Documente Academic
Documente Profesional
Documente Cultură
Service Manual
Version 1.44
IMPORTANT
Only the trained and authorized
serviceman is permitted to install
the system, following this Service
Manual.
Preface
Preface
This Automated Clinical Analyzer is an IVD medical equipment for indoor
use and is based on the colorimetric and turbidimetric measurement.
Its main application is expected to be colorimetric measurement, clinical
chemistry testing, immuno-serological testing, hematological testing and
urine testing in hospitals, clinical laboratories and other research
laboratories.
In this manual, the items you have to be careful for safety are
indicated by the following levels.
Note
0-1
2. Warranty
3. Safety Caution
0-2
Warning
Preface
Caution
Caution
Caution
Caution
Caution
0-3
Caution
Caution
Caution
Caution
0-4
Caution
Preface
Caution
(13) Piping
Caution
4. Caution in Operation
(1) Usage
Notice
Notice
1.
2.
Notice
0-5
Notice
1.
2.
3.
Notice
1.
2.
3.
Caution
0-6
Preface
Caution
Caution
(2) Servicing
Come in contact with us about servicing.
0-7
6. Installation Requirements
Only the trained and authorized serviceman is permitted to install or move
the system, following the requirements below.
50 cm
32 cm
50 cm
100 cm
(At the minimum)
(7) The following points should be taken into consideration to select the
installation position.
- No dust, good ventilation
- No direct sun light
- Small vibration
0-8
Preface
Indoor.
No direct sunlight.
No dust.
No upper load.
No dropping.
No throwing down.
Ambient temperature should be 1 ~ 45 Celsius degrees.
Ambient humidity should be 10 ~ 85% (no condensation).
0-9
Chapter 1
System configuration
1.1 Over view ------------------------------------------------------------------------------------- 1-2
1.2 Name of Each Unit ------------------------------------------------------------------------ 1-3
1.3 Panels ------------------------------------------------------------------------------------------ 1-6
1.4 Unit Layout ---------------------------------------------------------------------------------- 1-7
1.4.1 Unit Layout in the front view ------------------------------------------------ 1-7
1.4.2 Unit Layout in the right side view ----------------------------------------- 1-8
1.4.3 Unit Layout in the left side view ------------------------------------------- 1-9
1.4.4 Unit Layout in the rear view ------------------------------------------------- 1-10
1.4.5 Unit Layout of the first floor--------------------------------------------------- 1-11
1.4.6 Unit Layout of the second floor ---------------------------------------------- 1-12
1.4.7 Unit Location of the upper deck --------------------------------------------- 1-13
Chapter 2
Functions and Movement of the System
2.1 Functions and Movement ---------------------------------------------------------------- 2-2
2.2 Analysis Process and Optical Measurement Point ------------------------------ 2-4
2.3 Automatic Analysis Process ------------------------------------------------------------ 2-5
2.4 Reagent Dispensation Process----------------------------------------------------------- 2-8
2.5 Sample Dispensation Process ---------------------------------------------------------- 2-11
Chapter 3
Explanations about Units
3.1 Sampler (STF) -------------------------------------------------------------------------------- 3-2
3.2 Sample Pipette (STM)
3.3 Sample Pump (SP)
3.4 Reagent Tray (RGT)
------------------------------------------------------------------- 3-6
----------------------------------------------------------------------- 3-11
--------------------------------------------------------------------- 3-14
---------------------------------------------------------------------- 3-26
-------------------------------------------------------------------- 3-30
3.10
3.11
3.12
3.13
3.14
3.15
Chapter 4
Control PCB
4.1
4.2
4.3
4.4
UCNT-2 PCB-------------------------------------------------------------------------------4-12
4.5
4.6
4.7
4.8
4.9
4.10
Chapter 5
Hidden Screens of Maintenance
5.1
5.2
5.1.2
Password Setting
5.2.1
5.2.2
5.3
5.4
5.7
5.6.1
5.6.2
5.6.3
5.6.4
5.6.5
5.6.6
5.6.7
5.6.8
5.6.9
5.6.10
5.6.11
5.6.12
5.6.13
5.6.14
5.6.15
5.6.16
5.6.17
5.6.18
COVER (Sample & Top Covers Lock & Unlock) -------------------- 5-30
5.7.2
5.7.3
5.7.4
5.7.5
Appendix 1
Mechanical Assembly Drawing
1 System Piping Diagram -------------------------------------------------------------- ASSY-1
2 ISE Piping Diagram ------------------------------------------------------------------- ASSY-2
3 Sampler (STF) --------------------------------------------------------------------------- ASSY-3
4 Sample Pipette (STM) ------------------------------------------------------------------ ASSY-4
5 Sample Pump (SP) --------------------------------------------------------------------- ASSY-5
Appendix 2
Electrical Wiring Drawing
1 Electrical Block Diagram -1/2 -------------------------------------------------------- WIRING-1
Electrical Block Diagram -2/2 ------------------------------------------------------- WIRING -2
2 AC/DC Power Supply 1/3 ------------------------------------------------------------- WIRING -3
AC/DC Power Supply 2/3 ------------------------------------------------------------- WIRING -4
AC/DC Power Supply 3/3 ------------------------------------------------------------- WIRING -5
3 Sampler (STF) --------------------------------------------------------------------------- WIRING -6
4 STM Unit ---------------------------------------------------------------------------------- WIRING -7
5 Sample Pump (SP) --------------------------------------------------------------------- WIRING -8
6 Reagent Tray (RGT) ------------------------------------------------------------------- WIRING -9
7 Reagent Pipette (RTM) ---------------------------------------------------------------- WIRING -10
8 Reagent Pump (RP) -------------------------------------------------------------------- WIRING -11
9 Reaction Tray (RCT) ------------------------------------------------------------------- WIRING -12
10 Mixing Mecha.-1 (MU-1) -------------------------------------------------------------- WIRING -13
11 Mixing Mecha.-2 (MU-2) -------------------------------------------------------------- WIRING -13
12 Mixing Pump (MP) --------------------------------------------------------------------- WIRING -13
13 Spectrophotometer (SPM) ------------------------------------------------------------ WIRING -14
14 Cuvette Wash Station (CWS) ------------------------------------------------------- WIRING -15
Chapter 1
Service Manual
Chapter 1
System Configuration
Contents
1. 1.
1. 2.
1. 3.
1.4.
1.4.1.
1.4.2.
1.4.3.
1.4.4.
1.4.5.
1.4.6.
1.4.7.
1-1
SYSTEM CONFIGURATION
1.1.
Over View
Water reservoir
Reaction waste reservoir
Reaction sol. separating
collection reservoir (option)
Printer
Alkaline washing
solution reservoir
Top view
Top cover
Personal computer
Analyzer
Reaction waste liquid
reservoir (option)
Water reservoir
1-2
Chapter 1
Service Manual
1.2.
Top Cover
Inner Cover
Sampler Cover
Top Cover
Sampler Cover
Printer SW
Front Panel
1-3
Top Cover
ISE Door
1-4
Chapter 1
Service Manual
Ventilation Fan
Gas Damper
Top Cover
Control PCB
Reaction Cuvette
Washing Station
Reagent Tray
Sample Probe
Sampler
Reagent Probe
Reaction Tray
1-5
1.3.
Panels
Front Panel
Fixing Screw
Right Side
Fixing Screw
Front Panel
ISE Door
Top Panel
Fig. 1-8 System slant view
1-6
Chapter 1
Service Manual
1.4.
Unit Layout
1.4.1.
The drawing below shows front view when the front panel is removed.
4. ISE pump
5. Thermo module
1-7
1.4.2.
This drawing shows the right side view when the right side panel is removed.
3
1. Five-phase-motor driver
3. Five-phase-motor driver
5. Preheat (WSH)
6. ISE unit
7. Water supply and drainage panel Panel for washing solution supply and drainage
8. Slide stay (right)
1-8
Chapter 1
Service Manual
1.4.3.
This drawing shows the left side view when the left side panel is removed.
1
11
8
3
9
10
2
4
12
2. Noise filter
3. RS-232C connector
5. AC power input
6. Fuse
7. SSR AC PCB
8. SSR
1-9
1-10
10. Fuse
12. Fuse
Chapter 1
Service Manual
1.4.4.
This drawing shows the rear view when the rear panel is removed.
1
10
11
3. Ventilation fan
LDA150W-24-SNY type
LDA75F-15-SNY type
LDA15F-5-SNY type
7. Mother board
8. DP-1
9. DP-2
1-11
1.4.5.
1-12
1. Power transformer
2. DP-2
3. DP-1
4. Degasser
Chapter 1
Service Manual
1.4.6.
The drawing below shows the view when the upper chassis
is slanted and the upper panel and the plate are removed.
1
11
6
12
10
13
8. Reagent probe
9. Reaction tray
1-13
1.4.7.
The drawing below shows the control PCB attached to the front of system.
1-14
1. Lamp DC PCB
2. TCD PCB
3. MCPU PCB
4. UCN-1 PCB
5. UCN-2 PCB
6. SCPU PCB
7. PMDRV PCB
Service Manual
Chapter 2
Chapter 2
Contents
2.1 Functions and Movement --------------------------------------------- 2-2
2.2 Analysis Process and Optical Measurement Point------------- 2-4
2.3 Automatic Analysis Process------------------------------------------ 2-5
2.4 Reagent Dispensation Process--------------------------------------- 2-8
2.5 Sample Dispensation Process---------------------------------------- 2-11
2-1
MIXING-1
REACTION TRAY
CUVETTE WASHING STASION
SPECTROPHOTOMETER
REAGENT B.C
READER
SAMPLE PIPETTE
ISE MODULE
MIXING-2
WASHING
POT
REAGENT TRAY
REAGENT PIPETTE
SAMPLE PUMP
MIXING PUMP
A/D CONVERTER
LOG
CONVERTER
SAMPLE TRAY
CUVETTE
WASHING PUMP
REAGENT PUMP
DRAIN PUMP
PROBE
WASHING PUMP
PRINTER
SYSTEM
INTERFACE
TUBE LINE
CABLE
LINE
2-2
PC
Chapter 2
Service Manual
START
CUVETTE WASHING
WATER BLANK
2.7 min
EVACUATION
R1 DISPENSE
1.5 min
SAMPLE DISPENSE
MIXING 2 STAGES
5 min
1 ST Reaction
R2 DISPENSE
12 min
MIXING 2 STAGES
4 min
2 ND Reaction
5.5 min
MIXING 1 STAGE
END OF MEASUREMENT
CUVETTE WASHING
2.7 min
AUTO-STOP
2-3
2.2.
5min
5min
Water blank
ABS
Sub-reading points
Main-reding ponits
measuring point
This figure shows 2 points end reaction curve.
Explanation
S Sample dispense
R1 R1 dispense
R2 R2 dispense
M mixing
Analysis process
Cycle after start
Function
6, 7, 8
1, 2, 3
11
R1 dispense
17
Sample dispense
12
17, 18
12, 13
37
R2 dispense
32
38, 39
33, 34
53
48
60
55
2-4
Optical
measurement
Chapter 2
Service Manual
2.3.
reagent tray, reagent delivery, reaction tray, mixing units, cuvette washing unit,
spectrophotometer and so on.
The combined movements of mechanical units are explained here. Please refer to
Fig. 2-1 Function diagram and Fig. 2-2 Analysis flow diagram.
Sample cups/tubes with sample in it are set in the sample tray, and reagent
bottles are set in the reagent tray.
Test orders for these samples are entered in Order Entry screen.
When start button is clicked, cuvette washing unit starts washing from No.1
cuvette.
When No.1 cuvette advances to one step before the R1 dispensation position,
reagent tray rotates and transports the reagent bottle to the reagent aspiration
position.
Next, R1 probe moves to the aspiration position, above the reagent bottle, then
moves down to the reagent level.
*) This probe has level sensing function and it stops when the probe tip touches the
reagent. The designated amount of reagent is aspirated by reagent pump.
Next, R1 probe goes up and moves to the R1 dispense position and waits for the
cuvette coming to the dispense position.
When the cuvette comes to the position, R1 probe goes down to the designed level
and dispenses the reagent.
2-5
At the 5th cycle after the R1 dispense, sample tray rotates to transfer No.1 sample
cup/tube to the sampling position.
Then sample probe moves above No.1 cuvette and goes down to the R1 reagent level
in the cuvette, and dispenses sample.
At the next cycle, the cuvette is moved to the mixing station-1, where R1 and
sample are mixed by MU-1, mixing unit.
This mixing is repeated in the next cycle, two times in total.
Sample probe moves to probe washing pot and both inside and outside are washed
by de-ionized water, there.
When reaction
tray rotates and the cuvette passes the optical measurement position, light
absorption data of 12 wavelengths are measured.
After 5 minutes of R1 dispense, the cuvette moves to R2 dispense position, the same
position as R1 dispense, then R2 is dispensed by R2 probe, which has same function as
R1 probe.
At the next cycle, the cuvette is moved to the mixing station-1, where R2 and
reaction liquid are mixed by MU-1.
This mixing is repeated in the next cycle, two times in total.
R2 probe moves to probe washing pot and both inside and outside are washed by
de-ionized water.
After 4 minutes of R2 dispense, the cuvette comes to the mixing station-2, where the
particles in reaction liquid are dispersed again by MU-2.
After 10.7 minutes of sample dispense, reaction liquid in the cuvette is aspirated out
by the No.1 nozzle of the cuvette washing unit and transferred to the waste
reservoir, outside of the system.
2-6
Service Manual
Chapter 2
As the last step of the measurement, for the 1 st test of No. 1 sample, optical
absorption data are converted to concentration/activity data using calibration
curve.
But test results will not be printed out till whole test results of No.1
2-7
2.4.
2-8
Chapter 2
Service Manual
2-9
1. When R1 probe moves to above the reagent bottle, 20 micro-liter separation air is
aspirated by the action of reagent pump (RP-2).
2. R1 probe goes down till the tip detects the reagent level. Then RP-2 aspirates
10 micro-liter excess reagent volume than designated.
3. R1 probe goes up to the home position.
4. R1 probe rotates to the reaction cuvette position.
5. When R1 probe goes down into a cuvette and stops at the determined position,
designated volume of reagent is dispensed.
6. R1 probe once goes up to the home position and reaction tray rotates to the next
position. Then goes down again and stops. R2 probe, which is in the probe washing
pot, will be washed, following the procedure below.
- Solenoid valve EV-3 switches the line to R2 side and solenoid valve EV-2 will be
ON, which connects to probe washing pump.
- RP-2 dispenses excess 10 micro-liter reagent and separation air 20 micro-liter
into the washing pot.
- Probe washing pump (PWP) are actuated and RP-1 sends solution to wash probe
inside and PWP washes outside.
R1 probe washing is down within the same cycle as R2 washing.
R1 is washed right after the R2 probe reagent dispense.
The procedure of R1 washing is the same as R2 washing except R1 probe is in the
washing pot and solenoid valve EV-3 is switched to the R1 side.
7. After the probe washing, water droplet at the tip is wiped up by a wiping mechanism
at the top of the washing pot when the probe goes down.
2-10
Chapter 2
Service Manual
2.5.
2-11
2-12
Chapter 2
Service Manual
2-13
Chapter 3
Service Manual
Chapter 3
Contents
3. 1 Sampler (STF)-------------------------------------------------------------- 3-2
3. 2 Sample Pipette (STM) --------------------------------------------------- 3-6
3. 3 Sample Pump (SP) ------------------------------------------------------- 3-11
3. 4 Reagent Tray (RGT) -------------------------------------------------- 3-14
3. 5 Reagent Pipette (RTM) ---------------------------------------------- 3-22
3. 6 Reagent Pump (RP)--------------------------------------------------- 3-26
3. 7 Reaction Tray (RCT) ------------------------------------------------- 3-30
3. 8 Mixing Unit 1, 2 (MU-1, 2) ------------------------------------- 3-35
3. 9 Mixing Pump (MP) --------------------------------------------------- 3-39
3.10 Spectrophotometer (SPM)-------------------------------------------- 3-41
3.11 Reaction Cuvette Washing Mechanics (CWS) ------------------ 3-45
3.12 Reaction Cuvette Washing Pump (CWP) ------------------------ 3-49
3.13 Washing Solution Pre-heater (WSH) ----------------------------- 3-53
3.14 Probe Washing Pump (PWP) --------------------------------------- 3-57
3.15 Waste Liquid Pump 1, 2 (DP-1, 2) ---------------------------- 3-61
3-1
Cup Sensor
(inside)
Case
Cup Sensor
(outside)
Fig. 3-2
STF PCB
Connecter
Fig. 3-1
3.1.2.
Fig. 3-3
3-2
Fig. 3-4
Calibration Tray
(yellow)
Service Manual
Chapter 3
1. Sample trays
Patient tray (In case of No. 1 tray)
(Outside)
Patient sample (1 - 30) -------------30 samples
Patient sample (31 - 40) -----------10 samples
(Inside)
Control sample (C7 - C12) ---------- 6 samples
Washing
(W1, W2) -------------2 samples
STAT
(E1 - E5) ------------- 5 samples
ISE standard liq. (CAL) ------------ 1 sample
ISE washing liq. (WASH) ---------- 1 sample
3-3
3.1.3.
1. Function
The sample tray holds cups with patient sample, standard sample or control
sample in it. Then this tray is set to the sampler.
When the tray is set to the sampler, the tray number will be read by the tray
discriminating sensor (Fig. 3-2).
There are two aspiration positions of sample probe; samples set on the outside
of sample tray will be aspirated at A position, and samples set on the inside at
B position.
2. Operation (Refer to Fig. ASSY-3.)
Set sample tray, fitting the hole of sample tray into the guide pin of the
rotating disk (3)
When stepping motor (7) rotates, the rotation is transferred to the rotating disk
(3) and the sample tray through the plane gear (5).
The plane gear (5) has a shade pin, which shades the sensor (6) light at the
initialization. This detects home position.
When home position sensor detects home position, rotary encoder (united with
the motor) is reset and output will be zero.
During the measurement, sample tray rotates clockwise and carries sample
cups to the aspiration position A (in case of the outside), or to the aspiration
position B (in case of the inside) one by one.
Rotary encoder counts motor rotation angle, and sample tray rotation angle is
checked.
3-4
Chapter 3
Service Manual
3.1.4.
1. Unit dismantling
1. Turn off system power switch and PC
power switch.
2. Open the top cover.
3. Push over upper chassis gently.
(Fig. 3-5)
Remove fixing screws on the right
and left side panels in advance.
Confirm the click sound when a
stay is fixed.
Fig. 3-5 Front view
Connector
Case
J4102
J4103
J4104
3-5
Probe
Motor (up/down)
Piping Tube
Fig.3-7 Sample Probe outer view
3.2.
1. Function
Various samples, such as patient, standard and control samples, are aspirated
from sample cups at the amount of designated plus small excess then dispensed
designated amount into cuvettes.
Sample probe itself has level sensing function. When it detects sample level,
probe stops going down. Then the fixed amount (designated plus excess) will be
aspirated by sample pump.
After aspiration, probe goes up and rotates to above the cuvette position, goes
down into a cuvette, and again detects the first reagent level. Here the
designated amount of sample is dispensed into the reagent.
After the dispensation, probe goes up and rotates to above the washing pot and
then goes down into the pot. Then the both inside and outside of probe are
washed. The excess sample in the probe is washed out by inside washing water.
After that, water droplet at the probe tip is sucked up in the pot.
3-6
Chapter 3
Service Manual
variable
Cuvette
variable
fixed
down up
down up
aspiration
dispensation
down up
probe washing
droplet wipe out
variable
Cuvette
variable
down up
down up
aspiration
dispensation
fixed
down up
probe washing
droplet wipe out
3-7
During the measurement, rotary encoder counts motor rotation angle and sample
probe rotation position is checked.
- If rotation angle exceeds the limit or it consumed too much time, an alarm is
given and emergency stop is made.
Before rotation starts, system checks if up/down movement position is at the
home position.
- If it will not be at the home position within the fixed time, system gives alarm
and emergency stop.
3-8
Chapter 3
Service Manual
M6 Joint
3-9
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
1) Electric power should be off before the maintenance.
- If connecting or disconnecting while power is ON,
electronic parts may be damaged by electric shock.
2) Do not hold STM arm when lifting up the unit.
- The probe may be distorted if touched.
3) When removing M6 joint, wipe off the water flown from tube, before
moving the tube to the bottom of unit.
- If water is carried into STM section, electric parts such as sensors
may be damaged.
3-10
Chapter 3
Service Manual
SP-2
(for sample)
SP-1
(for washing)
Solenoid Valve
Motor
Motor
3.3.2.
1. Function
Various samples, such as patient, standard and control samples, are aspirated
from sample cups at the amount of designated plus small excess then dispensed
designated amount into cuvettes by the pump.
The amount of sample can be set as follows.
- 3 micro-liter 30 micro-liter (Variable by 0.5 micro-liter step)
Separation air volume, which prevents mixing of sample and water in the probe, and
excess sample amount to increase accuracy of dispensation volume are set as follows.
-
Separation air volume (To prevent mixing of washing water and sample.)
2 micro-liter (constant)
Excess amount (To increase dispensation accuracy.)
5 micro-liter (constant)
After air volume (To prevent sample splash while probe rotation.)
1 micro-liter (constant)
3-11
3-12
Chapter 3
Service Manual
3.3.3.
1. Unit dismantling
1)
2)
3)
4)
Joint of Piping
Fixing Screw
6)
Fig. 3-14 SP
2. Assembling
Assembling procedure is the reverse of the above.
3-13
3. Cautions
Electric power should be off before the maintenance.
- If connecting or disconnecting while power is ON,
electronic parts may be damaged by electric shock.
When removing M6 joints, wipe up the water flown from tube.
- If water is carried into electric parts such as sensors, they
may be damaged.
3.3.4.
Exchange of U-packing
This pump consists of SP-1 (for washing) and SP-2 (for samples).
The procedure of U-packing exchange is the same, so the following explanation is
done for SP-1.
1) Remove sample pump form the
system. (Refer to 3.3.3. Unit
dismantling)
2) Loosen pump head fixing screws (two)
and pull out them from unit to remove.
3) Loosen screws and remove the pump
from pump head.
4) Remove spacer form pump head with
tweezers.
5) Take out U-packing from pump head,
using tweezers
-
6)
Washer
Spacer
Fixing Screw
3-14
7)
8)
Chapter 3
Service Manual
3.4.
3.4.1.
Motor
Connector
Fan
3-15
3.4.2.
1.
Function
The set reagent bottles for the items (48) can be set on the tray.
Reagent bottles are set on the tray and the tray with the set reagent, dilluent
and washing solution bottles are transferred and rotated to the dispensation
position of reagent probe.
The tray can keep reagents cool as long as main power switch on the back of the
system is ON.
2.
3.
Temperature control
The sensor of temperature (13) attached to the lower part of bottle case (4) senses
temperature, and power supply to Peltier element (10) is controlled by digital
thermo.
Digital thermo is located on the left side panel of the system. (Refer to Fig. 1-11,
pages1-8.
Explanation of this digital thermo is omitted.
3-16
Chapter 3
Service Manual
3.4.3.
1. Unit dismantling
1.
2.
3.
Fig. 3-21
Bottle Case
Outer Case
4.
5.
Connector
J4707
J4701 J4702
6.
7.
2. Assembling
Assembling procedure is the reverse of the above.
3-17
3. Cautions
1) Electric power should be off before the maintenance.
- If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
2) Move reagent probe to cuvette.
- The probe may be distorted if touched.
3-18
Chapter 3
Service Manual
3.4.4.
2.
Fig. 3-23
Reagent Tray
3.
M3 screw
Revolving shaft
Fig. 3-24
Reagent Tray
4.
Fig. 3-25
Reagent Tray
3-19
5.
Insulator
Revolving shaft
Fig. 3-26
6.
Revolving Shaft
Fig. 3-27
7.
8.
Connector
Fig. 3-28
3-20
Chapter 3
Service Manual
9.
Fig. 3-29
Fig. 3-30
Connector
11. Set Peltier element as shown on the
left side, while holding down.
Be sure to put Peltier element
with the letters on it up.
12. Assemble with the reverse procedure
of the above.
Fig. 3-31
3-21
3.5.1.
Connector
Probe
Motor (up/down)
Fig. 3-32
Piping Tube
Fig. 3-32
Reagent Probe
Fig. 3-33
RLS Board
3.5.2.
1. Function
Reagent, dilluent and washing solution, etc. are aspirated from reagent bottle at
the amount of designated plus small excess, then dispensed designated amount
into reaction cuvettes.
Reagent probe itself has level sensing function. When the tip of probe detects
reagent level, probe stops going down. Then the fixed amount (designated plus
excess) will be aspirated by reagent pump.
After aspiration, probe goes up and rotates to above the cuvette position and goes
down into a cuvette, and stops at the set level. Here the designated amount of
reagent is dispensed by reagent pump.
After the dispensation, probe goes up and rotates to above the washing pot and
then goes down into the pot. Then the both inside and outside of probe are
washed and the excess reagent in the probe is washed out by inside washing
solution. After that, water droplet at the probe tip is sucked up in the pot.
3-22
Chapter 3
Service Manual
2.
3.
1) Normal operation
Reagent cup
variable
down up
aspiration
Cuvette
fixed
fixed
down up
dispensation
down up
probe washing
droplet wipe out
During the measurement, rotary encoder counts motor rotation angle and
reagent probe rotation position is checked.
- If rotation angle exceeds the limit or it consumed too much time, an alarm is
given and emergency stop is made.
Before rotation starts, system checks if up/down movement position is at the
home position.
- If it will not be at the home position within the fixed time, system gives an
alarm and emergency stop.
3-23
3.5.3.
1. Unit dismantling
1. Turn off system power switch and PC
power switch.
2. After opening top cover, push over
upper chassis gently to give a tilt
- Be sure to remove fixing screws on
left and right panels in advance.
- Confirm the click sound made by a
stay when fixed.
3. Remove front panel first, and then
upper panel.
Fig. 3-34 System front view
Fig. 3-35
ISE Pump
EV-3
5. Loosen M6 joints (two) connected to
both ends of solenoid valve (EV-3) and
disconnect tube.
- When disconnecting tube, water in the
tube flows out. Wipe it off with tissue
paper.
6.
Fig.3-36
3-24
Solenoid Valve
M6 joints
Chapter 3
Service Manual
Fixing Stay
RTM (enlarged)
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
1. Electric power should be off before the maintenance.
- If connecting or disconnecting while power is ON,
electronic parts may be damaged by electric shock.
2. Do not hold arm when lifting the unit up.
- The probe may be distorted if touched.
3. When removing M6 joint, wipe off the water flown from tube, before
moving the tube to the bottom of unit.
- If water is carried into electric parts such as sensors, they may
be damaged.
3-25
3.6.
3.6.1.
Solenoid valve
Motor
Motor
3.6.2.
1. Function
Reagent, dilluent and washing solution, etc. are aspirated from reagent bottle at
the amount of designated plus small excess, then dispensed designated amount
into reaction cuvettes.
The amount of reagent can be set as follows.
- 20 micro-liter 350 micro-liter (Variable by 5 micro-liter step)
Separation air volume which prevents mixing of reagent and water in the tube and
excess sample amount to increase accuracy of dispense volume are set as follows.
- Separation air volume (To prevent mixing of washing water and sample.)
20 micro-liter (constant)
- Excess amount (To increase dispense accuracy.)
10 micro-liter (constant)
At reagent aspiration, designated reagent volume (20 350 micro-liter) plus excess
volume (10 micro-liter) is aspirated after separation air 20 micro-liter is aspirated.
At reagent dispensation, designated reagent volume (20 350 micro-liter) is
dispensed into reaction cuvettes.
3-26
Service Manual
Chapter 3
3-27
3.6.3.
1. Unit dismantling
1. Turn off the system power source and
PC power switch.
2. Open top cover.
3. Pull and remove front panel.
- Refer to chapter 1, page 1-5.
Joint of Piping
Fixing Screw
4.
Joint of Piping
5.
6.
Fixing Screw
Fig. 3-40 RP
2. Assembling
Assembling procedure is the reverse of the above.
3-28
J4507
J4514
J4515
Chapter 3
Service Manual
3. Cautions
Electric power should be off before the maintenance.
-
3.6.4.
Exchange of U-packing
This pump consists of PR-1 (for washing) and PR-2 (for samples).
The explanation for exchange of U-packing is omitted, for the procedure is
the same as in sample pump. Refer to page 3-14.
3-29
3.7.
3.7.1.
Cuvette Holder
Motor
Fig. 3-41
-
3.7.2.
1. Function
This tray can hold sixty (60) reaction cuvettes and keep the temperature
in cuvettes at 37 Celsius degrees.
In these cuvettes, reagent and samples are made react.
Reaction cuvette is optical cell for absorbance measurement.
(Reaction cuvette direct photometry method)
3-30
Service Manual
Chapter 3
3. Temperature control
Temperature of heat block (10) is detected by two temperature sensors (9) fixed
to the side of heat block (10), and heater (11) is accurately controlled by PID
control of TCD PCB. (Heat block temperature : 37 plus-minus 0.1 Celsius
degrees)
As protecting function, thermostat (12) is installed to the lower part of heat
Block (7).
In case the temperature exceeds 70 Celsius degrees, electric current to heater
(11) is turned OFF.
3-31
3.7.3.
1. Unit dismantling
1. Turn off system power source and PC
power switch.
2. Open top cover and push over upper
chassis gently to give a tilt.
-
Nozzle Plate
7. After taking off nut fixing nozzle plate
of reaction cuvette washing
mechanism, remove nozzle plate.
- Put removed nozzle plate nearby
lightly with piping as it is.
3-32
Chapter 3
Service Manual
Fixing Nut
8.
Fig. 3-46
MU-1
Fixing Nut
Mixing Base
Fig. 3-47
MU-2
Joining Plate
Lamp House
10. Remove fixing screws (four) of
joining plate.
Fig. 3-48
3-33
2. Cautions
1. Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electronic parts may be damaged by electric shock.
2.
3.
When taking out reaction cuvettes, take great care not to touch optical
path face.
If touched, wipe off finger marks gently with soft paper.
4.
When taking lamp out, take great care not to touch the glass of lamp.
If touched, wipe off finger marks with soft paper soaked in alcohol.
3-34
Chapter 3
Service Manual
3.8.
3.8.1.
Fig. 3-52
MU-2
3.8.2.
1. Function
MU-1 performs mixing of R1 and sample, and mixing of Cal-1 and R2
in reaction cuvette by air pressure.
MU-2 performs re-mixing before the measurement by air pressure.
2. Action (Refer to ASSY-10 for assembly drawing.)
The shaft of stepping motor (7) slides up or down when pulse is given.
This sliding motion is transferred to mixing base (1) through ball spline (3).
Spline bearing (5) has a shade pin which shades home position sensor (6) light at
home position. This detects home position at initialization.
- Home position is at the top.
At descent, shaft of stepping motor (7) goes down, and mixing base also goes
down.
- At descent, it is moved about 1 mm lower than the distance between packing
and reaction cuvette, and this excess is absorbed by spring (10) for strict
adhesion of packing and reaction cuvette.
At ascent, shaft of stepping motor (7) goes up, and mixing base (1) also goes up.
In measurement, if mixing base will not be at the home position within the fixed
time, system gives an alarm and emergency stop.
3-35
3.8.3.
1. Dismantling
1.
2.
Plate
Fixing Nuts
Mixing Base
3. Loosen fixing nuts (two) of mixing
base of MU-1, and remove them.
And take out mixing base with
piping.
Fig. 3-54
Fixing Nut
MU-1
Mixing Base
4. Loosen fixing nuts (two) of mixing
base of MU-2, and remove them.
And take out mixing base with
piping.
Fig. 3-55
3-36
MU-2
Chapter 3
Service Manual
Fixing Screw
5. Loosen fixing screws (two) of MU-1
and remove them.
And then remove unit
Fig. 56
MU-1
Fig. 57
Fig. 58
MU-2
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
3-37
Mixing base
4. Cautions
Be sure to turn off the system power before operating.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
3-38
Chapter 3
Service Manual
Connector
Pump Head
3.9.2.
1. Function
There are four pumps and one of them is a spare.
Two pumps are to supply MU-1 (two) with air pressure (suction and
pressurization), and one is to supply MU-2 with air pressure (suction and
pressurization).
2. Action (Refer to ASSY-11 for assembly drawing.)
When stepping motor (10) rotates, the rotation movement is transformed into
linear motion through crank shaft (8) fixed to motor shaft, flange bush (9)
and bearing case (7). And the linear motion is transferred to Diaphragm (3).
Stepping motor (10) drives four pumps ; one side of the shaft drives two pumps
And another side drives two pumps.
ON OFF of motor driving is controlled by the time.
3-39
4.
5.
6.
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
3-40
Chapter 3
Service Manual
Lamp House
Lamp
Fig. 3-62 Reaction Tray periphery
-
Fig. 3-63
Spectrophotometer
3.10.2
1. Function
Rear spectrophotometry is adopted.
White light from light source lamp is converged by lens, etc. and is led to go through
reaction cuvette, a cell for measurement, and is diverged by concave diffraction
grating.
After the diffraction, light energy is measured by 12 individual photo-detectors in
each designed wavelength. Monochromatic measurement takes the output of one
detector out of twelve, and bi-chromatic measurement takes two.
And the absorbance or the absorbance difference is calculated by the information
amplified by an amplifier and then average deviation and logarithmic transformed
signals.
3-41
2. Specifications
Light source
Spectrum element
Light detector
Wavelength
340, 380, 405, 450, 505, 546, 570, 600, 660, 700, 750, 800 nm
Wavelength accuracy
Linearity range
3-42
: Plus-minus 2nm
: 0 - 2.5 Abs.
Chapter 3
Service Manual
1. Unit dismantling
1.
2.
Fig. 3-65
Joining Plate
Lamp House
4.
5.
6.
Fig. 3-66
3-43
Fixing Screw
7.
8.
Fig. 3-67
Spectrophotometer
9.
10.
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
- Electric power should be off before the maintenance.
If connecting or disconnecting even if power is ON,
electric parts may be damaged by electric shock.
- Be sure not to loosen the white screws of position determination plate.
If they are loosened, the position to reaction tray may be shifted at assembly
and it may cause inferior performance.
Do not loosen the
white screws. (two)
3-44
Chapter 3
Service Manual
3.11
3.11.1
Nozzle base
Connector
Dispensation nozzle
Aspiration nozzle
Motor
Fig. 3-68 Reaction Cuvette Washing Station
-
3-45
The direction of motor rotation is shown when seen from the motor shaft
side.
In measurement, if nozzle base will not be at the home position within the fixed
time, system gives an alarm and emergency stop.
3-46
Chapter 3
Service Manual
RCT periphery
Piping for washing (two)
Manifold
Fig. 3-70
Piping of CWS
3-47
Connector
Fig. 3-72
Fixing Screw
5.
6.
7.
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
3-48
Chapter 3
Service Manual
3.12
Pump Head
Motor
Manifold
Connector
3-49
Phi 10 plunger (12) connected to rack (13) makes a linear motion with it.
Rack (13) has a shade plate which shades home position sensor (9) light at home
position or stop sensor light. This detects home position or stop position.
In measurement, if rack will not be at the home position within the fixed time during
measurement, system gives an alarm and emergency stop.
3-50
Chapter 3
Service Manual
1. Unit dismantling
1.
2.
3.
Pump Head
4.
5.
Fig. 3-75
CWP periphery
Manifold
6.
Remove connectors.
J5201 J5203
7.
8.
9.
3-51
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
-
3.12.4.
Exchange of U-packing
3-52
Chapter 3
Service Manual
Heater
Thermostat
Level Sensor
For Acidic Solution
Connector
Fig. 3-77 Washing Solution Pre-heater
-
3.13.2.
1.
Function
This pre-heater is used for heating water (five lines), alkaline solution (one line)
and acidic solution (one line) used for washing of reaction cuvettes after
measurement to 37 Celsius degrees.
This washing water is dispended into reaction cuvettes from dispensation nozzle
of reaction cuvette washing station.
2. Temperature control
The temperature of water is detected by the sensor (fixed to the center of
station, and heater (4) is controlled accurately by PID control on TCD PCB.
3-53
3-54
Chapter 3
Service Manual
1.
2.
3.
4.
5.
Remove connectors.
J5401 J5402 J5403
6.
WSH
Fig. 3-78 System right side view
Joint of Piping
Fig. 3-79
WSH
Joint of Piping
WSH periphery
7.
WSH
3-55
2.
Assembling
Assembling procedure is the reverse of the above.
3.
Cautions
-
3-56
Chapter 3
Service Manual
3.14.
3.14.1.
Pump Head
Solenoid Valve
Motor
1. Function
Reagent probe and sample probe aspirate reagent or sample from reagent bottle,
and dispense into reaction cuvette. After finishing dispensation, both inside and
outside of reagent probe or sample probe are washed by washing water in probe
washing pot.
This pump provides this washing water.
The amount of the dispensation for one time is approximately 1800 micro-liter.
3-57
3-58
Chapter 3
Service Manual
1.
2.
3.
4.
Fixing Screw
Joint of Piping
Fixing Screw
5.
6.
Joint of Piping
3-59
2.
Assembling
Assembling procedure is the reverse of the above.
3.
Cautions
-
3.14.5.
Exchange of U-packing
3-60
Chapter 3
Service Manual
Connector
Pump Head
Pump Head
3-61
3.15.3.
1. Unit dismantling
DP-1
DP-2
Fixing Screw
Connector
3-62
Joint of Piping
Chapter 3
Service Manual
2. Assembling
Assembling procedure is the reverse of the above.
3. Cautions
- Electric power should be off before the maintenance.
If connecting or disconnecting while power is ON,
electric parts may be damaged by electric shock.
- When removing piping, wipe up the water flown from tube.
If water is carried into electric parts such as sensors, they
may be damaged
- Take a great care not to be infected with any infection when dealing
with waste water.
3-63
Chapter 4
Service Manual
Chapter 4
Control PCB
This chapter explains functions of control PCB and when it is broken down.
Contents
4. 1
4. 2
4. 3
4.4
4.5
4.6
4.7
4.8
4.9
4-1
4.1.
MCPU PCB
4.1.1.
1. Outside view
MCPU PCB
2. Location
MCPU PCB
PCB Box
- Refer to WIRING-18 MCPU PCB for connecting of this board.
4-2
Chapter 4
Service Manual
4.1.2.
1. Function
Items
Contents
CPU
External memory
External I/O
CPU peripheral
control
I/F
Notes
HD64F7044F28
HM6216255HJP-15
PC16552DV x 3
FPGA
EPF10K10
SH-2
SRAM 512KB
UART 6ch
Heater control
Use of FPGA
Mother board bus
HOST PC I/F
Bar code 1 I/F
Bar code 2 I/F
ISE unit I/F
Printer I/F
Spare I/F
I/F for debug
SCPU I/F
Heater control I/F
Temperature sensor I/F
+ 5V
+ 12V
- 12V
Input power
2. Diagram
SCPU
FOR DEBUG
CONTROL
SYNCHRONIZED
SERIAL(TTL)
RS232C
DRV/
RCV
ADDRESS
DATA
MAIN CPU
CONTROL
CONTROL
SH-2
TEMP.
SENSOR
AMP
CONTROL
CONTROL
SRAM
SIO
DRV/
RCV
RS232C
SIO
DRV/
RCV
RS232C
CONTROL
BARCODE
READER-1
SIO
DRV/
RCV
RS232C
CONTROL
BARCODE
READER-2
SIO
DRV/
RCV
RS232C
CONTROL
ISE UNIT
SIO
DRV/
RCV
RS232C
CONTROL
SIO
DRV/
RCV
RS232C
CONTROL
CONTROL
A/D Port
ADDRESS
MOTHER
BOARD
BUFFER
I/O CONTROL
I/O
CONTROL
PC
PRINTER
SRAM : HM6216255HJP-15
FPGA LOGIC
TEMP. CONTROL
CONTROL
COUNTER
SPARE
HEATER
CONTROL
4-3
When this board is broken down, one or more of the following phenomena
may occur.
4-4
Chapter 4
Service Manual
1. Outside view
SCPU PCB
2. Location
SCPU PCB
PCB box
-
4-5
4.2.2.
1. Function
Item
Contents
CPU
External memory
External I/O
CPU peripheral
control
A/D converter
Analog
I/F
Notes
H8/534
SRAM 512KB
UV-EPROM 128KB
DP-RAM 32KB
A/D converter control
Analog input gain control
Use of FPGA
HD6475348
HM628512BLFP-7
M27C1001-12F1
IDT7005L55PF
FPGA
FPGA
EPF10K10
16 bits
Conversion of data after A/D
conversion, with ROM (log
conversion)
14ch
Possible for gain from CPU,
and offset regulation
Mother board bus
MCPU I/F
CS5016-JP32
M27C256B-15F1 x 2
Input power
2. Diagram
SYNCHRONIZED
ADDRESS
MCPU
FOR DEBUG
DATA
RS232C
DRV/
RCV
SUB CPU
H8/300
BUFFER
BUFFER
DPRAM
DEC
CONTROL
CONTROL
CONTROL
ADDRESS
SRAM
CONTROL
ROM
I/O CONTROL
I/O
CONTROL
CONTROL
DATA
GAIN ADJUSTMENT
DPM
AMP
MPX
SPM
AMP
AMP
AMP
A/D
A/D
Bus
CONTROL
BLOCK
DPM
AMP
OFFSET ADJUSTMENT
ROM
ROM
SPECTROPHOTOMETER
FPGA LOGIC
CONTROL
DPM
CONTROL
SERIAL DATA
4-6
MOTHER
BOARD
Chapter 4
Service Manual
4-7
4.3.
UCNT-1 PCB
4.3.1.
1. Outside view
UCNT-1 PCB
2. Location
UCNT-1 PCB
PCB Box
-
4-8
Chapter 4
Service Manual
4.3.2.
1. Function
Items
Motor control
Input power
Contents
STF (Sampler revolution)
STM R (STM revolution)
RCT
(Reaction tray revolution)
Spare 1
MP (Mixing pump)
STM UD (STM Up/down)
SP1
(SP washing)
(SP sample)
SP2
Spare 2
Encoder pulse detection
Pulse count
CPLD
Mother board bus
Sample ID/Stop sensor I/F
5 phase/2 phase stepping
motor I/F
PMDRV
SP1/SP2 I/F
Encoder I/F
Level sensor I/F
Notes
PPMC-111CFP
PPMC-111CFP
PPMC-111CFP
PPMC-111CFP
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
ZEN2061F
MSM82C54-2GS-K
EPM7128S
All kinds of sensors
Motor control signal
Motor control signal
Motor control signal
Encoder pulse input
Level sensor signal
input
+ 5V
+ 12V
4-9
2. Diagram
MOTHER
BOARD
PPMC-111C
(STF)
BUFFER
DEC
CONT
PMDRV PCB
PPMC-111C
(STM R)
STEPPING
MOTOR
DRIVER
PPMC-111C
(RCT)
PLD LOGIC
PPMC-111C
(MP)
PPMC-111C
(SPARE)
82C54
(STM U/D)
82C54
(SP-1)
SAMPLE
PUMP
82C54
(SP-2)
SP-1/SP-2
82C54
(SPARE)
82C54
LEVEL
SENSOR
SLS/RLS
STF/STM/
RTM/RCT/
RGT
ENCODER
PARALLEL
PORT
4-10
SENSOR AND
BUFFER
OTHER I/O
Chapter 4
Service Manual
In case that STF motor, sensor, encoder, or mother driver has trouble,
STF motion failure may occur. Checking is necessary.
In case that ID sensor of STF has trouble, sample tray number reading
failure may occur. Checking is necessary.
In case that STM motor, sensor, encoder or motor driver has trouble,
STM motion failure may occur.
4-11
4.4.
UCNT-2 PCB
4.4.1.
1. Outside view
UCNT-2 PCB
2. Location
UCNT-2 PCB
PCB Box
- Refer to WIRING-21 UCNT-2 PCB for connecting of this board.
4-12
Chapter 4
Service Manual
4.4.2.
1. Function
Items
Motor control
Peripheral control
I/F
Input power
Contents
RTM R
(Reagent probe revolution)
RGT
(Reagent tray revolution)
Spare 3
RTM-UD
(Reagent probe up/down)
RP1
(Reagent probe washing)
RP2 (Reagent pump)
CWS (Reaction cuvette
washing station)
MU1 (Mixing unit 1)
MU2 (Mixing unit 2)
PWP (Probe washing pump)
CPLD
Mother board bus
Stop sensor I/F
Switch sensor I/F
5 phase/2 phase stepping
motor I/F
PMDRV I/F
RP1/RP2 I/F
SSRAC I/F
+ 5V
Notes
PPMC-111CFP
PPMC-111CFP
PPMC-111CFP
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
MSM82C54-2GS-K
EPM7128S
All kinds of stop sensors
All kinds of switch sensors
Motor control signal
Motor control signal
Motor control signal
Parallel port
4-13
2. Diagram
MOTHER
BOARD
STEPPING
MOTOR
DRIVER
PPMC-111C
(RTM R)
BUFFER
DEC
CONT
PPMC-111C
(RGT)
PPMC-111C
(CWS)
PMDRV PCB
PPMC-111C
(SPARE)
82C54
(RTM U/D)
82C54
(RP-1)
REAGENT
PUMP
82C54
(RP-2)
RP-1/RP-2
82C54
(MU1)
82C54
(MU2)
82C54
(PWP)
PARALLEL
PORT
SENSOR AND
BUFFER
OTHER I/O
2.
4-14
In case that RTM motor, sensor, encoder or five-phase motor driver has
trouble, RTM motion failure may occur.
In case that RGT motor, sensor, encoder or five phase motor driver has
trouble, RGT motor failure may occur.
Chapter 4
Service Manual
3.
Motion failure of the side for washing of reagent pump (RP 1) occurs.
Note
4.
5.
8.
7.
6.
4-15
4.5.
LAMPDC PCB
4.5.1.
1. Outside view
LAMPDC PCB
2. Location
LAMPDC PCB
PCB Box
-
4-16
Chapter 4
Service Manual
4.5.2.
1. Function
Items
Output I/F
Contents
Notes
Input power
Power
supply
1.7A
+24V: 600mA
1A
2. Diagram
FOR ANALOGUE
+15V
POWER INPUT
-15V
+12V
DROPOUT
LEGULATOR
CIRCUIT
-12V
FOR ANALOGUE
POWER OUTPUT
DROPOUT
LEGULATOR
CIRCUIT
FOR LAMP
POWER INPUT
+16V
DROPOUT
LEGULATOR
CIRCUIT
+12V
FOR LAMP
POWER OUTPUT
LAMP DOWN
DETECT
FOR PRINTER
POWER INPUT
FOR PRINTER
+5V
+16V
+24V
INSULATION TYPE
POWER OUTPUT
DC-DC CONVERTER
FOR ISE
POWER INPUT
+24V
INSULATION TYPE
DC-DC CONVERTER
+12V
FOR ISE
POWER OUTPUT
FOR MCPU
POWER OUTPUT
4-17
Even if you measure with clean cuvette, the reading of cell blank becomes
larger and data error mark, C, c comes out.
2.
3.
4.
5.
4-18
Chapter 4
Service Manual
4.6.
PM DRV PCB
4.6.1.
1. Outside view
PMDRV PCB
2. Location
PMDRV PCB
PCB Box
-
4-19
4.6.2.
1. Function
Items
Motor driver
Contents
STF
: TA8534H
STM UD : TA8534H
MU1
: TA8534H
MU2
: TA8534H
RTM UD : TA8534H
CWS
: TA8534H
MP
: TA8534H
: TA8534H
PWP
TD62003AP
2SD987
Isolation by photo
coupler
+ 5V
+ 24V
Notes
5ch
3ch
2. Diagram
UCNT-1 I/F
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
STF UNIT
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
STM U/D
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
MP UNIT
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
RTM U/D
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
MU-1 UNIT
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
MU-2 UNIT
PHOTOCOUPLER
CONTROL
CIRCUIT
MOTOR
DRIVER
CWS UNIT
UCNT-2 I/F
PHOTOCOUPLER
4-20
CONTROL
CIRCUIT
MOTOR
DRIVER
PHOTOCOUPLER
CONTROL
CIRCUIT
DC
SOLENOID
PHOTOCOUPLER
CONTROL
CIRCUIT
DC
SOLENOID
PWP
WSD UNIT
SPARE
Service Manual
Chapter 4
Sample tray (STF) rotation error comes out, or rotating sound becomes louder.
2.
Error comes out in up/down motion of sample probe (STM), or motion sound
becomes louder.
3.
Error comes out in up/down motion of mixing unit-1 (MU-1), or motion sound
becomes louder.
4.
Error comes out in up/down motion of mixing unit-2 (MU-2), or motion sound
becomes louder.
5.
Error comes out in up/down motion of reagent probe (RTM), or motion sound
becomes louder.
6.
Error comes out in up/down motion of reaction cuvette washing station (CWS), or
motion sound becomes louder.
7.
Mixing pump (MP) does not rotate, or rotating sound becomes louder.
4-21
4.7.
TCD PCB
4.7.1.
1. Outside view
TCD PCB
2. Location
TCD PCB
PCB Box
-
4-22
Chapter 4
Service Manual
4.7.2
1. Function
Items
Heater driver
Driver control
signal I/F
Input power
Contents
Notes
RCT
Isolation by photo coupler
+ 5V
+ 24V
2. Diagram
Main
CPU
Photo coupler
Driver
Smoothing
circuit
RCT Unit
4-23
4.8.
SSRAC PCB
SSRAC PCB
2. Location
SSRAC PCB
4-24
Chapter 4
Service Manual
4.8.2
1. Function
Items
SSR
Contents
CWP
Notes
: G3MC-202P-VD-1 DC50
Input power
Supply power
AC100V
2. Diagram
L
SSR
UCNT-2 I/F
SSR
SSR
SSR
SSR
F1 0.5A
(T-250V, 0.5A)
F2 0.5A
(T-250V, 0.5A)
F5 1A
(T-250V, 1A)
AC100V
INPUT
CWP UNIT
DP-1
DP-2
F6 1A
(T-250V, 1A)
F7 2A
(T-250V, 2A)
WSH UNIT
4-25
4-26
1.
Reaction cuvette washing pump (CWP) wont function, or wont stop functioning.
- Washing solution does not come out from dispensation nozzle of reaction
cuvette washing station (CWS), or is flowing all the time.
2.
3.
The temperature of washing solution does not rise to the set temperature,
or rises abnormally high.
Chapter 4
Service Manual
Mother Board
DC 5.18V - 5.20V
+ -
DC5V Input
* In case the voltage is not the reference value, adjust the voltage, referring to the
next page.
4-27
Volume
4-28
Chapter 4
Service Manual
4.10
Fuse
T-250V 6.3A.
AC Input
Power Supply Panel (Left rear)
4-29
Chapter 5
Service Manual
Chapter 5
Contents
5.1
5.2
Password Setting
5.2.1 Password Setting Procedure ------------------------------------------------ 5-5
5.2.2 Password Setting of Patient Information ----------------------------- 5-7
5.3
5-1
5.1
5.1.1
1. Click
Maint.
Double-click
little above the
boundary .
OK
button.
After checking the input against the registered word, the enlarged menu is
shown if it is identical.
Refer to the next section 5.2 Password Setting for setting the password.
5-2
Chapter 5
Service Manual
5-3
5.1.2
1. The followings show the screens mentioned in the enlarged sub-menu and their
functions.
Refer to the following pages for the details.
Temperature
Offset
The input screen of the offset values for the unit alignment.
Manual Function
5-4
?
The screen to see the information of encoder.
The monitor screen of the position sensor.
The same screen as the screen of Offset above.
The screen to combine the manual operations to
program.
The screen for the adjustment of the intensity
of light.
?
Chapter 5
Service Manual
5.2
Password Setting
5.2.1
1. Click the top right of the Main Control screen to close or minimize the screen.
Click
Selecting t224com.ini,
double-click.
Click
Click Prg.
5-5
5.2.2
Password Setting
1. Input Password of 16 characters or less in PSWD= of [Order] in Program
list. For example, PSWD= tbms.
is necessary.
2. Save this file.
File Overwritten
3. Close Memo.
File Close
5.2.3
Password Setting
5-6
Chapter 5
Service Manual
5.2.4
Password Setting
1. Input Password of 16 characters or less in pswd = of [System] in
Program list. For example, pswd = tms.
is necessary.
2. Save this file.
File Overwritten
3. Close Memo.
File Close
After completing setting of password, return the Main screen to the former size.
5-7
5.3
1.
2.
5-8
3.
Save it.
4.
Chapter 5
Service Manual
5.4
5.4.1
The temperature
display and power
display of the
reaction tray
Setting of the
time to plot.
2. When clicking
LevelPara
To the temperature
parameter screen.
Input of Temp.
parameter.
Input of Temp.
parameter.
To go back to
the temperature
monitor screen.
5-10
Set
Return
button.
Chapter 5
Service Manual
5.5
Choice between
CW and CCW
Update
Exit
button.
5-11
5.6
1. Selection
of unit.
Click, when
operating with
NO Safety.
2. Selection
of function.
3. Perform
function.
Sub-screens
Manual Function screen
For initialization
of all the units.
Unit
Unit
5-12
Function
RUN
(abbreviations)
1. STF
2. STM-R
3. STM-UD
4. SP
5. RGT
6. RTM-R
7. RTM-UD
8. SP
9. RCT
10. MU-1
11. MU-2
12. MP
13. CWS
(units)
Sampler
STM Rotation
STM up/down
Sample pump
Reagent tray
RTM rotation
RTM up/down
Reagent pump
Reaction tray
Mixing unit-1
Mixing unit-2
Mixing pump
Cuvette washing station
14. CWP
15. PWP
16.
17.
18.
19.
Drainage pump-1, 2
Colorimetry detector
?
Electromagnetic valve
monitor
DP 1/2
Detector
TEST
EVS
(movement)
Rotation of sampler
Rotation of sample probe
Up/Down movement of sample probe
Aspiration/Dispensation movement of sample pump
Chapter 5
Service Manual
5.6.1
STF
(Sampler)
1. Explanation of function
4. ADJUST
5. CW1Pitch(OUT)
6. CW1Pitch(IN)
7. CCW1Pitch(OUT)
8. CCW1Pitch(IN)
9. POSITION
10. Low Speed
5-13
5.6.2
STM-R
1. Explanation of Function
5-14
Chapter 5
Service Manual
5.6.3
STM-UD
1. Explanation of Function
3. ADJUST
8. Move Speed
9. LS(STM) -> ON
5-15
5.6.4
SP
(Sample Pump)
1. Explanation of Function
4. PUSH(1000uL)
5. Move Speed
6. EV1-ON
7. EV1-OFF
8. INIT
9. PULL(15uL)
10. PUSH(15uL)
5-16
Chapter 5
Service Manual
5.6.5
RGT
(Reagent Tray)
1. Explanation of Function
4. ADJUST
5. CW-1Pitch
6. CCW-1Pitch
7. CCW-1.6(BOTL12)
8. CW-1.6(BOTL1)
9. POSITION
10. Low Speed
5-17
5.6.6
RTM - R
1. Explanation of Function
5-18
Chapter 5
Service Manual
5.6.7
RTM - UD
1. Explanation of Function
3. ADJUST
8. Move Speed
9. LS(R1) - ON
11. LS(R2) - ON
5-19
5.6.8
RP
(Reagent Pump)
1. Explanation of Function
11. EV2-ON
12. EV2-OFF
13. EV3-ON
14. EV3-OFF
5-20
Chapter 5
Service Manual
5.6.9
RCT
(Reaction Tray)
1. Explanation of Function
2. INIT
3. OFFSET
4. ADJUST
5. CW-1/2 Pitch
6. CW-1 Pitch
7. CCW-1 Pitch
8. Stage-A
9. Stage-B
10. Stage-C
11. POSITION
12. Low Speed
13. High Speed
14. Pulse Count
15. Init Speed
5-21
5.6.10
MU-1
(Mixing Unit-1)
1. Explanation of Function
2. INIT
3. ADJUST
5. DOWN
6. Move Speed
5-22
Chapter 5
Service Manual
5.6.11
MU-2
(Mixing Unit-2)
1. Explanation of Function
2. INIT
3. ADJUST
5. DOWN
6. Move Speed
5-23
5.6.12
MP
(Mixing Pump)
1. Explanation of Function
One rotation
3. Low Speed
4. High Speed
5. Pulse Count
6. Init Speed
5-24
Chapter 5
Service Manual
5.6.13
CWS
1. Explanation of Function
2. INIT
3. ADJUST
5. DOWN
6. Low Speed
7. High Speed
8. Pulse Count
9. Init Speed
5-25
5.6.14
CWP
1. Explanation of Function
2. 1 Cycle
5-26
Chapter 5
Service Manual
5.6.15
PWP
1. Explanation of Function
2. INIT
3. PULL(1000uL)
4. PUSH(1000uL)
5. Volume Set
7. EV4-ON
8. EV4-OFF
5-27
5.6.16
DP-1/2
(Drainage Pump - 1, 2)
1. Explanation of Function
5-28
2. DP1 ON
DP-1 movement ON
3. DP1-OFF
4. DP2-ON
DP-2 movement ON
5. DP2-OFF
6. EV5-ON
7. EV5-OFF
8. EV6-ON
9. EV6-OFF
Chapter 5
Service Manual
5.6.17
1. Explanation of Function
2. Start Detector
Not ready.
3. Adjust Gain
Not ready.
4. Test(W1)
Not ready.
5. Test(W2)
Not ready.
6. Test(Ready)
Not ready.
7. Stage-C
Not ready.
8. POSITION
Not ready.
9. INT
Not ready.
Not ready.
11. CW 1 Pitch
Not ready.
Not ready.
5-29
5.7
5.7.1
1.
Encoder Screen
Unit names
The values
are shown.
Encoder screen
Encoder screen
5-30
1. Offset
2. Offset+Theory
3. Actual measurement
4. Error Limit
6. Position
7. Read
8. Exit
Chapter 5
Service Manual
5.7.2
1.
Sensor Screen
Electromagnetic
valves
Sensor screen
Sensor Names
1. Read
2. EXIT
5-31
5.7.3
1.
Select
program.
Manual Auto
5-32
1. Edit
2. New
3. Start
4. Stop
5. Exit
Chapter 5
Service Manual
or
New
5-33
5.7.4
1.
Gain Screen
2.
1) Click
Read All
2) Click
Offset
3) Clicking
Parameter
All CH Adjust
Input 1000.
5-34
Chapter 5
Service Manual
4. Input 1000 in the blank of Raw Data, and click Set button.
Clicking Set or Cancel button, the dialog box is closed.
5. Move the reaction cuvette with de-ionized water in it on the optic axis in
the following procedure.
Gain
Save
8. Move the reaction cuvette with de-ionized water in it to the original position as
follows.
- Click INIT button in the row of RCT Move.
- The reaction cuvette with de-ionized water in it goes back to the position
originally set.
9. Take out the reaction cuvette with de-ionized water in it, and set the original
reaction cuvette.
10. Go back to the manual function screen by clicking
Exit
button.
5-35
5.7.5
1.
PF BTN Screen
PF BTN screen
1. Button Name
2. File Name
3. Button comment
4. Set
5. Exit
6. Auto Gain Button
These numbers
correspond with
in the user maint.
screen.
5-36
:
:
:
:
Parts
Reaction Tray
Contents
Cleaning
Sampling Pump
Cleaning
Exchange of 10U-Packing
Reagent Pump
Cleaning
Exchange of 10U-Packing
Cleaning
10 packing (accessory)
Reaction Cuvette
Washing Pump
Cleaning
Applying of Silicon Oil to 10 plungers
Re-tightening of 10U-Packing for 6th month and exchange for 12th month..
10 packing (accessory)
Sampling Probe
Cleaning
Cleaning
Reagent Pump
Reaction Cuvette Wash Pump Movement Check (Checking on Leakage, Air Bubble and so on.)
Sampling Probe
Reagent Probe
Detector
Light Source Lamp (accessory)
Automated Sampler
Cleaning
Reagent Tray
Mixing Unit-1
Exchange of Packing
Mixing Unit-2
Contents
Parts
Reaction Cuvette Wash
Station
Exchange of Piping Tube
Water Supply Tube Set B (WSH - CWP Manifold, WSH - EV6 NC)
(Accessories)
Water Supply Tube Set C (WSH - SP EV1 NO, WSH - RP EV2 ON)
Water Supply Tube Set D WSH - PWP IN, PWP OUT - EV4 COM etc.)
Mixing Pump
Control PCB
Printer
Cleaning
Power Supply
Body
Total Inspection
Cleaning (Pot)
Sample Barcode
Reagent Barcode
Reading Test
Reading Test