AT235-AT235h - Technical Specifications

Page 1

Technical Specifications

AT235-AT235h
Impedance Audiometer

Item No. 8107027-1 07/2014

AT235-AT235h - Technical Specifications

Included and Optional Parts


Included Parts:
ATP-AT235u Universal probe system with shoulder strap and wrist strap
TSH39 Single contralateral headset
UPS400 external switch mode power supply
Power cable
BET55 Box of 65 assorted ear tips
TPR26 3 rolls of recording paper
PCR-AT235h dust cover
Operation Manual.
Multilingual CE instructions for use

Additional Parts:
TDH39 audiometric headsetACC226 Carrying Case
EARtone 5A Audiometric Insert Phones
ACC25 carrying case
CIR33 contralateral insert phone
APS2 patient signal
CAT40 calibration unit 0.2-0.5-2.0-5.0 ml
CAT227 calibration unit 0.2-0.5-1-2-3-4-5 ml
IES impedance ear simulator
USB cable 2m black

Page 2

AT235-AT235h - Technical Specifications

Page 3

General Technical Specifications


Standards:
Audiometer
Impedance
Safety

:
:
:

IEC60645-1, ANSI S3.6, Type 4


IEC 61027, ANSI S3.39, Type 2
IEC60601-1, Class 1, Type B, EMC: IEC60601-1-2

Impedance:
Probe Tone Frequency:
High Freq. Probe Tones:
Probe Tone Intensity :
Pressure Range
:
Accuracy
:
Compliance Range
:
High Frequency
:
Compliance Accuracy :

226 Hz 1%
678, 800 and 1000Hz.
85 dB SPL 1.5 dB
+300 to -600 daPa
5% or 10 daPa
0.1-6 ml.
0.3 12 mmho
5% or 0.1 ml.

Reflex:
Manual or automatic testing with 18 stimuli to each ear.
Auto Reflex Detection.
Multiple Reflex stimuli.
Stimulus Duration
:
1.0 sec.
Reflex Acceptance
:
Adjustable between 2% and 6%, or 0.05 0.15 ml change of
ear canal volume.
Ipsi Frequencies
:
500, 1000, 2000, 3000, 4000 Hz, wide band, high and low pass.
Intensity (Max.)
:
100, 105, 110 dBHL.
Contra Frequencies :
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz, wide band,
high and low pass.
Intensity (Max.)
:
90, 100, 120, dBHL.
Reflex Decay:
Stimulus Duration
Stimulus Frequency
Stimulus Intensity

:
:
:

10 sec.

Ipsi and Contra as above.

Audiometer:
Type 4 according to IEC60645-1, ANSI S3.6.
Frequencies Hz:
125
250
500
1000
2000
3000
4000
6000
8000

Intensities dBHL:
-10 to 90
-10 to 90
-10 to 120
-10 to 120
-10 to 120
-10 to 120
-10 to 120
-10 to 120
-10 to 110

Auto Threshold Determination (Modified Hughson Westlake).

AT235-AT235h - Technical Specifications

Tests:
Tympanometry
:
Reflexometry incl. decay:
Audiometry
:
Eustachian Tube

Page 4

Automatic or Manual
Automatic or Manual.
Manual or Auto Threshold (Patient controlled Hughson
Westlake test after ISO 8253)
Semi Automatic function test.
Williams Test (non perforated eardrums)

Calibration:
Impedance
:
IEC61027, ANSI S3.39
Audiometer
:
ISO 389-1, ANSI S3.6
Calibration is performed via the instrument's front panel and is stored in a permanent memory.
Computer Communication:
Built-in USB input/output computer interface. One mode allows the computer to both monitor and
control the AT235. Control actions will be indicated on the operation panel and the display.
Power supply:
100-240V~50/60Hz
Fuses: 2A
Consumption: 15VA, max. 45VA.
The AT235 is delivered with a power cord for the relevant national mains socket. But AT235 can be
powered by any national AC voltage - Before use, the correct attachment plug/mains cable must be
identified and installed.
Dimensions:
W x D x H: 32 x 32 x 10.2 cm, 16.5 x 12.5 x 4 inches. Weight: 4.1 kg / 9.0 lbs.