Information sheet
2. Overvoltage/installation categories
DVMs are categorized depending on the risk and severity of transient overvoltage that might occur at the point of test. Transients are short-lived burst of energy induced in a system, e.g. caused by lightning strike on a power line. On high-energy circuits this leads to hazardous situations as these circuits might provide enough current to feed an arc-over resulting in a plasma breakdown or even explosion. Higher CAT-identification refers to electrical environments with higher power and possibly higher energy transients.
A CAT I-rated meter is suitable for measurements on protected electronic circuits that are not directly connected to mains power, e.g. electronics circuits, control signals
A CAT II-rated meter is suitable for measurements in CAT I-environments and mono-phase appliances that are connected to the mains by means of a plug and circuits in a normal domestic environment, provided that the circuit is at least 10m apart from a CAT III- or 20m apart from a CAT IVenvironment. E.g. household appliances, portable tools
A CAT III-rated meter is suitable for measurements in CAT I- and CAT IIenvironments, as well as for measurements on (fixed) mono- or poly-phased appliances which are at least 10m apart from of a CAT IVenvironment, and for measurements in or on distribution level equipment (fuse boxes, lighting circuits, electric ovens).
A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and CAT IIIenvironments as well as on the primary supply level. Note that for all measurements on equipment for which the supply cables run outdoors (either overhead or underground) a CAT IV meter must be used.
Velleman
2/3
4. Gen neral
W Warnin g - Ca aution n
Disconnect the D t test lea ads from the test point ts before ch hanging the e position of o the fu unction (rot tary) switch h. D Disconnect t the test lea ads from the test point ts and remove the tes st leads from the m measuring terminals be efore replac cing the ba atteries/fuse es, or servicing the DV VM. W WARNING: To avoid electrical e sh hock alway ys disconne ect the test t leads prior r to opening the housing. To T prevent damage d or injury, only use batte eries and fu uses w with the sam me ratings as a specified d in this ma anual. Never conne ect a voltag ge source with w the fun nction (rotary) switch in / / / A / mA / A / Hz / Hf fe or C/F position. C Caution: ris sk of electro oshock whe en measurin ng voltages > 36VDC, 25VAC, cur rrents > 10mA, AC C power line es with indu uctive loads and AC power lines w with fluctuat ting po ower. A Avoid body contact wit th ground potential p (e. .g. metallic terminals, output soc ckets, le ead clamp ) while mea asuring. Ma ake sure to be electrica ally insulate ed from gro ound during meas surement. A Always use the device within its specified s range. O Only use the e included test t leads. When W dama aged, replace them wi ith test lead ds of th he same typ pe and with h the same specificatio ons. Calibration and a repair must m be per rformed by a qualified technician. . Refer to your y ocal dealer. lo D not expo Do ose the DVM M to extrem me tempera ature or hum midity levels.