Reduced Bias Voltage Lead Sets

196-3477-XX. Eight channel lead set with 100 Ω


series termination. Minimum pin spacing of 2 mm
(0.079 in).
196-3478-XX. Single channel lead set with 100 Ω
series termination. Minimum pin spacing of 2 mm
TLA700 Series (0.079 in).
P6417 & P6418 The reduced bias voltage lead sets contain a 20 kΩ
Eight-channel resistor connected between the signal lead and
Probe Accessory lead set with ground. Using these lead sets with the P6417 and
Instructions barrel connectors P6418 probes and the logic analyzer will reduce
the pull-up voltage from 2.2 V to 1.1 V. The
loading will change from 20 kΩ to 10 kΩ.
070-9948-02
Accessory Kits
The following accessory kits are available for the
P6417 and P6418 Probes.
2-to-1
KlipChip
connector P6417 & P6418 Probe accessories kit (020-2198-XX)
connector
Use these connectors to connect the lead sets to
square pins or KlipChip connectors.
2 Eight-channel lead sets (barrel connectors)
1 Single-channel lead set (barrel connector)
Figure 1: Lead set connections (barrel connectors shown) 20 SMT KlipChip connectors

P6417 & P6418 Probe 34-channel probe interface


kit barrel connectors (020-2199-XX)
Use these connectors to connect the lead sets to
square pins or KlipChip connectors.
4 Eight-channel lead sets (barrel connectors)
2 Single-channel lead sets (barrel connectors)
www.tektronix.com 1 Set of twelve 2-to-1 signal/ground lead sets
(barrel connectors)
1 Set of four 4-to-1 signal/ground lead sets
*P070994802* (barrel connectors)
070994802
P6417 & P6418 Probe 34-channel probe interface kit
Lead Set Connectors mini-PV connectors (020-3000-XX)
These instructions support several probe accessory Use these connectors to mass-terminate probe lead
kits, which contain lead set connectors for the sets into a mini-PV terminal connector. One 2 ¢
P6417 and P6418 Probes. The lead set connectors 20 mini-PV terminal connector can accommodate
simplify connections between the logic analyzer three eight-channel lead sets.
and your target system. See Figures 1 and 2 for The mini-PV terminal connectors are designed for
examples. use with 0.025-inch square pins set on 0.1 in
For further information about connecting the Figure 2: Single-channel lead set connection centers.
probe to the target system, refer to the TLA700 4 Eight-channel lead sets (mini-PV connectors)
online help or to the P6417 & P6418 Logic 2 Single-channel lead sets
Analyzer Probe Instructions. Lead Sets
(mini-PV connectors)
The following lead sets are available for the 1 Set of twelve 2-to-1 signal/ground lead sets
P6417 and P6418 Probes. (mini-PV connectors)
1 Set of four 4-to-1 signal/ground lead sets
General Purpose Lead Sets (mini-PV connectors)
196-3431-XX. Eight channel lead set with 100 Ω 1 Set of assorted mini-PV terminal connectors
series termination. Minimum pin spacing of 0.1 in.
196-3432-XX. Single channel lead set with 100 Ω Safety Summary
series termination. Minimum pin spacing of 0.1 in. Refer to the General Safety Summary in the
Tektronix Logic Analyzer Family User Manual for
Enhanced General Purpose Lead Sets applicable safety information.
196-3476-XX. Eight channel lead set with 100 Ω
series termination. Minimum pin spacing of 2 mm
(0.079 in).
196-3479-XX. Single channel lead set with 100 Ω
series termination. Minimum pin spacing of 2 mm
(0.079 in).
Connection Procedure 64 mm
(2.5 in)
Use the connection procedure on the next page for
the barrel connectors or the mini-PV connectors. 18 mm
Steps that are specific to only one type of (0.7 in)
connector are noted. Mini-PV lead
To avoid static discharge damage, use a grounding (crimped side
outward)
wrist and foot strap, and follow appropriate
antistatic precautions.
1. Power off the target system.
2. Connect the logic analyzer probe lead sets to
22 mm
the accessory lead set connectors. See Mini-PV (0.9 in)
Figure 3. Lead should terminal
CAUTION. Be sure that the logic analyzer snap into place. connector
podlets are connected to the 1¢ 8 signal
row and not to the ground row. Connecting
signals to the ground row can result in Figure 4: Mini-PV leads and terminal connector 7 mm
damage to the logic analyzer or the system (0.3 in)
under test.
Characteristics Figure 6: Eight-channel lead set (mini-PV connectors)
3. For mini-PV lead set connectors, insert the The characteristics listed below apply to all lead
leads into the mini-PV terminal connector. See sets and kit connectors unless otherwise noted.
Figure 4. A properly inserted lead snaps into
place and cannot be withdrawn. An Table 1: Characteristics 152 mm
improperly installed lead slides out easily. (6.0 in)
Recommended TTL and CMOS levels only.
If a lead is inserted in the wrong location, the usage Ground leads should be connected to 12 mm (0.46 in)
lead can be removed by gently lifting the ground of system under test. 18 mm (0.7 in) *
small plastic tab until the lead can be with-
drawn. The tab can be lifted with a small pick Not recommended for signals with
or pointed knife tip. Do not lift the tab too far edge rates > 1 V/ns. 5 mm (0.2 in)
or too often, or the tab will break off or Maximum clock 50 MHz 8 mm (0.3 in) *
permanently deform. speed 21 mm
4. If necessary, use the 2-to-1 or 4-to-1 AC loading < 5 pF per channel as seen by the (0.9 in)
connectors to connect multiple leads to a device under test (plus podlet)
single pin on the system under test. Figure 1 1.96 mm (0.077 in) *
on page 1 shows an example of a 2-to-1 DC loading None (General purpose and enhanced 3 mm
connector in use. general purpose lead set) (0.1 in)
5. Insert the lead set connector(s) onto the pins or
10 kΩ (Reduced bias voltage lead set 2.5 mm
on the system under test. when connected to a P6417 or P6418 1.96 mm
(0.1 in)
6. Power on the system under test. probe) (0.077 in) *
Looking at the end of the lead set
Ground lead Termination Each signal lead on the eight-channel
(black connector; terminates Channel leads (020-2198-XX lead set contains a 100 Ω series Figure 7: Single-channel lead set (barrel connectors)
in connector ground bus) and 020-2199-XX termination near the end of the barrel
only) connector to minimize signal
reflections.
64 mm
Dimensions See Figures 5 through 8. (2.5 in)
14 mm
Figures 5 through 8 show the dimensions of the (0.6 in)
lead set connectors. Both top and side views are
shown.
152 mm 5 mm
(6.0 in) (0.2 in)
25 mm
(1.0 in)

3 mm
(0.1 in)
Ground pins Signal pins
(bottom row) (top row) Figure 8: Single-channel lead set (mini-PV connectors)
26 mm (1.02 in)
27 mm (1.05 in) *
Figure 3: Eight-channel lead set connector * Physical characteristics for enhanced general purpose
21 mm
(0.9 in) lead sets and for reduced bias voltage lead sets

7.6 mm (0.3 in)


8.6 mm (0.34 in) *

Figure 5: Eight-channel lead set (barrel connectors)

Copyright  Tektronix, Inc.

S-ar putea să vă placă și