What is chromatography
Separation Science
Organizing by size
Organizing by boiling point
(vapor pressure)
Organizing by solubility
After
Before
GC Instrument
SAMPLE
PREPARATION
AND INJECTION
SYSTEMS
AUTOSAMPLERS
CHROMATOGRAPHIC
SEPARATION
SYSTEMS
GC BASIC UNITS
SAMPLE
PREPARATION
SYSTEMS
GC
Detector
DETECTION &
IDENTIFICATION
SYSTEMS
GC DETECTORS
INTEGRATORS
MS DETECTORS
Gaschromatog.
DATA SYSTEMS
GC Basic
Unit
Autosampler
Data
system
DATA
SYSTEMS
MS Detector
Autosampler
Data
system
GC Basic
Unit
GC/SA
Benchtop GC/MS
GC Instrument : injector
Injection techniques
Vaporizing
Injectors:
SSL
PPKD
PKD
PTV
No-vaporizing
Injectors:
On-column
Injectors
split/
splitless
PTV
Packed and
wide-bore
columns
PTV
PTV
PRIMARY COOLING
SYSTEM
SECONDARY COOLING
SYSTEM
Primary
cooling
system
SBP
Room
Temp.
temperature
Carrier
gas
Cooling fluid
(secondary
cooling)
Liquid sample
droplets
Liquid sample
plug
Liquid sample
film
Oven
temperature
Split Injector
Additional
Insulating
Septum
Purge
Carrier
Split Line
50 mm
Syringe
Glass Liner
5 mm ID
Splitting Point
Injector
Chamber
Liner
Length of Column
Inside Injector = 40 mm
Terminal
Fitting
Carrier
Pathway
Column
Capillary
Column
7001-1417
970810
Splitless Injector
Additional
Insulating
Septum
Purge
Carrier
Split Line
80 mm
Syringe
Splitless
Liner
Length of Column
Inside Injector = 64 mm
Capillary
Column
77001-1416
970805
PTV Injector
Septum purge line
Septum
Splitting
line
Carrier gas
inlet
Liner
Heating
element
FORCED AIR
COOLING
OVEN
Column
Carrier
Septum
purge
line
Splitting
line
GC Instrument
GC Instrument
Detectors Overview
Detectors Principles
Detectors Characteristics
RESPONSE MECHANISM
Concentration dependent: TCD, ECD, PID
Mass rate dependent: FID, NPD, FPD
NON DESTRUCTIVE
DESTRUCTIVE
RESPONSE TYPE
Universal: TCD, FID, (PID)
Selective: ECD, NPD, FPD, (PID)
Air
Column
effluent
+Hydrogen
Radioactive
source
Air
Interference
filter
Air
Column effluent
+ Hydrogen
Photomultipier tube
Heat sink
UV
lamp
Quartz cell
GC column
Exit tube
Coupling Chromatography to
Mass Spectrometry
Separation plus identification
Ion Source
Analytical
Quadrupole
Detector
Quadrupole Wiring
Series1
0.5
1.5
Stability Regions
YZ-stable
XZ-stable
Mathieu Parameters
RF =
m r2
e 2qz
DC =
m r2
e 4az
m = Mass in kg
r = radius = distance between
rod inner surface and center
of quadrupole
w = 2 * frequency (1100 kHz)
PCP