Documente Academic
Documente Profesional
Documente Cultură
2
3
Scene safety
BSi & Provide privacy
Presses ON* .
Prepares patient*
Explain the reason for ECG monitoring, reassuring the client that it allows immediate treatment
of abnormal rhythms if necessary. Explain alarms, their purpose, and possible causes such as
loose or disconnected lead wires Activity is permitted within ordered restrictions while on the
monitor
Bares patients chest
Prepares electrode sites with brisk dry rub
provide additional preparation needed for patient with excessive hair, oily or damp skin
*Check equipment for damage (i.e., fraying, bent, or broken wires). Connect lead wires to cable,
and secure the connections.
*Select electrode sites on the chest wall, considering lead to be monitored, skin condition, and
any incisions or catheters.
*Before Applies ECG electrodes
Confirms package is sealed and use by date has not passed
Attaches an electrode to each lead wire
Inspects electrode gel for moisture content
Identifies appropriate electrode sites for 3-wire cable in upper right,Upper left, lower left
torso areas; or 5-wire cable in upper right and left.
Applies electrodes flat to skin, smoothing tape outwardly.
Avoiding pressing on center of electrode Secures cable clasp to patients clothing
1
0
C
C
1.
5
1.
5
1.
5
1.
5
1.
5
1.
5
1.
5
C
1.
5
1.
5
C
1
4
1.
5
CARIAC MONITORINING
Note: candidate must complete all critical criteria . Critical action listed
above all as *
1
2
3
4
5
6
7
8
9
1
0
11
1
2
1
3
1
4
1
5
1
6
Scene safety
BSi & Provide privacy
Explain the procedure to the patient. Identifies patient, , ensures
written consent has been obtained if not emergency situation
.Identifies the indications for synchronized cardio version
*Establish IV access with a sodium chloride 0.9% running line .
Obtains baseline VS and rhythm strip, to include EKG if time allows.
Prepare airway ,suction and ventilation equipment .
Consider sedation as per sedation protocol or policy .
*Position the electrodes and pads as appropriate
Confirms the dysrhythmia
*Ensure that synchronizer is on .
*Confirm the synchronization is occurring on the R wave .change the
ECG amplitude on the monitor if necessary.
Ensure the patient is well oxygenated / ventilated prior to and after *
.sedation and cardio version
ENERGY SELECT" to choose the energy to be delivered. (monitor "
specific)
Perform a maximum of three attempted synchronized cardio *
versions in accordance with following joule sequence : 50, 100, 150
.joules or as per protocol
Turns on ECG recorder
C
C
C
C
1.5
C
C
1.5
1.5
C
1.5
1.5
1.5
C
1.5
C
1
7
1
8
1
9
2
0
2
1
2
2
1.5
C
1.5
1.5
Note: candidate must complete all critical criteria . Critical action listed
above all as *
S
1
2
3
4
5
6
Action
Scene safety
BSI
Positions the patient
Identifies indications for transcutaneous pacing.
Considers O2, IV, VS, rhythm strip with EKG if time allows.
If necessary, clip the hair over the areas of electrode placement. Do not
p
C
C
C
C
C
C
7
8
9
1
0
1
1
1
2
1
3
1
4
1
5
1
6
1
7
1
8
1
9
2
0
2
1
2
2
C
1.5
C
1.5
1.5
1.5
*Prepare to pace the heart. After making sure the energy output in
milliamperes (mA) is on 0, connect the electrode cable to the monitor
output cable.
*Gradually increase delivered current until capture is achieved (observes
pacer spikes
followed by wide QRS complexes at tall T waves)
*Check the waveform, looking for a tall QRS complex in lead II.
1.5
1.5
1.5
1.5
Monitor the patients heart rate and rhythm to assess ventricular response
to pacing.
Assess the patients vital signs, skin color, level of consciousness, and
peripheral pulses.
Take blood pressure in both arms.
*Document
all important
steps . Colleges for Health Sciences
AL-Ghad
International
TRANSCUTANEOUS PACING
Note: candidate must complete all critical criteria. Critical action listed
above all as *
C
1
Note: candidate must complete all critical criteria. Critical action listed
above all as *
5.
5.
5.
5.
5.
5.
5.
5.
C
1
5.
1
1
C
5.
1
1
1
C
1
Note: candidate must complete all critical criteria. Critical action listed
above all as *
1
1
5.
C
5.