Sunteți pe pagina 1din 20

Homemade Spectrophotometer

Aaron Russell 10 February 2014 Introduction Spectrophotometry is defined by nist.gov as the quantitative measurement of the absorption of a material as a function of a wavelength. Spectrometric analysis makes use of selected wavelengths of electromagnetic radiation for the qualitative or quantitative determination of a sample analysis. There are quite a few uses of spectrophotometry. This

science actually allows scientists to identify a materials identity or concentration without having to touch it. Spectrophotometry can be used to identify gases and many other materials as long as light is able to pass through the material. Spectrophotometry is used in the medical field to determine the seriousness of heart attacks, and it can also be used to identify different types of pigments in plants. A spectrometer can cost hundreds to thousands of dollars, but a much more basic version can be built. I will be assembling this basic home-made version.

Background Research I started off trying to find a science project by for advanced science fair projects on google.com. I found a plethora of ideas, but nothing sparked my interest. I found the idea for building a spectrophotometer on sciencebuddies.com in the Chemistry Section. I decided to do

Russell 2

this experiment because it seemed to be exciting and interesting. This experiment was inexpensive and Science Buddies provided adequate information on the project. This website provided me with a general idea of what I was going to build (Science Buddies, 2013). As I researched this topic I looked up the definition of a spectrophotometry on the internet. The definition of a spectrophotometry is the quantitative measurement of the absorption of a material as a function of a wavelength (NIST, 2013). I had a decent grasp on visible light and the electromagnetic spectrum because Mrs. Pearson had taught it to me. I knew that white light consisted of all the colors of the rainbow. I found out that a spectrophotometer analyzed this spectrum by measuring the intensity of light at different wavelengths. A wavelength is the distance that a wave will travel in one cycle. (Science Buddies, 2013) Variables I only had two types of variables: independent and dependent. My independent variables were the color of the solutions and the amount of dye in the solution. My dependent variables were the spectra. Hypothesis My hypothesis is very simple. My hypothesis is: A basic, inexpensive spectrometer can be built. Materials In order to start assembling the spectrophotometer, a couple of special items needed to be purchased. I ordered a linear 500 mm/lines diffraction gratings slide on Amazon.com, a 5

Russell 3

mm white LED light (part #RL5-W5020) on superbrightleds.com, and 20 plastic cuvettes, 1 cm, on ebay.com from the seller madlaboratory. All the other items are typical household items. List of Materials: Wood (Red Oak, x 3 planks) iPhone 4 (CMOS Camera) Canon EOS Rebel T3i (DSLR Camera) HP Computer (Windows Based) Cuvettes 1 cm Pathlength (Plastic) Food Coloring (Red, Green, Blue) LED light (diode, part #RL5-W5020) Battery, CR2032 3V Coin Darkened Room (For performing the experiments) Soldering Iron - Weller 25 watt (SP23L) by Cooper Tools Solder - Alphametals Resin Care (31604) 10 inch Miter saw (Craftsman) Toggle switch Fuse Holder (15 amp)

Russell 4

Crimp-on electrical connectors 16 gauge stranded electrical wire Plastic clamps Wood Screws Paint (white and black) Plywood (for the spectrometer base) Assembling the Temporary Spectrophotometer Two versions of the spectrophotometer were assembled. The first version was temporary and only used for the testing of the purchased materials. The second version was permanent and used for the final tests. The temporary version consisted of the diffraction gratings slide, LED light, the CR2032 battery, a cuvette filled with clear distilled water, a book, electrical tape, scotch tape, and a cell phone camera. The LED light has two prongs attached to it: a long one and a slightly shorter one. The longer prong was touched to the positive side of the battery, and the shorter prong touched the negative side. The electrical tape was used to provide a steady stream of light. This light contraption was elevated with a book. The diffraction gratings slide was then setup around 30 cm away from the light and at an angle. This was to try to limit the amount of exposure when taking pictures. The cuvette was then filled and images were taken with an iPhone 4.

Russell 5

This temporary version simply served as a baseline to determine the feasibility of the experiment. After a few photos were taken with adequate results, I decided to proceed to build a permanent version for final testing based upon on these results that I obtained from this temporary setup. I learned that I was going to need to find a way to accurately place each of the components on the base. I further realized that once the components were placed that I needed to make other changes due to the camera selection. Because of this, I made each component on a separate movable arrangement. The alignment of the components needed to be finalized. Assembling the Final Spectrophotometer The final version of the spectrophotometer was built as components and then assembled into the final test instrument. The components built were: a base, source, cuvette holder, diffraction grating holder, a slit plate. The instrument was built as components to allow for movement and placement of the individual components in order to achieve a functioning spectrophotometer. Alignment of the components was critical. Many trials were conducted to determine the best alignment. The Base was built using plywood and red oak planks. The base is 23.75 inches deep x 30.5 inches wide x 4 inches tall. The base was painted black. The Source was built using a LED light, a toggle switch, a battery, wire, and connectors. An electrical schematic is shown below. (Electronics Club, 2013) LED stand for light emitting diode. LEDs are quickly becoming a popular source of light and are replacing incandescent, halogen, and compact fluorescent bulbs. (ACDC LED Lights, 2012) A toggle switch was installed

Russell 6

to allow easy turning on and off of the LED. The fuse was installed for safety purposes. The battery is 3V. It was recommended by Science Buddies. A soldering iron was used to connect the wire to the LED, and electrical connectors were used for every other connection. One thing I noticed was that the LED will only work if the battery is if the battery is in the correct orientation. The positive side of the battery must be connected to the positive terminal of the LED or the circuit will not function.

Russell 7

The Cuvette Holder was built using red oak planks and a miter saw. The cuvette holder was made in a manner so that each time a new cuvette was placed it would be in the same orientation. This is important because the results could be affected by improper placement of the cuvette. A picture is below.

The Diffraction Grating Holder was built using Red Oak planks and a miter saw. A picture is below.

The Slit Plate was built using Red Oak planks and a miter saw. It was built in a manner that it was able to be moved around so the light could be focused. A picture is shown below.

Russell 8

In general, the spectrophotometer begins with a light source (LED). The light is focused into a beam by the tubing place over the LED bulb. This keeps any stray light from contacting the camera. The light then passes through the cuvette. After the light exits the cuvette it passes through the slit plate to further focus the light beam. The light then passes through the diffraction grating to disperse the white light into its spectrum. photographed by the detector to produce an image. The Spectrophotometer Software Program The Scheeline-Kelly cell phone spectrophotometer, developed by Alexander Scheeline and Kathleen Kelly, analyzes JPEG images from CMOS cameras which are very common in modern cell phones and tablets. The software uses reference and sample spectra, chosen by the user. The author took into consideration that no two identical frames could be taken and created an algorithm to linearly interpolate the intensity detected in a samples wavelength applicable to the wavelengths of the intensity detected with the sample This spectrum is then

absent but with solvent present. He included this in the transmittance computations. (Cell Phone Spectrophotometer, 2013)

Russell 9

This software has been used a total of three other times according to the author: a group of analysis students at the Faculty of Chemistry, Hanoi University of Science, a group of high school teachers from Illinois, and 26 high school stu dents attending a summer outreach program at Clark Atlanta University. He states that the classroom dynamics were the same in each of the reported cases. He also states that these cases did not involve controlled testing of the students. (Cell Phone Spectrometer, 2013) This software allows for a very inexpensive way to analyze samples. The author stated that one of the objectives of this software was to provide an inexpensive way to teach students the basics of spectrophotometry. (Cell Phone Spectrophotometer, 2013) I believe that this software is a very useful tool. Below is a screenshot from the software showing the sample scan on the left and the reference scan on the right. The resultant spectra is calculated and displayed on the screen in a variety of formats chosen by the user.

Russell 10

Optical Density / Beers Law / Key Terms The measurement principle of this spectrophotometer is known as absorption spectrophotometry. There are several key terms to understand such as transmittance and absorbance. Both of these key terms are based upon measurements from both a reference and a sample. Transmittance can be defined as:

Transmittance = T = P *100 P0 Where P = Intensity of the light passing through the Sample P0 = Intensity of the light passing through the Reference If no light is absorbed by the sample relative to the blank, the %T = 100%. If all the light is absorbed then %T = 0%. Another useful term is absorbance. Absorbance can be defined as: Absorbance = A = log P0 = log 100 P %T Where P = Intensity of the light passing through the Sample P0 = Intensity of the light passing through the Reference Absorbance is important because its variation is directly proportional to the concentration of the substance. It can also be referred to as optical density. The Beer-Lambert law, commonly known as Beers Law, is important in spectrometry. Beers Law is listed below: A=a*b*C Where a = Absorptivity b = particular wavelength and constant C = Concentration (Manahan, 1986)

Russell 11

Diffraction Grating A diffraction grating is a simple monochromator. It turns the incident polychromatic (white light) into its monochromatic components. The diffraction grating is the most common device for this purpose. A prism does the same thing; changes white light into its component colored wavelength. (Manahan, 1986)

As you can see from the two pictures above, a prism and a diffraction grating generally do the same thing. The difference between the two is accuracy. A prism bends light according

Russell 12

to refractive index and also absorbs some of the light. A diffraction grating bends light according to geometry. The bending is also much larger. The larger the bend, the more accurate your readings will be.

(Medway Optics LTD, 2006) Procedures and Data Test #1 The purpose of this test was to find out if the spectrometer and spectra analyzing software would work properly. I started this test by taking four 1cm cuvettes and filling them up with distilled water. I used distilled water to limit the amount of imperfections in the samples. I left one sample with just clear, distilled water. This would be my reference. The other three cuvettes were filled with two drops of different food color. These colors included red,

Russell 13

green, and blue. These would be my samples. I started off by testing my reference. The full spectrum was shown. I attempted to use the iPhone 4 as a camera, but the quality of it was not sufficient for accurate testing. I instead used a Canon EOS Rebel T3i. This camera still generates JPEG images making the images applicable to the software. The Rebel provided a much better image quality. Testing then resumed. A picture of the reference was then taken. Each sample was tested the same as the reference. The red sample only showed red on the spectrum. The blue sample only showed blue on the spectrum. The green sample only showed green on the spectrum. These results tell me that the spectrometer and diffraction grating worked properly. These images were placed into the spectra interpretation software. The software was very difficult to use and did not provide accurate results. The graphs that were computed from the software were saved. Conclusions were drawn from the results.

Reference

Blue Sample

Russell 14

Green Sample

Red Sample

Red Sample vs Reference

Russell 15

Blue Sample vs Reference

Green Sample vs Reference

You can conclude from these results that the spectrometer and diffraction grating worked well, but the spectra interpretation software was not very accurate. You could further explore the possibilities of the software in a future test, but I do not believe that the program is even functional. It is difficult to use and does not provide accurate results. This test does not disprove my hypothesis.

Russell 16

Test #2 The purpose of this test was to show how different intensities of a color can change the intensity of a spectrum. I started off by taking another picture of the reference spectrum. This was to ensure that this test would maintain the same state throughout. The reference again showed the complete spectrum. I then took a large container and filled it with distilled water. I added two drops of red food coloring to the container of distilled water. This diluted the solution and made a very light red. I then filled my cuvette with this solution. A photo was taken of the samples spectrum. The intensity of the samples spectrum was less than the intensity of the references spectrum. I then added two more drops of red food coloring to the cuvette directly to make a normal red sample. A photo was also taken of this samples spectrum. I could see a distinctive difference within the intensity of this spectrum compared to the references spectrum. Next I added even more red food coloring to the solution, which created a dark red solution. The intensity of the spectrum decreased even further. The spectrum started to show mainly red and was not very bright. I ended the experiment by adding a few more drops of red food coloring which made the red extremely dark. The intensity decreased even more. This test again proved that the spectrometer was quite functional. Pictures of the spectra are shown below.

Russell 17

Reference

Very Light Red

Normal Red

Dark Red

Very Dark Red

Extremely Dark Red

You can see from the images that the spectrometer proved to be functional once again. The darker the color of the sample the less intense the spectrum became. The software was unnecessary for this experiment. You could provide exact ratios of the amount of food coloring compared to the amount of water in a future experiment, but this did not seem necessary for a basic intensity test. This experiment did not disprove my hypothesis.

Russell 18

Conclusion The spectrometer worked great. The reference scan showed the entire spectrum of colors. The blue only showed blue. The green only showed green. The red showed only red. The intensity of the spectra changed as the samples became darker. The diffraction grating did its job effectively. However, the spectra interpretation software proved inaccurate. The graphs were inaccurate. The software was abstruse and not easy to use. Since the hypothesis stated that a basic, inexpensive spectrometer can be built, the hypothesis was not wrong. This project was successful.

Russell 19

Works Cited 1. "Circuit Symbols." Electronics Club. N.p., n.d. 15 Dec. 2013. <http://electronicsclub.info/circuitsymbols.htm>. 2. Manahan, Stanley E. Quantitative Chemical Analysis. Monterey, CA: Brooks/Cole Pub., 1986. Print. 3. Scheeline, Alexander, and Kathleen Kelley. "Cell Phone Spectrophotometer." Cell Phone Spectrometer. N.p., n.d. 15 Dec. 2013. <http://scheeline.scs.illinois.edu/~asweb/CPS/>. 4. "See the Light by Making a Cell Phone Spectrophotometer." See the Light by Making a Cell Phone Spectrophotometer. N.p., n.d. 15 Dec. 2013. <http://www.sciencebuddies.org/science-fairprojects/project_ideas/Chem_p100.shtml>. 5. "Spectrophotometry." NIST. N.p., 13 Nov. 2009. 15 Dec. 2013. <http://www.nist.gov/pml/div685/grp03/spectrophotometry.cfm>. 6. Vodopich, Darrell, and Randy Moore. "The Importance of Spectrophotometry." Online Learning Center. McGraw Hill, n.d. Web. 15 Dec. 2013. <http://highered.mcgrawhill.com/sites/0073031216/student_view0/exercise7/spectrophotometry_importance. html>.

Russell 20

S-ar putea să vă placă și