Introduction Before calibration, to do inspection of PCB : Please adjust mini jumper SWA1to LOCK, press ON/ZERO key to test LCD. Are there with shadow or word break? (After calibration the problems will be solved.)
Press OFF key to power off then press NET/GROSS and hold and press ON/ZERO key at same time. When screen displays into self testing module
Press ON/ZERO key Press UNITS key Press ON/ZERO key Press ON/ZERO key Press UNITS key Press UNITS key Press ON/ZERO key - 2 - 02005012
Internal value modal (must connect to Load cell with full bridge rectifier for testing) Press TARE , screen displays internal value. Please check internal value whether it during normal range (10000 ~ 150000 for empty) Please check backlight. Four corner modification (The internal value of four corner error should be under 10d) After four corner checking, please adjust mini jumper SWA1 to ADJ, then into calibration or press ON/ZERO key return to previous setting. The screen displays Keypad testing Press TARE key, the screen displays internal value of keypad: TARE key= 07 ! UNITS key= 06 ! NET/GROSS key =05 Press ON/ZERO key, return to previous setting, the screen displays
Program version display Press TARE key, the screen displays program version , Press TARE key again, the screen displays maintenance program version for 2 seconds. Press ON/ZERO key, return to previous setting, the screen displays
Return to previous setting Press TARE key, out of self testing and the scale power on automatically. - 3 - 02005012 ERROR MESSAGE EEPROM problem ( Cannot read data) Zero rang when switching on scale is too high. Zero range when switching on scale is too low. Internal value is unstable. Internal value is under Zero. AD problem (cannot read AD value) Weight over full capacity + 9 division AD is over the maximum resolution range.
WEIGHT UNIT (kg) 1 g = 0.001 kg (g) 1 g = 1 g (lb) 1 g = 0.002204623 lb (lb,oz) 1 g = 0.03527396 oz (oz) 1 g = 0.03527396 oz (GN) 1 g = 15.432358 GN (dwt) 1 g = 0.6430149 dwt (ct) 1 g = 5 ct
- 4 - 02005012 Press UNITS key INTERNAL FUNCTION SETTING 4 Please adjust mini jumper SWA1to ADJ (EEPROM UNLOCK), Switch on, into calibration function setting. The screen displays . After finishing calibration function setting, should adjust MINI JUMPER to LOCK If adjust MINI JUMPER to LOCK, will exit from setting model when doing calibration function setting.
Specification setting Function setting
Weighing calibration Acceleration of gravity
Linearity Calibration setting Press ON/ZERO key Press UNITS key Press ON/ZERO key Press ON/ZERO key Press ON/ZERO key Press ON/ZERO key Press UNITS key Press UNITS key Press UNITS key - 5 - 02005012 After key in 00!, press TARE key, return to previous setting. (Exit from specification setting) SPECIFICATION CALIBRATION
ON/ZERO key Upper key (Selects number 0,1,2,3,4,5,6,7,8,9) UNITS key Down key (Selects number 0,1,2,3,4,5,6,7,8,9) TARE key Right key, blinking word move to right NET/GROSS key Left key, blinking word move to left
Return to previous setting Specification setting Press TARE key Key in 00! or 01! - 6 - 02005012 kg Press TARE key See note 3!and key in number
Press TARE key Press TARE key See note 1!and key in number
Press TARE key, finish specification setting. Key in number 00! kg See note 2!and key in number
SPECIFICATION SETTING 4 This setting only use in unit kg, g, lb, oz! and the division is without limit.
.
.
ON/ZERO key Upper key (Selects number 0,1,2,3,4,5,6,7,8,9) UNITS key Down key (Selects number 0,1,2,3,4,5,6,7,8,9) TARE key Right key, blinking word move to right NET/GROSS key Left key, blinking word move to left
The screen displays max. capacity +9d and unit weight for 2 seconds, then back to previous setting function.
The screen displays previous max. capacity +9d and unit weight for 2 seconds.
Example: Amount of unit weights are 2 units kg!lb! and kg! is pointed unit weight.
Example: 015000 15.000 kg For example: 1 1 Division 3 Position of decimal at 3! Press TARE key - 7 - 02005012 Note 1
(a) (b) (c) (d) (e) (f ) (a) First unit weight (unit weight only kg!!g!!lb! , Please select one of number 0!1!2) (b) Second unit weight (Please select one of number 0 ~ 9) (c) Third unit weight (Please select one of number 0 ~ 9) (d) Fourth unit weight (Please select one of number 0 ~ 9) (e) Fifth unit weight (Please select one of number 0 ~ 9) (f) Amount of unit weight ( Please select one of number 1 ~ 5) The description of number: 0 kg 1 g 2 lb 4 lb!oz 5 oz
For example: Amount of unit weights are 2 units kg!lb! and kg! is pointed unit weight, then enter 020002 Note 2
(g) ( h) ( i ) ( j ) ( k) ( l ) Please enter maximum setting capacity, total is 6 numbers (not include 9d) For example: 15.000 kg enter 015000 1500.0 g enter 015000 6.000 lb enter 006000 Note 3
(m) ( n) (m) Division setting( please select one of number 1!2!5)
The description of number: 1 1 division 2 2 division 5 5 division
(n) The position of decimal ( 0 ~ 5) For example: 15.000 kg enter 3 1500.0 g enter 1 6.000 lb enter 3
- 8 - 02005012
STEPS OF SPECIFICATION SETTING Capacity Note 1 (Unit weight) Note 2 (Max. capacity setting) Note 3 (Division & decimal setting) Display 30kg X 10g 000001 003000 120000 30.01kg 6kg X 1g 100001 006000 100000 6001g 30kg X 5g 000001 030000 530000 30.005kg 3kg X 0.5g 100001 030000 510000 30005g 15kg X 5g 000001 015000 530000 15.005kg 6kg X 2g 100001 006000 200000 6002g 3kg X 1g 100001 003000 100000 3001g
- 9 - 02005012 After key in 00!, press TARE key, return to previous setting. (Exit from weight calibration)
WEIGHT CALIBRATION
ON/ZERO key Upper key (Selects number 0,1,2,3,4,5,6,7,8,9) UNITS key Down key (Selects number 0,1,2,3,4,5,6,7,8,9) TARE key Right key, blinking word move to right NET/GROSS key Left key, blinking word move to left
Return to previous setting Weight Calibration setting Gravity setting at manufacture
Press TARE key Key in 00! or 01! - 10 - 02005012 When the right word blinking, please put weight pan without any weight on scale then press TARE key kg kg Please put weight on the weight pan as screen displays value, press TARE key kg kg Take away all weight and press any key, finish weight calibration setting. Press TARE key, finishing weight calibration setting. WEIGHT CALIBRATION SETTING
.
.
.
ON/ZERO key Upper key (Selects number 0,1,2,3,4,5,6,7,8,9) UNITS key Down key (Selects number 0,1,2,3,4,5,6,7,8,9) TARE key Right key, blinking word move to right NET/GROSS key Left key, blinking word move to left
Press TARE key Screen displays weight value (The maximum capacity value of first unit) and unit weight and right word blinking. Scale read zero information, after stabling, the screen displays weight value with blinking. Scale read information of full capacity, after stabling, will beep! three times and screen displays weight value with blinking. Then take away all weight and test weighing. After stabling Key in number 00! - 11 - 02005012 Show last value Factory : Taipei Factory: shanghai Factory: others
!"# Gravity setting at manufacture 4 The gravity value is between equator and polar region. Gravity value at equator G E = 9.7803184558 m/sec 2
Gravity value at polar region G P = 9.8321772792 m/sec 2 Taipei " 9.78914 m/sec 2 Shanghai " 9.79585 m/sec 2
# TARE $ Press TARE press TARE Press TARE press TARE - 12 - 02005012 Confirm free of weight, press TARE ,enter linearity calibration Take off steel platter, remain plastic one, confirm free of load, press TARE key, record first point! After place first mas, enter mass value,press TARE key,record second point! Add second mass, enter mass value, press TARE key,record!third point! Add eighth mass, enter mass value, press TARE key,record!ninth point!
Add third mass, enter mass value, press TARE key,record!fouth point!
Linearity Calibration 4 After the linearity calibration, the weight calibration should be re-done. 4 The range of linearity calibration should include the weight calibration
ON/ZERO figure up key Enter 0,1,2,3,4,5,6,7,8,9 UNITS figure down key Enter 0,1,2,3,4,5,6,7,8,9 TARE shift right NET/GROSS shift left
*mass% - 13 - 02005012 *Mass proportion % 1. If the weight proportion of first mass is n!%&'()*+,-.*+/0n 2. the weight proportion of second mass is m!, then mass value should enter nm 3. the weight proportion of third mass is p!, then mass value should enter np 4 max value of mass is 15 ,enter F !
1: max calibration is 9 ( ~ ) during linearity calibration press TARE key, abort linearity calibration during linearity calibration type press NET/GROSS key, abort calibration during linearity calibration type press NET/GROSS key, abort calibration during linearity calibration type press NET/GROSS key, abort point 2 during linearity calibration type press NET/GROSS key, abort point 3 during linearity calibration type press NET/GROSS key, abort point 4 during linearity calibration type press NET/GROSS key, abort point 5 during linearity calibration type press NET/GROSS key, abort point 6 during linearity calibration type press NET/GROSS key, abort point 7 during linearity calibration type press NET/GROSS key, abort point 8 during linearity calibration type press NET/GROSS key, enter point 9, end
- 14 - 02005012 Function setting 4 environment parameter setting ! approval parameter setting and buzzer install setting
ON/ZERO figure up key Enter 0,1,2,3,4,5,6,7,8,9 UNITS figure down key Enter 0,1,2,3,4,5,6,7,8,9 TARE shift right NET/GROSS shift left
back to upper level environment parameter approval parameter buzzer install press TARE Press 00!01!02!03 After enter00!, press TARE back to upper level
- 15 - 02005012 Using ON/ZERO or UNITS to enter Using ON/ZERO or UNITS to enter Using ON/ZERO or UNITS to enter !%# Environment setting 4 if revise parameter at , then will be revise automatically
ON/ZERO figure up key Enter 0,1,2,3,4,5,6,7,8,9 UNITS figure down key Enter 0,1,2,3,4,5,6,7,8,9 TARE shift right NET/GROSS shift left
Press TARE Return zero point setting Using ON/ZERO or UNITSto enter Default setting = 0 0 show all 5 cover 5 d 1 cover 1 d 6 cover 6 d 2 cover 2 d 7 cover 7 d 3 cover 3 d 8 cover 8 d 4 cover 4 d 9 cover 9 d 4 weight value must over 1/3 full scale Digital switch &stable range parametersetting Screen show last setting Filter parameter setting Screen show last setting Press TARE to finish environment setting Enter 00! Return zero point setting display last setting
Filter parameter setting Using ON/ZERO or UNITS to enter Default setting = 5 Parameter 0 ~ 9 , as larger as quicker filter reaction. Digital switch &stable range parametersetting Using ON/ZERO or UNITS to enter Default setting = 0 Parameter 0 ~ 9 , As larger figure as easier filter stable Press TARE Press TARE Press TARE - 16 - 02005012
- 17 - 02005012 Using ON/ZERO or UNITS to enter !"# Approval setting
ON/ZERO figure up key Enter 0,1,2,3,4,5,6,7,8,9 UNITS figure down key Enter 0,1,2,3,4,5,6,7,8,9 TARE shift right NET/GROSS shift left
Press TARE Approval setting Screen show last display Press TARE to finish setting Enter 00! Approval setting Using ON/ZERO or UNITSto setting 0 non-approval, allow external calibration 1 OIML or NTEP, not allow external calibration 2 Sri Lanka(non-approval), allow external calibration 3 Sri Lanka(approval), not allow external calibration 4 when allow external calibration, E1!E2 range will not detect, Press TARE - 18 - 02005012 Using ON/ZERO or UNITS To setting !&# Buzzer install setting
ON/ZERO figure up key Enter 0,1,2,3,4,5,6,7,8,9 UNITS figure down key Enter 0,1,2,3,4,5,6,7,8,9 TARE shift right NET/GROSS shift left
Buzzer install setting screen show last setting Press TARE To finish setting Press00! Buzzer install setting Using ON/ZERO or UNITS to enter
0 (DC,PT-134OP) 1 (AC,UIC-4201)
Press TARE Press TARE - 19 - 02005012 Please key in gravity value with user position. !'#ACCELERATION OF GRAVITY SETTING
4 The value of acceleration of gravity has to between equator gravity and pole gravity.
ON/ZERO key Upper key (Selects number 0,1,2,3,4,5,6,7,8,9) UNITS key Down key (Selects number 0,1,2,3,4,5,6,7,8,9) TARE key Right key, blinking word move to right NET/GROSS key Left key, blinking word move to left