Sunteți pe pagina 1din 6

Body Infra Red Thermometer

User Manual
Contents

1 General Description2
1 Safety Information2
1 Features2
1 Intended Use3
1 Precaution Before Use3
1 Configuration
1 !ec"nical Specifications#
1 Use$
1 %easuring &peration'
1 (d)ice'
1 !rou*les"ooting+
2 General Description
Non-contact Body Infrared Thermometer is Specially designed to take the body
temperature of a person regardless of room temperature.
Depending on various skin types and thickness, there may be temperature difference.
hen the room temperature is significant difference, the body infrared thermometer
should be left in that room during !" to #$ minutes before using.
2 Safety Information
This device must only be used for the purposes described in this instruction manual.
This device must only be used in an ambient temperature range bet%een !$&and '$&.
Do not e(pose this thermometer to electric shocks.
Do not use the device in relative humidity higher than )"*+,
-lean the glass %ith a cotton bud lightly moistened %ith .$ alcohol.
Should a problem occur %ith your device, please contact your retailer.
Do not attempt to repair the device yourself.
3 Features
/recise non-contact measurements.
0ser selectable &or1.
Set 2larm value 3)&.
2utomatic Data ,old 4 2uto po%er off.
2utomatic selection range and Display +esolution $.!&$.!1.
Backlight 5-D display
Intended Use
The Non-contact Body Infrared Thermometer is designed for body surface and forehead
temperature measurement for infants and adult %ithout contact to human body. It can be
used by consumers in household environment in accordance %ith the 2ST6 !78'-!7))
Standard specification for Infrared Thermometers for Intermittent Determination of /atient
Temperature.
# Precaution Before Use
In order to obtain reliable and stable results, %hen changing the environment %ith a
significant room temperature difference, it is recommended to leave the Non-contact Body
Infrared Thermometer to this ambient temperature for !" to #$ minutes before using it.
Depending on various skin types and thickness, there may be temperature indifference.
Do not e(pose the thermometer to sunlight or to %ater.
2llo% a none minute interval bet%een t%o measurement.
Normal Temperatures According To Measurement Method
6easurement 6ethod Normal Temp & Normal Temp 1
+ectal 38.8 to 3) 7..) to !$$.'
9ral 3"." to 3.." 7".7 to 77."
2u(iliary 3'.. to 3..3 7'.' to 77.!
2uricular 3".) to 3) 78.' to !$$.'
Temporal:Thermo ;lash< 3".) to 3..) 78.' to !$$.$
The temperature of the human body varies throughout the day. It can also be influenced
by numerous e(ternal factors= age, se(, type and thickness of skin>
Normal Temperatures According To Age
2ge Temp & Temp 1
$-# years 38.' to 3) 7.." to !$$.'
3-!$years 38.! to 3..) 7..o to !$$.$
!!-8"years 3".7 to 3..8 78.8 to 77..
?8" years 3".) to 3.." 78.' to 77."
$ Configuration
1,I- Sensor
2,.aser sig"t
3,.CD Display
,Do/n Button
#,Up Button
$,%ode 0 1 2 Button
',Bac3 .ig"t1.aser Button
+,%easurement -igger
4,Battery Co)er
15,6andle
' !ec"nical Specifications
Normal -onditions of 0se
Display +esolution $.!&:$.!1<
9perating Temperature !$ to '$&"$ to !$'1
,umidity +ate @)"*
/o%er 7A:NBD2 !8$'2 or IB- 85+8!<
SiCe !$$ ( ') ( !8$ mm:5 ( ( ,<
eight !3$g
Measuring Range
6easuring +ange 3#.$ to '3.$&:7$ to !$7.'1<
2ccuracy D$.3&:$."'1 <
6easuring Distance "cm-!"cm # in -".7 in
2utomatic Stop . sec.
Non-contact Body Infrared Thermometer Precision
3# to 37."&E73.# to 78.81 D$.3&E$."1 2ccording to 2ST6
38 to 37&E78.) to!$#.#1 D$.#&E$.'1 Standard B!78"-
!77)
37to'#."E!$#.#to!$)."1 D$.3&E$."1 :#$$3<
+ Use
)-!.Install battery
)-#.;or the first use or %hen inserting ne% batteries %ait bet%een !$ minutes for the
%arm-up of the apparatus and %hen inserting the ne% batteries.
)-3 2im to%ards the forehead :see the diagram belo% for the Touch less Body
Thermometer positioning<, from a distance of "cm :# in< press the measuring keythe
temperature is displayed immediately. The temperature can also be taken behind the ear
lobe.
)-'.Before taking of the temperature make sure to remove hair and perspiration from the
forehead.
4 easuring &peration
7-!.Data 6emory
Data memory automatically after temperature measurements, %hich %ill display at the
right corner of 5-D. /ress F0/F or GD9NF button to display the last temperature
measurement.
7-#. Battery +eplacement
H 2s battery po%er is not sufficient, 5-D %ill display F-IF replacement %ith one ne%
battery type 7A is reJuired.
K9pen battery cover, then take out the battery from 7-Aoit battery and place the battery
cover back, instrument and replace %ith a ne%,
7-3. 5ongevity 0se
The Non-contact Body Infrared Thermometer %as conceived for an intense and
professional use, its longevity is guaranteed for '$,$$$ takings.
15, (d)ice
The protective glass over the lens is the most important and fragile part of the
thermometer, please take great care of it.
-lean the glass %ith cotton fabric, %et %ith %ater or .$& alcohol.
Do not use other batteries than mentioned batteries, do not recharge non rechargeable
batteries.
Do not e(pose the thermometer to sunlight or %ater.
11Trou!leshooting
If you happen to have one of the follo%ing problems %hile using youLre the Non-contact
Body Infrared Thermometer please refer to this breakdo%n service guide to help resolve
the problem, If the problem persists please contact our customer service.
The screen displays the message G,iF
hen using the Non-contact Body Infrared Thermometer the message ,I can sho% on the
screen.
The analysis is above the measurement range selected ,superior to '3&!$7.'1in
Body mode.
The screen displays the message G59F
hen using the Non-contact Body Infrared Thermometer the message 59 can sho% on the
screen.
The temperature analyCed is under the measuring range selected, less than
3#&7$1in Body mode.
This message displays in "arious cases-please find !elo# a list of the main cases$%&in 'urface mode
+easons for 59 message display 2dvice
Temperature reading hampered by hair,
perspiration>>
6ake sure that there is no obstruction prior to
taking a temperature
Temperature hampered by an air flu( 6ake sure there is no air flu( as this could
interfere %ith the infrared system.
Temperature readings too close together, the
Body Thermometer did not have the chance to
boot itself.
+espect the pause of !" seconds minimum
bet%een t%o readings 2 ! minute pause is
advised
The measuring distance is too far /lease respect the measuring distance
:bet%een " and !"cm-# in and ".7 in<.