Documente Academic
Documente Profesional
Documente Cultură
Learning Objectives:
At the completion of this session, the participant will be able to: Understand the medicolegal issues of clinical alcohol testing Understand good clinical laboratory practice for clinical alcohol testing Describe suitable specimen types and collection issues List assay methodologies Describe point-of-care tests for alcohol, including breath and saliva testing
Specimen Types
Plasma, serum, blood, urine, breath, and saliva are the commonly used specimens.
Specimen Collection
Use non-alcoholic disinfectant, such as benzalkonium chloride or povidone-iodine to cleanse the venipuncture site The laboratory requisition should contain:
Patients name or identification number Date and time of collection Ordering physicians name Phlebotomists name Container type
Specimen Preservative
Serum: Container with no anticoagulant. Allow specimen to clot. If analysis is delayed, add sodium fluoride (minimum 10 mg/ml) for storage
Plasma or whole blood: Potassium oxalate (5 mg/ml) and sodium fluoride (1.5 mg/ml) for 5C storage to 48 hours and 20C for long term storage. For storage at unrefrigerated temperature, use sodium fluoride at 10 mg/ml
Laboratory Report
Laboratory Report Should Contain the Following Information
Patients name or identification number Specimen number Date and time of specimen collection and
receipt in laboratory
Assay Methodologies:
Gas chromatography and enzymatic oxidation Gas chromatography Advantages: Specificity for ethanol. Enhanced with the use of multiple columns or varying chromatographic conditions
Gas Chromatography
Disadvantages:
Requires highly trained technical staff Analysis slower than enzymatic assay
The reaction is monitored following the absorbance of NADH at 340 nm or that of a color product at a higher (visible) wavelength formed by reacting NADH with a dye
.
Disadvantages: Not specific for ethanol. Other alcohols can interfere at high concentrations Will miss methanol and isopropanol overdoses
Assay Calibration
Calibrator concentrations should span those encountered in clinical practice and should also include the limits of assay linearity. Calibrators can be purchased commercially, or prepared in the laboratory using pure ethanol or analytical grade 95% ethanol. Assay accuracy should be checked against reference standard obtained from the National Institute of Standards and Technology (SRM 1828a) or against materials traceable to the NIST reference standard.
Frequency of Calibration
Enzymatic assays Follow manufacturers recommendations. Stability of calibration can be verified with the use of well-characterized controls in each run GC Assays Require more frequent calibration due to drifting of instrument operating conditions. The laboratory should validate calibration stability to determine the frequency of calibration. Calibration stability must be verified with stable, well characterized controls
Controls
Quality control program should be consistent with clinical laboratory standard of practice. Serum and blood controls are available commercially or can be prepared by the laboratory
Control concentrations should be chosen to monitor reliability near clinical decision points and assay linearity limits Accuracy can be assessed by participation in external proficiency survey
Blood collection is invasive Risk of injury or infection Requires specially trained personnel Turnaround time
The measured BrAC can be converted to BAC using the partition ratio: BAC = BrAC x 2100
Therefore, for a given blood alcohol concentration, an individual with a higher Hct will have a higher breath alcohol concentration than another individual with a lower Hct.
1. Convert BrAC to BAC using 2100:1 ratio. BAC in gram of alcohol per 100 ml of blood. Recommended. Physicians and nurses are more familiar with medical lab results in concentration units per 100 ml of blood 2. BrAC directly in gram of alcohol per 210 liter of breath. Define legal limits based on breath alcohol in g per 210L of breath
Gas chromatography
Electrochemical oxidation/fuel cell
Infrared spectrometry
IR spectrum(2 -25 m) is characteristic of ethanol. But BrAC IR analyzer is limited to 1 or 2 wavelength, resulting in loss of specificity The major concern is acetone interference
Chemical Interference
There are many potential chemical interfering compounds.
Acetone is the major interference. There are many clinical conditions which give elevated breath acetone Cases of interference by kerosene, menthol, tetrahydrofuran, toluene have been reported. Interpretation of a positive breath test includes assessing if the patient has had prior exposure to potentially interfering substances
Controls can be certified dry gas standards or prepared using a NHTSA-approved simulator and certified alcohol solutions
9. Record test date/time, device, QC results, patient ID, and test result 10. Prompt and accurate reporting
Non-invasive sample collection Sufficient sample quantity readily available Easy test performance Good approximation of BAC
Self-assessment Questions
1. Which of the following requirements for clinical alcohol testing is INCORRECT?
d. Specimen type must be specified and indicated in report 2. Which of the following statements is CORRECT? a. Serum alcohol conc. is higher than that of whole blood b. Saliva conc. is slightly lower than that of whole blood c. Whole blood conc. is higher than that of urine
3. The laboratory report should contain: a. Unique patient and specimen identification information
a. It is a quantitative assay
b. It is less specific for ethanol than the enzymatic assay c. It can detect methanol
d. Headspace analysis has lower through-put than the direct injection technique
5. Which of the following statements about the enzymic assay is CORRECT? a. It is specific for methanol and ethanol b. It is the most frequently used clinical alcohol assay c. It can be used for detecting isopropanol overdose d. In the assay, ethanol is converted to acetone 6. Which of the following statements on breath alcohol testing is INCORRECT? a. Breath collection is non-invasive
a. The blood:breath alcohol partition ratio of 2100 is constant at all times and for all test subjects
b. Breath analyzer displays result as blood concentration using a breath to blood conversion factor of 2100 c. Measured breath alcohol concentration is independent of the test subjects hematocrit d. Acetone is the only known interfering substance