instrument is necessary to measure the quantity in terms
of standard unit. It is the process of framing the scale of the instrument by applying some standardized signals. Calibration is a pre-measurement process, generally
carried out by manufacturers. It is carried out by making
adjustments such that the read out device produces zero output for zero measured input. Similarly, it should display an output equivalent to the known measured input near the full scale input value. The accuracy of the instrument depends upon the calibration. Constant use of instruments affects their accuracy. If the accuracy is to be maintained, the instruments must be checked and recalibrated if necessary. The schedule of such calibration depends upon the severity of use, environmental conditions, accuracy of measurement required etc. As far as possible calibration should be performed under environmental conditions which are very close to the conditions under which actual measurements are carried out. If the output of a measuring system is linear and repeatable, it can be easily calibrated.
CALIBRATION OF DIAL INDIACTOR
In most of the measurements, the dimensions to be measured are compared with a suitable reference standard. Such measurements involve the use of an instrument called comparator; which measures the deviations of a dimension from the standard reference value. Dial indicators measure the displacement of a plunger or stylus and
indicate the magnitude of the displacement on dial by
means of a rotating pointer. The comparators however, may sometime indicate faulty readings due to their inherent manufacturing errors or due to the wear of their parts during service. The comparator, therefore, needs to be calibrated periodically. Calibration of a dial indicator involves Range of the instrument to be calibrated, intervals of the readings to be taken, reference standard to be used etc. Clean the measuring faces of the dial calibration tester. The dial indicator is fixed on the dial tester with the plunger just touching the anvil of the micrometer. The pointer on the dial indicator should be adjusted such that the zero coincide the bezel. The circular disc of the micrometer is adjusted such that the zero coincides with the cross line, i.e. setting the main scale zero and the zero on the micrometer drum across the cross line. Rotate the thimble of the micrometer in clockwise direction, Note down the readings of the graduated
disc of the micrometer against the cross wire for each
interval of movement of the dial gage. Repeat the procedure for the range of the indicator, both in the forward and backward directions. Since the Dial Tester is used as reference standard. It should be justified. Choose a proper range of the dial tester and check against the slip gauges. Place the slip gauge between the micrometer head tip and the contact point of the electronic probe. Take the reading in the upward direction from 0.5 mm to 25 mm in a step size of 0.5 mm. Calculate the uncertainty as per guidelines by National Accreditation Board for testing and calibration Laboratory (NABL). Tabulate the results as divisions on the dial gage vs. the deviations from the actual readings note from the dial tester. Plot a graph of calibration with the dial indicator readings vs. the error for various turns Tabulate the results for justification of dial tester. Study the calibration graph and comment about sources of error in the dial indicator under test and its measurement capability. Comment about the capability of the Dial tester for its use as reference standard for calibration.
comparison.pdf for detailed calibrating procedure of gauge blocks. Refer file Micrometer Calibration.pdf for detailed example pertaining calibrating procedure of micrometer.