33%(3)33% au considerat acest document util (3 voturi)
672 vizualizări130 pagini
All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates.
All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates.
All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates.
of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Print No.: SiemensLandgraf AXIOM Artis FC/FA System Planning Guide AX 2001 AXA4-100.891.01.06.02 Replaces: AXA4-100.891.01.05.02 English Doc. Gen. Date: 01.08 AXIOM Artis FC/dFC/FA/dFA (e) 05904433 05904441 07412807 07555373 2 Revision / Disclaimer AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 2 of 130 1Revision / Disclaimer Document revision level The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed. Please contact your local Siemens office to order current revision levels. Disclaimer The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise autho- rized by Siemens or one of its affiliates to provide such services. Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service proce- dures. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Table of Contents 3 Page 3 of 130
Top-down View of System with MULTISPACE.F Option 0 The minimum room width increases due to the two additional working positions (*1/Fig. 8 / p. 14). *1 Dimensions without brackets Table rotation with the tabletop fully extended is not restricted. Dimensions in brackets Table rotation with the tabletop fully extended is restricted. *2 Dimensions without brackets for Peri, Tilt and OR tables (including 500 mm safety distance to the wall) Dimensions in brackets for the basic table 14 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 14 of 130 Fig. 8: Top-down view of system with MULTISPACE.F option Pos. 1 Head-end/left position rotation range (motorized movement) Pos. 2 Floor stand in the left position. Pos. 3 Orientation point/isocenter for floor; stand in the head-end position Pos. 4 Isocenter in the left position Pos. 5 Orientation point / floor fixpoint; patient table *1 Dimensions without brackets Table rotation with the tabletop fully extended is not restricted. Dimensions in brackets Table rotation with the tabletop fully extended is restricted. *2 Dimensions without brackets for Peri, Tilt and OR tables (including 500 mm safety distance to the wall) Dimensions in brackets for the basic table Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 15 Page 15 of 130 NOTE If the room size permits, the DCS should be shifted one ceil- ing grid to the left to enable unimpeded turning behind the table.
16 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 16 of 130 System with a second DCS Option 2.3 In addition to the existing DCS (1st DCS), in the future it will be possible to operate a sec- ond, optional DCS. Note the following for planning: For new plan- ning: The 1st DCS (max. 8 TFT monitors) should be planned longitudinally to the table. The 2nd DCS (max. 3 TFT monitors) should be planned transversely to the table. For a retrofit: The 1st DCS (max. 8 TFT monitors), if installed transversely to the table, should be planned longitudinally to the table. If already installed longitudinally, nothing changes. The 2nd DCS (max. 3 TFT monitors) should be planned transversely to the table, as with new planning. The following applies: 1. DCS longitudinally, 2nd DCS transversely to the table Limitation for 2nd DCS: A max. of only 3 monitors can be installed. Bypass images cannot be displayed and data display is not released. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 17 Page 17 of 130 Top-down View of System with a Second DCS 2.4 Fig. 9: Top-down view of system with second DCS Pos. 1 Cable fixpoint, 1st DCS Pos. 2 Cable fixpoint, 2nd DCS Pos. 3 Orientation point/isocenter for floor; stand in the head-end position Pos. 4 Isocenter in the left position Pos. 5 Orientation point / floor fixpoint; patient table For ceiling installation with a second DCS, see Detail Planning > Ceiling Installation with a second DCS. *1 Dimensions without brackets Table rotation with the tabletop fully extended is not restricted. Dimensions in brackets Table rotation with the tabletop fully extended is restricted. 18 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 18 of 130 Additional Work Positions with the MULTISPACE.F Option 0 This option enables manual rotation of the floor stand around the axis of rotation (see below). For medical applications, there are two new work positions, the right position and OR position. The option can be retrofitted. Fig. 10: Multispace.F Pos. A System in the right position Pos. B System in the OR position Pos. 1 Floor stand rotation axis *2 Dimensions without brackets for Peri, Tilt and OR tables (including 500 mm safety dis- tance to the wall) Dimensions in brackets for the basic table Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 19 Page 19 of 130 System: DCS Positioning 2.5 Fig. 11: DCS positioning Pos. A Version 1: Standard in combination with the MULTISPACE.F option Pos. B Version 2: only without the MULTISPACE.F option Variant 1 (standard in conjunction with the MULTISPACE.F option): For the room plan, see Room Planning > System with the DCS Option. For ceiling installation, see Installation Preparations > Ceiling Installation. Variant 2 (without MULTISPACE.F option only): For the room plan, see Detail Planning > DCS Positioning per Variant 2. For ceiling installation, see Detail Planning > System Ceiling Installation. 20 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 20 of 130 System: Side View 2.6 Fig. 12: System side view Pos. 1 Max. unit height *1 Orientation point/isocenter for floor stand in the head-end position *2 Admissible room height: DCS 1/2/3: 2500-3100 mm DCS 4/6/7: 2900-3100 mm D1 DCS (Option) (ceiling fixpoint) P1 Floor stand (ceiling fixpoint) T1 Patient table (floor fixpoint) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 21 Page 21 of 130 Planning Notes 2.7 Movement Ranges of the Components 0 When planning non-system components in the examination room, pay attention to the movement ranges of the system components (risk of collision!). Fig. 13: Movement ranges of the components Pos. 1 Floor stand movement range Pos. 2 DCS movement range Pos. 3 Patient table movement range (foot end) Pos. 4 Patient table movement range (head end) 22 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 22 of 130 NOTE If the room size permits, the DCS should be shifted one ceil- ing grid to the left to enable unimpeded turning behind the table.
Max. admissible Floor Unevenness 0 The maximum admissible floor unevenness of the base floor in the cross-hatched area is 8 mm. The cross-hatched area represents the rotation area of the stand and the movement range of the C-Arm. Fig. 14: Max. admissible floor unevenness Pos. 1 Wall Pos. 2 Orientation point/isocenter for floor stand in the head-end position *1 The MULTISPACE.F option enables manual turning of the stand about rotation point A. Within the turning range, the stand can lock in at 6 increments; motorized device movements can then be performed as long as the distance to the wall is at least 500 mm. Entering the room size into the system configuration during system start-up ensures that motorized movements will be blocked if this minimum dis- tance is not met. The technically possible manual rotation range with the MULTISPACE.F option is +60/-220. With a specified room width of 5930 mm, the rotation range is limited by the room with to -110. For medical applications however, this is of no importance. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 23 Page 23 of 130 *1 In the specified room plan (Fig. 13 / p. 21), the dashed range is intersected by the wall. If the wall extends further to the left, the dashed range has to be taken into account accordingly. 24 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 24 of 130 Floor Stand: Side View 2.8 Fig. 15: Floor stand: side view / A: with 20 x 20 flat panel detector / B: with 30 x 40 flat panel detector Pos. 1 Floor stand in the head-end position. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 25 Page 25 of 130 Floor Stand: Side View 2.9 with Image Intensifier 0 Fig. 16: Floor stand: side view with image intensifier Pos. 1 Floor stand in the head-end position *1 190 - 425 mm with 40 cm I.I. 26 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 26 of 130 Floor Stand: Top-down View of dFC/FC/FA/dFA 2.10 Fig. 17: Floor Stand: Top-down View / A: dFC/FC/FA / B: dFA Pos. 1 Head-end/left position rotation range (motorized movement) Pos. 2 P1 cable fixpoint on the ceiling Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 27 Page 27 of 130 Pos. 3 Orientation point / isocenter, head-end position Pos. 4 Isocenter, left position Pos. 5 Rotation axes For additional manual rotation range with the MULTISPACE.F option, see Room Planning > Planning Notes. 28 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 28 of 130 Patient table 2.11 Versions 0 There are 4 different versions of the Artis patient table that can be shipped: 1. Basic Table 2. Peri table (like the basic table with additional motorized longitudinal movement of the tabletop) 3. Tilting table (like the Peri table with an additional tilting function in longitudinal direction) 4. OR table (like the tilting table with an additional tilting function in transverse direction) The Artis patient tables all have the same exterior dimensions. Notes regarding the OR table 0 An additional control cabinet (SC2) is shipped with the OR table. The cabinet primarily contains the motor controller. Unlike other Artis patient tables, the motor controller can- not be installed in the table base, as this would limit the tilting and canting movement functions.. The OR table is splash-proof (IPX4 according to IEC 60 601-2-46). Use in the operating room If the table is to be used for surgical interventions, according to IEC 60 601-2-46 it also has to be able to operate even in the event of a power failure. In this case, system con- trol cabinet SC1 has to be connected to an on-site uninterruptible power supply (UPS) . Since all system components, with the exception of the generator, are supplied power via the control cabinet, the units, image system, etc. remain operational even in the event of a power failure. For detailed information on the UPS, see PG Axiom Artis emergency power supplies, print no. AXA4-770.891.02... Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 29 Page 29 of 130 Views 0 Fig. 18: Views of patient table 1 = View x 2 = Card tabletop 3 = Basic table / Peri table (no tilting) 30 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 30 of 130 *1 Power outlet (option) See the on-site electrical installation -> on-site power cables *2 Panel for syringe connection *3 Panel with nipple for compressed air connection *4 Accessory rails *5 Longitudinal table tilt (option) *6 Transverse table tilt (option) *7 Table Lift *8 Longitudinal table movement *9 Transverse table movement *10 Table rotation range T1 Patient table, floor fixpoint / orientation point / center of rotation Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 31 Page 31 of 130 Positioning of Cabinets and Components 2.12 Fig. 19: Arrangement of cabinets and components / A=recommendation 1, B=recommendation 2 Pos. 1 Generator cabinet Pos. 2 Cable cabinet (option) Pos. 3 System control cabinet #1 Pos. 4 System control cabinet #2 (only with OR pat. table) Pos. 5 Cooling unit for tube unit Recommendation 1: Generator (PU1), system control cabinets (SC1/SC2), and cooling unit (CU1) are in the equipment room (due to noise and heat generation). Recommendation 2: Generator (PU1) and system control cabinets (SC1/SC2) are in the examination room, cooling unit (CU1) is in the equipment room. NOTE Note the heat and noise generation of the generator (PU1).
Image system (IS): in the control room (standard). It may be placed in the equipment room only if specified in the order. Take note of the heat dissipation and temperature ranges of the cabinets and compo- nents (see Chapter 5). If necessary, air condition the equipment room. Systems with flat panel detector: The maximum permitted height difference between the detector and SC1 control cabinet is 10 m. 32 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 32 of 130 The cabinets can be set up as shown. The following must be observed: - The cabinets may only be positioned apart from one another or at a corner at the locations indicated by the arrows. Here, take note of the System Connections Cable List > Overview of System Connections. - When positioning at a corner, maintain a minimum 40 mm distance to the next cabi- net (necessary to open the cabinet doors). - The PU1 and cable cabinet can also be positioned to the right of the SC1/SC2. - For additional information, see (Generator / p. 91) and Detail Planning (System Con- trol Cabinets #1 / #2 / p. 94). Cable cabinet (option) - The cable cabinet is required if excessive cable lengths cannot be housed in any other manner (e.g., under computer flooring). It has the same dimensions as the gen- erator cabinet. For cable routing from the top left/right, see (System Connections > Notes regarding Laying Cables) into the cabinet base from below and from all sides (except from the front) is possible. - If you require a cable cabinet, mark it in the checklist sent to you by Logistics. The cable cabinet is provided free of charge. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 33 Page 33 of 130 Control Room 2.13 Fig. 20: Control room Pos. 1 Hardcopy Pos. 2 Image system Pos. 3 System control console Pos. 4 Container (if applicable, obtain locally) Pos. 5 Measuring station Pos. 6 Option: Multi-Modality Viewing Container (not shown) Plan the control room at the foot end of the examination room. Operating console and a second display (not shown; Detail Planning > Second Data Display (Option). Control modules (option), ECC (option) - The control modules for the floor stand, collimator, and patient table are not suited for freestanding installation. The modules have to be left in the console. The cutout is to be created on-site by a cabinet maker with a router (for details see Detailed Planning > Tabletop cutout for control modules and ECC) - An emergency Stop button is shipped in conjunction with the control modules (not with the ECC). It has to be installed within reach of the operator. Position components near the control room distributor (see (System Connections) (Control Room Details). Plan the image system in the control room (standard). It may be placed in the equipment room only if specified in the order. For the Multi Modality Viewing Container (option), see PG AXA4-100.891.14... 34 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 34 of 130 Fig. 21: Control room Pos. 1 17" color display for DVD recorder Pos. 2 Remote Power Archive Unit (RPAU) Pos. 3 System Live display Pos. 4 Reference display Pos. 5 Emergency Stop Switch Pos. 6 Image system Pos. 7 UPS Pos. 8 Control room distributor Pos. 9 2nd footswitch Pos. 10 DVD recorder Pos. 11 Container (obtain locally) Pos. 12 Control console Pos. 13 Keyboard + mouse Pos. 14 Control modules Pos. 15 ECC Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 35 Page 35 of 130 For control room distributor details, see Detail Planning > (Control Room Distributor (BTEE) / p. 101). *1 Option *2 Image system in the control room = standard, optionally in the equipment room; must be stated in the order; for details see Detail Planning > Image System. *3 Plan the UPS (option) for the image system in the immediate proximity of the image system. A minimum distance of 25 mm on all sides is required for air circulation. *4 The RPAU is required if the image system is set up outside the control room. *5 Emergency Stop Switch 36 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 36 of 130 Static Requirements 2.14 Floor Stand 0 Fig. 22: Floor stand installation plate Pos. A Installation plate Numbered holes 1-7 for installation directly on the floor; other holes are spare holes. This maximum value does not occur simultaneously at all mounting points. The tensile forces depend on the working position and system movement. The installation plate is anchored in the reinforced concrete with 7 Hilti heavy-duty anchors bolts. Minimum concrete quality: C20/25 Max. tensile force: F1-F7 3.9 kN Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 37 Page 37 of 130 Floor Stand Center of Gravity Fig. 23: Floor stand center of gravity Pos. A Installation plate Point load (stand + installation plate) = 6.6 kN DCS 0 Vertical ceiling load, Fy max : *1 Range of point load center of gravity from head (N) in the Peri position (P). *2 Additional point load movement range with the MULTISPACE.F option (option can be retrofitted) *3 Orientation point/isocenter for floor stand in the head-end position 38 Room Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 38 of 130 Max. 5.4 kN per attachment point( = 2 screws; the screws can be under differing loads, max. 4.2 kN) The dynamic load is exerted on each attachment point, since it is a single moving load. For additional information see PG AXA4-700.891.01... Patient table 0 The maximum extraction forces occur with rotated table (+/- 45) with fully extended table- top at the head-end. The maximum extraction forces were determined with a patient weight of 250 kg and table accessories weighing 40 kg. The installation plate is anchored in the reinforced concrete with 4 Hilti heavy-duty anchors. Minimum concrete quality: C20/25 Foot end load per attachment point: Max. tensile force: 4.5 kN NOTE With the basic table, the max. admissible patient weight is 250 kg. With the other tables, it is 200 kg.
Center of gravity of the patient table The movement of the center of gravity during the rotation of the table was determined under the following conditions: Tabletop completely extended at the head-end. Tabletop extended to the right to the end-stop. Tabletop not tilted. Tabletop loaded with a patient weight of 250 kg. Table accessories weighing 40 kg hooked into the right accessory rail. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Room Planning 39 Page 39 of 130 Fig. 24: Center of gravity of the patient table 1 = Movement of center of gravity 2 = Center of gravity with table rotated 120 3 = Orientation point / center of rotation 4 = Center of gravity with table rotated -120 5 = Center of gravity with table in 0 position *1 A point load of max. 8 kN is exerted on the center of gravity 40 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 40 of 130 3- 3Installation Preparations General Information 0 Required Aids 0 Transport castor set, Part No. 11 53 654 If the system is installed outside of Europe, the transport castor set, if not already avail- able, must be ordered in advance. Installation template, Part No. 65 72 627 G144E for aligning the installation plates of the floor stand and patient table - Within Europe, the template can be borrowed from the Hegele Company. Contact: Tel.: +49 / (0)9191-72 05 610 e-mail: juergen.hoeflinger@hegele.de Spacer gauge, Part No. 70 004 422 X2278, to set the correct distance of the DCS longitudinal rails Floor Stand 0 The pre-installation parts are part of the pre-installation shipment. The pre-installation parts include the installation plate, along with the installation materi- als and the ceiling outlet (Detail Planning > Floor Stand Ceiling Outlet). Preinstallation in Computer Flooring Recommendation: Floor stand installation plate on filled concrete or on the on-site sub- structure. Patient Table 0 The pre-installation parts are part of the pre-installation shipment. The pre-installation parts include the installation plate and the mounting materials. Preinstallation in computer flooring Recommendation: Floor stand installation plate on filled concrete or on the on-site sub- structure. Panels are present on the foot end of the table for (see Room Planning > Views) - Syringe connection - Compressed air connection - Power outlet (option) Injector Floor Outlet Option 0 see Installation Preparations > Injector Floor Outlet Option Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 41 Page 41 of 130 DCS (Option) 0 The pre-installation parts are part of the pre-installation shipment. The pre-installation parts include the longitudinal rails and the installation materials. Transverse mounted ceiling profiles (installation ceiling) are the prerequisite for install- ing the DCS longitudinal rails. Grid spacing for ceiling mounting is max. 675 mm. Control Room Distributor 0 The control room distributor is positioned in the control room, either under the operat- ing console (option) or on the wall. For details, see Detail Planning > Control Room Dis- tributor (BTEE). If available, plan the second footswitch and/or control modules (floor stand, patient table) in the control room such that, when activated, visual and acoustic contact is maintained with personnel, the patient, and the injector. DVD Recorder + 17" Color Display (Option) 0 When setting up the DVD recorder, make sure there is good ventilation. The clear space to walls from stationary fixtures or other components must be at least 3 cm. Reference Display (Option) 0 Located in the DCS (standard) and also possibly in the control room (Room Planning > Control Room). Second Data Display (Option) 0 Positioned in the control room or examination room on the wall or on a console. For details, see Detail Planning > Second Data Display (Option). Control Cabinets 0 For floor and wall anchoring as well as cable routing, see Detail Planning > (Generator / p. 91) and (System Control Cabinets #1 / #2 / p. 94). Store excess cable lengths in the cable cabinet (option), under computer flooring or in cable ducts, etc. (System Connections > Notes regarding Laying Cables). Cooling System 0 For additional information, see the Klver Cooling Unit PG, RA71-020.891.01... 42 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 42 of 130 Lower Body Radiation Shield (Option) 0 The lower body radiation shield is installed on the left or the right of the table. Upper Body Radiation Shield, Surgery Lamp, Injector (Option) 0 1. Overview 2. Upper Body Radiation Shield 2 versions can be shipped: - Upper body radiation shield with fixed installation - Movable upper body radiation shield All mounting parts are included in the shipment. For ordering, see the applicable SIR PM. 3. Surgery Lamp (Option) For electrical connection, see Installation Preparations > On-site Power Cable 4. Injector (Option) There are three ways to install the injector rack mount unit. For details, see Detail Plan- ning > Injector Rack Mount Unit (Option). Wireless Footswitch (Option) 0 The "wireless footswitch" is comprised of the following components: Footswitch Released Accessories Ceiling column with fixed installation (For Dimensional Drawing, see Detail Planning> Fixed Upper Body Radiation Shield, Surgery Lamp; for posi- tioning, see Installation Preparations > DCS Version 1 with Fixed Upper Body Radiation Shield (Option), Detail Planning > Ceiling Installation with Fixed Upper Body Radiation Shield (Option)) Upper body radiation shield or injector stand or Upper body radiation shield and surgery light or Ceiling column installed on a carriage (For Dimensional Drawing, see Detail Planning > Mov- able Upper Body Radiation Shield, Surgery Lamp (Option); for positioning, see Installation Preparations > Ceiling Installation, Detail Planning > System Ceiling Installation) Upper body radiation shield and/or sur- gery light Second carriage for injector stand pos- sible (separate Part No.) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 43 Page 43 of 130 Charging unit for footswitch, consisting of: Wall holder for charging cable (7 m); Plug-in power supply with cable (2 m). Notes for Planning: Plan the wall holder in proximity of the patient table in the examination room (max. 5 m from the patient table). Attach with screws/anchors or with (adhesive) Velcro tape. All mounting hardware is included in the shipment. Weight of wall holder: < 1 kg. Power consumption: < 10 W. Power outlet for the charging unit (no fuse protection in the Siemens system). The outlet always must be live to charge the footswitch battery overnight. Wall holders and footswitches may not be installed or used in the control room. 44 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 44 of 130 Component Installation 3.1 Floor Mounting 0 Floor Stand/Patient Table The two installation plates for the stand and the patient table must be installed on solid sur- face with sufficient load capacity. Cut out the existing floor covering. If applicable, material that is not capable of handling the load capacity in the area of the installation plates must be replaced with filled concrete. Hilti heavy-duty anchors are included for installation of the installation plates: Longer heavy-duty anchor bolts are available from the Hilti Company on request. If needed, the installation plates can also be secured through the ceiling of the room below using M12 threaded rods, min. hardness rating 8.8 according to the ISO Norm (obtain the threaded rods, etc. on site). Fig. 25: Floor stand/patient table Pos. 1 Floor stand support base plate Pos. 2 Floor stand installation plate Pos. 3 Orientation point/isocenter for floor stand in the head-end position Pos. 4 Patient table installation plate Pos. 5 Patient table orientation point *1 Patient table cable lead-in: with installation directly on solid concrete, lead in the cables through a break in the ceiling from the room below (see Installation Preparations > Patient Table). with installation on filled concrete, through a cable channel (see Installation Preparations > Patient Table). Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 45 Page 45 of 130 Installation directly on Solid Concrete Floor Stand Fig. 26: Floor stand Pos. 1 Mounting points for floor stand Pos. 2 Orientation point/isocenter for floor stand in the head-end position Pos. 3 Insulation parts 46 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 46 of 130 Pos. 4 Solid concrete Pos. 5 Installation plate Pos. 6 Support base plate Pos. 7 Insulation plate Installation plate with 13 holes ( 25 mm): Anchored directly in the concrete with 7 Hilti HSL-3 M12/50 anchors; 7 evenly distrib- uted holes should be used (see Room Planning > Static Requirements). No more than two holes may remain open between the attachment points. Patient Table *1 Hilti HSL-3 M12/50 anchors (7 included in the shipment) *2 Floor anchoring with threaded rods through the ceiling possible (threaded rods must be purchased locally) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 47 Page 47 of 130 Fig. 27: Patient table Pos. 1 Head end Pos. 2 Installation plate Pos. 3 Solid concrete Insulation parts (insulation plate, etc.) not shown. *1 Anchored with 4 Hilti HSL-3 M12/25 anchors *2 Area for cable inlet in table base *3 Opening in the installation plate *4 Cable lead-in through ceiling opening from the room below *5 Patient table orientation point 48 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 48 of 130 Installation on Filled Concrete Floor Stand Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 49 Page 49 of 130 Fig. 28: Floor stand Pos. 1 Mounting points for floor stand Pos. 2 Size of the concrete filling Pos. 3 Orientation point/isocenter for floor stand in the head-end position Pos. 4 Insulation parts Pos. 5 Installation plate Pos. 6 Support base plate Pos. 7 Insulation plate Pos. 8 Solid concrete Installation plate with 13 holes ( 25 mm): Anchored on filled concrete with 7 Hilti HSL-3-G M12/100 anchors; 7 even-distributed holes should be used (see Room Planning > Static Requirements). No more than two holes may remain open between two attachment points. *1 Hilti HSL-3-G M12/100 anchors (7 included in the shipment) *2 Floor anchoring with threaded rods through the ceiling possible (threaded rods must be purchased locally) 50 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 50 of 130 Patient Table Fig. 29: Floor stand Pos. 1 Cable duct, e.g. 200 x 70 mm Pos. 2 Head end Pos. 3 Area of the concrete filling Pos. 4 Self-leveling grout Pos. 5 Insulation Pos. 6 Filled concrete Pos. 7 Installation plate Pos. 8 Solid concrete Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 51 Page 51 of 130 Insulation parts (insulation plate, etc.) not shown. Ceiling Installation 0 DCS Version 1 with Movable Upper Body Radiation Shield (Option) Recommended ceiling grid: 675 mm *1 Anchored with 4 Hilti HSL-3 M12/100 anchors *2 Area for cable inlet in table base; cable routing via cable duct or opening in ceiling from room below. When routing via cable duct: To ensure the cable is not bent severely at the table inlet, plan only cable ducts with a minimum height of 70 mm. *3 Opening in the installation plate *4 Patient table orientation point 52 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 52 of 130 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 53 Page 53 of 130 Fig. 30: DCS version 1 with movable upper body radiation shield (option) Pos. 1 Grid spacing Pos. 2 Do not plan any lighting components, ventilation outlets, sprinkler elements, etc. between the ceiling rails. Pos. 3 Radiation shield Pos. 4 DCS NOTE If the room depth permits it, the DCS should be moved one ceiling grid to the left to make unrestricted pivoting behind the table possible.
DCS Version 1 with Fixed Upper Body Radiation Shield (Option) Not recommended in conjunction with the MULTISPACE.F option *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 DCS ceiling outlet *4 observe the maximum admissible rail overhang. *5 The ceiling rails (option) for the radiation shield, surgery lamp, injector stand (see Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (option)); length when shipped: 4 m, can be shortened. IN1 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option), (see Installation Preparations > On-site Power Cables, Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (Option). TR1 Control console cart (option) (see Detail Planning > Control Console Cart (Option)) 54 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 54 of 130 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 55 Page 55 of 130 Fig. 31: DCS version 1 with fixed upper body radiation shield (option) Pos. 1 Do not plan any lighting components, ventilation outlets, sprinkler elements, etc. between the ceiling rails. Pos. 2 Radiation shield installation plate Pos. 3 DCS NOTE If the room depth permits it, the DCS should be moved one ceiling grid to the left to make unrestricted pivoting behind the table possible.
*1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 DCS ceiling outlet *4 Installation plate (option, see Detail Planning > Fixed Upper Body Radiation Shield, Surgery Lamp (Option)) for Radiation Shield and Surgery Lamp or Injector Stand *5 With a smaller ceiling grid, move the installation plate to the side until it is centered to the ceiling grid. IN2 Injector head with fixed installation on the ceiling (option) SU Surgery lamp (option) (see Installation Preparations > On-site Power Cables, Detail Planning > Fixed Upper Body Radiation Shield, Surgery Lamp (Option)). TR1 Control Console Cart (option) 56 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 56 of 130 On-site Electrical Installation 3.2 The detailed regulations regarding electrical installations must be observed. In Germany, the most important standards used are in the series VDE 0100 and VDE 0100-710; in other countries, the applicable national regulations. For power line connection voltages, see (Electrical Data / p. 87). Notes regarding the recommendations on the following pages for electrical instal- lation in accordance with DIN VDE 0100-710 0 RCD Ground Fault Circuit Interrupters Only use ground fault circuit interrupters for alternating and pulsed DC fault currents, with no surge current, with I N = 30 mA. An RCD ground fault circuit interrupter cannot be used in power networks without an N cable. RCD ground fault circuit interrupter for system control cabinet (SC1): I N = 40 A/ I = 30 mA RCD ground fault circuit interrupter for generator: According to DIN EN 50178 clas- sification VDE 0160, "Ausrstung von Starkstromanlagen mit elektronischen Betrieb- smitteln" (Equipping of high-voltage systems with electrical resources), the following FI circuit breakers are to be used exclusively. Line voltage RCD Pre-fusing for U N = 3 400/415 V: I N = 63A I = 30 mA Purchased via electrical wholesaler: Order no. 5SM3 346-4 allstrom Purchased via SPH (Med depart- ment: Part no. 49 54 470 Y7933 (Width of the RCD ground fault cir- cuit interrupter is 72 mm = 4TE, mounting on a standard rail) 63 A slow-blow for U N greater than 3 415 V: U N = 500V I N = 125A I = 30 mA Purchased via SPH (Med depart- ment): Part no. 51 41,168 Y7933 63 A slow-blow Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 57 Page 57 of 130 On/Off switch (with pilot light) Provide an on-off switch for the system. Preferably, this on/off switch should be located in the equipment room (if configured). The installation height is 180cm. Emergency Off switch (with lock) For fast system shutdown in hazardous situations, at least one emergency off switch (with lock) is to be provided in the examination room, control room, and equipment room. It should be positioned 1.8 m high is such a way that it is easy to reach but without the pos- sibility of accidental activation. Recommendation for Application Group 1 0 Electrical installation on site per DIN VDE 0100-710 Fig. 32: Electrical installation: Recommendation application group 1 A = Suggestion for room distributor B = Power supply cable for X-ray C = 5x according to calculation 1 Potential ground wire to external conductive parts 2 Power cable to fixed or movable equipment 58 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 58 of 130 NOTE The short circuit pre-fusing values apply to the SIEMENS ground fault circuit interrupters only
NOTE If using a system UPS, the specifications in the Emergency Power Supply PG must be observed for the emergency OFF circuit.
Recommendation for Application Group 2 0 Electrical installation on site per DIN VDE 0100-710 3 Power cable to fixed or movable equipment, e.g., developer machine 4 Ground Fault Circuit Interrupter 4a Ground fault circuit interrupter for system control cabinet 4b Ground fault circuit interrupter for generator 5 System breaker 6 Isolation transformer with overload monitoring conforming to DIN VDE 0100-710, Section 710.512.1.6.2 (output power 3.15 - 8 kVA, isolation test volt- age 4 kV). 7 For MMV connection, see PG: AXA4-100.891.14... (option) EAT On/Off switch with pilot lamp for X-ray system AT1 Emergency off button in examination room AT2 Emergency off button in control room AT3 Emergency off button in equipment room [] Figures are free cable lengths in m *1 Surgery light connection (115 V or 230 V ) (option), see Installation Prepa- rations > On-site Electrical Installation > On-site Power Cables. *2 Table outlet connection (option), see Installation Preparations > On-site Elec- trical Installation > On-site Power Cables > Table Outlet. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 59 Page 59 of 130 Fig. 33: Electrical installation: Recommendation for Application Group 2 Pos. 1 Power supply Pos. 2 Suggestion for room distributor Pos. 3 Emergency Power Supply SV Pos. 4 Second cable Pos. 5 First cable Pos. 6 Power supply cable for X-ray Pos. 7 5x per the calculation *1 Surgery light connection (115 V or 230 V ) (option), see Installation Preparations > On-site Elec- trical Installation > On-site Power Cables. *2 Table outlet connection (option), see Installation Preparations > On-site Electrical Installation > On-site Power Cables > Table Outlet. *3 Required only for the CATHCOR (not for Sensis). 1 Voltage monitoring and switching device according to VDE 0100-710 2 Isolation transformer with overload monitoring conforming to DIN VDE 0100-710, Section 710.512.1.6.2 (output power 3.15 - 8 kVA, isolation test voltage 4 kV). 3 Isolation monitoring device 60 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 60 of 130 NOTE The short circuit pre-fusing values apply to the SIEMENS ground fault circuit interrupters only
NOTE If using a system UPS, the specifications in the Emergency Power Supply PG must be observed for the emergency OFF circuit.
On-site Power Cables 0 A maximum of four on-site power cables are required: Power cables to the system control cabinet (SC1) - Power cable, 5x 16 mm 2 . - The connection terminals are designed for a maximum of 16 mm 2 . 4 Report and test combination 5 Potential ground wire to external conductive parts 6 Power cable to fixed or movable equipment 7 Power to fixed or movable equipment, e.g., developer machine 8 Ground Fault Circuit Interrupter 8a Ground fault interrupter switch for the system control cabinet, see the note (Installation Preparations > On-site Electrical Installation) 8b Ground fault interrupter switch for the generator, see the note (Instal- lation Preparations > On-site Electrical Installation) 8c Pre-fusing for the generator, see (Installation Preparations > On-site Electrical Installation) 9 System breaker 10 4-socket connection strip for AXIOM Sensis (option) 11 Gnd pin socket 12 For the MMV connection, see PG: AXA4-100.891.14... (option) EAT On/Off switch with pilot lamp for X-ray system AT1 Emergency off button in examination room AT2 Emergency off button in control room AT3 Emergency off button in equipment room D1 ECG display option in DCS [] Figures are free cable lengths in m Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 61 Page 61 of 130 Generator Power Cable (PU1) - Power cable 5x per the calculation. Here, take note of the internal line resistance: a) Measure the internal power line resistance, R i , at the sub-distributor with an internal line resistance meter. b) Size the power cable so that the max. admissible R imax (see table) is not exceeded. Measured internal power line resistance plus resistance of the power cable R imax . * Resistance value in Ohm at U N , -15% ** Resistance value in Ohm at U N , -10% Resistance values in Ohms at UN, +/- 10% - Maximum power output for the generator (Angio Generator M or POLYDOROS 100A) is 100 kW. With a larger internal line resistance than is listed in the above table for 100 kW, the generator power is automatically reduced. For the X-ray tubes currently being shipped, a max. generator power of 80 kW is suf- ficient. - Minimum Ri: 0.02 Ohm - Max. cable cross section that can be connected: 70 mm 2 . Angio Generator M U N P 100 kW 80 kW 400 V* 0.10 Ohm 0.14 Ohm 415 V** 0.14 Ohm 0.17 Ohm 440 V** 0.16 Ohm 0.21 Ohm 480 V** 0.20 Ohm 0.27 Ohm POLYDOROS A100 U N P 100 kW 80 kW 380 V 0.08 Ohm 0.10 Ohm 400 V 0.09 Ohm 0.11 Ohm 420 V 0.09 Ohm 0.12 Ohm 440 V 0.10 Ohm 0.14 Ohm 480 V 0.12 Ohm 0.16 Ohm 62 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 62 of 130 Connecting the Surgery Lamp (Option) Fig. 34: Connecting the surgery lamp (option) Pos. 1 Room distributor Pos. 2 Surgery lamp, 1 bulb Pos. 3 Surgery lamp, 3 bulbs Surgery lamp: (1 bulb) Power consumption 50 W can be connected to 115 V or 230 V (inter- nal transformer) Surgery lamp: (3 bulbs) Power consumption 160 W can be connected to 115 V to 247 V (external transformer) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 63 Page 63 of 130 If necessary, obtain all cables, switches, and relays locally. Table Power Outlet (Option) An optional power outlet is available for the patient table. - If the power outlet was ordered, it has already been installed in the table at the fac- tory. - All patient tables are shipped ex factory with the 2 x 1.5 mm 2 power cable pulled in. Connecting the Table Power Outlet: in examination rooms according to application group 1 with MMV: Power supply to the table power outlet via MMV or an external isolation transformer with overload monitoring (see (Recommendation for Application Group 1 / p. 57)). in examination rooms according to application group 1 without MMV: Power supply to the table power outlet via an external isolation transformer with over- load monitoring (Recommendation for Application Group 1 / p. 57). in examination rooms according to application group 2: Regardless of whether an MMV is configured, the connection to the hospital backup power supply is provided via an isolation transformer with isolation monitoring in accor- dance with EC 60364-7-710, Section 710.413.1.5 (output power 3.15 - 8 kVA, isolation test voltage 4 kV). Secondary fusing of the outlet at 115 V: 5 A, for 230 V: 2.5 A(Recom- mendation for Application Group 2 / p. 58) *1 For the connection to line voltage, see Installation Preparations > (Recommendation for Application Group 1 / p. 57), Installation Preparations > (Recommendation for Application Group 2 / p. 58), *1 *2 The light can be connected: directly or via a wall switch (*2a), an additional footswitch, Part No. 17 41 313 (*2b), via the potential-free contact of the Z58/M102 to the line voltage. The surgery light is switched on and off with fluoro / exposure via the potential-free contact of the Z58/M102. *3 The three-bulb surgery lamp (option) operates via an external transformer. The trans- former (dimensions L 280 x W 200 x H 160 mm; weight approx. 8 kg) should be installed near the SC1. *4 For systems shipped prior to 6/2005, the K4 relay must be planned on site, e.g. in the room distributor. 64 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 64 of 130 Siemens Remote Services (SRS) 3.3 For detailed information on SIEMENS Remote Services (SRS), see PG TD00-000.891.01... (on the Intranet under: CS/ForService/ProductInformation/Siemens Remote Service Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 65 Page 65 of 130 Network 3.4 The image system is equipped with a network card for Fast Ethernet (100 Base T). Network connection with system switch (in delivery volume) Fig. 35: Network 66 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 66 of 130 DICOM 3.5 For detailed information on DICOM, see PG TD00-000.891.02... Can be found on the Intranet under: CS/ForService/ProductInformation/Planning/Gen- eral... Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Installation Preparations 67 Page 67 of 130 Injector Floor Outlet Option 3.6 Fig. 36: Injector floor outlet (option) Pos. 1 Patient table installation plate Pos. 2 Table longitudinal axis Pos. 3 Area in which the outlet is installed. The outlet must be planned centered to the table longitudinal axis. Pos. 4 Head end NOTE An injector cable may not be pulled through the table base. It is absolutely necessary to route the injector cable via the floor outlet.
The following manufacturers have been released: 0 Medrad: Josh Sevick (Cardiovascular Product Manager) Medrad Inc. One Medrad Drive Indianola, PA 15051-0780 jsevick@medrad.com 68 Installation Preparations AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 68 of 130 Tyco: Pat Dyer Site Planning Dept. 2111 E. Galbraith Rd. Cincinnati, Ohio 45237 pat.dyer@tycohealthcare.com +1 (513) 761-2700 ext.5721 Detailed information on planning / project planning can be obtained from the addresses indicated Fig. 37: Medrad floor outlet Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 69 Page 69 of 130 4- 4System Connections General Information 0 All system-specific cables (including the protective conductor) are included in the ship- ment. The system cables consist of standard cables and cable wiring harnesses of a fixed length. They may not be shortened. Excessive cable lengths can be stored in the cable cabinet (option), cable ducts, under computer floors, etc. Notes regarding on-site electrical installation are included (Recommendation for Appli- cation Group 1). 70 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 70 of 130 Overview of System Connections 4.1 Fig. 38: Overview of system connections Pos. 1 Ethernet Notes: RV = Room Distributor The room distributor and the power cables are not included in the shipment. The illustration includes the cable cabinet (option). The wiring harnesses can also be laid to the fixpoints directly, i.e., without routing them through the cable cabinet. The excess cable lengths must then be stored in another way (in cable conduits, under computer flooring, etc.). The option "monitor trolley (fixpoint D2)" is not shown. For a definition of the fixpoints, see the System Connections > List of Fixpoints. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 71 Page 71 of 130 List of Fixpoints 4.2 Maximum Fixpoint Distances and Planning Cable Channel Sizes 0 Wiring Harness No. From fix- point To fix- point Cable channel cross-sec tion in mm2 Con- duit, clear space in inches Mini- mum open- ing in mm Maximum fix- point dis- tance in m Comment 1 P1 PU1 780 2 32 16 2a P1 PU1 500 3 42 16 High-voltage cables 2b P1 PU1 500 3 42 16 High-voltage cables 3 P1 SC1 3125 3 70 x 50 15 min. bending radius: 80 mm 4 P1 CU1 2000 2,5 50 30 min. bending radius: 50 mm 5 SC1 CR1 780 3 60 x 35 19 for control room options (control mod- ules, footswitches, dis- play, ECG). 6 SC1 CR1 780 3 55 x 55 19 7 SC1 T1 3125 3 80 x 55 14 skipped with OR tables 8 SC1 SC2 780 2 50 x 25 2 only with OR tables 9a SC2 T1 3125 3 80 x 55 14 9b SC2 T1 3125 3 80 x 55 14 10 SC1 CU1 780 2 35 30 11 SC1 PU1 780 2 35 x 35 5 12 SC1 IS 780 3 55 x 35 19 13a SC1 D1 780 3 50 x 40 19 DCS (Option) 13b IS D1 780 2,5 40 x 25 19 14a SC1 D12 780 3 50 x 40 19 2. DCS (Option) 14b IS D12 780 2,5 40 x 25 19 15 SC1 D2 780 2,5 40 x 25 19 Monitor Trolley (Option) * 16 IS D2 780 2,5 40 x 25 19 72 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 72 of 130 17a IS CR1 780 3 50 x 25 19 only if the image sys- tem is located outside of the control room. 17b IS CR1 780 2 25 19 17c IS CR1 780 2 25 19 17d IS CR1 780 2 35 x 15 19 18 SC1 TR1 780 3 60 x 35 16 Control Console Cart (Option) ** 19 IS Ethernet cable (thick wire or twisted pair) Ethernet connection (option), on-site cable 19a CR1 Net- work socket Ethernet cable (included in shipment) see Installation Prepa- rations > Network; 5 m and 30 m included in shipment 20 SC1 IW 780 3 60 x 35 19 Injector wall connec- tion (option) 21 SC1 IW 780 3 55 x 55 19 22 CR1 Dis- play 290 2 30 x 20 8 Second display (option) 23 CR2 D1 150 2,5 40 x 40 27 Measuring station (option) 24 CR2 SC1 100 3 70 x 17 27 Measuring station (option) 25 CR2 T1 120 2 45 x 15 27 Measuring station (option) T1 CR1 Ultrasound Integra- tion option; for details, see the Ultrasound PG. RV PU1 Power cable Generator (Power Unit 1) -> Installation Preparations > Recommendation for Application Group 1 On-site cables RV SC1 Power cable System control cabinet -> Installation Preparations > Recommendation for Application Group 1 On-site cables RV SU Power cable Surgery Lamp (Option) -> Installation Preparations > Recommendation for Application Group 1 On-site cables Wiring Harness No. From fix- point To fix- point Cable channel cross-sec tion in mm2 Con- duit, clear space in inches Mini- mum open- ing in mm Maximum fix- point dis- tance in m Comment Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 73 Page 73 of 130 List of Fixpoints Used 0 with- out the MMV RV with- out the MMV T1 125 1/2 10 28 Table power outlet option, see Installation Preparations > Rec- ommendation for Application Group 1; Installation Prepara- tions > Recommenda- tion for Application Group 2; Installation Preparations > On-site Power Cables with the MMV MMV with the MMV T1 125 1/2 10 28 Wiring Harness No. From fix- point To fix- point Cable channel cross-sec tion in mm2 Con- duit, clear space in inches Mini- mum open- ing in mm Maximum fix- point dis- tance in m Comment * 6 m from the ceiling outlet to the monitor cart ** 6 m from the ceiling outlet to the control console cart Fixpoint Subsystem Comment CR1 Control room (X-ray control room) Fixpoint on the bottom of the control room distributor CR2 Measuring station (control room, E-medicine) (option) Floor fixpoint CU1 Cooling unit for X-ray tube unit cooling (Cooling Unit 1) Floor fixpoint D1, D12 DCS (Option) Ceiling fixpoint D2 Monitor Trolley (Option) Ceiling fixpoint IN1 Injector head on the injector ceiling column in the radiation shield support arm system (option) Ceiling fixpoint IN2 Injector head with stationary installation on the ceil- ing (option) Ceiling fixpoint IS Image System Floor/wall fixpoint IW Injector wall connection in the examination room (option) (Injector Wall Outlet) Wall fixpoint 74 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 74 of 130 P1 Floor stand (Plane 1) Ceiling fixpoint SC1 System control cabinet #1 (system control) Floor fixpoint SC2 System control cabinet #2 (only with the Axiom Artis OR patient table) Floor fixpoint PU1 Generator #1 (Power Unit #1) Floor fixpoint SU Surgery lamp (option) Ceiling fixpoint T1 Patient table (Table) Floor fixpoint TR1 Control console cart (trolley) (option) Ceiling fixpoint) Fixpoint Subsystem Comment Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 75 Page 75 of 130 Laying Cables 4.3 Notes: Lay cables/wiring harnesses separately in conduits or in closed cable channels. If laying in open cable channels, lay cables separately using metal dividers or a similar device. The cable channel depth should be at least 70 mm; if there are cross-overs, it may be necessary to provide for greater depth. Calculating the Required Cable Channel Cross Section 0 The total of the cross sections of the cables/wiring harnesses present should apply for all cable channel/conduit cross sections. The individual cross sections can be taken from the cable list. The required fill factor are already included in this. If excess cable lengths are to be stored in the cable channels, this must be taken into account when sizing. Remarks regarding Laying Cables 0 Laying excess cable lengths: in the cable cabinet (option; service friendly) in cable channels (take note of accessibility) in ceiling/wall boxes (take note of accessibility) under computer flooring (service friendly) The following must be observed: Cable coils are not allowed! 76 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 76 of 130 Fig. 41: Routing excess cable length in the false floor Pos. 1 Meander cable in false floor Fig. 39: Routing excess cable length in the cable cabinet Pos. 1 Meander cable in cable cabinet Fig. 40: Routing excess cable length in the cable duct Pos. 1 Meander cable in cable duct Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 77 Page 77 of 130 Control Room Details 4.4 Image System in the Control Room (Standard) 0 Fig. 42: Optional image system in the control room (standard) Pos. 1 Display Pos. 2 Keyboard Pos. 3 Mouse Pos. 4 Ref. display Pos. 5 Handswitch Pos. 6 Control modules Pos. 7 ECC Pos. 8 Emergency stop Pos. 9 Footswitch Pos. 10 DVD recorder Pos. 11 Image system in the control room Pos. 12 Control room distributor 78 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 78 of 130 Image System optionally in the Equipment Room 0 If the image system is located in the equipment room, the video cables to the monitor, etc. are routed via the control room distributor. * Video cables ** Monochrome or color displays can be shipped. Type of Display max. distance to image system in m 17 color display 7,5 18 color display 3,5 18 monochrome display 3,5 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 79 Page 79 of 130 Fig. 43: Image system in the control room Pos. 1 Remote Power and Archive Unit Pos. 2 Display Pos. 3 Keyboard Pos. 4 Mouse Pos. 5 Ref. display Pos. 6 Handswitch Pos. 7 Control modules Pos. 8 ECC Pos. 9 Emergency stop Pos. 10 Footswitch Pos. 11 DVD recorder Pos. 12 Control room distributor * Monochrome or color displays can be shipped. 80 System Connections AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 80 of 130 Room Light, Radiation Display, Door Contact, EPS 4.5 For room light control, etc., only use flexible cables within the control cabinet (SC1). Fig. 44: Room Light, Radiation Display, Door Contact, EPS Pos. 1 System control cabinet Pos. 2 Room light Pos. 3 External radiation display Pos. 4 Potential-free door contact (24 VDC / 20 mA) *1, door closed = contact closed Pos. 5 Potential-free contact for EPS (24 VDC / 20 mA) *3, emergency power "on" = contact closed Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX System Connections 81 Page 81 of 130 *1 If needed, a switch to block radiation can be installed in series with the door con- tact. The switch (24 VDC / 20 mA) must be provided on site. Control using +24 V from the system control cabinet Plan the switch so that unintentional operation is not possible. The switch must be provided with an appropriate symbol (or designation) from which the function can be recognized. *2 If the 2 A current load is insufficient, plan an external power relay through which the room light can be switched on and off. *3 When emergency power is "on", exposure mode is blocked and only emergency fluoroscopy is possible. *4 With systems shipped prior to 6/2005, the k4/K5 relays must be provided on site. Control of the relay is with 24 V DC (max. 0.2 A). A free-wheeling diode must be circuited over the relay coil. SC1 Fixpoint system control cabinet 82 Technical Data AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 82 of 130 5- 5Technical Data Dimensions, Weights and Heat Dissipation,... 0 Examination Room 0 Fixpoint Sub-system Dimensions (W x D x H) [mm] Weight [kg] Heat Dissipa- tion [kW] D1 DCS (Option) see the DCS PG D2 Monitor carriage with display (option) see the PG P1 Floor stand (with installation plate) Installation plate see Room Plan- ning > Floor Stand: Side View See Installation Preparations > Installation directly on Solid Concrete < 665 approx. 45 0,2 n.a. T1 Patient table (incl. installation plate, 22 kg and table accessories, 40 kg): Basic table Peri table Tilt table OR table see Room Plan- ning > Views 452 482 530 550 0,2 0,2 0,2 0,1 TR1 Control console cart See Detail Plan- ning > Control Console Cart (Option). approx. 26 n.a. SU Fixed upper body radiation shield, surgery lamp (option) See Detail Planning > Fixed Upper Body Radiation Shield, Surgery Lamp (Option) n.a. SU Movable Upper Body Radiation Shield, Surgery Lamp (Option) See Detail Planning > Mov- able Upper Body Radiation Shield, Surgery Lamp (Option) n.a. IW Injector Wall Connection Box (Option) 324 x 102 x 257 approx. 5 n.a. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Technical Data 83 Page 83 of 130 Equipment Room 0 Control Room 0 Fixpoint Sub-system Dimensions (W x D x H) [mm] Weight [kg] Heat Dissipation [kW] CU1 Cooling Unit (Cooling Unit 1) 479 x 398 x 481 42 2,4 SC1 System Control Cabinet #1: System with image intensifier Systems with Flat Panel Detec- tor 600 x 434 x 2202 235 270 0,9 max. 1.6 SC2 System Control Cabinet #2 (only with the OR table) 600 x 434 x 2202 125 0,2 PU1 Angio Generator M 800 x 434 x 2202 350 1,0 PU1 Polydoros A100 800 x 434 x 2202 300 1,0 n.a. Cable cabinet (option) 800 x 434 x 2202 120 n.a. Fixpoint Sub-system Dimensions (W x D x H) [mm] Weight [kg] Heat Dissipation [kW] IS Image system 689 x 800 x 720 87 0,7 n.a. UPS for Image System (Option) 170 x 439 x 216 19 0,03 n.a. Remote Power Archive Unit (only when image system is in equipment room) 216 x 368 x 222 9 < 0,1 CR1 Control Room Distributor 1054 x 207 x 409 29 0,1 n.a. Keyboard 447 x 157 x 20/55 < 1 0,1 n.a. 17" display or 17 color display or 18 color display or 18 monochrome display 413 x 230 x 413 413 x 230 x 413 413 x 230 x 413 413 x 230 x 413 9,0 9,0 9,0 9,8 < 0,07 < 0,07 < 0,07 < 0,07 n.a. DVD Recorder 214 x 365 x 102 4,6 0,1 n.a. 17" Color Display for DVD Recorder (Option) 413 x 230 x 413 9,0 < 0,07 n.a. Second Examination Control Con- sole (ECC) (Option) 370 x 225 x 70/90 approx. 3 n.a. n.a. Patient Table Control Console (Option) 91 x 161 x 65/85 < 1 n.a. 84 Technical Data AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 84 of 130 Floor stand control console in the col- limator control console (option) 216 x 161 x 65/85 < 2 n.a. n.a. Control console (option) (height adjustable from 705 to 730 mm) 1392 x 800 x 720 approx. 60 n.a. D1 Second Data Display (Option) See Detail Planning > Second Data Dis- play (Option). 9 n.a. Fixpoint Sub-system Dimensions (W x D x H) [mm] Weight [kg] Heat Dissipation [kW] Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Technical Data 85 Page 85 of 130 Climatic Conditions 5.1 NOTE Observe the limited temperature range in the examination room as well as in the equipment room (SC1 system control cabinet) with the systems with the flat panel detector (see *2).
Examination and control room Temperature range: Relative humidity: 10 C... 35C *1 (recommended tempera- ture 22 C) 15 C... 30 C *2 20-75% without condensation Image System (IS) Temperature range: Relative humidity: Max. temperature gradient: Air flow rate: Max. noise level: 10 C... 35 C 20-75% without condensation 10 C/h 657 m 3 /h 50 dB(A) Angio Generator M or POLYDOROS A100 (PU1) Temperature range: Relative humidity: Max. temperature gradient: Air flow rate: Max. noise level: 10 C... 35 C 20-75% without condensation 5 C/h 160 m 3 /h 55 dB(A) System control cabinet #1 (SC1) Temperature range: Relative humidity: Max. temperature gradient: Air flow rate: Max. noise level: 10 C... 35 C *1 15 C... 30 C *2 20-75% without condensation 5 C/h 500 m 3 /h 48 dB(A) System control cabinet #2 (SC2) Temperature range: Relative humidity: Max. temperature gradient: Air flow rate: Max. noise level: 10 C... 35 C 20-75% without condensation 5 C/h n.a. n.a. 86 Technical Data AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 86 of 130 Cooling unit (CU1) (for Megalix X-ray tube) Cooling air: Air flow: Max. noise level: 5 C... 30 C (frost-free room) 1100 m3/h 55 dB(A) at 50 Hz, 59 dB(A) at 60 Hz Stand with flat detector Max. temperature gradient: Barometric pressure: Mechanical impact: Vibrations: 5 C/h 70 kPa... 104 kPa max. 10 g/ 16 ms max. 0.1 g/ 10-200 Hz *1: System with image intensifier *2: Systems with Flat Panel Detector Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Technical Data 87 Page 87 of 130 Electrical Data 5.2 NOTE If using on-site voltage stabilization, contact Customer Solu- tions.
Nominal connec- tion volt- age * [V] +/- 10% Frequency [Hz] +/- 1 Hz Internal fusing [A] Exter- nal fus- ing [A] Max. power con- sump- tion [kVA] Connection value ** [kVA] Angio Generator M (shipped until 09/2005) (Power Unit PU1) 3 ph. 400 50/60 63 *** 18 for fluo- roscopy 160 for exposure 43,6 3 ph. 415 50 45,3 3 ph. 440 50/60 48,0 3 ph. 480 60 52,4 POLYDOROS A100 (Power Unit PU1) 3 ph. 380 50/60 50 63A slow-b low 8 for fluo- roscopy 160 for exposure 32,9 3 ph. 400 34,6 3 ph. 420 36,4 3 ph. 440 38,1 3 ph. 480 41,6 System control cabi- net (System Control SC1) 3 ph. 380 50/60 35 50
slow-b low 8,5 23,0 3 ph. 400 50/60 24,2 3 ph. 440 50/60 26,7 3 ph. 480 60 29,1 * Max. permissible voltage between phases (L1, L2, L3) and Gnd: 300 V ** The connection value is calculated using the following formula: Connection value = 3 x nominal connection voltage x internal fusing *** Depends on the type of ground fault interrupter, see Installation Prep- arations > On-site Electrical Installation 88 Technical Data AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 88 of 130 Paint Information 5.3 Primary color Medical White, C610 Combination Color Medical Blue, C750 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Technical Data 89 Page 89 of 130 Transport and Storage 5.4 System Temperature range: -20 C ... 60 C Relative humidity: 10 % ... 95% without condensation Barometric pressure: 50 kPa ... 106 kPa Mechanical shocks max. 10 g/ 16 ms max. 25 g/ 6 ms (in original packaging) Vibrations max. 2 g/ 58-150 Hz max. 10 g/ 58-150 Hz (in original packaging) 90 Transport and Storage Conditions AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 90 of 130 6- 6Transport and Storage Conditions Packaging and Transport Routes 0 Largest crate with packaging, weight dFC/FC/FA W 2760 x D 1020 x H 2090 mm, approx. 1200 kg dFA W 2800 x D 950 x H 2020 mm, approx. 1200 kg Largest single part with transport car- riage (min. door opening), weight dFC/FC/FA W 2500 x D 820 x H 1900 mm, approx. 760 kg dFA W 2620 x D 820 x H 1880 mm, approx. 760 kg Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 91 Page 91 of 130 7- 7Detail Planning Generator 0 Fig. 45: Generator 92 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 92 of 130 A = Front View B = Side View C = Top-down view 1 = Exhaust air 2 = Cabinet base 3 = Air intake 4 = For service 5 = Cabinet doors, removable *1 Removable base panels on all sides. *2 Distance to wall is not required. *3 When setting up across from one another, the min. distance to the cable cabinet is 40 mm (required to open the cabinet doors). *4 Wall mounting is required, because the assemblies can be pivoted out. Secure the cabinet to the wall with wall anchors and the supplied brackets. Obtain anchors locally. If needed, the cabinet can also be secured to the floor with anchors (see next page). Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 93 Page 93 of 130 Fig. 46: Generator Pos. A Top-down view of cabinet Pos. B Top-down view of cabinet base Pos. C Side view of cabinet base Pos. D Rear view of cabinet base Pos. 1 Ventilation *1 14 mm holes for floor mounting Obtain anchors locally. *2 Cable lead-in To ensure that the power cable can be strain-relieved, it must be lead into the cabinet base as far as possible on the right (direction of arrow). The power cable cannot be lead in from the left side. 94 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 94 of 130 System Control Cabinets #1 / #2 7.1 Fig. 47: System control cabinet #1 / #2 A = Front View B = Side View Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 95 Page 95 of 130 C = Top-down view 1 = Exhaust air 2 = Cabinet base 3 = Air intake 4 = For service 5 = Cabinet doors, removable *1 Removable base panels on all sides. *2 Distance to wall is not required. *3 When setting up across from one another, the min. distance to the cable cabinet is 40 mm (required to open the cabinet doors). *4 Wall mounting is required, because the assemblies can be pivoted out. Secure the cabinet to the wall with wall anchors and the supplied brackets. Obtain anchors locally. If needed, the cabinet can also be secured to the floor with anchors (see next page). 96 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 96 of 130 Fig. 48: System control cabinets #1 / #2 Pos. A Top-down view of cabinet Pos. B Top-down view of cabinet base Pos. C Side view of cabinet base Pos. D Rear view of cabinet base Pos. 1 Exhaust air *1 14 mm holes for floor mounting Obtain anchors locally. *2 Surface (area with thick outline) for cable feed from the floor (cable duct). *3 The cables can be laid over the floor to the base from all sides after removing the base panels, but not from the front. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 97 Page 97 of 130 Image System 7.2 Fig. 49: Image system Pos. A Front view Pos. B Side view Pos. C Top-down view Pos. 1 Exhaust air Pos. 2 Air intake Pos. 3 Castors Pos. 4 Wall Pos. 5 Cable loop approx. 2 m on the floor under the image system for service 98 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 98 of 130 *1 Cable outlet; cable lead-in is possible through the floor or wall outlet. *2 Under normal circumstances, no side spacing to walls or other components is nec- essary. The image system can be moved into a position for service using the castors so that it is accessible from all sides (approx. 0.5 m on each side). Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 99 Page 99 of 130 Floor Stand Ceiling Outlet 7.3 Fig. 50: Floor stand ceiling outlet: A = cable outlet in the installation ceiling / B = cable outlet edge trim Pos. 1 M10 threaded rods Pos. 2 Drop ceiling Pos. 3 Housing Pos. 4 Cover cap Pos. 5 Ceiling opening Pos. 6 Tabs can be installed in the 25 mm radius area Pos. 7 Outlet dia. approx. 100 mm 100 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 100 of 130 *1 Ceiling mounted, tensile force per screw is less than 1000 N. *2 Threaded rod spacing *3 Cable lead-in trim. Cable inlet is cut into the housing. Cable channel inside dimensions are min. 60 x 160 mm Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 101 Page 101 of 130 Control Room Distributor (BTEE) 7.4 Installation location: Below the console or on the wall. If needed, the control room distributor can also be set up directly on the floor (wall anchoring required). Fig. 51: Control Room Distributor (BTEE) Pos. 1 Console plate Pos. 2 Cable lead-in for cables from the system possible from the right and left (see arrow) Pos. 3 Cable inlet Pos. 4 Cable lead-in (from above) for monitors, ECC, mouse and keyboard, optional control consoles, remote power archive unit (only if the image system is in the equipment room) *1 Mounting spacing, obtain mounting materials on site. 102 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 102 of 130 Control Consoles (Option) 7.5 Fig. 52: Control console (option) Pos. 1 Control room distributor Pos. 2 Height adjustable Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 103 Page 103 of 130 Tabletop Opening for Control Modules and ECC 7.6 Fig. 53: Tabletop cutout for the control modules and ECC Pos. 1 Table cutout Pos. 2 Work surface Provide an opening in the work table conforming to the configured control modules and, if applicable, the ECCs: Table Control Module (TCM) Floor Stand Control Module (SCM) Collimator Control Module (CCM) EEC the EEC can also be installed in a different location, separately from the control mod- ules. Unline with the control modules, the ECC is also suitable for free setup, i.e., it does not have to be recessed into the tabletop. 104 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 104 of 130 Second Data Display (Option) 7.7 The display can be placed on an on-site console or installed on the wall. Fig. 54: Second Data Display (Option) Pos. 1 Mounting points Pos. 2 Data Display Pos. 3 Lateral pivot range *1 The mounting materials (four 8 x 70 wood screws and screw anchors) are included in the shipment. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 105 Page 105 of 130 Stationary Upper Body Radiation Shield, Surgery Lamp (Option)7.8 With installation of an injector stand, it is not permissible to also add a surgery lamp or a radiation shield. The spacing between the bottom edge of the radiation shield ceiling column and the floor is RH - 820 mm (RH = room height). Recommendation: For room heights measuring less than 2900 mm, it is better to use the movable upper body radiation shield and not the sta- tionary version. 106 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 106 of 130 Fig. 55: Stationary upper body radiation shield, surgery lamp (option) Pos. 1 Installation plate Pos. 2 Surgery lamp, 1 or 3 beam Pos. 3 Ceiling column Pos. 4 Studs for radiation shield or injector stand Pos. 5 Double articulated arm Pos. 6 Large pane Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 107 Page 107 of 130 Installation plate approx. 23 kg 1. Ceiling column with extender, spring-loaded arm and large disk or 2. ceiling column with extender, spring-loaded arm and ceiling lamp or 3. maximum configuration (1-beam lamp and large disk) 4. maximum configuration (3-beam lamp and large disk) approx. 36 kg approx. 30 kg approx. 50 kg approx. 51 kg 108 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 108 of 130 Movable Upper Body Radiation Shield, Surgery Lamp (Option) 7.9 Fig. 56: Movable upper body radiation shield, surgery lamp (option) Pos. 1 Movable cart Pos. 2 Longitudinal rail Pos. 3 Surgery lamp, 1 or 3 beam Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 109 Page 109 of 130 Pos. 4 Ceiling column Pos. 5 Double articulated arm Pos. 6 Large disk *1: Alternatively, the short column A second cart for the injector head is possible (separate Part No.). Longitudinal rail approx. 30 kg Movable cart approx. 5 kg 1. Ceiling column with extender, spring-loaded arm and large disk or 2. ceiling column with extender, spring-loaded arm and ceiling lamp or 3. maximum configuration (1-beam lamp and large disk) 4. maximum configuration (3-beam lamp and large disk) approx. 36 kg approx. 30 kg approx. 50 kg approx. 51 kg 110 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 110 of 130 Control Console Cart Ceiling Outlet 0 Fig. 57: Monitor or control cart ceiling outlet Pos. 1 e.g. Unistrut profile Pos. 2 Installation parts (included in shipment) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 111 Page 111 of 130 Pos. 3 Cable outlet Pos. 4 Bracket (included in shipment) Pos. 5 Bracket installation on the profile (on-site) System Ceiling Installation with Monitor Trolley Option 0 Ceiling Installation with Movable Upper Body Radiation Shield (Option) Recommendation: ceiling grid of 675 mm 112 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 112 of 130 Fig. 58: Ceiling installation with mobile upper body radiation shield (option) Pos. 1 Grid spacing Pos. 2 Do not plan any lighting components, ventilation outlets, sprinkler elements, etc. between the longitudinal rails Pos. 3 Radiation shield Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 113 Page 113 of 130 Ceiling Installation with Stationary Upper Body Radiation Shield (Option) Recommendation: ceiling grid of 675 mm *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 Monitor trolley ceiling outlet *4 observe the maximum admissible rail overhang. *5 The ceiling rails (option) for the radation shield, surgery lamp, injector stand (see Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (option)); length when shipped: 4 m, can be shortened. IN1 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option), see Installation Preparations > On-site Power Cables, Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (Option). TR1 Control console cart (option) (see Detail Planning > Control Console Cart (Option)) 114 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 114 of 130 Fig. 59: Ceiling installation with stationary upper body radiation shield (option) *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 Monitor trolley ceiling outlet Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 115 Page 115 of 130 *4 Installation plate (option, see Detail Planning > Stationary Upper Body Radiation Shield, Surgery Lamp (Option)) for Radiation Shield and Surgery Lamp or Injector Stand *5 With a smaller ceiling grid, move the installation plate to the side until it is centered to the ceiling grid. IN2 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option) (see Installation Preparations > On-site Power Cables, Detail Planning > Stationary Upper Body Radiation Shield, Surgery Lamp (Option)). TR1 Control console cart (option) (see Detail Planning > Control Console Cart (Option)) 116 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 116 of 130 Control Console Cart (option) 7.10 The control console cart can be moved on castors. Cable lead-in is via a ceiling outlet. The cable outlet is the same design as the cable outlet for the monitor trolley (see Detail Planning > Monitor Trolley Ceiling Outlet). Fig. 60: Control Console Cart (option) Pos. 1 Clamp rail for ECC Pos. 2 clamp rail for control modules Pos. 3 Cable holder Pos. 4 Corrugated hose 30 (6 m to ceiling outlet) Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 117 Page 117 of 130 Injector Wall Connection Box (Option) 7.11 Fig. 61: Injector wall outlet (option) Pos. 1 Cable duct *1 Mounting points, obtain mounting materials on site. 118 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 118 of 130 Planning Recommendation for the Injector Wall Connection 0 Fig. 62: Planning recommendation for the injector wall connection IW Injector wall connection box (option, injector wall outlet) *1 Orientation point/isocenter for floor stand in the head-end position Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 119 Page 119 of 130 Injector Rack Mount Unit (Option) 7.12 Versions - With table stand: For space reasons, the control cable between the table stand-injector control cannot be lead through the table - or with a fixed ceiling column - or with a movable ceiling column (used in the rail system for the radiation shield) Allowed Injectors: Installation Position see Installation Preparations-> Upper Body Radiation Shield, Sur- gery Lamp, Injector (Option). Injector control unit on an on-site wall console, etc.; if possible, place in the equipment room. Cable lengths: Fig. 63: Pos. 1 ULI (in control room distributor or wall outlet) Pos. 2 Injector control console Pos. 3 Injector control unit Pos. 4 Injector head Injector Comment ANGIOMAT ILLUMENA Table stand: In combination with the tilt or OR table, the injector is released only with the air detector. MEDRAD MARK V PROVIS with ODU connector MEDRAD AVANTA Table installation only 120 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 120 of 130 NOTE The current data for the allowed injectors are described in the SIR PM AX 39...
Injector Stand Cable lengths (m) l 1 l 2 l 3 ANGIOMAT ILLUMENA 5,0 24,0 24,0 MEDRAD MARK V PROVIS 3,0 max.30.5 max.30.5 MEDRAD AVANTA 3,0 20,0 20,0 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 121 Page 121 of 130 Connection of non-Siemens ECGs 7.13 If no Sensis is configured, triggering of exposure can be synchronized to a non-Sie- mens ECG signal (QRS pulse). In addition, the display of the analog ECG signal is pos- sible. Here, the non-Siemens measuring station must meet the following specifications: - galvanic separation of the electrical signals - Cabling meets the requirements defined by EN 60601-1-2 (electro-magnetic interfer- ence...) ECG trigger signal: Input impedance [] Signal level [V] High pulse width [msec] LOW (min...max) HIGH (min...max) (min...max) 1 - 1.0 ... + 0,8 + 2,0 ... + 7,0 1 ...50 Connection of the ECG signal cable is made to the RTC in the system control cabinet. Analog ECG signal: Input impedance [] Amplitude [V] Offset [V] (min...max) (min...max) 100 + 0,6... + 1,3 (typical, 1 Vpp) - 5...+ 5 Adaptation of the ECG signal is made in the system control cabinet on the ECG board in the RTC (A). 122 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 122 of 130 DCS Location per Version 2 7.14 System without the MULTISPACE.F Option 0 Fig. 64: System without the MULTISPACE.F option Pos. 1 P1 cable fixpoint on the ceiling for the floor stand Pos. 2 D1 cable fixpoint on the ceiling for the DCS *1 Dimensions without brackets: Dimensions in brackets: Table longitudinal movement not restricted Table longitudinal movement limited by the room width *2 Dimensions without brackets: for Peri, Tilt and OR tables (including 500 mm safety distance to the wall) Dimensions in brackets: for the basic table *3 Orientation point/Isocenter Floor stand in the head-end position. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 123 Page 123 of 130 System Ceiling Installation 0 Ceiling Installation with Movable Upper Body Radiation Shield (Option) Recommendation: ceiling grid of 675 mm *4 Isocenter in the Left Position *5 Orientation point/Floor fixpoint Patient table 124 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 124 of 130 Fig. 65: Ceiling installation with movable upper body radiation shield (option) Pos. 1 Grid spacing Pos. 2 Do not plan any lighting components, ventilation outlets, sprinkler elements, etc. between the longitudinal rails. Pos. 3 DCS Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 125 Page 125 of 130 Pos. 4 Radiation shield *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 DCS ceiling outlet *4 observe the maximum admissible rail overhang. *5 The ceiling rails (option) for the radation shield, surgery lamp, injector stand (see Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (option)); length when shipped: 4 m, can be shortened. IN1 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option) (see Installation Preparations > On-site Power Cable, Detail Planning > Movable Upper Body Radiation Shield, Surgery Lamp (Option)). TR1 Control console cart (option) (see Detail Planning > Control Console Cart (Option)) 126 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 126 of 130 Ceiling Installation with Stationary Upper Body Radiation Shield (Option) Fig. 66: Ceiling installation with movable upper body radiation shield (option) Pos. 1 Do not plan any lighting components, ventilation outlets, sprinkler elements, etc. between the longitudinal rails. Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 127 Page 127 of 130 Pos. 2 DCS Pos. 3 Radiation shield *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor stand ceiling outlet (see Detail Planning > Floor Stand Ceiling Outlet) *3 DCS ceiling outlet *4 observe the maximum admissible rail overhang. *5 Installation plate (option, see Detail Planning > Stationary Upper Body Radiation Shield, Surgery Lamp (Option)) for Radiation Shield and Surgery Lamp or Injector Stand *6 With a smaller ceiling grid, move the installation plate to the side until it is centered to the ceiling grid. IN2 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option) (see Installation Preparations > On-site Power Cables, Detail Planning > Stationary Upper Body Radiation Shield, Surgery Lamp (Option)). TR1 Control console cart (option) (see Detail Planning > Control Console Cart (Option)) 128 Detail Planning AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 128 of 130 Ceiling Installation with a second DCS (Option) 7.15 Siemens AXA4-100.891.01.06.02 AXIOM Artis FC/FA 01.08 CS CD AX Detail Planning 129 Page 129 of 130 Fig. 67: Ceiling installation with second DCS (option) Pos. 1 Grid spacing Pos. 2 DCS Pos. 3 Radiation shield *1 Orientation point/isocenter for floor stand in the head-end position *2 Floor Stand Ceiling Outlet (see) *3 DCS ceiling outlet *4 observe the maximum admissible rail overhang. *5 The ceiling rails (option) for the radation shield, surgery lamp, injector stand (see); length when shipped: 4 m, can be shortened. IN2 Injector stand installed on the injector ceiling column in the radiation shield rails (option). SU Surgery lamp (option) TR1 Control Console Cart (option) 130 Changes to Previous Version AXIOM Artis FC/FA AXA4-100.891.01.06.02 Siemens 01.08 CS CD AX Page 130 of 130 8- 8Changes to Previous Version Chapter Section Changes Complete document revised editorially. Cover page General Information Room Planning Installation Preparations System Connections Technical Data Transport and Storage Detail Planning Changes to Previous Version