Temperature calibrators for

the servicing sector


Series TP 18000 and TP 18000 S
Temperature calibrators, Series TP 18 000 and TP 18 000 S
The professional solution for the servicing sector
Exact temperature measurement and
monitoring are "musts" in applications
crucial to operational safety of machi-
nery, ships and industrial installations.
Regul ar i nspecti on of the temperature
sensors used i n these appl i cati ons i s
absol utel y essenti al for economi c and
techni cal -safety reasons and i s al rea-
dy prescri bed as obl i gatory i n many
sectors.
The servicing temperature cali-
brators, type-tested by Ger mani scher
LLoyd (GL), are approved for use on
sea-going ships and inland waterway
vessel s and i n the offshore area.
Description:
The calibrators of series TP 18000 and TP 18000 S operate with an elec-
tronically controlled, heatabl e and cool abl e aluminium or brass block (de-
pendent on temperature range).
The requi red temperature i s set usi ng a decade swi tch. Preci se adj ust-
ment is performed using a rotary knob.
The bl ock temperature i s di spl ayed on the 3 1 / 2 digit digital display. The
haeting block features a bore hole (diameter on depth, see adjacent table)
servi ng to accommodate the test speci men. Transi ti on sl eeves are used f or
test specimens with smaller diameter.
The right unit for any temperature range:
GL No. 11297-98 HH
Servicing temperature
calibrators for applications
in the:
The temperature calibrators of
series TP 18000 and TP 18000 S, de-
signed for "in-situ inspection", are al -
ready a part of the standard
equi pment of the servi ng t echni can i n
the above l i sted today.
These compact devi ces are easy to
transport and easy to operate and
have all performance features requi -
red for "in-situ inspection".
Servicing calibrators for inspection of:
Thermometers/ machine thermometers
Inspecti on i s performed by compari son of the temperature measured by
the test specimen and the block temperature indicated by the unit.
Temperature switches/thermostats
The test specimen is connected to the two terminal sockets of the unit.
The switch setting respective to the switch point is signaled by the check
indicator.
Resistance thermometers and thermocouples
A separate temperature-measuri ng i nstrument i s requi red for i nspecti on.
We recommend the use of our 4
1
/2-digit temperature-measuring devices of
Series TT 7200 and TT 31048. The inspection is performed by comparison
of the temperature indicated on the external measuring instrument with the
bl ock temperature.
Shipping industry
Energy-production and energy
distribution sector
Chemical and petrochemical in-
dustry
Pharmaceutical industry
Foodstuffs industry
And a great deal more
Technical data, temperature calibrators TP 18000 and TP 000 S
Device type TP 18150 TP 18200 TP 18600 TP 18850
Temperature range -30C to+150C Ambient temperature up
to +200C
Ambient temperature
up to +600C
Ambient temperature up
to +850C
Block temp.setting with decade switch
Block temp, tolerance 1 C
Heating block
control/reference sensor
PID-controller/ Pt 100
Display
Block temp, display 3
1
/2-digit, 7-segment
LED, 20 mm high
3
1
/2-digit, 7-segment
LED,11 mm high
3
1
/2-digit, 7-segment
LED, 20 mm high
3
1
/2-digit, 7-segment
LED, 20 mm high
1 Display range -30C to +150C 0 C to +200 C 0 C to +600 C 0 C to +850 C
Resolution 1C
Test specimen fixture
Heating block material Aluminum Aluminum Brass HT- alloy
18 18 18 (Option 28) 18 (Option 28)
Inspection bore hole diameter
Insertion depth
100 100 100 (Option 200) 100 (Option 200)
Adapter sleeves Inside diameter: 1,5 to 15 mm (25 mm) in 0,5- steps
General data
Power supply 230 V, 50/60 Hz (Option 115V, 50/60 Hz)
Power consumption ca. 150VA ca. 220 VA ca. 1300VA ca. 2000 VA
Dimensions 288x1 44x300 mm 1 92 x 96 x 200 mm 288 x 1 44 x 300 mm 384 x 192 x 350 mm
Weight ca. 1 0 kg ca. 3,3 kg ca. 9 kg ca. 14,5kg
Device type TP 18150 S TP 18200 S TP 18600 S
Temperature range -30C to150C Ambient temp, to +200C Ambient temp, to +600C
Block temp, display resolution 1/1 0C 1/10 C 1/10 C to 199,9C
1C to600C
Block temp, tolerance 0,2C 0,3C 0,3C
All other technical data corresponds to that of Series TP 1 8 000
Heating and cooling times
Device Type Heating time* (min) Cooling time* (min)
TP 18150 / TP 18 150 S 20C to 80C: 3
80C to 150C: 10
+20C to 0C: 2
0C to-30C: 12
TP 18 200 / TP 18 200 S 20C to 100C: 9
100C to 200C: 21
200C to120C: 36
120C to90C: 42
TP18600 / TP18600S 20C to 300C: 9
300C to 600C: 18
600C to 300C: 25
300C to 100C: 47
TP 18850 20C to 400C: 11
400C to 850C: 38
850C to 400C: 31
400C to100C: 69
* Also allow for an additional settling time of approx. 5 min.
ICS SCHNEIDER MESSTECHNIK GmbH Tel.: +49 3303 504066
Briesestrae 59 Fax: +49 3303 504068
16562 Bergfelde e-mail: info@ics-schneider.de