Documente Academic
Documente Profesional
Documente Cultură
Document No.
We here by confirm that all contents (including changes) of the operational qualification procedures described in this
document have been approved in the presence of the performer, reviewer and the manager.
Print / /
Title
Certificate No.
(Shimadzu-certified Performer only)
Company
Print / /
Title
Company
Print / /
Title
Company
Operational Qualification Table of Contents
Table of Contents
1. Definitions......................................................................................................................... 3
1-1 Documentation ..............................................................................................................................3
1-2 Scope .............................................................................................................................................4
1-3 Entries and Revisions ....................................................................................................................4
2. Preliminary Checks ......................................................................................................... 7
2-1 Reviewer Confirmation .................................................................................................................7
2-2 Performer Identification & Approval .............................................................................................7
2-3 Planning Items to be Evaluated in Operational Qualification........................................................7
2-4 Calibration Certificate and Instruments Used ................................................................................8
2-5 Other Tools Used ...........................................................................................................................9
3. Operational Qualification Outline ............................................................................... 10
3-1 Outline of Inspection Procedures................................................................................................. 10
4. Operational Qualification Protocol .............................................................................. 14
4-1 UVProbe/Personal Computer and Peripherals ............................................................................. 14
4-2 Spectrophotometer ....................................................................................................................... 25
4-2-1 UV-1700 UV-VIS Spectrophotometer .............................................................................. 25
4-2-2 UV-1800 UV-VIS Spectrophotometer .............................................................................. 34
4-2-3 UV-1601PC/UV-1650 UV-VIS Spectrophotometer ......................................................... 44
4-2-4 UV-2401PC/UV-2450 UV-VIS Spectrophotometer ......................................................... 53
4-2-5 UV-2501PC/UV-2550 UV-VIS Spectrophotometer ......................................................... 62
4-2-6 UV-3600 UV-VIS Spectrophotometer .............................................................................. 71
4-3 Syringe Sipper N/CN ................................................................................................................... 81
4-4 Automatic Sample Changer ASC-5 ............................................................................................. 87
4-5 Elution Test 6-Cell Holder DIS-1200 .......................................................................................... 89
4-6 Elution Test 8-Cell Holder DIS-8 ................................................................................................ 90
4-7 Cell Positioner CPS-240A/B ....................................................................................................... 91
4-8 Thermoelectrically Temperature Controlled Cell Holder TCC-240A ......................................... 95
4-9 Screen Copy Printer HCP-1C ...................................................................................................... 96
4-10 Screen Copy Printer DPU-414................................................................................................... 97
4-11 Screen Copy Printer MPU ......................................................................................................... 98
4-12 ESC/P Compatible Screen Copy Printer .................................................................................... 99
4-13 PCL Compatible Screen Copy Printer .....................................................................................100
4-14 Time Course (Multi λ) Program Pack ......................................................................................101
4-15 DNA/Protein Quantitation Program Pack................................................................................102
4-16 Protein Quantitation Program Pack .........................................................................................103
4-17 CLASS-Agent..........................................................................................................................104
4-18 Validation Software Package ................................................................................................... 112
4-19 Color Measurement Software COL-UVPC ............................................................................. 113
4-20 Film Thickness Measurement Software FLM-UVPC ............................................................. 115
4-21 Invalidating Inspection Data .................................................................................................... 117
5. Change List.................................................................................................................... 118
1. Definitions
1-1 Documentation
1. This document stipulates the procedures and the documentation of the Operational Qualification (OQ) of
Shimadzu Corporation’s UV-VIS Spectrophotometers. When all procedures are executed and all items pass
the inspection, it is verified that the system operates to satisfy the intended purpose. The contents of this
document must be checked and approved by the Performer, Reviewer and Manager before implementing
the Operational Qualification.
2. The following are established as the responsible parties in the execution of inspections.
· Performer
The Performer is the individual who actually performs the procedures stipulated in the Installation
Qualification Protocol and the Operational Qualification Protocol.
The Performer shall be an individual who, having been trained by the Quality Assurance Department of the
Analytical & Measuring Instruments Division of Shimadzu Corporation, has been certified as an IQ/OQ
inspector by the General Manager of the Quality Assurance Department, or an individual recognized by the
Manager to be equipped with the ability to conduct this inspection.
· Reviewer
The Reviewer is an individual designated by the Manager charged with responsibility for the system.
The Reviewer verifies the contents of the Installation Qualification Protocol and Operational Qualification
Protocol. In addition, the Reviewer shall attend the inspection procedures together with the Performer, and
evaluate the procedures.
The Manager can serve concurrently as the Reviewer.
· Manager
The Manager is the individual responsible for administering the entire UV-VIS system being inspected.
As well as verifying the contents of the Installation Qualification Protocol and Operational Qualification
Protocol, the Manager shall also grant final approval to the inspection procedures.
3. The Performer shall first check the content of this document. After checking all items, the Performer
shall sign and date the cover sheet of this document and submit this document as the Operational
Qualification Protocol report. (OQ must be performed by a trained and knowledgeable individual in
accordance with 21 CFR Part 211 Section 211.25 (cGMP, USA promulgated 29 September 1978)).
4. The Reviewer shall first check the content of this entire document. After checking all items, the Reviewer
shall sign and date the cover sheet of this document as approval of the Operational Qualification Protocol.
5. The Manager shall first check the content of this entire document. After checking all items, the Manager
shall enter a unique document control number in the Document No. column at the top right of the cover sheet,
and then sign and date the cover sheet of this document as final approval of the Operational Qualification
Protocol.
7. This document contains important information and any form of duplication, photography, or reproduction
without the permission of the General Manager of the Quality Assurance Department, Analytical &
Measuring Instruments Division, Shimadzu Corporation, is prohibited.
1-2 Scope
The Operational Qualification Protocol shall include the following.
· Spectrophotometer Operation Check
The spectrophotometer operation and some functions will be verified and the results will be checked against
pass criteria specified in this document.
· Personal Computer and Special Accessories Operation Check
Operation check will be conducted for the personal computer that controls the spectrophotometer, its
peripherals and installed special accessories.
· Software and firmware check
The Operational Qualification also checks the software and firmware provided.
The firmware checking is conducted based on version display and the Software Certificate of Compliance.
The software and firmware must be properly managed and change procedures must be properly clarified.
Analysis methods and data acquisition systems such as networks are not verified in the OQ of this system. These
must be separately verified by the customer.
5. Revising entries
If the need to alter the contents of an entry arises, change the entry in accordance with “3. Deletion and
correction of mistakenly filled items”, and then write the reason for the change in the Change List of this
document.
When adding a new description, write it in the margins of the relevant page, and write “Change List No. _”,
the date and signature. Then, write the reason for the change in the Change List.
8. Attaching data
Attached data are added by assigning page numbers as “A, B, C…” for each inspection item. If, for example,
there is a total of 3 sheets for attachment A, notate the pages as “73-A-1/3 - A-3/3”, ensuring that no
numbers are omitted. The Performer shall date and sign the attached data.
9. Other
Every item entered into the Change List shall be approved by the Reviewer and Manager.
Uncertain value
(underlined) Uncertain value
(underlined)
Uncertain value
(underlined)
The value 1.25 was measured only to
the 1/100 digit position.
Consequently, the result becomes
As the digit “0” already contains
5.62 (to three significant digits).
uncertainty, the result becomes 13.0
(to three significant digits).
<References>
1. Writing a value as 0.00123000 represents six significant digits.
2. Subtracting 1.230 (four significant digits) from 1.231 (four significant digits) produces the result 0.001,
which is one significant digit (1 x 10-3).
3. Dividing 2.462 (four significant digits) by 1.231 (four significant digits) produces the result 2.000, which
is four significant digits (not 2).
If the number of significant digits of the measured result exceeds the number of significant digits of the pass
criteria, round the final result down to the number of significant digits of the pass criteria.
2. Preliminary Checks
* Standard filters
This is used when verifying the “photometric accuracy” and “photometric repeatability” of the
spectrophotometer.
When conducting evaluation according to the instrument specifications, either NIST (National Institute of
Standard and Technology) filters or semi-standard optical filters that have been calibrated to the NIST
filters must be used.
In addition, these standard filters shall have an accuracy of ±1% (value relative to transmittance) as a
guaranteed performance specification when conducting evaluation based on criteria specified by the
Japanese Pharmacopoeia, and the filters shall be calibrated using the slit widths that can be set in the
instrument being inspected.
* Thermometers
The thermometer used to measure the CPS-240A/B and TCC-240A set temperature accuracy shall have
a precision rating of ±0.4°C in the 10°C to 60°C temperature range required for the unit inspection as a
guaranteed performance specification.
The thermometer used to measure the CN syringe sipper temperature accuracy shall have a precision
rating of ±1°C at the 37°C required for the unit inspection as a guaranteed performance specification.
* Thermocouples
The thermocouples used to measure the CN type syringe sipper temperature control accuracy shall
conform to JIS Class 1 and have a precision rating of ±0.6°C at the 37°C required for the unit inspection
as a guaranteed performance specification.
* Electronic balance
This is used only when conducting the operation qualification tests for the Syringe Sipper N/CN.
Requires a calibrated balance able to weigh down to 0.001g.
The equipment and solvent below are used only for the operational qualification of the N/CN syringe sipper.
They are not needed if no syringe sipper is installed.
* Equipment
Beakers (2): capacity approx. 50 mL
* Solvent
Distilled water: 2L
The equipment and reagents below are used only for the operational qualification of specific devices. They are
not needed if applicable devices are not installed.
* Reagents
Potassium dichromate K2Cr2O7, 50mg: for N/CN syringe sipper
Perchloric acid HClO4(60%), 200 mg: for N/CN syringe sipper
* Other
Semi-micro cell [P/N 200-66501]: for N/CN syringe sipper
4. ASC-5
5. DIS-1200
6. DIS-8
8. TCC-240A
9. HCP-1C
10. DPU-414
11. MPU
17. CLASS-Agent
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
UVProbe/Personal Computer and Peripherals. Refer to the Instruction Manual for details about the operation
procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
User Registration of Performer
If the UVProbe GLP or security mode is used, the system administrator must complete user registration for the
Performer who conducts the work (OQ User) before this Operational Qualification is conducted.
Simultaneously, register the Windows user ID to allow the OQ User to log in to Windows as a normal operator.
If the Class-Agent operation software is not used, user registration and authority setting is required for
UVProbe only.
Disable the Performer's user registration when all inspections are complete.
Login ID UVProbe
User name Password CLASS-Agent Authority Group
(User ID) Authority Group
Operator,
OQ_USER_1 OQ_Operator ******* Operator
but with additional “Approval”
authority for Agent Manager
OQ_USER_2 OQ_Approver ******* Administrator Administrator
The administrator shall issue the initial user password and inform the user.
● UVProbe Login
Use the User ID “OQ_USER_2” to log in UVProbe if not specified otherwise.
Test 1-1
(Item) Computer boot-up check
(Outline) Checks that the computer boots up and runs normally. This test uses the self-test during
computer start-up to ensure that no memory errors occur.
(Procedure) 1) Turn on the PC.
2) Check that Windows starts up normally.
* If more than one OS is installed in the personal computer, follow the test procedure above to
check that each OS boots up normally.
Test 1-2
(Item) Time and date check
(Outline) Checks that the pc date and time are correct.
(Procedure) 1) Open the Windows Control Panel.
2) Double-click [Date/Time] in the Control Panel to display the Date/Time Properties.
3) Check that the date and time are correct. Check the time against an accurate clock, such
as the telephone time service.
(Criteria) Date and time are correct.
Test 1-3
(Item) Hard disk drive check
(Outline) Checks that the hard disk operates correctly.
(Procedure) In order to perform this test, the user must log into Windows with Administrator authorization.
In addition, when performing step 5) of the procedure, do not open or close any files.
Test 1-4
(Item) Print driver check
(Outline) Prints the print driver test page and checks the version number.
(Procedure) 1) Select [Printers] from the Windows Control Panel to display the printers used.
2) Right-click the printer used and select [Properties].
3) Click the [Print Test Page] button on the Properties [General] tab to print the test page.
* If more than one OS is installed in the personal computer, follow the test procedure above
to print the test page under each OS.
4) Check that the test page printed correctly and record the print driver version number
printed in it.
Test page prints correctly.
(Criteria)
The driver version number matches the version number on record.
Test 1-5
(Item) Network connection test
This inspection is required only for network connected PCs.
(Outline) Checks the network connection between the inspected computer and the reference computer
(e.g., the server computer).
(Procedure) 1) Run the Command Prompt and enter “C: ¥>PING***.***.***.***” (substituting the IP
address of the server computer or other reference computer for “***.***.***.***”).
2) Data communication is conducted with the designated computer and the results are
displayed.
3) Check that no packets were lost.
4) Maximize the screen. Hold down the [Alt] key and press [PrtSc] to copy the screen.
Paste the content into the Windows Paint accessory and print it out.
(Criteria) No packets lost.
Test 2-1
(Item) Software version check
(Outline)
Checks the version number of the UVProbe software and Shimadzu User Authentication Tool.
(Procedure) 1) Check the version number printed in the test results from the alteration test in Test 2-2.
2) Check that the version number matches the version number on record.
(Criteria)
Version number matches the version number on record.
Test 2-2
(Item) Program alteration check
(Outline) Checks that no modifications have occurred to the contents, date, or checksum of the files
(*.dll, *.exe, *.ocx, etc.) in the UVProbe or Shimadzu User Authentication Tool folders.
(Procedure) UVProbe Alteration Check Procedure
1) Click on the Windows [Start] button and select [Programs] -> [Shimadzu] -> [UVProbe –
Check the Program Files] to run the alteration check program.
2) When the login screen appears, enter the user ID “OQ_USER_2”.
3) When the [Check the Program] screen appears, select (highlight) “UVProbe VER.XXX”
and click the [Check] button.
* If the supplemental disk is installed, 2 UVProbe lists are displayed on the screen. In this
situation, be sure to select the one in which the version numbers match.
Example) If Supplemental Disk Ver2.00SU1 is installed,
-> select “UVProbe Ver.2.00SU1”.
4) Click the [Run] button to run the program alteration check.
5) When the check is complete, click the [NotePad] button.
6) The results are displayed as a text file. Ensure that all items are OK and print the results.
Test 2-3
(Item) Shimadzu User Authentication Tool connection check
(Outline) Checks the Shimadzu User Authentication Tool connection using database access and log
registration.
(Procedure)
1) Open the Windows Control Panel and double-click on the Shimadzu User Authentication
Tool icon to open the main menu.
2) Record the server name displayed at the bottom of the main menu and confirm that the
connected database is correct.
* Conduct steps 3) and 4) below after conducting Test 2-4-4.
3) Log in to [Log Display] as “OQ_USER_2” from the main menu. Set the test date and test
computer name in [Search Conditions]. Search the log and print out the results.
4) Confirm that the user lockout generated in Test 2-4-4 is registered and that the software
name is UVProbe.exe.
Test 2-4
(Item) 2-4-1 Password log management function check
2-4-2 Number of password characters check
2-4-3 Failed login detection
2-4-4 User lockout function check
2-4-5 Failed login record
(Outline) Checks that the security functions operate, including restrictions on the number of password
characters.
Before conducting this test, verify that the configuration settings of Shimadzu User
(Preparation)
Authentication Tool are as listed below. Change the settings for items (3) and (4) as indicated
below.
Item Setting
(1) Minimum password length (Check only; set to non-zero value)
(2) Permission to use previous No check
password
(3) User lockout time 1 minute
(4) User login failures (Check only; set to non-zero value)
(5) PC Login failures Value the same as or larger than
“User login failures” setting
(Procedure) 1) Start up UVProbe, and log in as the “OQ_USER_1” test user.
2) Select [Security] in the [Edit] menu.
3) Enter the current password, and also enter the same password as the new password.
4) Confirm that the new password is not accepted.
5) Enter the current password and enter a new password that is shorter than the minimum
number of characters set for the Shimadzu User Authentication Tool software.
6) Confirm that the change is not accepted, as the password is too short.
7) Enter a password equal in length to the minimum number of characters and confirm that the
change is accepted.
8) Close UVProbe.
9) Start up UVProbe again, and enter “OQ_USER_1” as the test user ID.
10) Enter a password that is not set and click [OK].
11) Confirm that login is not permitted and a message is displayed warning that the password is
incorrect.
12) Repeat steps 9) to 11) for the number of times designated for “User login failures”.
13) After the designated numbers of login failures, confirm that the “OQ_USER_1” is locked
out and cannot log in for the designated lockout time.
14) After the designated lockout time elapses, start up UVProbe again. Enter “OQ_USER_1” as
the login ID and correctly enter the set password.
15) Confirm that login is possible.
16) Close UVProbe, and then log in again as the “OQ_USER_2” test user.
17) Select [Security] in the [Edit] menu.
18) Select the [Event Log] tab, and confirm that the login failures in the procedure above and
the corresponding times appear in the displayed log. Close UVProbe.
19) If any settings in the Shimadzu User Authentication Tool configuration were changed in the
Preparation above, return them to their original values.
(Criteria) 2-4-1 Duplicate password is not accepted.
2-4-2 Change to a password shorter than the minimum number of characters is not
accepted.
Change to a password with the minimum number of characters is accepted.
2-4-3 Login prohibited with an incorrect password but possible with the correct
password.
2-4-4 User locked out after the designated number of login failures.
Login prohibited for the designated lockout time.
2-4-5 Login failures recorded in the log.
Test 2-5
(Item) User access rights check
(Outline) Confirms that software use is restricted according to differences in registered user access rights.
This procedure is performed only when UVProbe is run in Security or GLP mode.
(Procedure)
1) Start UVProbe.
2) Log in with the user ID “OQ_USER_2”
3) Select [Method] from the [Edit] menu, and confirm that the content of the analysis method
can be changed. Close UVProbe.
4) Start UVProbe again, and log in as “OQ_USER_1”.
5) Confirm that the [Method] command in the [Edit] menu is disabled (grayed out).
(Criteria) OQ_USER_2 can edit the analysis method, but it is disabled for OQ_USER_1.
Test 2-6
(Item) Lock function check
(Outline) Checks that during UVProbe startup, the manual access lock function operates normally. This
procedure is performed only when UVProbe is run in GLP mode.
(Procedure)
Perform this procedure as described below. For detail on operation, refer to the UVProbe
Instruction Manual (Tutorial).
1) Start up UVProbe
2) Select [Lock] from the [Window] menu.
3) The Login screen is displayed. Confirm that the UVProbe menus cannot be accessed.
4) Log in using the “OQ_USER_2” password to close the Login screen, and confirm that the
UVProbe menus have become accessible.
(Criteria) Login screen is displayed.
UVProbe can be used after entering password.
Test 2-7
If data from this computer is being registered to the CLASS-Agent database, certain check and implementation
items must be performed beforehand. Refer to cautions described below.
1) Close all of the data loaded in UVProbe, and delete the data file copied in “Preparation”
above using Explorer. In addition, if any of the UVProbe automatic registration settings
were changed, return the settings to their original values.
Test 2-7-1
(Item) Peak detection / point pick check
(Outline) Checks that peak detection and peak pick functions of the Spectrum module operate normally.
(Procedure)
1) Start up UVProbe. (Spectrophotometer need not be connected to PC.)
2) Using the [Open] command in the [File] menu of the Spectrum module, load the
following data.
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Test¥Anthracene.spc
3) Using the [Save As] command in the [File] menu, save the data to the same folder as
“Anthra.spc”.
4) Select [Peak Detection] from the [Data Processing] menu.
5) Display the shortcut menu by right-clicking on the peak detection table, select
[Properties], and in the [General] page, set the peak detection parameters as follows
6) Select [Settings] in the [View] menu, and for the number of decimal places to be
displayed for the X and Y axes, set 1 place and 4 places, respectively.
7) Using [Save As] in the [File] menu, save the data as “Anthra1.spc”.
8) Confirm that the detected peaks and valleys match those shown below.
11) Using [Save As] in the [File] menu, save the data as “Anthra2.spc”.
12) Print the report using the report template (OQ_ProbeTest3.rpt) copied to the same folder
at step 2). If the point pick table or spectrum data set log object is not contained in
“OQ_ProbeTest3.rpt”, insert them using the report generator.
13) Verify that the report records the file name changes from steps 3), 5), 7), 10) and 11), as
well as the detection parameter changes, date and user name in the data set log.
(Criteria) Peak detection and point pick results match in confirmation steps 8) and 10).
File name changes in steps 3), 5), 7), 10) and 11) and the peak detection parameter
changes are recorded in the data log along with the date and user name*1.
*1) The user name of the user logged into UVProbe is recorded as the data set log user name.
Therefore, if the UVProbe application mode is Normal, the user name is left blank.
Test 2-7-2
(Item) Quantitation function check
(Outline) Checks that the quantitation function in the Photometric module operates normally.
(Procedure)
1) Start up UVProbe. (Spectrophotometer need not be connected to PC.)
2) Select [Photometric] in the [Window] menu.
3) Select [Method] in the [Edit] menu, in the [Wavelengths] page, register Measurement
Wavelength: 452 nm, and in the [Calibration] page, select WL1:WL452.0. (Leave the
other settings at their default values.)
4) In the [Measurement Parameters (Standard)] page, change the [Data Acquired By] setting
to “User Entry”. (Leave the other settings at their default values.)
5) Complete the measurement method editing, leaving the file name as the default name.
6) Select [Settings] in the [View] menu, and for the number of decimal places to be
displayed in the Standard Sample and Sample tables, set 5 places and 3 places,
respectively.
7) Enter the sample IDs, concentrations and absorbances in the Standard Sample table as
follows.
Sample ID Concentration Absorbance(WL452.0)
Std_1 1 0.03314
Std_2 2 0.06407
Std_3 4 0.12666
Std_4 10 0.33617
8) Select [Standard Curve Statistics] in the [Graph] menu, and place check marks for
“Equation” and “Correlation Coefficient”.
9) Verify that the calibration curve function and correlation coefficient displayed at the
bottom of the calibration curve graph match the values shown below.
Calibration curve function : Y=0.03383x - 0.00379
Correlation coefficient r2 : 0.99935
10) Select [Save As] in the [File] menu, and after selecting [Standard file (*.std)] as the file
type, save the file as “OQ_Test1-3-2.std”.
11) Load the following file using the [Open] command in the [File] menu.
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Data¥Ver2Data¥B-carotene.unk
12) Import the standard sample file created in step 10) using the [Open] command in the
[File] menu.
13) Change the file name as follows, and save it as a photometric file.
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Test¥OQ_Test1-3-2.Pho
14) Verify that the sample concentrations in the sample table match those listed below.
15) Open the following file using the Report Generator, and print the report after inserting the
photometric log and photometric summary objects on the second page (Add New).
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Reports¥Standard Curve.rpt
16) Verify that the printed report records the standard sample file importation, file save
content, date and user name implemented in steps 12 and 13) above.
17) Return the settings and objects that were changed in steps 6) and 15) to their original
state.
(Criteria) The calibration curve function and correlation coefficient, and quantitation results
match in steps 9) and 14).
The standard sample file importation and file save content of steps 12) and 13) are
recorded in the photometric log along with the date and user name.
Test 2-8
(Item) Communication check
(Outline)
Checks that spectrophotometer initialization settings are performed using UVProbe, and that
communication operates normally.
(Procedure)
Perform the procedure as described below.
For details on operation, refer to the UVProbe Instruction Manual (Tutorial).
4-2 Spectrophotometer
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the UV-1700
Ultraviolet-Visible Spectrophotometer. The procedures shown here employ the Validation functions provided
as standard with the UV-1700. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
The UV-1700 incorporates a standard semi/full-automatic inspection validation function for checking the
instrument performance. Using that validation function, the procedures for performing tests 1 - 11 are shown
here.
The following Validation function test items are used for performing the tests.
Test 1: Firmware version check ….Checks the ROM version printed in results printout.
Test 2: Initial setup check …. Full-automatic test “7. Initialization results record”
Test 3: Wavelength accuracy check …. Full-automatic test “1. Wavelength accuracy”
Test 4: Wavelength setting repeatability check …. Full-automatic test “2. Wavelength repeatability”
Test 5: Photometric accuracy check …. Semi-automatic test “1. Photometric accuracy”
Test 6: Photometric repeatability check …. Semi-automatic test “2. Photometric repeatability”
Test 7: Baseline flatness check …. Full-automatic test “5. Baseline flatness”
<Printer Selection>
9) In the UV-1700 mode selection screen, press the [F3] key (Mainte.).
10)In the instrument’s Maintenance screen, select [1. Validation].
11)In the Validation menu screen, press the [F4] key (Test Settings).
1) When the test settings screen is displayed, press the [2] key, to set [2. Auto print] to
“ON”.
2) Press the [RETURN] key to return to the Validation menu screen.
(Criteria)
Test 1: The version number matches the controlled version number.
Test 2: Each item of the initial settings is completed as “OK”.
Test 3: Measurement deviation of D2 emission lines 656.1nm and 486.0nm is within
±0.3nm.
Test 4 :
Instrument specifications …. Deviation from average of 3 values is within ±0.1nm
Pharmacopoeia …. Deviation from average of 3 values is within ±0.2nm
Test 5:
Instrument specifications
Absorbance near 1Abs : Deviation from reference value is within ±0.004 Abs
Absorbance near 0.5Abs: Deviation from reference value is within ±0.002 Abs
Note) The instrument specifications apply only to specifications determined with a
NIST-traceable visible region filter, or a semi-standard optical filter whose value is
calibrated to a NIST filter.
Pharmacopoeia….Deviation from standard value is within ±0.008 Abs
Test 6:
Instrument specifications
Absorbance near 1Abs : Deviation from average of 3 values is within ±0.002 Abs
Absorbance near 0.5Abs : Deviation from average of 3 values is within ±0.001 Abs
Pharmacopoeia
Absorbance near 1Abs : Deviation from average of 3 values is within ±0.004 Abs
Absorbance near 0.5Abs : Deviation from average of 3 values is within ±0.002 Abs
Test 7: Baseline flatness is within ±0.002 Abs
* For the measurement methods and pass criteria values in Tests 3 - 11, refer to Additional
Information from page 31.
1) In the UV-1700 mode selection screen, press the [F3] key (Mainte.).
2) In the instrument Maintenance screen, select [1. Validation].
3) In the Validation menu screen, press the [1] key (Semi-auto Items).
4) Enter the set password.
5) The Semi-auto items setting menu screen is displayed, and the items that are set for
inspection will all be highlighted. At this time, if [3. Stray light] inspection is set to be
performed, perform steps 6) - 8).
6) In the Semi-auto items setting menu screen, press the [3] key (Stray Light).
7) In the Stray Light conditions setting screen, press the [1] key (Inspection) to toggle the
setting to “No”.
8) Press the [RETURN] key to return to the Semi-auto items setting menu screen.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance tolerance at
440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance range of
0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the rounding
errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted for the
photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion of
(reference value ± 0.008 Abs) is adopted.
The number of measurements and the measurement method for Test 6 conforms with the 15th
Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in order
to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of the
average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
In the UV-1700, from the viewpoint of noise exclusion, the measured baseline is subjected to
smoothing, and the absorbance difference with the 0 Abs line is set as the baseline curvature
(flatness).
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the UV-1800
Ultraviolet-Visible Spectrophotometer. The procedures shown here employ the Validation functions provided
as standard with the UV-1800. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
The UV-1800 incorporates a standard semi/full-automatic inspection validation function for checking the
instrument performance. Using that validation function, the procedures for performing tests 1 - 11 are shown
here.
The following Validation function test items are used for performing the tests.
Test 1: Firmware version check ….Checks the ROM version printed in results printout.
Test 2: Initial setup check …. Validation Settings “3. Print init. status”
Test 3: Wavelength accuracy check …. Full-automatic test “WL accuracy D2”
Test 4: Wavelength setting repeatability check …. Full-automatic test “WL repeatability D2”
Test 5: Photometric accuracy check …. Semi-automatic test “Photo accuracy”
Test 6: Photometric repeatability check …. Semi-automatic test “Photo repeatability”
Test 7: Baseline flatness check …. Full-automatic test “Baseline flatness”
<Printer Selection>
9) In the UV-1800 mode selection screen, press the [F3] key (Mainte.).
10) In the instrument’s Maintenance screen, select [1. Validation].
11) In the Validation menu screen, press the [F4] key (Settings).
12) When the settings screen is displayed, press the [1] key, to set [1. Auto print] to “ON”.
13) Press the [3] key to set [3. Print init. status ] to “ON”.
14) Press the [RETURN] key to return to the Validation menu screen.
(Criteria)
Test 1: The version number matches the controlled version number.
Test 2: Each item of the initial settings is completed as “OK”.
Test 3:
Instrument specifications Measurement deviation of D2 emission lines 656.1nm is
within ±0.1nm.
Measurement deviation of D2 emission lines 486.0nm is
within ±0.3nm.
Pharmacopoeia Measurement deviation of D2 emission lines 656.1nm
and 486.0nm is within ±0.3nm.
Test 4 :
Instrument specifications …. Deviation from average of 3 values is within ±0.1nm
Pharmacopoeia …. Deviation from average of 3 values is within ±0.2nm
Test 5:
Instrument specifications
Absorbance near 1Abs : Deviation from reference value is within ±0.004 Abs
Absorbance near 0.5Abs: Deviation from reference value is within ±0.002 Abs
Note) The instrument specifications apply only to specifications determined with a
NIST-traceable visible region filter, or a semi-standard optical filter whose value is
calibrated to a NIST filter.
Pharmacopoeia….Deviation from standard value is within ±0.008 Abs
Test 6:
Instrument specifications
Absorbance near 1Abs : Deviation from average of 3 values is within ±0.001 Abs
Absorbance near 0.5Abs : Deviation from average of 3 values is within ±0.001 Abs
Pharmacopoeia
Absorbance near 1Abs : Deviation from average of 3 values is within ±0.004 Abs
Absorbance near 0.5Abs : Deviation from average of 3 values is within ±0.002 Abs
Test 7: Baseline flatness is within ±0.0010 Abs
* For the measurement methods and pass criteria values in Tests 3 - 11, refer to Additional
Information from page 41.
15) In the Validation Function Settings screen, press the [1] key (Semi-auto Items).
<Additional Information>
· About Displayed Wavelength Value Accuracy, and Wavelength Setting Repeatability
The pass criteria and procedures used for the displayed wavelength value accuracy and
wavelength setting repeatability inspections of Test 3 and Test 4, respectively, are derived from
the Japanese Pharmacopoeia Fifteenth Edition.
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
After pre-adjusting the spectrophotometer, confirm that the wavelength and transmittance comply with
the following tests.
Using optical filters, the wavelength is within ... Testing can be conducted using a low-pressure
mercury lamp ... or deuterium discharge tube 486.00 nm and 656.10 nm emission lines. The
measured wavelength and emission line wavelength shall differ by no more than ±0.3 nm. When
conducting three measurements, all measured values shall be within ± 0.2 nm of the average value.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance tolerance at
440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance range of
0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the rounding
errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted for the
photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion of
(reference value ± 0.008 Abs) is adopted.
The number of measurements and the measurement method for Test 6 conforms with the 15th
Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in order
to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of the
average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
In the UV-1800, from the viewpoint of noise exclusion, the measured baseline is subjected to
smoothing, and the absorbance difference with the 0 Abs line is set as the baseline curvature
(flatness).
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
Ultraviolet-Visible Spectrophotometer UV-1601PC/UV-1650 (main unit).
Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1
(Item) Firmware version check
Test 2
(Item) Initial setup check
(Outline) Checks that the instrument initial setup was completed correctly.
6) Display the unit history in the output window on the UVProbe screen.
* The initial setup items are displayed in this window in a different order from the setting and
diagnosis procedure.
7) The initial setup information appears in the output window. Select all this information and
copy it to the Clipboard.
8) Paste and print the initial setup information using the Report Generator or another
software application.
Test 3
(Item) Wavelength accuracy check
(Outline) Uses the emission-line wavelengths (486.0, 656.1 nm) of the internal D2 lamp to check the
accuracy of the displayed wavelengths.
(Criteria) At D2 lamp 486.0 nm and 656.1 nm emission lines, measurement error ≤ ± 0.3 nm
Test 4
(Item) Wavelength setting repeatability check
(Outline) Measures the wavelength of the D2 lamp emission lines three times each and then compares
the average value against each measured value to check the repeatability of the wavelength
setting.
The pass criteria (Pharmacopoeia) for the wavelength accuracy check and the pass criteria
and measurement methods for the wavelength setting repeatability check (Test 3 and Test
4) are prescribed in the 15th Japanese Pharmacopoeia.
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
After pre-adjusting the spectrophotometer, confirm that the wavelength and transmittance
comply with the following tests.
Using optical filters, the wavelength is within ... Testing can be conducted using a
low-pressure mercury lamp ... or deuterium discharge tube 486.00 nm and 656.10 nm
emission lines. The measured wavelength and emission line wavelength shall differ by no
more than ±0.3 nm. When conducting three measurements, all measured values shall be
within ±0.2 nm of the average value.
Test 5
(Item) Photometric accuracy check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy.
6) For the measurement wavelengths, registered the wavelengths specified in the optical
filter test results document.
7) Set the other parameters, as follows:
[Calibration]
Calibration curve method : Photometric measurement
[Instrument parameters]
Photometry type : Absorbance or transmittance (According to pass
criteria)
Slit width (nm) : (Slit width used in optical filter test results document)
8) Click the [Attachments] tab, set the following parameters, and click the [Initialize] button:
Attachments : CPS
Number of cells : (1 + number of set filters)
* The operations above are necessary for the UVProbe software to recognize that the CPS-240A
accessory is connected. The CPS-240A does not operate when the [Initialize] button is clicked.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as
follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance
tolerance at 440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance
range of 0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the
rounding errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted
for the photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion
of (reference value ± 0.008 Abs) is adopted.
Test 6
(Item) Photometric repeatability check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy over three repeated
measurements.
The number of measurements and the measurement method for Test 6 conforms with the
15th Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of
the average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
Test 7
(Item) Baseline flatness check
(Outline) Checks the baseline flatness in the range from 190 nm to 1100 nm.
11) Print the measurement results. Check the maximum and minimum values, using the
central value of the noise level (excluding shock noise due to light-source or filter
switching) as the baseline absorbance.
For a Shimadzu UV-VIS Spectrophotometer, from the viewpoint of noise elimination, the
central value of the noise level (excluding shock noise due to light-source or filter
switching) is defined as the baseline absorbance and the maximum and minimum values
define the baseline deflection.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
Ultraviolet-Visible Spectrophotometer UV-2401PC/UV-2450 (main unit).
Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1
(Item) Firmware version check
Test 2
(Item) Initial setup check
(Outline) Checks that the instrument initial setup was completed correctly.
6) Display the unit history in the output window on the UVProbe screen.
* The initial setup screen items are displayed in this window in a different order from the setting and
diagnosis procedure.
7) The initial setup screen information appears in the output window. Select all this
information and copy it to the Clipboard.
8) Paste and print the initial setup information using the Report Generator or another
software application.
Test 3
(Item) Wavelength accuracy check
(Outline) Uses the emission-line wavelengths (486.0, 656.1 nm) of the internal D2 lamp to check the
accuracy of the displayed wavelengths.
(Procedure)
Follow the procedure below.
Refer to the UVProbe Instruction Manual (Tutorial) for details about these operations.
(Criteria) At D2 lamp 486.0 nm and 656.1 nm emission lines, measurement error ≤ ± 0.3 nm
Test 4
(Item) Wavelength setting repeatability check
(Outline) Measures the wavelength of the D2 lamp emission lines three times each and then compares
the average value against each measured value to check the repeatability of the wavelength
setting.
Test 5
(Item) Photometric accuracy check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy.
6) For the measurement wavelengths, register the wavelengths specified in the optical filter
test results document.
7) Set the other parameters, as follows:
[Calibration]
Calibration curve method : Photometric measurement
[Instrument parameters]
Photometry type: Absorbance or transmittance (According to pass criteria)
Slit width (nm) : (Slit width used in optical filter test results document)
8) Click the [Attachments] tab, set the following parameters, and click the [Initialize] button:
Attachments: CPS
Number of cells: (1 + number of set filters)
* The operations above are necessary for the UVProbe software to recognize that the CPS-240A
accessory is connected. The CPS-240A does not operate when the [Initialize] button is clicked.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as
follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance
tolerance at 440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance
range of 0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the
rounding errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted
for the photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion
of (reference value ± 0.008 Abs) is adopted.
Test 6
(Item) Photometric repeatability check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy over three repeated
measurements.
The number of measurements and the measurement method for Test 6 conforms with the
15th Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of
the average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
Test 7
(Item) Baseline flatness check
(Outline) Checks the baseline flatness in the range from 190 nm to 900 nm.
11) Print the measurement results. Check the maximum and minimum values, using the
central value of the noise level (excluding shock noise due to light-source or filter
switching) as the baseline absorbance.
For a Shimadzu UV-VIS Spectrophotometer, from the viewpoint of noise elimination, the
central value of the noise level (excluding shock noise due to light-source or filter
switching) is defined as the baseline absorbance and the maximum and minimum values
define the baseline deflection.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
Ultraviolet-Visible Spectrophotometer UV-2501PC/UV-2550 (main unit).
Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1
(Item) Firmware version check
Test 2
(Item) Initial setup check
(Outline) Checks that the instrument initial setup was completed correctly.
6) Display the unit history in the output window on the UVProbe screen.
* The initial setup screen items are displayed in this window in a different order from the setting and
diagnosis procedure.
7) The initial setup screen information appears in the output window. Select all this
information and copy it to the Clipboard.
8) Paste and print the initial setup information using the Report Generator or another
software application.
Test 3
(Item) Wavelength accuracy check
(Outline) Uses the emission-line wavelengths (486.0, 656.1 nm) of the internal D2 lamp to check the
accuracy of the displayed wavelengths.
(Procedure)
Follow the procedure below.
Refer to the UVProbe Instruction Manual (Tutorial) for details about these operations.
(Criteria) At D2 lamp 486.0 nm and 656.1 nm emission lines, measurement error ≤ ± 0.3 nm
Test 4
(Item) Wavelength setting repeatability check
(Outline) Measures the wavelength of the D2 lamp emission lines three times each and then compares
the average value against each measured value to check the repeatability of the wavelength
setting.
Test 5
(Item) Photometric accuracy check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy.
6) For the measurement wavelengths, register the wavelengths specified in the optical filter
test results document.
7) Set the other parameters, as follows:
[Calibration]
Calibration curve method: Photometric measurement
[Instrument parameters]
Photometry type: Absorbance or transmittance (According to pass criteria)
Slit width (nm): (Slit width used in optical filter test results document)
8) Click the [Attachments] tab, set the following parameters, and click the [Initialize] button:
Attachments: CPS
Number of cells: (1 + number of set filters)
* The operations above are necessary for the UVProbe software to recognize that the CPS-240A
accessory is connected. The CPS-240A does not operate when the [Initialize] button is clicked.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as
follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance
tolerance at 440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance
range of 0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the
rounding errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted
for the photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion
of (reference value ± 0.008 Abs) is adopted.
Test 6
(Item) Photometric repeatability check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy over three repeated
measurements.
The number of measurements and the measurement method for Test 6 conforms with the
15th Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of
the average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
Test 7
(Item) Baseline flatness check
(Outline) Checks the baseline flatness in the range from 190 nm to 900 nm.
11) Print the measurement results. Check the maximum and minimum values, using the
central value of the noise level (excluding shock noise due to light-source or filter
switching) as the baseline absorbance.
For a Shimadzu UV-VIS Spectrophotometer, from the viewpoint of noise elimination, the
central value of the noise level (excluding shock noise due to light-source or filter
switching) is defined as the baseline absorbance and the maximum and minimum values
define the baseline deflection.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
Ultraviolet-Visible Spectrophotometer UV-3600.
Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1
(Item) Firmware version check
Test 2
(Item) Initial setup check
(Outline) Checks that the instrument initial setup was completed correctly.
6) Display the unit history in the output window on the UVProbe screen.
* The initial setup screen items are displayed in this window in a different order from the setting and
diagnosis procedure.
7) The initial setup screen information appears in the output window. Select all this
information and copy it to the Clipboard.
8) Paste and print the initial setup information using the Report Generator or another
software application.
Test 3
(Item) Wavelength accuracy check
(Outline) Uses the emission-line wavelengths (486.0, 656.1 nm) of the internal D2 lamp to check the
accuracy of the displayed wavelengths.
9) Print the measurement results (peak detect table) and check the accuracy of the
wavelengths.
Test 4
(Item) Wavelength setting repeatability check
(Outline) Measures the wavelength of the D2 lamp emission lines three times each and then compares
the average value against each measured value to check the repeatability of the wavelength
setting.
The pass criteria (Pharmacopoeia) for the wavelength accuracy check and the pass criteria
and measurement methods for the wavelength setting repeatability check (Test 3 and Test
4) are prescribed in the 15th Japanese Pharmacopoeia.
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
After pre-adjusting the spectrophotometer, confirm that the wavelength and transmittance
comply with the following tests.
Using optical filters, the wavelength is within ... Testing can be conducted using a
low-pressure mercury lamp ... or deuterium discharge tube 486.00 nm and 656.10 nm
emission lines. The measured wavelength and emission line wavelength shall differ by no
more than ±0.3 nm. When conducting three measurements, all measured values shall be
within ±0.2 nm of the average value.
Test 5
(Item) Photometric accuracy check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy.
6) For the measurement wavelengths, register the wavelengths specified in the optical filter
test results document.
7) Set the other parameters, as follows:
[Calibration]
Calibration curve method : Photometric measurement
[Instrument parameters]
Photometry type: Absorbance or transmittance (According to pass criteria)
Slit width (nm) : (Slit width used in optical filter test results document)
Time constant (sec) : 1.0
8) Click the [Attachments] tab, set the following parameters, and click the [Initialize] button:
Attachments : CPS
Number of cells : (1 + number of set filters)
* The operations above are necessary for the UVProbe software to recognize that the CPS-240A
accessory is connected. The CPS-240A does not operate when the [Initialize] button is clicked.
The measurement method for photometric accuracy conforms with the 15th Japanese
Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within
1% of the upper and lower limits of the relative accuracy indicated in the test report… .
(Explanation)
The actual transmittance tolerance at the wavelength of each optical filter for transmittance
calibration can be determined, as described below. The actual transmittance tolerance at the
wavelength of each optical filter for transmittance calibration can be determined, as described
below.
Example: For a 30% ND optical filter with reference value and guaranteed accuracy as
follows:
The Japanese Pharmacopoeia prescribes a transmittance tolerance relative value of ±2% (1%
added to the guaranteed accuracy upper and lower limits), such that the transmittance
tolerance at 440 nm is:
23.92 x (2/100) ≈ 0.48
Thus, the transmittance tolerance is 23.92 ± 0.48% (23.44% to 24.40%). Converting this to
absorbance using absorbance = log10 (100/transmittance), gives an absorbance tolerance
range of 0.613 to 0.630 Abs.
Based on the above information, the Test 5 Pass Criteria/ (Pharmacopoeia ) are defined as
follows:
The tolerance range is 29.40% to 30.60% at 30% transmittance (0.523 Abs absorbance).
Converting this to absorbance gives 0.514 to 0.532 (0.523 ± 0.009). However, due to the
rounding errors during conversion, a pass criterion of (reference value ± 0.008 Abs) is adopted
for the photometric accuracy near 0.5 Abs.
Similarly, as the tolerance range is 0.991 to 1.009 Abs at 10% transmittance, a pass criterion
of (reference value ± 0.008 Abs) is adopted.
Test 6
(Item) Photometric repeatability check
(Outline) Uses filter for transmittance calibration to check the photometric accuracy over three repeated
measurements.
The number of measurements and the measurement method for Test 6 conforms with the
15th Japanese Pharmacopoeia:
Japanese Pharmacopoeia Fifteenth Edition - General Test Methods
2.24. UV-VIS Spectrophotometer Measurement Methods
Wavelength, transmittance, or absorbance calibration
Transmittance and absorbance are measured using optical filters for transmittance calibration.
When testing is conducted under the test conditions in the test report attached to each filter in
order to read the transmittance at the reference wavelength value indicated in the test report, the
displacement between the measured transmittance and reference transmittance shall lie within 1%
of the upper and lower limits of the relative accuracy indicated in the test report. Additionally,
after the measurement is conducted three times, all measured absorbance values (or
measured transmittance values converted to absorbance values) shall lie within ±0.002 of
the average value for absorbance up to 0.500 and ±0.004 for absorbance over 0.500.
Test 7
(Item) Baseline flatness check
(Outline) Checks the baseline flatness in the range from 185 nm to 3300 nm.
11) To eliminate fluctuations such as noise in the above measurement results, perform
smoothing using the parameters displayed below.
12) Print out the measurement results, record the maximum and minimum values of the
processed peaks, and verify that the respective values are within the criteria standard.
Specifications at Inspection
Baseline max. value & min. value ≤ ±0.008 Abs (185 nm to 200 nm)
≤ ±0.002 Abs (200 nm to 3000 nm)
≤ ±0.01 Abs (3000 nm to 3300 nm)
For a Shimadzu UV-VIS Spectrophotometer, from the viewpoint of noise elimination, the
central value of the noise level (excluding shock noise due to light-source or filter
switching) is defined as the baseline absorbance and the maximum and minimum values
define the baseline deflection.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Syringe
Sipper N/CN. Tests 1 to 4 are conducted for an N-type syringe sipper; tests 1 to 5 are conducted for a CN-type
syringe sipper. However, Test 4 (Carry-over check) applies only to units using a flow-through cell recommended
for a Shimadzu syringe sipper. In addition, no pass criterion applies to Test 5; the measured value is noted as a
verification item. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1
(Item) Initial operation check
(Outline) Checks that the syringe pump unit initial operation is correct.
(Procedure) If UVProbe is not used with the UV-1700, perform the test using the following
procedure.
For details on operation, refer to the UV-1700 Spectrophotometer instruction manual
(Operation Guide) and the syringe sipper instruction manual.
1) Check that the spectrophotometer initial setup is completed normally.
2) Select [1. Photometric] from the Mode selection screen.
3) Press the [F2] key (Sample Control).
4) Select the Syringe Sipper as an accessory, and when the [F1] key (Initial) is pressed,
verify that the syringe sipper initialization operation is conducted.
If UVProbe is used for instrument control, perform the test using the following
procedure.
Refer to the Syringe Sipper Instruction Manual and UVProbe Instruction Manual (Tutorial)
for details about these operations.
Test 2 & 3
(Item) Test 2: Intake volume accuracy check
Test 3: Intake volume repeatability check
(Outline) Test 2: Conducts three intake operations on distilled water in a beaker and determines the
actual intake volume by weighing the beaker before and after.
Test 3: Conduct three intake operations on distilled water in a beaker. Determine the intake
repeatability from the discrepancy with the average intake volume.
(Procedure) If UVProbe is not used with the UV-1700, perform the test using the following
procedure.
For details on operation, refer to the UV-1700 Spectrophotometer instruction manual
(Operation Guide) and the syringe sipper instruction manual.
1) Prepare the following devices and sample.
a) Container : beaker
b) Sample : distilled water
c) Balance : Calibrated balance capable of measurement to 0.001g
5) Press the [F1] key (Initial) to conduct the sipper initialization operation.
6) Fill the beaker half-way with distilled water, weigh the beaker and its contents and record
the weight.
7) Take the sipper nozzle that is installed in the spectrophotometer sample compartment, and
insert it into the distilled water in the beaker.
8) Press the [F4] key (Manual sip) to begin sipping the distilled water.
9) When the sipping operation is completed, weigh the beaker on the balance and record the
weight.
10) Repeat steps 7) – 9) three times, and confirm the sip volume accuracy.
11) Take the average value of the 3 sipping operation results, and determine the deviation of
each sipping value from the average value to verify the sip volume repeatability.
(Procedure) If UVProbe is used for instrument control, perform the test using the following
procedure.
Follow the procedure below.
Refer to the Syringe Sipper Instruction Manual and UVProbe Instruction Manual (Tutorial)
for details about these operations.
5) Click the [Reset] button on the Method Setting screen to conduct the syringe sipper initial
operation.
6) Half fill the beaker with distilled water. Weigh the beaker on the electronic balance and
record the weight.
7) Insert the intake nozzle mounted to the spectrophotometer sample compartment into the
distilled water.
8) Click on the [Sip] button to take in distilled water.
9) After taking in the sample, weigh the beaker on the electronic balance again and record
the weight.
10) Repeat steps 7) to 9) three times to determine the intake volume accuracy.
11) Determine the average of the three measured results. Determine the deviation of each
measured value from the average value to confirm the intake volume repeatability.
Test 4
(Item) Carry-over check
(Outline) Checks the carry-over value using an aqueous solution of potassium dichromate.
(Procedure) If UVProbe is not used with the UV-1700, perform the test using the following
procedure.
For details on operation, refer to the UV-1700 Spectrophotometer instruction manual
(Operation Guide) and the syringe sipper instruction manual.
1) Prepare the following devices and sample.
a) Container : 2 beakers
b) Sample : distilled water and potassium dichromate aqueous solution (diluted to
obtain absorbance near 0.5 Abs at 350 nm)
7) Press the [F1] key (Initial) to conduct the sipper initialization operation.
8) Press the [F4] key (Manual sip), sip distilled water from the sample compartment nozzle,
and press the [RETURN] key to return to the Measurement screen.
9) Press the [AUTO ZERO] key to perform the auto zero operation.
10) After measuring the distilled water 3 times, measure the potassium dichromate aqueous
solution 3 times.
11) Repeat step 10) two more times.
12) Print the measurement results, and check the carryover.
(Procedure) If UVProbe is not used with the UV-1700, perform the test using the following
procedure.
Follow the procedure below.
5) After the initialization operation is complete, click the [Sip] button to take distilled water
into the sample compartment intake nozzle. Click the [Close] button to close the
measurement method.
6) Click the [Go TO WL] button and set the wavelength to 350.0 nm.
7) Click the [Auto Zero] button to autozero.
8) Enter the numbers 1 to 18 in the Sample ID column to create an 18-line sample table.
9) Measure distilled water three times and then measure the potassium dichromate aqueous
solution three times.
10) Repeat step 9) two more times.
11) Print the measurement results (sample table) and check the carry-over value.
<Additional Preparation procedure for weakly acidic aqueous solution of potassium dichromate
Information> (1) Dilute 167 mg HClO4 (60%) in 1L distilled water to create a 0.001 mol/L (0.001N)
perchloric acid solution.
(2) Dissolve 47 mg potassium dichromate (K2Cr2O7) to the 0.001N perchloric acid solution
created at step (1) above to create 1L solution.
Test 5
(Item) Temperature control accuracy check
(Outline) Checks the difference between the circulating water and cell holder temperatures.
(Procedure) Follow the procedure below.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Automatic
Sample Changer ASC-5. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1 & 2
(Item) Test 1: Firmware version check
Test 2: Initial operation check
(Outline) Test 1: Checks the program version.
Test 2: Checks that the initial operation is correct.
(Procedure) Follow the procedure below.
Refer to the ASC-5 Instruction Manual for details about these operations.
1) Turn on the ASC-5.
2) The LEDs display the program checksum value (flashing). Record the checksum value
and use it to confirm the version number from the table in Additional Information below.
3) Confirm that the intake nozzle automatically returns to the home position and the ASC-5
reverts to its initial status (TEACHING and READY LEDs not lit).
(Criteria) Test 1: The program checksum value is displayed and matches the version number on
record.
Test 2: Initial movements complete normally with no error displayed.
<Additional · Relationship between the firmware version and checksum value
Information>
The relationship between the ASC-5 firmware checksum values and version numbers is
listed below.
Checksum value Version
02A ; Ver 1.00
Test 3
(Item) Manual operation check
(Outline) Uses ASC-5 manual key operation to check that the intake nozzle moves smoothly in the X
and Y directions one pitch at a time.
(Procedure)
Follow the procedure below.
Refer to the ASC-5 Instruction Manual for details about these operations.
1) Turn on the ASC-5.
2) After the initial movements are complete, press the HOME key to set the ASC-5 into
ready status (READY LED lit).
3) Mount the standard test-tube rack (ASC-5 accessory) to the ASC-5 sample stage.
4) Insert standard test-tubes (ASC-5 accessory) into the four corners of the standard
test-tube rack.
5) Press the down-arrow key while holding down the [FILE] key and set “F-0” (File 0).
6) Press the [START/STOP] key while holding down the [FILE] key and load “F-0”.
7) Confirm that the intake nozzle is at the #1 test-tube position (X:1, Y:1).
8) Press the arrow keys repeatedly to move the intake nozzle one pitch at a time and check
that the nozzle moves to each test-tube position.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Elution Test
6-Cell Holder DIS-1200. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
Test 1 & 2
(Item) Test 1: Initial operation check
Test 2: Cell position check
(Outline) Test 1: Checks that the cell holder home position is correctly detected when the Initialize
command is used.
Test 2: Conduct three transmittance measurements at each cell position to verify that the
optical beam is not obstructed.
(Procedure) Perform the operations described in the procedure below.
For details on operation, refer to the DIS-1200 instruction manual and UV-1700 Instruction
Manual (Operation Guide).
9) Press the [RETURN] key to return to the measurement conditions setting screen.
10) Press the [AUTO ZERO] key to perform auto zero correction (100% correction).
11) Press the [START/STOP] key to start transmittance measurement of cell holders 1 – 6.
12) Repeat step 11) two more times, for a total of 3 measurements.
13) Print the measurement results, and verify that the values are within the standard value.
(Criteria) Test 1: Initialization causes cell holder 1 to move to the optical beam position.
Test 2: Transmittance at each cell position is 100.0 ±1.0%.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Elution Test
8-Cell Holder DIS-8. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Initial operation check Checks that the cell holder home position is correctly detected when
the Initialize command is used.
2 Cell position check Conduct three transmittance measurements at each cell position to
verify that the optical beam is not obstructed.
3. Inspection Procedures
Test 1 & 2
(Item) Test 1: Initial operation check
Test 2: Cell position check
(Outline) Test 1: Checks that the cell holder home position is correctly detected when the Initialize
command is used.
Test 2: Conduct three transmittance measurements at each cell position to verify that the
optical beam is not obstructed.
(Procedure) Perform the operations described in the procedure below.
For details on operation, refer to the DIS-8 instruction manual and UV-1700 Instruction
Manual (Operation Guide).
9) Press the [RETURN] key to return to the measurement conditions setting screen.
10) Press the [AUTO ZERO] key to perform auto zero correction (100% correction).
11) Press the [START/STOP] key to start transmittance measurement of cell holders 1 – 8.
12) Repeat step 11) two more times, for a total of 3 measurements.
13) Print the measurement results, and verify that the values are within the standard value.
(Criteria) Test 1: Initialization causes cell holder 1 to move to the optical beam position.
Test 2: Transmittance at each cell position is 100.0 ±1.0%.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Cell
positioner CPS-240A/B. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Initial operation check Checks that the cell holder home position is correctly detected when
the power is turned on.
2 Cell position check Conducts three transmittance measurements at each cell position to
confirm that the optical beam is not interrupted by thermal insulation,
etc.
3 Temperature accuracy check Checks the cell holder base temperature accuracy with respect to the
set temperature.
3. Inspection Procedures
Test 1 & 2
(Item) Test 1: Initial operation check
Test 2: Cell position check
(Outline) Test 1: Confirms that the cell holder home position is correctly detected when the power is
turned on.
Test 2: Conducts three transmittance measurements at each cell position to confirm that the
optical beam is not interrupted by thermal insulation, etc.
(Procedure) If UVProbe is not used with the UV-1700, perform the test using the following
procedure.
For details on operation, etc., refer to the UV-1700 instruction manual (Operation Guide).
* The operations above are required in order for the CPS-240A/B to be recognized as an accessory. Even
if the [Initial] button is clicked, the CPS-240A/B will not operate.
11) Press the [RETURN] key to return to the measurement conditions setting screen.
12) Press the [AUTO ZERO] key to perform auto zero correction (100% correction).
13) Press the [START/STOP] key to start transmittance measurement of cell holders 1 – 6.
14) Repeat step 11) two more times, for a total of 3 measurements.
Print the measurement results, and verify that the values are within the standard value.
(Procedure) If UVProbe is used for instrument control, perform the test using the following
procedure.
Follow the procedure below.
Refer to the CPS-240 Instruction Manual and UVProbe Instruction Manual (Tutorial) for
details about these operations.
(Criteria) Test 1: Cell 1 automatically moves to the optical beam position when CPS-240A/B is
turned on.
±1.0%.
Test 2: Transmittance at each cell position is within 100.0±
Test 3
(Item) Temperature accuracy check
(Outline) Checks the cell holder base temperature accuracy with respect to the set temperature.
1) Place the temperature sensor against the base of the CPS-240A (or CPS-240B) Cell 6
holder.
2) Set the cell holder to 16°C at the CPS temperature controller.
3) While watching the thermometer connected to the sensor, wait until the sensor temperature
stabilizes.
4) Once the sensor’s temperature stabilizes, record the temperature read from the
thermometer.
Also record the actual measurement temperature obtained by subtracting the temperature
deviation (correction value), which is determined by the measurement value (Ω) specified
in the temperature sensor inspection report and the JIS table, from the thermometer
reading.
5) Determine the deviation of the actual measurement temperature from the set temperature
(16°C), and verify that it falls within the criteria.
6) Repeat steps 2) to 5) to determine the deviation at set temperatures of 37°C and 60°C.
(Criteria) The deviation from the set temperatures (16°°C, 37°°C, 60°°C) is within ±(0.5+A+B) °C
A: Thermocouple test accuracy
B: Thermometer accuracy rating*
Example) If the temperature rating is ± (0.1% of rdg + 0.3°C), the accuracy rating for the set
temperatures 16°C and 37°C is ±0.3°C, and ±0.4°C for 60°C.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
Thermoelectrically Temperature Controlled Cell Holder TCC-240A. Refer to the Instruction Manual for
details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Temperature accuracy check Checks the cell holder base temperature accuracy with respect to the
set temperature.
3. Inspection Procedures
Test 1
(Item) Temperature accuracy check
(Outline) Checks the cell holder base temperature accuracy with respect to the set temperature.
Example) If the temperature rating is ± (0.1% of rdg + 0.3°C), the accuracy rating for the set
temperatures 10°C and 37°C is ±0.3°C, and ±0.4°C for 60°C.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Screen Copy
Printer HCP-1C. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Printing test Performs test printout, and checks control circuit operation, print
head/printer mechanism functions and print quality.
3. Inspection Procedures
Test 1
(Item) Printing test
(Outline) Performs test printout, and checks control circuit operation, print head/printer mechanism
functions and print quality.
(Procedure) Perform the operations described in the procedure below.
For details on operation, refer to the HCP-1C instruction manual.
1) Set recording paper in the HCP-1C, and position the recording paper set lever at [Print].
2) Turn on the HCP-1C and UV instrument, and wait for the initialization process to be
completed.
3) At the UV instrument, select the [9. Utilities] menu.
4) Select [5. Printer], and select HCP-IC.
5) Following the on-screen messages, set the conditions to the desired values.
6) When the printer settings are completed, press the [RETURN] key to return to the mode
selection screen.
7) Set the HCP-1C to the ON LINE state.
8) Press the [PRINT] key to print out a hard copy of the mode selection screen.
(Criteria) No problems with printing condition.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Screen Copy
Printer DPU-414. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Printing test Performs test printout, and checks control circuit operation, print
head/printer mechanism functions and print quality.
3. Inspection Procedures
Test 1
(Item) Printing test
(Outline) Performs test printout, and checks control circuit operation, print head/printer mechanism
functions and print quality.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Screen Copy
Printer MPU. Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain
points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Printing test Performs test printout, and checks control circuit operation, print
head/printer mechanism functions and print quality.
3. Inspection Procedures
Test 1
(Item) Printing test
(Outline) Performs test printout, and checks control circuit operation, print head/printer mechanism
functions and print quality.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the ESC/P
Compatible Screen Copy Printer. Refer to the Instruction Manual for details about the operation procedure
and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Printing test Performs test printout, and checks control circuit operation, print
head/printer mechanism functions and print quality.
3. Inspection Procedures
Test 1
(Item) Printing test
(Outline) Performs test printout, and checks control circuit operation, print head/printer mechanism
functions and print quality.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the PCL
Compatible Screen Copy Printer. Refer to the Instruction Manual for details about the operation procedure
and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Printing test Performs test printout, and checks control circuit operation, print
head/printer mechanism functions and print quality.
3. Inspection Procedures
Test 1
(Item) Printing test
(Outline) Performs test printout, and checks control circuit operation, print head/printer mechanism
functions and print quality.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Time Course
(Multi λ) Program Pack. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Program version check Checks the program version.
2 Checksum value check Checks the checksum value, and checks that there is no problem with
the program.
3. Inspection Procedures
Test 1&2
(Item) Test 1: Program version check
Test 2: Checksum value check
(Outline) Test 1: Checks the program version using the spectrophotometer IC card check function.
Test 2: Verifies that the program’s check sum value matches the original value using the
spectrophotometer IC card check function.
(Procedure) Perform the operations described in the procedure below.
For details on operation, refer to the UV-1700 instruction manual (Installation/Maintenance
and Operation Guides).
1) Check that the spectrophotometer is turned on.
2) Insert the program pack in the slot on the right side of the spectrophotometer.
3) In the mode selection screen, press the [F2] key (IC card).
4) In the data/program pack management screen, select [3. Program pack check].
5) Check that the check sum value matches, and check the version number.
6) Press the [PRINT] key to print the results screen (hard copy).
(Criteria) Test 1 : Program version number matches the controlled version number.
Test 2 : Checksum value matches.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the
DNA/Protein Quantitation Program Pack. Refer to the Instruction Manual for details about the operation
procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Program version check Checks the program version.
2 Checksum value check Checks the checksum value, and checks that there is no problem with
the program.
3. Inspection Procedures
Test 1 & 2
(Item) Test 1: Program version check
Test 2: Checksum value check
(Outline) Test 1: Checks the program version using the spectrophotometer IC card check function.
Test 2: Verifies that the program’s check sum value matches the original value using the
spectrophotometer IC card check function.
(Procedure) Perform the operations described in the procedure below.
For details on operation, refer to the UV-1700 instruction manual (Installation/Maintenance
and Operation Guides).
1) Check that the spectrophotometer is turned on.
2) Insert the program pack in the slot on the right side of the spectrophotometer.
3) In the mode selection screen, press the [F2] key (IC card).
4) In the data/program pack management screen, select [3. Program pack check].
5) Check that the check sum value matches, and check the version number.
6) Press the [PRINT] key to print the results screen (hard copy).
(Criteria) Test 1 : Program version number matches the controlled version number.
Test 2 : Checksum value matches.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Protein
Quantitation Program Pack. Refer to the Instruction Manual for details about the operation procedure and to
clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
# (Item) (Outline)
1 Program version check Checks the program version.
2 Checksum value check Checks the checksum value, and checks that there is no problem with
the program.
3. Inspection Procedures
Test 1 & 2
(Item) Test 1: Program version check
Test 2: Checksum value check
(Outline) Test 1: Checks the program version using the spectrophotometer IC card check function.
Test 2: Verifies that the program’s check sum value matches the original value using the
spectrophotometer IC card check function.
(Criteria) Test 1 : Program version number matches the controlled version number.
Test 2 : Checksum value matches.
4-17 CLASS-Agent
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Class-Agent.
Refer to the Instruction Manual for details about the operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Procedures
User Registration of Performer
The system administrator must complete user registration for the Performer who conducts the work (OQ User)
before this Operational Qualification is conducted. Simultaneously, register the Windows user ID to allow the
OQ User to log in to Windows as a normal operator.
Disable the Performer's user registration when all operations are complete.
Login ID UVProbe
User name Password CLASS-Agent Authority Group
(User ID) Authority Group
Operator,
OQ_USER_1 OQ_Operator ******* Operator
but with additional “Approval”
authority for Agent Manager
OQ_USER_2 OQ_Approver ******* Administrator Administrator
The administrator should issue the initial user password and inform the user.
Test 1-1
(Item) Software version check
(Outline) Checks that the software is not changed to an inappropriate version.
(Procedure) 1) Check the version number printed in the test results from the alteration test in Test 1-2.
2) Check that the version number matches the version number on record.
(Criteria) Version number matches the version number on record.
Test 1-2
(Item) Program alteration check
(Outline) Conducts the program alteration check.
(Procedure) 1) Click on the Windows [Start] button and select [Programs] -> [Class-Agent Software] ->
[Program Files Check] to run the alteration check program.
2) A login window is displayed. Login using the test user ID “OQ_USER_2”.
3) The Program Alteration Check screen is displayed. After selecting a program to be
checked (it becomes highlighted), click the [Check] button.
4) Click the [Run] button. File comparison is conducted and the results indicate whether
program alteration has occurred.
5) Click the [Notepad] button.
6) The results are displayed as a text file. Ensure that all items are OK and print the results.
7) Similarly, check other installed programs (not Shimadzu User Authentication Tool).
Test 1-3
(Item) Shimadzu User Authentication Tool connection check
(Outline) Checks the Shimadzu User Authentication Tool connection using database access and log
registration.
(Procedure) 1) Open the Windows Control Panel and double-click on the Shimadzu User Authentication
Tool icon to open the main menu.
2) Record the server name displayed at the bottom of the main menu and confirm that the
connected database is correct.
* Conduct steps 3) and 4) below after conducting Test 3-4-4.
3) Log in to [Log Display] as “OQ_USER_2” from the main menu. Set the test date and test
computer name in [Search Conditions]. Search the log and print out the results.
4) Confirm that the user lockout generated in Test 3-4-4 is registered and that the software
name is AgentMan2.exe.
Test 1-4
(Item) 1-4-1 Password log management function check
1-4-2 Number of password characters check
1-4-3 Failed login detection
1-4-4 User lockout function check
1-4-5 Failed login record
(Outline) Checks that the security functions operate, including restrictions on the number of password
characters.
(Preparations) Before conducting this test, verify that the configuration settings in the Shimadzu User
Authentication Tool are set as shown below. Be sure to change items (3) and (4) to the settings
shown below.
Item Setting
(1) Minimum password length (Verify only; set to non-zero value)
(2) Previous password is permitted No check mark
(3) User lockout duration 1 minute
(4) User lockout after # bad logon attempts Check only; set to non-zero value
(5) PC lockout after # bad logon attempts The same number as the user logon
attempts, or higher
(Procedure) 1) Run CLASS-Agent Manager, and log on as test user “OQ_USER_1”.
2) Select [Change Password] in the [Tools] menu.
3) Enter the current password and enter the same password as the new password.
4) Confirm that the new password is not accepted.
5) Enter the current password and enter a new password that is shorter than the minimum number
of characters set in the Shimadzu User Authentication Tool.
6) Confirm that the change is not accepted, as the password is too short.
7) Enter a password equal in length to the minimum number of characters and confirm that the
change is accepted.
8) Log out.
9) Click [Login] in the [File] menu, and log in using the test user ID “OQ_USER_1”.
10) Enter a password that is not set and click OK.
11) Check that login is not permitted and a message is displayed warning that the password is
incorrect.
12) Repeat steps 9) to 11) for the number of times designated for “User logon attempts”.
13) After the designated numbers of login failures, confirm that a message indicates that the current
user is locked out and confirm that the user cannot log in for the designated lockout time.
14) After the designated lockout time elapses, select [File] -> [Login]. Enter “OQ_USER_1” as the
login ID and correctly enter the set password.
15) Confirm that login is possible.
16) Click [SYSTEM] -> [SYSTEM LOG] -> [LOG2.MDB] in the CLASS-Agent Manager
database tree.
17) Click [OK] on the “Search Log Information” screen.
18) Confirm that the login failures in the procedure above and the corresponding times appear in
the displayed log.
19) If any settings were changed in the Configuration settings of the User Authentication Tool as
indicated in (Preparations), return the settings to their original values.
(Criteria)
1-4-1 Duplicate password is not accepted.
1-4-2 Change to a password shorter than the minimum number of characters is not
accepted.
Change to a password with the minimum number of characters is accepted.
1-4-3 Login prohibited with an incorrect password but possible with the correct password.
1-4-4 User locked out after the designated number of login failures.
Login prohibited for the designated lockout time.
1-4-5 Login failures are recorded in the log.
Test 1-5
(Item) User access rights check
(Outline) Checks that software use is restricted according to differences in registered user access rights.
(Procedure) 1) Run CLASS-Agent Manager.
2) Enter “OQ_USER_2” as the login ID, and correctly enter the set password.
3) Select [Tools] and check that the “User Administration” item is enabled (displayed as
solid black characters).
4) Log out of CLASS-Agent Manager.
5) Select [File] -> [Login] once more. Enter “OQ_USER_1” as the login ID. Correctly enter
the set password.
6) Select [Tools] and check that the “User Administration” item is disabled (displayed as
gray characters).
(Criteria) “User Administration” item is enabled for OQ_USER_2 but disabled for OQ_USER_1.
Test 1-6
(Item) Test of function to prevent duplicate user registration
(Outline) Checks that a duplicate user name cannot be registered.
(Procedure) 1) Run CLASS-Agent Manager.
2) Log in by entering the user ID “OQ_USER_2” and the corresponding password.
3) Select [Tools] -> [User Administration].
4) Select “OQ_USER_1” and click [Active/Invalid].
5) A message indicates that Login ID: OQ_USER_1 is invalidated. Click [OK].
6) Enter “OQ Test” in the Reasons column and click [OK].
7) Click the [New] button.
8) Enter the following information and click [OK].
Login ID OQ_USER_1
User name OQ_Operator
9) Confirm that a message indicates that the user cannot be registered because the Login ID
or User Name already exists.
10) After this test, select “OQ_USER_1” and click [Active /Invalid] to validate this user.
(Criteria) Duplicate user cannot be registered and a message indicates that the setting is not possible
because the Login ID or User Name already exists.
Test 1-7
(Item) Data display, search, and sort check
(Outline) Checks that the sample search and sort work correctly and that the data is displayed normally.
(Procedure) 1) Run CLASS-Agent Manager. Login by entering the user ID “OQ_USER_1” and the
corresponding password.
2) Display the contents of the database registered in UVProbe Test 2-7-1.
3) Click in the sample display area, select [Edit] -> [Filter Database] and perform a search
using the following conditions:
(1) Reg. date (from): Date that UVProbe Test 2-7-1 was implemented
(2) Reg. date (to): Date that UVProbe Test 2-7-1 was implemented
(3) PC name: Computer name used for Test 1-5.
Name of the inspected computer if UVProbe Test 2-7-1 was implemented with the
inspected computer. Otherwise, designate the name of the computer used to register
data to the displayed database for the UVProbe OQ.
(4) File information 1: Anthra
4) Check that the samples registered during UVProbe Test 2-7-1 are found and displayed.
5) Click in the sample information display area. Select [Edit] -> [Sort]. Enter “File
information 1” for [Sort Item 1] and select “Ascending.”
6) Confirm that the samples are displayed in File Information 1 in the sample information
display area in the sequence: “Anthra.scp”, “Anthra1.scp”, “Anthra2.scp”.
7) Double-click on “Link file 1” for the data “Anthra2.scp” registered in the UVProbe Test
2-7-1, display the peak detection and point pick results, and print them out. Confirm that
the printed output content is identical to that in the report printed in UVProbe Test 2-7-1.
8) Duble-click on “Link file 3” for the same data, display the data set log and print it out.
Confirm that the print contnet is identical to that in the report printed in UVProbe Test
2-7-1.
Test 1-8
(Item) Data recovery
(Outline) Checks that the data files registered in a database can be correctly recovered.
(Procedure) 1) Conduct this test while logged in for Test 1-7.
2) Select the data (Anthra2.spc) registered during UVProbe Test 2-7-1 and select
[Operations] -> [Browse Compressed File].
3) Click [Select All] and then click [Extract]. When a message prompts whether to
overwrite files, click [Overwrite All].
4) Using Explorer, check that the following file is saved in the folder in
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Test:
C:¥ProgramFiles¥Shimadzu¥UVProbe¥Test¥Anthra2.spc
Test 2
This test uses data registered in the database during UVProbe Test 2-7-1 and data recovered in CLASS-Agent
Manager Test 1-8.
Peak detection/point pick results and the contents of the data set log are identical to
(Criteria)
those of UVProbe Test 2-7-1.
Re-registration is accomplished without overwriting
Data set log displayed content is identical to that of the UVProbe report.
Test 3
This test uses data recovered in CLASS-Agent Manager Test 1-8.
Test 4
(Item) Electronic signature check
(Outline) Checks the CLASS-Agent Manager electronic signature function.
(Procedure) 1) Run CLASS-Agent Manager.
2) Login as the “OQ_USER_2” test user.
3) Select [Tools] -> [Configuration] and open the [Approval Settings] tab. If the approvals
level is not set to 1, set it to 1, click [Save Settings], and click [Close] to exit
CLASS-Agent Manager. Proceed to step 6).
4) If the approval level is already set to 1, set it to 3 and click [Save]. Next, revert the setting
to 1, click [Save] again, and then click [Close].
Note: This step is required for Test 7 System Log Recording Check.
5) Exit CLASS-Agent Manager.
6) Run CLASS-Agent Manager and log in as “OQ_USER_1”.
7) Click in the sample display area, Select [Edit] -> [Filter Database] and perform a search
using the following conditions:
(1) Reg. date (from): Date that UVProbe Test 2-7-1 was implemented
(2) Reg. date (to): Date that Test 3 was implemented
(3) PC name: Computer name used for UVProbe Test 2-7-1.
Name of the inspected computer if UVProbe Test 2-7-1 was implemented with
the inspected computer. Otherwise, designate the name of the computer used to
register data to the displayed database for the UVProbe OQ.
(4) File information 1: Anthra2
8) Click one of the displayed data files and click [Approve] to approve it.
9) Click the sample information display area. Select [File] -> [Print]. Confirm that a print
preview window showing the printed image opens and that the printed image shows the
details of the sample information table. Adjust the column widths, if necessary, and print.
10) Confirm that “OQ_USER 1” is printed in the Approver cell and that the correct date and
time are printed in the Approval Date cell.
11) Log off and then log in as “OQ_USER_2”.
12) Select the data file approved at step 8) and click [Reject]. Enter the comment “Data was
invalidated due to the OQ Test data.” in the Comment field and click [OK].
13) If the Approval level was changed at step 3), return it to the original level.
Test 5
(Item) System log recording check
(Outline) Checks that the changes to the CLASS-Agent Manager system setting conditions are recorded to
the system log.
(Procedure)
1) Run CLASS-Agent Manager.
2) Login as the “OQ_USER_2” test user.
3) Select the “SYSTEM LOG” database.
4) Select [Filter of Log Info.], enter the date that Test 4 was implemented for [DATE (from)]
and [DATE (to)], enter the name of the inspected computer as the [PC name], and enter
“Config” for [Operation]. Click [OK].
5) Select [File] -> [Print]. Place a check at all options and then click [Printing]. If necessary,
adjust the column widths.
6) Check that the change to the approval level in Test 4 is correctly recorded.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Validation
Software Package. Refer to the Instruction Manual for details about the operation procedure and to clarify
uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2 Inspection Procedures
Test 1
(Item) Software version check
(Outline)
Checks the version number of the validation software package.
(Procedure)
1) Check the version number printed in the program alteration check results from Test 2.
2) Check that the version number matches the controlled version number.
(Criteria) Version number matches the version number on record.
Test 2
(Item) Program alteration check
(Outline) Checks that no modifications have occurred to the contents, date, or checksum of the files
(*.dll, *.exe) in the validation software folder.
(Procedure) 1) Click on the Windows [Start] button and select [Programs] -> [Shimadzu] -> [UVPC
Perfval - Alteration Check Software] to run the alteration check program.
2) When the Login screen opens, log in using the login ID and password of Admin.
3) The Program Alteration Check screen is displayed. Select (highlight) “UVPC Validation
Software” and click the [Check] button.
4) Click the [Execute] button to run the program alteration check.
5) When the check is complete, click the [NotePad] button.
6) The results are displayed as a text file. Ensure that all items are OK and print the results.
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Color
Measurement Software COL-UVPC. Refer to the Instruction Manual for details about the operation procedure
and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2 Inspection Procedures
Test 1
(Item) Software version check
(Outline) Checks the version number of the color measurement software package.
(Procedure)
1) Check the version number printed in the program alteration check results from Test 2.
2) Check that the version number matches the controlled version number.
(Criteria) Version number matches the version number on record.
Test 2
(Item) Program alteration check
(Outline) Checks that no modifications have occurred to the contents, date, or checksum of the files
(*.dll, *.exe) in the color measurement software folder.
(Procedure)
1) Click on the Windows [Start] button and select [Programs] -> [Shimadzu] -> [UVPC
Perfval - Alteration Check Software] to run the alteration check program.
2) When the Login screen opens, log in using the login ID and password shown below, as
appropriate.
3) When the [Program Alteration Check] screen is displayed, select (highlight) “UVColor”,
and click the [Check] button.
4) Click the [Run] button to execute the Program Alteration Check program.
5) When the check is complete, click the [NotePad] button.
6) The results are displayed as a text file. Ensure that all items are OK and print the results.
Test 3
(Item) Function check
(Outline) Checks the results of running the color measurement software data calculation functions to
ensure that the software functions operate normally.
(Procedure)
Follow the procedure below.
1) Run COL-UVPC.
2) Select [Illuminant/Obs Parameters] from the [Configure] menu and set the following
parameters:
Illuminant :C
Std Observer : 2°
Sample ID repeat :1
Hitch Std : NO
3) Select [Color Scales] from the [Configure] menu and select the [CIE XYZ…] X, Y and Z
calculation items.
4) Read the following installed file as sample data:
Data name Date and time
Silcon.spc 93/03/04 11:30
5) Automatic calculation is conducted and the results are displayed in a table on the screen.
Confirm that the values match the values below and print out the table.
X Y Z
24.52 35.56 65.34 (when the version is over 3.00)
24.52 35.56 65.33 (when the version is under 2.70)
(Criteria) Steps 1) to 4) run correctly using the file in Step 4) and the results match the values
shown in Step 5).
1. Introduction
This section presents the standard operating procedure (SOP) for Operational Qualification of the Film
Thickness Measurement Software FLM-UVPC. Refer to the Instruction Manual for details about the
operation procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2 Inspection Procedures
Test 1
(Item) Software version check
(Outline) Checks the version number of the film thickness measurement software package.
(Procedure) 1) Check the version number printed in the program alteration check results from Test 2.
2) Check that the version number matches the controlled version number.
(Criteria) Version number matches the version number on record.
Test 2
(Item) Program alteration check
(Outline) Uses the alteration check software to check that no modifications have occurred to the
contents, date, or checksum of the files (*.dll, *.exe) in the film thickness measurement
software folder.
(Procedure) 1) From the Windows Start menu, select [Programs] -> [Film Thickness Measurement] ->
[Program Alteration Check] to run the Program Alteration Check program.
2) When the Login screen opens, log in using the login ID and password shown below, as
appropriate
3) When the [Program Alteration Check] screen is displayed, select (highlight) “UVFilm”,
and click the [Check] button.
4) Click the [Run] button to execute the Program Alteration Check program.
5) When the check is complete, click the [NotePad] button.
6) The results are displayed as a text file. Ensure that all items are OK and print the results.
Test 3
(Item) Function check
(Outline) Checks the results of running the film thickness measurement software data calculation
functions to ensure that the software functions operate normally.
(Procedure)
Follow the procedure below.
1) Run FLM-UVPC.
2) Select [Calculating Parameters] from the [Configure] menu and set the following
parameters:
Wavelength (nm) : 400.0 to 800.0
Refractive index : 1.500
Incident Angle : 5.0
Smoothing dWL : Enable
Delta lambda (interval x) : 10
Threshold of Peak : 0.1000
Repetition :1
Auto Plot Out : Disable
4) Automatic calculation is conducted and the results are displayed in a table on the screen.
Confirm that the values match the values below and print out the table.
Film thickness Peak SD
2.397 15.87
(Criteria) Steps 1) to 3) run correctly using the file in Step 3) and the results match the values
shown in Step 4).
1. Introduction
This section presents the standard operating procedure (SOP) to invalidate the inspection data stored in the
database using the CLASS Agent Manager. Refer to the Instruction Manual for details about the operation
procedure and to clarify uncertain points.
Follow section 1-3 Entries and Revisions to change or delete the contents of the procedures.
2. Inspection Outline
Uses the CLASS-Agent Manager to invalidate inspection data stored in the CLASS-Agent database.
# (Item) (Outline)
1 Invalidating Inspection Data Uses the CLASS-Agent Manager to invalidate inspection data stored
in the CLASS-Agent database.
3. Inspection Procedures
Test 1
(Item) Invalidating Inspection Data
(Outline) Uses the CLASS-Agent Manager to invalidate inspection data stored in the CLASS-Agent
database.
(Procedure) 1) Run CLASS-Agent Manager and log in as “OQ_USER_2”.
2) Display the database where the data is registered and click the sample information display
area. Select [Edit] -> [Search] and search for the following data conditions.
(1) Reg. date (from): Date that the test was implemented.
(2) Reg. date (to): Date that the test was implemented.
(3) PC name: Computer name used for the test.
The inspected computer is the same as the computer used to conduct
this procedure.
(4) Instrument name : Name of analytical instrument inspected here.
3) Select all displayed data.
4) Click [Reject] in Agent Manager and enter the comment “Data was invalidated after the
OQ Test.” in the Comment row to invalidate the data.
5. Change List
When changing or deleting the contents of this document, or adding new information to this document, record
change list number, page number, reason, date, and name of the Performer in the Change List below, and obtain
the approval of the Reviewer and the Manager.
Change No: Enter the change number assigned in sequence from 1 onwards in accordance with
“1.3 Entries and Revisions”.
Page: Enter the page number in which the change, deletion or addition was made.
Reason: Enter the reason for the change, deletion or addition.
Date: Enter the date on which the change, deletion or addition was made.
Change
Page Reason Date Performer Reviewer Manager
No.
Change
Page Reason Date Performer Reviewer Manager
No.
Change
Page Reason Date Performer Reviewer Manager
No.