1. ProductConcepts Speedy HighQualityImage


EasyandQuick

FirstCanonBrandedOCTislaunched!

2. AppealPoints
21.3m depthresolution 22.70,000Ascan/sec 23.10mmscanwidth 24.Just2clicksforacquiringtomography 25.Autotrucking 26.Followupexams 27.Autoalignment/Autofocus/AutoCgate 28.FlyingspotSLO 29.8typesofscanmodes 210.Noisereductionbyaveragingupto50images 211.10layerboundaryrecognition 212.Vitreous/Choroid mode 213.Versatilereports 214.Flexiblelayout 215.Retinalcameraimageimport 216.Anteriorsegmentadapter 217.Normativedatabase 218.InstallationImage 219.Specification 220.ProductComponents&Options 221.ComparisonchartwithCompetitorsproducts

Speedy

Highqualityimage

21.3mDepthresolution
Highimagequalityisarchivedbym Depth resolution.
AMD AMD

(L) AMD Male age 76

Averaging (50/50) Averaging(50/50)

Default display: Fit

22. 70,000Ascan/sec
by70,000 Ascan/sec.

Highspeedscan&Highimagequalityisrealized

(R) Macular hole Female age 61 DC Averaging (10/10)

23.10mmScanWidth
DC Averaging(50/50)

10mm :MaxScanWidth Widescannedareaisdisplayed atonce.


DE Averaging (10/10)

24.

Just2clicksforacquiringtomography

Start

Just2clicksgiveimagesFirstclickmakesstartingautoalignment.Secondclick makescapturingimage.Rightsideclickorpushingspacekeytwicemakesame resultasabove. OCT-HS100 Sales book V1.0

25.Autotrucking
Autotruckingfunctionensuretheexactcapturing image
AnteriorChamberAutoTrucking: Autodetectedpositionascenterofthepupilor designatedpositionby usercouldbetruckedasthecenterofscanarea. RetinalAutoTruckingWhiledisplayingretinalliveviewimage,Scanareais truckedautomaticallybycalculatedcenterpositionsmovingdistance.
Anterior AutoTrucking

ON/OFFselectablerespectively 2sizesofinternalfixationlightisselectable

2mm
RetinalAutoTrucking

6mm

26.FollowupExam
RetinalAutoTruckingfunctionenableuserstocapture samepointwithpastexaminationdata.Difference:lessthan100m Whenanexaminationisselectedinthepatientscreenorwhencapture screenisopenedfromthereportscreen,followupexamsetis automaticallyselected.
Thefollowingitemsshouldbesetto sameconditionsasaprevious examination. RightorLefteye ScanMode ScanPosition ScanSize(width,distance) NumberofAveraging AutomaticorManualMode InternalorExternalFixationTarget InternalFixationTargetSize InternalFixationTargetPosition CGateorientation

Capturedpointisregistered.

27.AutoAlignment/Autofocus/AutoCgate
Variousauto functionsmakeexaminationseasy&short.

AutoAlignment Centerofthepupilisdetectedandsetascenterofimage AutoRefsfunction/WorldsfirstfunctionasOCT AutofocusThe focusiskeptgoodlevelautomatically. AutoCgateThepositionoflivescanimageisadjustedautomatically.


Autoalignment

Theotherautoadjustmentfunctions AP(AutoPolarization) APisnotimplementedsincenoadjustment isregardedasunnecessaryforeachpatient ARP(AutoReferencePower) ARPisimplementedtomonitorthereference powerandfeedbacktocontrolNDfilter(for eachpatient,onlyforthefirstexamination)

AutoCgate AutoFocus

28.FlyingSpot SLO
Retinal image Preview (H44 V33)

Live image (No delay) High resolution For disease High resolution For blood vessel Simple mechanism

Whileradiatingbeamonretina, usercangaintheinformationofitby scanning

fsSLO hasbetterresolutionthanLSLOandcanobtain bettercontrastimages. (Nidek andHeidelberg adoptsfsSLOaswell.) LSLOisbetterinquickscanbutnotsomuch differenceasfsSLOthesedays (Zeiss adoptsLSLO.)

29.8 typesofscanmodes
Scan Mode Scan Direction Horizontal Fixation Target Macula C-Gate Orientation Vitreous or Choroid Scan Size Description Macula 3D 10x10mm For macula analysis of macula disease. This mode supports to change C-Gate orientation for choroidal disease (AMD) and other macula disease. For macula analysis of glaucoma. This mode has a vertical scanning direction for symmetrical diagnostic. For disc analysis of glaucoma. Volume scan,12 axis by 15 degrees extracted. For general usage. This mode does not support analysis reports for specific disease. For multi purpose Averaging: 5, 10 B-scans For high image quality Averaging: 5, 10, 20, 50 B-scans For anterior imaging For anterior imaging Averaging: 5, 10, 20, 50 B-scans

Glaucoma 3D

Vertical

Macula

Vitreous

10x10mm

Disc 3D

Horizontal

Disc

Vitreous

6x6mm

Custom 3D

Horizontal or Vertical Horizontal and Vertical Horizontal and Vertical Horizontal Horizontal and Vertical

Macula or Disc Macula or Disc Macula or Disc Center Center

Vitreous or Choroid Vitreous or Choroid Vitreous or Choroid -

3x3mm to 10x10mm 3x3mm to 10x10mm 3x3mm to 10x10mm 6x6mm 3x3mm to 6x6mm

Multi Cross

Cross

Anterior 3D Anterior Cross

210.Noisereductionbyaveragingupto50images
HighqualityimagesareprovidedbyMAX50images averagingfunction.

1image

10images

20 images

50 images

Averaging:MultiCrossScan(5,10Bscans), CrossScan( 5,10,20,50Bscans)

210.Noisereductionbyaveragingupto50images
M210DC vs RS-3000Nidek
Canon HS-100 DC averaging(50/50) NIDEK RS-3000 averaging(7/10)

(L) ERM Female age 75


Canon HS-100 DC averaging(50/50)

NIDEK RS-3000 averaging(7/10)

211. 10layerboundaryrecognition
OCTHS100softwarecanrecognizeboundaryof Bruchsmembrane(BM).

Recognize10boundaries
ILM RNFL/GCL GCL/IPL IPL/INL OPL/ONL OPL/ONL IS/OS OS/RPE RPE/Choroid BM

15

212. Vitreous/Choroid mode


Usercandiagnosebychangingviewmodes suitable forvitreousorchoroid.
Vitreous is shown in over gate Vitreous Vitreous mode mode

Choroid is shown deeper than usual in under gate. Choroid Choroid Mode Mode (EDI) (EDI) Availablewhenyouchoose: Macula3D,MultiCross Custum 3D,Cross

213. Versatilereports
Easytochangethereporttype
View Mode Single Both Eyes Comparison Progression General 3D A standard analysis report for a single examination. A report for comparison of right and left eyes. OCT-HS100 software search examination of another side eye automatically. A report for comparison of 2 examinations. OCT-HS100 software search suitable examination automatically. A report for confirmation of disease progress. OCT-HS100 software search suitable examinations automatically. Advanced visualization for single 3D examination. 3D visualization for single 3D examination.

Single

Both eyes

Comparison

Progression

General

3D

17

213. Versatilereports
Easytomakereportsformaculadiseaseorglaucoma
Scan mode Macula 3D Glaucoma 3D Disc 3D Custom 3D Multi Cross Cross
Macula analysis for macula disease Scan Scan modeMacula 3D

Report type Macula analysis for macula disease Macula analysis for glaucoma Disc analysis for glaucoma. General 3D report General 2D report General 2D report
Macula analysis for glaucoma Scan modeGlaucoma 3D Disc analysis for glaucoma Scan modeDisc 3D

Focusonthethicknessofcornea

FocusonthethicknessofNFL+GCL+IPL

Focusontheparametermeasurementson NFL andONH

214.Flexiblelayout
Inspectorandpatientcanbepositionedflexibly.
Supporting communicationbetween Inspectorandpatients Savingspaceand increasingefficiencyfor usersworkflow Competitors layout

Sided layout

45 degree layout

Nidek difficult to see patients face

L mark layout

Confronted layout Zeiss position is fixed

215.RetinalCameraImageImport
CaptureFundusimage,possibleto display(oroverlay)insteadofthe SLOimage
ImportJPEG orBMP retinalcameraimages Softwaretrimstheretinal imagesinaformatof4:3. SLOimageisoverlaidonto theretinalcameraimage automaticallyormanually sothattheposition matches. Transparencycanbe adjustedfortheoverlay.

216. Anteriorsegmentadapter
Byattachingananterioradapter(optional),itspossible totakeOCTimageoftheanterioroftheeye

217. NormativeDatabase
Normativedatabasewillbesupportedinthenextversionof software. TBI
Normativedatabasewillworkwiththefollowingitems:
Macular 3D Deviation Map Significance Map Thickness Sector Thickness Profile Thickness Time-line Graph Thickness Measurement Table Glaucoma D NFL+GCL+IPL Deviation Map NFL+GCL+IPL Significance Map NFL+GCL+IPL Sector NFL+GCL+IPL Thickness Profile NFL+GCL+IPL Time-line Graph NFL+GCL+IPL Measurement Time-line Table Disk D RNFL Deviation Map RNFL Significance Map RNFL Thickness Sector ONH Measurement Table RNFL Thickness Profile RFNL Thickness Time-line Graph ONH Measurement Time-line Table

White base display when normative database is not applicable No database display when normative database is not applicable Sample of reports when normative database is not applicable

218.InstallationImage

23

219.BasicSpecifications
Retinal observationCapture Scan speed Cross lateral direction resolving power Depth resolving power OCT wavelength The smallest pupil diameterfor Working distance Method Range of observation 3D Macula for Macula scan pattern Disease 3D Macula for Glaucoma 3D Disc for Glaucoma 3D Custom for Advance Max 70,000 A-scan per second 20um 3.0um 855nm5nm 3.0mm 35mm flying spot SLO 3333OCT / 4433SLO 1024Ascan x 12810x10mm, Horizontal, Vitreous or Choroid 1024Ascan x 12810x10mm, Vertical, Vitreous 512Ascan x 2566x6mm, Horizontal, Vitreous default : 1024Ascan x 128 9x9mm, Horizontal, Vitreous, Macula X and Y range : 3-9mm(adjustable independently) C-Gate : Vitreous or Choroid Fixation Target : Macula or Disc default : 1536Ascan X and Y range : 3-9mm(adjustable independently) averaging : 1-10 C-Gate : Vitreous or Choroid Fixation Target : Macula or Disc default : 1536Ascan X and Y range : 3-9mm(adjustable independently) averaging : 1-50 C-Gate : Vitreous or Choroid Fixation Target : Macula or Disc Default is 2mm"X"at fundus Orange590nm 387x499x474 29kg

Multi Cross B-Scan

Cross B-scan

Internal fixation target DimensionsWeight Dimensions(WxDxH: mm) Weight

220.Productcomponents&Options
Basiccomposition OCTHS100MainUnit ACPowercordset Objectivelenscap Externaleyefixationlamp sameasCR2 ChinrestPaper(100sheets) DustCover Synchronouscable DVDR(CanonOCTSoftware) FixingpartsforCameralinkcable CableTie FerriteCore UserManuals(Mainunit,OCTSoftware) Standardaccessory Cameralinkboard PCIe1433 2Cameralinkcables Option AnteriorAdapter ForeheadPadforAnteriorAdapter ChinRestforAnteriorAdapter AnteriorAdaptermainunit Storagecase