Documente Academic
Documente Profesional
Documente Cultură
Introduction
pressure and enable a smooth transition from one mode or pressure to another
EasyCare Tx Software
Tx Link
System Components
VPAP Tx Therapy Device A flexible, all-in-one device makes adjustments to Tx Link Provides connectivity between the software and
therapy device
Relays real-time signals measured by the therapy
the mask at the bedside and ensures the selected mask has a proper seal before starting therapy
Prescription Report Seamlessly incorporates the final
mask leak data, eliminating the need to reference charts to calculate appropriate leak values
VPAP Tx delivers all ResMed therapy modes and clinically superior algorithms
CPAP Fixed pressure delivered with optional T (Timed) The fixed respiratory rate and the fixed
inspiration/expiration time set by the clinician are supplied regardless of patient effort
VAuto Automatically adjusts pressure in response to
to snore, flow limitation and obstructive sleep apneas with no pressure support
Bilevel Delivers two treatment pressuresone for
inspiration (IPAP) and one for expiration (EPAP)and provides control over the following bilevel therapy modes:
S (Spontaneous) Follows the patients natural
flow limitation, snore and apneas; pressure support (PS) is fixed throughout the night and can be set by the clinician
Adaptive Servo-Ventilation Treats central sleep apnea
breathing pattern, allowing them to breathe at their regular respiratory rate and rhythm
S/T (Spontaneous/Timed) Augments any breaths
initiated by the patient, but will also supply additional breaths if the breath rate falls below the clinicians set backup respiratory rate
TEChniCaL SPECiFiCaTionS
Pressure Ranges CPAP: 420 cm H2O (EPR 0-3) AutoSet: APAP 420 cm H2O (EPR 0-3) Bilevel (S, S/T, T): EPAP 325 cm H2O IPAP 430 cm H2O VAuto: EPAP 425 cm H2O IPAP 425 cm H2O ASV: Pressure Support 316 cm H2O EEP 415 cm H2O Dimensions 5.6 (l) x 9.8 (w) x 11.5 (h) 14.2 cm (l) x 24.8 (w) x 29.3 cm (h) Weight 8.4 lbs (3.7 kg) Filter Two-layered, powder-bonded, polyester non-woven fiber altitude Compensation Automatic Electrical Requirements 100240 V DC Power Direct connect cord
*EPR available in CPAP and AutoSet modes
EaSyCaRE TX SoFTWaRE
minimum hardware and Software Requirements PC (Mac not supported) Pentium 1 Ghz CPU 1 GB RAM 1024 x 768 display resolution 10/100 Mbps Ethernet Port Microsoft Windows 2000 SP4 or Windows XP SP2 or Windows Vista Microsoft .NET Framework 2.0 Cat 5 cable or available network port between patient room and control room
Getting Started
VPAP Tx
1
Home
2 4 6 8
3 5 7
1 2
Returns to Ready screen from anywhere in the menu except during power-up, Learn Circuit and Mask Fit. The yellow or red LED illuminates if an alarm is triggered. A message will also appear on the LCD. Performs the function indicated by text displayed above it (eg, entering sub-menus, applying settings and initiating actions). Performs function indicated by text displayed above it (eg, exiting sub-menus and cancelling actions). Increases settings options and scrolls through the menu.
Left
Right
Up
Down
Start/ Stop
Starts or stops treatment. Extended hold for at least two seconds starts the Mask Fit function. The Start/Stop key is backlit continuously. Mutes an audible, active alarm. The mute key backlight will flash when an audible alarm is triggered. This backlight will stay on while the alarm is active but muted.
Alarm Mute
The LCD is backlit whenever any key is pressed. The lighting turns off Automatically if no key is pressed for two minutes.
Mask Type
Checking and changing the mask type
The mask type should be checked when the VPAP Tx is turned on or a mask is connected to the air tubing. To ensure accurate prescriptions, ResMed mask systems are ideal for use with this therapy device.
The mask selection should be changed whenever the mask type is changed to a new group (ie, pillows, full face or nasal).
Mask Fit
Optimizing mask fit
The VPAP Tx mask-fitting feature delivers air pressure for checking and adjusting mask fit to minimize leaks.
The LCD displays a Mask Fit star rating from zero to five stars. Three to five stars indicates a good fit or better. Zero or two stars indicates that the mask needs to be adjusted.
3. Adjust the mask, cushion and headgear to achieve optimal fit. After three minutes, treatment will begin. If you do not wish to wait three minutes, hold down the Start/Stop key for three seconds and treatment will begin immediately.
10
EPR x mm.m
ee.e-ii.i mm.m
is the EPAP value for the current breath as determined by the algorithm is the IPAP value for the current breath as determined by the algorithm is the unintentional leak value
11
VPAP Tx Menus
Navigating the VPAP Tx menus
The VPAP Tx provides a set of functions which are arranged in menus and submenus. These functions are viewed on the LCD screen. Many of these functions are also accessible remotely by EasyCare Tx. The menus provided are outlined in the following table.
Setup Therapy Data Select mask and perform Learn Circuit Adjust therapy modes and settings. Refer to the EasyCare Tx Online Help for details. Access real-time measures of system leak, mask pressure, patient flow and ventilatory parameters. These figures are updated at least once every breath. The best place to view this data is via EasyCare Tx. Advanced Service Save a set of preferred default settings and restore these settings if required. Also, adjust user preferences such as language and data units. Access information about the VPAP Tx, including software version, serial number and run hours.
Starting a Session
Before you start titrating a patient, you need to start EasyCare Tx and then start a titration session.
To start a titration session:
1. Double-click the EasyCare Tx icon on the Desktop. The EasyCare Tx toolbar is displayed and the default Tx Link is automatically connected. If the Tx Link is not automatically connected, connect to a Tx Link. 2. Configure Mask and Humidifier Settings. 3. Click the Therapy Start/Stop icon. Titration begins and the therapy indicator turns green. Note: The Therapy ON/OFF indicator turns green during therapy and grey when therapy is off.
12
To connect to Tx Link:
1. From the Menu drop-down, click Connect. The Select Device window is displayed. 2. Select the required Tx Link from the Connect To drop-down list. 3. Click OK. A window indicating that EasyCare Tx is establishing a connection with Tx Link is displayed.
Bed 1 - EasyCare Tx
Connectivity issues
You may experience connectivity issues in the following circumstances:
EasyCare Tx is unable to connect to the Tx Link EasyCare Tx loses connectivity with the Tx Link Tx Link is unable to connect to the therapy device Tx Link loses connectivity with the therapy device.
In such instances, a window indicating the connectivity status is displayed and this helps you to take the appropriate action to restore connectivity. For instructions on resolving these issues, refer to troubleshooting.
13
1 2 3 4 5 6 7 8 9 10 11 12
Name of the connected Tx Link Therapy controls Therapy controls drop-down Real-time data Real-time data items drop-down Start/Stop Therapy button Therapy device connected Therapy ON/OFF indicator Humidifier ON/OFF indicator1 Menu drop-down Therapy mode drop-down Locked/unlocked status
The Humidifier icon is only displayed if the connected therapy device has a humidifier that can be remotely controlled.
14
Therapy settings
When adjusting individual parameters displayed on the toolbar, the changes are applied instantly. If a confirmation is not sent from the therapy device within two seconds, the parameter will revert the original value. Alternatively, using the therapy settings window, changes are made to one or more parameters related to a therapy and on pressing OK.
15
Titration Protocols
Monitor patient
Yes
No
Increase CPAP 1 cm H2O every 5 mins for obstructive apneas, hypopneas, RERAs and at least 3 min of loud or unambiguous snoring
Yes
Continue monitoring patient Decrease CPAP by 1 cm H2O and wait 20 mins. Consider Adapt SV if centrals persist and patient meets criteria
No
Observe patient and document final settings; be sure to document the final CPAP pressure, EPR setting (if any) and ramp time
19
CPAP Prescription:
q S9 Elite w/ Climate Control q S8 Elite ii q S8 Escape ii
q EPR: 1 2 Ramp Time: Interface: 3 min(s) (OFF - 45 min)
Sample
20
The graph illustrates the effect of EPR inhalation settings with Easy-Breathe.
21
EPR
Comfort levels
EPR provides three comfort settings. Each comfort setting correlates to an exact drop in pressure relief: Setting 1: Mild reduction (1 cm H2O) Setting 2: Medium reduction (2 cm H2O) Setting 3: Maximum reduction (3 cm H2O).
Note: The numeric value (1, 2 or 3 cm H2O) for each EPR setting represents the maximum pressure drop during CPAP therapy expiration. Therapy pressure will never drop below 4 cm H2O. So, for example, if therapy pressure during Ramp Time is 5 cm H2O and EPR is set at level 3, then the pressure will only reduce to 4 cm H2O. For this reason, EPR is a unique comfort feature that ensures therapy effectiveness at all times. EPR allows patient comfort without compromise because the selected setting offers a defined pressure drop value that never exceeds the setting value.
22
Yes
No
Observe patient and document final settings, including IPAP/EPAP pressures and TiControl settings if altered from default
23
VPAP S Prescription:
q VPaP S
q IPAP: cm H2O (4-25 cm H2O) EPAP: cm H2O (3-25 cm H2O) Ramp Time: min(s) (OFF - 45 min) Interface:
Sample
24
Ti Min
Cycle Window
Unique to the VPAP series, Vsync manages leaks automatically, minimizing any loss of pressure due to leak while still ensuring reliable breath detection.
Rise time
25
Once adjusted, triggering and cycling levels are maintained by the automatic leak compensation feature Vsync. Unique to ResMed, the Vsync algorithm automatically adjusts for leak as it occurs. This provides accurate reporting of patient flow and ensures optimum synchronization and patient support.
26
COPD Ti Max
Normal Ti Max
30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12
1.0 1.0 1.1 1.1 1.2 1.2 1.3 1.3 1.4 1.4 1.5 1.6 1.7 1.8 1.9 2.0 2.1 2.3 2.5
0.5 0.5 0.5 0.6 0.6 0.6 0.6 0.7 0.7 0.7 0.8 0.8 0.8 0.9 0.9 1.0 1.1 1.2 1.3
0.7 0.7 0.7 0.7 0.8 0.8 0.8 0.9 0.9 0.9 1.0 1.0 1.1 1.2 1.2 1.3 1.4 1.5 1.7
1.0 1.0 1.1 1.1 1.2 1.2 1.3 1.3 1.4 1.4 1.5 1.6 1.7 1.8 1.9 2.0 2.1 2.3 2.5
27
Setting Ti Min
1. Measure the respiratory rate (see instructions below). 2. Refer to the table column entitled Restrictive Ti Min. 3. Set Ti Min to the value that corresponds to the resting respiratory rate from step 1. 4. If the patient complains that the inspiratory time is too long or too short, verify the respiratory rate and Ti Min setting or adjust the Ti Min setting until the patient feels the inspiratory time is slightly too long. 5. Reduce the Ti Min setting by 0.2 to 0.3 seconds from the setting in the previous step.
Setting Ti Max
1. Measure the respiratory rate (see instructions below). 2. Refer to the table column entitled Restrictive Ti Max. 3. Set Ti Max to the value that corresponds to the resting respiratory rate from step 1. 4. If the patient complains that inspiratory time is too short, verify the respiratory rate and Ti Max setting.
Setting Ti Max
1. Measure the respiratory rate (see instructions below). 2. Refer to the table column entitled COPD Ti Max. 3. Set Ti Max to the value that corresponds to the resting respiratory rate from step 1. 4. If the patient complains that the inspiratory time is too short, verify the respiratory rate and Ti Max setting.
Setting Ti Min
Most patients with OSA and COPD do not have problems with premature cycling, therefore the Ti Min setting can remain at the default setting of 0.3 seconds.
Setting Ti Max
1. Measure the respiratory rate (see instructions below). 2. Refer to the table column entitled Normal Ti Max. 3. Set Ti Max to the value that corresponds to the resting respiratory rate from step 1. 4. If the patient complains that the inspiratory time is too short, verify the respiratory rate and Ti Max setting.
Setting Ti Min
Most patients with OSA do not have problems with premature cycling, therefore the Ti Min setting can remain at the default setting of 0.3 seconds.
Measuring Respiratory Rate: Measure the patients resting respiratory rate during assisted ventilation. If you think the patients respiratory rate will decrease during sleep and you are unable to measure it, then estimate the nocturnal respiratory rate and use this figure to determine the Ti Max and Ti Min setting. Refer to the section above that most appropriately matches the patients condition. In some instances, recommended settings may need to be fine-tuned based on patient feedback and/or ongoing clinical assessment.
note: If a high leak condition is noted (leak 24 L/min) the nasal mask fitting should be adjusted, and/or a Mirage Quattro Full Face Mask used in the event of a mouth leak.
28
Observe patient and document final pressure settings (IPAP/EPAP pressures, respiratory rate and TiControl settings) if altered from default
29
VPAP ST Prescription:
q VPaP ST
q Spont Mode q Spont/Timed Mode IPAP: cm H2O (4-25 cm H2O) for VPAP ST EPAP: cm H2O (3-25 cm H2O) Rate: BPM (5-30 BPM) Ramp Time: min(s) (OFF - 45 min) Interface:
Sample
30
Start therapy at default settings and observe patient for 20-40 minutes Throughout titration adjust mask as needed to maintain leak less than 24 L/min
No
Yes
No
Yes
Increase EEP by 1 cm H2O every 20 minutes until obstructive events are eliminated
31
Note:
Before making changes, always verify leak is below 24 L/min. After changing a setting, always wait 20 minutes before making another change.
If patient exhibits:
Obstructive events Residual events: Snoring Respiratory arousals Desaturations Flow limitation
ICD-9
327.21 327.22 327.26 327.27 Primary CSA* High altitude periodic breathing Sleep-related hypoventilation/hypoxemia in conditions classifiable elsewhere (requires underlying diagnosis code) CSA in conditions classified elsewhere (requires underlying diagnosis code) 786.04 CSA due to CheyneStokes respiration (CSR)
*Most commonly used for complex sleep apnea patients. These are examples of diagnoses that may be associated with the above mentioned technology. Physicians must determine the appropriate ICD-9 diagnosis based on individual patient needs during the initial exam or through a history and physical.
After hours support available MondayFriday 6pm3am (PT) by calling 1-888-AUTOSET (1-888-288-6738)
32
Airflow
APNEA
VPAP Adapt SV
VPAP Adapt SV
VPAP Adapt SV and adaptive servo-ventilation With its advanced adaptive servo-ventilation algorithm, the ResMed VPAP Adapt SV provides ventilatory support to rapidly treat all forms of CSA, mixed apnea and periodic breathing, commonly known as CheyneStokes respiration (CSR). The adaptive servo-ventilation algorithm delivers customized therapy Ventilation to a moving target To determine the degree of pressure support needed, the adaptive servo-ventilation algorithm continuously calculates a target ventilation. Based on respiratory rate and tidal volume, the target is 90% of the patients recent average ventilationthat means that ventilation can vary gradually and naturally over the course of the night. Patientmachine synchronization The algorithm uses three factors to achieve synchronization between pressure support and the patients breathing. 1. The patients own recent average respiratory rate, including the ratio of inspiration to expiration and the length of any expiratory pause. 2. The instantaneous direction, magnitude and rate of change of the patients airflow, which are measured at a series of set points during each breath. 3. A backup respiratory rate of 15 breaths per minute.
Late mid-inspiration Early end-inspiration
Airflow
Time
To ensure ventilatory support is synchronized to the patients effort, the VPAP Adapt SV relies on factors 1 and 2. When a central apnea or hypopnea occurs, support initially continues to reflect the patients recent breathing pattern. However, as the apnea/ hypopnea persists, the device increasingly uses the backup respiratory rate. Minimal support during stable breathing The adaptive servo-ventilation algorithm starts working even before CSA events occur, from the moment the patient lies down, puts on the mask and switches on the VPAP Adapt SV. As long as ventilation is at or above the target: The magnitude of the pressure support remains minimal, the default minimum value of 3 cm H2O provides enough support to do about half the work of breathing.
33
The underlying end expiration pressure (EEP) can be adjusted by the clinician from 4 to 15 cm H2O, which helps reduce obstructive events and can also reduce central events. Support when its needed When a central hypopnea or apnea occurs and ventilation suddenly drops below the target: Pressure support rapidly increases over a few breaths to keep ventilation at the target. The default maximum pressure support value of 15 cm H2O will fully ventilate a centrally apneic patient with an open airway and normal lungs. Adaptive support when breathing resumes As breathing resumes and total ventilation exceeds the target: the minimum 3 cm H2O.
Why adaptive servo-ventilation instead of bilevel? Most CSA/CSR patients have trouble tolerating conventional bilevel ventilatory support. The pressure must be adjusted to a constant high level to adequately support a patient during apnea/hypopnea. This leads to over-ventilation during periods of normal breathing or hyperpnea, which causes arousals and discomfort. It can even propagate more CSA events. The adaptive servo-ventilation algorithm stabilizes patient breathing By ventilating the patient appropriately during periods of hypopnea and apnea and reducing support during periods of hyperventilation and normal breathing, the advanced adaptive servo-ventilation algorithm rapidly stabilizes breathing patterns and arterial blood gases and minimizes the discomfort and arousals often associated with bilevel treatment.
34
35
Reimbursement
MD sleep specialist
ICD-9 Codes
327.21 Primary CSA* 327 .22 CSA due to high altitude periodic breathing 327 .26 Sleep-related hypoventilation/hypoxemia in conditions classifiable elsewhere (requires underlying diagnosis code) 327 .27 CSA in conditions classified elsewhere (requires underlying diagnosis code) 786.04 CSA due to CSR
* Most commonly used for complex sleep apnea patients. These are examples of diagnoses that may be associated with the above mentioned technology. Physicians must determine the appropriate ICD-9 diagnosis based on individual patient needs during the initial exam or through a history and physical.
CPT Codes
95810 PSG; sleep staging with four or more additional parameters of sleep, attended by a technologist 95811 PSG; sleep staging with four or more additional parameters of sleep, with initiation of continuous positive airway pressure therapy or bilevel ventilation, attended by a technologist (94770 carbon dioxide, expired gas determination by infrared analyzer)
Definitions
Respiratory Insufficiency Impairment in respiratory function severe enough to prohibit certain activities that the patient might normally pursue, and to interfere with daily living; occurring in association with measurements of respiratory mechanics and/or gas exchange that are markedly abnormal. Complex sleep apnea (CompSA) is a form of CSA specifically identified by the persistence or emergence of central apneas or hypopneas upon exposure to CPAP or an E0470 device when obstructive events have disappeared. These patients have predominantly obstructive or mixed apneas during the diagnostic sleep study occurring at greater than or equal to five times per hour. With use of a CPAP or E0470, they show a pattern of apneas and hypopneas that meets the definition of CSA. CSA is defined as: (1) An apneahypopnea index greater than five; and (2) Central apneas/hypopneas greater than 50% of the total apneas/hypopneas; and (3) Central apneas or hypopneas greater than or equal to five times per hour; and (4) Symptoms of either excessive sleepiness or disrupted sleep.1
References 1 Centers for Medicare & Medicaid Services, LCD for Respiratory Assist Devices (L11493) U.S. Department of Health and Human Services, http://www.cms.hhs.gov/mcd/viewlcd. asp?lcd_id=11493&lcd_version=22&show=all (revision effective date 4/01/2006) Current Procedural Terminology 2005 American Medical Association. All Rights Reserved. Current Procedural Terminology (CPT) is copyright 2005 American Medical Association. All Rights Reserved. No fee schedules, basic units, relative values, relative values, or related listings are included in CPT. The AMA assumes no liability for the data contained herein. Applicable FARS/DFARS restrictions apply to government use.
39
If No, a new initial face-to-face clinical evaluation is required but not a new sleep test. A new 3-month trial would begin for use of the bilevel. If Yes, the patient is qualified for an E0470 device (bilevel without a backup rate, such as VPAP Auto 25)
ahi is defined as the average number of episodes of apnea and hypopnea per hour of sleep.
40
ii. CoPD
ABGs (done while awake) PaCO2 52 mm Hg (patients prescribed FiO2)
ABGs (done while awake) PaCO2 45 mm Hg (patients prescribed FiO2) or Sleep oximetry Oxygen saturation 88% for 5 minutes, minimum 2 hours recording time (patients prescribed FiO2) or For neuromuscular disease only, either FVC < 50% of predicted or MIP < 60 cm H2O
Sleep oximetry Oxygen saturation 88% for 5 minutes, minimum 2 hours recording time (on 2 L/min O2 or patients prescribed FiO2, whichever is higher)
OSA and CPAP treatment has been considered and ruled out
(E0470)
For CoPD patients to qualify for a RaD with backup rate (E0471):
Situation 1 After period of initial use of an E0470, ABG (done while awake) shows PaCO2 worsens 7 mm Hg compared to original ABG result (on patients prescribed FiO2); PSG demonstrates oxygen saturation 88% for 5 minutes, minimum 2 hours recording time, on an E0470, not caused by obstructive upper airway events (ie, AHI < 5). Situation 2 No sooner than 61 days after initial use of E0470, ABG (done while awake) shows PaCO2 52 mm Hg (on patients prescribed FiO2); Sleep oximetry on an E0470 demonstrates oxygen saturation 88% for 5 minutes, minimum 2 hours recording time (on 2 L/min O2 or patients prescribed FiO2, whichever is higher).
Respiratory assist device (RAD) documentation requirements for continued coverage Patients on an E0470 or E0471 device must be re-evaluated no sooner than 61 days after initiating therapy. Required documentation Progress of relevant symptoms Signed and dated statement by treating physician declaring patient using average 4 hours per 24-hour period and patient benefiting from use
41
iV. hypoventilation
ABGs (done while Covered by E0470 being used
awake) PaCO2 45 mm Hg
(patients prescribed FiO2)
Settings that will be prescribed for initial use at home Patients prescribed FiO2
ABGs
(done during sleep or immediately upon awakening)
ABGs
(done while awake)
PaCO2 worsens 7 mm Hg compared to ABG result used to qualify for E0470 (patients prescribed FiO2) or
PSG demonstrates
oxygen saturation 88% for 5 minutes, minimum 2 hours recording time not caused by obstructive upper airway events (ie, AHI < 5)
oxygen saturation 88% for 5 minutes, minimum 2 hours recording time, on E0470, not caused by obstructive upper airway events (ie, AHI < 5)
(E0470)
(E0471)
hypopneas upon exposure to CPAP or an E0470 device when obstructive events have disappeared
during the diagnostic sleep study occurring at 5 times per hour and hypopneas that meets the definition of CSA
This information is provided as of the date listed, and all coding and reimbursement information is subject to change without notice. It is the providers responsibility to verify coding and coverage with payors directly. For a full description of the policy go to www.cms.hhs.gov. ResMed reimbursement hotline, dial 1-800-424-0737 and select option 4.
42
ResMed Corp San Diego, CA, USA +1 858 836 5000 or 1 800 424 0737 (toll free). ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). See www.resmed.com for other ResMed locations worldwide. TiControl, VPAP Adapt SV and Vsync are trademarks of ResMed Ltd. AutoSet, EPR, HumidAire, Mirage, Mirage Quattro, S9 and VPAP are trademarks of ResMed Ltd and are registered in the U.S. Patent and Trademark Office. 2010 ResMed. 1013904/1 10 06
www.resmed.com