EXAMINATION
Gonioscopy
I.
II.
III.
IV.
V.
VI.
VII.
Purpose of gonioscopy
Principle of gonioscopy
Method of gonioscopy
Angle structures
Grading of angle width
Developmental abnormalities
Key points
Purpose of gonioscopy
Why do I need to perform gonioscopy?
Fundamental part of comprehensive exam
Most important factor in correct diagnosis
Omission of gonioscopy is a common cause
of misdiagnosis
Purpose of gonioscopy
What can I achieve with gonioscopy?
Visualisation of the anterior chamber angle
Angle topography: scenery, landscape,
geography
Purpose of gonioscopy
What should I look for in gonioscopy?
Recognise angle landmarks and consider:
Ultrasound
biomicroscopy
Anterior segment
optical coherence
tomography
Principle of gonioscopy
Total internal reflection
Fibre optic cable
Principle of gonioscopy
Overcoming total internal reflection
Leuenberger EU
Asian Eye Institute
Method of gonioscopy
Direct gonioscopy
Indirect gonioscopy
Without indentation
With indentation
Without indentation
Jain S
Direct gonioscopy
Koeppe lens
50D
Leuenberger EU
Asian Eye Institute
Leuenberger EU
Asian Eye Institute
Leuenberger EU
Asian Eye Institute
Direct gonioscopy
Advantages
Straight-on view
Angle of visualisation
variable
More panoramic
Angle recession:
comparison
Ritch R
Direct gonioscopy
Disadvantages
Inconvenient
Special equipment needed
Indirect gonioscopy
Goldmann three-mirror lens
Leuenberger EU
Asian Eye Institute
Indirect gonioscopy
Slit lamp technique
Leuenberger EU
Asian Eye Institute
Indirect gonioscopy
Slit lamp technique
Goldmann three-mirror lens
Leuenberger EU
Asian Eye Institute
Indirect gonioscopy
Posner four-mirror lens
Indirect gonioscopy
Sussman four-mirror lens
Leuenberger EU
Asian Eye Institute
Indirect gonioscopy
Slit lamp technique
Four-mirror lens
Leuenberger EU
Asian Eye Institute
Leuenberger EU
Asian Eye Institute
Indirect gonioscopy
Slit lamp technique
Insertion of lens
Click on
screen to
play video
Indirect gonioscopy
Advantages
Preferred by most
Quick, convenient
No special equipment needed
Slit lamp
Variable magnification and illumination
Indirect gonioscopy
Disadvantages
Mirror image can be
confusing
Inadvertent pressure on
the cornea:
exaggerates the degree of
angle narrowing in the
Goldmann lens
opens the angle in fourmirror lenses
Leuenberger EU
Asian Eye Institute
Pale angles
Alward WLM. Color Atlas of Gonioscopy. Foundation of the American Academy of Ophthalmology, 2000.
Paintings by Lee Allen University of Iowa. Reproduced with permission.
Jain S
Angle structures:
ciliary body band
Iris inserts into
concave face
Wide in:
angle recession
(scan circumference)
cyclodialysis (cleft)
Ritch R
Angle recession
Jain S
Leuenberger EU
Asian Eye Institute
Angle structures:
trabecular meshwork
More pigmented with age
Flow is through posterior
TM
Pigment is intracellular
More pigment inferiorly
ID by Schwalbes line
ID by blood in Schlemms
canal
Anterior
trabecular
meshwork
Posterior
trabecular
meshwork
Leuenberger EU
Asian Eye Institute
Angle structures:
Schwalbes line
Termination of
Descemets membrane
Pigmented
Sampaolesis line
Landmark for
identification of TM
in narrow angles
Leuenberger EU
Asian Eye Institute
Schwalbes line
Jain S
Trabecular meshwork
Schwalbes
line
Flat iris
inserted at ciliary body band
Scleral
spur
Neovascularisation
Fine
Arborising
Crosses scleral spur
Differentiating between
iris processes and synechiae
Iris processes
Fine
Extend into scleral spur
Follow concavity of recess
Underlying structures seen
Iris moves with indentation
Broken with angle recession
Ritch R
Synechiae
Broad
Extend beyond scleral spur
Bridge concavity of recess
Obscure structures
Resist movement
Intact
Leuenberger EU
Asian Eye Institute
Indentation gonioscopy
Useful when iris surface is convex
Done when recognition of angle structures
is difficult
Plateau iris
Indentation gonioscopy
Over-the-hill gonioscopy
EGS 2003
Leuenberger EU
Asian Eye Institute
EGS 2003
Leuenberger EU
Asian Eye Institute
Open angle
Chew PTK, Aquino MC
Appositional closure
Synechial closure
Indentation gonioscopy:
synechial closure
EGS 2003
EGS 2003
Synechial closure
Leuenberger EU
Asian Eye Institute
Indentation gonioscopy:
plateau iris configuration
Curvature of iris along
anteriorly placed
ciliary processes
Curvature of iris
along lens surface
Jain S
Indentation gonioscopy:
ciliary body (pseudoplateau iris)
Ritch R
Ritch R
Indentation gonioscopy:
lens-induced angle closure
Lim A
EGS 2003
No
Yes
Do indentation gonioscopy
Any synechiae?
Yes
No
IOP raised?
Yes
No
1
EGS 2003
2
EGS 2003
10
20
Narrow
30
40
Wide
Nolan W
3
EGS 2003
0: none
4
4+: maximal
Drawing reproduced with permission from the EGS.
Nolan W
Shaffer
Modified
Shaffer
Grade
0
Grade
I
Grade
II
Grade
III
Grade
IV
Closed
10
20
30
40
Schwalbes
line not
visible
Schwalbes
line visible
Anterior
TM visible
Scleral
spur
visible
Ciliary
band
visible
Developmental abnormalities
Congenital glaucoma
Leuenberger EU
Asian Eye Institute
Leuenberger EU
Asian Eye Institute
Developmental abnormalities
Posterior embryotoxon
Developmental abnormalities
Axenfelds anomaly
Leuenberger EU
Asian Eye Institute
Key points
Gonioscopy
Fundamental part of ophthalmic evaluation
Confirmation of normal angle structures
Determination of narrowness or closure
of anterior chamber angle
Grading of angle width
Pathological findings