DOCUMENT REVISION SUMMARY
Title:LIQUID PARTICLE COUNT (LPC-S02)
Page No:
Current Revision
Date
1 of 1
09
10 May 2014
Prepared By :
OGY SDN.BHD
N SUMMARY
Doc No : CM-LAB-009
Remark
Approved By :
DOC. NO
CM-LAB- 009
KHOO WEI LIANG
PREPARED BY
ROSHAZAIMI
APPROVED BY
REF. DOC. NO
92-004230 Rev AK
ANALYTICAL LAB
TEST
DATE
10-May-14
PAGE
1 of 1
Fig. 1
Ultrasonic Extraction Conditions:
Component
Base
Cover
Antidisk
Fig. 2
in a clean
water tube.
hydrophone
measurement
nd DC light
he level used
ute before
endix XXX.
rge it into
ally in the
ately 1/2'' from
ker.
Fig. 2
Fig. 3
DOC. NO
CM-LAB- 009
KHOO WEI LIANG
PREPARED BY
SRI RAHAYU
APPROVED BY
REFERENCE DOC NO
92-004230 Rev AK
ANALYTICAL LAB
TEST
DATE
10-May-14
PAGE
1 OF 1
METHOD
a) Fill Ultrasonic tank with volume of DO water according to volume of water extraction. Make sure water level in tank same as wate
b) DEGAS the water in ultrasonic tank by turning on the power to 100% of full power for 10 min.
3. Blank Determination
a) Set the ultrasonic power (2.900 0.2 V AC for 2L beaker/ 3.100 0.2 V AC) using a WD approved hydrophone.
b) Then place the beaker filled with extraction water into the ultrasonic bath.
c) Lower appropriate fishing line fixture into the beaker and ensure it is equal to the level used in sample determination
d) Start the timer and sonicate for 2 minutes 2 seconds
e) Transfer the beaker to the LPC stir base.
f) Put the stir bar into the beaker by using retriever to pull and move down slowly.
g) Turn the stir bar rate to 75-150rpm for 1 minute.
h) Place the sample capillary into the beaker at the center and measure the sample counts.
Note: DC light value should not exceed 0.5, discard the blank and perform sensor cleaning as per instruction from manufacturer.
4. Sample Determination
d) Lower the fishing line fixture into the beaker and ensure that the part is horizontally in the center of the extraction beaker and th
1/2'' from surface water.
e) Repeat step d to h in Section 1.
f) Calculate the result based on calculation given below.
SKETCH/REMARKS
RESULT CALCULATION (SEAGATE).
Fig. 1
Ultrasonic Extraction Conditions:
Component
Base
Cover
Antidisk
Fig. 2
METHOD
until overflow. Start timer to record the time taken to obtain 40 2% DO level water
his time during sample extraction.
ng to volume of water extraction. Make sure water level in tank same as water level in beaker.
sure that the part is horizontally in the center of the extraction beaker and the top part is vertically
w.
SKETCH/REMARKS
Fig. 2
Fig. 3