- 1 - 02005012

Press UNITS key


Introduction
Before calibration, to do inspection of PCB :
Please adjust mini jumper SWA1to LOCK, press ON/ZERO key to test LCD.
Are there with shadow or word break? (After calibration the problems will be
solved.)

Press OFF key to power off then press NET/GROSS and hold and press
ON/ZERO key at same time. When screen displays into self
testing module




















Press ON/ZERO key Press UNITS key
Press ON/ZERO key
Press ON/ZERO key
Press UNITS key
Press UNITS key
Press ON/ZERO key
- 2 - 02005012


Internal value modal (must connect to
Load cell with full bridge rectifier for testing)
Press TARE , screen displays internal value.
Please check internal value whether it during normal range
(10000 ~ 150000 for empty)
Please check backlight.
Four corner modification (The internal value of four corner error should be
under 10d)
After four corner checking, please adjust mini jumper SWA1 to ADJ, then
into calibration or press ON/ZERO key return to previous setting. The screen
displays
Keypad testing
Press TARE key, the screen displays
internal value of keypad: TARE key= 07 ! UNITS key= 06 !
NET/GROSS key =05
Press ON/ZERO key, return to previous setting, the screen displays

Program version display
Press TARE key, the screen displays program version ,
Press TARE key again, the screen displays maintenance program version
for 2 seconds.
Press ON/ZERO key, return to previous setting, the screen displays

Return to previous setting
Press TARE key, out of self testing and the scale power on automatically.
- 3 - 02005012
ERROR MESSAGE
EEPROM problem ( Cannot read data)
Zero rang when switching on scale is too high.
Zero range when switching on scale is too low.
Internal value is unstable.
Internal value is under Zero.
AD problem (cannot read AD value)
Weight over full capacity + 9 division
AD is over the maximum resolution range.

WEIGHT UNIT
(kg) 1 g = 0.001 kg
(g) 1 g = 1 g
(lb) 1 g = 0.002204623 lb
(lb,oz) 1 g = 0.03527396 oz
(oz) 1 g = 0.03527396 oz
(GN) 1 g = 15.432358 GN
(dwt) 1 g = 0.6430149 dwt
(ct) 1 g = 5 ct



- 4 - 02005012
Press UNITS key
INTERNAL FUNCTION SETTING
4 Please adjust mini jumper SWA1to ADJ (EEPROM UNLOCK),
Switch on, into calibration function setting.
The screen displays .
After finishing calibration function setting, should adjust MINI JUMPER to
LOCK
If adjust MINI JUMPER to LOCK, will exit from setting model when doing
calibration function setting.





























Specification setting Function setting

Weighing calibration Acceleration of gravity

Linearity Calibration setting
Press ON/ZERO key Press UNITS key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press UNITS key
Press UNITS key
Press UNITS key
- 5 - 02005012
After key in 00!, press
TARE key, return to previous
setting. (Exit from
specification
setting)
SPECIFICATION CALIBRATION


































ON/ZERO key Upper key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key Right key, blinking word move to right
NET/GROSS key Left key, blinking word move to left


Return to previous setting
Specification setting
Press TARE key
Key in 00! or 01!
- 6 - 02005012
kg
Press TARE key
See note 3!and key in number

Press TARE key
Press TARE key
See note 1!and key in number

Press TARE key,
finish specification setting.
Key in number 00!
kg
See note 2!and key in number

SPECIFICATION SETTING
4 This setting only use in unit kg, g, lb, oz! and the division is without limit.



















































.













.



ON/ZERO key Upper key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key Right key, blinking word move to right
NET/GROSS key Left key, blinking word move to left




The screen displays max. capacity
+9d and unit weight for 2 seconds,
then back to previous setting function.

The screen displays previous
max. capacity +9d and unit weight
for 2 seconds.

Example:
Amount of unit weights are 2 units
kg!lb! and kg! is pointed unit
weight.

Example:
015000 15.000 kg
For example:
1 1 Division
3 Position of decimal at 3!
Press TARE key
- 7 - 02005012
Note 1

(a) (b) (c) (d) (e) (f )
(a) First unit weight (unit weight only kg!!g!!lb! , Please select one of number 0!1!2)
(b) Second unit weight (Please select one of number 0 ~ 9)
(c) Third unit weight (Please select one of number 0 ~ 9)
(d) Fourth unit weight (Please select one of number 0 ~ 9)
(e) Fifth unit weight (Please select one of number 0 ~ 9)
(f) Amount of unit weight ( Please select one of number 1 ~ 5)
The description of number:
0 kg
1 g
2 lb
4 lb!oz
5 oz

For example: Amount of unit weights are 2 units kg!lb! and kg! is pointed
unit weight, then enter 020002
Note 2

(g) ( h) ( i ) ( j ) ( k) ( l )
Please enter maximum setting capacity, total is 6 numbers (not include 9d)
For example: 15.000 kg enter 015000 1500.0 g enter 015000
6.000 lb enter 006000
Note 3

(m) ( n)
(m) Division setting( please select one of number 1!2!5)

The description of number:
1 1 division
2 2 division
5 5 division

(n) The position of decimal ( 0 ~ 5)
For example: 15.000 kg enter 3 1500.0 g enter 1 6.000 lb enter 3





- 8 - 02005012

STEPS OF SPECIFICATION SETTING
Capacity Note 1
(Unit weight)
Note 2
(Max. capacity setting)
Note 3
(Division & decimal setting)
Display
30kg X 10g 000001 003000 120000 30.01kg
6kg X 1g 100001 006000 100000 6001g
30kg X 5g 000001 030000 530000 30.005kg
3kg X 0.5g 100001 030000 510000 30005g
15kg X 5g 000001 015000 530000 15.005kg
6kg X 2g 100001 006000 200000 6002g
3kg X 1g 100001 003000 100000 3001g

- 9 - 02005012
After key in 00!, press
TARE key, return to previous
setting. (Exit from weight
calibration)

WEIGHT CALIBRATION






























ON/ZERO key Upper key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key Right key, blinking word move to right
NET/GROSS key Left key, blinking word move to left





Return to previous setting
Weight Calibration setting
Gravity setting at manufacture

Press TARE key
Key in 00! or 01!
- 10 - 02005012
When the right word blinking,
please put weight pan without
any weight on scale then
press TARE
key
kg
kg
Please put weight on the
weight pan as screen displays
value, press
TARE key
kg
kg
Take away all weight and press
any key, finish weight
calibration setting.
Press TARE key,
finishing weight calibration
setting.
WEIGHT CALIBRATION SETTING




















































.



.

.



ON/ZERO key Upper key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key Right key, blinking word move to right
NET/GROSS key Left key, blinking word move to left




Press TARE key
Screen displays weight value (The
maximum capacity value of first unit) and
unit weight and right word blinking.
Scale read zero information, after
stabling, the screen displays
weight value with blinking.
Scale read information of full capacity,
after stabling, will beep! three times
and screen displays weight value with
blinking. Then take away all weight and
test weighing.
After stabling
Key in number 00!
- 11 - 02005012
Show last value
Factory : Taipei Factory: shanghai Factory: others

!"# Gravity setting at manufacture
4 The gravity value is between equator and polar region.
Gravity value at equator G
E
= 9.7803184558 m/sec
2

Gravity value at polar region G
P
= 9.8321772792 m/sec
2
Taipei " 9.78914 m/sec
2
Shanghai " 9.79585 m/sec
2









































$1:
Use ON/ZERO or UNITS key
To choose factory

.

.







ON/ZERO Choose figure
0,1,2,3,4,5,6,7,8,9
UNITS Choose figure
0,9,8,7,6,5,4,3,2,1
TARE shift right,
NET/GROSS shift left,

# TARE $
Press TARE
press TARE
Press TARE
press TARE
- 12 - 02005012
Confirm free of weight,
press TARE ,enter linearity
calibration
Take off steel platter, remain plastic
one, confirm free of load, press TARE
key, record first point!
After place first mas, enter mass
value,press TARE key,record
second point!
Add second mass, enter mass value,
press TARE key,record!third point!
Add eighth mass, enter mass value,
press TARE key,record!ninth point!

Add third mass, enter mass value,
press TARE key,record!fouth point!

Linearity Calibration
4 After the linearity calibration, the weight calibration should be
re-done.
4 The range of linearity calibration should include the weight
calibration







































ON/ZERO figure up key
Enter 0,1,2,3,4,5,6,7,8,9
UNITS figure down key
Enter 0,1,2,3,4,5,6,7,8,9
TARE shift right
NET/GROSS shift left





*mass%
- 13 - 02005012
*Mass proportion %
1. If the weight proportion of first mass is n!%&'()*+,-.*+/0n
2. the weight proportion of second mass is m!, then mass value should enter
nm
3. the weight proportion of third mass is p!, then mass value should enter np
4 max value of mass is 15 ,enter F !


1:
max calibration is 9 ( ~ )
during linearity calibration press TARE key, abort linearity calibration
during linearity calibration type press NET/GROSS key, abort
calibration
during linearity calibration type press NET/GROSS key, abort calibration
during linearity calibration type press NET/GROSS key, abort point 2
during linearity calibration type press NET/GROSS key, abort point 3
during linearity calibration type press NET/GROSS key, abort point 4
during linearity calibration type press NET/GROSS key, abort point 5
during linearity calibration type press NET/GROSS key, abort point 6
during linearity calibration type press NET/GROSS key, abort point 7
during linearity calibration type press NET/GROSS key, abort point 8
during linearity calibration type press NET/GROSS key, enter point 9,
end


- 14 - 02005012
Function setting
4 environment parameter setting ! approval parameter
setting and buzzer install setting































ON/ZERO figure up key
Enter 0,1,2,3,4,5,6,7,8,9
UNITS figure down key
Enter 0,1,2,3,4,5,6,7,8,9
TARE shift right
NET/GROSS shift left





back to upper level
environment parameter
approval parameter
buzzer install
press TARE
Press 00!01!02!03
After enter00!,
press TARE back to upper level

- 15 - 02005012
Using ON/ZERO or UNITS
to enter
Using ON/ZERO or UNITS
to enter
Using ON/ZERO or UNITS
to enter
!%# Environment setting
4 if revise parameter at , then will be revise
automatically


































































ON/ZERO figure up key
Enter 0,1,2,3,4,5,6,7,8,9
UNITS figure down key
Enter 0,1,2,3,4,5,6,7,8,9
TARE shift right
NET/GROSS shift left


Press TARE
Return zero point setting
Using ON/ZERO or UNITSto enter
Default setting = 0
0 show all 5 cover 5 d
1 cover 1 d 6 cover 6 d
2 cover 2 d 7 cover 7 d
3 cover 3 d 8 cover 8 d
4 cover 4 d 9 cover 9 d
4 weight value must over 1/3 full scale
Digital switch &stable range
parametersetting
Screen show last setting
Filter parameter setting
Screen show last setting
Press TARE to finish
environment setting
Enter 00!
Return zero point setting display last
setting

Filter parameter setting
Using ON/ZERO or UNITS to enter
Default setting = 5
Parameter 0 ~ 9 , as larger as quicker filter
reaction.
Digital switch &stable range parametersetting
Using ON/ZERO or UNITS to enter
Default setting = 0
Parameter 0 ~ 9 ,
As larger figure as easier filter stable
Press TARE
Press TARE
Press TARE
- 16 - 02005012


- 17 - 02005012
Using ON/ZERO or UNITS
to enter
!"# Approval setting























































ON/ZERO figure up key
Enter 0,1,2,3,4,5,6,7,8,9
UNITS figure down key
Enter 0,1,2,3,4,5,6,7,8,9
TARE shift right
NET/GROSS shift left



Press TARE
Approval setting
Screen show last display
Press TARE to finish
setting
Enter 00!
Approval setting
Using ON/ZERO or UNITSto setting
0 non-approval, allow external
calibration
1 OIML or NTEP, not allow
external calibration
2 Sri Lanka(non-approval), allow
external calibration
3 Sri Lanka(approval), not allow
external calibration
4 when allow external calibration,
E1!E2 range will not detect,
Press TARE
- 18 - 02005012
Using ON/ZERO or UNITS
To setting
!&# Buzzer install setting





















































ON/ZERO figure up key
Enter 0,1,2,3,4,5,6,7,8,9
UNITS figure down key
Enter 0,1,2,3,4,5,6,7,8,9
TARE shift right
NET/GROSS shift left





Buzzer install setting
screen show last setting
Press TARE
To finish setting
Press00!
Buzzer install setting
Using ON/ZERO or UNITS to enter

0 (DC,PT-134OP)
1 (AC,UIC-4201)

Press TARE
Press TARE
- 19 - 02005012
Please key in gravity
value with user position.
!'#ACCELERATION OF GRAVITY SETTING






















4 The value of acceleration of gravity has to between equator gravity
and pole gravity.

Gravity of Equator : G
E
= 9.7803184558 m/sec
2

Gravity of Pole : G
P
= 9.8321772792 m/sec
2

Taipei " 9.78914 m/sec
2
Shanghai " 9.79585 m/sec
2














.

ON/ZERO key Upper key
(Selects number 0,1,2,3,4,5,6,7,8,9)
UNITS key Down key
(Selects number 0,1,2,3,4,5,6,7,8,9)
TARE key Right key, blinking word move to right
NET/GROSS key Left key, blinking word move to left

.


Press TARE key
Press TARE key