Service Manual
COPYRIGHT
Beijing Choice Electronic Technology Co., Ltd. (hereinafter called Beijing Choice) owns
all right to this unpublished work and intends to maintain this work as confidential. Beijing
Choice may also seek to maintain this work as an unpublished copyright. This publication is
to be used solely for the purpose of reference, operation, maintenance, or repair of Choice
equipment. No part of this can be disseminated for other purposes.
In the event of inadvertent or deliberate publication, Beijing Choice intends to enforce its
right to this work under copyright laws as a published work. Those having access to this work
may not copy, use, or disclose the information in this work unless expressly authorized by
Beijing Choice to do so.
All information contained in this publication is believed to be correct. Choice shall not be
liable for errors contained herein nor for incidental or consequential damages in connection
with the furnishing, performance, or use of this material. This publication may refer to
information and protected by copyrights or patents and does not convey any license under
the patent rights of Beijing Choice, nor the rights of others. Beijing Choice does not assume
any liability arising out of any infringements of patents or other rights of third parties.
CONTENT
Chapter 1 Introduction 5
General 5
Description of abbreviation 5
Explanation of symbols 5
Special feature 6
Appearance of monitor 7
Specification 13
Chapter 2 Special operation in menu 17
Chapter 3 Monitor function and principle 19
CSN801 board 19
Component 21
Structure 21
Chapter 5 Troubleshooting 30
Diagram of solutions to various troubles 30
Troubleshooting 40
Error Message 43
Alarm test 45
ECG test 46
Respiration test 47
Temperature test 47
MMED60000DP (S6)
Service Manual
NIBP test 47
SPO2 test 49
ECG adjustment 50
NIBP adjustment 50
SpO2 adjustment 50
IBP Calibration 52
ET-CO2 calibration 54
Monitor Disassembly 55
Battery Removal 55
Maintenance check 57
General cleaning 57
Sterilization 58
Chapter 10 Warranty 61
Customer guarantee 62
Repackaging 62
MMED60000DP (S6)
Service Manual
MMED60000DP (S6)
Service Manual
CHAPTER 1 INTRODUCTION
1.1 General
MMED6000DP is a Portable Patient Monitor that has abundant monitoring functions and is
used for the clinical monitoring of adult, pediatric and neonate. Besides, the user may select
the different parameter configuration according to different requirements.
MMED6000DP can be connected to the central working station to form a network monitoring
system.
MMED6000DP can monitor vital signals as ECG, RR, SpO2, PR, NIBP, TEMP, Dual-channel
IBP (optional), CO2 (optional). It integrates parameter measuring modules, display and
recorder in one device, featuring in compactness, lightweight and portability. Large true color
display provides clear view of 6 waveforms and full monitoring parameters.
BF type equipment
Rotated knob
Power on/off Counter-clockwise
and clockwise
Sound on Silence
Equipotential grounding AC
terminal 50/60 HZ
Fuse
This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The
unit displaying this symbol contains an F-type isolated (floating) patient applied
part providing a high degree of degree of protection against shock, and is suitable
for use during defibrillation.
Fig.1-1
The display of MMED6000DP patient monitor is TFT LCD, which displays the patient
parameters, waveforms, historical data and monitor status.
The screen is divided into five areas: (1) monitor status, (2) waveforms, (3) the table of
historical data, (4) parameters (5) menu as well as (6) frame of menu status.
⑴ ⑵ ⑶ ⑷
MMED60000DP (S6)
Service Manual
⑹ ⑿
⑺
⒀
⑻
⒁
⒆ ⒅ ⒄ ⒃ ⒂
⒇
Fig. 1-2
Monitoring status frame
⑴ Bed No.: indicates the bed number of the patient being monitored when the patient
monitor is connected with the workstation.
⑵ DEMO: indicates the device is in DEMO monitoring mode now. In clinical application,
this function is not recommended because the DEMO will mislead the hospital
workers to treat the waveform and parameter as actual data of the patient, which may
result in delay of treatment or mistreatment.
⑶ Start time: indicates the time powered on the monitor, it shows in “16:25”.
⑷ Current time: indicates the current date and time, it shows in “05/16/2004, 10:10”
Menu
The menu always occupies the fixed position on the bottom of the screen, user can
set the system and perform operation.
⒅ Trend: the function is for operator observing the patient’s latest 72 hours change,
including historical data table, HR, NIBP, SpO2, ST segment.
⒆ System setup: you can configure various aspects of the monitor, including system
time, simulation, print setup, color, display wave, language, color, etc..
⒇ Menu bar: it shows the different menus of every operation.
Other modes:
IBP EXT:
MMED60000DP (S6)
Service Manual
GRAPH EXT:
ECG EXT:
(1) Alarm
ALARM button, pressing the button will prohibit all the technically audio alarm and
physically audio alarm for 3 minutes. The audio alarm function will be restored
automatically after 3 minutes or be activated when new alarm occurs. As pressing the
button, the icon in menu status shows , which indicates all the audio alarms
NOTE
When “ ” mark appears, the system can not give out the audio alarm
(2)Silence
SILENCE button, when pressing the button, this mark appears in the
menu status, indicating that all kinds of sound including the audio alarm and heart beep
have been manually muted until the button has been pressed again, the system will
immediately restore the normal status, and the icon shows as
NOTE
When pressing “SILENCE”, the system cannot give out the audio alarm and
heart beep; therefore, the operator should be considerate in using the
function.
(3)Freeze
FREEZE button, pressing the button can freeze the waveform of ECG. The screen
displays two traces of ECG waveform. When one trace is active, pressing the FREEZE
button will freeze another trace. Pressing “FREEZE” button again will restore the
normal monitor status.
(4)Start/Cancel
MMED60000DP (S6)
Service Manual
START/CANCEL BP measuring button, pressing the button will inflate the cuff to start a
new NIBP measurement. When measuring, press it to cancel the measurement and
deflate the cuff, including the “CYCLE” and “STAT (NOT STOP)” measuring mode.
(5)Print
PRINT button, pressing this button will motivate the recorder or desktop printer to
output the result if the monitor is equipped with them.
(6) Menu
Main MENU button, pressing the button will exit the submenu and refresh the screen.
(7)Rotary Knob
The operator uses the rotary knob to select the menu
item and modify the setup. It can be rotated clockwise
or counter-clockwise and pressed like other buttons.
The operator uses the knob to realize the operations.
The method of using the knob to execute the operation: Fig. 1-4
Rotary knob is just like the cursor of computer. When operator rotates the knob on
the icon where the operation is wanted, the icon will be automatic highlight. Then
pressing the knob, operator will open the setup menu of the corresponding
parameter so as to set up the menu.
(8) Power Supply Indicator
Indicating the power supplying state of the monitor
Orange: 220Vac main power supplying
Orange and flash: 220Vac main power supplying and battery charging
Green and flash: internal battery power supplying and 50% power has been
consumed.
(3)
(4)
(7)
(5) (6)
Fig.1-5
(1) ECG socket:connect ECG 5-lead wire with 6 PIN connector
MMED60000DP (S6)
Service Manual
(8)
Fig.1-6 Fig.1-7
1.5.4 Recorder
Fig. 1-8
A thermal dot matrices recorder with standard 50mm (+1/-1) wide printout paper is used for
MMED6000DP portable patient monitor.
NOTE
The thermal array recorder should be installed by the adequate technician.
(1)
(2)
(3)
(5)
(6)
(4)
Fig.1-9
(1) Power switch
(2) NET: Internet access point: connecting with the central monitoring system through
the standard RJ-45.
(3) CRT: exteriorly connecting screen and standard VGA.
(4) Printer: exteriorly connecting printer, which type is designated by the manufacturer.
(5) AC Power supply: 100-230 (VAC), 50/60(Hz); Fuse: standard –T 1.6A
(6) Equivalent electric ground access for connecting with the hospital’s grounding system.
1.6 Specification
Environment
Display
Displayed Parameters
Pulse Rate
TEMP
Technique Resistance
Channel 2 (T1 and T2)
Scales F. Or ℃
Probes Resistive; recta and skin (reusable and disposable)
YSI 400 Series types
Range 0-50℃
Revolution 0.1℃
Accuracy ±0.1℃
Alarm delay Body temperature high and low limits alarm delay<7s
CO2 (Optional)
Technique
Side Stream, non-dispersive infrared (NDIR) absorption technique. Including multiple
water trapped/filtration system and microprocessor logic control of sample handling and
calibration.
IBP
MMED60000DP (S6)
Service Manual
Power Requirements
Some operations of monitor aim at the dealer but not the doctor, such as “Language setup”
or “Hz selection”, the dealer should note the following points before monitor getting end user.
NOTE: you must keep the correct setup according customer’s condition before using
the equipment.
In monitoring status, rotate the knob to highlight the MENU icon, and then press the
knob, the menu bar of “system setup” appears on the bottom of screen.
Fig.2-1
Rotate the knob and select “EXTEND” item to access the sub-menu shown as below Fig.1-2
Fig.2-2
The functions of the items in the submenu please refer to section 6.5 of chapter 6 system
menu in Operator’s manual.
Select “SYS-ACC” item to access the sub-menu shown as Fig.2-3,
Fig.2-3
Then Input “2” in each of the items, press PW-ENTER to enter another sub-menu shown as
Fig.2-5:
Fig.2-4
Fig.2-5
Rotate the knob to select “SETUP” menu to enter in the sub-menu shown as Fig.2-6
Fig.2-6
Language Setup
The system offers two languages: English and Chinese. Technician is able to set the
required menu as per customer’s demand.
Screen Selection
MMED60000DP (S6)
Service Manual
If the screen of monitor is 12.1 inch, you should select “800*600” for the item, otherwise
please select “640*480” for 10.4 inch.
ACCHOICE
It is the AC 50HZ or 60HZ filter.
CHAPTER 3 MONITOR FUNCTION and PRINCIPLE
Power supply unit consist of AC-DC adapter, battery and Power board. The AC-
DC adapter converts 220VAC into 15VDC, and the Power board provides 5VDC
to mainboard and CSN801 board, 12VDC to TFT displayer and fan, power
Supply Indication signal to LED, charge battery port to battery, and speaker
signal to speaker.
3.2.1 Power
Board input voltage is 5-6v, through the DC-DC, the voltage is changed into
±9v, though manostat voltage is changed into ±5v and +3.3v.
3.2.2 CPU
The board includes ECG part, SpO2 part, respiration part, temperature part,
NIBP part, and there are 3 CPU in the board altogether. ATEMEGA128L (clock
8MHz) is used as major CPU. 128K Bytes of In-System programmable Flash, it
can endure 1,000 Write/Erase Cycles.4K Bytes EEPROM, and endure 100,000
Write/Erase Cycles. It is responsible for the whole control of the board. Using
the timer interrupt, CPU is detecting the R wave and ST segment from ECG
waveform input by ADC, calculating heart rate, the respiration rate, and
receiving the data from NIBP part, temperature part and SpO2 part, and
transferring the data to the host computer.
Z80 is another CPU, and responsible for NIBP measurement .The third CPU is
8051F007; it is responsible for SpO2 measuring.
There are two serial RS232 I/O ports, one for transferring data to the host
computer, and another is for receiving the NIBP data from the NIBP part.
Signal through the ECG lead wires is inputted into three multiplexer. One
multiplexer can selects the I, II, III, avR, avL, avF, V1, V2, V3, V4, V5, V6 lead
and CAL(1mV scale voltage), the second multiplexer can select filter mode
including monitor mode, diagnosis mode, operation mode, the software filter
mode, and the third select feedback and driver. After through the low pass filter
and high pass filter, the signal made 1000 times as large by the amplifier, and
then output to the A/D Conversion of CPU.
There are two ways to detect the leadoff. One is from the input circuits, and the
other is from the feedback circuits. When one of two conditions occurs, the
leadoff is detected.
3.2.4.3 Filter
Hardware filter consists of the low pass filter and the high pass filter. The
bandwidths are different according the various filter modes. In Diagnosis mode
it is 0.05-100Hz, in Monitor mode is 0.5-75Hz, in Operation mode is 1-25Hz.
After calculating and processing from the CPU, the ECG data are transferred to
the host board.
3.2.5 NIBP
Blood pressure measurement module can be used for monitoring the blood
pressure from neonates to adults. Measurement adopts the oscillimetric
method, CPU controls the pump, valve to inflate and deflate the cuff, and the
sensor transforms the pressure signals into the electrical signals. After that, the
signals are amplified by the AD623, and sampled by 1293AD. At the same time,
1293A/D converter sampled the pluses wave. Based on the pluses wave, the
systolic, diastolic and mean pressures are confirmed. According to the
difference of Neonate mode and Adults mode, overpressure protection circuit
resets the pressure value of blood pressure module to ensure the patients’
safety.
3.2.6 SpO2
3.2.7 TEMP
The resistances is sampled by ADS1100, and transferred into the CPU, through
matching the resistance values, the temperature values are calculated.
3.2.8 RESP
⑵ Option:
No. Production Description Standard Quantity
1 IBP module 1 piece
2 Et-CO2 module (side-stream) 1 piece
3 Thermal array recorder 1 piece
4.2 Structure
4.2.1 Front panel
MMED60000DP (S6)
Service Manual
3 1
1
7
2
6 5 4
5 4 3 2 1
Fig.4-9 Profile
1. CSN801 board
2. Pump and valve fix board
3. Power board
4. Mainboard
5. SSD
Fig.4-11
MMED60000DP (S6) Service Manual
TXD
DATA
SSD P2 ETCO2 module
Pc104 COM1
RXD
DOC COM4
DATA Recorder
32
MMED60000DP (S6) Service Manual
SSD
AC/DC DC 5V DC 5V
Power Supply Main board
DC 5V DC 12V Inverter
Key
DC 12V Board
LED
&Volume
Signal
LED
Battery
charge
Battery circuit
DC 12VFan
33
MMED60000DP (S6) Service Manual
34
MMED60000DP (S6) Service Manual
CHAPTER 5 TROUBLESHOOTING
5.1 Diagram of solutions to various troubles
A.
NO DISPLAY WITH AC MAIN POWER
ON
NO
NO
Is Power Supply
Indicator (LED)
Is fuse blown? lighted? A
NO 12
YES
YES
Replace fuse NO NO
Is TFT
screen Replace the
Are there any backlig inverter.
loose
YES YES
35
MMED60000DP (S6) Service Manual
YES
A
12
NO
Does AC/DC Change
power supply AC/DC POWER SUPPLY
output 15V?
YES
A
2
Change
the power board
36
MMED60000DP (S6) Service Manual
NO NO
Can you hear the Is mainboard Change
voice of mainboard supply voltage the power board
power-on self- OK?
YES
YES
B.
NO BLOOD PRESSURE
READING
NO
Is connector Reconnect cuff/hose
connected connectors.
properly?
YES
NO NO
37
MMED60000DP (S6) Service Manual
NO
B Change the CSN801
Is there air leak? board
1
YES
B
2
B
1
NO
NO
Is pressure inflated Change the CSN801
Is there air leak? board
to pre-set valve?
YES
YES
B
NO 2
Is deflation
speed OK?
YES
38
B
2
MMED60000DP (S6) NO
Service Manual
Are CUFF and Change CUFF and
AIR HOSE OK? AIR HOSE
YES
Change the CSN801
board
NO
Is 5V supplied to Change the
the valves? CSN801 board
YES
B
3
NO NO
Change the CSN801
Does PUMP Is 5V supplied board
work? to the pump?
YES
END
Change the pump
39
MMED60000DP (S6) Service Manual
40
MMED60000DP
(S6) Service Manual
C.
NO ECG MEASUREMENT
NO
Is there “ECG
ERROR” alarm?
YES
NO
Change
Is 5V supplied Power board
voltage is OK?
YES
Change
CSN801 board
NO
Change ECG cable and
Are ECG cable Lead wire
and lead wires
YES
NO
Replace the electrodes
Are electrodes refer to operation manual
placed OK?
YES
D.
NO TEMPERATURE
READING
NO
Is temperature Replace PROBE
PROBE OK?
YES
NO
Reconnect connector
Is connection
OK?
YES
NO
Check power source
Is supply voltage
OK?
YES
Change
CSN801 board
MMED60000DP
(S6) Service Manual
E.
NO RESPIRATION
READING
NO
Are ECG cable Replace PROBE
and lead wires
YES
NO
Replace electrodes
Are electrodes
be placed OK?
YES
NO
Is respiration Select the respiration
type same as the type correctly
internal setup?
YES
Change
CSN801 board
MMED60000DP
(S6) Service Manual
F.
NO SPO2 READING
NO
Is SPO2 sensor Replace SPO2 sensor
OK?
YES
NO
Reconnect connector
Is connection
OK?
YES
NO
Is supply voltage Change power source
OK?
YES
Change
CSN801 board
MMED60000DP
(S6) Service Manual
5.2 Troubleshooting
5.2.1 The main trouble of ECG/HR
Failure Possible cause Solution
Display shows “ECG CSN801 board has Replace the CSN801 board
ERROR” when you turn on problem.
the equipment
No ECG waveform 1 Poor connection of ECG 1 Use new electrode to
electrode films ensure good contact.
suspended
damaged.
4 Note the precondition
4 The patient’s skin is
of skin.
too dry.
4 CSN801 board is
4 Replace it.
damaged.
Only one or some leads 1 Some electrodes are Connect the lead with ECG
could display waves, some not contacted well. simulator, if the signal is
leads display straight lines. normal, the trouble lies in
the electrodes or leads
2 The lead wires are
(identify which electrode
should be changed) if the
damaged.
signal is abnormal still,
The CSN801 board change the CSN801 board.
3
has problem.
ECG waveform displays 1 Patient is moving 1 Keep patient quiet.
disorderly constantly.
monitor
grounded receptacle
HR data fluctuate Patient is moving Keep patient quiet.
remarkably constantly.
ECG base line is not Power supply is inferior Do not use the same power
smooth panel with high power
equipment.
constantly.
3 Replace it.
3 CSN801 board is
damaged.
The data change 1 The dryness of skin 1 Note the precondition of
remarkably and have skin
deviation from the real data.
2 Inferior quality of 2 Replace the new
electrode electrodes.
probe.
Not enter the monitor status Password is locked Pressing “START” twice,
“PRINT” once release lock.
Start the monitor, the Mainboard’s BIOS” is rewrite “BIOS” programme
monitor can check damaged or replace the main board
automatically, but it still
stays “starting Rom-Dos”
User Limited. Enter 1,the expire date is earlier 1,change the setting
password than the current date
2,change the setting
2, the runtime is over the 3,replace it
usetime
3. the mainboard is bad
Mercury Manometer
wave at 10Hz)
For internal battery Operation, confirm that the LED Power Indicator is
illuminated Green.
Confirm that the Screen is displayed without any error messages when
1 Rotate the knob in the clockwise direction while verifying that the highlight moves
2 Rotate the knob in the counter-clockwise direction while verifying that the highlight
3 Highlight any parameter icon, rotate the knob clockwise in the parameter area,
4 Highlight any parameter icon, rotate the knob counter-clockwise in the parameter
5 Highlight the RETURN menu item, press the knob and verify that the menu is return
1 Press ALARM key, observe that the alarm icon in frame of menu status will
change.
2 Press SILENCE key, observe the silence icon in frame of menu status will change.
3 Press FREEZE key, one channel of the ECG waveform should be frozen,
4 Press START key will start NIBP measurement, and press the key again will cancel
this measurement.
5 Press PRINT key, it can print the relevant waveform and measured data if the
printer is installed.
Set the simulation item ON, the detailed operation please refer to the Section
Select “ON” for SOUND in ECG menu, the detailed operation please refer to
① Connect the ECG Patient Simulator to the ECG input on the monitor side panel.
② Set the Patient Simulator to output a 1mV QRS pulse at 60bpm, set the lead of
ECG submenu as II
③ Set the heart rate high limit as 58bpm, low limit as 30bpm.
⑤ Verify the heart rate visual and auditory alarms, the heart rate data should flash
and “dudu” voice should be heard.
① Connect the NIBP Patient Simulator to the NIBP input on the monitor side panel.
⑤ Verify the systolic visual and auditory alarms, the systolic data should flash and
“dudu” voice should be heard.
① Connect SPO2 Probe to the SPO2 connector on the monitor side panel.
② Insert the operator’s finger into the finger sensor, the SPO2 measured value of
healthy person should be 96%.
⑤ Verify the SPO2 visual and auditory alarms, the SPO2 data should flash and
“dudu” voice should be heard.
② Set the patient simulator to output a 1mV , 10Hz sine wave, set the wave of
ECG submenu as 10mm/Mv, the ECG waveform should show the sine wave
that the height is 10mm.
③ Set the patient simulator to output a 1mV QRS pulse at 60bpm, set the lead of
ECG submenu as II.
④ Verify the ECG waveform moving QRS pulses without jittering, the number
indicate an ECG count of 60bpm.
⑤ Set the patient simulator to output a 1mV QRS pulse at 120bpm, verify the ECG
MMED60000DP
(S6) Service Manual
waveform moving QRS pulses without jittering and the number indicate an
ECG count of 120bpm.
⑥ Set the patient simulator to output a 1mV QRS pulse at 240bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an
ECG count of 240bpm.
⑦ Set the patient simulator to output a 0.5mV QRS pulse at 30bpm, verify the
ECG waveform moving QRS pulses without jittering and the number indicate
an ECG count of 30bpm.
⑧ Set the patient simulator to output a 2mV QRS pulse at 240bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an
ECG count of 240bpm.
6.7.1 ECG LEAD OFF test
Perform the following procedure to verify if the function of ECG Lead Off works
orderly.
① Attach the lead to the ECG patient simulator.
② Set the patient Simulator to output a 1.0mV QRS pulse at 60BPM.
④ Remove one lead from simulator in turn (that pertains to the selected lead) until
all the leads have been tested. In this situation, verify that “LEADOFF”
message appears in ECG displaying frame of parameter area in 12S.
⑤ Disconnect the Patient Simulator from the monitor front panel, the “LEADOFF”
message appears.
6.7.2 ST segment test
Perform the following procedure to verify if ST segment program works orderly.
1) 30℃±0.1℃
2) 35±0.1℃
3) 40±0.℃
80 50 ±5mmHg
120 80 ±5mmHg
150 100 ±10mmHg
200 150 ±10mmHg
NOTE: the uncertainty specifications of the NIBP Simulator must be added to
the monitor tolerances for proper accuracy verification.
as following figure.
4) Inflating the cuff manually by pressing the rubber ballonet. With the
NOTE
The MMED6000DP Monitor contains a great number of Static Sensitive circuits.
All service procedures must be done by grounded personnel.
ECG gain is fixed. User cannot adjust it. If ECG gain dose not meet the standard
demand. Please contact the service person of Choice Co.
Fig. 7-1
VR1
Fig. 7-2
Rotate the knob to call up “ALARM” submenu of NIBP, set the high limit and low limit
to maximum. For example, if you want to adjust systolic, set the high limit of systolic
to 255, low limit to 220, then press RETURN to return the superior menu, next press
RETURN of the superior menu again back the main menu, at that time, you will find
NIBP menu has changed as Fig.7-3:
Fig.7-3
SYSQUTY: quotiety of diastolic and systolic value
SYSOFFS: offset of diastolic and systolic value
Fig.7-4
② Then press ALARM icon to enter ALARM submenu, set the high limit to
100(maximum), and low limit to 99(maximum).
Fig.7-5
③ Press RETURN to enter the SpO2 main menu, which has changed as Fig.7-6,
“OFFSET” item displays on the screen. You can adjust the value in this item. For
Example, if the SpO2 value is 3 lower than standard, you can adjust OFFSET
value as “3” (Figure 4); If the value is higher than standard, you can adjust value
MMED60000DP
(S6) Service Manual
as “–2”.
Fig.7-6
WARNING :
Close the one way of the transducer’s three-way connected to the patient before
zeroing.
Put the transducer at the same level with the patient’s cardio, namely
The zeroing must be done before the monitor used to monitor the patients
and at least one time everyday. (The zeroing must be done after inserting or
taking out the power supply wire.)
1. Connect the transducer with cable, install DOME. The both ends of DOME should be
connected with air (e.g.: IBP1)
2. Set the pressure revising value to zero.
3. The pressure value will be shown on the IBP channel1, S, D value also will be shown,
and please take the shown S value as criterion
4. The S value should be 0, 1, 2, or 3. The tolerance is 0.4 kPa (3mmHg).
5.If the error exceeds the tolerance, please enter into IBP1 pressure revising item to
adjust IBP1 revising value and let S value be zero.
Fig.7-7
The purpose of the calibration is to insure that the system can give accurate
measure result.
You must zero the transducer before the mercury manometer calibration.
MMED60000DP
(S6) Service Manual
should be:
Standard sphygmomanometer
3-way switch
About 25 centimeter pipeline
Calibration steps
Fig.7.9
When presses “ZERO”, the menu shows as follow:
Fig.7.10
6. Then press the “ADJUST” item, the pressure value will be shown on the IBP channel
1, S, D value also will be shown. And take the S value as criterion.
MMED60000DP
(S6) Service Manual
Fig.7-11
7.5.2 Then press ALARM icon to enter ALARM submenu, set the high limit to 75
(maximum), and low limit to 70(maximum), next please press RETURN to back
the former menu.
Fig.7-12
7.5.3 Enter into the CO2SETUP again, OFFSET item is appeared on the menu, then
you can adjust the offset from –20 to +20 to calibrate the CO2 value.
Fig.7-13
MMED60000DP
(S6) Service Manual
MMED60000DP
(S6) Service Manual
Screw 1
Screw 2
Fig.8-1
③ After unscrewing 3 screws, the front panel and rear panel are separated.
Battery
AC-DC
adapter
Fig.8-2
2 Unscrew the 6 screws on the metal shell; the battery and adapter are
disassembled together from inner. Draw the battery out the shell.
3 Unplug the battery cable from the power board, the battery can be removed.
Fig.8-3
MMED60000DP
(S6) Service Manual
NOTE
Before cleaning the monitor or the sensor, make sure to turn off the power and
disconnect the AC power.
3. Check all the functions of the monitor to make sure that the monitor is in good
condition.
If finding any damage on the monitor, stop using the monitor on patient.
6. The overall check of the monitor, including the safety check, should be performed
only by qualified person once every 6 to 12 month and each time after fix up.
2. It is recommended to regularly cleaning the monitor shell and the screen. Use only
NOTE
MMED6000DP:
2. Most cleaning agents must be diluted before use. Follow the manufacturer's
MMED60000DP
(S6) Service Manual
4. Do not let the cleaning agent enter into the chassis of the system. Do not
5. Do not leave the cleaning agents on any part of the device surface.
6. Except for those cleaning agents listed in “NOTE” part, following disinfectants
NOTE
diluted sodium hyocihlo depends on how many organisms (blood, mucus) on the
Hydrogen Peroxide 3%
Alcohol
Isopropanol
NOTE
MMED60000DP
(S6) Service Manual
9.3 Sterilization
CAUTION
1. Follow the manufacturer’s instruction to dilute the solution, or adopt the lowest
possible density.
5. Use a moistened cloth to wipe off any agent remained on the monitor.
7.Appropriate disinfecting materials for ECG lead, SpO2 sensor, blood pressure
cuff, TEMP probe, IBP sensor are introduced Operator’s Manual respectively.
NOTE
Before cleaning the monitor or the sensor, make sure to turn off the power and
1 Cleaning
MMED6000DP monitor and sensor surface can be cleaned with hospital-grade ethanol
2 Sterilization
Ethylate: 70%
ethanol: 70%
Acetaldehyde
4 Disinfection
1. The TEMP probe should not be heated above 100℃ (212℉). It should only be
4. The rectal probes should be used, if possible, in conjunction with a protective rubber
cover.
5. To clean the probe, hold the tip with one hand and with the other hand rubbing the
probe down in the direction of the connector using a moist lint-free cloth.
NOTE
disposed of properly.
NOTE
1. Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the
surface of the sensor, and then dry it with a cloth. This cleaning method can also be
2. The cable can be cleaned with 3% hydrogen dioxide, 7% isopropanol, or other active
reagent. However, connector of the sensor shall not be subjected to such solution.
MMED60000DP
(S6) Service Manual
CHAPTER 10 WARRANTY
⑴Repair response time: AM9:00 to PM17: 30 on Monday to Friday except legal holiday.
the instrument when you dial our service line (86-10-86863551). Professional
repair engineer online provides technical support.
Field inspecting: we will send engineers to repair the instrument if necessary.
Certified engineers of our company or local repair team trained by our company
provide this service.
⑶ Spare machine for repair: it is used to replace the damaged machine for customer
using, customer should send the damaged machine to us to repair.
⑷ Repair for sponsoring and contributing machine: customer should send the machine
to us to repair.
⑴ Warranty does not apply to the damage or loss sustained due to well-known
MMED60000DP
(S6) Service Manual
⑵ Non-service items:
① The cost and insurance of dismantling and testing, overhauling, reinstall, transfer,
moving the instrument or parts.
② Damage or loss sustained due to inspected or repaired by other institute that is
not certified
③ Damage or alteration by anyone other than our company authorized service
personnel.
⑶ The damage or lose sustained due to connection to peripheral equipment (such as
printer, computer etc.), that are not provided by our company are not covered by the
warranty.
⑷Obligation restriction: In the duration of warranty, if the operators use other fittings
that are not provided by us, we reserve the right to cancel warranty.
Due to we stipulate the warranty period according to the relevant electronic regulation of
country, which we stipulate is on year, accessory is three months. When customer
requires to extending the warranty period, you should consider whether it is reasonable.
Because electronic product quickly replace, as to the warranty period over three years,
purchased accessories may be out of stock. In this case, we will adopt to entirely
upgrade or replace the old, you should pay the minimum acceptable cost of renewed
device.
10.6 Repackaging
Remove all the detectors, leads and accessories and put them into the plastic bag.
Try to use the original packaging case and materials. Any damage due to the improper
packaging during the transportation shall be responsible by the user.
If you are still within the period of warranty, please present the warranty card and one
copy of the invoice or receipt.
Please present a written note detailing all the troubles when repairing the
instrument.