0 evaluări0% au considerat acest document util (0 voturi)
25 vizualizări6 pagini
A novel preterm respiratory mechanics active simulator to test the performances of neonatal pulmonary ventilators is proposed. The implemented prototype is capable of reproducing values of the parameters of pulmonary mechanics of healthy newborns and preterm pathological infants. Accuracy and reliability on the order of 1% is estimated.
A novel preterm respiratory mechanics active simulator to test the performances of neonatal pulmonary ventilators is proposed. The implemented prototype is capable of reproducing values of the parameters of pulmonary mechanics of healthy newborns and preterm pathological infants. Accuracy and reliability on the order of 1% is estimated.
A novel preterm respiratory mechanics active simulator to test the performances of neonatal pulmonary ventilators is proposed. The implemented prototype is capable of reproducing values of the parameters of pulmonary mechanics of healthy newborns and preterm pathological infants. Accuracy and reliability on the order of 1% is estimated.
A novel preterm respiratory mechanics active simulator to test
the performances of neonatal pulmonary ventilators
Paolo Cappa a),b) Department of Mechanics and Aeronautics, University of Rome LA SAPIENZA, Via Eudossiana 18, 00184 Rome, Italy Salvatore Andrea Sciuto a),c) Department of Mechanical and Industrial Engineering, University of Rome ROMA TRE, Via della Vasca Navale 79, 00146 Rome, Italy Sergio Silvestri a),d) School of Biomedical Engineering, University CAMPUS BIO-MEDICO, Via Longoni 83, 00155 Rome, Italy Received 28 November 2001; accepted for publication 26 March 2002 A patient active simulator is proposed which is capable of reproducing values of the parameters of pulmonary mechanics of healthy newborns and preterm pathological infants. The implemented prototype is able to: a let the operator choose the respiratory pattern, times of apnea, episodes of cough, sobs, etc., b continuously regulate and control the parameters characterizing the pulmonary system; and, nally, c reproduce the attempt of breathing of a preterm infant. Taking into account both the limitation due to the chosen application eld and the preliminary autocalibration phase automatically carried out by the proposed device, accuracy and reliability on the order of 1% is estimated. The previously indicated value has to be considered satisfactory in light of the eld of application and the small values of the simulated parameters. Finally, the achieved metrological characteristics allow the described neonatal simulator to be adopted as a reference device to test performances of neonatal ventilators and, more specically, to measure the time elapsed between the occurrence of a potentially dangerous condition to the patient and the activation of the corresponding alarm of the tested ventilator. 2002 American Institute of Physics. DOI: 10.1063/1.1480453 I. INTRODUCTION Nearly 60% of newborn babies with severe respiratory diseases need ventilatory support performed by means of as- sisted or controlled mechanical ventilation. This task is gen- erally fullled by pulmonary ventilators specically de- signed for neonatal use. The constraints for infant pulmonary ventilators are denitely more severe in terms of reliability, stability, and accuracy, due to the following indicated rea- sons. First, the small volumes delivered to the patient, in the range of about 4 ml per breath, represent a strict constraint to the design of the tester. Second, further constraints are: a the extreme delicateness and vulnerability of the neonatal respiratory system, mainly due to both the reduced mass of a preterm infant which can be in the order of 0.5 kg, and b the highest tolerable lung pressures that should be less than about 1.01.5 kPa 1015 cm H 2 O. 14 Third, the typical time length of the treatment, that can even last longer than three months, induces further limitation to the design of the experimental setup. As is well known, the ventilation system Fig. 1 is com- posed of a pulmonary ventilator, an inspiration and expira- tion circuit, a humidier, an endotracheal tube, and a patient. Depending on both the specic behavior of each component and the high number of parameters characterizing the men- tioned system, difculties immediately emerge in the evalu- ation of the performances of this intensive care equipment in terms of the actual amount of gaseous mixture that reaches the alveolus-blood capillary interface, especially when the patients are very small i.e., premature infants at 56 months gestational period. Moreover, most of the commercially available neonatal ventilators measure pressure and ow only at ventilator level, i.e., upstream the inspiratory circuit see Fig. 1. As a consequence, the actual values of air pres- sure in the bronchial tree and volume delivered to the alveoli are not directly measured in situ but calculated on the basis of the parameter values measured upstream the circuit and correcting them to take into account a standard pressure drop and a standard compressible volume along the inspiratory and expiratory tube. Therefore, it seems obvious that in the case of very small patients, like preterm infants, the ne regulation of the pulmonary ventilator parameters requires specic attention because it can cause pathologies of the pul- monary apparatus. Neonatal pediatricians are particularly worried about iatrogenic damages that frequently occur in newborn babies subjected to mechanical ventilation, mainly because of their critical conditions and the vulnerability of a Present address: Clinical Engineering Service, Childrens Hospital BAM- BINO GESU ` of Rome, piazza S. Onofrio 4, 00165, Rome, Italy. b Electronic mail: paolo.cappa@uniroma1.it c Electronic mail: sciuto@uniroma3.it d Electronic mail: s.silvestri@unicampus.it REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 73, NUMBER 6 JUNE 2002 2411 0034-6748/2002/73(6)/2411/6/$19.00 2002 American Institute of Physics their cardiorespiratory apparatus. 5,6 Finally, during the treat- ment, unpredictable events, such as coughs, sobs, and sighs, can cause the patient to breath against the ventilator that is supposed to recognize such events and adapt its operating conditions in order to avoid dangerous situations. 7 The aforementioned reasons, in addition to fatalities caused by an inadequate ventilation 810 showed the necessity to test the performances of neonatal ventilators by means of an active mechanical simulator of the respiratory system spe- cic for a preterm infant. Such a device should be able to: a provide similar physiological mechanical characteristics of the pulmonary system, even in pathological conditions, b simulate unexpected events, and, nally, c monitor online respiratory parameters such as airway pressure, volume, and ow. The analysis carried out on specic scientic literature and on instrumentation presently in the market outlined the lack of testers able to simulate mechanical properties of lungs in the whole range of premature infant respiratory pa- thologies. This issue is conrmed by the specic needs pre- sented by some pathologies associated with an elevated stiff- ness of the respiratory system, such as neonatal respiratory distress syndrome NRDS associated with a low pulmonary compliance on the order of 1 ml/kPa 0.1 ml/cm H 2 O, in connection to small volumes delivered per breath. In fact, commercially available active simulators generally use com- pliant rubber bellows, 11,12 often equipped with springs, 13,14 or rigid containers 15,16 to simulate pulmonary compliances but in a limited range of variations 17 and are not able to reach low compliance values such as those previously indicated. Moreover, passive simulators, mainly devoted to test venti- lator performances in an endurance test, are currently cov- ered by International 18 and American 19,20 standards and are based on mechanical systems composed by tubes, resis- tances, and compliances, such as rigid volumes, manufac- tured with standard dimensions. International Standards Or- ganization ISO standard, 18 for example, denes the values of a pulmonary mechanical tester for infant and adult venti- lators; more specically, the values of compliance C and resistance R indicated for infant simulation are respectively equal to 10 ml/kPa 1 ml/cm H 2 O and 5 kPa s l 1 50 cm H 2 O s l 1 . However, these values are not representative of the complete range of possible pathological conditions of premature newborns: 21,22 in fact, for these patients, C can reach 23 values that lie in the range of 140 ml/kPa 0.14.0 ml/cm H 2 O and R can vary in the range 24 from 3 to 15 kPa s l 1 30 to 150 cm H 2 O s l 1 . Furthermore, some very repu- table worldwide manufacturers of testers for biomedical in- strumentation commercialize pulmonary simulators not in- tended for exclusive infant use. However, none of these devices is able to reach the herein reported C and R values, they are only able to simulate C values in the range from 10 to 1019 ml/kPa 1 to 100 ml/cm H 2 O and a lower value for tidal volume, i.e., the amount of air inspired during each breath, equal to 5 ml only while, in the presence of NRDS or broncopulmonary disease BPD, it can be lower than 4 ml. From the aforementioned synthesized considerations, the usefulness of a device able to simulate the mentioned values of respiratory parameters specic for healthy and pathologi- cal conditions for neonatal patients emerges. Therefore, the main task of the present article is to design an innovative tester for pulmonary ventilators specic for infant use. Then, a prototype has been implemented and metrologically char- acterized in order to be utilized to test infant pulmonary ven- tilators. II. INSTRUMENT DESCRIPTION A novel device is proposed and realized in order to ac- curately and reliably reproduce the preterm respiratory me- chanics. The simulator is composed of two main parts, sche- matically represented in Fig. 2, the electronic unit EU and the pneumatic/mechanical unit PMU. The EU consists of a personal computer PC able to operate a software control of the PMU through a digital to analog D/A 12-bit converter and an analog to digital A/D 16-bit converter; both convert- ers are available on the same multi-input/output plug-and- play data conversion board. The software was specically designed to allow the operator to set all the parameters useful to characterize a chosen respiratory pattern, i.e., tidal vol- ume, frequency, inspiratory and expiratory time, type of simulated pathology, apnea periods, sobs, coughs, and other unexpected events, through an easy-to-use graphical user in- terface. The respiratory pattern is then repeated any number of times according to the chosen test time length. The D/A converts the programmed digital pattern, representing the respiratory volume changes, into an analog voltage signal to drive the shaker S through a power amplier PA. The airway pressure and ow are measured by means of specic transducers T. Signals provided by transducers are then l- FIG. 1. Scheme of the ventilation system. FIG. 2. Schematic representation of the infant pulmonary simulator. A/D and D/A, analog to digital and digital to analog converters; B, bellows; CR, clamp resistance; EF, elliptic lters; EU, electronic unit; T, ow and pres- sure transducers; PA, power amplier; PC, personal computer; PMU, pneumatic/mechanical unit; PV, pulmonary ventilator; S, electromechanical shaker; SC, three-way stopcock; and VAO, ventilator analog outputs. 2412 Rev. Sci. Instrum., Vol. 73, No. 6, June 2002 Cappa, Sciuto, and Silvestri tered by means of an elliptic low-pass lter 25 EF in order to drastically reduce noise due to air humidity and ow turbu- lence, and to avoid aliasing in data acquisition. Pressure, ow, and alarm signals provided as ventilator analog outputs VAO are directly gathered by the data conversion board. Once digitized, signals are postprocessed to calculate pulmo- nary parameters to be displayed on the PC monitor. Methods and equations to calculate the whole set of respiratory me- chanical parameters from pressure and ow wave forms are software implemented to online visualize and continuously update the parameter mean values with the associated stan- dard deviations. As a nal task, the simulator allows the storing and printing of the respiratory parameter trend as a function of time for further evaluation by physicians and technicians. As further processing, a program routine evalu- ates: a the delay time of ventilator apnea alarm since the beginning of simulated apnea by the tester and b the differ- ences between the set respiratory parameters and those mea- sured and visualized by the examined pulmonary ventilator. A ow chart of the methodology of the device is reported in Fig. 3. As regards the PMU, it is necessary to observe that pre- vious studies 26 conrm that, even though each of the millions of alveoli could act as a separate mechanical unit, the me- chanics of normal or pathological lungs during quiet breath- ing can be represented by a single or double compartment model. As a consequence, a double compartment model has been chosen to simulate, by means of lumped parameters, the mechanical behavior of two lungs with different mechanical characteristics. Thus, the PMU of the device consists of two silicone bellows B of compliance certied by the manufacturer 27 both connected to an electromechanical shaker S that is able at the initial phase to drive their ex- pansion during inspiratory phase, by means of the volume variation law generated by the software, and then to induce their collapse during the expiratory phase. One of the two bellows is placed in a specically designed damping system described more fully in the following section in order to have a compartment with variable compliance and to avoid high- frequency turbulent ow oscillations that led previous efforts in making infant pulmonary simulators to fail due to a poor data repeatability. 11,28 In particular, the variation of the bel- low volume as a function of time is indicated as critical for the simulation of ne compliance variations in respiratory patterns associated with pathological conditions. Bellows are connected to the articial trachea through two tubes and a three-way stopcock SC which permits the exclusion of one of two bellows in order to simulate sym- metric pulmonary diseases. The three-way SC can be substi- tuted by a simple Y piece in the case of simulation of asym- metric diseases, i.e., the frequent case of a pathology affecting only one lung, where an intrapulmonary shunt is realized by means of the difference of the mechanical char- acteristics of the lungs in order to actuate the typical low- frequency oscillations in pressure and ow signals. The pressure transducer 29 is placed between the articial trachea and the bellows to measure, at lung level, the relative pressure induced by the volume change of the pneumatic system. As tidal volume was chosen as the control variable for actuator feedback, a high accuracy level is required on its measurement. Tidal volume is then measured by digital inte- gration of air ow passing through the articial trachea by means of a volume measurement procedure implemented in the proposed system and accomplished by means of a high accuracy owmeter. 30 The chosen owmeter is based on a double hot wire anemometer 31 specically designed for in- fant respiratory measurements and placed in a rigid 5 mm diameter cylinder. One of the anemometers is shielded on one side in order to allow this device to discriminate the direction of ow. The integrator was previously calibrated by means of a 10 ml syringe placed into an infusion pump able to generate a constant 0.1 ml/s ow. In consideration of the metrological performances of the owmeter, the overall ac- curacy in volume variation measurements was equal to about 0.02 ml for a breathing frequency range from 20 to 120 breath per minute bpm. In consideration of the aforemen- tioned observations, the chosen ow transducer presents the following advantages: a the accuracy is adequate to mea- sure small ows like those produced by preterm infant breathing, 32 b it allows one to discriminate the direction of ow, c due to its reduced dimensions, it adds an unnotice- able dead space to the breathing circuit, d introduces a low pressure drop, 33 and, nally, e depending on the operating temperature of the wires, about 400 C, it is not affected by condensate. During system calibration, temperature and pressure test areas have been monitored and their effects on transducer calibration curves were found negligible, conrming data FIG. 3. Flow chart of the infant pulmonary simulator methodology. 2413 Rev. Sci. Instrum., Vol. 73, No. 6, June 2002 Active neonatal breathing simulator sheet values provided by manufacturers. With reference to Fig. 4, one bellow was placed in a cylindrical container 34 partially lled up with about 300 ml of oil, in order to obtain low values for the simulated pulmonary compliance. The controlled damping system was implemented with two per- forated disks aligned with the bellows; the lower disk LD was xed to the container and the upper disk UD could be rotated by an axial rod. The relative position of the UD with respect to the LD allows one to set the orices span OS where oil ows through during bellow expansion or collapse. As is well known, pressure drop across an orice is depen- dent on uid dynamic viscosity, , and cross-sectional area of the orice itself. As dynamic uid viscosity strongly de- pends on temperature, a diathermic oil 35 was chosen and its temperature was continuously monitored during preliminary tests, performed in accordance to the conguration shown in Fig. 4. The mentioned tests are conducted in order to evalu- ate the relationship between the damping force of the system, the orice cross-sectional area, and the operating oil tem- perature. The data collected showed negligible effects in- duced by temperature variations in the range of the observed test area temperature, i.e., 235 C. Moreover, the men- tioned preliminary tests outlined that the amount of oil in the container and, more specically its level upon the perforated disks, is critical due to three main reasons: a the thermal capacity and the heat dissipation of the oil, b the hydro- static pressure on the bellow, and, nally, c a minimum level is necessary to reduce boundary effects of oil efux through the holes during breathing simulation in such a man- ner as to increase the system repeatability. During these tests, a xed amount of oil was gradually added to the container and an optimal level of 20 mm over the disks was found leading to standard deviations less than 1% for each of the monitored respiratory parameters. To complete the simulator setup, a variable clamp resis- tance CR, see Fig. 4, was placed along the articial trachea AT to simulate the upper airway resistance to ow. The lowest value of air ow, in the order of 0.5 l/min, is reached by manually reducing the AT section. III. PERFORMANCE EVALUATION AND DISCUSSION In order to evaluate the performances of the realized tester, it is important to outline that, as previously mentioned, the respiratory mechanics of the preterm patient are simu- lated by means of a rst-order system due to the almost negligible mass of the delivered air to the patient. Thus, simulator behavior can be characterized completely by means of a single characteristic parameter, i.e., the time con- stant , representing the product of an elastic term, the lung compliance, and a resistive one, the airway resistance. There- fore, once the pathology and the patient characteristics to simulate are chosen, the tidal volume and are unequivo- cally dened. As a consequence, the evaluation of the met- rological performances of the implemented device was con- ducted by paying particular attention to the simulation of pulmonary compliance that is the most critical parameter of an infant respiratory simulator. In fact, when is individual- ized, it is possible to impose a C value, with the resulting value of R consequently determined. It should be noted that, even though with the described prototype, it is not possible to change R and C independently, this does not represent a limitation. In fact, as reported in Ref. 25, airway resistance in pulmonary function tests is calculated, at rest breathing rates, as the ratio R/C, and, then, R is not representative of the only hydrodynamic airway resistance to ow but, more gen- erally, of the dissipation of the whole respiratory system. In particular, the proposed prototype allows one to regulate the compliance value in two different ways: a by changing disk OS, which leads to a different damping force exerted on the bellows during breathing volume variations, and b by modifying the shape of the exponential signal that drives bellow collapse during the expiratory phase, i.e., varying its time constant . At the initial stage, it was decided to conduct prelimi- nary acquisition on healthy patients by means of a commer- cially available ow transducer and examining the available scientic literature. So, on the basis of the previously con- ducted tests, best signals simulating the normal respiratory volume variation as a function of time were found as com- posed by a sequence of two exponential functions. The rst wave form, simulating the inspiratory phase, is characterized by a positive slope and a time constant i and it is represen- tative of muscle integrity and the ability to drive lung expan- sion; the latter has a negative slope and a time constant e , representing a passive expiration, i.e., the free evolution of a rst-order system, such as lungs and chest wall at low breathing rates, and depends only on the mechanical charac- teristics of the system. The implemented software allows one to modify both values in order to simulate diseases affect- ing respiratory muscles or pulmonary parenchyma. Several FIG. 4. Representation of the PMU. AT, articial trachea; CR, clamp resis- tance; FT, ow transducer; LD, lower disk; OS, orice span gray area; PT, pressure transducer; PV, pulmonary ventilator; and UD, upper disk. 2414 Rev. Sci. Instrum., Vol. 73, No. 6, June 2002 Cappa, Sciuto, and Silvestri tests have been carried out to determine the relationship be- tween the simulated pulmonary compliance calculated by MeadWhittenberger method 36 and the orice opening percentage. 37 OS was increased by 10% steps and measure- ments were carried out during ve day tests for each OS value at environmental laboratory conditions. Results re- ported in Fig. 5 show an almost linear relation between com- pliance and OS with R 2 0.9334 and a standard deviation always lower than 8% that is generally assumed as acceptable. 38 Despite each value associated with a specic OS percentage result within the limits stated by ISO, the measured values were considered too high to fulll the constraint imposed by adequate compliance accuracy for a simulated range from 2 to 6ml/kPa 0.2 to 0.6 ml/cm H 2 O. Thus, it was decided to choose 0%, 30%, 60% and 100% OS values as a coarse compliance regulation, adopting as a ne regulation the variation of e relative to the expiration driv- ing signal in order to achieve compliance values even less than 2.0 ml/kPa 0.20 ml/cm H 2 O with an accuracy in the order of 0.5 ml/kPa 0.05 ml/cm H 2 O. As a further remark, it is necessary to observe that in case of OS equal to 100%, the simulated compliance higher value corresponds to the one characteristic of the bellow. Finally, whereas the reduc- tion of OS implies a reduction of tidal volume due to higher friction forces developed also in the inspiratory phase, the reduction of e relative to the expiratory wave form leaves the tidal volume reached in the inspiratory phase unchanged and causes a quicker bellow collapse which simulates a stiffer lung. The realized device is also applicable to test further ven- tilator features: in fact, a widely utilized diagnostic tool to identify human respiratory diseases is the shape recognition of the pressurevolume loop PV loop and owvolume loop FV loop generated by the patient during forced res- piration. Moreover, automatic shape recognition systems have been recently developed 39 also by means of neural net- work algorithms in order to support medical diagnosis. For these reasons, the proposed simulator was designed to be effectively used to plot PV and FV loops according to the characteristics of a chosen pathology by means of an ad- equate regulation of OS percentage and variations. Figure 6 shows PV loops produced by varying the aforementioned two parameters. Curves a, b, c, and d1 were obtained with 100%, 60%, 30%, and 0% OS respectively, corresponding to pulmonary compliances equal to 5.6, 4.2, 3.6, and 2.3 ml/kPa 0.55, 0.41, 0.35, and 0.23 ml/cm H 2 O. Curves d1 and d2 instead were obtained by modifying expiratory signal via software setting equal to 140 ms and 50 ms, respectively: in particular, curve d2 corresponds to 1.5 ml/kPa 0.15 ml/cm H 2 O compliance. In the clinical practice, curve a is recognized as characteristic of respiratory functionality of a healthy preterm infant at 8 months gestational age, while curve d2 is called bird-shaped PV loop and indicates the presence of the pathological condition known as NRDS, i.e., low volumes ventilated with strong muscular efforts due to lung stiffening. From an analysis of curves d1 and d2, a small compliance variation C0.8 ml/kPa 0.08 ml/cm H 2 O, induced by a variation equal to 90 ms, produces two easily distinguishable PV loops that, being digitized and stored, can also be analyzed by an automatic pattern recog- nition system. As a result, the lowest compliance value the TABLE I. Comparison of the performances of realized device Tester A to those of a commercially available tester for pulmonary ventilators Tester B. Tester A Tester B Intended use neonatal, for premature infants for adults, pediatric, and neonatal Respiratory frequency 40200 bpm 10150 bpm Compliance C 1.55.9 ml/kPa 0.150.60 ml/cm H 2 O 101529 ml/kPa 1150 ml/cm H 2 O Compliance accuracy 0.5 ml/kPa 0.05 ml/cm H 2 O unknown Time constant 50300 ms unknown Resistance R 629 kPa s l 1 60300 cm H 2 O s l 1 ) 0.5, 2.0, 4.9, 19.6, and 49.1 kPa s l 1 5, 20, 50, 200, and 500 cm H 2 O s l 1 ) Resistance accuracy dependent on C and accuracies unknown Tidal volume 27 ml 52000 ml Tidal volume standard deviation 1% unknown FIG. 5. Simulated compliance vs OS percentage tests results. is standard deviation associated to OS% acceptable values. FIG. 6. PV loops obtained by varying OS percentage span and expiratory . Curves a, b, c, and d1 correspond to 100%, 60%, 30%, and 0% OS, curves d1 and d2 correspond to 140 ms and 50 ms and 0% OS. 2415 Rev. Sci. Instrum., Vol. 73, No. 6, June 2002 Active neonatal breathing simulator proposed system is able to simulate is 1.5 ml/kPa 0.15 ml/cm H 2 O. In consideration of the herein determined simulator per- formances, once the pathology to simulate is chosen and both the breathing mechanical parameters and the values for the described regulations are specied individually, the pul- monary simulator generates a respiratory pattern by repeat- ing the wave form synthesized according to time parameters set by the operator and continuously measures the delivered volume. Before proceeding with the test, a preliminary cali- bration procedure is then actuated for up to 5 min while software routines continuously compare the measured vol- ume signal to the one set, and automatically adjust software coefcients until the difference between measured and set values is less than 1%, i.e., an accuracy of 0.05 ml in corre- spondence to a volume peak value of about 5 ml. As previously mentioned, the simulated airway resis- tance is unequivocally determined once and C are set and it can range from 6 to 29 kPa s l 1 60 to 300 cm H 2 O s l 1 , in correspondence to and C variations in the range from 50 to 300 ms and from 1.5 to 5.9 ml/kPa 0.15 to 0.60 ml/cm H 2 O, respectively. In Table I are reported the performances of the realized prototype Tester A compared to those of a widely 11,12 uti- lized reference tester for pulmonary ventilators Tester B. From an analysis of the data reported in Table I, it is possible to observe that a reduction of the limits of utilization of the proposed device determines an accuracy increase of the simulated mechanical parameters. In particular, low compli- ance values can be obtained only by means of novel design of the pulmonary ventilator tester, specically intended for neonatal utilization. Moreover, whereas resistance variations are obtained in tester B by manual substitution of a tube representing the AT, the proposed tester A is able to change the working characteristic by means of the variation of me- chanical parameter values, performed almost exclusively via software by the choice of and C. The use of a stepper motor coupled with the axial rod for the continuous regulation of the OS percentage will then be examined in the ongoing research phase. Finally, the possibility of simulating apnea periods, sighs, coughs, etc., can easily be implemented in the proposed simulator by means of the control software, in or- der to reproduce, in a more realistic way, the mechanical behavior of infant respiratory apparatus in physiopathologi- cal conditions. ACKNOWLEDGMENTS The authors wish to thank Professor Francesco Paolo Branca for providing suggestions and Rocco Conti for assist- ing with experimental tests. They express their appreciation to the Clinical Engineering Service of the Childrens Hospi- tal Bambino Gesu ` of Rome for the experimental facilities provided. 1 Despite the fact that the International Unit for pressure is pascal, due to the examined topic that mainly concerns medical applications, we have chosen to report pressure values also in cm H 2 O centimeters of water in order to allow the immediate comparison of the obtained results with those reported in scientic medical literature where this old unit is still widely used. 2 Le Systeme International dUnites (SI) The International System of Units SI, 6th ed. Bur. Intl. Poids et Mesures, Svres, France, 1991. 3 Natl. Inst. Stand. Technol. Spec. Publ. 330, 1991 Edition, The Interna- tional System of Units SI, edited by B. N. Taylor U.S. Government Printing Ofce, Washington, 1991. 4 ISO 1000:1992, SI units and recommendations for the use of their mul- tiples and of certain other units, International Organization for Standard- ization, Geneva, Switzerland, 1992. 5 M. Hauschild et al., Klin. Padiatr. 206, 167 1994 in German. 6 A. R. Stark et al., J. Pediatr. St. Louis 94, 439 1979. 7 A. Greenough et al., J. Pediatr. St. Louis 103, 769 1983. 8 S. K. Pingleton, Am. Rev. Respir. Dis. 137, 1463 1988. 9 D. Dreyfuss et al., Am. Rev. Respir. Dis. 137, 1159 1988. 10 D. Dreyfuss and G. Saumon, Am. J. Resp. Crit. Care Med. 157, 294 1998. 11 L. F. Samodelov and K. J. Falke, Intensive Care Med. 14, 632 1988. 12 Y. A. Konyukov et al., Intensive Care Med. 22, 363 1996. 13 T. Takahashi et al., Chest 100, 1030 1991. 14 M. Nishimura, D. Hess, and M. Kacmarek, Am. J. Resp. Crit. Care Med. 152, 1901 1995. 15 G. S. H. Sassoon, and S. E. Cruer, Intensive Care Med. 21, 159 1995. 16 G. Belforte et al., Measurement 27, 241 2000. 17 Commercially available infant lung simulators allow the user to set com- pliances in the range 10204 ml/kPa 120 ml/cm H 2 O and this setting is usually step variable. Step resolution varies depending on the manufac- turer; our survey outlined an average resolution of about 51 ml/kPa 5 ml/cm H 2 O. In most cases, manufacturers do not even guarantee the resolution at very low compliances and do not specify, for instance, the associated accuracy; moreover the devices intended for infant use are not suitable for preterm infants. 18 ISO 10651-1. 19 ASTM F1100, Ventilators for use during critical care. 20 ASTM F920, Resuscitators for use with humans. 21 T. Hansen and A. Corbet, in Disease of the Newborn, 7th ed., edited by H. W. Taeusch, R. A. Ballard, and M. E. Avery Saunders, Philadelphia, 1998, pp. 634643. 22 A. Greenough, C. J. Morley, and N. R. C. Robertson, in Textbook of Neonatology, edited by N. R. C. Robertson Churchill Livingstone, Edin- burgh, 1991, pp. 385486. 23 The lower value, i.e., 1 ml/kPa 0.1 ml/cm H 2 O, is reached in case of respiratory distress syndrome RDS, frequent in premature infants. 24 The higher value of 15 kPa s l 1 150 cm H 2 O s l 1 ) is typical of BPD. 25 0200 Hz bandwidth and 80 dB/dec. 26 J. R. Ligas and F. P. Primiano, in Encyclopedia of Medical Devices and Instrumentation, edited by J. G. Webster Wiley, New York, 1988, pp. 25502573. 27 Compliance and volume are respectively equal to 6.00.5 ml/kPa 0.60 0.05 ml/cm H 2 O and about 40 ml. This volume value, that is higher than one specic for the lungs of a neonatal patient, is not relevant to the simulator performance: in fact, its functioning only depends on the current compliance, resistance, tidal volume, that is on the order of 4 ml, ow, and pressure values. 28 E. Calzia et al., Intensive Care Med. 24, 931 1998. 29 Measuring range 2.5 kPa 25 cm H 2 O, overall accuracy 50 Pa 0.5 cm H 2 O. 30 Measuring range 4200 ml/s, overall accuracy 1.5 ml/s. 31 12 m diameter of platinum wires, 2.4 mm wire length, velocity range 0.210.0 m/s, and overall accuracy 0.1 m/s. 32 1030 ml/s corresponding to a delivered air volume per minute of 0.20.8 l/min in the range of breathing frequency from 20 to 160 bpm. 33 P10 Pa s ml 1 0.1 cm H 2 O s ml 1 . 34 6.0 cm internal diameter, open in the upper side. 35 0.120.01 kg m 1 s 1 in temperature range 10 C80 C. 36 J. Mead and J. L. Whittenberger, J. Appl. Physiol. 31, 934 1971. 37 0%, orices completely closed; 100%, orices completely open. 38 ISO standards require an error less than 10%15% on simulated param- eters. 39 S. Leonhardt et al., Control. Eng. 5, 1355 1997. 2416 Rev. Sci. Instrum., Vol. 73, No. 6, June 2002 Cappa, Sciuto, and Silvestri