM:


n Kodak DryView 8300 LASER IMAGER......................................................................... 29
n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 30
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 31
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 33
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 35
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 37
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 39
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 41
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 43
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 45
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5) ................................................................. 47
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4,5) ................................................................. 49
n Kodak DryView 8900 LASER IMAGER......................................................................... 51
n Kodak DryView 8150 LASER IMAGER......................................................................... 53
n Kodak Color Medical Imager 1000 ................................................................................. 55

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
Mana
nput
ulti-I
0M
880
3M

yView
00 Dr
3M 87

Fiber Cable
8700 Dual

8800 MIM

Requirements: DSI rel. R4.2.1, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8700 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8700 Plus

Requirements: Philips DSI rel. R4.2.1, 8700 Plus rel. 1.15

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: Philips DSI rel. R4.2.1, 8700 Standard rel. 1.15

Setup: on Philips DSI rel. R4.2.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
Mana
nput
ulti-I
0M
880
3M

yView
00 Dr
3M 87

Fiber Cable
8500 Dual

8800 MIM

Requirements: DSI rel. R4.2.1, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8500 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8500 Plus

Requirements: Philips DSI rel. R4.2.1, 8500 Plus rel. 1.22

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: DSI rel. R4.2.1, 8500 Standard rel. 1.22

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
nag
ut Ma
lti-Inp
00 Mu

Fiber Cable 3M88

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: DSI rel. R4.2.1, 8800 rel. 5.1, 8300 rel. 4.2

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0

8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif=Yes
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips DSI
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


I mage Data OUT

Requirements: Philips DSI rel. R4.2.1, 8300 rel. 4.2

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Odd
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak Digital Science 969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M H Q L a ser Im ag er

M969 HQ

Requirements: Philips DSI rel. R4.2.1, M969 HQ rel. 3.0

Setup:

n on Philips DSI rel. R4.2.1 install Camera Config.File 3M


n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT 3 M H Q L a s er Im a g e r

Philips 37-pin
Data Cable
I mage Data OUT

Requirements: Philips DSI rel. R4.2.1, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak Digital Science 959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a s e r I m a g er

M959 XL

Requirements: Philips DSI rel. R4.2.1, M959 XL rel. 3.8

Setup:
n on Philips DSI rel. R4.2.1 install Camera Config.File 3M
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu:
n 959 Default
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak DryView 8100 LASER IMAGER

DryView 8100
RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips DSI Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

Requirements: DSI rel. R4.2.1, 8100 v1.1

Setup: on DSI rel. R4.2.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=10, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=11
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips DSI Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 DryV
iew Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: DSI rel. R4.2.1, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips DSA model DSI with LASER IMAGER Products

on I/O CARD SETUP, DPI4 Digital Menu:


n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=dsi, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips DSI Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: DSI rel. R4.2.1, 9410 v1.6, 8100 v2.4.1

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=dsi, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips DSA model DSI with LASER IMAGER Products

n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On


n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips DSI Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: DSI rel. R5.1.1


Requirements: MIM 200 v6.1, KELI 160 v7.2

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=Kodak,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=DSI DF 50Hz:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dsi,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=1024, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips DSA model DSI with LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips DSI Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 2180 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: DSI rel. R5.1.1


Requirements: MIM 200 v6.1, KELP 2180 v2.17H

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=Kodak,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=DSI DF 50Hz:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dsi,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=1024, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips DSA model DSI with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips DSI Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 1120 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: DSI rel. R5.1.1


Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=Kodak,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=DSI DF 50Hz:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dsi,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=1024, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 2400, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips DSA model DSI with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips DSI Control Adaptor Cable
Control Cable 00 DryV
iew
3M 87

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: DSI rel. R5.1.1, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM 200 Destinations

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=DSI DF 50Hz:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dsi,
n Mod.Name=dsi, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=1024, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips DSA model DSI with LASER IMAGER Products

n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips DSA model DSI with LASER IMAGER Products

Philips DSI with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips DSI Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: DSI rel. R5.1.1, MIM200 v6.1, 8100 v2.4.1

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=DSI DF 50Hz:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=dsi,
n Mod.Name=dsi, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1280, Lines=1024, Parity=Odd, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips DSA model DSI with LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8300

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.2


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT rel. 3.0.3B

Philips DSI Setup:


n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14 (10x8 is not released)
n film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8300=IMN_8300
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=8300, ADDRESS=IMN_8300, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_8300] set:
- PORT_NUMBER=104, HOST_NAME=IMN_8300
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=0
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0 i.e., Auto

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips DSA model DSI with LASER IMAGER Products

n Film Layout : Format=1 on 1, Mode=Landscape, 2:1L, 4:1L


n Note: due to DSI matrix 1024x1280, 4:1 allowed only with Film Layout: 4:1L
n on 8300 Service Menu into Configure User set: Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Kodak 190 camera
Merge Comm Board
Control 422OUT

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4, 190 rel. 1.7.29

Philips DSI Setup: on DSI install Camera Config.File 3M

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 190=NER_ANY
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=190, ADDRESS=NER_ANY, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_ANY] set:
- PORT_NUMBER=5040, HOST_NAME=NER_ANY
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS575\CN0\PD1.2, Polarity=NORMAL

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips DSI
Control Cable Merge Adaptor Cable Fiber Cable 3M 87
00 DryV
iew

Control 422OUT Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
9410 PACS Link
DryView 8700 Plus

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Philips DSI Setup:


n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips DSA model DSI with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
9410 PACS Link DryView 8100

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.2, 9410 v1.6, 8100 v2.4.1

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC, Smoothing_Type=11
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,9 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips DSA model DSI with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, Contrast Group=0
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Req.Image Size=0, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER
DICOM Network

Control 9-pin
COPCIL Board
Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board Kodak DryView 8300
Kodak 9410 PACS Link

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.2


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Philips DSI Setup:


n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14 (10x8 is not released)
n film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 9410=IMN_9410
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=9410, ADDRESS=IMN_9410, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [IMN_9410] set:
- PORT_NUMBER=1024, HOST_NAME=IMN_9410
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=0
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips DSA model DSI with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type =831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=IMN_9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Landscape, 2:1L, 4:1L
n Note: due to DSI matrix 1024x1280, 4:1 allowed only with Film Layout: 4:1L
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips DSA model DSI with LASER IMAGER Products

DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips DSI
Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 160 camera
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: MIM 200 v6.1, KELI 160 v7.2

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=ANY
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=ANY, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [ANY] set:
- PORT_NUMBER=5040, HOST_NAME=ANY
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips DSA model DSI with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 2180 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.4


Requirements: MIM200 v6.1, 2180 v2.17H

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_2180
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_2180, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_2180] set:
- PORT_NUMBER=5040, HOST_NAME=NER_2180
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Medium_Type=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips DSA model DSI with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Merge Comm Board Kodak 1120 camera
Control 422OUT
Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: MIM200 v6.1, 1120 v2.40/032

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_1120
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM, ADDRESS=NER_1120, PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_1120] set:
- PORT_NUMBER=5040, HOST_NAME=NER_1120
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=ANY, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=NORMAL, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=CS435\CN0\PD1.2, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips DSA model DSI with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips 37-pin RS422 37-to-25 pin Merge DPI


Philips DSI
Control Cable Merge Adaptor Cable
Merge Comm Board
Control 422OUT Crossed
Network Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8200

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: MIM200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8200
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8200,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8200] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8200
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=LUT=0,9, Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips DSA model DSI with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI 00 Dr


yView
3M 87
Control Cable Merge Adaptor Cable Fiber Cable
Control 422OUT Merge Comm Board

Fiber Cable
Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200
Kodak DryView 8700 Plus

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8700
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8700,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8700] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8700
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK

Author: gian.valesano@kodak.com, Edition: May ´05 46


Philips DSA model DSI with LASER IMAGER Products

- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO


- Configuration_ Information=LUT=0,9, Polarity=NORMAL
MIM200 Setup:
n on Configure, Network:
n Node=MIM, IPAddr=accordingly, Subnet=accordingly, Gateway=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On, Landscape=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 47


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
DICOM Network

RS422
Control 9-pin 9-pole Plug
COPCIL Board
Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board
Kodak MIM 200 Kodak DryView 8100

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

DSI Setup:
n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14, film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8100
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8100,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8100] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8100
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information=LUT=0,9, Polarity=NORMAL

MIM200 Setup:
n on Configure, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 48


Philips DSA model DSI with LASER IMAGER Products

n Node Name=MIM, IPAddr=accordingly


n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 49


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER
DICOM Network

Control 9-pin
COPCIL Board
Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI
Copper Cable
Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin Data 37-pin


Data Cable
Image Data OUT Merge Data Board Kodak DryView 8300
Kodak MIM 200

Requirements: Philips DSI rel. 4.2.1, Merge DPI rel. 2.2


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

Philips DSI Setup:


n on DSI press F4 and select Service, Start Service Mode
n on TCOP software select 2 (Testsoftware) / Run / Config / HCU
n set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
n HCU with remote control=yes, filmsize=17x14 (10x8 is not released)
n film feedback=yes
n cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of MIM=NER_8300
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=MIM,ADDRESS=NER_8300,PROTOCOL=DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [NER_8300] set:
- PORT_NUMBER=5040, HOST_NAME=NER_8300
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=LANDSCAPE
- Film_Size=8INX10IN, Magnification_Type=CUBIC, Smoothing_Type=2
- Border_Density=BLACK, Empty_Image_Density=BLACK, Max_Density=300
- Trim=NO, Configuration_ Information=LUT=0,10 , Polarity=NORMAL

Author: gian.valesano@kodak.com, Edition: May ´05 50


Philips DSA model DSI with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 51


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8900

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

DSI Setup:
- on DSI press F4 and select Service, Start Service Mode
- on TCOP software select 2 (Testsoftware) / Run / Config / HCU
- set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
- HCU with remote control=yes, filmsize=17x14, film feedback=yes
- cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8900= 8900
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8900,ADDRESS= 8900,PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8900] set:
- PORT_NUMBER=5040, HOST_NAME= 8900
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8900 Setup:
n on Configuration, System, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 52


Philips DSA model DSI with LASER IMAGER Products

n Host Name=DV8900, IPAddr=accordingly


n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=11 (Always=N), Image Polarity=Normal (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=dsi

Author: gian.valesano@kodak.com, Edition: May ´05 53


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin
Data Cable
Image Data OUT Merge Data Board
Kodak DryView 8150

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

DSI Setup:
- on DSI press F4 and select Service, Start Service Mode
- on TCOP software select 2 (Testsoftware) / Run / Config / HCU
- set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
- HCU with remote control=yes, filmsize=17x14, film feedback=yes
- cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of 8150= 8150
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME=DV8150,ADDRESS= 8150,PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [8150] set:
- PORT_NUMBER=5040, HOST_NAME= 8150
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=14INX17IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

8150 Setup:
n on Configuration, System, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 54


Philips DSA model DSI with LASER IMAGER Products

n Host Name=DV8150DRE, IPAddr=accordingly


n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=dsi

Author: gian.valesano@kodak.com, Edition: May ´05 55


Philips DSA model DSI with LASER IMAGER Products

Philips DSI via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips DSI Philips 37-pin RS422 37-to-25 pin Merge DPI


Control Cable Merge Adaptor Cable
Control 422OUT Merge Comm Board

Philips 37-pin
Data Cable Kodak CMI 1000
Image Data OUT Merge Data Board

Requirements: Philips DSI rel. 4.2.1, Merge DPI v2.4


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

DSI Setup:
- on DSI press F4 and select Service, Start Service Mode
- on TCOP software select 2 (Testsoftware) / Run / Config / HCU
- set: HCU board=yes, hardcopy=local copy, HCU protocol=3M,
- HCU with remote control=yes, filmsize=17x14, film feedback=yes
- cycle power DSI

Merge DPI Setup:


n into usr/home/merge/xpi/config/xpi_digital_952.ini under [HOST_PORT] set:
- BAUD=1200, PARITY=EVEN, DATA_BITS=8, STOP_BITS=1
- RS_232_422=RS-422, DTE_DCE=DTE
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [MACH] set: PARITY=ODD
n into usr/home/merge/xpi/config/xpi_digital_952.ini under [SC] set: ACQ_TYPE=DI
n into /etc/hosts set: IP addr. of CMI1000= CMI1000
n into usr/home/merge/xpi/config/targets.ini under [TARGET_2] set:
- NAME= CMI1000,ADDRESS= CMI1000,
- PROTOCOL= DICOM BASIC PRINT
- FORMAT_FILE=default.fm, DEFAULT=YES
n into usr/home/merge/xpi/config/tk_app.ini under [CMI1000] set:
- PORT_NUMBER=5040, HOST_NAME= CMI1000
- SERVICE_LIST=MergeXPI_Service_List
n into usr/home/merge/xpi/config/dbp.ini under [SOURCE_OVERRIDE] set:
- Film_Destination=PROCESSOR, Film_Orientation=PORTRAIT
- Film_Size=8INX10IN, Magnification_Type=CUBIC
- Smoothing_Type=11, Border_Density=BLACK
- Empty_Image_Density=BLACK, Max_Density=300, Trim=NO
- Configuration_ Information= LUT=0,9, Polarity=NORMAL

CMI-1000 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 56


Philips DSA model DSI with LASER IMAGER Products

at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 57


Philips MR Infinion with LASER IMAGER Products

Setup for Philips MR Infinion


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips MR Infinion with LASER IMAGER Products

Philips MR Infinion via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips MR Infinion

Kodak DryView 8300


Requirements: Infinion VIA4.0
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8300,
n Remote AE=8300, Remote Host=IPaddr.accordingly, Port=104, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=0,10
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8300, Format=6 on 1

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips MR Infinion with LASER IMAGER Products

Philips MR Infinion via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips MR Infinion Kodak 190 camera

Requirements: Infinion VIA4.0


Requirements: 190 rel. 1.7.29

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=190,
n Remote AE=190, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config= CS435\CN0
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=190, Format=20 on 1

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips MR Infinion
Fiber Cable

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Infinion VIA4.0
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=9410, Remote description=8700,
n Remote AE=9410, Remote Host=IPaddr.accordingly, Port=1024, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8700, Format=20 on 1

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips MR Infinion with LASER IMAGER Products

n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2


n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR Infinion

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100
Requirements: Infinion VIA4.0
Requirements: 9410 v1.6, 8100 v2.4.1

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=9410, Remote description=8100,
n Remote AE=9410, Remote Host=IPaddr.accordingly, Port=1024, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8100, Format=20 on 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips MR Infinion with LASER IMAGER Products

n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips MR Infinion with LASER IMAGER Products

Philips MR Infinion via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR Infinion COPCIL board
Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Infinion VIA4.0
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=9410, Remote description=8300,
n Remote AE=9410, Remote Host=IPaddr.accordingly, Port=1024, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=0,10
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8300, Format=6 on 1

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips MR Infinion with LASER IMAGER Products

n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips MR Infinion
Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, KELI 160 v7.2

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=160,
n Remote AE=160, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config= CS435\CN0
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=160, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips MR Infinion with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips MR Infinion
Kodak 2180 camera

Fiber Cable

Kodak MIM 200


Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, 2180 rel. 2.17H

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=2180,
n Remote AE=2180, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config= CS435\CN0
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=2180, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips MR Infinion with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips MR Infinion
Kodak 1120 camera

Fiber Cable

Kodak MIM 200


Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=1120,
n Remote AE=1120, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config= CS435\CN0
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=1120, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips MR Infinion with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Kodak DryView 8200

Philips MR Infinion

Crossed Network Cable

Kodak MIM 200


Requirements: Infinion VIA4.0
Requirements: MIM 200 v6.1, DV8200 v3.5

Note: This setup is basically also valid for 25 PS computer

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8200,
n Remote AE=8200, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8200, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips MR Infinion with LASER IMAGER Products

n Interpolation: Cubic Spline=On, Cubic Convolution 2=On


n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips MR Infinion
Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8700,
n Remote AE=8700, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8700, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips MR Infinion with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR Infinion

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8100,
n Remote AE=8100, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8100, Format=20 on 1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips MR Infinion with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR Infinion COPCIL board
Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Infinion VIA4.0
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8300,
n Remote AE=8300, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=0,10
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8300, Format=6 on 1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips MR Infinion with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips MR Infinion

Kodak DryView 8900


Requirements: Infinion VIA4.0
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8900,
n Remote AE=8900, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8900, Format=20 on 1

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips MR Infinion with LASER IMAGER Products

n Formatting: True Landscape=On, Rotation=On


n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=2 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=mr

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips MR Infinion

Kodak DryView 8150


Requirements: Infinion VIA4.0
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description=8150,
n Remote AE=DV8150, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera=8150, Format=20 on 1

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips MR Infinion with LASER IMAGER Products

n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=2 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=mr

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips MR Infinion with LASER IMAGER Products

MR Infinion via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips MR Infinion

Kodak CMI 1000


Requirements: Infinion VIA4.0
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

MR Infinion Setup:
n login as service to go to Unix level
n at Unix level type: cd /Picker/commands
n at Unix level type: dicom_manager, click OK, click New
n set Manufacturer=Kodak, Model=190, Remote description= CMI1000,
n Remote AE= CMI1000, Remote Host=IPaddr.accordingly, Port=5040, PDU=4
n click Previous, under DICOM Print SCU set Min.Density=20,
n Max.Density=300, Manufacturer’s Config=LUT=Pmsv3k.w87,9
n click Save, click Yes, click Exit, click OK
n at Unix level type: logout
n enter MR Application, select Film, set Camera= CMI1000, Format=6 on 1

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips MR Infinion with LASER IMAGER Products

n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)


n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Setup for Philips MR NT / Intera / Achieva


with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual.............................................................. 3
n Kodak DryView 8700 LASER IMAGER / Plus............................................................... 4
n Kodak DryView 8700 LASER IMAER / Standard .......................................................... 5
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual.............................................................. 6
n Kodak DryView 8500 LASER IMAGER / Plus............................................................... 7
n Kodak DryView 8500 LASER IMAGER / Standard ....................................................... 8
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual.............................................................. 9
n Kodak DryView 8300 LASER IMAGER / Standalone.................................................... 10
n Kodak Digital Science 969 HQ LASER IMAGER ...................................................... 11
n Kodak Digital Science 969 HQT LASER IMAGER ................................................... 12
n Kodak Digital Science 959 XL LASER IMAGER ...................................................... 13
n Kodak DryView 8100 LASER IMAGER......................................................................... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 15
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 21
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) . 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4).................................................................... 27
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4).................................................................... 29

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: June ´05 1


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER


(not possible, incompatibility during image transfer)
n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 31
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 32
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 34
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 36
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 38
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 40
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 42
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 44
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 46
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5) ................................................................. 48
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4,5) ................................................................. 50
n Kodak DryView 8900 LASER IMAGER......................................................................... 52
n Kodak DryView 8150 LASER IMAGER......................................................................... 54
n Kodak Color Medical Imager 1000 ................................................................................. 56

Author: gian.valesano@kodak.com, Edition: June ´05 2


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
putM
ulti-In
0M
880
3M

iew
0 DryV
3M 870

Fiber Cable
8700 Dual

8800 MIM

Requirements: MR NT rel. 4.5, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8700 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on EIB Menu:
n Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 3


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

iew
0 DryV
3M 870

8700 Plus

Requirements: Philips MR NT rel. 4.5, 8700 Plus rel. 1.15

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 4


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT
yView
00 Dr
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: Philips MR NT rel. 4.5, 8700 Standard rel. 1.15

Setup: on Philips MR NT rel. 4.5 install Camera Config.File 3M +

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 5


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
putM
ulti-In
0M
880
3M

iew
0 DryV
3M 870

Fiber Cable
8500 Dual

8800 MIM

Requirements: MR NT rel. 4.5, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8500 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 6


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

iew
0 DryV
3M 870

8500 Plus

Requirements: Philips MR NT rel. 4.5, 8500 Plus rel. 1.22

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 7


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT
yView
00 Dr
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: MR NT rel. 4.5, 8500 Standard rel. 1.22

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 8


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual
Philips 37-pin RS422 37-to-9 pin
Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
r
age
Man
put
ulti-In

Fiber Cable
00M
3M88

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: MR NT rel. 4.5, 8800 rel. 5.1, 8300 rel. 4.2


Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
RS422 37-to-9 Host 37 pin side 2 20 3 21 1
Adaptor Cable pinout 9 pin side 8 3 7 2 1
8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0
8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 9


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips MR NT
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


I mage Data OUT

Requirements: Philips MR NT rel. 4.5, 8300 rel. 4.2

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Even
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: June ´05 10


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak Digital Science 969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M H Q L a ser Im a g er

M969 HQ

Requirements: Philips MR NT rel. 4.5, M969 HQ rel. 3.0

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 11


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT 3 M H Q L a ser Im ag er

Philips 37-pin
Data Cable
I mage Data OUT

Requirements: Philips MR NT rel. 4.5, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 12


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak Digital Science 959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a se r I m ag er

M959 XL

Requirements: Philips MR NT rel. 4.5, M959 XL rel. 3.8

Setup:
n on Philips MR NT rel. 4.5 install Camera Config.File 3M +
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu:
n 959 Default
n on EIB Menu:
n Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 13


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak DryView 8100 LASER IMAGER

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips MR NT Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

DryView 8100

Requirements: MR NT rel. 4.5, 8100 v1.1

Setup: on MR NT rel. 4.5 install Camera Config.File 3M+

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 14


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips MR NT Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 Dr
yView Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: MR NT rel. 4.5, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Setup:
n on MR NT rel. 4.5 install Camera Config.File 3M+
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=60, Parity=Even, Memory=10
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=mrnt, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA

Author: gian.valesano@kodak.com, Edition: June ´05 15


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)


n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 16


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT with
Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER
Philips 37-pin RS422 37-to-9 pin
Philips MR NT Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: MR NT rel. 4.5, 9410 v1.6, 8100 v2.4.1


Setup:
n on MR NT rel. 4.5 install Camera Config.File 3M+
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=60, Parity=Even, Memory=10
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=mrnt, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

Author: gian.valesano@kodak.com, Edition: June ´05 17


Philips MR NT / Intera / Achieva with LASER IMAGER Products

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 18


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips MR NT / Intera Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: MR NT / Intera rel. 4.5


Requirements: MIM 200 v6.1, KELI 160 v7.2

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=mrnt,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8

Author: gian.valesano@kodak.com, Edition: June ´05 19


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2


n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 20


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips MR NT / Intera Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 2180 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: MR NT / Intera rel. 4.5


Requirements: MIM 200 v6.1, KELP 2180 v2.17H

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=mrnt,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2

Author: gian.valesano@kodak.com, Edition: June ´05 21


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Formatting: Trim Width Enabled=2, Trim Width Disabled=2


n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 22


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips MR NT / Intera Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 1120 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: MR NT / Intera rel. 4.5


Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=mrnt,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0

Author: gian.valesano@kodak.com, Edition: June ´05 23


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Formatting: Trim Width Enabled=0, Trim Width Disabled=0


n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 24


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips MR NT / Intera Control Adaptor Cable
Control Cable 0 DryVie
w
3M 870

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: MR NT rel. 4.5, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM 200 Destinations

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File 3M+

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=mr,
n Mod.Name=mr, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: June ´05 25


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 26


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips MR NT / Intera Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: MR NT rel. 4.5, MIM200 v6.1, 8100 v2.4.1

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File 3M+

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=mr,
n Mod.Name=mr, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4, 1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: June ´05 27


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Formatting: Pixel Pitch=12.8205


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 28


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board Control 9-pin
Philips MR NT / Intera Control Adaptor Cable
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8300

Requirements: MR NT rel. 4.5, MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Setup: on MR NT / Intera rel. 4.5 install Camera Config.File 3M+

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=Gyroscan NT-5/10/15:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=nt,
n Mod.Name=nt, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048

Author: gian.valesano@kodak.com, Edition: June ´05 29


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3


n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8300, Media Size Select=8x10in
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=10
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=u1_109b.w8u, Auto Study=Off, Console=On, Annotation=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: June ´05 30


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT / Intera / Achieva via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips MR NT Kodak 190 camera

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 190 rel. 1.7.29

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup:


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=190, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=190, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=190, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=NORMAL,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=CS435\CN0, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

190 Setup:
n On 190 Control Panel press Menu, Setup:

Author: gian.valesano@kodak.com, Edition: June ´05 31


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2


n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: June ´05 32


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

8700 Plus

Philips MR NT
Fiber Cable
iew
00 DryV
3M 87

Fiber Cable

9410 PACS Link

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=9410, port=1024,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer

Author: gian.valesano@kodak.com, Edition: June ´05 33


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n set AETitle=9410, IP addr.=accordingly, port=1024, Return,1=Exit,2=Logout


n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8700, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=9410, Port=1024, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsv3k.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 34


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR NT

Data 37-pin

9410 PACS Link


DryView 8100
Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)
Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 9410 v1.6, 8100 v2.4.1

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=9410, port=1024,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=9410, IP addr.=accordingly, port=1024, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 35


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8100, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=9410, Port=1024, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsb.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 36


Philips MR NT / Intera / Achieva with LASER IMAGER Products

Philips MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR NT / Intera COPCIL board
Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)
Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 8x10 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=9410, port=1024,
n min dens=20, max dens=300, config info=LUT=0,10
n trim=NO, size_id=8INx10IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=0,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8300, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=9410, Port=1024, IP=accordingly, Medium Type=BLUE FILM,

Author: gian.valesano@kodak.com, Edition: June ´05 37


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n set Image Diplay Format=1,4, Film Destination=PROCESSOR,


n set Photometric Interpretation=MONOCHROME2, Film Size ID=8INX10IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=0,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=0,10, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=8INX10IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: June ´05 38


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips MR NT / Intera Kodak 160 camera


Crossed
Network Cable

Kodak MIM 200


Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)
Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM 200 v6.1, KELI 160 v7.2

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=keli (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie keli
n modify parameters eg IP address=accordingly, scp_ae_title=160, port=5040,
n min density=20, max density=300, config info= CS435\CN0, trim=NO,
n size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=NORMAL,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie keli
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=160, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 39


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=160, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=160, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=NORMAL,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=CS435\CN0, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)

Author: gian.valesano@kodak.com, Edition: June ´05 40


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n NeventReports=Off (Always=Y), Warning Status=On


n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 41


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips MR NT / Intera Kodak 2180 camera

Fiber Cable

Kodak MIM 200


Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)
Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM200 v6.1, 2180 rel. 2.17H

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=kelp (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie kelp
n modify parameters eg IP address=accordingly, scp_ae_title=2180, port=5040,
n min density=20, max density=300, config info= CS435\CN0, trim=NO,
n size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=NORMAL,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie kelp
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=2180, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 42


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=2180, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=2180, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=NORMAL,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=CS435\CN0, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)

Author: gian.valesano@kodak.com, Edition: June ´05 43


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)


n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 44


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips MR NT / Intera
Kodak 1120 camera

Fiber Cable

Kodak MIM 200


Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)
Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=kelp (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie kelp
n modify parameters eg IP address=accordingly, scp_ae_title=1120, port=5040,
n min density=20, max density=300, config info= CS435\CN0, trim=NO,
n size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=NORMAL,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie kelp
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=1120, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 45


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=1120, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=1120, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=NORMAL,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=CS435\CN0, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)

Author: gian.valesano@kodak.com, Edition: June ´05 46


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)


n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 47


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips MR NT / Intera
Crossed
Network Cable

Kodak MIM 200 Kodak DryView 8200

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM 200 v6.1, 8200 v3.5

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=8200, port=5040,
n min density=20,max density=300,config info=LUT=0,9,
n trim=NO,size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=8200, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print is already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 48


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8200, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=8200, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsb.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On

Author: gian.valesano@kodak.com, Edition: June ´05 49


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: June ´05 50


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips MR NT / Intera
Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination
Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=8700, port=5040,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit

Author: gian.valesano@kodak.com, Edition: June ´05 51


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n into System: select 7=DICOM Management, 5=Config DICOM Printer


n set AETitle=8700, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8700, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=8700, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsv3k.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205

Author: gian.valesano@kodak.com, Edition: June ´05 52


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1


n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 53


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR NT / Intera

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer


Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=8100, port=5040,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=8100, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable

Author: gian.valesano@kodak.com, Edition: June ´05 54


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Patient->Select Exam->Proceed (note at this point image to print already selected)


n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8100, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=8100, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsb.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:

Author: gian.valesano@kodak.com, Edition: June ´05 55


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)


n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 56


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR NT / Intera COPCIL board
Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer


Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 8x10 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=8300, port=5040,
n min dens=20, max dens=300, config info=LUT=0,10
n trim=NO, size_id=8INx10IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=0,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note at this point image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer

Author: gian.valesano@kodak.com, Edition: June ´05 57


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n set Printer name=8300, Copy definition from=ISGDICOMkodak, click Apply/Next


n set SCP AE Title=8300, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=8INX10IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=0,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=0,10, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=8INX10IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109b.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)

Author: gian.valesano@kodak.com, Edition: June ´05 58


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: June ´05 59


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips MR NT / Intera

Kodak DryView 8900

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=8900, port=5040,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=8900, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 60


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8900, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AE Title=8900, Port=5040, IP=accordingly, Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsv3k.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=11 (Always=N), Image Polarity=Normal (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)

Author: gian.valesano@kodak.com, Edition: June ´05 61


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)


n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=mrnt

Author: gian.valesano@kodak.com, Edition: June ´05 62


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips MR NT / Intera

Kodak DryView 8150

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v10.3 (based on Windows platform)
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Note: on MR (Unix platform) only a DICOM camera could be installed

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name=dryview (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie dryview
n modify parameters eg IP address=accordingly, scp_ae_title=DV8150, port=5040,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=14INx17IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie dryview
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=DV8150, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 63


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=8150, Copy definition from=ISGDICOMkodak, click Apply/Next
n set SCP AETitle=DV8150,Port=5040,IP=according,Medium Type=BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2, Film Size ID=14INX17IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsb.w87,9, Polarity=NORMAL
n set Supported Medium Types=BLUE FILM, Film Sizes=14INX17IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,

Author: gian.valesano@kodak.com, Edition: June ´05 64


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)


n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=mr

Author: gian.valesano@kodak.com, Edition: June ´05 65


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR NT / Intera / Achieva via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips MR
- NT
- Intera
- Achieva

Kodak CMI 1000

Requirements: Philips MR NT / Intera v7.1.2 (based on Unix platform)


Requirements: Philips MR NT / Intera v11.1 (based on Windows platform)
Requirements: CMI-1000 v1.1 (DRE=v1.1, MIMCore=v2.0.b6)

Note: on MR (Unix platform) only a DICOM camera could be installed


Note: MR is not able to print color images.

MR (Unix platform) Setup (modifying 190 Config.file ie 14x17 only):


n into gyrotest do: define/job gyro_test=true
n into Figman: select 6=Console Config, 2=Copy Facilities
n select 1=Add a DICOM Printer Type
n enter name= CMI (no numbers are allowed)
n select an existing DICOM printer eg ISGDICOMkodak190
n select 3= Modify DICOM Printer Attributes, choose the created printer ie CMI
n modify parameters eg IP address=accordingly, scp_ae_title= CMI1000, port=5040,
n min dens=20, max dens=300, config info=LUT=0,9
n trim=NO, size_id=8INx10IN,
n orientation=PORTRAIT, magnification=CUBIC, smoothing=11,
n border_density=BLACK, formats=1,4, polarity=NORMAL
n select 4=Select a Printer/Camera, choose the created printer ie CMI
n cancel, cancel, exit
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

MR (Unix platform) Setup (using 190 Config.file ie 14x17 only):


n into Figman: select 6=Console Config, 2=Copy Facilities
n set P2P camera=None, DICOM camera=KD_MLP190, Proceed,Proceed,Exit
n into System: select 7=DICOM Management, 5=Config DICOM Printer
n set AETitle=CMI1000, IP addr.=accordingly, port=5040, Return,1=Exit,2=Logout
n restart application, wait a beep, check that HC->Camera Control->Output=Enable
n Patient->Select Exam->Proceed (note here image to print already selected)
n HC Screen-> Choose Film Layout->Print Job->Hide (ie which means Proceed)

Author: gian.valesano@kodak.com, Edition: June ´05 66


Philips MR NT / Intera / Achieva with LASER IMAGER Products

MR (Windows platform) Setup:


n press Windows Startmenu key
n enter into MR System Management/Config/Config.Application, click OK
n enter User name=at least 1 char, click OK, enter Type Action=Installation,
n enter Reason Action=1 char, Description Action=1 char, click OK
n select Config/Hardcopy Facilities/ Add DICOM Printer
n set Printer name=CMI, Copy definition from=ISGDICOMkodak,
n click Apply/Next
n set SCP AETitle=CMI1000,Port=5040,IP=accordingly,
n set Medium Type= BLUE FILM,
n set Image Diplay Format=1,4, Film Destination=PROCESSOR,
n set Photometric Interpretation=MONOCHROME2 (RGB not supported by MR),
n set Film Size ID=8INX10IN,
n set Film Orientation=PORTRAIT, Magnification=CUBIC, Smoothing=11,
n set Border Density=BLACK, Empty Density=BLACK
n set Min Density=20, Max Density=300, Trim=NO,
n set Configuration Information=LUT=Pmsv3k.w87,9, Polarity=NORMAL
n set Supported Medium Types= BLUE FILM, Film Sizes=8INX10IN
n click Apply/OK/Yes/OK/OK/OK/Logoff/Yes/OK/OK/Close/Close/Restart
n to select SMPTE go into application under System/Advanced Tools/Service
n go under System/Print Job Control, select Destination=CMI, Printer Enabled=On
n to print press button HC-Screen

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y),TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role

Author: gian.valesano@kodak.com, Edition: June ´05 67


Philips MR NT / Intera / Achieva with LASER IMAGER Products

n Image Resizing=Minify (Always=N),


n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: June ´05 68


Philips MR Panorama with LASER IMAGER Products

Setup for Philips MR Panorama


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips MR Panorama

Kodak DryView 8300

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 8300, Remote Host Port=104
n Calling AE Title=Panorama, Called AE Title=8300
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=6, Copies=1

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=2
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips MR Panorama Kodak 190 camera

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 190 rel. 1.7.29

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=Kodak, Property Template=None
n Remote Host Name=IP addr. of 190, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=190
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=Kodak, Format=20, Copies=1

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips MR Panorama
Fiber Cable

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 9410, Remote Host Port=1024
n Calling AE Title=Panorama, Called AE Title=9410
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips MR Panorama with LASER IMAGER Products

n Trim=No, Attribute Warning=No


n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR Panorama

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 9410 v1.6, 8100 v2.4.1

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 9410, Remote Host Port=1024
n Calling AE Title=Panorama, Called AE Title=9410
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips MR Panorama with LASER IMAGER Products

n on SIB board (ie external switch below Control plugs): SW1=Down


n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR Panorama COPCIL board
Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 9410, Remote Host Port=1024
n Calling AE Title=Panorama, Called AE Title=9410
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=6, Copies=1

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips MR Panorama with LASER IMAGER Products

n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto


n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips MR Panorama
Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, KELI 160 v7.2

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=Kodak, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=160
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=Kodak, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips MR Panorama with LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips MR Panorama
Kodak 2180 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, 2180 rel. 2.17H

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=Kodak, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=2180
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=Kodak, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips MR Panorama with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips MR Panorama
Kodak 1120 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=Kodak, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=1120
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=Kodak, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips MR Panorama with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips MR Panorama
Crossed
Network
Cable

Kodak MIM 200 Kodak DryView 8200

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM 200 v6.1, DV8200 v3.5

Note: This setup is basically also valid for 25 PS computer

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=8200
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips MR Panorama with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips MR Panorama
Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=8700
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips MR Panorama with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips MR Panorama

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=8100
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips MR Panorama with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips MR Panorama COPCIL board
Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of MIM, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=8300
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=6, Copies=1

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips MR Panorama with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips MR Panorama

Kodak DryView 8900

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 8900, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=8900
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips MR Panorama with LASER IMAGER Products

n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4


n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=N), Contrast=9 (Always=N)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=mr

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips MR Panorama

Kodak DryView 8150

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name=DryView, Property Template=None
n Remote Host Name=IP addr. of 8150, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title=DV8150
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera=DryView, Format=20, Copies=1

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips MR Panorama with LASER IMAGER Products

n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10


n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=mr

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips MR Panorama with LASER IMAGER Products

Philips MR Panorama via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips MR Panorama

Kodak CMI 1000

Requirements: Philips MR Panorama v5.1 (based on Win2000 platform)


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

MR Panorama Setup:
n into Tools/Options/DICOM/Printer/New set:
n Printer Name= CMI1000, Property Template=None
n Remote Host Name=IP addr. of CMI1000, Remote Host Port=5040
n Calling AE Title=Panorama, Called AE Title= CMI1000
n click OK/Close/OK/Exit (to restart)
n to print into HardCopy set Camera= CMI1000, Format=6, Copies=1

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips MR Panorama with LASER IMAGER Products

n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No


n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Setup for Philips Nuc.Med. Odyssey


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Philips Nuc.Med. Odyssey via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Odyssey

Kodak DryView 8300


Requirements: Odyssey v1.2.1
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=104, Remote Applic.Title=8300, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8300,
n Printer type=Kodak MLP190, Printer’s queue name=8300
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8300, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Philips Nuc.Med. Odyssey via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips Nuc.Med. Kodak 190 camera


Odyssey

Requirements: Odyssey v1.2.1


Requirements: 190 rel. 1.7.29

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=190, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=190,
n Printer type=Kodak MLP190, Printer’s queue name=190
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=190, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Nuc.Med.
Fiber Cable
Odyssey

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Odyssey v1.2.1
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=1024, Remote Applic.Title=9410, Remote node name=9410
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8700,
n Printer type=Kodak MLP190, Printer’s queue name=8700
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8700, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n SCU AETitle= accordingly, SCU IPAddr=accordingly


n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Nuc.Med.
Odyssey

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100
Requirements: Odyssey v1.2.1
Requirements: 9410 v1.6, 8100 v2.4.1

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=1024, Remote Applic.Title=9410, Remote node name=9410
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8100,
n Printer type=Kodak MLP190, Printer’s queue name=8100
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8100, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No


n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Philips Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Nuc.Med. COPCIL board
Odyssey Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Odyssey v1.2.1
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=1024, Remote Applic.Title=9410, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8300,
n Printer type=Kodak MLP190, Printer’s queue name=8300
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8300, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No


n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Odyssey Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, KELI 160 v7.2

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=160, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=160,
n Printer type=Kodak MLP190, Printer’s queue name=160
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=160, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips Nuc.Med.
Kodak 2180 camera
Odyssey

Fiber Cable

Kodak MIM 200


Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, 2180 rel. 2.17H

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=2180, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=2180,
n Printer type=Kodak MLP190, Printer’s queue name=2180
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=2180, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Format Types: Custom=On, Slide=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips Nuc.Med.
Kodak 1120 camera
Odyssey

Fiber Cable

Kodak MIM 200


Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=1120, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=1120,
n Printer type=Kodak MLP190, Printer’s queue name=1120
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=1120, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Format Types: Custom=On, Slide=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Nuc.Med. Crossed


Odyssey Network
Cable

Kodak MIM 200 Kodak DryView 8200


Requirements: Odyssey v1.2.1
Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=8200, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8200,
n Printer type=Kodak MLP190, Printer’s queue name=8200
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8200, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Supported Item: Remote Image Deletion=On, Rotation=On


n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Nuc.Med.
Fiber Cable
Odyssey

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=8700, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8700,
n Printer type=Kodak MLP190, Printer’s queue name=8700
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8700, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None


n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Nuc.Med.
Odyssey

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=8100, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8100,
n Printer type=Kodak MLP190, Printer’s queue name=8100
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8100, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Nuc.Med. COPCIL board
Odyssey Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Odyssey v1.2.1
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=8300, Remote node name=MIM
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8300,
n Printer type=Kodak MLP190, Printer’s queue name=8300
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8300, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Odyssey

Kodak DryView 8900


Requirements: Odyssey v1.2.1
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=8900, Remote node name=DV8900
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8900,
n Printer type=Kodak MLP190, Printer’s queue name=8900
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8900, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Formatting: Req.Size Scale=On, Req.Size Mix Page=On


n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=2 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=nm

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Odyssey

Kodak DryView 8150


Requirements: Odyssey v1.2.1
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=DV8150, Remote node name=DV8150
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=8150,
n Printer type=Kodak MLP190, Printer’s queue name=8150
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=8150, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n Formatting: True Landscape=On, Rotation=On


n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=2 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=nm

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips Nuc.Med. Odyssey with LASER IMAGER Products

Nuc.Med. Odyssey via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Nuc.Med.
Odyssey

Kodak CMI 1000


Requirements: Odyssey v1.2.1
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

NM Odyssey Setup:
n with the pointer on the screen background, middle-click on the mouse
n select Configure/DICOM Print/Add
n set Port=5040, Remote Applic.Title=CMI1000, Remote node name=CMI1000
n click Done, click Yes-proceed, set IP=accordingly, click Done, OK
n type Unix root password and note assigned Dest.Number, click Done
n select Imprint Configuration/Add a printer, set Menu name for printer=CMI1000,
n Printer type=Kodak MLP190, Printer’s queue name=CMI1000
n click Update, Yes-proceed, OK, type Unix root password, Done, Done
n with the pointer on the screen background, middle-click on the mouse
n select Imprint Film, select Printer=CMI1000, select Options/set Printer controls
n Polarity=Positive, Border=Black, Trim=Off, Orient=Landscape, Mag=Cubic
n Smooth=Current, Copies=1, Density Min=0.2, Density Max=3.0, click Done

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips Nuc.Med. Odyssey with LASER IMAGER Products

n under More Quality Settings:


n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Setup for Philips Nuc.Med. Pegasys


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Philips Nuc.Med. Pegasys via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Pegasys

Kodak DryView 8300


Requirements: Pegasys v4.20
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8300, IP=accordingly and click again Add Host
n Printer Name=8300, Port=104, Called AE Title=8300, Copies=1,
n Config.Info=LUT=0,10, Min.Density=20, Max.Density=300, Smoothing=0
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=8INX10IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8300 printer, press Return, Return

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Philips Nuc.Med. Pegasys via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips Nuc.Med. Kodak 190 camera


Pegasys

Requirements: Pegasys v4.20


Requirements: 190 rel. 1.7.29

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=190, IP=accordingly and click again Add Host
n Printer Name=190, Port=5040, Called AE Title=190, Copies=1,
n Config.Info=CS435\CN0, Min.Dens.=20, Max.Dens.=300, Smoothing=NORMAL
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 190 printer, press Return, Return

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Nuc.Med.
Fiber Cable
Pegasys

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Pegasys v4.20
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8700, IP=accordingly and click again Add Host
n Printer Name=8700, Port=1024, Called AE Title=9410, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8700 printer, press Return, Return

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5


n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Nuc.Med.
Pegasys

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100
Requirements: Pegasys v4.20
Requirements: 9410 v1.6, 8100 v2.4.1

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8100, IP=accordingly and click again Add Host
n Printer Name=8100, Port=1024, Called AE Title=9410, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8100 printer, press Return, Return

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2


n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Philips Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Nuc.Med. COPCIL board
Pegasys Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Pegasys v4.20
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8300, IP=accordingly and click again Add Host
n Printer Name=8300, Port=1024, Called AE Title=9410, Copies=1,
n Config.Info=LUT=0,10, Min.Dens.=20, Max.Dens.=300, Smoothing=0
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=8INX10IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8300 printer, press Return, Return

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0


n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Pegasys Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, KELI 160 v7.2

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=160, IP=accordingly and click again Add Host
n Printer Name=160, Port=5040, Called AE Title=160, Copies=1,
n Config.Info=CS435\CN0, Min.Dens.=20, Max.Dens.=300, Smoothing=NORMAL
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 160 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Format Types: Custom=On, Slide=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips Nuc.Med.
Kodak 2180 camera
Pegasys

Fiber Cable

Kodak MIM 200


Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, 2180 rel. 2.17H

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=2180, IP=accordingly and click again Add Host
n Printer Name=2180, Port=5040, Called AE Title=2180, Copies=1,
n Config.Info=CS435\CN0, Min.Dens.=20, Max.Dens.=300, Smoothing=NORMAL
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 2180 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips Nuc.Med.
Kodak 1120 camera
Pegasys

Fiber Cable

Kodak MIM 200


Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=1120, IP=accordingly and click again Add Host
n Printer Name=1120, Port=5040, Called AE Title=1120, Copies=1,
n Config.Info=CS435\CN0, Min.Dens.=20, Max.Dens.=300, Smoothing=NORMAL
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 1120 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Nuc.Med. Crossed


Pegasys Network
Cable

Kodak MIM 200 Kodak DryView 8200


Requirements: Pegasys v4.20
Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8200, IP=accordingly and click again Add Host
n Printer Name=8200, Port=5040, Called AE Title=8200, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8200 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)


n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Nuc.Med.
Fiber Cable
Pegasys

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8700, IP=accordingly and click again Add Host
n Printer Name=8700, Port=5040, Called AE Title=8700, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8700 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n General: Database Name=8700, Model=8700, Memory=32


n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Nuc.Med.
Pegasys

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8100, IP=accordingly and click again Add Host
n Printer Name=8100, Port=5040, Called AE Title=8100, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8100 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60


n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Nuc.Med. COPCIL board
Pegasys Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Pegasys v4.20
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8300, IP=accordingly and click again Add Host
n Printer Name=8300, Port=5040, Called AE Title=8300, Copies=1,
n Config.Info=LUT=0,10, Min.Dens.=20, Max.Dens.=300, Smoothing=0
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=8INX10IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8300 printer, press Return, Return

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60


n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Pegasys

Kodak DryView 8900


Requirements: Pegasys v4.20
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8900, IP=accordingly and click again Add Host
n Printer Name=8900, Port=5040, Called AE Title=8900, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8900 printer, press Return, Return

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=2 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=nm

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Nuc.Med.
Pegasys

Kodak DryView 8150


Requirements: Pegasys v4.20
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=8150, IP=accordingly and click again Add Host
n Printer Name=8150, Port=5040, Called AE Title=DV8150, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=14INX17IN, Color=Grayscale, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to 8150 printer, press Return, Return

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Formatting: Req.Size Scale=On, Req.Size Mix Page=On


n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=2 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=nm

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips Nuc.Med. Pegasys with LASER IMAGER Products

Nuc.Med. Pegasys via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Nuc.Med.
Pegasys

Kodak CMI 1000


Requirements: Pegasys v4.20
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

NM Pegasys Setup:
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select Patient Utilities/DICOM Utilities/DICOM Config
n click Configure=Printer, click Add,
n click Add Host and set Host=CMI1000, IP=accordingly and click again Add Host
n Printer Name= CMI1000, Port=5040, Called AE Title= CMI1000, Copies=1,
n Config.Info=LUT=Pmsv3k.w87,9, Min.Dens.=20, Max.Dens.=300, Smoothing=2
n Format=STANDARD\2,2, Orientation=LANDSCAPE, Maedia=BLUE FILM,
n Film Size=8INX10IN, Color=Color, Border=BLACK, Empty=BLACK,
n Transpose=NO, Magnif=CUBIC, Trim=NO, Dest=PROCESSOR, Priority=MED
n Save All & Exit
n with the pointer on the screen background, right-click on the mouse
n with further right-clicks select System Maintenance/Set Default Devices
n with further right-clicks select Change Default Settings/Change Default Formatter
n choose the number corresponding to CMI1000 printer, press Return, Return

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips Nuc.Med. Pegasys with LASER IMAGER Products

n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)


n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips Ultrasound Envisor with LASER IMAGER Products

Setup for Philips Ultrasound Envisor


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips Ultrasound
Envisor

Kodak DryView 8300


Requirements: Philips Envisor vC.0
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8300, IP Addr.=accordingly, Port=104
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=2, Rows=3, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=8in x 10in, Media=Blue Film,
n Config.Printer Specific Data=LUT=1,2, Max Density=260,
n Page Size: sizeX=2256, sizeY=2676, Resolution=11.81, click OK
n click Done, click OK, click Close

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips Ultrasound Kodak 190 camera


Envisor

Requirements: Philips Envisor vC.0


Requirements: 190 rel. 1.7.29

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=K190, AE Title=190, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=K190
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=CS435\CN0, Max Density=260,
n Page Size: sizeX=4096, sizeY=5120, Resolution=12.67, click OK
n click Done, click OK, click Close

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Ultrasound
Fiber Cable
Envisor

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Philips Envisor vC.0
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=9410, IP Addr.=accordingly, Port=1024
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4096, sizeY=5216, Resolution=12.82, click OK
n click Done, click OK, click Close

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips Ultrasound Envisor with LASER IMAGER Products

n on Device Setup, SCP PRINTERS, Registered Users:


n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Ultrasound
Envisor

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100
Requirements: Philips Envisor vC.0
Requirements: 9410 v1.6, 8100 v2.4.1

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=9410, IP Addr.=accordingly, Port=1024
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4096, sizeY=5216, Resolution=12.82, click OK
n click Done, click OK, click Close

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips Ultrasound Envisor with LASER IMAGER Products

n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low


n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Ultrasound COPCIL board
Envisor Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Philips Envisor vC.0
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=9410, IP Addr.=accordingly, Port=1024
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=2, Rows=3, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=8in x 10in, Media=Blue Film,
n Config.Printer Specific Data=LUT=upmsv3k.w8u,2, Max Density=260,
n Page Size: sizeX=2256, sizeY=2676, Resolution=11.81, click OK
n click Done, click OK, click Close

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips Ultrasound Envisor with LASER IMAGER Products

n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low


n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips Ultrasound
Envisor Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, KELI 160 v7.2

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=K160, AE Title=160, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=K160
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=CS435\CN0, Max Density=260,
n Page Size: sizeX=4360, sizeY=5348, Resolution=12.82, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips Ultrasound Envisor with LASER IMAGER Products

n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On


n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips Ultrasound
Kodak 2180 camera
Envisor

Fiber Cable

Kodak MIM 200


Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, 2180 rel. 2.17H

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=K2180, AE Title=2180, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=K2180
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=CS435\CN0, Max Density=260,
n Page Size: sizeX=4096, sizeY=5156, Resolution=12.67, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips Ultrasound Envisor with LASER IMAGER Products

n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips Ultrasound
Kodak 1120 camera
Envisor

Fiber Cable

Kodak MIM 200


Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=K1120, AE Title=1120, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=K1120
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=CS435\CN0, Max Density=260,
n Page Size: sizeX=4096, sizeY=5132, Resolution=12.67, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips Ultrasound Envisor with LASER IMAGER Products

n Interpolation: NoMag=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Ultrasound Crossed


Envisor Network
Cable

Kodak MIM 200 Kodak DryView 8200


Requirements: Philips Envisor vC.0
Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8200, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4360, sizeY=5216, Resolution=12.82, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips Ultrasound Envisor with LASER IMAGER Products

n Supported Item: Collation=On, Common Text=On, Tonescaling=On


n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Ultrasound
Fiber Cable
Envisor

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8700, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4096, sizeY=5216, Resolution=12.82, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips Ultrasound Envisor with LASER IMAGER Products

n Media Info: Media=35x43cm, 4096x5223, 5133, Blue


n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Ultrasound
Envisor

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8100, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4096, sizeY=5216, Resolution=12.82, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips Ultrasound Envisor with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Ultrasound COPCIL board
Envisor Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Philips Envisor vC.0
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8300, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=2, Rows=3, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=8in x 10in, Media=Blue Film,
n Config.Printer Specific Data=LUT=upmsv3k.w8u,2, Max Density=260,
n Page Size: sizeX=2256, sizeY=2676, Resolution=11.81, click OK
n click Done, click OK, click Close

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips Ultrasound Envisor with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On


n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Ultrasound
Envisor

Kodak DryView 8900


Requirements: Philips Envisor vC.0
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=8900, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4488, sizeY=5464, Resolution=12.82, click OK
n click Done, click OK, click Close

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips Ultrasound Envisor with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Image Size=On


n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=2 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=us

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Ultrasound
Envisor

Kodak DryView 8150


Requirements: Philips Envisor vC.0
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=DryView, AE Title=DV8150, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select B&W Printer SCP=DryView
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=3, Rows=5, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=14in x 17in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=4438, sizeY=5390, Resolution=12.82, click OK
n click Done, click OK, click Close

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips Ultrasound Envisor with LASER IMAGER Products

n Formatting: Req.Size Mix Row=On, Trim=On,


n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=2 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=us

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips Ultrasound Envisor with LASER IMAGER Products

Philips Ultrasound Envisor via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Ultrasound
Envisor

Kodak CMI 1000


Requirements: Philips Envisor vC.0
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Envisor Setup:
n press Setup-button
n by pressing Enter-button go into System/DICOM/Servers & Roles
n under Servers section, click New and set:
n Name=CMI1000, AE Title=CMI1000, IP Addr.=accordingly, Port=5040
n click Done
n under Roles section, click Modify, select Color Printer SCP=CMI1000
n click Advanced
n into Basic set: Copies=1, Orient.=Portrait, Cols=2, Rows=3, click OK
n click Advanced
n into Advanced set: Dest.=Processor, Size=8in x 10in, Media=Blue Film,
n Config.Printer Specific Data=LUT=Pmsv3k.w87,5, Max Density=260,
n Page Size: sizeX=2508, sizeY=2954, Resolution=12.35, click OK
n click Done, click OK, click Close

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips Ultrasound Envisor with LASER IMAGER Products

n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)


n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Setup for Philips Ultrasound HDI 5000


with LASER IMAGERS
Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: June ´05 1


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Philips Ultrasound HDI 5000 via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips Ultrasound
HDI 5000

Kodak DryView 8300


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

HDI 5000 Setup:


n Press Setups-button, select Image Management and press Select-button
n select Configuration Files and press Select, select Edit and press Select
n select Host Table and press Select, go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8300 and delete the // in front of this line
n at the side of Imation_(3M)8300 insert IP-address of 8300 and Port=104
n note that at the side of Port the entry for Device Table is 3m_8300_Printer_v3
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8300_Printer_v3 that: Type=Image_Printer, File=3m_8300.v3
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8300.V3 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8300, Film Size=8x10, Film Format=2x3, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=300, Empty Density=300
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8300, Copies=1, B&W Film Format=2x3
n Close
n To print press Patient Data, press Print for every image, press End Exam

8300 Setup:
n on 8300 Operator Menu into Configure User set: D=16, C=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set: Modality Group=0

Author: gian.valesano@kodak.com, Edition: June ´05 2


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Philips Ultrasound HDI 5000 via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips Ultrasound Kodak 190 camera


HDI 5000

Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17


Requirements: 190 rel. 1.7.29

HDI 5000 Setup:


n Press Setups-button, select Image Management and press Select-button
n select Configuration Files and press Select, select Edit and press Select
n select Host Table and press Select, go to $SAM_IMAGEPRINTERS
n choose KELP2180 and delete the // in front of this line
n at the side of KELP2180 insert IP-address of 190 and Port=5040
n note that at the side of Port the entry for Device Table is Kodak_KELP2180_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for Kodak_KELP2180_v2 that: Type=Image_Printer, File=kelp2180.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file KELP2180.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=KELP2180, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=KELP2180, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: June ´05 3


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Ultrasound
Fiber Cable
HDI 5000

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of 9410 and Port=1024
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8700.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=320, Empty Density=320
n Close
n under Setups\Image Management\Printer Selection check that

Author: gian.valesano@kodak.com, Edition: June ´05 4


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Printed Image Destination=B/W to B/W Printer


n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=260, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 5


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Ultrasound
HDI 5000

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17


Requirements: 9410 v1.6, 8100 v2.4.1

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of 9410 and Port=1024
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8700.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=320, Empty Density=320
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 6


Philips Ultrasound HDI 5000 with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=260, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 7


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Philips Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Ultrasound COPCIL board
HDI 5000 Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8300 and delete the // in front of this line
n at the side of Imation_(3M)8300 insert IP-address of 9410 and Port=1024
n note that at the side of Port the entry for Device Table is 3m_8300_Printer_v3
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8300_Printer_v3 that: Type=Image_Printer, File=3m_8300.v3
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8300.V3 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8300, Film Size=8x10, Film Format=2x3, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=300, Empty Density=300
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8300, Copies=1, B&W Film Format=2x3
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 8


Philips Ultrasound HDI 5000 with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: June ´05 9


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips Ultrasound
HDI 5000 Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM200 v6.1, KELI 160 v7.2

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose KELP2180 and delete the // in front of this line
n at the side of KELP2180 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is Kodak_KELP2180_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for Kodak_KELP2180_v2 that: Type=Image_Printer, File=kelp2180.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file KELP2180.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=KELP2180, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=KELP2180, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 10


Philips Ultrasound HDI 5000 with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 11


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips Ultrasound
Kodak 2180 camera
HDI 5000

Fiber Cable

Kodak MIM 200


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM200 v6.1, 2180 rel. 2.17H

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose KELP2180 and delete the // in front of this line
n at the side of KELP2180 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is Kodak_KELP2180_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for Kodak_KELP2180_v2 that: Type=Image_Printer, File=kelp2180.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file KELP2180.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=KELP2180, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=KELP2180, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 12


Philips Ultrasound HDI 5000 with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 13


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips Ultrasound
Kodak 1120 camera
HDI 5000

Fiber Cable

Kodak MIM 200


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose KELP1120 and delete the // in front of this line
n at the side of KELP1120 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is Kodak_KELP1120_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for Kodak_KELP1120_v2 that: Type=Image_Printer, File=kelp1120.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file KELP1120.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=KELP1120, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=KELP1120, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 14


Philips Ultrasound HDI 5000 with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: June ´05 15


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Ultrasound Crossed


HDI 5000 Network
Cable

Kodak MIM 200 Kodak DryView 8200


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8700.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=320, Empty Density=320
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5

Author: gian.valesano@kodak.com, Edition: June ´05 16


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Close
n To print press Patient Data, press Print for every image, press End Exam

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=2.60 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=Y), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87

Author: gian.valesano@kodak.com, Edition: June ´05 17


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n on LOCAL PANEL Menu: Contrast=6, Density=3.00


n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: June ´05 18


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Ultrasound
Fiber Cable
HDI 5000

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM200 v6.1, 8700 v1.34
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8700.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=320, Empty Density=320
n Close
n under Setups\Image Management\Printer Selection check that

Author: gian.valesano@kodak.com, Edition: June ´05 19


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Printed Image Destination=B/W to B/W Printer


n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=2.60 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=Y), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: June ´05 20


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Ultrasound
HDI 5000

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8700.V2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=320, Empty Density=320
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5

Author: gian.valesano@kodak.com, Edition: June ´05 21


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Close
n To print press Patient Data, press Print for every image, press End Exam

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=2.60 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=Y), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off

Author: gian.valesano@kodak.com, Edition: June ´05 22


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87


n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: June ´05 23


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips Ultrasound COPCIL board
HDI 5000 Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8300 and delete the // in front of this line
n at the side of Imation_(3M)8300 insert IP-address of MIM200 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8300_Printer_v3
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8300_Printer_v3 that: Type=Image_Printer, File=3m_8300.v3
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3M_8300.V3 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8300, Film Size=8x10, Film Format=2x3, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=300, Empty Density=300
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8300, Copies=1, B&W Film Format=2x3

Author: gian.valesano@kodak.com, Edition: June ´05 24


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Close
n To print press Patient Data, press Print for every image, press End Exam

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=Y), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: June ´05 25


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Ultrasound
HDI 5000

Kodak DryView 8900


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of 8900 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3m_8700.v2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 26


Philips Ultrasound HDI 5000 with LASER IMAGER Products

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=2 (Always=Y), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=2.60 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=us

Author: gian.valesano@kodak.com, Edition: June ´05 27


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Ultrasound
HDI 5000

Kodak DryView 8150


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

HDI 5000 Setup:


n Press Setups-button
n select Image Management and press Select-button
n select Configuration Files and press Select-button
n select Edit and press Select-button
n select Host Table and press Select-button
n go to $SAM_IMAGEPRINTERS
n choose Imation_(3M)8700 and delete the // in front of this line
n at the side of Imation_(3M)8700 insert IP-address of 8150 and Port=5040
n note that at the side of Port the entry for Device Table is 3m_8700_Printer_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for 3m_8700_Printer_v2 that: Type=Image_Printer, File=3m_8700.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file 3m_8700.v2 and change parameter, exit with Save
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer=3M_8700, Film Size=14x17, Film Format=3x5, Film Type=Blue
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=260,
n Border Density=399, Empty Density=399
n Close
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer
n B/W Printer=3M_8700, Copies=1, B&W Film Format=3x5
n Close
n To print press Patient Data, press Print for every image, press End Exam

Author: gian.valesano@kodak.com, Edition: June ´05 28


Philips Ultrasound HDI 5000 with LASER IMAGER Products

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=2 (Always=N)
n Dmin=0.00 (Always=N), Dmax=2.60 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=us

Author: gian.valesano@kodak.com, Edition: June ´05 29


Philips Ultrasound HDI 5000 with LASER IMAGER Products

Ultrasound HDI 5000 via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Ultrasound
HDI 5000

Kodak CMI 1000


Requirements: Philips HDI 5000 rel. 4252-0913-15-190.17
Requirements: CMI-1000 v1.1 (DRE=v1.1, MIMCore=v2.0.b6)

HDI 5000 Setup:


n Press Setups-button, select Image Management and press Select-button
n select Configuration Files and press Select, select Edit and press Select
n select Host Table and press Select, go to $SAM_IMAGEPRINTERS
n choose Codonics and delete the // in front of this line
n at the side of Codonics insert IP-address of CMI1000 and Port=5040
n note that at the side of Port the entry for Device Table is Codonics_v2
n Save, (note: to go back, select Setups Directory and press Select-button)
n
n just as a check, select Device Table and press Select-button
n check for Codonics_v2 that: Type=Image_Printer, File= Codonics.v2
n exit with Don’t Save
n
n just if a parameter change is needed, select Devices and press Select-button
n double-click on file Codonics.v2 and change parameter, exit with Save
n (here you can see that for Codonics both Grey and Color are enabled)
n
n under Setups\Image Management\B&W Printer Setups check that
n B&W Printer= Codonics, Film Size=8x10, Film Format=2x3, Film Type=Paper
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=300, Empty Density=300
n Close
n
n under Setups\Image Management\Color Printer Setups check that
n Color Printer=Codonics, Film Size=8x10, Film Format=2x3, Film Type=Paper
n Destination=Processor, Orientation=Portrait, Min Density=0, Max Density=300,
n Border Density=300, Empty Density=300
n Close
n
n under Setups\Image Management\Printer Selection check that
n Printed Image Destination=B/W to B/W Printer Color to Color Printer
n B/W Printer=Codonics, B/W Copies=1, B&W Film Format=2x3

Author: gian.valesano@kodak.com, Edition: June ´05 30


Philips Ultrasound HDI 5000 with LASER IMAGER Products

n Color Printer=Codonics, Color Copies=1, Color Film Format=2x3


n Close
n
n To print press Patient Data, press Print for every image, press End Exam
n (to have Color images printed, the blue-button Color must be pressed)

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=Y), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: June ´05 31


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Specific Setup for Philips Ultrasound iU22


with LASER IMAGERS

OEM Data
Manufacturer Philips
Modality US
Model iU22
sw version v1.1.0.138
Implement. Class UID 1.3.46.670589.5.2.10
Implement. Version Name ACP1.1L4

Kodak Data
DICOM-interface MIM internal / external or 9410 as below reported
DICOM-intf sw version as below reported
Printer model All cameras as below reported
Printer sw version as below reported

What’s new:
The setup of a Kodak printer for a given OEM Modality was split in 2 separate documents:

- the generic Setup Catalog for Kodak Cameras


- the specific Setup Manual for a given OEM Modality

************* the generic Setup Catalog for Kodak Cameras:


the Setup Catalog for Kodak Cameras contains that part of a printer installation which is Generic, ie common to
all DICOM connections independently from the Specific modality connected.

Therefore, as first step, setup the printer as per the correspondent page of the Setup Catalog for Kodak Cameras
(ie go to the Setup Catalog doc and do that first, then come back
here and finish install as per this Specific Setup Manual doc).

************ the specific Setup Manual for a given OEM Modality:


the Setup Manual for a given OEM Modality contains that part of a printer installation which is Specific to that
particular OEM Modality to be connected, ie the OEM console Settings and the printer Image Quality
parameters.

Therefore, as second and final step, setup the printer as per the indications reported here into the Setup Manual
for that particular OEM Modality.
Note that the Setup Manual has 2 sections ie the DryView section and the Wet section.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

What’s inside this document:


n OEM Modality Setup (ie how to configure,into the OEM console,the Kodak printer) 2
n DryView correspondance with Kodak Setup Catalog for DryView cameras................ 3
n Image Quality setup into MIM (or 9410) for DryView cameras ............................... 4
n More info on OEM Modality when connected to DryView cameras .............................. 5
n Wet correspondance with Kodak Setup Catalog for Wet cameras .................................. 6
n Image Quality setup into MIM (or 190) for Wet cameras ............................................... 7
n More info on OEM Modality eg Page-Mode and IQ parameters sent out by OEM ........ 8
n Miscellaneous: Deviations from Kodak Setup Catalog / Known Problems / Notes........ 9

OEM Modality Setup


ie how to configure into the OEM console the Kodak printer:

Note: MIM setup is basically also valid for 9410 by using AE=9410, Port=1024
Note: for each DryView camera select into the Modality the correspondent Prototype,
in particular for 8900 choose the Kodak 8900 standard,
in particular for CMI1000 choose the Kodak CMI1000 or Codonics 1660MD.

iU22 Setup:
n on keyboard sliding out from below iU22 console press Setups-button
n by pressing Enter-button go into PrintNetwork/Global Config./Devices
n click New Device
n into Step1 select Device Type=DICOM BW Printer
n into Step2 select Printer Type=Kodak-DryView-8150 (or 8200)-DICOM and set:
n Name=DryView, AE Title=DV8150, Port#=5040, IP Addr.=accordingly
n click OK
n by pressing Enter-button go into PrintNetwork/Device Selection/Printer
n highlight DryView and click Edit and set:
n Copies=1, Size=14x17, MediumType=Blue Film, Format=2x3, Orient.=Portrait
n click OK
n under Associate to Print select DryView
n at the bottom click Done to end the configuration

Author: gian.valesano@kodak.com, Edition: May ´05 2


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

DryView section:
Setup for Kodak DryView cameras ie
8900, 8700, 8610, 8600, 8500, 8300, 8200, 8150, 8100, 1000

Correspondance with Kodak Setup Catalog for DryView cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

MIM (inside printer) 8900 8150


into Kodak catalog 19 20
see Setup as per page
MIM sw version 5.6.b5 5.7.b8
Printer sw version Rev.E ie 2.11 1.0.2.b04

MIM external 8700, 8500 8100 8300 8610 (8600) 8200 1000
into Kodak catalog 15 16 17 18 14 21
see Setup as per page
MIM sw version 6.1 6.1 6.1 6.1 6.1 2.0.b3
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5) 4.9.1 1.0.b8

9410 external 8700, 8500 8100 8300 8610 (8600)


into Kodak catalog 7 8 9 10
see Setup as per page
9410 sw version 1.6 1.6 1.6 1.6
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5)

Author: gian.valesano@kodak.com, Edition: May ´05 3


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 9410) for DryView cameras:

MIM (or 9410) 8900, 8150, 8700, 8300 8610, 8600


Parameters 8500, 8200, 8100, (see Note2 below,
1000 confirm Y/N)
(see Note1 below)
TFT/ULUT Pmsv3k.w87 u1_109s.w8u (gsdfaa0p.w8u)
(8610 Modality/Contrast Group) 6
(Always Apply=Y/N) (Y) (Y)
Contrast 9 10 3
(Always Apply=Y/N) (Y) (Y) (Y)
Max.Density 300 300 350
(Always Apply=Y/N) (Y) (Y) (Y)
Smoothing 0 1 1
(Always Apply=Y/N) (Y) (Y) (Y)
Presentation ie PLUT=Y/N N N N
(not available on 9410 )
8610 Convolution / Kernel Group Not available Not available 1
8610 Gamma Table Not available Not available Transmittance

Note1: in particular for 8100, 8200, 8150 use TFT=Pmsb.w87.

Note2: OEM Modality tested also with 8300 standalone Yes


ie without MIM or 9410 in front? (Yes/No, Mod.Gr.=0, D=16, C=10, smooth=1
parameters)

Author: gian.valesano@kodak.com, Edition: May ´05 4


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to DryView cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg TFT=Wrkstn2a.w87, Dmax=3.00, C=2.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for Value sent out Value sent out
Parameters Modality 8900, 8150, 8700, for 8300 for 8610, 8600
(Yes/No) 8500, 8200, 8100,
1000
Config. Information Yes LUT=0,9 LUT=0,10 LUT=0,3
(LUT=TFT,Contrast)
Max. Density Yes 300 300 300
Magnification Yes NONE NONE NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 5


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Wet section:
Setup for Kodak Wet cameras ie 160, 2180, 1120, 190

Correspondance with Kodak Setup Catalog for Wet cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

160 2180 1120 190


into Kodak catalog 11 12 13 6
see Setup as per page
MIM sw version 6.1 6.1 6.1 Not present
Printer sw version 7.2 2.17H 2.40 / 032 1.7.29

Note: in regards to 969 parameters use the same as per 8700 setup.

Author: gian.valesano@kodak.com, Edition: May ´05 6


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 190) for Wet cameras:

MIM (or 190) Parameters 160, 2180, 1120, 190


(see Note below)
Curve Shape 3
(Always Apply=Y/N) (Y)
Contrast 0
(Always Apply=Y/N) (Y)
Max.Density 300
(Always Apply=Y/N) (Y)
PLUT=Y/N N
(not available on 190 )

Note: reported setup was tested eg for 160 and tentatively is valid also for other cameras ie
2180, 1120, 190.

Author: gian.valesano@kodak.com, Edition: May ´05 7


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to Wet cameras:

Works the OEM Modality in Page-Mode? (Yes/No) Yes


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg CurveShape=0,Dmax=3.00,Contrast=0.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for
Parameters Modality 160, 2180, 1120,
(Yes/No) 190
Config. Information Yes CS435\CN0
(CurveShape\Contrast)
Max. Density Yes 300
Magnification Yes NONE
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 8


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Miscellaneous:

Deviations in respect to the printer setup reported into Kodak Catalog:

e.g. disable N-Event ie NeventReports=No (Always=Y)

Known Problems or Limitations:

Notes:

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

Setup for Philips workstation EasyVision DX (Sectra)


with LASER IMAGERS

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

Philips EasyVision DX (Sectra) via DICOM with


Kodak DryView 8300 LASER IMAGER
DICOM Network

Philips EasyVision DX
Workstation

Kodak DryView 8300


Requirements: EasyVision DX v8.2
Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8300 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8300_8x10 {
node <IP-address accordingly> # Must be configured
port <104> # Must be configured
Label <8300> # Must be configured
called_app_title <8300>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <8INX10IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=0,10>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <2256>
Size <191:227>
}

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0
n on 8300 Service Menu into Configure User set: Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

Philips EasyVision DX (Sectra) via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER
DICOM Network

Philips EasyVision DX Kodak 190 camera


Workstation

Requirements: EasyVision DX v8.2


Requirements: 190 rel. 1.7.29

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 190 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_190_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <190> # Must be configured
called_app_title <190>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <CS435\CN0>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <327:409>
}

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips EasyVision DX
Fiber Cable
Workstation

Fiber Cable

Kodak 9410 PACS Link


Kodak DryView 8700 Plus
Requirements: EasyVision DX v8.2
Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8700 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8700_14x17 {
node <IP-address accordingly> # Must be configured
port <1024> # Must be configured
Label <8700> # Must be configured
called_app_title <9410>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsv3k.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <319:407>

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips EasyVision DX
Workstation

Data 37-pin

Kodak DryView 8100


Kodak 9410 PACS Link
Requirements: EasyVision DX v8.2
Requirements: 9410 v1.6, 8100 v2.4.1

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8100 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8700_14x17 {
node <IP-address accordingly> # Must be configured
port <1024> # Must be configured
Label <8100> # Must be configured
called_app_title <9410>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsb.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <319:406>
}

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952


n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

Philips EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Kodak DryView 8300


Control 9 pin
Philips EasyVision DX COPCIL board
Workstation Copper Cable

Data 37 pin

Kodak 9410 PACS Link


Requirements: EasyVision DX v8.2
Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8300 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8300_8x10 {
node <IP-address accordingly> # Must be configured
port <1024> # Must be configured
Label <8300> # Must be configured
called_app_title <9410>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <8INX10IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=0,10>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <2256>
Size <191:227>
}

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even


n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips EasyVision DX
Workstation Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200


Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, KELI 160 v7.2

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 160 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_160_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <160> # Must be configured
called_app_title <160>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <CS435\CN0>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4360>
Size <340:417>
}

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n Subnet Mask=accordingly, Gateway Addr=accordingly


n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
DICOM Network

Philips EasyVision DX
Kodak 2180 camera
Workstation

Fiber Cable

Kodak MIM 200


Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, 2180 rel. 2.17H

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 2180 printer
n (note that per each filmsize a separate printer must be configured, ie
n 14x14: Resolution <4096>, Size <323:326>
n 11x14: Resolution <3236>, Size <255:326>
n 8x10: Resolution <2286>, Size <180:224>).
n copy and paste an existing printer and set parameters as per below example
KODAK_2180_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <2180> # Must be configured
called_app_title <2180>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <CS435\CN0>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <323:407>
}

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips EasyVision DX
Kodak 1120 camera
Workstation

Fiber Cable

Kodak MIM 200


Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 1120 printer
n (note that per each filmsize a separate printer must be configured, ie
n 14x14: Resolution <4096>, Size <327:328>
n 11x14: Resolution <4096>, Size <327:258>
n 8x10: Resolution <2304>, Size <184:232>).
n copy and paste an existing printer and set parameters as per below example
KODAK_1120_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <1120> # Must be configured
called_app_title <1120>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <CS435\CN0>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <327:410>
}

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER
DICOM Network

Philips EasyVision DX Crossed


Workstation Network
Cable

Kodak MIM 200 Kodak DryView 8200


Requirements: EasyVision DX v8.2
Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8200 printer
n (note that per each filmsize a separate printer must be configured, ie
n 14x14: Resolution <4360>, Size <340:331>
n 11x14: Resolution <3272>, Size <255:340>).
n copy and paste an existing printer and set parameters as per below example
KODAK_8200_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8200> # Must be configured
called_app_title <8200>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsb.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4360>
Size <340:407>
}

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off
8200 Setup on CONFIGURATION Menu:
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
DICOM Network

iew
00 DryV
3M 87

Philips EasyVision DX
Fiber Cable
Workstation

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, 8700 v1.34
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8700 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8700_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8700> # Must be configured
called_app_title <8700>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsv3k.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <319:407>
}

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips EasyVision DX
Workstation

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8100 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8700_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8100> # Must be configured
called_app_title <8100>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsb.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4096>
Size <319:406>
}

MIM200 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips EasyVision DX COPCIL board
Workstation Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300


Requirements: EasyVision DX v8.2
Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B
Note: This setup is basically also valid for 25 PS computer

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8300 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_8300_8x10 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8300> # Must be configured
called_app_title <8300>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <8INX10IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=0,10>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <2256>
Size <191:227>
}

MIM200 Setup:
n on Configure, Network:

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n Node Name=MIM, IPAddr=accordingly


n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips EasyVision DX
Workstation

Kodak DryView 8900


Requirements: EasyVision DX v8.2
Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8900 printer
n (note that per each filmsize a separate printer must be configured, ie
n 14x14: Resolution <4488>, Size <350:350>
n 11x14: Resolution <3514>, Size <274:350>
n 10x12: Resolution <3184>, Size <248:299>
n 8x10: Resolution <2528>, Size <197:248>).
n copy and paste an existing printer and set parameters as per below example
KODAK_8900_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8900> # Must be configured
called_app_title <8900>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsv3k.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4488>
Size <350:427>
}

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=0 (Always=Y), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips EasyVision DX
Workstation

Kodak DryView 8150


Requirements: EasyVision DX v10.1
Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure 8150 printer
n (note that per each filmsize a separate printer must be configured, ie
n 14x14: Resolution <4438>, Size <346:344>
n 11x14: Resolution <3440>, Size <268:346>).
n copy and paste an existing printer and set parameters as per below example
KODAK_8150_14x17 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <8150> # Must be configured
called_app_title <8150>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <14INX17IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=Pmsv3k.w87,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <4438>
Size <346:421>
}

8150 Setup:

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n on Configuration, System, Network:


n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=0 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

EasyVision DX (Sectra) via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips EasyVision DX
Workstation

Kodak CMI 1000


Requirements: EasyVision DX v10.1
Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

EasyVision DX (Sectra) Setup:


n into C:\Program Files\Philips\EasyVision\config\local_config\_all_\general\def
n very carefully edit the file dicom_printer.def to configure CMI1000 printer.
n copy and paste an existing printer and set parameters as per below example
KODAK_1000_8x10 {
node <IP-address accordingly> # Must be configured
port <5040> # Must be configured
Label <1000> # Must be configured
called_app_title <CMI1000>
copies <1>
priority <MED>
medium_type <BLUE FILM>
film_destination <PROCESSOR>
film_orientation <PORTRAIT>
film_size <8INX10IN>
magnification_type <CUBIC>
max_density <300>
configuration_information <LUT=0,9>
border_density <BLACK>
empty_image_density <BLACK>
trim <NO>
polarity <NORMAL>
Resolution <2508>
Size <203:239>
}

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips workstation EasyVision DX (Sectra) with LASER IMAGER Products

n under More Settings: DICOM Port=5040


n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=2 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Setup Philips workstation EasyVision/EasyRAD with LASER IMAGERS


(note: models EasyDiagnost, MultiDiagnost, OmniDiagnost, Eleva could
have either the setup of Philips DSI or the setup of Philips EasyVision)

Connection POINT-TO-POINT:
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8700 LASER IMAGER / Dual... 3
n Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual .................................................................................. 4
n Kodak DryView 8700 LASER IMAGER / Plus......................................................................................... 5
n Kodak DryView 8700 LASER IMAGER / Standard ................................................................................. 6
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8500 LASER IMAGER / Dual... 7
n Kodak DryView 8500 LASER IMAGER / Plus......................................................................................... 8
n Kodak DryView 8500 LASER IMAGER / Standard ................................................................................. 9
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8300 LASER IMAGER / Dual... 10
n Kodak DryView 8300 LASER IMAGER / Standalone.............................................................................. 11
n Kodak DryView 8300 LASER IMAGER / Standard via SCSI .................................................................. 12
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8700 LASER IMAGER / Dual /
Kodak DryView 8300 LASER IMAGER / Dual........................................................................................ 13
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8700 LASER IMAGER / Dual /
Kodak DryView 8500 LASER IMAGER / Dual........................................................................................ 14
n Kodak DryView 8800 MULTI-INPUT MANAGER / Kodak DryView 8500 LASER IMAGER / Dual /
Kodak DryView 8300 LASER IMAGER / Dual........................................................................................ 15
n Kodak Digital Science 969 HQ LASER IMAGER ................................................................................ 17
n Kodak Digital Science 969 HQT LASER IMAGER ............................................................................. 18
n Kodak Digital Science 959 XL LASER IMAGER ................................................................................ 19
n Kodak DryView 8100 LASER IMAGER................................................................................................... 20
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) ( 1) ......................................................... 21
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER ( 1).............................................................................................. 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER ( 2,3,4) ........................................................................................... 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) ........................................................................................ 27
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) ........................................................................................ 29
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
4
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) ( )................................................ 31
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
4
Kodak DryView 8100 LASER IMAGER ( )............................................................................... 33
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
4
Kodak DryView 8300 LASER IMAGER ( )............................................................................... 35
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
4
Kodak DryView 8200 LASER IMAGER ( )............................................................................... 37

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER................................................................................................... 39


n Kodak Digital Science 190 MEDICAL LASER PRINTER................................................................... 40
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) ( 1) ......................................................... 41
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual ( 1)............................................................................. 43
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER ( 1).............................................................................................. 45
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER ( 1).............................................................................................. 47
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600) ( 1) ...................................................................................................................................... 49
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER ( 2,3,4) ........................................................................................... 51
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) ........................................................................................ 53
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) ........................................................................................ 55
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak 3600 DESKTOP MEDICAL IMAGER ( 2,3,4) ................................................................................ 57
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
3,4,5
Kodak DryView 8200 LASER IMAGER ( ) .......................................................................... 59
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
4,5
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ) ( ).............................................. 61
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8800 MULTI-INPUT MANAGER /
4,5
2 x Kodak DryView 8700 LASER IMAGER / Dual ( ) .............................................................. 63
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
4,5
Kodak DryView 8100 LASER IMAGER ( )............................................................................. 65
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
4,5
Kodak DryView 8300 LASER IMAGER ( )............................................................................. 67
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
4,5
(see also 8600) ( ) .............................................................................................................. 69
n Kodak DryView 8900 LASER IMAGER................................................................................................... 71
n Kodak DryView 8150 LASER IMAGER................................................................................................... 73
n Kodak Color Medical Imager 1000 ........................................................................................................... 75

5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips
EasyVision / EasyRAD Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
anag
putM
i-In
Mult
8800
3M

iew
00 DryV
3M 87

Fiber Cable
8700 Dual

8800 MIM

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 rel. 4.2, 8700 rel. 1.15

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual
Philips
EasyVision / EasyRAD Philips 37-pin RS422 37-to-9 pin
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
anag
putM
i-In
Mult
8800
3M

iew
00 DryV
3M 87

Fiber Cable
8700 Dual
iew
00 DryV
3M 87

Fiber Cable

8800 MIM
8700 Dual

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 rel. 5.1, 8700 rel. 1.22
EasyVision / EasyRAD Setup (so the 2 cameras are treated independently):
n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface type=PCI/SBus, Printer type=3M 8700, Name=dv8700nbr1
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n into Advanced Printer Properties: Shared Interface=On, Dest. Nbr=0, Proceed, Exit
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface type=PCI/SBus, Printer type=3M 8700, Name=dv8700nbr2
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n into Advanced Printer Properties: Shared Interface=On, Dest. Nbr=1
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle
Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
8800/8700 Dual Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8700 Plus

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8700 Plus rel. 1.15

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnif.=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8700 Standard rel. 1.15

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnification=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

ger
Mana
nput
ulti-I
0M
880
3M

yView
00 Dr
3M 87

Fiber Cable
8500 Dual

8800 MIM

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 rel. 5.1, 8500 rel. 1.22

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8500, Name=dv8500
n into Config. Media List: 11x14 G8=On (8-bit-print), 11x14 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

yView
00 Dr
3M 87

8500 Plus

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8500 Plus rel. 1.22

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8500, Name=dv8500
n into Config. Media List: 11x14 G8=On (8-bit-print), 11x14 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnif.=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT
iew
00 DryV
Philips 37-pin 3M 87

Data Cable
I mage Data OUT

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8500 Standard rel. 1.22

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8500, Name=dv8500
n into Config. Media List: 11x14 G8=On (8-bit-print), 11x14 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnif.=No
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
nag
ut Ma
lti-Inp
00 Mu

Fiber Cable 3M88

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 rel. 5.1, 8300 rel. 4.2

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8300 Portrait, Name=dv8300
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8300 Setup:
n on 8300 Operator Menu set: Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None, Modality Group=0

8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips EasyVision
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


I mage Data OUT

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8300 rel. 4.2

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8300 Portrait, Name=dv8300
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Odd
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via SCSI with


Kodak DryView 8300 LASER IMAGER

Philips
SCSI Cable to be
EasyVision/EasyRAD
connected in P4 P4
SCSI OUT

SCSI Termination
to be plugged in P5
P5

Requirements: EasyVision 2.3 or EasyRAD 2.3


Requirements: 8300 v4.9.1, 8300 LUT v3.0.3B, 8300 memory=16 MB

Note: inside 8300 on SCSI Module a short flat cable


must be installed between 2 connectors
Note: to set Density and Contrast on 8300 SCSI Module, due to a sw restriction,
a 8300 MO driver is needed, which could be returned after proper setup
Note: please use on 8300 sw version 4.9.1, otherwise a white sheet film
will be printed sporadically.

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=SCSI, SCSI Bus Printer Type=3M 8300 SCSI Portrait, Name=dv8300
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8300 Setup:
n on 8300 Operator Menu into User Nbr. & Params (User nbr.=2) set:
n Nbr. B&W copies=1, Nbr. Color copies=1
n Density=16, Contrast=10, Sharp/Smooth=1
n Polarity=Positive, Print Options=Print All
n B&W Format=1 on 1, Color Format=1 on 1
n Film Numbering=Off, Autoprint=On, Print Aspect Ratio=1.00
n on 8300 Service Menu into Configure User set:
n SCSI Interface: Role=Target, Unit ID=4, Bus Termination=On
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual /
Kodak DryView 8300 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable
9-to-9-pin Copper Cable
Fiber Cable
37-pin Digital Cable POEIB nag
er
ut Ma
lti-Inp
0 Mu
880
3M

Fiber Cable

iew
0 DryV
3M 870
8300 Dual

8800 MIM

8700 Dual

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 v.5.1, 8700 v.1.22, 8300 v.4.2

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M MIM, Name=dv8800
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8300 Setup:
n on 8300 Operator Menu set: Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu: Serial: Baud=1200, Data=8, Stop=1, Parity=Even, Protocol=3M
n on 8300 Service Menu set: Digital Interface: Parity=None, Modality Group=0

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correct=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable
Fiber Cable
8700 Dual 3M 87
00 Dr
yView

ger
Mana
nput
ulti-I
0M
880
3M

Fiber Cable

yView
00 Dr
3M 87

8800 MIM

8500 Dual

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8800 v.5.1, 8700/8500 v.1.22

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M MIM, Name=dv8800
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8800/8700 Dual/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correction=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual /
Kodak DryView 8300 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable
9-to-9-pin Copper Cable
Fiber Cable
37-pin Digital Cable POEIB nag
er
ut Ma
lti-Inp
0 Mu
880
3M

Fiber Cable

iew
0 DryV
3M 870
8300 Dual

8800 MIM

8500 Dual

Requirements EasyVision 4.2 or EasyRAD 2.3, 8800 v.5.1, 8500 v.1.22, 8300 v.4.2

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M MIM, Name=dv8800
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0

8800/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu: Pxl.Correct=No, Scl.Format=No, Scl.Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak Digital Science 969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M H Q L a ser Im ag er

M969 HQ

Requirements: EasyVision 4.2 or EasyRAD 2.3, M969 HQ rel. 3.0

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M HQ 969, Name=hq969
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST: Pixel Correction=No, Scale Format=No, Scale Magnif.=No, Slides=Super
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT 3 M H Q L a s er Im a g e r

Philips 37-pin
Data Cable
I mage Data OUT

Requirements: Philips EasyVision rel. 4.2, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M HQ 969, Name=hqt969
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=No, Scale Format=No, Scale Magnification=No
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak Digital Science 959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
I mage Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a s e r I m a g er

M959 XL

Requirements: EasyVision 4.2 or EasyRAD 2.3, M959 XL rel. 3.8

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 959 XL, Name=xl959
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup:
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu: 959 Default
n on EIB Menu: Pixel=8, Header/Line, Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak DryView 8100 LASER IMAGER

DryView 8100

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips EasyVision Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

Requirements: EasyVision 4.2 or EasyRAD 2.3, 8100 v1.1

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8100
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=180, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ)

Philips EasyVision Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00 DryV
iew Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: EasyVision 4.2 or EasyRAD 2.3, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into 9410 Local Printers and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD Camera Config.File

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=ev, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips EasyVision Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: EasyVision 4.2 or EasyRAD 2.3, 9410 v1.6, 8100 v2.4.1

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8100
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=22
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=ev, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=Off


n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14, Bit=8

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.3


Requirements: MIM 200 v6.1, KELI 160 v7.2

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELI-160 (or MLP-190), Name=k160
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 2180 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.3


Requirements: MIM 200 v6.1, KELP 2180 v2.17H

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELP 2180, Name=k2180
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 1120 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.3


Requirements: MIM 200 v6.1, KELP 1120 v2.40/032

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=Kodak KELP 1120, Name=k1120
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ )

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable 00 DryV
iew
3M 87

Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: EasyVision 4.2 or EasyRAD 2.3, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into MIM 200 Destinations and in case of 8500 and HQ
by installing 8500 and HQ into EasyVision/EasyRAD Camera Config.File

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8700, Name=dv8700
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=MOV
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips EasyVision Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: EasyVision 4.2 or EasyRAD 2.3, MIM200 v6.1, 8100 v2.4.1

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8100 (or 8700), Name=dv8100
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=MOV
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board Control 9-pin
Philips EasyVision Control Adaptor Cable
Control Cable Copper Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8300

Requirements: EasyVision 4.2 or EasyRAD 2.3


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8300 Portrait, Name=dv8300
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=MOV
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8300, Media Size Select=8x10in
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=10
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=u1_109b.w8u, Auto Study=Off, Console=On, Annotation=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips EasyVision Control Adaptor Cable
Control Cable
Crossed
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200 Kodak DryView 8200

Requirements: EasyVision rel. 5.1, MIM 200 v6.1, 8200 v3.5

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=PCI/SBus, Printer=3M 8200, Name=dv8200
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=EasyVision Wkst.:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=ev,
n Mod.Name=ev, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2048, Parity=Odd, Timer=59, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=MOV
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8200, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=0, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 38


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300

Philips EasyVision

Requirements: EasyVision or EasyRAD v4.2, 8300 v4.9.1, 8300 DICOM board v3.3

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8300, port=104, AETitle=8300
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=4K or 16K,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=8300
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16
n Contrast=10
n Sharp/Smooth=1
n Film Layout : Format=1 on 1, Mode=Portrait
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 39


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips EasyVision Kodak 190 camera

Requirements: Philips EasyVision rel. 4.2 or EasyRAD rel. 4.2, 190 rel. 1.7.29

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 190, port=5040, AETitle=190
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak MLP-190, Name=k190,
n AE title=190, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 40


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ)

DICOM Network

Kodak DryView 8700 Plus

Philips EasyVision
Fiber Cable
iew
00 DryV
3M 87

Fiber Cable

Kodak 9410 PACS Link

Requirements: EasyVision rel. 4.2 or EasyRAD rel.4.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into 9410 Local Printers and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD DICOM Camera Config.File
(if HQ DICOM Camera Config.File not yet available it will also work with 8700)

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8700, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 41


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 42


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual

DICOM Network

8800 MIM 8700 Dual

Philips EasyVision
Fiber Cable 00 DryV
iew
nag
er 3M87
ut Ma
lti-Inp
Mu
8800

Fiber Cable
3M

Fiber Cable
iew
00 DryV
3M 87

9410 PACS Link Fiber Cable

8700 Dual

Requirements: EasyVision 4.2 or EasyRAD 4.2, 9410 v1.6, 8800 v5.2, 8700 v1.34

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=8700nbr1, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1025, AETitle=9411
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed

Author: gian.valesano@kodak.com, Edition: May ´05 43


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=8700nbr2, AE title=9411
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700_1, Model=8800/8700 Dual, Output=Fiber, 8800 Setup=Printer1
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700_1, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700_1, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700_2, Model=8800/8700 Dual, Output=Fiber, 8800 Setup=Printer2
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700_2, AE Title=9411, Port=1025, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8700_2, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

Author: gian.valesano@kodak.com, Edition: May ´05 44


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

8800/8700 Dual Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 45


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips EasyVision

Data 37-pin

9410 PACS Link


DryView 8100

Requirements: EasyVision rel. 4.2 or EasyRAD rel. 4.2, 9410 v1.6, 8100 v2.4.1

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8100, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title=9410, Port=1024, Max Associations=5
on Device Setup, SCP PRINTERS, Registered Users:

Author: gian.valesano@kodak.com, Edition: May ´05 46


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n SCU AETitle=accordingly, SCU IPAddr=accordingly


n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 47


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300
Control 9 pin
Philips EasyVision COPCIL board
Copper Cable

Data 37 pin

9410 PACS Link

Requirements: EasyVision rel. 4.2 or EasyRAD rel.4.2


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=9410
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title=9410, Port=1024, Max Associations=5

Author: gian.valesano@kodak.com, Edition: May ´05 48


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n on Device Setup, SCP PRINTERS, Registered Users:


n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No, Print Always=Yes
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 49


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips EasyVision COPCIL board
Copper Cable

Data 37 pin

Kodak 9410 PACS Link

Requirements: EasyVision rel. 5.1 or EasyRAD rel.4.2


Requirements: 9410 v1.6, 8610 v2.4 (if 8600 v2.5)

Note: This setup is basically also valid for 8600 camera

EasyVision Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8610, Name=dv8610, AE title=9410
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print) (Off if 8600)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 50


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8610, Model=8600/8610, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=12 (8 if 8600), Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=1, Installed TFT/ULUT=don’t care
n Max.Density=300, Contrast=5, Border Density=0
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8610, AE Title=9410, Port=1024, Max Associations=5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle=accordingly, SCU IPAddr=accordingly
n SCP=8610, Copies=1, Contrast Table=5, Max.Density=300
n Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Gamma Table=Linear in Transmittance, Contrast Group=1, Kernel Group=1
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma Table=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 51


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips EasyVision
Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.2


Requirements: MIM200 v6.1, KELI 160 v7.2

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=160
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELI-160 (or MLP-190), Name=k160
n AE title=160, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 52


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 53


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips EasyVision
Kodak 2180 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.2


Requirements: MIM200 v6.1, 2180 rel. 2.17H

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=2180
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELP2180 (or MLP-190), Name=k2180,
n AE title=2180, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 54


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 55


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips EasyVision
Kodak 1120 camera

Fiber Cable

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.3 or EasyRAD rel. 4.2


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=1120
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=Kodak KELP1120 (or MLP-190), Name=k1120,
n AE title=1120, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 56


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 57


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak 3600 DESKTOP MEDICAL IMAGER

DICOM Network

Philips EasyVision
Kodak DMI 3600 camera
Crossed
Network Cable

Kodak MIM 200

Requirements: Philips EasyVision rel. 4.2, MIM200 v6.1, DMI 3600 rel. 2.0
Important: set 3600 Media Menu correctly, respecting which size in which tray

EasyVision Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=3600
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer Type=Kodak DMIS DICOMRGB
n Name=DMI3600, AE title=3600
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., KodakDMI3600, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM5
n on Configure, Destination, Add, Print, Direct, Qualified, DMI 3600:
n General: Log.Name=DMI3600, IPAddr=100.100.100.20
n General: Database Name=DMI3600, Model=DMI3600
n Media=11x14in, 3128x3854, 3778, Film

Author: gian.valesano@kodak.com, Edition: May ´05 58


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Media=8.5x11in, 2436x3032, 2957, Paper


n Media=8x10in, 2228x2654, 2579, Film
n Media=4x6, 1084x1596, 1521, Paper
n Media=A4, 2380x3242, 3164, Paper
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Pivot=On, Print Quality=On, Rotation=On
n Interpolation: No Magn.=On, Bilin=On, Repl=On, Cubic=On
n Format Types: Custom=On, Row Symmetric=On, Slide Format=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=5, Min Vert Sep=5
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.8111
n Image Processing: Color=On, Color/Grayscale Same Page=On
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=1.7
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape=3 i.e., 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=On

DMI 3600 Setup: press Menu:


SUPPLIES MENU =Supplies ok
COLOR MENU: Document Type=Medical Image, Print Quality=High 1200dpi
MEDIA MENU: Media Size: Upr Tray=8x10, Lwr Tray=A4
MEDIA MENU: Media Type: Upr Tray=Medical Film, Lwr Tray=Medical Paper
MEDIA MENU: Media Source: Upr Tray / Lwr Tray (set on which media will be printed)
FINISHING MENU: Separator Sheets=None
SETUP MENU: Jam Recovery=Auto, Print Area=Normal, Ink Alert=Off
POSTSCRIPT MENU: Print PS Error=Off
NETW.MENU: NETW.OPT.1: PCL Smart Switch=Off
NETW.MENU: NETW.OPT.1: PS Smart Switch=Off
NETW.MENU: NETW.OPT.1: NPA Mode=Auto
NETW.MENU: NETW.OPT.1: Network Buffer=Auto
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: Active=Yes
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: Enable BOOTP=No

Author: gian.valesano@kodak.com, Edition: May ´05 59


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: Enable DHCP=No


NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: Enable RARP=No
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: IP Address =100.100.100.20
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: IP Netmask=255.255.255.0
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: TCP/IP: IP Gateway=100.100.100.1
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: NETWARE: Active=No
NETW.MENU: NETW.OPT.1: NETW.1 SETUP: Apple Talk: : Active=No

Author: gian.valesano@kodak.com, Edition: May ´05 60


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Kodak DryView 8200

Philips EasyVision

Crossed Network Cable

Kodak MIM 200

Requirements: EasyVision rel. 5.1 or EasyRAD rel. 4.2


Requirements: MIM200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8200
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM,Printer=3M 8200 (or 8700),Name=k8200,AE title=8200
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 61


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:


n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space=On, Variable Border=On, PLUT Service=On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smoothing=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 62


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual, 8500 Plus, 8800/8500 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips EasyVision
Fiber Cable

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: EasyVision rel. 4.2 or EasyRAD rel.4.2, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual, 8500 Plus,
8800/8500 Dual and HQ cameras by basically using the correspondent
Model into MIM200 Destination and in case of 8500 and HQ by installing
8500 and HQ into EasyVision/EasyRAD DICOM Camera Config.File
(if HQ DICOM Camera Config.File not yet available it will also work with 8700)

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8700
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8700, Name=dv8700, AE title=8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

Author: gian.valesano@kodak.com, Edition: May ´05 63


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 64


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8800 MULTI-INPUT MANAGER /
2 x Kodak DryView 8700 LASER IMAGER / Dual
DICOM Network
Kodak DryView 8700 Dual

Kodak 8800 MIM


Philips
Fiber Cable
EasyVision / EasyRAD Mu
lti-I
npu
t Ma
nager
3M 87
00 DryV
iew

8800

Fiber Cable
3M

Fiber Cable
iew
00 DryV
3M 87

Kodak MIM 200


Fiber Cable

Kodak DryView 8700 Dual

Requirements: EasyVision 4.2 or EasyRAD 4.2


Requirements: MIM200 v6.1, 8800 v6.1, 8700 v1.34
Note: This setup is basically also valid for 25 PS computer

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8700A
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8700, Name=8700nbr1, AE title=8700A
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8700B
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed

Author: gian.valesano@kodak.com, Edition: May ´05 65


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8700, Name=8700nbr2, AE title=8700B
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup to program first 8700 destination:


n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8800/8700, Set:
n General: Log.Name=8700A
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Output Printer=1
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=0 (Always=N), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

Author: gian.valesano@kodak.com, Edition: May ´05 66


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

MIM200 Setup to program second 8700 destination:


n on Configure, Destination, Add, Print, Direct, Qualified, 8800/8700, Set:
n General: Log.Name=8700B
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Output Printer=2
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=0 (Always=N), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8800/8700 Dual Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 67


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips EasyVision

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: EasyVision rel. 4.2 or EasyRAD rel. 4.2, MIM200 v6.1, 8100 v2.4.1
Note: This setup is basically also valid for 25 PS computer

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8100
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8100 (or 8700),
n Name=dv8100, AE title=8100
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32

Author: gian.valesano@kodak.com, Edition: May ´05 68


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Media Info: Media=35x43cm, 4096x5223, 5133, Blue


n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 69


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips EasyVision COPCIL board
Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: EasyVision rel. 4.2 or EasyRAD rel.4.2


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B
Note: This setup is basically also valid for 25 PS computer

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Configured Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8300
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration i.e., do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Configured Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer=3M 8300, Name=dv8300, AE title=8300
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32

Author: gian.valesano@kodak.com, Edition: May ´05 70


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Media Info: Media=8x10in, 2256x2676, 2596, Blue


n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.811
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT=u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 i.e., Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 71


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips COPCIL board
EasyVision / EasyRAD Copper Cable

Data 37 pin

Kodak MIM 200


Requirements: EasyVision rel. 5.1 or EasyRAD rel. 4.2 Level 4
Requirements: MIM200 v6.1, 8610 v2.4 (if 8600 v2.5)
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8600 camera

EasyVision Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8610
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8610, Name=dv8610, AE title=8610
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8610 (or 8600), Set:

Author: gian.valesano@kodak.com, Edition: May ´05 72


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n General: Log.Name=8610 (or 8600)


n General: Database=8610 (or 8600), Model=8610 (or 8600), Memory=32
n Media Info: Media=8x10in, 5025x6200, 6020, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12 (8 if 8600), Header/Line,Transfer=2000,Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=25.8065
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.5
n M952 Usage: Min H/V Mag=0.26, Max H/V Mag=16, Min Dmax 952=2
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= upmsv3kp.w8u (Always=N), Contrast (dry)=5 (Always=N)
n Modality Group=1, Conv.Group=1, Gamma Table=Linear in Transmittance
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=1 (Always=Y), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None (Always=N)
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 73


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips
EasyVision / EasyRAD

Kodak DryView 8900

Requirements: EasyVision v6.1 Level 2 or EasyRAD v4.2 Level 5


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8900, port=5040, AETitle=8900
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface= DICOM, Printer= Kodak 8900 standard
n (or 3M / Kodak 8900 and in this case 2mm Border=No, or 3M 8200),
n Name=dv8900, AE title=8900
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any

Author: gian.valesano@kodak.com, Edition: May ´05 74


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Media Types: Media=35x35cm,9100x9087, 8831, Any


n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=0 (Always=Y), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ev

Author: gian.valesano@kodak.com, Edition: May ´05 75


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips
EasyVision / EasyRAD

Kodak DryView 8150

Requirements: EasyVision v5.2 Level 2 or EasyRAD v4.2 Level 5


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8150, port=5040, AETitle=DV8150
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface= DICOM, Printer=Kodak 8150 DICOM (or 3M 8200),
n Name=dv8150, AE title=DV8150
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=Off (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any

Author: gian.valesano@kodak.com, Edition: May ´05 76


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

n Media Types: Media=11x14in, 3440x4438, 4310, Any


n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=0 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ev

Author: gian.valesano@kodak.com, Edition: May ´05 77


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

Philips EasyVision / EasyRAD via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips
EasyVision / EasyRAD
workstation

Kodak CMI 1000

Requirements: EasyVision v5.2 Level 2 or EasyRAD v4.2 Level 5


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

EasyVision / EasyRAD Setup:


n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of CMI-1000, port=5040, AETitle=CMI1000
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Print SOP Classes set:
n Basic Greyscale=On, Basic Color=On
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface= DICOM,
n Printer= Kodak CMI-1000 DICOM (or Codonics-NP1660MD-DICOM),
n Name=CMI1000, AE title=CMI1000
n into Config. Media List:
n if Kodak CMI-1000 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, A4 Color RGB=On, all others Off
n if Letter inside CMI-1000 Paper Tray, A Color RGB=On, others=Off
n if Codonics-NP1660 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, set A4 P Col=On, others=Off
n if Letter inside CMI-1000 Paper Tray, set A P Col=On, others=Off
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):

Author: gian.valesano@kodak.com, Edition: May ´05 78


Philips workstation EasyVision / EasyRAD with LASER IMAGER Products

at Local Panel, under Menu, Imager Setup:


n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=N), TFT Set (From list)=Pmsv3k.w87 (Always=N)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=0 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 79


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Specific Setup for Philips workstation ENsphere


with LASER IMAGERS

OEM Data
Manufacturer Philips
Modality workstation
Model ENsphere
sw version v1.1
Implement. Class UID 1.2.276.0.7230010.3.0.3.4.1
Implement. Version Name OFFIS_DCMTK_341

Kodak Data
DICOM-interface MIM internal / external or 9410 as below reported
DICOM-intf sw version as below reported
Printer model All cameras as below reported
Printer sw version as below reported

What’s new:
The setup of a Kodak printer for a given OEM Modality was split in 2 separate documents:

- the generic Setup Catalog for Kodak Cameras


- the specific Setup Manual for a given OEM Modality

************* the generic Setup Catalog for Kodak Cameras:


the Setup Catalog for Kodak Cameras contains that part of a printer installation which is Generic, ie common to
all DICOM connections independently from the Specific modality connected.

Therefore, as first step, setup the printer as per the correspondent page of the Setup Catalog for Kodak Cameras
(ie go to the Setup Catalog doc and do that first, then come back
here and finish install as per this Specific Setup Manual doc).

************ the specific Setup Manual for a given OEM Modality:


the Setup Manual for a given OEM Modality contains that part of a printer installation which is Specific to that
particular OEM Modality to be connected, ie the OEM console Settings and the printer Image Quality
parameters.

Therefore, as second and final step, setup the printer as per the indications reported here into the Setup Manual
for that particular OEM Modality.
Note that the Setup Manual has 2 sections ie the DryView section and the Wet section.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

What’s inside this document:


n OEM Modality Setup (ie how to configure,into the OEM console,the Kodak printer) 2
n DryView correspondance with Kodak Setup Catalog for DryView cameras................ 3
n Image Quality setup into MIM (or 9410) for DryView cameras ............................... 4
n More info on OEM Modality when connected to DryView cameras .............................. 5
n Wet correspondance with Kodak Setup Catalog for Wet cameras .................................. 6
n Image Quality setup into MIM (or 190) for Wet cameras ............................................... 7
n More info on OEM Modality eg Page-Mode and IQ parameters sent out by OEM ........ 8
n Miscellaneous: Deviations from Kodak Setup Catalog / Known Problems / Notes........ 9

OEM Modality Setup


ie how to configure into the OEM console the Kodak printer:

Note: MIM setup is basically also valid for 9410 by using AE=9410, Port=1024

Ensphere Setup:
n enter PreferencesConfiguration//Administrator/Site Configuration
n into Printer set:
n Name=DryView, Type=DICM, Spool Directory=accordingly,
n Color=MONO2, IP Address=accordingly, port=5040, AETitle=DV8150
n Click Exit, click Yes
n When you print set Film Type=Blue Film, Filmsize=14x17, Magnification=Cubic

Author: gian.valesano@kodak.com, Edition: May ´05 2


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

DryView section:
Setup for Kodak DryView cameras ie
8900, 8700, 8610, 8600, 8500, 8300, 8200, 8150, 8100, 1000

Correspondance with Kodak Setup Catalog for DryView cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

MIM (inside printer) 8900 8150


into Kodak catalog 19 20
see Setup as per page
MIM sw version 5.6.b5 5.7.b8
Printer sw version Rev.E ie 2.11 1.0.2.b04

MIM external 8700, 8500 8100 8300 8610 (8600) 8200 1000
into Kodak catalog 15 16 17 18 14 21
see Setup as per page
MIM sw version 6.1 6.1 6.1 6.1 6.1 2.0.b3
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5) 4.9.1 1.0.b8

9410 external 8700, 8500 8100 8300 8610 (8600)


into Kodak catalog 7 8 9 10
see Setup as per page
9410 sw version 1.6 1.6 1.6 1.6
Printer sw version 1.34 2.4.1 3.5 2.4 (2.5)

Author: gian.valesano@kodak.com, Edition: May ´05 3


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 9410) for DryView cameras:

MIM (or 9410) 8900, 8150, 8700, 8300 8610, 8600


Parameters 8500, 8200, 8100, (see Note2 below,
1000 confirm Y/N)
(see Note1 below)
TFT/ULUT Pmsv3k.w87 u1_109s.w8u (gsdfaa0p.w8u)
(8610 Modality/Contrast Group) 6
(Always Apply=Y/N) (Y) (Y)
Contrast 9 10 3
(Always Apply=Y/N) (Y) (Y) (Y)
Max.Density 300 300 350
(Always Apply=Y/N) (Y) (Y) (Y)
Smoothing 2 2 1
(Always Apply=Y/N) (Y) (Y) (Y)
Presentation ie PLUT=Y/N N N N
(not available on 9410 )
8610 Convolution / Kernel Group Not available Not available 1
8610 Gamma Table Not available Not available Transmittance

Note1: in particular for 8100, 8200, 8150 use TFT=Pmsb.w87.

Note2: OEM Modality tested also with 8300 standalone Yes


ie without MIM or 9410 in front? (Yes/No, Mod.Gr.=0, D=16, C=10, smooth=2
parameters)

Author: gian.valesano@kodak.com, Edition: May ´05 4


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to DryView cameras:

Works the OEM Modality in Page-Mode? (Yes/No) No


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg TFT=Wrkstn2a.w87, Dmax=3.00, C=2.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for Value sent out Value sent out
Parameters Modality 8900, 8150, 8700, for 8300 for 8610, 8600
(Yes/No) 8500, 8200, 8100,
1000
Config. Information No
(LUT=TFT,Contrast)
Max. Density No
Magnification Yes CUBIC CUBIC CUBIC
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 5


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Wet section:
Setup for Kodak Wet cameras ie 160, 2180, 1120, 190

Correspondance with Kodak Setup Catalog for Wet cameras:

Note: the Kodak Setup Catalog contains that part of a printer installation which is Generic,
ie common to all DICOM connections independently from the Specific modality connected.
Therefore, as first step, setup the printer as per the correspondent page of the Kodak Setup
Catalog, and then, as second and final step, setup the Image Quality parameters as per the
Setup Manual for the Specific OEM Modality here below reported.

160 2180 1120 190


into Kodak catalog 11 12 13 6
see Setup as per page
MIM sw version 6.1 6.1 6.1 Not present
Printer sw version 7.2 2.17H 2.40 / 032 1.7.29

Note: in regards to 969 parameters use the same as per 8700 setup.

Author: gian.valesano@kodak.com, Edition: May ´05 6


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Image Quality setup into MIM (or 190) for Wet cameras:

MIM (or 190) Parameters 160, 2180, 1120, 190


(see Note below)
Curve Shape 3
(Always Apply=Y/N) (Y)
Contrast 0
(Always Apply=Y/N) (Y)
Max.Density 300
(Always Apply=Y/N) (Y)
PLUT=Y/N N
(not available on 190)

Note: reported setup was tested eg for 160 and tentatively is valid also for other cameras ie
2180, 1120, 190.

Author: gian.valesano@kodak.com, Edition: May ´05 7


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

More info on OEM Modality when connected to Wet cameras:

Works the OEM Modality in Page-Mode? (Yes/No) No


Note: see Page-Mode definition and matrixes into Kodak cameras setup catalog.

Are Image Data lutted by the OEM Modality, No


ie Linear-in-Density LUT into Camera? (Yes/No)
Note: Linea-in-Density LUT into Camera means eg CurveShape=0,Dmax=3.00,Contrast=0.

Which are the Image Quality parameters sent out by the OEM Modality:
DICOM Sent out by Value sent out for
Parameters Modality 160, 2180, 1120, 190
(Yes/No)
Config. Information No
(CurveShape\Contrast)
Max. Density Yes
Magnification Yes CUBIC
Smoothing No
PLUT No
Illumination (PLUT) No
Reflective Ambient No
Light (PLUT)

Author: gian.valesano@kodak.com, Edition: May ´05 8


Specific Setup Manual for a given OEM Modality via DICOM with Kodak cameras

Miscellaneous:

Deviations in respect to the printer setup reported into Kodak Catalog:

e.g. disable N-Event ie NeventReports=No (Always=Y)

Known Problems or Limitations:

Notes:

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation Integris 3D RA with LASER IMAGER Products

Setup for Philips workstation Integris 3D RA


with LASER IMAGERS

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) . 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER (1).................................................................... 8
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) ................................................................. 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (2,3,4) .............................................................. 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (2,3,4) .............................................................. 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (3,4,5)................................................................ 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(4,5) 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4,5).................................................................. 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (4,5)..................................................................22
n Kodak DryView 8900 LASER IMAGER......................................................................... 24
n Kodak DryView 8150 LASER IMAGER......................................................................... 26
n Kodak Color Medical Imager 1000 ................................................................................. 28

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
5
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips workstation Integris 3D RA with LASER IMAGER Products

Philips Integris 3D RA via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

DryView 8300

Philips 3D RA
Workstation

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Imation 8300 Generic Printer and set:
n Printer Name=8300 , IP Addr.=addr. of 8300, Port=104, AETitle=8300
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakDryView8300Generic Printer and set:
n Printer Name=8300 , IP Addr.=accordingly, Port=104, AETitle=8300
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=0, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=0 ie Auto
n Film Layout : Format= 1 on 1, Mode= Portrait
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips workstation Integris 3D RA with LASER IMAGER Products

Philips Integris 3D RA via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips 3D RA Kodak 190 camera


Workstation

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 190 rel. 1.7.29

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=190 , IP Addr.=addr. of 190, Port=5040, AETitle=190
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu, select KodakMLP190Generic Printer and set:
n Printer Name=190 , IP Addr.=accordingly, Port=5040, AETitle=190
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=NORMAL, Min.Density=20, Print Priority=HIGH
n Store parameters

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

8700 Plus

Philips 3D RA
Fiber Cable
Workstation 00 DryV
iew
3M 87

Fiber Cable

9410 PACS Link

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 9410 v1.6, 8700 v1.34

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only
Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Imation Generic Printer and set:
n Printer Name=8700 , IP Addr.=addr. of 9410, Port=1024, AETitle=9410
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu, select Kodak9410PACSGeneric Printer and set:
n Printer Name=8700 , IP Addr.=accordingly, Port=1024, AETitle=9410
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips workstation Integris 3D RA with LASER IMAGER Products

n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5


n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips 3D RA
Workstation

Data 37-pin

9410 PACS Link


DryView 8100

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 9410 v1.6, 8100 v2.4.1

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Imation Generic Printer and set:
n Printer Name=8100 , IP Addr.=addr. of 9410, Port=1024, AETitle=9410
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu, select Kodak9410PACSGeneric and set:
n Printer Name=8100 , IP Addr.=accordingly, Port=1024, AETitle= 9410
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips workstation Integris 3D RA with LASER IMAGER Products

n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87


n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips workstation Integris 3D RA with LASER IMAGER Products

Philips Integris 3D RA via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Kodak DryView 8300


Control 9 pin
Philips 3D RA COPCIL board
Workstation Copper Cable

Data 37 pin

Kodak 9410 PACS Link

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 9410 v1.6, 8300 v 4.9.1, 8300 LUT v3.0.3B

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Imation 8300 Generic Printer and set:
n Printer Name=8300 , IP Addr.=addr. of 9410, Port=1024, AETitle=9410
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakDryView8300Generic Printer and set:
n Printer Name=8300 , IP Addr.=accordingly, Port=1024, AETitle=9410
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=0, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips workstation Integris 3D RA with LASER IMAGER Products

n SCU AETitle= accordingly, SCU IPAddr=accordingly


n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=No, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1:1 Portrait, 2:1P, 4:1P, 6:1P
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips 3D RA
Kodak 160 camera
Workstation

Crossed Network Cable

Kodak MIM 200

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM 200 v6.1, KELI 160 v7.2

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=160, IP Addr.=addr.of MIM200, Port=5040, AETitle=160
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu, select KodakKELI160Generic Printer and set:
n Printer Name=160, IP Addr.=accordingly, Port=5040, AETitle=160
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=NORMAL, Min.Density=20, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips workstation Integris 3D RA with LASER IMAGER Products

n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips 3D RA
Kodak 2180 camera
Workstation

Fiber Cable

Kodak MIM 200

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 2180 rel. 2.17H

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=2180, IP Addr.=addr. of MIM200, Port=5040, AETitle=2180
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMKELPGeneric or KodakDPSServerGeneric Printer and set:
n Printer Name=2180, IP Addr.=accordingly, Port=5040, AETitle=2180
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=NORMAL, Min.Density=20, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips workstation Integris 3D RA with LASER IMAGER Products

n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)


n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips 3D RA
Kodak 1120 camera
Workstation

Fiber Cable

Kodak MIM 200

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=1120, IP Addr.=addr. of MIM200, Port=5040, AETitle=1120
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMKELPGeneric or KodakDPSServerGeneric Printer and set:
n Printer Name=1120, IP Addr.=accordingly, Port=5040, AETitle=1120
n Film Size=14x17, Medium Type=Blue Film, Config.Info= CS435
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=NORMAL, Min.Density=20, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips workstation Integris 3D RA with LASER IMAGER Products

n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)


n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic,
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips 3D RA
Workstation
Crossed
Network Cable

Kodak MIM 200


Kodak DryView 8200

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 8200 v3.5

Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=8200 , IP Addr.=addr. of MIM200, Port=5040, AETitle=8200
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=8200 , IP Addr.=accordingly, Port=5040, AETitle=8200
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips workstation Integris 3D RA with LASER IMAGER Products

n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)


n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips 3D RA
Fiber Cable
Workstation

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only
Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=8700, IP Addr.=addr.of MIM200, Port=5040, AETitle=8700
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=8700, IP Addr.=accordingly, Port=5040, AETitle=8700
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips workstation Integris 3D RA with LASER IMAGER Products

n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips 3D RA
Workstation

Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer


Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=8100, IP Addr.=addr.of MIM200, Port=5040, AETitle=8100
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=8100, IP Addr.=accordingly, Port=5040, AETitle=8100
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips workstation Integris 3D RA with LASER IMAGER Products

n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even


n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips 3D RA COPCIL board
Workstation Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer


Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Imation 8300 Generic Printer and set:
n Printer Name=8300, IP Addr.=addr.of MIM200, Port=5040, AETitle=8300
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakDryView8300Generic Printer and set:
n Printer Name=8300, IP Addr.=accordingly, Port=5040, AETitle=8300
n Film Size=8x10, Medium Type=Blue Film, Config.Info= LUT=0,10
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=2, Min.Density=don’t care, Print Priority=HIGH
n Store parameters

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips workstation Integris 3D RA with LASER IMAGER Products

n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even


n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.811
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N)
n Border=Black (Always=N), Film Base=Blue Film (Always=N)
n Film Size=8INx10IN (Always=N), Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips 3D RA
Workstation

Kodak DryView 8900

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)
Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=8900, IP Addr.=addr.of 8900, Port=5040, AETitle=8900
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=8900, IP Addr.=accordingly, Port=5040, AETitle=8900
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips workstation Integris 3D RA with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On


n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT= Pmsv3k.w87 (Always=Y), Contrast=9 (Always=Y)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=3dra

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips 3D RA
Workstation

Kodak DryView 8150

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)
Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=8150, IP Addr.=addr.of 8150, Port=5040, AETitle=DV8150
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=8150, IP Addr.=accordingly, Port=5040, AETitle=DV8150
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips workstation Integris 3D RA with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Image Size=On


n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips workstation Integris 3D RA with LASER IMAGER Products

Integris 3D RA via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips 3D RA
Workstation

Kodak CMI 1000

Requirements: Integris 3D RA v1.1 or v3.2


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)
Note: on 3D RA image matrix=512, possible formats 1:1, 2:1, 4:1, 6:1 only

Integris 3D RA v1.1 Setup:


n into Print Selection/Change menu, select Kodak Generic Printer and set:
n Printer Name=CMI1000, IP Addr.=addr.of CMI1000,
n Port=5040, AETitle= CMI1000
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300 and Store parameters

Integris 3D RA v3.2 Setup:


n into Print Selection/Change menu
n select KodakMIMDryViewGeneric Printer and set:
n Printer Name=CMI1000, IP Addr.=accordingly, Port=5040, AETitle=CMI1000
n Film Size=14x17, Medium Type=Blue Film, Config.Info= LUT=0,9
n Magnification=CUBIC, Max.Density=300, Film Destination=PROCESSOR
n SmoothingType=11, Min.Density= don’t care, Print Priority=HIGH
n Store parameters

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips workstation Integris 3D RA with LASER IMAGER Products

n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=11 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips workstation ViewForum with LASER IMAGER Products

Setup for Philips workstation ViewForum


with LASER IMAGERS

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual).................... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 8
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 10
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600) ................................................................................................................. 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual)(1)................ 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 26
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................28
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600) (1) ............................................................................................................30
n Kodak DryView 8900 LASER IMAGER......................................................................... 32
n Kodak DryView 8150 LASER IMAGER......................................................................... 34
n Kodak Color Medical Imager 1000 ................................................................................. 36

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips workstation ViewForum with LASER IMAGER Products

Philips ViewForum via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips ViewForum
workstation

Kodak DryView 8300

Requirements: ViewForum v3.1


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8300, port=104, AETitle=8300
n click Automatic Config., after a while the result should be Print as SCP,
n (note: in case of DICOM connection problem set Max PDU=4K or 16K), Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8300 DICOM, Name=dv8300, AE title=8300
n into Config. Media List: 8x10 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=1
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips workstation ViewForum with LASER IMAGER Products

Philips ViewForum via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Philips ViewForum Kodak 190 camera


workstation

Requirements: ViewForum v3.2


Requirements: 190 rel. 1.7.29

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 190, port=5040, AETitle=190
n click Automatic Config., after a while the result should be Print as SCP,
n (note: in case of a DICOM connection problem set Max PDU=65536), Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= Kodak MLP-190 DICOM, Name=k190,
n AE title=190, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87
Philips ViewForum
Fiber Cable
workstation

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8500 Plus

Requirements: ViewForum v3.1


Requirements: 9410 v1.6, 8500 v1.34

Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into 9410 Local Printers

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8500 DICOM, Name=dv8500, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8500, Model=8500, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips workstation ViewForum with LASER IMAGER Products

n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0


n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8500, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8500, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips ViewForum
Fiber Cable
workstation

Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus

Requirements: ViewForum v3.1


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle= 9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8700 DICOM, Name=dv8700, AE title= 9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips workstation ViewForum with LASER IMAGER Products

n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0


n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips ViewForum
workstation

Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Requirements: ViewForum v3.1


Requirements: 9410 v1.6, 8100 v2.4.1

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8100 DICOM, Name=dv8100, AE title=9410
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips workstation ViewForum with LASER IMAGER Products

on Device Setup, SCP PRINTERS, Registered Users:


n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=0
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation ViewForum with LASER IMAGER Products

Philips ViewForum via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips ViewForum COPCIL board
workstation Copper Cable

Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300

Requirements: ViewForum v3.1


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle= 9410
n click Automatic Config., after a while the result should be Print as SCP,
n (note: in case of a DICOM connection problem set Max PDU=65536), Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8300 DICOM, Name=dv8300, AE title=9410
n into Config. Media List: 8x10 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips workstation ViewForum with LASER IMAGER Products

n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No


n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips workstation ViewForum with LASER IMAGER Products

Philips ViewForum via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips ViewForum COPCIL board
workstation Copper Cable

Data 37 pin

Kodak 9410 PACS Link


Requirements: ViewForum v3.2
Requirements: 9410 v1.6, 8610 v2.4 (if 8600 v2.5)

Note: This setup is basically also valid for 8600 camera

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8610 DICOM, Name=dv8610, AE title=9410
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8610, Model=8600/8610, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=12 (8 if 8600), Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=1, Installed TFT/ULUT= don’t care

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips workstation ViewForum with LASER IMAGER Products

n Max.Density=300, Contrast=5, Border Density=0


n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8610, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8610, Copies=1, Contrast Table=5, Max.Density=300
n Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Gamma=Linear in Transmittance, Contrast Group=1, Kernel Group=1
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=127
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips ViewForum
workstation Kodak 160 camera
Crossed
Network Cable

Kodak MIM 200

Requirements: ViewForum v3.1


Requirements: MIM200 v6.1, KELI 160 v7.2

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=160
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= Kodak KELI-160 DICOM, Name=k160
n AE title=160, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips workstation ViewForum with LASER IMAGER Products

n Supported Item: Collation=On, Text=On, Tonescaling on Image=On


n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips ViewForum
Kodak 2180 camera
workstation

Fiber Cable

Kodak MIM 200

Requirements: ViewForum v3.1


Requirements: MIM200 v6.1, 2180 rel. 2.17H

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=2180
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= Kodak KELP2180 DICOM, Name=k2180,
n AE title=2180, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips workstation ViewForum with LASER IMAGER Products

n Media Info:Media=35x35cm,4136x4136,4006, (Blue)


n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips ViewForum
Kodak 1120 camera
workstation

Fiber Cable

Kodak MIM 200


Requirements: ViewForum v3.1
Requirements: MIM200 v6.1, 1120 rel. 2.40/032

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=1120
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= Kodak KELP1120 DICOM, Name=k1120,
n AE title=1120, into Config. Media List: 14x17 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips workstation ViewForum with LASER IMAGER Products

n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)


n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips ViewForum
workstation
Crossed
Network Cable

Kodak MIM 200 Kodak DryView 8200

Requirements: ViewForum v3.1


Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8200
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer=3M 8200 DICOM, Name=k8200, AE title=8200
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips workstation ViewForum with LASER IMAGER Products

n General: Database Name=8200, Model=8200, Memory=32


n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On
n Format Types: Custom=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87

Philips ViewForum
Fiber Cable
workstation

Fiber Cable

Kodak MIM 200 Kodak DryView 8500 Plus

Requirements: ViewForum v3.1


Requirements: MIM200 v6.1, 8500 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into MIM200 Destination

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8500
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8500 DICOM, Name=dv8500, AE title=8500
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips workstation ViewForum with LASER IMAGER Products

n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9


n on Configure, Destination, Add, Print, Direct, Qualified, 8500, Set:
n General: Log.Name=8500
n General: Database Name=8500, Model=8500, Memory=32
n Media Info: Media=11x14in, 3388x4283, 4193, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=6
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n ICCProfile=Off (Always=N), Modality=OT (Always=N)
n BodyPart=Default (Always=N), Modality Group=6
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=11INx14IN (Always=N)
n Negotiate Color Print=Off

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips ViewForum
Fiber Cable
workstation

Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus

Requirements: ViewForum v3.1


Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8700
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8700 DICOM, Name=dv8700, AE title=8700
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips workstation ViewForum with LASER IMAGER Products

n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9


n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips ViewForum
workstation

Data 37-pin

Kodak MIM 200


Kodak DryView 8100
Requirements: ViewForum v3.1
Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8100
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8100 DICOM, Name=dv8100, AE title= 8100
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips workstation ViewForum with LASER IMAGER Products

n General: Database Name=8100, Model=8100, Memory=32


n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
Philips ViewForum COPCIL board
workstation Copper Cable

Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: ViewForum v3.1


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle= 8300
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8300 DICOM, Name=dv8300, AE title=8300
n into Config. Media List: 8x10 G8=On (8-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips workstation ViewForum with LASER IMAGER Products

n General: Database Name=8300, Model=8300, Memory=32


n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=0 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips ViewForum COPCIL board
workstation Copper Cable

Data 37 pin

Kodak MIM 200


Requirements: ViewForum v3.2
Requirements: MIM200 v6.1, 8610 v2.4 (if 8600 v2.5)

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8600 camera

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8610
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM, Printer= 3M 8610 DICOM, Name=dv8610, AE title=8610
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips workstation ViewForum with LASER IMAGER Products

n on Configure, Destination, Add, Print, Direct, Qualified, 8610 (or 8600), Set:
n General: Log.Name=8610 (or 8600)
n General: Database=8610 (or 8600), Model=8610 (or 8600), Memory=32
n Media Info: Media=8x10in, 5025x6200, 6020, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12 (8 if 8600), Header/Line,Transfer=2000,Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=25.8065
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.5
n M952 Usage: Min H/V Mag=0.26, Max H/V Mag=16, Min Dmax 952=2
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= upmsv3kp.w8u (Always=N), Contrast (dry)=5 (Always=N)
n Modality Group=1, Conv.Group=1, Gamma Table=Linear in Transmittance
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=1 (Always=Y), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None (Always=N)
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=127
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips ViewForum
workstation

Kodak DryView 8900

Requirements: ViewForum v3.4


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8900, port=5040, AETitle=8900
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM,
n Printer= 3M 8900-Standard DICOM (or 3M 8200 DICOM),
n Name=dv8900, AE title=8900
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips workstation ViewForum with LASER IMAGER Products

n Media Types: Media=35x35cm,9100x9087, 8831, Any


n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=0 (Always=Y), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=vf

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips ViewForum
workstation

Kodak DryView 8150

Requirements: ViewForum v3.4


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8150, port=5040, AETitle=DV8150
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM,
n Printer= Kodak 8150 DICOM (or 3M 8200),
n Name=dv8150, AE title=DV8150
n into Config. Media List: 14x17 G8=On (8-bit-print), 14x17 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any

Author: gian.valesano@kodak.com, Edition: May ´05 34


Philips workstation ViewForum with LASER IMAGER Products

n Media Types: Media=35x35cm, 4438x4414,4286, Any


n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2
n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=0 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 35


Philips workstation ViewForum with LASER IMAGER Products

ViewForum via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips ViewForum
workstation

Kodak CMI 1000

Requirements: ViewForum v4.1V1L2


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

ViewForum Setup:
n enter into Configuration
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of CMI-1000, port=5040, AETitle=CMI1000
n click Automatic Config., after a while the result should be Print as SCP,
n (Note: in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Print SOP Classes set:
n Basic Greyscale=On, Basic Color=On, Presentation LUT=Off
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the VF Node and click Edit
n the whole menu is called VF Config, select Peripherals settings
n below window containing list of Printers defined for VF, click New
n set Interface=DICOM,
n Printer= Kodak CMI-1000 DICOM (or Codonics-NP1660MD-DICOM),
n Name=CMI1000, AE title=CMI1000
n into Config. Media List:
n if Kodak CMI-1000 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, A4 Paper RGB_8=On, all others Off
n if Letter inside CMI-1000 Tray, 8_5INx11IN Paper RGB_8=On, others=Off
n if Codonics-NP1660 was choosed then, depending on what inside CMI-1000, set:
n if A4 inside CMI-1000 Paper Tray, set A4 Paper RGB_8=On, others=Off
n if Letter inside CMI-1000 Paper Tray, set A Paper RGB_8=On, others=Off
n Proceed, Exit, Proceed, Execute, Proceed, click on triangle

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):

Author: gian.valesano@kodak.com, Edition: May ´05 36


Philips workstation ViewForum with LASER IMAGER Products

at Local Panel, under Menu, Imager Setup:


n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=N), TFT Set (From list)=Pmsv3k.w87 (Always=N)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)
n Smoothing=0 (Always=N), Polarity=Positive (Always=N)
n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 37


Philips workstation Xcelera with LASER IMAGER Products

Setup for Philips workstation Xcelera


with LASER IMAGERS

Connection via DICOM:

n Kodak DryView 8300 LASER IMAGER......................................................................... 2


n Kodak Digital Science 190 MEDICAL LASER PRINTER......................................... 3
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) ...... 4
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual).................... 6
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER......................................................................... 8
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER......................................................................... 10
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER .......................................................................... 12
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ....................................................................... 14
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ....................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER (1).................................................................... 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) .. 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus (see also 8800/8500 Dual)(1)................ 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (1).................................................................... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER (1)....................................................................26
n Kodak DryView 8900 LASER IMAGER......................................................................... 28
n Kodak DryView 8150 LASER IMAGER......................................................................... 30
n Kodak Color Medical Imager 1000 ................................................................................. 32

1
Setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: May ´05 1


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak DryView 8300 LASER IMAGER

DICOM Network

Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound

Kodak DryView 8300

Requirements: Xcelera v1.2


Requirements: 8300 v4.9.1, 8300 DICOM board v3.3, 8300 LUT v3.0.3B

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8300, Port number=104
n Display Format (cols x rows)=2x3, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On could cause problems)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

8300 Setup:
n on 8300 Operator Menu into Configure User set:
n Density=16, Contrast=10, Sharp/Smooth=2
n on 8300 Service Menu into Configure User set:
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: May ´05 2


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak Digital Science 190 MEDICAL LASER PRINTER

DICOM Network

Kodak 190 camera


Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound

Requirements: Xcelera v1.2


Requirements: 190 rel. 1.7.29

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=190, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On could cause problems)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

190 Setup:
n On 190 Control Panel press Menu, Setup:
n Printer Density=3.00, Curve Shape=3.0, Contrast=0, Contrast Pivot Density=1.2
n IPAddress=accordingly, Subnet Mask=accordingly, Gateway Address=accordingly
n Port=5040, Network Node Name=Kodak MLP190, Connection Type=RJ45

Author: gian.valesano@kodak.com, Edition: May ´05 3


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Xcelera Fiber Cable


cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8700 Plus

Requirements: Xcelera v1.2


Requirements: 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=9410, Port number=1024
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On doesn’t work)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8700, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8700, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87

Author: gian.valesano@kodak.com, Edition: May ´05 4


Philips workstation Xcelera with LASER IMAGER Products

n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11


n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 5


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87

Philips Xcelera Fiber Cable


cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Fiber Cable

Kodak 9410 PACS Link Kodak DryView 8500 Plus

Requirements: Xcelera v1.2


Requirements: 9410 v1.6, 8500 v1.34

Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into 9410 Local Printers

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=9410, Port number=1024
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On doesn’t work)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8500, Model=8500, Output=Fiber
n Command Set=Superset, Baud=19200, Data=8, Stop=1, Parity=None
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8500, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8500, Copies=1, Contrast Table=9, TFT/ULUT=Pmsv3k.w87

Author: gian.valesano@kodak.com, Edition: May ´05 6


Philips workstation Xcelera with LASER IMAGER Products

n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11


n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Enable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 7


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound
Data 37-pin

Kodak 9410 PACS Link


Kodak DryView 8100

Requirements: Xcelera v1.2


Requirements: 9410 v1.6, 8100 v2.4.1

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=9410, Port number=1024
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On doesn’t work)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8100, AE Title= 9410, Port= 1024, Max Associations= 5
on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8100, Copies=1, Contrast Table=9, TFT/ULUT=Pmsb.w87
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=11
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No

Author: gian.valesano@kodak.com, Edition: May ´05 8


Philips workstation Xcelera with LASER IMAGER Products

n Contrast Test Mode=No, Contrast Test Mode Density=0

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 9


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
COPCIL board
Philips Xcelera
Copper Cable
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Data 37 pin

Kodak 9410 PACS Link Kodak DryView 8300

Requirements: Xcelera v1.2


Requirements: 9410 v1.6, 8300 v4.9.1, 8300 LUT v3.0.3B

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=9410, Port number=1024
n Display Format (cols x rows)=2x3, Orientation=Portrait, Annotation=Off
n (Note: Annot.=Off means no patient name, Annot.=On doesn’t work)
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8300, Model=8300 Copper, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=8, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=0, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=upmsv3k.w8u
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8300, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8300, Copies=1, Contrast Table=10, TFT/ULUT=u1_109b.w8u
n Max.Density=300, Border Density=0, Mag.Type=8, Smoothing=2
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Interceptor=Yes, N-Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No

Author: gian.valesano@kodak.com, Edition: May ´05 10


Philips workstation Xcelera with LASER IMAGER Products

n Contrast Test Mode=No, Contrast Test Mode Density=0

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 11


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

DICOM Network

Philips Xcelera
Kodak 160 camera
cardiac PACS workstation Crossed
for images coming from Network Cable
Nuc.Med. / Ultrasound

Kodak MIM 200

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, KELI 160 v7.2

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=160, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On

Author: gian.valesano@kodak.com, Edition: May ´05 12


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On


n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=N), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation= Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 13


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

DICOM Network

Philips Xcelera Kodak 2180 camera


cardiac PACS workstation
for images coming from
Fiber Cable
Nuc.Med. / Ultrasound

Kodak MIM 200

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, 2180 rel. 2.17H

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=2180, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On

Author: gian.valesano@kodak.com, Edition: May ´05 14


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Size Scale=On


n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Modality=OT (Always=N)
n Border=3.99 (Always=N), Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Film Output Dest=Bin 9 (Always=N)
n Negotiate Color Print=Off

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 15


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

DICOM Network

Philips Xcelera Kodak 1120 camera


cardiac PACS workstation
for images coming from
Fiber Cable
Nuc.Med. / Ultrasound

Kodak MIM 200

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, 1120 rel. 2.40/032

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=1120, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On

Author: gian.valesano@kodak.com, Edition: May ´05 16


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: DICOM Annotation Box=On, Req.Size Scale=On


n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=Curve Shape (Always=Y)
n Curve Shape= 3 ie 435 (Always=Y), Contrast (wet)=0 (Always=Y)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Interpolation=Cubic, Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: May ´05 17


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8200 LASER IMAGER

DICOM Network

Philips Xcelera Crossed


cardiac PACS workstation Network
for images coming from Cable
Nuc.Med. / Ultrasound

Kodak MIM 200 Kodak DryView 8200

Requirements: Xcelera v1.2


Requirements: MIM 200 v6.1, 8200 v3.5

Note: This setup is basically also valid for 25 PS computer

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8200, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8200, Set:
n General: Log.Name=8200, IPAddr=100.100.100.40,
n General: Database Name=8200, Model=8200, Memory=32
n Media Info: Media=35x43cm, 4361x5223, 5133, Any (Blue,Clear)
n Media Info: Media=35x35cm, 4361x4247, 4157, Any (Blue,Clear)
n Media Info: Media=11x14in, 3272x4361, 4271, Any (Blue,Clear)
n Supported Item: Collation=On, Common Text=On, Tonescaling=On
n Supported Item: Remote Image Deletion=On, Rotation=On
n Interpolation: No Magnification=On, Bilin=On, Repl=On
n Interpolation: Cubic Spline=On, Cubic Convolution 2=On

Author: gian.valesano@kodak.com, Edition: May ´05 18


Philips workstation Xcelera with LASER IMAGER Products

n Format Types: Custom=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n SCU: Density Space= On, Variable Border=On, PLUT Service= On
n SCU: Curve Shape Tonescaling=On, Perception LUT Tonescaling=On
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8200 Setup on CONFIGURATION Menu:


n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on DICOM Menu: SCU Config.: AE=MIM, TFT=VER693C0, C=6, Smooth=0
n on DICOM Menu: DICOM Config.: AE=NER_ANY, Port=1025

Author: gian.valesano@kodak.com, Edition: May ´05 19


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

DICOM Network

iew
00 DryV
3M 87

Philips Xcelera Fiber Cable


cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Fiber Cable

Kodak MIM 200 Kodak DryView 8700 Plus


Requirements: Xcelera v1.2
Requirements: MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8700, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On

Author: gian.valesano@kodak.com, Edition: May ´05 20


Philips workstation Xcelera with LASER IMAGER Products

n Format Types: Custom=On, Slide=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=Y), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 21


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8500 LASER IMAGER / Plus
(see also 8800/8500 Dual)

DICOM Network

iew
00 DryV
3M 87

Philips Xcelera Fiber Cable


cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Fiber Cable

Kodak MIM 200 Kodak DryView 8500 Plus

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, 8500 v1.34

Note: This setup is basically also valid for 25 PS computer


Note: This setup is basically also valid for 8800/8500 Dual camera
by basically using the correspondent Model into MIM200 Destination

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8500, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8500, Set:
n General: Log.Name=8500
n General: Database Name=8500, Model=8500, Memory=32
n Media Info: Media=11x14in, 3388x4283, 4193, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 22


Philips workstation Xcelera with LASER IMAGER Products

n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On


n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsv3k.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N),
n Film Base=Blue Film (Always=N), Film Size=11INx14IN (Always=N)
n Negotiate Color Print=Off

8500 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Port=0, Parity=Disable Even, Timeout=60

Author: gian.valesano@kodak.com, Edition: May ´05 23


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

DICOM Network

RS422
COPCIL Board 9-pole Plug
Copper Cable Control 9-pin
Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound
Data 37-pin

Kodak MIM 200


Kodak DryView 8100

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, 8100 v2.4.1

Note: This setup is basically also valid for 25 PS computer

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8100, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 24


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= Pmsb.w87 (Always=N), Contrast (dry)=9 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=11 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Negotiate Color Print=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=90, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=Ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: May ´05 25


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8300 LASER IMAGER

DICOM Network

Control 9 pin
COPCIL board
Philips Xcelera
Copper Cable
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound Data 37 pin

Kodak MIM 200 Kodak DryView 8300

Requirements: Xcelera v1.2


Requirements: MIM200 v6.1, 8300 v4.9.1, 8300 LUT v3.0.3B

Note: This setup is basically also valid for 25 PS computer

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8300, Port number=5040
n Display Format (cols x rows)=2x3, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:
n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9
n on Configure, Destination, Add, Print, Direct, Qualified, 8300, Set:
n General: Log.Name=8300
n General: Database Name=8300, Model=8300, Memory=32
n Media Info: Media=8x10in, 2256x2676, 2596, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=8, Header/Line, Transfer=2000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symm.=On, SuperSlide=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On

Author: gian.valesano@kodak.com, Edition: May ´05 26


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=11.9048
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3
n M952 Usage: Min H/V Mag=1, Max H/V Mag=16, Min Dmax 952=1.5
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= u1_109s.w8u (Always=Y), Contrast (dry)=10 (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Print Priority=Normal (Always=N), Image Polarity=Normal (Always=N)
n Smoothing Type=2 (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8300 Setup:
n on Operator Menu set: D=16, C=6, Sharp/Smooth=0 ie Auto
n on Operator Menu set: Film Layout: 1 on 1=Portrait
n on Service Menu set: Ratio=1.00, Serial=1200,8,1,Even,Comma,3M, Memory=16
n on Service Menu set: Digital Interface: Parity=Even, Modality Group=1

Author: gian.valesano@kodak.com, Edition: May ´05 27


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak DryView 8900 LASER IMAGER

DICOM Network

Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound

Kodak DryView 8900

Requirements: Xcelera v1.2


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=8900, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On

Author: gian.valesano@kodak.com, Edition: May ´05 28


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On


n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=N), Contrast=9 (Always=N)
n Smoothing Type=11 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=N), Warnings=No
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Blue Film (Always=N)
n Film Size=14INx17IN (Always=N), Sorter Bin=Bin1 (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 29


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak DryView 8150 LASER IMAGER

DICOM Network

Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound

Kodak DryView 8150

Requirements: Xcelera v1.2


Requirements: 8150 v0.08.b6 (DRE=DV8150_DRE_0.08.b6, MIM=MIM_v5.7.b6)

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=DV8150, Port number=5040
n Display Format (cols x rows)=3x4, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

8150 Setup:
n on Configuration, System, Network:
n Host Name=DV8150DRE, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM
n on Configuration, Destination:
n AE Title: Log.Name=DV8150
n Media Types: Media=35x43cm, 4438x5390, 5262, Any
n Media Types: Media=35x35cm, 4438x4414,4286, Any
n Media Types: Media=11x14in, 3440x4438, 4310, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0

Author: gian.valesano@kodak.com, Edition: May ´05 30


Philips workstation Xcelera with LASER IMAGER Products

n Formatting: Max Img Rows=0, Pixel Pitch=12.8000, Min Horiz Sep=2


n Formatting: Min Vert Sep=2, Trim Width Enabled=2
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.10
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No,
n TFT Set Name=Pmsb.w87 (Always=Y)
n Contrast=9 (Always=N), Smoothing Type=11 (Always=N)
n Dmin=0.00 (Always=N), Dmax=3.00 (Always=N)
n Image Polarity=Normal (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No,
n Film Base=Blue Film (Always=N), Film Size=14INx17IN (Always=N)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No,
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N),
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: May ´05 31


Philips workstation Xcelera with LASER IMAGER Products

Philips Xcelera via DICOM with


Kodak Color Medical Imager 1000

DICOM Network

Philips Xcelera
cardiac PACS workstation
for images coming from
Nuc.Med. / Ultrasound
Kodak CMI 1000

Requirements: Xcelera v1.2


Requirements: CMI-1000 v1.0.b8 (DRE=v1.0.b8, MIMCore=v2.0.b3)

Xcelera Setup:
n enter File/Preferences/Hardcopy printer and set:
n Hostname= IP Address accordingly, AETitle=CMI1000, Port number=5040
n Display Format (cols x rows)=2x3, Orientation=Portrait, Annotation=On
n click OK
n to print select a study, if it is a run click the stop-button below the image
n drag-and-drop the wanted image into the image folder
n point on the image into the image folder and right-click on the mouse
n select Hardcopy Print

CMI-1000 Setup:
at Local Panel, under Media, set A4 or Letter ie what inside Paper Tray
(CMI-1000 sw considers only this field and uses it as A4 or Letter (Always=Y):
at Local Panel, under Menu, Imager Setup:
n Network Setup:
n IPAddr=accordingly, Subnet Mask=accordingly, Gateway=accordingly
n under More Settings: DICOM Port=5040
n Edit Host Name/AE Title=CMI1000
n Modality (SCU) Config (or after modality recognition):
n Select Modality=accordingly, Modality Config.Assistant=Yes, Reset Config=No
n Note: Color IQ settings used when images sent with Color DICOM SOP
n under Image Quality, Color Image Quality: (no need to adjust, leave defaults)
n Match Monitor Colors=default ie sRGB Color Space Profile.icm
n Adjust Color Tones=default ie Default_Contrast_100.icm
n Note: Grayscale IQ settings used when images sent with Grayscale DICOM SOP:
n under Image Quality, Grayscale Image Quality:
n Contrast=9 (Always=Y), TFT Set (From list)=Pmsv3k.w87 (Always=Y)
n Wet Contrast=0 (Always=N), Curve Shape=429 (Always=N)
n under More Quality Settings:
n Tonescaling=Presentation LUT (Always=N),
n Image Dmax=220 (Always=N), Image Dmin=10 (Always=N)

Author: gian.valesano@kodak.com, Edition: May ´05 32


Philips workstation Xcelera with LASER IMAGER Products

n Smoothing=11 (Always=N), Polarity=Positive (Always=N)


n Interpolation=CUBIC (Always=N), Trim=No (Always=N)
n Page Orientation=PORTRAIT (Always=N), Negotiate PLUT=No
n under Modality Defaults:
n Media Size=8.5x11 (Always=N), don’t care, this field plays no role
n Image Resizing=Minify (Always=N),
n Priority=Normal (Always=N), NeventReports=No (Always=Y),
n Warnings=Yes, Color=Yes
n (set Color=No when only b/w images are printed and set it to be sure
n to apply to the b/w images the Greyscale Image Quality settings)
n Page Annotation:
n Hosp.Name=No, Modality ID=No, Page Nbr=No, Date=No, Time= No
n Save Changes and Exit

Author: gian.valesano@kodak.com, Edition: May ´05 33


Philips CR Thoravision with some LASER IMAGER Products

Setup Manual for Philips CR Thoravision


with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual................................................................ 2
n Kodak DryView 8700 LASER IMAGER / Plus ................................................................. 3
n Kodak DryView 8700 LASER IMAGER / Standard ......................................................... 4
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual................................................................ 5
n Kodak DryView 8500 LASER IMAGER / Plus ................................................................. 6
n Kodak DryView 8500 LASER IMAGER / Standard ......................................................... 7
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual................................................................ 8
n Kodak DryView 8300 LASER IMAGER / Standalone ...................................................... 9
n Kodak Digital Science 969 HQ LASER IMAGER ....................................................... 10
n Kodak Digital Science 969 HQT LASER IMAGER..................................................... 11
n Kodak Digital Science 959 XL LASER IMAGER ........................................................ 12
n Kodak DryView 8100 LASER IMAGER.......................................................................... 13
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ)(1) ...... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1) ..................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) .................................................................. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ( 2,3,4)............................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ( 2,3,4 )............................................................... 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) ..... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4) ..................................................................... 26

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: November ´04 1


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
tM
npu
l ti- I
Mu
0
88
3M

DryView
3M 8700

Fiber Cable
8700 Dual

8800 MIM

Requirements: Thoravision rel. 3.3.1, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 2


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

Dr yView
3M 8700

8700 Plus

Requirements: Philips Thoravision rel. 3.3.1, 8700 Plus rel. 1.15

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 3


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT
DryView
Philips 37-pin 3M 8700

Data Cable
Image Data OUT

Requirements: Philips Thoravision rel. 3.3.1, 8700 Standard rel. 1.15

Setup: on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8700 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 4


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
tM
npu
l ti- I
Mu
0
88
3M

DryView
3M 8700

Fiber Cable
8500 Dual

8800 MIM

Requirements: Thoravision rel. 3.3.1, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/O FF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8800/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 5


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

Dr yView
3M 8700

8500 Plus

Requirements: Philips Thoravision rel. 3.3.1, 8500 Plus rel. 1.22

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 6


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT
DryView
Philips 37-pin 3M 8700

Data Cable
Image Data OUT

Requirements: Thoravision rel. 3.3.1, 8500 Standard rel. 1.22

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8500 Standard Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 7


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual
Philips 37-pin RS422 37-to-9 pin
Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
ng
a
t Ma
npu
l ti- I
Mu

Fiber Cable
800
M8
3

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: Thoravision rel. 3.3.1, 8800 rel. 5.1, 8300 rel. 4.2
Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
RS422 37-to-9 Host 37 pin side 2 20 3 21 1
Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout: Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0
8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 8


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


Image Data OUT

Requirements: Thoravision rel. 3.3.1, 8300 rel. 4.6

Setup: on Thoravision rel. 3.3.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout: Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Delimiter=Comma, Protocol=3M
n Memory=20 Meg
n Digital Interface: Parity=None
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: November ´04 9


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak Digital Science  969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M H Q L a se r Im a g e r

M969 HQ

Requirements: Philips Thoravision rel. 3.3.1, M969 HQ rel. 3.0

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQ Setup:
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on EIB Menu:
n Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 10


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT 3M H Q L a se r I m a g e r

Philips 37-pin
Data Cable
Image Data OUT

Requirements: Philips Thoravision rel. 3.3.1, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 11


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak Digital Science  959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a se r I ma g er

M959 XL

Requirements: Philips Thoravision rel. 3.3.1, M959 XL rel. 3.8

Setup:
n on Philips Thoravision rel. 3.3.1 install Camera Config.File 3M
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

M959 XL Setup:
n on COMM Menu:
n 959 Default
n on COMM Menu:
n 831 Cmd Set Beta Tables: Horiz.Smooth=11, Vert.Smooth=11
n on EIB Menu:
n Parity=Enable Odd, Timeout=255

Author: gian.valesano@kodak.com, Edition: November ´04 12


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak DryView 8100 LASER IMAGER

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips Thoravision Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

DryView 8100

Requirements: Thoravision rel. 3.3.1, 8100 v1.1

Setup: on Thoravision rel. 3.3.1 install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=180, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 831-Beta-Smooth=11
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: November ´04 13


Philips CR Thoravision with some LASER IMAGER Products

Thoravision with
Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips Thoravision Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable UKEIB
Host Control 422 OUT Host
Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00Dr
yView
Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Thoravision rel. 3.3.1, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 3M is defined
- ItemName=SiteInformation set Preferred Printer= 3M

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: November ´04 14


Philips CR Thoravision with some LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=10
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=thorav, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14, Bit=8
n into Defaults: Beta Y =11

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 15


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips Thoravision Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Thoravision rel. 3.3.1, 9410 v1.6, 8100 v2.4.1

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitio ns verify 3M is defined
- ItemName=SiteInformation set Preferred Printer= 3M

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: November ´04 16


Philips CR Thoravision with some LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=255, Parity=Odd, Memory=10
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=thorav, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14, Bit=8
n into Defaults: Beta Y =11

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 17


Philips CR Thoravision with some LASER IMAGER Products

Thoravision with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

Kpad / Host Control


RS422 37-to-15 pin
Philips Thoravision Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Thoravision rel. 3.3.1, MIM 200 v6.1, KELI 160 v7.2

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak is defined
- ItemName=SiteInformation set Preferred Printer=Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=ThoraVision:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=thorav,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2560, Parity=Odd, Timer=5, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: November ´04 18


Philips CR Thoravision with some LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 19


Philips CR Thoravision with some LASER IMAGER Products

Thoravision with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

Kpad / Host Control


RS422 37-to-15 pin
Philips Thoravision Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 2180 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Thoravision rel. 3.3.1, MIM 200 v6.1, KELP 2180 v2.17H

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak is defined
- ItemName=SiteInformation set Preferred Printer=Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=ThoraVision:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=thorav,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2560, Parity=Odd, Timer=5, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: November ´04 20


Philips CR Thoravision with some LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 21


Philips CR Thoravision with some LASER IMAGER Products

Thoravision with
Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

Kpad / Host Control


RS422 37-to-15 pin
Philips Thoravision Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Host Control 422 OUT Kodak 1120 camera
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Thoravision rel. 3.3.1, MIM 200 v6.1, KELP 1120 v2.40/032

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify Kodak is defined
- ItemName=SiteInformation set Preferred Printer=Kodak

MIM200 Setup:
n on Configure, Network:
n Computer Name=e.g., Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=ThoraVision:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=thorav,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2560, Parity=Odd, Timer=5, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: November ´04 22


Philips CR Thoravision with some LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 23


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips Thoravision Control Adaptor Cable
Control Cable 700 Dry
View
3 M8
Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37- pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Thoravision rel. 3.3.1, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 3M is defined
- ItemName=SiteInformation set Preferred Printer= 3M

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=ThoraVision:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=thorav,
n Mod.Name=thorav, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2560, Parity=Odd, Timer=5, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On

Author: gian.valesano@kodak.com, Edition: November ´04 24


Philips CR Thoravision with some LASER IMAGER Products

n Format Types: Custom=On, Slide=On, Row Symmetric=On


n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 25


Philips CR Thoravision with some LASER IMAGER Products

Philips Thoravision with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER

RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37 -pin Octopus Adaptor Cable COPCIL Board
Philips Thoravision Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37 -to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Thoravision rel. 3.3.1, MIM200 v6.1, 8100 v2.4.1

Thoravision Setup:
n on Thoravision into System/Service/Configuration/Complete System:
- ItemName=SystemOptions check OptHCU=Yes
- ItemName=Unstructured Items/printerAttached check Value=Yes
- ItemName=Imager Definitions verify 3M is defined
- ItemName=SiteInformation set Preferred Printer= 3M

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect, Config.Add, Digital, Qual.=ThoraVision:
n Board/Channel=Digital Board 1 Channel 1, Logical Name=thorav,
n Mod.Name=thorav, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=2048, Lines=2560, Parity=Odd, Timer=5, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: November ´04 26


Philips CR Thoravision with some LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination=8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 27


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Setup Manual for Philips CT Tomoscan SR / AV


with LASER IMAGERS
Connection POINT-TO-POINT (dig/422 host control):

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual (dig/422 hc) ............................................ 3
n Kodak DryView 8700 LASER IMAGER / Plus (dig/422 hc) ............................................. 4
n Kodak DryView 8700 LASER IMAGER / Standard (dig/422 hc)...................................... 5
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual (dig/422 hc) ............................................ 6
n Kodak DryView 8500 LASER IMAGER / Plus (dig/422 hc) ............................................. 7
n Kodak DryView 8500 LASER IMAGER / Standard (dig/422 hc)...................................... 8
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual (dig/422 hc) ............................................ 9
n Kodak DryView 8300 LASER IMAGER / Standalone (dig/422 hc)................................... 10
n Kodak Digital Science 969 HQ LASER IMAGER (dig/422 hc).................................... 11
n Kodak Digit al Science 969 HQT LASER IMAGER (dig/422 hc).................................. 12
n Kodak Digital Science 959 XL LASER IMAGER (dig/422 hc)..................................... 13
n Kodak DryView 8100 LASER IMAGER (dig/422 hc) ...................................................... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ) (dig/422 hc) ( 1)............................................................ 15
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (dig/422 hc) (1).................................................. 17
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (dig/422 hc) (2,3,4)............................................... 19
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (dig/422 hc) ( 2,3,4)............................................ 21
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (dig/422 hc) ( 2,3,4)............................................ 23
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ) (dig/422 hc) ( 4)............................................................ 25
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (dig/422 hc) (4).................................................. 27

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: November ´04 1


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Connection POINT-TO-POINT (video/keypad):

n Kodak DryView 8100 LASER IMAGER (v/kpad)............................................................ 29


n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ) (v/kpad) (1) ................................................................. 30
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (v/kpad) ( 1)....................................................... 32
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (v/kpad) (2,3,4) .................................................... 34
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER (v/kpad) (2,3,4) ................................................. 36
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER (v/kpad) (2,3,4) ................................................. 38
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ) (v/kpad) (4) ................................................................. 40
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (v/kpad) ( 4)....................................................... 42

Author: gian.valesano@kodak.com, Edition: November ´04 2


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
tM
npu
l ti- I
Mu
0
88
3M

DryView
3M 8700

Fiber Cable
8700 Dual

8800 MIM

Requirements: Tomoscan SR rel. 1.4F, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8800/8700 Dual Setup:


n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 3


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

Dr yView
3M 8700

8700 Plus

Requirements: Philips Tomoscan SR rel. 1.4F, 8700 Plus rel. 1.15

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8700 Plus Setup:


n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 4


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT
DryView
Philips 37-pin 3M 8700

Data Cable
Image Data OUT

Requirements: Philips Tomoscan SR rel. 1.4F, 8700 Standard rel. 1.15

Setup: on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M

8700 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 5


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

ger
ana
tM
npu
l ti- I
Mu
0
88
3M

DryView
3M 8700

Fiber Cable
8500 Dual

8800 MIM

Requirements: Tomoscan SR rel. 1.4F, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8800/8500 Dual Setup:


n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 6


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

Dr yView
3M 8700

8500 Plus

Requirements: Philips Tomoscan SR rel. 1.4F, 8500 Plus rel. 1.22

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8500 Plus Setup:


n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 7


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT
DryView
Philips 37-pin 3M 8700

Data Cable
Image Data OUT

Requirements: Tomoscan SR rel. 1.4F, 8500 Standard rel. 1.22

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8500 Standard Setup:


n on TDB/C board:
n Jumpers=RS422, Switch=Center
n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 8


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
er
ng
a
t Ma
npu
l ti- I
Mu

Fiber Cable
800
M8
3

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: Tomoscan SR rel. 1.4F, 8800 rel. 5.1, 8300 rel. 4.2

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0

8800 Setup:
n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 9


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 37-pin RS422 37-to-9 pin
Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422 OUT

Philips 37-pin Data Cable


Image Data OUT

Requirements: Philips Tomoscan SR rel. 1.4F, 8300 rel. 4.2

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=4800, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Odd
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: November ´04 10


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak Digital Science  969 HQ LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR
Control Cable Host Adaptor Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M H Q L a se r Im a g e r

M969 HQ

Requirements: Philips Tomoscan SR rel. 1.4F, M969 HQ rel. 3.0

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

M969 HQ Setup:
n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 11


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 37-pin RS422 37-to-9 pin
Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT 3M H Q L a se r I m a g e r

Philips 37-pin
Data Cable
Image Data OUT

Requirements: Philips Tomoscan SR rel. 1.4F, M969 HQT rel. 1.0


Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M

M969 HQT Setup:


n on TDB/C board: Jumpers=RS422, Switch=Center
n on COMM Menu: 4800, 8, Enable Even, 1, 959 Set, CR, ACK/NACK, Pace=0
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on HOST Menu:
n Override 952 Beta=Yes, 831 Cmd Set Beta Tables: Horiz. Sharp=15
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 12


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak Digital Science  959 XL LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3 M L a se r I ma g er

M959 XL

Requirements: Philips Tomoscan SR rel. 1.4F, M959 XL rel. 3.8

Setup:
n on Philips Tomoscan SR rel. 1.4F install Camera Config.File 3M
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center
n on XL Input Module install LUT 322A0

M959 XL Setup:
n on COMM Menu:
n 4800, 8, Enable Even, 1, 959 Command Set, CR, ACK/NACK, Pace=0
n on EIB Menu:
n Parity=Enable Odd, Timeout=10

Author: gian.valesano@kodak.com, Edition: November ´04 13


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak DryView 8100 LASER IMAGER

RS422 9-pole
Philips 37-pin RS422 37-to-9 pin plug
Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host Control 422 OUT
Philips 37-pin
Data Cable
Image Data OUT

DryView 8100

Requirements: Tomoscan SR rel. 1.4F, 8100 v1.1

Setup: on Tomoscan SR rel. 1.4F install Camera Config.File 3M

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 4800 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=10, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=13
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Odd

Author: gian.valesano@kodak.com, Edition: November ´04 14


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips CT Tomoscan SR Philips 37-pin RS422 37-to-9 pin


Control Cable Host Adaptor Cable UKEIB
Host Control 422 OUT Host
Imager

Philips 37-pin Copper Cable


Data Cable DPC 25-9 DPC-UKEIB Cable
Image Data OUT

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


00Dr
yView
Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Tomoscan SR rel. 1.4F, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
n on ALIIF board set switch DSW3 (1-8)=OFF/ON/OFF/ON/ON/OFF/ON/ON (3M)

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

Author: gian.valesano@kodak.com, Edition: November ´04 15


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=4800
on Device Setup, HOST KEYPADS:
n Kpad=sr, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Dest.=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 16


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 37-pin RS422 37-to-9 pin


Philips CT Tomoscan SR Host Adaptor Cable
Control Cable
Host UKEIB
Host Control 422 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Tomoscan SR rel. 1.4F, 9410 v1.6, 8100 v2.4.1

Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
n on ALIIF board set switch DSW3 (1-8)=OFF/ON/OFF/ON/ON/OFF/ON/ON (3M)

Setup: on UKEIB set OFF/ON/OFF/ON/ON/ON, Switch=Center

RS422 37-to-9 Host 37 pin side 2 20 3 21 1


Adaptor Cable pinout 9 pin side 8 3 7 2 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=10, Parity=Odd, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=4800
on Device Setup, HOST KEYPADS:
n Kpad=sr, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On

Author: gian.valesano@kodak.com, Edition: November ´04 17


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On


n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 18


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Tomoscan SR Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Crossed Kodak 160 camera
Host Control 422 OUT
Network Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, KELI 160 v6.1

Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
on ALIIF board set switch DSW3 (1-8)=ON/OFF/ON/ON/ON/OFF/ON/ON (Kodak)

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add, Dig.,Qual.=Tomoscan Expanded:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ctsr,
n Primary Control: Type= Hitachi, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 4800, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: C ustom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 19


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 20


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Tomoscan SR Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 2180 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 2180 v2.17H

Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
on ALIIF board set switch DSW3 (1-8)=ON/OFF/ON/ON/ON/OFF/ON/ON (Kodak)

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add, Dig.,Qual.=Tomoscan Expanded:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ctsr,
n Primary Control: Type= Hitachi, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 4800, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: November ´04 21


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 22


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Kpad / Host Control
RS422 37-to-15 pin
Philips CT Tomoscan SR Philips 37-pin Octopus Adaptor Cable
Control Adaptor Cable
Control Cable
Kodak 1120 camera
Host Control 422 OUT
Fiber Cable
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 1120 v2.40/032

Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
on ALIIF board set switch DSW3 (1-8)=ON/OFF/ON/ON/ON/OFF/ON/ON (Kodak)

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add, Dig.,Qual.=Tomoscan Expanded:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ctsr,
n Primary Control: Type= Hitachi, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio =1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 4800, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On

Author: gian.valesano@kodak.com, Edition: November ´04 23


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: True Landscape=On


n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optic al Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 24


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Kpad / Host Control


RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable
Philips CT Tomoscan SR Control Adaptor Cable
Control Cable 700 Dry
View
3 M8
Fiber Cable
Host Control 422 OUT
RS422 37-to-50 pin
Philips 37- pin Fiber Cable
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination
Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
on ALIIF board set switch DSW3 (1-8)=OFF/ON/OFF/ON/ON/OFF/ON/ON (3M)

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add, Dig.,Qual.=Tomoscan Expanded:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ct,
n Mod.Name=ct, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4, 4800, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On

Author: gian.valesano@kodak.com, Edition: November ´04 25


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Req.Size Mix in Row=On, Trim=On


n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 26


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR digital with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Kpad / Host Control 9-pole Plug
RS422 37-to-15 pin
Philips 37-pin Octopus Adaptor Cable COPCIL Board
Philips CT Tomoscan SR Control Adaptor Cable Control 9-pin
Control Cable Copper Cable
Host Control 422 OUT
RS422 37 -to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 8100 v2.4.1

Note: same setup valid for Tomoscan CT of SR series and AV series

Tomoscan SR Setup:
on ALIIF board set switch DSW3 (1-8)=OFF/ON/OFF/ON/ON/OFF/ON/ON (3M)

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add, Dig.,Qual.=Tomoscan Expanded:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ct,
n Mod.Name=ct, Primary Control: Type=M952/P831, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=1024, Lines=1024, Parity=Odd, Timer=10, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4, 4800, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: November ´04 27


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 28


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak DryView 8100 LASER IMAGER

Keypad

RS422 9-pole
Keypad 9-to-9-pin plug
Philips CT Tomoscan SR UKEIB
Copper Cable
Imager

Video Cable
Image Video OUT

DryView 8100

Requirements: Tomoscan SR rel. 1.4F, 8100 v1.1

Setup: on UKEIB set ON/OFF/ON/ON/ON/ON, Switch=Center

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=10, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=13
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Video
n on VIDEO Menu: Parameters as per autosync

Author: gian.valesano@kodak.com, Edition: November ´04 29


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)
Philips CT Tomoscan SR
Keypad UKEIB
Keypad
Imager

Video Cable Copper Cable


Image VideoOUT

Imagraph Card
Octopus Cable

DryView 8700 Plus 3M 870


0 Dr yVie
w
Fiber Cable
RS422 Copper
Fiber Cable Converter Cable

9410 PACS Link

Requirements: Tomoscan SR rel. 1.4F, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers
Note: same setup valid for Tomoscan CT of SR series and AV series

Setup: on UKEIB set ON/OFF/ON/ON/ON/ON, Switch=Center

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, IMAGRAPH Video Menu:
n Vert.Tot=1125, Frame Period=36, Interlace=ON, Vert.Back Porch=41.0
n Image Height=1024, Horiz.Tot=1792, Horiz.Back Sync=436
n Image Width=1294, Horiz.Sync Width=2518, Horiz.Back Porch=421
on I/O CARD SETUP, SERIAL Ports Menu:
n COM3: Protocol=Superset, Parity=None, Baud Rate=19200
on Device Setup, IMS KEYPADS:
n Kpad=sr, Hosp=philips, Serial Port=COM3, Image Card=Imagraph, Channel=A1

Author: gian.valesano@kodak.com, Edition: November ´04 30


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Available Printers=8700, Film Destination=Processor, TFT/ULUT=Ver693c0.w87


n into Defaults: Dest=8700, Size=14x17, C=9, D=14, Dmax=3.0, Bit=8

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 31


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR vide o with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips CT Tomoscan SR
UKEIB
Keypad Keypad
Imager

Copper Cable
Video Cable
ImageVideo OUT

Imagraph Card
DryView 8100 Octopus Cable
RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable RS422 Copper
Converter Cable

Data 37 pin
9410 PACS Link

Requirements: Tomoscan SR rel. 1.4F, 9410 v1.6, 8100 v2.4.1

Setup: on UKEIB set ON/OFF/ON/ON/ON/ON, Switch=Center

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, IMAGRAPH Video Menu:
n Vert.Tot=1125, Frame Period=36, Interlace=ON, Vert.Back Porch=41.0
n Image Height=1024, Horiz.Tot=1792, Horiz.Back Sync=436
n Image Width=1294, Horiz.Sync Width=2518, Horiz.Back Porch=421
on I/O CARD SETUP, SERIAL Ports Menu:
n COM3: Protocol=Superset, Parity=None, Baud Rate=19200
on Device Setup, IMS KEYPADS:
n Kpad=sr, Hosp=philips, Serial Port=COM3, Image Card=Imagraph, Channel=A1
n Available Printers=8100, Film Destination=Processor, TFT/ULUT=Ver693c0.w87
n into Defaults: Dest=8100, Size=14x17, C=9, D=14, Dmax=3.0, Bit=8

Author: gian.valesano@kodak.com, Edition: November ´04 32


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 33


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Keypad Kpad / Host Control
Octopus Adaptor Cable
Philips CT Tomoscan SR Kodak 160 camera
Crossed
Network Cable
Video Card
Octopus Cable
Video Cable
ImageVideo OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, KELI 160 v7.2

Note: same setup valid for Tomoscan CT of SR series and AV series

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config., Input, Direct Connect, Config.Add,Video,Qual.=Tomoscan SR6000:
n Board/Channel=Video60 Board 1 Channel 1, Logical Name= ctsr,
n Primary Control Type= Conv.Kpad, Com Port=9
n Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n HL=31.36, HD=22.75, Clk=57.14, Tot.Pxls=1792, Image Pxls=1300
n HS=3, HFP=0.61, HBP=5, HB=9.61
n Image Lines=1024, Tot.Lines=1124, VBP=38, VS=4, PreE=PostE=4
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: November ´04 34


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Pixel Pitch=12.8


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 35


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Keypad Kpad / Host Control
Octopus Adaptor Cable
Philips CT Tomoscan SR Kodak 2180 camera

Fiber Cable
Video Card
Octopus Cable
Video Cable
ImageVideo OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 2180 v2.17H

Note: same setup valid for Tomoscan CT of SR series and AV series

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gate way Addr=accordingly
n on Config., Input, Direct Connect, Config.Add,Video,Qual.=Tomoscan SR6000:
n Board/Channel=Video60 Board 1 Channel 1, Logical Name= ctsr,
n Primary Control Type= Conv.Kpad, Com Port=9
n Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n HL=31.36, HD=22.75, Clk=57.14, Tot.Pxls=1792, Image Pxls=1300
n HS=3, HFP=0.61, HBP=5, HB=9.61
n Image Lines=1024, Tot.Lines=1124, VBP=38, VS=4, PreE=PostE=4
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 36


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Bla ck, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 37


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Keypad Kpad / Host Control
Octopus Adaptor Cable
Philips CT Tomoscan SR Kodak 1120 camera

Fiber Cable
Video Card
Octopus Cable
Video Cable
ImageVideo OUT

Kodak MIM 200

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 1120 v2.40/032

Note: same setup valid for Tomoscan CT of SR series and AV series

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config., Input, Direct Connect, Config.Add,Video,Qual.=Tomoscan SR6000:
n Board/Channel=Video60 Board 1 Channel 1, Logical Name= ctsr,
n Primary Control Type= Conv.Kpad, Com Port=9
n Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n HL=31.36, HD=22.75, Clk=57.14, Tot.Pxls=1792, Image Pxls=1300
n HS=3, HFP=0.61, HBP=5, HB=9.61
n Image Lines=1024, Tot.Lines=1124, VBP=38, VS=4, PreE=PostE=4
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 38


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 39


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Keypad Kpad / Host Control


Octopus Adaptor Cable ryView
Philips CT Tomoscan SR 3M
87 00 D

Fiber Cable

Video Card
Fiber Cable
Octopus Cable
Video Cable
ImageVideo OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM200 Destination
Note: same setup valid for Tomoscan CT of SR series and AV series

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config., Input, Direct Connect, Config.Add,Video,Qual.=Tomoscan SR6000:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ct,
n Primary Control Type= Conv.Kpad, Com Port=9
n Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n HL=31.36, HD=22.75, Clk=57.14, Tot.Pxls=1792, Image Pxls=1300
n HS=3, HFP=0.61, HBP=5, HB=9.61
n Image Lines=1024, Tot.Lines=1124, VBP=38, VS=4, PreE=PostE=4
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: November ´04 40


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 41


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

Philips CT Tomoscan SR video with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Keypad Kpad / Host Control 9-pole Plug
Octopus Adaptor Cable
Philips CT Tomoscan SR Control 9-pin
COPCIL Board
Copper Cable
Video Card
Octopus Cable
Video Cable
ImageVideo OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Tomoscan SR rel. 1.4F, MIM200 v6.1, 8100 v2.4.1

Note: same setup valid for Tomoscan CT of SR series and AV series

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config., Input, Direct Connect, Config.Add,Video,Qual.=Tomoscan SR6000:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= ct,
n Primary Control Type= Conv.Kpad, Com Port=9
n Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n HL=31.36, HD=22.75, Clk=57.14, Tot.Pxls=1792, Image Pxls=1300
n HS=3, HFP=0.61, HBP =5, HB=9.61
n Image Lines=1024, Tot.Lines=1124, VBP=38, VS=4, PreE=PostE=4
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 42


Philips CT Tomoscan SR / AV with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8100, Media Size Sele ct=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=13, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 43


Philips Mammo Diagnost with LASER IMAGER Products

Setup Manual for Philips Mammo Diagnost


with LASER IMAGERS
Connection via DICOM:

n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /


Kodak DryView 8610 LASER IMAGER
(see also 8600) (1) ............................................................................................................. 2
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGER
(see also 8600) (2,3 ) ........................................................................................................... 4
n Kodak DryView 8900 LASER IMAGER with Mammography Accessory.......................... 6

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.
3
setup valid also for Kodak PACS LINK 25 PRINT SERVER.

Author: gian.valesano@kodak.com, Edition: November ´04 1


Philips Mammo Diagnost with LASER IMAGER Products

Philips Mammo Diagnsot via DICOM with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8610 LASER IMAGING SYS TEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips COPCIL board
Mammo Diagnost Copper Cable
PCR-reader
EasyRAD workstation Data 37 pin

Kodak 9410 PACS Link


Requirements: Philips EasyRAD rel. 4.2 Level 4
Requirements: 9410 v1.6, 8610 v2.4 (if 8600 v2.5)
Note: This setup is basically also valid for 8600 camera

EasyRAD Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 9410, port=1024, AETitle=9410
n click Automatic Config., after a while the result should be Print as SCP,
n (Note : in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8610, Name=dv8610, AE title=9410
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print) (Off if 8600)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

9410 Setup:
n on Device Setup, LOCAL PRINTERS:
n Printer=8610, Model=8600/8610, Output=Copper
n Command Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Pixel=12 (8 if 8600), Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831

Author: gian.valesano@kodak.com, Edition: November ´04 2


Philips Mammo Diagnost with LASER IMAGER Products

n Interp Num=1, Installed TFT/ULUT= don’t care


n Max.Density=300, Contrast=5, Border Density=0
n on Device Setup, SCP PRINTERS, Current Providers:
n Printer=8610, AE Title= 9410, Port= 1024, Max Associations= 5
n on Device Setup, SCP PRINTERS, Registered Users:
n SCU AETitle= accordingly, SCU IPAddr=accordingly
n SCP=8610, Copies=1, Contrast Table=5, Max.Density=300
n Border Density=0, Mag.Type=8, Smoothing=1
n Polarity=1, Orientation=1, Set Req.Image Size=No, Priority=Low
n Gamma=Linear in Transmittance, Contrast Group=1, Kernel Group=1
n Interceptor=Yes, N -Event-Print Job=No, N-Event Printer=No
n Trim=No, Attribute Warning=No
n Contrast Test Mode=No, Contrast Test Mode Density=0

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: November ´04 3


Philips Mammo Diagnost with LASER IMAGER Products

Philips Mammo Diagnost via DICOM with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8610 LASER IMAGING SYSTEM / for Mammography
(see also 8600)

DICOM Network

Kodak DryView 8610


Control 9 pin
Philips COPCIL board
Mammo Diagnost Copper Cable
PCR-reader
EasyRAD workstation Data 37 pin

Kodak MIM 200

Requirements: Philips EasyRAD rel. 4.2 Level 4


Requirements: MIM200 v6.1, 8610 v2.4 (if 8600 v2.5)
Note: This setup is basically also valid for 25 PS computer
Note: This setup is basically also valid for 8600 camera

EasyRAD Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of MIM200, port=5040, AETitle=8610
n click Automatic Config., after a while the result should be Print as SCP,
n (Note : in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= 3M 8610, Name=dv8610, AE title=8610
n into Config. Media List:
n 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Configure, Input, Network, Add:

Author: gian.valesano@kodak.com, Edition: November ´04 4


Philips Mammo Diagnost with LASER IMAGER Products

n Protocol=DICOM 3.0, Log.Name=DICOM, Type/Com=Keypad/COM9


n on Configure, Destination, Add, Print, Direct, Qualified, 8610 (or 8600), Set:
n General: Log.Name=8610 (or 8600)
n General: Database=8610 (or 8600), Model=8610 (or 8600), Memory=32
n Media Info: Media=8x10in, 5025x6200, 6020, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12 (8 if 8600), Header/Line,Transfer=2000,Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=25.8065
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.5
n M952 Usage: Min H/V Mag=0.26, Max H/V Mag=16, Min Dmax 952=2
n on Configure, SCP Services, select the specific modality, Modify:
n PLUT=N, ToneScaling=PerceptionLUT (Always=Y)
n TFT/ULUT= upmsv3kp.w8u (Always=N), Contrast (dry)=5 (Always=N)
n Modality Group=1, Conv.Group=1, Gamma Table=Linear in Transmittance
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n Smoothing Type=1 (Always=Y), Image Polarity=Normal (Always=N)
n NeventReports=Off (Always=Y), Warning Status=On
n Print Date=Off, Print Time=Off, Page Number=Off, Hosp.Name=Off
n Modality ID=Off, Annotation Override=Off (Always=N)
n Minify/Scale=On, Crop=Off, Text Box Relocation=None (Always=N)
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n Film Base=Blue Film (Always=N), Film Size=8INx10IN (Always=N)
n Negotiate Color Print=Off

8610 Setup: on 8610 Operator Menu into Test Menu, Software Version, verify:
n Imager=2.4, VRTX32=1.08, Contrast=6.P.2, Conv.Kernel=M.1.0
n (if 8600: Imager=2.5, Contrast=6.G.7, Conv.Kernel=1.0.5)
n on 8610 Operator Menu into Configure User set:
n Density=16, Contrast=6, Sharp/Smooth=1, Film Layout: 1:1P
n Gamma=Xmittance, Image Size=Scale To Fit, Detector Size=don’t care
n on 8610 Service Menu into Configure User set:
n Ratio=1.00, Serial Control Intf.=1200,8,1,Even,Comma,3M, Memory=255
n Digital Interface: Parity=Even, Modality Group=1, Smoothing Group=1

Author: gian.valesano@kodak.com, Edition: November ´04 5


Philips Mammo Diagnost with LASER IMAGER Products

Author: gian.valesano@kodak.com, Edition: November ´04 6


Philips Mammo Diagnost with LASER IMAGER Products

Philips Mammo Diagnost via DICOM with


Kodak DryView 8900 LASER IMAGER with Mammography Accessory

DICOM Network

Philips
Mammo Diagnost
PCR-reader
EasyRAD workstation

Kodak DryView 8900

Requirements: Philips EasyRAD rel. 4.2 Level 5


Requirements: 8900 rev.E ie v2.11 (ie DRE v2.11, MIM v5.6.b5, MCS v2.10)

EasyRAD Setup:
n after power up select 4=Install Config, 2=Start Config
n below window containing list of Config.Nodes, click New, choose DICOM Node
n set IP addr.=addr.of 8900, port=5040, AETitle=8900
n click Automatic Config., after a while the result should be Print as SCP,
n (Note : in case of a DICOM connection problem set Max PDU=65536,
n on SCP Settings: do not run Automatic Configuration ie do not click on it
n on SCP Settings: de-click all others possibilities, leave selected only Print as SCP
n on SCP Settings, Extended Settings, on SCP Transfer Syntaxes:
n de-click all others possibilities, leave selected only Implicit Little Endian, Exit)
n Proceed
n inside window containing list of Config.Nodes select the EV Node and click Edit
n the whole menu is called EasyVision Config, select Peripherals settings
n below window containing list of Printers defined for EV, click New
n set Interface=DICOM, Printer= Kodak 8900 double (or 3M HQ 969),
n Name=dv8900, AE title=8900
n into Config. Media List: 8x10 G8=On (8-bit-print), 8x10 G12=On (12-bit-print)
n Proceed, Exit, Proceed, Execute, Proceed, click on handle, click on triangle

8900 Setup:
n on Configuration, System, Network:
n Host Name=DV8900, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr= accordingly
n on Configuration, System, Software License:
n License Status=Mammo Key is ON
n on Configuration, DICOM SCP, Communications:
n Port=5040, Association Limit=12, Source Name=DICOM

Author: gian.valesano@kodak.com, Edition: November ´04 7


Philips Mammo Diagnost with LASER IMAGER Products

n on Configuration, Destination:
n General: Log.Name=8900
n Media Types: Media=35x43cm, 9100x11037, 10781, Any
n Media Types: Media=35x35cm,9100x9087, 8831, Any
n Media Types: Media=11x14in, 7137x9100, 8844, Any
n Media Types: Media=10x12in, 6490x7767, 7511, Any
n Media Types: Media=8x10in, 5167x6490, 6234, Any
n Formatting: Aspect Ratio Mix Page=On, Aspect Ratio Mix Row=On
n Formatting: Image Size Mix in Std Fmt=On, Image Size Mix Row=On
n Formatting: DICOM Annotation Box=On, Req.Image Size=On
n Formatting: Req.Size Scale=On, Req.Size Mix Page=On
n Formatting: Req.Size Mix Row=On, Trim=On,
n Formatting: True Landscape=On, Rotation=On
n Formatting: Minify/Crop=On, Tonescaling on Image Basis=On
n Formatting: Max Img Band=7, Max Band Page=7, Max Img Cols=0
n Formatting: Max Img Rows=0, Pixel Pitch=25.5910, Min Horiz Sep=4
n Formatting: Min Vert Sep=4, Trim Width Enabled=4
n Formatting: Trim Width Disabled=0, Maximum DMAX=3.30
n on Configuration, DICOM SCP, SCP Services, select specific modality, Modify:
n Negotiate Presentation LUT=No, Negotiate Color Print=No
n TFT=Pmsv3k.w87 (Always=Y), Contrast=9 (Always=N)
n Smoothing Type=0 (Always=N), Image Polarity (Always=N)
n Image Dmin=0.00 (Always=N), Image Dmax=3.00 (Always=N)
n NeventReports=No (Always=Y), Warnings=Yes
n Print Priority=Normal (Always=N), Border=3.99 (Always=N)
n 2mm Border=Yes, Film Clip Border=No, Film Base=Mammo Blue Film (Always=Y)
n Film Size=8INx10IN (Always=N), Sorter Bin=Bin1 (Always=Y)
n Print Date=No, Print Time=No, Page Nbr=No, Hosp.Name=No
n Modality ID=No, Annotation Override=No, Density Patch=No
n Image Resizing=Minify/Scale (Always=N)
n Text Box Location=None, Source ID=ws

Author: gian.valesano@kodak.com, Edition: November ´04 8


Philips MR Gyroscan T5 with some LASER IMAGER Products

Setup Manual for Philips MR Gyroscan T5


(old Philips modality, no more in production)
with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual................................................................ 2
n Kodak DryView 8700 LASER IMAGER / Plus ................................................................. 3
n Kodak DryView 8700 LASER IMAGER / Standard ......................................................... 4
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual................................................................ 5
n Kodak DryView 8500 LASER IMAGER / Plus ................................................................. 6
n Kodak DryView 8500 LASER IMAGER / Standard ......................................................... 7
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual................................................................ 8
n Kodak DryView 8300 LASER IMAGER / Standalone ...................................................... 9
n Kodak Digital Science 969 HQ LASER IMAGER ....................................................... 10
n Kodak Digital Science 969 HQT LASER IMAGER..................................................... 11
n Kodak Digital Science 959 XL LASER IMAGER ........................................................ 12
n Kodak DryView 8100 LASER IMAGER.......................................................................... 13
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) ..... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1) ..................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) .................................................................. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ( 2,3,4)............................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ( 2,3,4)............................................................... 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) ..... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4) ..................................................................... 26

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: November ´04 1


Philips MR Gyroscan T5 with some LASER IMAGER Products

Author: gian.valesano@kodak.com, Edition: November ´04 2


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 25-pin RS232 25-to-9 pin


Philips MR Gyroscan T5 Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

r
na
ge
Ma
put
ti -I n
ul
00M
88
3M

ry Vie w
700 D
3M8

Fiber Cable
8700 Dual

8800 MIM

Requirements: Gyroscan T5 rel. R3.6.2, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8800/8700 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 3


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 25-pin RS232 25-to-9 pin


Philips MR Gyroscan T5
Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232OUT
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

View
0 0 Dry
3 M 87

8700 Plus

Requirements: Gyroscan T5 rel. R3.6.2, 8700 Plus rel. 1.15

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 4


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard
Philips 25-pin RS232 25-to-9 pin
Philips MR Gyroscan T5
Control Cable Host Adaptor Cable
Host Control 232 OUT
yView
700 D r
3M 8

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroscan T5 rel. R3.6.2, 8700 Standard rel. 1.15

Setup: on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8700 Standard Setup:

n on TDB/C board:
n Jumpers=RS232, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 5


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 25-pin RS232 25-to-9 pin


Philips MR Gyroscan T5 Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
ag
n
t Ma
npu
lti I-
0Mu
88
0
3M

ie w
0Dr yV
3 M870

Fiber Cable
8500 Dual

8800 MIM

Requirements: Gyroscan T5 rel. R3.6.2, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8800/8500 Dual Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 6


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 25-pin RS232 25-to-9 pin


Philips MR Gyroscan T5
Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232OUT
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

View
0 0 Dry
3 M 87

8500 Plus

Requirements: Gyroscan T5 rel. R3.6.2, 8500 Plus rel. 1.22

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8500 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 7


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard
Philips 25-pin RS232 25-to-9 pin
Philips MR Gyroscan T5
Control Cable Host Adaptor Cable
Host Control 232 OUT
yView
700 D r
3M 8

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroscan T5 rel. R3.6.2, 8500 Standard rel. 1.22

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8500 Standard Setup:


n on TDB/C board:
n Jumpers=RS232, Switch=Center
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 8


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips MR Gyroscan T5 Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
ager
an
pu
tM
i- In
ult

Fiber Cable
00M
88
3M

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: Gyroscan T5 rel. R3.6.2, 8800 rel. 5.1, 8300 rel. 4.2

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0

8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 9


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 25-pin RS232 25-to-9 pin
Philips MR Gyroscan T5
Control Cable Host Adaptor Cable
Host Control 232 OUT

Philips 37-pin Data Cable


Image Data OUT

Requirements: Gyroscan T5 rel. R3.6.2, 8300 rel. 4.5

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Even
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: November ´04 10


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak Digital Science  969 HQ LASER IMAGER

Philips MR Gyroscan T5 Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable
Fiber Cable

3 M H Q L as e r I m a ge r

M969 HQ

Requirements: Gyroscan T5 rel. R3.6.2, M969 HQ rel. 3.0

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

M969 HQ Setup:
n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu:
n Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 11


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 25-pin RS232 25-to-9 pin
Philips MR Gyroscan T5
Control Cable Host Adaptor Cable 3 M H Q L a ser I m ag e r

Host Control 232 OUT

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroscan T5 rel. R3.6.2, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 12


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak Digital Science  959 XL LASER IMAGER

Philips MR Gyroscan T5 Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M L a s er I m ag er

M959 XL

Requirements: Gyroscan T5 rel. R3.6.2, M959 XL rel. 3.8

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center
n on XL Input Module install LUT 322A0

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pino ut 9 pin side 4 6 1

M959 XL Setup:
n on COMM Menu:
n 959 Default
n on EIB Menu:
n Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 13


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak DryView 8100 LASER IMAGER

RS232 9-pole
Philips 25-pin RS232 25-to-9 pin plug
Philips MR Gyroscan T5 Host Adaptor Cable
Control Cable
Host Control 232 OUT
Philips 37-pin
Data Cable
Image Data OUT

DryView 8100

Requirements: Gyroscan T5 rel. R3.6.2, 8100 v1.1

Setup: on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=5, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 14


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips MR Gyroscan T5 Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 232 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC 25-9 DPC-UKEIB Cable

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


0 0 Dry
View
Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Gyroscan T5 rel. R3.6.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

Author: gian.valesano@kodak.com, Edition: November ´04 15


Philips MR Gyroscan T5 with some LASER IMAGER Products

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=5, Parity=Even, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=mrt5, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 16


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 25-pin
RS232 25-to-9 pin
Philips MR Gyroscan T5 Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232 OUT Imager

Philips 37-pin
Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Gyroscan T5 rel. R3.6.2, 9410 v1.6, 8100 v2.4.1

Setup:
n on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=5, Parity=Even, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=mrt5, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

Author: gian.valesano@kodak.com, Edition: November ´04 17


Philips MR Gyroscan T5 with some LASER IMAGER Products

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 18


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER
Philips MR
Philips 25-pin 25-to-25 Null Kpad / Host Control
Gyroscan T5 RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 160 camera
Crossed
Host Control Converter Box
Network Cable
232 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Gyroscan T5 rel. R3.6.2, MIM200 v6.1, KELI 160 v7.2

Setup: on Gyroscan T5 install Camera Config.File KODAK, LUT=NONE


Note: a Null Modem Cable may be needed for RS232 connection

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroscan ACS T5 MRI:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= mrt5,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=722, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2

Author: gian.valesano@kodak.com, Edition: November ´04 19


Philips MR Gyroscan T5 with some LASER IMAGER Products

n Formatting: Trim Width Enabled=2, Trim Width Disabled=0


n Formatting: Pixel Pitch=12.8
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 20


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER
Philips MR
Philips 25-pin 25-to-25 Null Kpad / Host Control
Gyroscan T5 RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 2180 camera
Host Control Converter Box
Fiber Cable
232 OUT
RS422 37 -to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Gyroscan T5 rel. R3.6.2, MIM200 v6.1, 2180 v2.17H

Setup: on Gyroscan T5 install Camera Config.File KODAK, LUT=NONE


Note: a Null Modem Cable may be needed for RS232 connection

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroscan ACS T5 MRI:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= mrt5,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=722, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 21


Philips MR Gyroscan T5 with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=2
n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 22


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER
Philips MR
Philips 25 -pin 25-to-25 Null Kpad / Host Control
Gyroscan T5 RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 1120 camera
Host Control Converter Box
Fiber Cable
232 OUT
RS422 37-to-50 pin
Philips 37 -pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Gyroscan T5 rel. R3.6.2, MIM200 v6.1, 1120 v2.40/032

Setup: on Gyroscan T5 install Camera Config.File KODAK, LUT=NONE


Note: a Null Modem Cable may be needed for RS232 connection

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroscan ACS T5 MRI:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= mrt5,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=722, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 1200, RS422, 8, 1, Even
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0

Author: gian.valesano@kodak.com, Edition: November ´04 23


Philips MR Gyroscan T5 with some LASER IMAGER Products

n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0
n Formatting: Trim Width Enabled=0, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 24


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips MR
Philips 25-pin 25-to-25 Null Kpad / Host Control
Gyroscan T5 RS23-to-RS422 Dry Vie
w

Control Cable Modem Cable Octopus Adaptor Cable 3M 8


70 0

25-to-15 pin
Fiber Cable
Host Control Converter Box
232 OUT
RS422 37-to-50 pin Fiber Cable
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Gyroscan T5 rel. R3.6.2, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM 200 Destinations

Setup: on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroscan ACS T5 MRI:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= mr,
n Mod.Name= mr, Primary Control: Type=M952/P831 , Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=722, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: November ´04 25


Philips MR Gyroscan T5 with some LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 26


Philips MR Gyroscan T5 with some LASER IMAGER Products

Philips MR Gyroscan T5 with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Philips MR Philips 25-pin 25-t o-25 Null Kpad / Host Control 9-pole Plug
Gyroscan T5 RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable Control 9-pin
25-to-15 pin COPCIL Board
Host Control Converter Box Copper Cable
232 OUT
Philips 37-pin RS422 37-t o-50 pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Gyroscan T5 rel. R3.6.2, MIM200 v6.1, 8100 v2.4.1

Setup: on Gyroscan T5 install Camera Config.File LI3M Plus, LUT=NONE

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroscan ACS T5 MRI:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= mr,
n Mod.Name= mr, Primary Control: Type=M952/P831 , Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=722, Lines=528, Parity=Even, Timer=5, Sync=On, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4,1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media =35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=O n, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: November ´04 27


Philips MR Gyroscan T5 with some LASER IMAGER Products

n Formatting: Pixel Pitch=12.8205


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 28


Philips workstation Gyroview with some LASER IMAGER Products

Setup for Philips workstation Gyroview


(old Philips modality, no more in production)
with LASER IMAGERS
Connection POINT-TO-POINT:

n Kodak DryView 8800 MULTI-INPUT MANAGER /


Kodak DryView 8700 LASER IMAGER / Dual................................................................ 2
n Kodak DryView 8700 LASER IMAGER / Plus ................................................................. 3
n Kodak DryView 8700 LASER IMAGER / Standard ......................................................... 4
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual................................................................ 5
n Kodak DryView 8500 LASER IMAGER / Plus ................................................................. 6
n Kodak DryView 8500 LASER IMAGER / Standard ......................................................... 7
n Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual................................................................ 8
n Kodak DryView 8300 LASER IMAGER / Standalone ...................................................... 9
n Kodak Digital Science 969 HQ LASER IMAGER ....................................................... 10
n Kodak Digital Science 969 HQT LASER IMAGER..................................................... 11
n Kodak Digital Science 959 XL LASER IMAGER ........................................................ 12
n Kodak DryView 8100 LASER IMAGER.......................................................................... 13
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (1) ..... 14
n Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER (1) ..................................................................... 16
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER (2,3,4) .................................................................. 18
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER ( 2,3,4)............................................................... 20
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER ( 2,3,4)............................................................... 22
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus (see also 8800/8700 Dual and HQ) (4) ..... 24
n Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER (4) ..................................................................... 26

1
Setup valid also for Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.3 and v1.5.1.
2
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v3.2.1.
3
Setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v4.0.
4
setup valid also for Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v5.0, v5.2, v6.0.

Author: gian.valesano@kodak.com, Edition: November ´04 1


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8700 LASER IMAGER / Dual

Philips 25-pin RS232 25-to-9 pin


Philips Gyroview Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

er
ag
n
t Ma
npu
lti I-
0Mu
88
0
3M

ie w
0Dr yV
3 M870

Fiber Cable
8700 Dual

8800 MIM

Requirements: Gyroview rel. 2.1.2, 8800 rel. 4.2, 8700 rel. 1.15

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8800/8700 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 2


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8700 LASER IMAGER / Plus

Philips 25-pin RS232 25-to-9 pin


Philips Gyroview
Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232OUT
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

View
0 0 Dry
3 M 87

8700 Plus

Requirements: Gyroview rel. 2.1.2, 8700 Plus rel. 1.15

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8700 Plus Setup:


n on COMM Menu:
n 959 Default
n on IMAGE Menu:
n TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu:
n Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 3


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8700 LASER IMAGER / Standard

8700 Standard
Philips 25-pin RS232 25-to-9 pin
Philips Gyroview
Control Cable Host Adaptor Cable
Host Control 232 OUT
yView
700 D r
3M 8

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroview rel. 2.1.2, 8700 Standard rel. 1.15

Setup: on Gyroview install Camera Config.File 3M +

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8700 Standard Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 4


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8500 LASER IMAGER / Dual

Philips 25-pin RS232 25-to-9 pin


Philips Gyroview Host Adaptor Cable
Control Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

r
na
ge
Ma
put
ti -I n
ul
00M
88
3M

ry Vie w
700 D
3M8

Fiber Cable
8500 Dual

8800 MIM

Requirements: Gyroview rel. 2.1.2, 8800 rel. 5.1, 8500 rel. 1.22

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8800/8500 Dual Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 5


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8500 LASER IMAGER / Plus

Philips 25-pin RS232 25-to-9 pin


Philips Gyroview
Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232OUT
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

View
0 0 Dry
3 M 87

8500 Plus

Requirements: Gyroview rel. 2.1.2, 8500 Plus rel. 1.22

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8500 Plus Setup:


n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 6


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8500 LASER IMAGER / Standard

8500 Standard
Philips 25-pin RS232 25-to-9 pin
Philips MR Gyroscan T5
Control Cable Host Adaptor Cable
Host Control 232 OUT
yView
700 D r
3M 8

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroview rel. 2.1.2, 8500 Standard rel. 1.22

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8500 Standard Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 7


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8800 MULTI-INPUT MANAGER /
Kodak DryView 8300 LASER IMAGER / Dual

Philips Gyroview Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB
Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

8300 Dual
ager
an
pu
tM
i- In
ult

Fiber Cable
00M
88
3M

9-to-9-pin Copper Cable

37-pin Digital Cable POEIB

8800 MIM

Requirements: Gyroview rel. 2.1.2, 8800 rel. 5.1, 8300 rel. 4.2

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8300 Setup:
n on 8300 Operator Menu set:
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=None
n Modality Group=0

8800 Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: Density=16, Contrast=10
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 8


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8300 LASER IMAGER / Standalone

8300
Philips 25-pin RS232 25-to-9 pin
Philips Gyroview
Control Cable Host Adaptor Cable
Host Control 232 OUT

Philips 37-pin Data Cable


Image Data OUT

Requirements: Gyroview rel. 2.1.2, 8300 rel. 4.5

Setup:
n on Gyroview install Camera Config.File 3M +

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8300 Setup:
n on 8300 Operator Menu set:
n Density=16
n Contrast=10
n Film Layout : Format=1on 1, Mode=Portrait
n on 8300 Service Menu set:
n Serial Control Interface: Baud Rate=1200, Nbr. of Data bits=8,
Nbr. of Stop bits=1, Parity=Even, Protocol=3M
n Digital Interface: Parity=Even
n Modality Group=0

Author: gian.valesano@kodak.com, Edition: November ´04 9


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak Digital Science  969 HQ LASER IMAGER

Philips Gyroview Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable
Fiber Cable

3 M H Q L as e r I m a ge r

M969 HQ

Requirements: Gyroview rel. 2.1.2, M969 HQ rel. 3.0

Setup:
n on Gyroview install Camera Config.File 3M +
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

M969 HQ Setup:
n on COMM Menu: 959 Default
n on IMAGE Menu: TFT=PMSV3K, Dmax=3.00, Contrast=9, Density=14
n on HOST Menu:
n Pixel Correction=Yes, Scale Format=Yes, Scale Magnification=Yes
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 10


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak Digital Science 969 HQT LASER IMAGER

M969 HQT
Philips 25-pin RS232 25-to-9 pin
Philips Gyroview
Control Cable Host Adaptor Cable 3 M H Q L a ser I m ag e r

Host Control 232 OUT

Philips 37-pin
Image Data OUT Data Cable

Requirements: Gyroview rel. 2.1.2, M969 HQT rel. 1.0

Note: in case of image quality problems update HQT to rel. 2.0

Setup:
n on Gyroview install Camera Config.File 3M +

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

M969 HQT Setup:


n on TDB/C board: Jumpers=RS232, Switch=Center
n on COMM Menu: 959 Default
n on HOST Menu: Pixel Correction=Yes, Scale Format=Yes, Scale Magnif.=Yes
n on IMAGE Menu: (note: values suitable for 60 secs Kodak APS chemistry)
n TFT=PMS, Contrast=15, Density=7, on OIC Menu: Chemistry=2
n on SYSTEM Menu: Memory Management=Best Fit
n on EIB Menu: Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 11


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak Digital Science  959 XL LASER IMAGER

Philips Gyroview Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host Control 232 OUT Host UKEIB

Philips 37-pin
Data Cable
Image Data OUT DEIB

Fiber Cable

Fiber Cable

3M L a s er I m ag er

M959 XL

Requirements: Gyroview rel. 2.1.2, M959 XL rel. 3.8

Setup:
n on Gyroview install Camera Config.File 3M +
n on XL Control Panel set Density & Contrast depending on chemistry
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center
n on XL Input Module install LUT 322A0

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

M959 XL Setup:
n on COMM Menu:
n 959 Default
n on EIB Menu:
n Parity=Enable Even, Timeout=5

Author: gian.valesano@kodak.com, Edition: November ´04 12


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak DryView 8100 LASER IMAGER

Philips 25-pin RS232 9-pole


Philips Gyroview RS232 25-to-9 pin plug
Control Cable Host Adaptor Cable
Host Control 232 OUT

Philips 37-pin
Image Data OUT Data Cable

DryView 8100

Requirements: Gyroview rel. 2.1.2, 8100 v1.1

Setup: on Gyroview install Camera Config.File 3M+

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=5, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=On, Pixel Correct=On, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=PMSB
n on LOCAL PANEL Menu: Contrast=9, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=8, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 13


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips Gyroview Philips 25-pin RS232 25-to-9 pin


Control Cable Host Adaptor Cable
Host UKEIB
Host Control 232 OUT Imager

Philips 37-pin Copper Cable


Image Data OUT Data Cable DPC 25-9 DPC-UKEIB Cable

DPI4-to-DPC Cable

DryView 8700 Plus 3M 87


0 0 Dry
View
Fiber Cable

Fiber Cable 9410 PACS Link

Requirements: Gyroview rel. 2.1.2, 9410 v1.6, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into 9410 Local Printers

Setup:
n on Gyroview install Camera Config.File 3M+
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8700, Model=8700, Output=Fiber
n Baud=19200, Parity=None, Data=8, Stop=1, Command Set=SuperSet
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=2 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=5, Parity=Even, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=gv, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA

Author: gian.valesano@kodak.com, Edition: November ´04 14


Philips workstation Gyroview with some LASER IMAGER Products

n Available Printer=8700, Film Destination=Processor, Printer1=8700(Processor)


n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsv3k.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 15


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK 9410 ACQUISITION SYSTEM Software v1.6 /
Kodak DryView 8100 LASER IMAGER

Philips 25-pin
RS232 25-to-9 pin
Philips Gyroview Control Cable
Host Adaptor Cable
Host UKEIB
Host Control 232 OUT Imager

Philips 37-pin
Image Data OUT Data Cable DPC DPC-to-UKEIB Cable

DryView 8100 DPI4 Cable


RS422
9-pole Plug Control 9-pin
COPCIL Board
Copper Cable

9410 PACS Link


Data 37 pin

Requirements: Gyroview rel. 2.1.2, 9410 v1.6, 8100 v2.4.1

Setup:
n on Gyroview install Camera Config.File 3M+
n on UKEIB set OFF/ON/OFF/OFF/ON/OFF, Switch=Center

RS232 25-to-9 Host 25 pin side 2 3 7


Adaptor Cable pinout 9 pin side 4 6 1

9410 Setup:
on Device Setup, LOCAL PRINTERS:
n Printer=8100, Model=8100 4096 pxls/line, Output=Copper
n Baud=1200, Parity=Even, Data=8, Stop=1, Command Set=952
n Pixel=12, Header/Line, Timeout=60, Transfer Rate=1 MHz
n Film Mismatch=Print Always, Interpolation Type=831
n Interp Num=2, Max.Density=300, Contrast=6, Border Density=0
n TFT/ULUT Emulation=On, Installed TFT/ULUT=Ver693c0.w87
on I/O CARD SETUP, DPI4 Digital Menu:
n Image Data Port: Header, 8 bit, Timeout=5, Parity=Even, Memory=7
n Serial Port: Protocol=952, Parity=Even, Baud Rate=1200
on Device Setup, HOST KEYPADS:
n Kpad=gv, Hosp=philips, Serial=DPI4 A Host, Image=DPI4 A DATA
n Available Printer=8100, Film Destination=Processor, Printer1=8100(Processor)
n 300 DPI=Off, Scale Magnif.=Off, Scale Format / Pixel Correction=On
n TFT/ULUT=Pmsb.w87, Mem.Full =On, Image Size =On, Alarm Mode=On
n into Defaults: Destination=Printer 1, Media Size=14x17, C=9, D=14

Author: gian.valesano@kodak.com, Edition: November ´04 16


Philips workstation Gyroview with some LASER IMAGER Products

8100 Setup on CONFIGURATION Menu:


n on SIB board (i.e., external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 17


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with|


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 160 LASER IMAGER

Philips Gyroview Philips 25-pin 25-to-25 Null Kpad / Host Control


RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 160 camera
Crossed
Host Control Converter Box
Network Cable
232 OUT
RS422 37-to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Gyroview rel. 2.1.2, MIM200 v6.1, KELI 160 v7.2

Setup: on Gyroview install Camera Config.File KODAK

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak160, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroview Workstation:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= gv,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=540, Parity=Even, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELI 160, Set:
n General: Log.Name=KELI 160, IPAddr=100.100.100.30,
n General: Database Name=KELI 160, Model=KELI 160
n Media Info: Media=35x43cm, 4361x5348, 5228, Any (Blue,Clear)
n Supported Item: Collation=On, Text=On, Tonescaling on Image=On
n Supported Item: Pivot Density=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0

Author: gian.valesano@kodak.com, Edition: November ´04 18


Philips workstation Gyroview with some LASER IMAGER Products

n Formatting: Pixel Pitch=12.8


n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination=KELI 160, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

160 Setup: no modality-specific setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 19


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 2180 LASER PRINTER

Philips Gyroview Philips 25-pin 25-to-25 Null Kpad / Host Control


RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 2180 camera
Host Control Converter Box
Fiber Cable
232 OUT
RS422 37 -to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requireme nts: Gyroview rel. 2.1.2, MIM200 v6.1, 2180 v2.17H

Setup: on Gyroview install Camera Config.File KODAK

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak2180, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroview Workstation:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= gv,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=540, Parity=Even, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 2180, Set:
n General: Log.Name=KELP2180, Database Name=KELP 2180
n General: Model=KELP 2180, Memory=24, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4136x5160, 5030, (Blue)
n Media Info:Media=35x35cm,4136x4136,4006, (Blue)
n Media Info: Media=11x14in, 3240x4136, 4006, (Blue)
n Media Info: Media=8x10in, 2290x2840, 2710, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bil=On, Repl=On,Cubic=On, Cubic 2=On
n Format Types: Custom=On, Slide=On, Row Symmetric =On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2

Author: gian.valesano@kodak.com, Edition: November ´04 20


Philips workstation Gyroview with some LASER IMAGER Products

n Formatting: Trim Width Enabled=2, Trim Width Disabled=2


n Formatting: Pixel Pitch=12.6582
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.2
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 2180, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

2180 Setup:
n Optical Interface board will automatic configure into 2180, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 21


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak Ektascan 1120 LASER PRINTER

Philips Gyroview Philips 25-pin 25-to-25 Null Kpad / Host Control


RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable
25-to-15 pin Kodak 1120 camera
Host Control Converter Box
Fiber Cable
232 OUT
RS422 37 -to-50 pin
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200

Requirements: Gyroview rel. 2.1.2, MIM200 v6.1, 1120 v2.40/032

Setup: on Gyroview install Camera Config.File KODAK

MIM200 Setup:
n on Configure, Network:
n Computer Name=eg Kodak1120, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroview Workstation:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= gv,
n Primary Control: Type=KCL, Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=540, Parity=Even, Timer=1, Sync=O ff, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n 9600, RS422, 8, 1, None
n on Configure, Destination, Add, Print, Direct, Qualified, KELP 1120, Set:
n General: Log.Name=KELP1120, Database Name=KELP 1120
n General: Model=KELP 1120, Memory=20, Node ID=6 or accordingly
n Media Info: Media=35x43cm, 4096x5132, 5034, (Blue)
n Media Info: Media=35x35cm, 4096x4108, 4010, (Blue)
n Media Info: Media=11x14in, 4096x3232, 3134, (Blue)
n Media Info: Media=8x10in, 2304x2900, 2802, (Blue)
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=3, Min Vert Sep=0

Author: gian.valesano@kodak.com, Edition: November ´04 22


Philips workstation Gyroview with some LASER IMAGER Products

n Formatting: Trim Width Enabled=0, Trim Width Disabled=0


n Formatting: Pixel Pitch=12.5
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=2.8
n M952 Usage: Min H/V Mag=0, Max H/V Mag=0, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup:
n Select Destination= KELP 1120, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Film Contrast=+0, Interpolation=Cubic Spline
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n Curve Shape=3.0, Auto Study=Off, Console=On
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=CurveShape

1120 Setup:
n Optical Interface board will automatic configure into 1120, no setup needed

Author: gian.valesano@kodak.com, Edition: November ´04 23


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8700 LASER IMAGER / Plus
(see also 8800/8700 Dual and HQ)

Philips Gyroview Philips 25-pin 25-to-25 Null Kpad / Host Control


RS23-to-RS422 Dry Vie
w

Control Cable Modem Cable Octopus Adaptor Cable 3M 8


70 0

25-to-15 pin
Fiber Cable
Host Control Converter Box
232 OUT
RS422 37-to-50 pin Fiber Cable
Philips 37-pin
Digital Adaptor Cable
Data Cable
Image Data OUT

Kodak MIM 200


Kodak DryView 8700 Plus

Requirements: Gyroview rel. 2.1.2, MIM200 v6.1, 8700 v1.34

Note: This setup is basically also valid for 8800/8700 Dual and HQ cameras
by basically using the correspondent Model into MIM 200 Destinations

Setup: on Gyroview install Camera Config.File 3M+

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroview Workstation:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= gv,
n Mod.Name=gv, Primary Control: Type=M952/P831 , Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=540, Parity=Even, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4, 1200, RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8700, Set:
n General: Log.Name=8700
n General: Database Name=8700, Model=8700, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd=SuperSet,Baud=19200,Data=8,Stop=1,Parity=None
n Output Params: Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On

Author: gian.valesano@kodak.com, Edition: November ´04 24


Philips workstation Gyroview with some LASER IMAGER Products

n Formatting: Req.Image Size=On, Req.Size Mix on Page=On


n Formatting: True Landscape=On
n Formatting: Max Img per Band=20, Max Band per Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
n Formatting: Trim Width Enabled=2, Trim Width Disabled=0
n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8700, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Typ e=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsv3k.w87, Auto Study=Off, Console=On, Annotation=Off

8700 Plus Setup:


n on COMM Menu: 969 Default
n on IMAGE Menu: TFT=VER693C0, Dmax=3.00, Contrast=6, Density=14
n on EIB Menu: Pixel=12, Header/Line, Parity=Enable Even, Port=0, Timeout=60

Author: gian.valesano@kodak.com, Edition: November ´04 25


Philips workstation Gyroview with some LASER IMAGER Products

Philips Gyroview with


Kodak PACS LINK MEDICAL IMAGE MANAGER 200 Software v6.1 /
Kodak DryView 8100 LASER IMAGER
RS422
Philips Gyroview Philips 25-pin 25-t o-25 Null Kpad / Host Control 9-pole Plug
RS23-to-RS422
Control Cable Modem Cable Octopus Adaptor Cable Control 9-pin
25-to-15 pin COPCIL Board
Host Control Converter Box Copper Cable
232 OUT
Philips 37-pin RS422 37-t o-50 pin
Digital Adaptor Cable
Data Cable
Image Data OUT
Data 37 pin
Kodak MIM 200
Kodak DryView 8100

Requirements: Gyroview rel. 2.1.2, MIM200 v6.1, 8100 v2.4.1

Setup: on Gyroview install Camera Config.File 3M+

MIM200 Setup:
n on Configure, Network:
n Node Name=MIM, IPAddr=accordingly
n Subnet Mask=accordingly, Gateway Addr=accordingly
n on Config, Input, Direct Connect,Config.Add,Dig.,Qual.=Gyroview Workstation:
n Board/Channel=Digital Board 1 Channel 1, Logical Name= gv,
n Mod.Name=gv, Primary Control: Type=M952/P831 , Com Port=4
n Aux.Control=Conv.Kpad, Com Port=5, Ratio=1,1, Page Prealloc.=Off
n on Configure, Input, Direct Connect, Parameters Acquisition, Modify, Set:
n Pixels=720, Lines=540, Parity=Even, Timer=1, Sync=Off, Pixel=8
n on Configure, Input, Direct Connect, Parameters Autofilming, Modify, Set:
n COM4, 1200 , RS422, 8, 1, Even, Mem Full=BSY
n on Configure, Destination, Add, Print, Direct, Qualified, 8100, Set:
n General: Log.Name=8100
n General: Database Name=8100, Model=8100, Memory=32
n Media Info: Media=35x43cm, 4096x5223, 5133, Blue
n Output Params: Cmd Set=952, Baud=1200, Data=8, Stop=1, Parity=Even
n Output Params: Pixel=12, Header/Line, Transfer=1000 KHz, Acquire=60
n Supported Item: Collation=On, Text=On, Tonescaling=On, Rotation=On
n Interpolation: NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
n Format Types: Custom=On, Slide=On, Row Symmetric=On
n Format Types: 1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
n Formatting: Aspect Ratio Mix Page=On, Image Size Mix in Std Fmt=On
n Formatting: DICOM Annotation Box=On, Req.Size Scale=On
n Formatting: Req.Size Mix in Row=On, Trim=On
n Formatting: Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
n Formatting: Req.Image Size=On, Req.Size Mix on Page=On
n Formatting: True Landscape=On
n Formatting: Max Img Band=20, Max Band Page=20, Max Img Col=0
n Formatting: Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2

Author: gian.valesano@kodak.com, Edition: November ´04 26


Philips workstation Gyroview with some LASER IMAGER Products

n Formatting: Trim Width Enabled=2, Trim Width Disabled=0


n Formatting: Pixel Pitch=12.8205
n Image Processing: Grayscale=On, Minify=On, Max Image Dmax=3.1
n M952 Usage: Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
n at MIM Kpad: press Control (Ready to Store appears), Menu, Service Menu:
n into Machine Setup Menu set: Tonescaling=Perception LUT
n at MIM Kpad: press Control (Ready to Store appears), Menu, Destination Setup
n Select Destination= 8100, Media Size Select=35x43
n at MIM Kpad: press Control (Ready to Store appears), Menu, Imaging Options:
n Orientation=P, Copies=01, Smoothing Type=11, Film Contrast=9
n Image Dmax=3.00, Border=Black, Collation=Off, Auto Print=On
n TFT/ULUT=Pmsb.w87, Auto Study=Off, Console=On, Annotation=Off

8100 Setup on CONFIGURATION Menu:


n on SIB board (ie external switch below Control plugs): SW1=Down
n on COMMUNICATIONS Menu:
n Parity=Even, Stop=1, Data=8, 1200 baud, EOM=CR, Protocol=831/952
n Mem.Full=BSY, Alarm=Old, Acquire Timeout=60, P1,…,P6=0
n on IMAGE QLTY-> Mag Scaling=Off, Pixel Correct=Off, 952 Override=Off
n on IMAGE QUALITY-> TFT FILES Menu: TFT=ver693c0.w87
n on LOCAL PANEL Menu: Contrast=6, Density=3.00
n on SYSTEM Menu: Acquire=Digital, Max.Img.Columns=4361
n on DIGITAL Menu: Pixel Depth=12, Header/Line, Parity=Even

Author: gian.valesano@kodak.com, Edition: November ´04 27

S-ar putea să vă placă și