Sunteți pe pagina 1din 45

MRIDeviceCompliance

MartinVogel,PhD KimberleyPoling ApplicationEngineeringTeam EasternUSA


1 2011ANSYS,Inc. October24,2011

Overview

HighsimulationefficiencyforMRI MethodthatenablesanonEEtoanalyzethethermaleffects ofanMRIscanonanimplant

2011ANSYS,Inc.

October24,2011

Maintools
HFSS 3DFullwaveElectromagneticsimulator Designer CircuitsimulatorthatinteractswithHFSS ANSYSProfessionalNLT ThermalsimulatorthatuseselectromagneticlossesfromHFSS asheatloads

2011ANSYS,Inc.

October24,2011

1.Simulationefficiency

MinimizeRAMand simulationtime
4 2011ANSYS,Inc. October24,2011

Genericbirdcagecoil
Ring with 24 ports, phases -180 165 deg

Ring with 24 ports, phases 0 345 deg

Around the coil is a cylindrical shield with open ends.

2011ANSYS,Inc.

October24,2011

Savingcomputerresources
Abirdcagecoilcanbecreatedwithmany,e.g.48,sourcesthat imposetherequiredrotatingfield. Arealisticbirdcagecoilhasfewsourcesandmanycapacitors. Suchacoilcanbesimulatedwithlessresources. WithacoupleofANSYStools,wehavecreatedsuchrealisticcoils, onefor1.5Tandonefor3T.

2011ANSYS,Inc.

October24,2011

Realisticcoilfeednetwork
Feed points simplified and optimized with ports and capacitors
Bot_1:1 C4 0 PNUM=1 RZ=50ohm IZ=0ohm C5 C6 Cap C8 Cap C20 Cap 0 C19 C18 Cap C16 Cap C26 Cap 0 C24 Cap Cap Bot_2:1 Bot_3:1 Bot_4:1 C7 Cap Bot_5:1 Bot_6:1 Cap Bot_7:1 Bot_8:1 Bot_9:1 Bot_10:1 C17 Cap Bot_11:1 Bot_12:1 CapC237 Bot_13:1 Bot_14:1 C25 Cap Bot_15:1 Bot_16:1 C23 Cap Bot_17:1 C240 Bot_18:1 C39 Cap Bot_19:1 Cap Bot_20:1 C37 Cap Bot_21:1 Bot_22:1 C35 Cap Bot_23:1 C34 Cap Cap Cap Bot_24:1 Top_1:1 Top_2:1 Top_3:1 Top_4:1 Top_5:1 Cap C41 Top_6:1 Top_7:1 Cap Top_8:1 Top_9:1 Top_10:1 Top_11:1 Cap C54 Top_12:1 Top_13:1 Cap C64 Top_14:1 Top_15:1 Cap C62 Top_16:1 Top_17:1 Cap C60 Top_18:1 Top_19:1 Cap C70 Top_20:1 Top_21:1 Cap C68 Top_22:1 Top_23:1 Cap C66 Top_24:1 Cap Cap Cap C65 Cap C67 Cap C69 0 Cap C59 Cap C61 Cap C63 0 PNUM=4 RZ=50ohm IZ=0ohm Cap C53 C56 Cap C55 Cap C57 0 Cap C40 C43 Cap C42 C44 0 PNUM=3 RZ=50ohm IZ=0ohm

PNUM=2 RZ=50ohm IZ=0ohm

C38 0 C36 Cap

Feed points are driven with phases 0, 90, 180 and 270 degrees.
7 2011ANSYS,Inc. October24,2011

S11 coilswithcapacitors
127.6 MHz

63.8 MHz

2011ANSYS,Inc.

October24,2011

FinetunecapacitorsforMRIimagequality
B+/B ratioisatleast20

1.5 T 124.5 pF
9 2011ANSYS,Inc. October24,2011

3T 26.5 pF

RotatingHwithcapacitors

Feed points

Capacitors have been finetuned in HFSS for best rotating H after optimization in Designer.
10 2011ANSYS,Inc. October24,2011

Applythisdesigntechniqueto birdcagecoilwithhumanbodymodel
RAM12.5GB Elapsedtime58minutes

11

2011ANSYS,Inc.

October24,2011

RotatingBforbodyinbirdcagewith48sources Comparewithresultsofothersimulations.

12

2011ANSYS,Inc.

October24,2011

RotatingBforhomogeneousbodyinbirdcage with4sourcesand44capacitorsisquitesimilar.
RAM5.5GB Elapsedtime22minutes

13

2011ANSYS,Inc.

October24,2011

Isitpossibletosimulatejustthebluevolumeand IMPOSEfieldsfromaseparatecoilsimulation? YES,youcandothat!

14

2011ANSYS,Inc.

October24,2011

Withimposedfields,therotatingBagainlooksverysimilar!
RAM1.6GB Elapsedtime7.3minutes

15

2011ANSYS,Inc.

October24,2011

Nowthatwehavesuchanefficientmethod,useitwitha complicatedheterogeneoushumanbodymodelanda pacemaker.

RAM4.3GB Elapsedtime20 minutes

16

2011ANSYS,Inc.

October24,2011

Nowthatwehavesuchanefficientmethod,useitwitha complicatedheterogeneoushumanbodymodelanda pacemaker.

RAM4.3GB Elapsedtime20 minutes

17

2011ANSYS,Inc.

October24,2011

LocalSARincutplane(3Tcoil)

18

2011ANSYS,Inc.

October24,2011

R13Schematic

19

2011ANSYS,Inc.

October24,2011

Thermalresult
Maximumoccursonneck,notatimplant

20

2011ANSYS,Inc.

October24,2011

Contoursof38.10Candhigher

Tissue around pacemaker

21

2011ANSYS,Inc.

October24,2011

Pacemaker,leadandheart

22

2011ANSYS,Inc.

October24,2011

2.Thermaleffectsof MRIscanonimplant

Minimizetheimplant designersexposureto electromagnetics


23 2011ANSYS,Inc. October24,2011

Genericbirdcagecoil

Around the coil is a cylindrical shield with open ends.


24 2011ANSYS,Inc. October24,2011

Whereshouldimplantsbeplaced?
Implantsshouldhaveworstcaselocationandorientation. Nextfewslidesillustratehowthisisachieved.

25

2011ANSYS,Inc.

October24,2011

Phantomplacement
|B| is almost uniform within coil (left). However, |E| is not (right). Make sure implants be located in a region with relative high E-field.

|B|

|E|

26

2011ANSYS,Inc.

October24,2011

WorstcaseSAR:implantnearphantomwall; doesnotstickoutofRFcoil

27

2011ANSYS,Inc.

October24,2011

Worstcase:implantparallelwithelectricfieldvectors

28

2011ANSYS,Inc.

October24,2011

Workflowforimplanttesting
Havereusablecoil+phantom modelinHFSS Havephantom+parameterized implant(s)inseparategeometry system Thisseparategeometrywillbeexposedtothefieldsfromthe coils. Thermalsimulatorwilldeterminetemperaturesof phantom+implant(s)duetoRFlosses. Nextslidesshowthisinmoredetail.

29

2011ANSYS,Inc.

October24,2011

WorkflowinR14Schematic
NoneedtoopenHFSS.

CAD

30

2011ANSYS,Inc.

October24,2011

CreateinotherCADtool importwith parametersintoprepared HFSSdesign

31

2011ANSYS,Inc.

October24,2011

Transfergeometrytothermalsimulator
NomoreneedforFile/ExportandFile/Import

Notetheparametersfrom TheCADtoolareavailable!
32 2011ANSYS,Inc. October24,2011

Libraryofthermalmaterialproperties isavailable

33

2011ANSYS,Inc.

October24,2011

Havepreparedasubsetwiththesame materialnamesasinHFSS.

34

2011ANSYS,Inc.

October24,2011

Objectscomeinwithcorrectmaterial assignment!

35

2011ANSYS,Inc.

October24,2011

DragandDropHFSSSolutiontoThermal Setup

DropandDrag

ResultingSchematicwithextraLink

36

2011ANSYS,Inc.

October24,2011

AssignLossesfromHFSStoObjectsinThermal

37

2011ANSYS,Inc.

October24,2011

EffectimplantsonlocalSAR

38

2011ANSYS,Inc.

October24,2011

Temperaturesonimplantsafter900s

39

2011ANSYS,Inc.

October24,2011

Addfinalmaximumtemperatureasanoutput parameter

40

2011ANSYS,Inc.

October24,2011

Inputparameter: Implantlength

Outputparameter: Max.temperature

41

2011ANSYS,Inc.

October24,2011

AttachDXtoinvestigatedesignvariations efficiently

Varylengthandinspectfinaltemperatures
42 2011ANSYS,Inc. October24,2011

Maximumtemperatureasafunctionof implantlength
(atreducedpower,nottobecomparedwithpreviousresult)

406080100120140160 Lengthimplant(mm)
43 2011ANSYS,Inc. October24,2011

180

Observations
Temperaturestendtodecreasewhenimplantisburieddeeper. Thismakessense,aselectromagneticfieldsdecaywithdepth. Forsmallimplants,temperaturesincreasewithimplantlength. Thismakessense,becausealongerimplantcanfunctionasamore effectivereceivingantenna(aslongasimplant /2atthisfrequency andinthisenvironment). Forlargeimplants,fieldinhomogeneitymakespredictionshardertomake. Resultsaresensitivetothegelselectricalandthermalmaterialproperties.

44

2011ANSYS,Inc.

October24,2011

Summary
Wehavedemonstrated

(1)Awaytoachieveveryefficientsimulations; (2)AsimulationflowthatenablesnonEMexpertstodetermine MRIinducedheatingfordifferentimplantdesigns.

45

2011ANSYS,Inc.

October24,2011

S-ar putea să vă placă și