Sunteți pe pagina 1din 90

M

e
a
s
u
r
e
,
w
h
a
t

|
s

m
e
a
s
u
r
a
b
|
e
,
a
n
d

m
a
k
e

m
e
a
s
u
r
a
b
|
e
t
h
a
t

w
h
|
c
h

|
s

n
o
t
.
G
a
|
|
|
e
o

G
a
|
|
|
e
|

1
5
6
4
-
1
6
4
2
}
Service Manual
COBRIX3 & COBRIX3 multibev
On-line Beverage Analyzer

Service Manual
COBRIX3 & COBRIX3 multibev
On-line Beverage Analyzer

While every precaution has been taken in the preparation of this document, Anton Paar GmbH assumes
no responsibility for technical or printing errors or omissions.
Nor is any liability assumed for damages resulting from the use of the information contained in this
instruction manual.
Anton Paar GmbH does not make a commitment to update the information in this manual. Specifications
are subject to change without notice.
All rights reserved (including translation). No part of this document may be translated, reproduced or
distributed in any form (print, photocopy, microfilm or any other process) without the prior written
permission of Anton Paar GmbH.
Trade marks may be used in this instruction manual without being marked as such. These are the
property of their respective owners and are legally protected.
Published by Anton Paar GmbH. Printed in Austria.
Copyright 2004 Anton Paar GmbH, Graz, Austria
Contact: Anton Paar GmbH
Anton-Paar-Str. 20
A-8054 Graz / Austria - Europe
Tel: +43 316 257-0
Fax: +43 316 257-257
E-mail: info@anton-paar.com
Web: www.anton-paar.com
Date: 31.12.2004
Document number: B43IB20C.fm
B43IB20 5
Contents
1 System Overview............................................................................................................................. 7
1.1 Versions of COBRIX3.............................................................................................................. 7
1.2 Functional Components/Spare Part List .................................................................................. 8
2 Safety Instructions......................................................................................................................... 17
3 Symbols in the Service Manual .................................................................................................... 19
4 Reference Manuals........................................................................................................................ 21
5 Service and Maintenance.............................................................................................................. 23
6 Cleaning the COBRIX3 .................................................................................................................. 25
7 Disconnecting the COBRIX3......................................................................................................... 27
7.1 Stopping the COBRIX3.......................................................................................................... 27
7.2 Disconnecting the Power Supply........................................................................................... 27
7.3 Disconnecting the Compressed Air Supply ........................................................................... 28
7.4 Closing the Bypass Valve...................................................................................................... 28
7.5 Disconnecting the COBRIX3 from the Product Line.............................................................. 29
7.6 Removing the Stainless Steel Covers ................................................................................... 29
7.7 Opening the Stainless Steel Housing Rear Door................................................................... 29
8 Checking the Pump Motor Overload Relay ................................................................................. 31
9 Relay for the Liquid Sensor .......................................................................................................... 35
10 24V Switched Mode Power Supply............................................................................................... 39
11 Line Protector................................................................................................................................. 41
12 Additional Alarm Relay.................................................................................................................. 43
13 Alarm/Warning Light...................................................................................................................... 47
14 LED Display.................................................................................................................................... 49
15 Replacing the CARBO 2100 CO2 Transducer ............................................................................. 53
16 Replacing the DPRn or DSRn Transducer................................................................................... 57
17 Replacing the Shaft Seal of the Pump......................................................................................... 61
18 Replacing the Pump ...................................................................................................................... 67
19 Disconnecting the Pump Cable.................................................................................................... 69
20 Wetted Parts................................................................................................................................... 75
21 List of Spare Parts for 1 Year of Operation................................................................................. 77
22 List of Tools.................................................................................................................................... 79
23 Troubleshooting............................................................................................................................. 81
24 Document Numbers....................................................................................................................... 83
Appendix A: Service/Maintenance Report .................................................................................... 84
Appendix B: Feedback Form.......................................................................................................... 90
6 B43IB20
1 System Overview
B43IB20 7
1 System Overview
1.1 Versions of COBRIX3
There are 2 versions of the COBRIX3:
COBRIX3 multibev Beverage Analyzer:
All-round version with DSRn for
sugared and diet beverages
sugared beverages with sugar inversion compensation
beverages with alcohol (alcoholic lemonades, malternatives, etc.)
beer
wine
COBRIX3 Beverage Analyzer:
Basic version with DPRn for sugared and diet beverages.
1 System Overview
8 B43IB20
1.2 Functional Components/Spare Part List
Fig. 1.1 - 1 Side view of COBRIX3


Alarm lamp
Warning lamp
Main switch
Switch box including
mPDS evaluation unit
Bypass with liquid
Production
line
Stainless steel (SS)
protection cover
Front door
CO
2
transducer
CARBO 2100
DPRn or DSRn
density/sound velocity
transducer (see warning 3
sensor and valves
Rear door
Lock (see warning 1 below)
Pump (see
warning 2
below)
below)
(see warning 4
below)
Handles (see warning 5 below)
1 System Overview
B43IB20 9
Warning1:
High voltage. Danger of electric shock. Before removing outside covers,
switch off the instrument and disconnect it from the mains.
Service procedures may only be carried out by authorized and fully trained
service personnel.
Warning 2:
The pump may be under pressure. Before opening the pump, disconnect the
pump from the product line and exhaust it.
Service procedures may only be carried out by authorized and fully trained
service personnel.
Warning 3:
The product line may be under pressure. Before opening the DPRn or DSRn,
disconnect it from the product line.
Service procedures may only be carried out by authorized and fully trained
service personnel.
Warning 4:
The product line may be under pressure. Before opening the CARBO 2100,
disconnect it from the product line.
Service procedures may only be carried out by authorized and fully trained
service personnel.
Warning 5:
To remove the cover, use the handles. Do not use the handles for carrying the
COBRIX3.
Warning 6:
Hot product or hot CIP liquid may come out of the sample valve.
Wear protective gloves to open and clsoe the sample valve.
Take care not to get in touch with the hot rinsing liquid or hot product.
1 System Overview
10 B43IB20

Alarm lamp
Measuring value



Optional displays for Brix / %Diet,
CO
2
and alcohol or temperature
Warning lamp

Pump start

storage button Alarm Reset Button
button
Main switch
Optional product
selector switch
Pump ON/OFF
switch
Fig. 1.1 - 2 Operator Interface of COBRIX3
1 System Overview
B43IB20 11
Fig. 1.1 - 3 Parts of COBRIX3

12 B43IB20















7045




















12864 11040












11039


Fig. 1.1 - 4 Options for COBRIX3
Bottle counter
Flow monitor Product selector switch
3 LED Displays
1 System Overview
B43IB20 13
COBRIX3 MultiBev beverage analyzer with Density-Sound Transducer:
10765 COBRIX3 MULTIBEV BEVERAGE ANALYZER 115V WITH DSRn
10764 COBRIX3 MULTIBEV BEVERAGE ANALYZER 230V WITH DSRn
Consists of:
DSRn 427 density/sound velocity transducer
CARBO 2100 CO
2
transducer
mPDS 2000 V3
By-pass Product Pump
Liquid Sensor
COBRIX3 beverage analyzer with Density Transducer:
10763 COBRIX3 BEVERAGE ANALYZER 115V WITH DPRn
10762 COBRIX3 BEVERAGE ANALYZER 230V WITH DPRn
Consists of:
DPRn 427 density transducer
CARBO 2100 CO
2
transducer
mPDS 2000 V3
By-pass Product Pump
Liquid Sensor
Option:
75504 DAVIS DATA ACQUISITION SOFTWARE
7045 BOTTLE COUNTER
12864 FLOW MONITOR COBRIX3
11039 3 LED DISPLAYS COBRIX3 MOUNTED
11040 EXTERNAL PRODUCT SELECTOR SWITCH COBRIX3
12865 MOUNTING SET SAWA PUMP COBRIX3
Accessories:
1151 ELBOW TUBE 45 DN 25 TRICLAMP
1 System Overview
14 B43IB20
77390 ELBOW TUBE 90 DN 25 TRICLAMP
5107 TRI-CLAMP FILTER DN25 EPDM
12995 CONVERTER KIT PC RS232-RS485
Spare parts for "Speck pump NPY 2051":
77385 CENTRIFUGAL PUMP NPY-2051 115V 50/60HZ
77506 CENTRIFUGAL PUMP NPY-2051 230V 50/60HZ
77953 SHAFT SEAL FOR NPY-2051
O-RING FOR NPY-2051 CASE
77954
77957 IMPELLER NPY-2051 50/60 Hz
77507 WASHER G3/4" PCTFE
Spare parts for "Sawa Pump":
8428 CENTRIFUGAL PUMP SAWA 230V 50HZ
8429 CENTRIFUGAL PUMP SAWA 115V 60HZ
9404 SHAFT SEAL SLIP RING SAWA PUMP
9405 SHAFT SEAL LIP RING SAWA PUMP
8433 TRICLAMP WASHER SMALL 25MM EPDM
8432 TRICLAMP BRACKET DA 25mm 1.4301
10394 FLEXIBLE HOSE COBRIX3 SAWA PUMP-DPR DN10X260
8424 FLEXIBLE HOSE COBRIX3 MAIN LINE-SAWA PUMP DN12X800
Spare parts for COBRIX3 - bypass etc.:
70768 CLAMP 25mm DIN 32676 1.4301
70769 WASHER 25MM DIN 32676 EPDM
10767 LIQUID SENSOR RELAY 115V
10766 LIQUID SENSOR RELAY 230V
10790 LIQUID SENSOR COBRIX3
77514 WASHER G3/8" PCTFE
1910 BUTTERFLY VALVE DN25 CLAMP / WELDING END
10393 FLEXIBLE HOSE COBRIX3 MAIN LINE-SPECK PUMP DN12X830
4254 FLEXIBLE HOSE PUMP/DPR 3/4" DN10 3/8" 400 MM
10396 FLEXIBLE HOSE COBRIX3 CARBO-MAIN LINE DN10X1500
1 System Overview
B43IB20 15
10775 SAMPLE VALVE COBRIX3 TRICLAMP DN10
10395 FLEXIBLE HOSE COBRIX3 DPR-CARBO DN10X395
16 B43IB20
2 Safety Instructions
B43IB20 17
2 Safety Instructions
This service manual does not claim to address all of the safety issues
associated with the service and operation of the COBRIX3 and samples. It is
the service technicians responsibility to ensure all health and safety
regulations are met for the persons concerned prior to starting service and
maintenance work.
Before servicing the COBRIX3, read this manual completely.
Anton Paar GmbH only warrants the proper functioning of the COBRIX3 if
no unauthorized adjustments have been made to mechanical parts,
electronic parts and software, and the following points are adhered to.
Follow all hints, warnings and instructions in the manual to ensure the
correct and safe functioning of the COBRIX3.
Do not handle the COBRIX3 other than described in the manual. Anton Paar
GmbH is not liable for damages caused by incorrect handling of the
COBRIX3.
The installation procedure must only be carried out by authorized personnel
who are familiar with the installation instructions.
Do not operate the COBRIX3 if a malfunction is suspected, or damages,
injuries or loss of life cannot be excluded under all circumstances.
The COBRIX3 is not an explosion-proof instrument and therefore must not
be operated in areas where there is a risk of explosion.
Use only original spare parts and parts explicitly indicated as parts for repair.
Service or repair work must only be carried out by qualified personnel,
authorized service centers, professional workshops or by Anton Paar GmbH.
Use only recommended and appropriate tools.
Ensure that all operators are fully trained to use the COBRIX3 correctly and
safely.
Due to the nature of the measurement, the measuring results do not only
depend on the correct use and functioning of the COBRIX3, but may also be
influenced by other factors. We therefore advise that the analysis results are
plausibility tested before consequential actions are taken.
Repair and service procedures may only be carried out by authorized
personnel or by Anton Paar GmbH.
18 B43IB20
If the safety appliances of the COBRIX3 are opened, changed or interrupted
during the service, a safety check must be performed after the repair.
Never install the COBRIX3 Beverage Analyzer and the evaluation unit in
areas where strong electromagnetic fields are present.
Before carrying out any service procedures, the COBRIX3 Beverage
Analyzer must be disconnected from the mains, product line and
compressed air supply.
The COBRIX3 Beverage Analyzer contains electrostatically sensitive
devices. When handling modules/components of the instrument they must
be protected from electrostatic discharge. Use a grounded wristband when
handling these components. If you return modules for repair, use appropriate
packaging which protects these modules from electrostatic discharge.
Do not touch areas marked with this symbol (yellow) when the power is
turned on. This "Warning" sign on the instrument means "Carefully read
the accompanying documents".
3 Symbols in the Service Manual
B43IB20 19
3 Symbols in the Service Manual
The following symbols are used in the instruction manual:
Warning:
The "Warning" sign indicates a hazard.
It calls attention to an operating procedure, practice, etc. which, if not correctly
performed or adhered to, could result in injury or loss of life.
Do not proceed beyond a "Warning" sign until the indicated conditions are
fully understood and met.
Important:
The "Important" sign indicates a hazard.
It calls attention to an operating procedure, practice, etc. which, if not correctly
performed or adhered to, could result in damage or destruction of the
instrument or parts of it.
Do not proceed beyond an "Important" sign until the indicated conditions are
fully understood and met.
Hint:
The "Hint" sign calls attention to any additional information which might be
of use to the operator.
20 B43IB20
4 Reference Manuals
B43IB20 21
4 Reference Manuals
The standard COBRIX3 on-line beverage analyzer consists of:
the mPDS 2000V3 evaluation unit
the DPRn density transducer or DSRn density/sound velocity transducer
(multibev)
the CARBO 2100 CO
2
transducer
For detailed information concerning these instruments please refer to the
following manuals.
mPDS 2000V3
Instruction Manual
mPDS 2000V3
Program Description Beverages
mPDS 2000V3
Service Manual
CARBO 2100 CO
2
transducer
Instruction Manual
CARBO 2100 CO
2
transducer
Service Manual
DPRn density transducer
Instruction Manual
DSRn density & sound velocity transducer
Instruction Manual
22 B43IB20
5 Service and Maintenance
B43IB20 23
5 Service and Maintenance
To ensure reliable and trouble-free operation, the COBRIX3 must be checked,
serviced and cleaned periodically.
The COBRIX3 service and maintenance intervals depend on the operating
conditions. The following is recommended as a general maintenance procedure.
Daily
Check daily for error messages on the display of the evaluation unit.
Once a week
Check the inside of the stainless steel cover (SS cover) once a week for
leaks.
Check (visually) the condensate level in the pressure regulator.
Check the set pressure of the pressure regulator. The set pressure must be
4.5 bar.
Process instruments are often exposed to extreme working conditions such
as pressure shock, sudden changes in temperature, aggressive samples,
etc. which may cause drifts or failure. Therefore, it is essential to periodically
compare the measured results to reference values and readjust the
measuring system if required.
Check the indicated concentration of the product.
Check the indicated CO
2
concentration.
Check the indicated temperature of the product
Check the indicated line pressure.
Pump maintenance:
The maintenance interval depends on the process and operating conditions of the
pump. We recommend checking the following:
- Sample residues on the pump and motor
Clean the pump and motor
Although the pump and motor are protected by stainless steel covers,
product or rinsing liquid may get under the cover e.g. through the opening
5 Service and Maintenance
24 B43IB20
on the bottom. Remove product residues from the pump and motor using
warm water. This prevents the motor becoming blocked and ensures
proper cooling of the motor.
- Leakage at the COBRIX3 hoses
Tighten all fittings
We use high quality seal rings for all fittings of the COBRIX3. Immediately
after start up some leakage may occur due to the initial compression of
the seal rings.
- Leakage at the shaft seal
Replace the shaft seal
During normal operation there should be no leakage at the shaft seal.
Foreign substances or wear can damage the shaft seal causing leakage.
In this case, the shaft seal must be replaced.
Continuous leakage at the shaft seal may destroy the motor of the pump.
Every 6 month
Replace the diaphragms:
The diaphragms of the CARBO 2100 are mechanically stressed and will
wear out. They should be replaced after a period of 4 months up to half a
year. The period depends on the operating conditions.
Check the CARBO 2100 impeller visually and replace it if necessary.
Check the drives of the diaphragm valves visually and replace if necessary.
For more detailed service and maintenance information concerning the
CARBO 2100 CO
2
transducer, mPDS 2000V3 evaluation unit and DPRn or DSRn
transducer, please refer to the respective manual:
CARBO 2100 CO
2
transducer service manual.
mPDS 2000V3 service manual.
DPRn density transducer instruction manual or DSRn density & sound
velocity transducer instruction manual.
6 Cleaning the COBRIX3
B43IB20 25
6 Cleaning the COBRIX3
The COBRIX3 cleaning interval depends on the operating conditions. We
recommend cleaning the COBRIX3 at least once a week.
Check if any sample is visible on the instrument and clean if necessary.
Use water or warm water as the cleaning agent.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.6: Removing the Stainless Steel Covers.
26 B43IB20
7 Disconnecting the COBRIX3
B43IB20 27
7 Disconnecting the COBRIX3
7.1 Stopping the COBRIX3
Before carrying out any service procedures, it is necessary to STOP the
COBRIX3. Perform the following steps:
Inform the production manager that the COBRIX3 will be stopped
Turn off the power for the rotary pump using the Pump ON/OFF switch (see
Fig. 1.2).
Turn off the main switch (see Fig. 1.2)
7.2 Disconnecting the Power Supply
1. Turn off the main switch
The lower part of the COBRIX3 contains the transducers. For service or
maintenance work on the lower part of the COBRIX3, turn off the main switch of
the COBRIX3 (see Fig. 1.2)
2. Turn off the circuit breaker
Turn off the main switch. The upper part of the COBRIX3 contains the evaluation
unit and the electrical equipment. For service or maintenance work on the upper
part of the COBRIX3, the circuit breaker must be turned off. The circuit breaker is
located in a switch cabinet outside the COBRIX3. Please contact the plant
electrician.
Warning:
Before carrying out any service procedures, the COBRIX3 Beverage Analyzer
must be disconnected from the mains, product line and compressed air
supply.
7 Disconnecting the COBRIX3
28 B43IB20
7.3 Disconnecting the Compressed Air Supply
Disconnect the compressed air hose from the inlet at the pressure regulator of the
CARBO 2100 transducer.
7.4 Closing the Bypass Valve
Close the bypass valves to separate the COBRIX3 from the product line (see
Fig. 1.1-3).
Open the sample valve to relieve the pressure in the bypass (see Fig. 1.1-3).
Warning:
Only fully trained and authorized technicians may carry out this
procedure.
The COBRIX3 is supplied with line voltage which may be life-threatening if
handled improperly.
Make sure that the line voltage is turned off when handling the power
cable.
Warning:
Compressed air can be hazardous.
Turn off the compressed air supply before disconnecting the hose!
Fig. 7.3-1 Pressure regulator
7 Disconnecting the COBRIX3
B43IB20 29
For service or maintenance for the transducers, first separate them from the
product line.
7.5 Disconnecting the COBRIX3 from the Product
Line
For a safe separation of the COBRIX3 from the product line, we recommend
disconnecting the bypass from the product line completely.
7.6 Removing the Stainless Steel Covers
Tools:
Allen screwdriver 3mm
Remove the stainless steel (SS) covers for service or maintenance work on the
lower part of the COBRIX3.
Open the screws of the SS protection covers.
Use the handles to remove the SS protection covers (see Fig. 1.1-3).
7.7 Opening the Stainless Steel Housing Rear Door
Tools:
Key for the lock of the electronic housing
For service or maintenance procedures on the electronics it is necessary to open
the SS housing rear door.
Open the lock of the rear door (see Fig. 1.1-1).
After shipment the key for the lock is located behind the front door (see Fig.
7.7-1).
W
Warning:
The product line may be under pressure.
Close the bypass valves.
Open the sample valve to relieve the pressure.
Disconnect the bypass from the product line by opening the Triclamp at
the bypass hoses (see Fig.1.1-3).
7 Disconnecting the COBRIX3
30 B43IB20
Warning:
High voltage. Danger of electric shock. Before opening the rear door, switch
off the main switch and the circuit breaker. Service procedures may only be
carried out by authorized and fully trained service personnel.
Fig. 7.7 - 1 Key for rear door
8 Checking the Pump Motor Overload Relay
B43IB20 31
8 Checking the Pump Motor Overload
Relay
Tools:
Key for the lock of the housing rear door (see chapter 7.7).
W
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.7: Opening the Stainless Steel Housing Rear Door.
Fig. 8 - 1
32 B43IB20
A The trip current must be adjusted to the rated current of the pump motor
Typical values are mentioned in the following table:
B Push to test the overload relay

C B A
D
Fig. 8 - 2
Warning:
The rated current of the pump must be taken from the label directly on the
pump. A wrong setting can destroy the pump in case of a malfunction.
Type of pump (cat. No, voltage, ...) Typical current
77506,230 VAC, SPECK NPY 2051 2,5 A
77385, 115 VAC,SPECK NPY 2051 3,8 A
8428, 230 VAC, SAWA 1,8 A
8429, 115 VAC, SAWA 4,35 A
8 Checking the Pump Motor Overload Relay
B43IB20 33
C Trip indicator window: If the yellow indicator is visible, the overload relay has
tripped.
D Man mode of operation is selected. Push to reset the relay.
34 B43IB20
9 Relay for the Liquid Sensor
B43IB20 35
9 Relay for the Liquid Sensor
Spare parts:
10766 LIQUID SENSOR RELAY 230V
10767 LIQUID SENSOR RELAY 115V
Tools:
Key for the lock of the housing rear door (see chapter 7.7).
Slotted screwdriver 3mm
A The resistance value is set to 20 kOhm.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.7: Opening the Stainless Steel Housing Rear Door.
Fig. 9 - 1

A
36 B43IB20
Function:
After the relay is supplied with power, the LED (U) is ON. The liquid sensor is
connected to terminals C and MAX. If product is in the sensor, the relay contact
will be closed (terminals 11 and 14) and the LED (R) is ON. It is possible to check
this function using a short circuit between terminals C and MAX.
Fig. 9 - 2

U
Fig. 9 - 3

R
Warning:
High voltage. Danger of electric shock. The terminals A1 and A2 are
supplied with high voltage which is life-threatening. Service and maintenance
may only be carried out by authorized and fully trained personnel.
9 Relay for the Liquid Sensor
B43IB20 37
Replacement:
See chapter 7.2: Disconnecting the Power Supply.
Reconnect the cables to the correct terminals.
The cable ties are used to make a safe separation between circuits with high
and SELV voltage. Therefore, the cable ties must not be changed.
Mount the plastic screw to terminal 12 of the new relay.
38 B43IB20
10 24V Switched Mode Power Supply
B43IB20 39
10 24V Switched Mode Power Supply
Spare parts:
10782 POWER SUPPLY 85-264VAC 24VDC 2,1A
Tools:
Key for the lock of the housing rear door (see chapter 7.7).
A The switched mode power supply has a rated power of 50W. It is the power
supply for the:
- CARBO 2100 15W
- Alarm and Warning Lamps 20W
- COBRIX3 Electronics 10 W
- mPDS2000V3 24V circuits 5W
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.7: Opening the Stainless Steel Housing Rear Door.
Fig. 10 - 1

A

G

R
40 B43IB20
Function:
Green LED (G) is ON if the output voltage is active.
Red LED (R) is ON if the output has a short circuit.
11 Line Protector
B43IB20 41
11 Line Protector
Tools:
Key for the lock of the housing rear door (see chapter 7.7).
The line protector has a rated current of 10A. If the current is higher the protector
will trip.
A Lever
Position up: the protector is ON.
Position down: tripped.
B Trip indicator window:
Red: the protector is ON.
Green: tripped.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.7: Opening the Stainless Steel Housing Rear Door.
Fig. 11 - 1

A

B
42 B43IB20
12 Additional Alarm Relay
B43IB20 43
12 Additional Alarm Relay
Spare parts:
8434 RELAY 24V 1A EMG 12-REL/KSR
Tools:
Key for the lock of the housing rear door (see chapter 7.7).
Slotted screw driver 3mm
A Additional alarm relay
Terminal 13 and 14.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.7: Opening the Stainless Steel Housing Rear Door.
Fig. 12 - 1

A
12 Additional Alarm Relay
44 B43IB20
Function:
The additional alarm relay works simultaneously to the alarm Light. It can be used
to connect the alarm light signal to a PLC or other process control unit. A LED
within the relay indicates when the contact is closed.
Technical data:
Requirements for the circuit connected to this relay:
- Rated voltage: 24V
- Rated power: 10W
- Safety: connect only SELV circuits
- Safety: the isolation between the additional alarm relay circuit and circuits
with live voltage must be double insulation.
Fig. 12 - 2
12 Additional Alarm Relay
B43IB20 45
Replacement:
See chapter 7.2: Disconnecting the Power Supply.
Reconnect the conductors to the correct terminals.
The cable ties are used to make a safe separation between circuits with high
and SELV voltage. Therefore, the cable ties must not be changed.
Warning:
Use a cable with double insulation when installing the additional alarm
circuit. Use cable ties to make a safe separation between the alarm circuit
and circuits with high voltage.
The installation may only be carried out by authorized and fully trained
personnel.
46 B43IB20
13 Alarm/Warning Light
B43IB20 47
13 Alarm/Warning Light
Spare parts:
10779 LAMP 24V 10 W
To replace the lamp of the alarm or warning light, open the light housing:
3. Rotate the black ring in the direction "open".
4. Lift the housing.
5. The lamp is located at the top of the housing.
6. Replace the lamp.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
Fig. 13 - 1
13 Alarm/Warning Light
48 B43IB20

Lamp
Fig. 13 - 2
14 LED Display
B43IB20 49
14 LED Display
If it is necessary to change the default settings of the optional LED displays follow
the instructions bellow:
Settings of the mPDS2000V3 (e.g.: AO#1)
1. Set the format of the analog output #1 to 0-20 mA.
[Config] [AO-Format] [AO1-Form] [0-20mA]
2. Assign one measuring value to the analog output.
[Config] [Assign] [AnalogOut] [AO1]
3. Enter the product specific scaling of the selected value.
[Prod-Spec] [AO-Scale] [AO1-0/4mA]/ [AO1-20mA]
Connection mPDS 2000 to LEDs
It is not recommended to change the wiring of the LEDs:
the left LED is connected to AO#1
the LED in the middle is connected to AO#2 and
the right LED is connected to AO#3 of the mPDS
Set the position of the decimal point
Open the protection cover on the rear of the LED.
It is possible to change the position of the decimal point with jumpers 10, 11 and
12.
Important:
The scaling must be entered for all products in use.
Jumper on position 10: 000.0
Jumper on position 11: 00.00
Jumper on position 12: 0.000
14 LED Display
50 B43IB20
Adjust the scaling of the display
Example I.:
0-20 mA 0.0.0.0 - 0.9.9.9 or 0.0.0.0 1.9.9.9
Set jumpers 2 and 8, and connect a current of about 20 mA to input terminals
4 and 5 of the display.
Remove the red front cover of the display.
There are 3 adjustment potentiometers:
Adjust the displayed value with the ADJ screw on the front until it matches
the target value. The target value is the one displayed on the mPDS 2000 or
if you use an external current source, it is the value on a reference ampere
meter in the current loop.
Example II:
0-20 mA 0.0.0.0 0.5.0.0
Set jumpers 3 and 8, and carry out the steps described in the above
example.
Put on the cover again.
ADJ maximum range adjustment
LZ minimum range adjustment screw (zero point, not used)
BRIGHT contrast adjustment
Important:
The LED is able to display numbers between 0 1999.
Numbers 2000 can only be displayed with 3 digits (0200).
Make sure to assemble the protection cover on the rear and the front
cover after finishing the setting.
14 LED Display
B43IB20 51
Short description of the manufacturer
- +
24VDC
analog output
of mPDS
52 B43IB20
15 Replacing the CARBO 2100 CO2 Transducer
B43IB20 53
15 Replacing the CARBO 2100 CO
2

Transducer
Spare parts:
77514 WASHER G3/8" PCTFE
Tools:
Open-end wrench 10, 14 and 19 mm
Disconnect the hose at the inlet (I).
Disconnect the hose at the outlet (O).
Disconnect the analog and digital cables (see Fig. 14-2).
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.3: Disconnecting the Compressed Air Supply.
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.5: Disconnecting the COBRIX3 from the Product Line.
See chapter 7.6: Removing the Stainless Steel Covers.
Fig. 14 - 1

I
O
15 Replacing the CARBO 2100 CO2 Transducer
54 B43IB20
Replace the CARBO 2100 CO
2
transducer. The transducer is fixed with four
nuts.
To open the nut located at the lower left side (DL) it is necessary to remove
the air regulator. To remove the air regulator open the black nut (BN) of the
air regulator.
Disconnect the analog and digital cables.
Fig. 14 - 2
Fig. 14 - 3

BN
15 Replacing the CARBO 2100 CO2 Transducer
B43IB20 55
Fig. 14 - 4
DL
Hint:
Use new seals when remounting the hoses. Part number: 77514 WASHER
G3/8" PCTFE
Important:
Install a connection cable between terminals 17 and 20 on the mPDS 2000V3
if there is not already a connection cable.
Fig. 14 - 5
56 B43IB20
16 Replacing the DPRn or DSRn Transducer
B43IB20 57
16 Replacing the DPRn or DSRn
Transducer
Spare parts:
77514 WASHER G3/8" PCTFE
Tools:
Open-end wrench 14, 17 and 19 mm
Slotted screwdriver 3 mm
Pozidrive screwdriver (PZ 2x100)
Disconnect the hose at the inlet (I: see Fig. 15-1).
Disconnect the hose at the outlet (O: see Fig. 15-1).
Disconnect the DPRn or DSRn cable.
Warning.
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.3: Disconnecting the Compressed Air Supply.
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.5: Disconnecting the COBRIX3 from the Product Line.
See chapter 7.6: Removing the Stainless Steel Covers.
Fig. 15 - 1

O

I
16 Replacing the DPRn or DSRn Transducer
58 B43IB20
Replace the DPRn or DSRn transducer. The transducer is fixed with four
nuts and screws (4N). .
Fig. 15 - 2
Fig. 15 - 3
O
I
4N
Warning:
As the DPRn or DSRn transducer is very heavy (30 kg), two persons are
necessary to replace the transducer: One person fixes the transducer while
the other person removes the nuts. Afterwards, both people replace the old
transducer with the new transducer.
Hint:
Use new seals when remounting the hoses. Part number: 77514 WASHER
G3/8" PCTFE
16 Replacing the DPRn or DSRn Transducer
B43IB20 59
The standard DPRn or DSRn transducer is delivered with a PG9 cable gland. In
this case the cable gland must be replaced by the M12 plug. Connect the GND
cable to the housing (G), the + wire to the terminal with the red label (R) and the
"- wire to the terminal with the blue label.
Important:
When ordering a DPRn or DSRn transducer as a replacement, please specify
that it is required for a COBRIX3. It will then be shipped with the plug (P) for
the DPR cable connector.
Fig. 15 - 3

P
Fig. 15 - 4
16 Replacing the DPRn or DSRn Transducer
60 B43IB20
Fig. 15 - 5

R

G
17 Replacing the Shaft Seal of the Pump
B43IB20 61
17 Replacing the Shaft Seal of the Pump
Spare Parts:
77953 SHAFT SEAL for NPY 2051
77954 O-RING FOR NPY-2051 CASE
77957 IMPELLER NPY-2051 50/60 Hz
77507 WASHER G3/4" PCTFE
Tools:
Open-end wrench 10mm
Slotted screw driver 3mm
Cutting pliers
Retaining ring pliers
Disconnect the hose at the inlet (I).
Disconnect the hose at the outlet (O).
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.5: Disconnecting the COBRIX3 from the Product Line.
See chapter 7.6: Removing the Stainless Steel Covers
Fig. 16 - 1

I

O
17 Replacing the Shaft Seal of the Pump
62 B43IB20
(1): Lock ring
(2): Disk
(3): Spring
(4): Shaft seal part
(5): Counter ring
(6): O-ring
Fig. 16 - 2
Fig. 16 - 3
Open the pump cover:
Remove the hexagon
screws from the pump
casing. Take off the casing
cover using the two
screwdrivers from the pump
casing.
Fig. 16 - 4
Remove the Impeller:
Remove the feather key. Pull off
the impeller using the two
screws. Remove the locking
ring (1), disk (2) and spring (3).
17 Replacing the Shaft Seal of the Pump
B43IB20 63
Fig. 16 - 5
Fig. 16 - 6
Remove the casing:
Remove the pump casing
from the motor shaft using
the two screwdrivers.
Fig. 16 - 7
17 Replacing the Shaft Seal of the Pump
64 B43IB20
Press out the counter ring:
Press the counter ring (5) of the
mechanical seal out the pump
casing.
Fig. 16 - 8
Clean parts:
Clean the pump parts and the
motor shaft. Clean the shaft with
a lint-free cloth.
Fig. 16 - 9
Mount the shaft seal counter
ring (5):
Lubricate the O-ring (6) with oil
and press it into the pump
casing. Place the pump casing
on the motor flange and center
it. Press the counter ring (5)
over the motor shaft into the
casing.
Fig. 16 - 10
17 Replacing the Shaft Seal of the Pump
B43IB20 65
Check the smooth running of the pump by rotating the fan.
Fig. 16 - 11
Lubricate the seal rubber (4):
Lubricate the rubber of the shaft
seal part (4) with oil. Caution:
Do not lubricate the slide faces.
Mount the shaft seal part (4):
Press the part (4) slightly and
without interruption up to the
shaft seal counter part (5).
Fig. 16 - 12
Fig. 16 - 13
Mount the spring (3), disk (2)
and lock ring (1). Insert the
feather key into the shaft.
Caution: Push to fit. Place the
cover with the O-ring on the
casing. Caution Observe the
direction of the channel.
Tighten the hexagon screws
over cross.
17 Replacing the Shaft Seal of the Pump
66 B43IB20
Hint:
Use new seals when remounting the pump:
77954 O-RING FOR NPY-2051 CASE
77507 WASHER G3/4" PCTFE
Check the impeller visually and replace it if necessary:
77957 IMPELLER NPY-2051 50/60 Hz
18 Replacing the Pump
B43IB20 67
18 Replacing the Pump
Spare Parts:
77385 CENTRIFUGAL PUMP NPY-2051 115V 50/60HZ
77506 CENTRIFUGAL PUMP NPY-2051 230V 50/60HZ
77507 WASHER G3/4" PCTFE
Tools:
Open-end wrench 10, 19 and 32 mm
Cutting pliers
Snap-blade knife
Disconnect the hose at the inlet (I).
Disconnect the hose at the outlet (O).
Disconnect the pump cable (C).
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply.
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.5: Disconnecting the COBRIX3 from the Product Line.
See chapter 7.6: Removing the Stainless Steel Covers

I

C

O
Fig. 17 - 1
18 Replacing the Pump
68 B43IB20
Open the four screws (4S).
Replace the pump.
Important:
To disconnect the pump cable it is not necessary to open the electronic box of
the pump. Refer to chapter 18 Disconnecting the Pump Cable.
Fig. 17 - 2

4S
Hint:
Use new seals when remounting the pump:
77507 WASHER G3/4" PCTFE
19 Disconnecting the Pump Cable
B43IB20 69
19 Disconnecting the Pump Cable
Tools:
Open-end wrench 19 mm
Cutting pliers
Snap-blade knife
For the pump cable a quick connection system is used. Therefore, the cable (C)
can be removed without opening the electronic box of the pump.
Removing the cable:
Open the gland.
Remove the conductor.
Warning:
See chapter 7: Disconnecting the COBRIX3.
See chapter 7.1: Stopping the COBRIX3.
See chapter 7.2: Disconnecting the Power Supply
See chapter 7.4: Closing the Bypass Valve.
See chapter 7.5: Disconnecting the COBRIX3 from the Product Line.
See chapter 7.6: Removing the Stainless Steel Covers
Fig. 18 - 1

C
70 B43IB20
The conductor can be removed from the terminal point by pulling on the
conductor.
Surplus insulation must be removed before repeated connection.
Connecting the cable:
Observe in particular the following steps:
Strip the sheath to a length of approx. 15 mm (see Fig. 18-2 part 1) and slide
the union nut (1), the cap (2) and the rubber seal (3) onto the conductor.
Slide the rubber seal as far as the rim of the insulation and then slide the cap
onto the rubber seal. This provides the strain relief for the conductor (see
Fig. 18-2 part 2).
Insert the core ends into the feedthrough of the splice ring (see Fig. 18-2 part
3). To guarantee a clear assignment of the cores, the individual core
feedthroughs of the splice ring are numbered (1, 2, ..).
Cut off the projecting core ends. Make sure that the core ends are flush with
the splice ring: They can protrude by up to 3 mm but must not be too short.
Insert the prepared conductor into the contact carrier (4) (see Fig. 18-2 part
1). Turn the conductor with splice ring until the coding pins fit exactly into the
corresponding guides (see Fig. 18-2 part 4).
Close the gland by tightening the union nut. This presses the core ends into
the insulation displacement terminal blocks, cuts open the core insulation
and creates the electrical contact.
19 Disconnecting the Pump Cable
B43IB20 71
union nut
cap
rubber seal
contact carrier
splice ring
Fig. 18 - 2 Connecting the pump cable
19 Disconnecting the Pump Cable
72 B43IB20
Core assignment for COBRIX3 with Serial Numbers below 672
Connect the pump cable to the contacts as explained above. The internal
connection in the electronic box of the pump is already carried out and must not
be changed.
Core assignment for COBRIX3 with Serial Numbers 672 and higher
Connect the pump cable to the contacts as explained above. The internal
connection in the electronic box of the pump is already carried out and must not
be changed.
1
2
3
Front view:
Connect the wires of the pump cable to the
cutting contacts as shown at the drawing.
The wires muat be inserted at the following
positions at the splice ring:
1 ... L
2 ... Protective conductor
3 ... N
Protective conductor
N
L
19 Disconnecting the Pump Cable
B43IB20 73
3
4
2 1
Front view:
Conect the wires of the pump cable to the
cutting contacts as shown at the drawing .
The wires must be inserted at the following
positions at the splice ring:
1 ... L
2 ... Protective conductor
3 ... N
4 ... not used
N
Protective conductor
not used
L
74 B43IB20
20 Wetted Parts
B43IB20 75
20 Wetted Parts
Part.No Item Material
1 1910 BUTTERFLY VALVE DN25 SS 1.4571
SS 1.4302
EPDM
2 70769 WASHER 25MM DIN 32676 EPDM EPDM
3 10790 LIQUID SENSOR COBRIX3 PSU
SS 1.4571
EPDM
4 77390 ELBOW TUBE 90 DN 25 TRICLAMP SS 1.4404
SS 1.4301
5 / FLEXIBLE HOSES SS 1.4571
PTFE
6 77385
77506
CENTRIFUGAL PUMP 115V
CENTRIFUGAL PUMP 230V
SS 1.4581
SS 1.4122
SiC/ C/ Viton
EPDM
Peek
7 / DPRn or DSRn TRANSDUCER Hastelloy
C276
8 / CARBO 2100 CO
2
TRANSDUCER
Refer to the CARBO 2100 service
manual
/
9 10775 SAMPLE VALVE COBRIX3 SS 1.4571
SS 1.4404
76 B43IB20
21 List of Spare Parts for 1 Year of Operation
B43IB20 77
21 List of Spare Parts for 1 Year of
Operation
Wearing parts:
The washers are wearing parts which should be replaced after opening the
connection.
Part.No Item Quantity/
1 System
Quantity/
5 Systems
77385 CENTRIFUGAL PUMP NPY-2051
115V 50/60HZ
1
77506 CENTRIFUGAL PUMP NPY-2051
230V 50/60HZ
1
77953 SHAFT SEAL
FOR NPY-2051
1 2
77954 O-RING
FOR NPY-2051 CASE
1 2
77507 WASHER G3/4" PCTFE 2 6
77514 WASHER G3/8" PCTFE 4 12
70769 WASHER
25MM DIN 32676 EPDM
3 9
10767 LIQUID SENSOR RELAY 115V 1
10766 LIQUID SENSOR RELAY 230V 1
10790 LIQUID SENSOR COBRIX3 1
10782 POWER SUPPLY 85-264VAC
24VDC 2.1A
1
10779 LAMP 24V 10 W
Warning, Alarm
3
8434 RELAY 24V 1A EMG 12-REL/KSR 2
Hint:
For spare parts of the CARBO 2100 refer to the service manual CARBO 2100.
21 List of Spare Parts for 1 Year of Operation
78 B43IB20
All other spare parts:
We suggest you keep the spare parts listed above on stock. Process equipment
is often exposed to extreme working conditions which may cause unexpected
failure.
22 List of Tools
B43IB20 79
22 List of Tools
Item Fig.
Key for the lock of the electronic housing 21-1
Cutting pliers 21-1
Retaining ring pliers 21-1
Snap-blade knife 21-1
Slotted screwdriver (3mm) 21-2
Allen screwdriver (3mm) 21-2
Pozidrive screwdriver (PZ 2x100) 21-2
Open-end wrench 10 mm 21-3
Open-end wrench 14 mm 21-3
Open-end wrench 17 mm 21-3
Open-end wrench 19 mm 21-3
Open-end wrench 32 mm 21-3
Fig. 21 - 1
22 List of Tools
80 B43IB20
Fig. 21 - 2
Fig. 21 - 3
23 Troubleshooting
B43IB20 81
23 Troubleshooting
General requirements:
- Liquid sensor
The pump starts automatically if the liquid sensor defects sensor detects
liquid in the line. Make sure that the liquid sensor is properly installed and
that it switches the pump on when liquid is in the line.
- Empty lines at line shut down
The COBRIX3 and bypass hoses must be installed so that the product runs
out of the COBRIX3 bypass when the main line is empty.
If the COBRIX3 bypass is not completely empty the liquid sensor switches
the pump on and off frequently within a short time.
This may cause the pump to overheat.
- Cleaning before emptying the line
After production and before line shut down, all tubes and hoses must be
properly cleaned with CIP. If CIP is not carried out product residues may
stay in the pump and could clog the shaft seal. If a pump with clogged shaft
seal is switched on, the retaining ring holding the shaft seal is stressed
beyond specifications. This may destroy the pump.
- Electrical ratings of the pump motor
Make sure that the power supply lies within the specified range given by
the label on the motor. Frequent power failures may cause the motor to
overheat.
Possible causes for motor failures and how to avoid them:
- Liquid entering the motor
The motor will be destroyed by liquid entering its housing.
The cover of the terminal box must be screwed on tightly.
Cable glands in the terminal box must be tight.
Absolutely no liquid must be splashed on the motor because liquid may
enter the pump through the bearings.
- Damaged motor bearings
If product or other liquids are splashed on the motor, the bearings may clog.
Always use the pump cover and do not splash any liquid on the motor.
Do not use steam jet cleaner!
- The pump periodically switching on and off
If the pump switches on and off several times within a short period the
motor may be overheated. Make sure that the liquid in the COBRIX3
23 Troubleshooting
82 B43IB20
bypass runs off when the main line is empty.
- Power supply
Frequent power failures can cause the motor to overheat.
Make sure that the specified power ratings are not exceeded.
If the mains voltage is unstable, use a UPS or take other measures to
stabilize the voltage.
- Clogged shaft seals
If the main line is emptied before cleaning, product residues may remain in
the pump.
This may clog the shaft seal. If a pump with clogged shaft seal is switched
on, the retaining ring holding the shaft is stressed beyond specifications.
This may destroy the pump.
Clean the bypass with CIP before line shut down!
- Solid particles in the pump
Solid particles (metal filings, etc.), as may occur during putting into
operation or after repair work, can damaged the impeller inside the pump.
Use the Triclamp filter when putting into operation and when changing the
piping anywhere along this main line.
Please refer to the following instruction manuals for further details:
1. COBRIX3 instruction manual
2. mPDS 2000V3 instruction manual
3. DPRn density transducer or
4. DSRn density sound velocity transducer instruction manual
5. CARBO 2100 CO
2
transducer instruction manual
24 Document Numbers
B43IB20 83
24 Document Numbers
Document number Date Comment
B43IB20A 10.04.2003 First release
B43IB20B 22.10.2003 Minor changes
B43IB20C 23.06.2004 Change the default settings of the
optional LED Display
Appendix A: Service/Maintenance Report
84 B43IB20
Appendix A: Service/Maintenance Report
see next pages:
Appendix A: Service/Maintenance Report
B43IB20 85
COBRIX3 Preventive Maintenance Protocol
Maintenance protocol
COBRIX serial number
Maintenance contract No.
Field of operation
Temperature range
Bypass pump
Type S/no.
Density/velocity of sound transducer
DPRn 427
DSRn 427
Serial number
CO
2
transducer
CARBO 2100
Serial number of transducer
COBRIX evaluation unit mPDS 2000 V
Serial number
Application program
Company:
Name:
Address:
Post code
City:
COBRIX
Anton Paar GmbH
Anton-Paar-Str. 20, A-8054 Graz, Austria Europe
Tel: +43 316 257-0, Fax: +43 316 257-257
E-mail: info@anton-paar.com
Instruments for: - Density and concentration measurement
- Rheometry and viscometry
- High-precision temperature measurement
- X-ray structure analysis
- Microhardness testing
- Sample preparation
- CO2 measurement
- Colloid science
Web: www.anton-paar.com
B43IE31B-E-MaintenanceProtocol-Cobrix3.doc Page 1 of 5
Appendix A: Service/Maintenance Report
86 B43IB20
COBRIX3 Preventive Maintenance Protocol
Cobrix
COBRIX
Accessories (connected to the Anton Paar COBRIX evaluation unit):
Flow monitor Filler stopped signal
Bottle counter 3# LED displays
Product selector Davis Software
Version
Additional measurements Type
Control functions Type
Regulating functions Type
External condition of the instrument before maintenance:
COBRIX housing and rack: OK not OK cleaned
CO
2
transducer: OK not OK cleaned
DPR/DSR transducer: OK not OK cleaned
Bypass pump: OK not OK cleaned
Comment:
Checking the configuration / Setting the parameters:
CARBO 2100 settings:
Unit: g/l vol
Prod.: BE SD MW SW CS
Presetting: Carbo-dV Carbo-dT Carbo-dP
Adjustment factors: CO
2
-Offset CO
2
-Gain
COBRIX evaluation unit:
Instrument constants DPR/DSR transducer
Acc. to data sheet from: OK not OK
Comment:
System data saved: yes no not applicable
File name:
After discussing with the customer, the COBRIX measuring system has been separated from the
production line
B43IE31B-E-MaintenanceProtocol-Cobrix3.doc Page 2 of 5
Appendix A: Service/Maintenance Report
B43IB20 87
COBRIX3 Preventive Maintenance Protocol
Cobrix
COBRIX
Checking / maintenance of the Carbo 2100
Checking / cleaning Cable connections Control valve for compressed air
Pneumatic connections Pneumatic valves Magnetic valves
Stirrer Temperature sensor Pressure sensor
Functional test
Checking / maintenance of the COBRIX evaluation unit
Visual check OK not OK
Contacts maintenance carried
out
yes no
Buffer battery OK not OK not measurable VDC
Digital inlets OK not OK not applicable
Analog inlets OK not OK not applicable
Digital outlets OK not OK not applicable
Analog outlets OK not OK not applicable
Counter inlets OK not OK not applicable
Relay outlets OK not OK not applicable
DPR/DSR inlet OK not OK not applicable
Product selection OK not OK not applicable
Checking / maintenance of the DPRn/DSRn transducer
Power supply voltage OK not OK VDC
Visual check of measuring electronics OK not OK
Seal on electronic housing OK not OK replaced
Checking / maintenance of the bypass pump
Mechanical seal replaced yes no
Functional test OK not OK
B43IE31B-E-MaintenanceProtocol-Cobrix3.doc Page 3 of 5
Appendix A: Service/Maintenance Report
88 B43IB20
COBRIX3 Preventive Maintenance Protocol
Cobrix
COBRIX
The customer was informed that the transducer has been separated from the production line for the
maintenance procedures and can only be operated again after successful cleaning (CIP):
yes no
Checking the Brix/Diet adjustment
Current product:
Reference method
Brix/Diet conc. reference Temp. C
Brix/Diet conc. COBRIX Temp. C
Deviation
Checking the CO
2
adjustment
Current product
Reference method
CO
2
conc. reference
CO
2
conc. Carbo P-Line bar P-Gas bar Temp. C
CO
2
deviation
Other maintenance carried out / comments:
Replaced parts for CARBO 2100:
PTFE pcs. 6924 Stirrer Igus/Two Wing pcs. 12745
Stirrer motor Igus pcs. 10745 valve Hubbegr. pcs. 6923
Absolute pressure sensor pcs. 6522 valve EPDM pcs. 12758
Temperature sensor pcs. 6521 EPDM pcs. 12759
Interface card new pcs. 9116 EPDM pcs. 13881
Magnetic valve pcs. 6525 EPDM pcs. 16772
B43IE31B-E-MaintenanceProtocol-Cobrix3.doc Page 4 of 5
Appendix A: Service/Maintenance Report
B43IB20 89
COBRIX3 Preventive Maintenance Protocol
Cobrix
COBRIX
Replaced parts (DPR/DSR/mPDS...):
Seal 120x220 pcs. 6812 O-ring 290x6 FPM 80 pcs. 54241
Coil A43 pcs. 60691 Battery mPDS4000 pcs. 59746
Liquid sensor capacitive pcs.76435
Replaced parts for the bypass pump NPY-2051 ( Speck pump):
Mechanical seal pcs. 77953 O-Ring for Speck pump pcs. 77954
Impeller pcs. 77957
Other replaced parts:
For the service engineer:
Name:
Company:
Date: Signature:
For the customer:
Name:
Company:
Date: Signature:
B43IE31B-E-MaintenanceProtocol-Cobrix3.doc Page 5 of 5
Appendix B: Feedback Form
90 B43IB20
Appendix B: Feedback Form
Please help us improve this service manual and our service to you!
If you have any suggestions, comments or problems concerning the contents of
this service manual, please do not hesitate to contact us:
E-mail: info@anton-paar.com or copy and fax this form to: +43-316-257-257
Service manual for instrument type:...................................................................
Name and address: .......................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
The service manual should contain more information on:
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
The following sections are difficult to understand:
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
Other comments:
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
...............................................................................................................................
.............................................................................................................................
Thank you for your feedback!

S-ar putea să vă placă și