Sunteți pe pagina 1din 7

Hardware Integrity Validation (C76-C85) Rev-01 Page 1 of 7

ON-SITE REPAIR INSTRUCTIONS


Hardware Integrity Validation
Product Calisto, Calisto 2
S/Ns C76 C85 (10620 and up)
NOTICE:
The following document contains six sections, 1.0 Electronic Board Verification, 2.0
Power Supply Source Validation, 3.0 Power Supply Input Validation, 4.0 Power Switch
Validation, 5.0 Thermal Fuse Validation and 6.0 Power Supply Validation.
TIME REQUIRED: - 15 minutes
TOOLS NEEDED: - 3/8" flat screwdriver
- Multimeter measuring up to 300VAC and up to 25 VDC
CAUTI ON
Please follow the procedure carefully to prevent any accidents. The monitor
must be powered on to determine any malfunction. For this reason, electrical
shocks are highly possible.
Make sure the door is closed when you are not working directly on the
monitor.
Make sure the electrical tools and equipment you use are properly insulated.
To prevent yourself from being the unintentional path to ground, make sure
the monitor is well grounded.
Hardware Integrity Validation (C76-C85) Rev-01 Page 2 of 7
1.0 ELECTRONIC BOARD VERIFICATION
NOTICE:
At this step a validation that the Calisto is working properly will be made.
Check the status of the LEDs on the electronic board assembly.
LED ACTIVITY
Power LED Always ON
Cell I/O LED Always ON
CPU LED Blinking
If the LEDs do not behave as described above, please follow the steps in the next sections.
Power
LED
(13.2V)
CPU
LED
CELL I/O
LED
Hardware Integrity Validation (C76-C85) Rev-01 Page 3 of 7
2.0 POWER SUPPLY SOURCE VALIDATION
NOTICE:
At this step a validation that the Calisto receives enough power and also that the power supply
is working properly will be made.
Please refer to the figure below for the correct installation of the power cord. Note that, besides AC
NEUTRAL and AC LIVE, a third feed line must be provided, the EARTHline.
Using a multimeter set on voltage measurement, measure the voltage between pin 1 and pin 2 of the
TB2 connector, see image above. The voltage must be in a range of 100 V AC and 240 V AC.
V
Hardware Integrity Validation (C76-C85) Rev-01 Page 4 of 7
3.0 POWER SWITCH VALIDATION
NOTICE:
At this step the integrity of the rocker switch will be validated.
1. Power on Calisto (if not already done).
2. Be sure that the main fuse is functional, see Section 3.0.
3. Measure the voltage between pin 3 of TB7 connector and pin 1 of TB3 connector then
measure the voltage between pin 4 of TB7 connector and pin 2 of TB3 connector.
Both voltages must be in a range of 100 V AC and 240 V AC. If its not the case then please, contact
Morgan Schaffer at support@morganschaffer.com with the code Rocker switch replacement.
V
V
Hardware Integrity Validation (C76-C85) Rev-01 Page 5 of 7
4.0 THERMAL FUSE VALIDATION
NOTICE:
At this step the integrity of the thermal fuse will be validated.
1. Power on Calisto (if not already done).
2. Check the continuity between pin 1 and pin 2 of the TB7 connector. See figure below:
If there is no continuity then please, contact Morgan Schaffer at support@morganschaffer.com
with the code Thermal fuse replacement.
Hardware Integrity Validation (C76-C85) Rev-01 Page 6 of 7
5.0 POWER SUPPLY INPUT VALIDATION
NOTICE:
At this step the integrity of the power supply input will be validated. The goal is to test the
functionality of the main fuse.
1. Power on Calisto (if not already done).
2. Measure the voltage between pin 1 and pin 2 of the TB8 connector. See figure below:
If there is no voltage then please, contact Morgan Schaffer at support@morganschaffer.com with
the code main fuse replacement.
V
Hardware Integrity Validation (C76-C85) Rev-01 Page 7 of 7
6.0 POWER SUPPLY VALIDATION
NOTICE:
At this step the integrity of the internal Power Supply will be validated.
1. Power on Calisto (if not already done).
2. Using a multimeter set on V DC, measure the voltage between the two pins of the J11 connector
(see figure below). The voltage should be between 13.1 and 13.25 VDC and steady.
If the voltage is under 12 V, or is not steady, it means that the power supply is no longer valid
and in this case contact Morgan Schaffer at support@morganschaffer.com the code Power
supply replacement.
J11 Connector
Position 2 on J11
Position 1 on J11

S-ar putea să vă placă și