SPECULAR MICROSCOPE
SP-3000P
TOPCON CORPORATION
R-SP3000P-1-0601-01
CONTENTS
Page
1. INTRODUCTION ..................................................................................................................................1
1-1
1-2
1-3
1-4
1-5
Nomenclature...........................................................................................................................................3
Optical System (Arrangement) ................................................................................................................5
Electric System ........................................................................................................................................6
Principles .................................................................................................................................................8
PCB Arrangement..................................................................................................................................10
4. TOOL LIST............................................................................................................................................72
4-1 Special Repair Tools..............................................................................................................................72
4-2 General Repair Tools .............................................................................................................................73
R-SP3000P-1-0601-01
1. INTRODUCTION
1-1 Features of Product
(1) Photography of the corneal endothelium and the measurement of the corneal thickness can be performed
at the same time.
(2) The auto-alignment function ensures quick and easy photography and measurement.
(3) Simplified cell analysis enables the user to easily understand the rough value of the cell density.
0.25 0.5mm
Photography/measuring point
Photography magnification
0.01mm
Alignment
Monitor
Photography mode
AUTO/MANUAL/SEMI-AUTO switching
Flash level
HIGH/LOW switching
Image memory
Working distance
25mm
RS-232C IN
Operating temperature
10 ~ 40C
40mm
92mm
28mm
60mm
Source voltage
AC100-240V, 50Hz/60Hz
Power input
190VA
Dimensions
Weight
22kg
* Subject to change in specifications and appearance without advance notice for future improvement.
R-SP3000P-1-0601-01
Initial inspection
Applicable item
Page
P. 16 ~ 23
P. 24 ~ 33
Troubleshooting
P. 34 ~ 43
Trouble
Inspection
OK
NO
Trouble location
check
Repair operations
Inspection
NO
P. 24 ~ 33
P. 44 ~ 71
P. 16 ~ 23
Normal
YES
End
Photography principles
P. 8 ~ 9
1-5 Cautions
Refer to the illustrations of the parts list for the disassembly and assembly of the unit which is not
described in this manual. Refer to the instruction manual for replacing the consumables.
If the optical system is soiled when repairing, photographing is adversely affected. Be careful not to touch
or soil the optical system when disassembling, assembling or adjusting.
R-SP3000P-1-0601-01
2. PRODUCT OUTLINE
2-1 Nomenclature
1
4
6
No.
Name
Function
On this monitor, the aligned anterior eye segment image, the aligned
corneal endothelium image and the photography result are displayed.
Photography switch
Press this switch after alignment while observing the corneal endothelium, and it is possible to take a picture. Press this switch when the
memory pictures are displayed, and the memory pictures will be
switched by turns. This switch is also available for resetting when the
power saving function is working.
Joystick
Locking knob
R-SP3000P-1-0601-01
10
11
12
13
16
17
15
No.
14
Name
10
Forehead rest
11
Photography window
12
13
Chinrest
14
Power cord
15
Fuse holder
16
Mouse
17
Power switch
Function
R-SP3000P-1-0601-01
No.
Name
Function
XY LED
XY alignment sensor
Z LED
Observation slit
Photographing slit
Xenon
Z alignment sensor
10
Mask lens
11
Knife edge
12
13
CCD camera
14
15
R-SP3000P-1-0601-01
CONTROL SW
PCB UNIT
CHINREST CONTROL
PCB UNIT
CPU-I/O
CABLE 2 UNIT
CONDENSER UNIT
LCD YC CABLE
STEPPING MOTOR
FOR Z AXIS
XENON UNIT
Z AXIS MOTOR
CABLE UNIT
STEPPING MOTOR
FOR X AXIS
PD SENSOR UNIT
USB I/F
CABLE UNIT
USB MOUSE
CABLE UNIT
RS-232C OUT
CABLE UNIT
RS-232C IN
CABLE UNIT
EXTERNAL SYSTEM
CABLE UNIT
VIDEO OUT
CABLE UNIT
SOLENOID UNIT
Z LED
PCB UNIT
XY LED
PCB UNIT
XY SENSOR
PCB UNIT
Z SENSOR
PCB UNIT
LCD POWER CABLE
CPU-I/O
CABLE 1 UNIT
3P SW CABLE UNIT
6
R-SP3000P-1-0601-01
Xenon tube
Sensor unit
CCD camera
Solenoid
Base unit
Chinrest unit
Chinrest
motor
Power unit
Video OUT connector
Commercial
power supply
RS-232C IN connector
R-SP3000P-1-0601-01
2-4 Principles
2-4-1 Photography principles
Slit light which enters the cornea gets through it and reaches the aqueous humor. Part of the light is reflected
on the front of the cornea in the boundary whose refraction index is different.
When the light reflected on the backside of the cornea is magnified, it can be observed as an image, because
the cells on the backside of the cornea are uneven.
Cornea
Photography
system (P)
22.5
22.5
R'
R
Illumination
system (I)
S'
S
Arrange the illumination system (I) and the photography system (P) so that the normal line which is perpendicular to the cornea may divide into I and P equally. Consequently the light is reflected on the backside of
the cornea. Move the instrument while keeping the angles of I and P uniform so that the normal line may
divide into I and P equally. So the front reflection of the cornea (S'-S) and the backside reflection of the cornea (endothelium cells R'-R) can be observed.
Front of cornea
Backside reflection
L
Reflected
light
Front reflection
Line sensor
Slit light
Corneal
thickness
R-SP3000P-1-0601-01
R-SP3000P-1-0601-01
Z SENSOR PCB
J502
Z SENSOR PCB
J501
XY ANALOG PCB
J401
VIDEO OUT
EXTERNAL
SYSTEM
D-sub
CONTROL SW
PCB
J702
IO PCB
J204
IO PCB
J203
USB MOUSE
10
R/L SENSOR
RS-232C IN
Din
RS-232C OUT
D-sub
USB I/F
R-SP3000P-1-0601-01
X AXIS MOTOR
Z AXIS MOTOR
Y AXIS MOTOR
X AXIS ZERO
DETECTION
SENSOR
Z AXIS ZERO
DETECTION
SENSOR
CCD CAMERA
POWER SUPPLY
LCD POWER
SUPPLY
PD SENSOR
DETECTION
CHINREST
CNTROL PCB
J601
Y AXIS
LOWER LIMIT
DETECTION
SENSOR
SWITCHING
POWER SUPPLY
CN6
Y AXIS
UPPER LIMIT
SENSOR
SWITCHING
POWER SUPPLY
CN3
CPU PCB
J102
SWITCHING
POWER SUPPLY
CN4
11
CONDENSER UNIT
XENON UNIT
CPU PCB
J101
R-SP3000P-1-0601-01
XY SENSOR PCB
J701
SOLENOID UNIT
XY LED PCB
Z LED PCB
CENTER FIXATION LED
PCB
VERTICAL FIXATION LED
PCB
LATERAL FIXATION LED
PCB
ANTERIOR EYE SEGMENT
ILLUMINATION LED PCB
LAMP HOUSE COVER
SENSOR
CPU PCB
J105
12
R-SP3000P-1-0601-01
XY ANALOG PCB
J402
CPU PCB
J107
CPU PCB
J106
13
R-SP3000P-1-0601-01
I/O PCB
J202
I/O PCB
J218
3P SWITCH
I/O PCB
J201
CHINREST DC
MOTOR
I/O PCB
J216
14
R-SP3000P-1-0601-01
CPU PCB
J110
JOYSTICK SW UNIT
15
R-SP3000P-1-0601-01
3. REPAIR WORK
3-1 FACTORY MENU Program
Carry out the following check to find a trouble early and to adjust the instrument.
Program name
Contents
-----
6. XY LED ADJUST
Sets a test eye and stores the mask slit inside dimension.
Initial value: 0.1985
2. Z SENSOR ADJUSTMENT
3. PACHOMETRY ADJUST
4. OPTICAL ADJUST
5. OPTICAL CHECK
2. SWITCH CHECK
3. LED CHECK
4. XENON CHECK
16
R-SP3000P-1-0601-01
Checks the operation of the X, Y and Z motors and the photo interrupter.
X drive - Reciprocation : OK or NG is displayed.
Y drive - Reciprocation : OK or NG" is displayed.
Z drive - Reciprocation : OK or "NG" is displayed.
7. SOLENOID CHECK
Checks the charge and discharge of the condenser under the cover
IN/OUT condition.
Checks the operation of the lamp house cover photo interrupter.
The following data are displayed by opening/closing the lamp house
cover.
COVER
: IN/OUT OK or NG is displayed.
CHARGE ON : OK or NG is displayed.
CHARGE OFF : OK or NG is displayed.
9. AGING TEST
2. ID INPUT
3. ADVANCED MORPHOLOGY
4. MACHINE PARAM
17
R-SP3000P-1-0601-01
Clear switch
Delete switch
(2) The following display appears on the screen. Press the [Image transfer] switch.
(3) Press the [Photography] switch. Move the cursor to the item to be checked and press the [Image transfer]
switch.
The item list is divided into four displays. Use the [Image selector] switch to move to the next display.
(4) On each checking display, press the [Clear] switch to return to the item list display.
(5) Turn off the power switch to end FACTORY MENU.
18
R-SP3000P-1-0601-01
Caution:If the programs of "FACTORY MENU GROUP-A/B" are executed without installing the special tools, the memory data may be changed. Be careful.
This chapter describes the programs 1 ~ 12 of "FACTORY MENU GROUP-C" which are used to check the
functions. Refer to the items of "3-4 Assembly and Adjustment Procedures" for the "FACTORY MENU" programs which are used for adjustment.
Clear switch
Delete switch
Menu switch
Image transfer switch
Delete switch
Menu switch
Mask #1
The alignment frame is displayed at the center.
Mask #2
Mask #3
The perpendicular line is displayed at the center.
Mask #4
This is the display for correcting the slit inside dimension.
19
R-SP3000P-1-0601-01
Clear switch
: Turns on/off the mask image display.
Image transfer switch : Lights the xenon and takes a picture only when observing the endothelium.
Image selector switch: Opens and closes the solenoid.
The following table shows the relation between the light sources in this case.
Anterior eye segment observation
Endothelium observation
XY LED
ON
OFF
Z LED
OFF
ON
20
R-SP3000P-1-0601-01
Display on monitor
E2PROM CHECK
SWITCH CHECK
PCB.
LED CHECK
XY LED
LED.
(2) Make sure that each LED
blinks every 2 seconds on
the monitor.
21
R-SP3000P-1-0601-01
Check program
Display on monitor
XENON CHECK
out.
SOLENOID CHECK
22
R-SP3000P-1-0601-01
Check program
Display on monitor
not switched:
Check the photo interrupter.
Check the CPU PCB.
Check the I/O PCB.
RS232C CHECK
RS-232C output
terminal
RS-232C cable
RS-232C input
terminal
23
R-SP3000P-1-0601-01
3
9
7
6
10
12
Illustration
11
13
9
Proc.
1
10 14
14
Disassembly item
Removing
cover 2
Removing
right cover
the
the
front
head
12
Removal method
.
.
.
.
10 .
12 .
14 .
24
R-SP3000P-1-0601-01
Illustration
Proc.
1
Disassembly item
Removing the
video monitor 2
color
Removal method
25
R-SP3000P-1-0601-01
5
4
Illustration
3
Proc.
1
Disassembly item
Removing
PCB 1
the
CPU
Removal method
Remove the spacer 2 .
Remove the two screws
26
R-SP3000P-1-0601-01
Illustration
1
3
4
Proc.
Disassembly item
Removal method
Removing
cover 1
base
the
27
R-SP3000P-1-0601-01
Illustration
10
9
Proc.
Disassembly item
Removal method
28
for each
It is necessary to adjust
each PCB.
R-SP3000P-1-0601-01
Illustration
Proc.
Disassembly item
Removing
camera 3
the
Removal method
1
CCD
It is necessary to adjust
the CCD camera position.
29
R-SP3000P-1-0601-01
5
Illustration
Proc.
Disassembly item
Removal method
.
.
30
R-SP3000P-1-0601-01
1
9
10
9
Illustration
8
7
10
Proc.
Disassembly item
Removal method
31
R-SP3000P-1-0601-01
Illustration
1
2
Proc.
Disassembly item
Removal method
Remove the connector J406.
Remove the two screws 4 .
Remove the two screws 6 .
Remove the two screws 2 at each
place.
32
R-SP3000P-1-0601-01
Illustration
7
4
6
3
1
2
Proc.
Disassembly item
Removal method
33
R-SP3000P-1-0601-01
3-3 Troubleshooting
3-3-1 Troubles related with the color video monitor
(1) No image is displayed.
(2) No character is displayed.
3-3-2 Error messages are displayed on the screen.
(1) XENON OVER
(2) XENON POWER PCB ERROR
(3) CABLE UNCONNECTED
(4) RS232C FAIL
(5) USB FAIL
(6) CHANGE BATTERY
(7) COVER OUT
3-3-3 Troubles at alignment
(1) Automatic alignment does not operate.
(2) Automatic alignment operates but the endothelium is not photographed. (The image cannot be changed to
the endothelium one or it is difficult to change.)
(3) The image is stored but flash is not done. (The photographed image is very dark.)
(4) The image is not stored and flash is not done. (Alignment still continues.)
3-3-4 After photographing
(1) The image cannot be transferred for IMAGEnet.
(2) Data cannot be inputted or outputted by RS-232C.
(3) Data cannot be outputted by USB.
(4) Zooming and simplified cell analysis cannot be done.
34
R-SP3000P-1-0601-01
J103
CPU PCB
LCD
J101102
J104
VOLUME
PCB
J203J204
DATA
I/O PCB
J206
+12V
J205
+12V
CCD camera
Image signal
CCD camera
A/D conversion
DATA
CPU PCB
A/D conversion
35
Image
signal
R-SP3000P-1-0601-01
36
R-SP3000P-1-0601-01
37
R-SP3000P-1-0601-01
38
R-SP3000P-1-0601-01
Connect the synchroscope to the following test point of the I/O PCB, which corresponds to the relevant
motor. Start MOTOR P.I. CHECK and check if the signal is changed (LOW level HIGH level).
X motor: TP214
LOW level (near 0V) in normal status
HIGH level (5V) when the original point is detected
Z motor: TP217
LOW level (near 0V) in normal status
HIGH level (5V) when the original point is detected
Y motor: TP215
LOW level (near 0V) in normal status
HIGH level (5V) when the upper limit is detected
Y motor: TP216
LOW level (near 0V) in normal status
HIGH level (5V) when the lower limit is detected
If the signal of the test point is changed, it is highly possible that the I/O PCB malfunctions.
If the signal is not changed, it is judged that the relevant original point sensor malfunctions.
Replace the relevant original point sensor.
If the instrument is still abnormal though the above-mentioned check is done, it is judged that the relevant motor malfunctions.
Replace the relevant motor.
If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or XY analog PCB.
Z alignment : Replace the Z sensor PCB.
39
R-SP3000P-1-0601-01
(2) Automatic alignment operates but the endothelium is not photographed. (The image cannot be changed to
the endothelium one or it is difficult to change.)
[CHECK POINT]
1 Start FACTORY MENU Program and perform SOLENOID CHECK to check if the image is
changed.
If the solenoid is not turned ON/OFF, check the connection of J410 on the XY analog PCB.
If there is no problem in the light path selector mechanism, it is judged that the solenoid unit malfunctions. Replace the solenoid unit.
If the solenoid is turned ON/OFF, check if the XY alignment or Z alignment is properly adjusted.
If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or the XY analog PCB.
Z alignment : Replace the Z sensor PCB.
40
R-SP3000P-1-0601-01
(3) The image is stored but flash is not done. (The photographed image is very dark.)
[CHECK POINT]
1 Start FACTORY MENU Program and perform XE CHECK to check if flashing is done.
If the xenon lamp does not flash, check the connection of J215 on the I/O PCB.
If flashing is not done though the xenon lamp is replaced, it is judged that the I/O PCB malfunctions.
Replace the I/O PCB.
If the xenon lamp flashes, replace the xenon lamp or check the optical system light path and check if the
xenon lamp is stained.
(4) The image is not stored and flash is not done. (Alignment still continues.)
[CHECK POINT]
1 Check if XY alignment or Z alignment is properly adjusted.
If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or the XY analog PCB.
Z alignment : Replace the Z sensor PCB.
41
R-SP3000P-1-0601-01
I/O PCB
J112
1
2
VIDEO OUT
DG
1
2
3
4
5
6
7
8
9
J113
1
2
3
4
5
SIGNAL
D-SUB 9PIN
CONTROL
DG
[CHECK POINT]
1 Is the VIDEO signal outputted?
2 Is D-SUB9PIN connected from J113 of the I/O PCB?
42
R-SP3000P-1-0601-01
Replace the mouse with a new one and check if it operates normally.
If the mouse does not operate normally, it is judged that the CPU PCB malfunctions.
Replace the CPU PCB.
43
R-SP3000P-1-0601-01
Unit number
41903 66300
Page
45
voltage
I/O PCB
CPU PCB
41903 72000
41903 71000
Replacement of EEPROM
57
52
56
58
59
sion
XY analog PCB
41903 74000
XY sensor PCB
41903 75100
Z sensor PCB
41903 75500
XY LED PCB
41903 76100
Z LED PCB
41903 76200
49
51
52
50
51
52
53
54
55
56
71
69
When replacing or adjusting the PCB's mentioned above, be sure to carry out the checks or adjustments in the
following order.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
R-SP3000P-1-0601-01
[Procedure]
Check the switching power supply output voltage so that the voltage at each unit can meet the
standard of the element being used during operation.
Make sure that the power supply voltage at each PCB is within the standard.
1
Make sure that the output voltage of the switching power supply is within the following
standard.
Output voltage
Standard
DC+24V
DC+285V
Make sure that all the cables of the electric parts are correctly connected. Then, make sure
that the output voltage at each PCB is within the standard.
Output voltage
Standard
PCB to be measured
Pin No.
DC+18V output
+18V10%
TP204
DC+15V output
+15V5%
TP207
DC+12V output
+12V0.5V
TP208
DC-15V output
-15V5%
TP210
DC+5.0V output
+5.0V0.5V
TP205
DC+3.3V output
+3.3V0.3V
TP209
DC+1.5V output
+1.5V0.1V
45
R-SP3000P-1-0601-01
[Procedure]
Store the initial value of each correction value before adjusting the electricity of SP-3000P.
Unless this setting is done, all the adjustments are adversely affected hereafter (for example, the
character display position is not proper on the monitor screen).
1
2
3
4
5
6
7
Start the "FACTORY MENU" program and make sure that the selection display appears on
the screen.
Set SW201 to the WP-OFF side on the I/O PCB to cancel WRITE-PROTECT. (LED208 is
turned on.)
Press the [Photography] switch. Select "A-1. EEPROM INITIALIZE" and press the [Image
transfer] switch.
The display as shown below appears on the monitor screen. Press the [Image transfer]
switch, and EEPROM is initialized.
When the initialization is finished normally, "EEPROM INITIALIZATION DONE" is displayed on the monitor screen.
When the initialization fails, "EEPROM INITIALIZATION FAILED" is displayed on the
monitor screen.
After the initialization is finished, press the [Image transfer] switch to return to the selection
display.
46
R-SP3000P-1-0601-01
Make the light intensity of the XY alignment LED uniform for each instrument.
1
2
3
Adjustment place
47
R-SP3000P-1-0601-01
48
R-SP3000P-1-0601-01
[Procedure]
Tune the synchronized frequency of the band-pass filter circuit in the XY alignment (PSD sensor) circuit in the LED modulated frequency (3.6kHz). Unless this adjustment is done correctly,
the filter characteristics are deteriorated by temperature change.
1
2
3
4
Adjustment place
X1 axis
VR6
TP5
X2 axis
VR5
TP4
Y1 axis
VR1
TP2
Y2 axis
VR2
TP1
49
R-SP3000P-1-0601-01
[Procedure]
Adjust the XY alignment light to be approximately at the center of the XY sensor. Unless this
adjustment is done correctly, the alignment detection range is not well-balanced.
1
2
3
4
5
6
Carry out this adjustment just after "3-4-6 Adjustment of the XY alignment circuit frequency" without changing the condition.
Connect the probes of the oscilloscope as shown below.
Observation probe (CH1) : TP5 (XY analog PCB)
(CH2) : TP4 (XY analog PCB)
Ground
: TP7 (XY analog PCB)
Set "CHOP MODE" for the oscilloscope. Move the XY sensor PCB in the X direction to
adjust its position so that the level of TP5 may be the same as TP4.
Connect the probes (CH1) and (CH2) of the oscilloscope to TP2 and TP1. Move the XY
sensor PCB in the Y direction to adjust its position so that the level of TP2 may be the same
as TP1.
Repeat the above 3 and 4 several times. Adjust the XY sensor PCB position until each
level for TP5 and TP4 and for TP2 and TP1 is within the standard.
Press the [Image transfer] switch to return to the selection display.
Standard
Adjustment place
XY sensor PCB
50
R-SP3000P-1-0601-01
[Procedure]
Adjust the circuit sensitivity so that the XY alignment can be detected correctly when the
patient's eye is aligned.
1
2
3
4
Carry out this adjustment just after "3-4-7 Adjustment of the XY alignment sensor position"
without changing the condition.
Select "GROUP-B XY SENSOR ADJUST" in the "FACTORY MENU" program.
The gain level values of the XY alignment circuit are displayed. Adjust VR8, VR7, VR4
and VR3 (XY analog PCB) so that the level values may be within the standard.
After adjustment, press the [Image transfer] switch to return to the selection display.
Standard
The displayed value on the
monitor:
1075
Adjustment place
X1 axis
VR8
X2 axis
VR7
Y1 axis
VR4
Y2 axis
VR3
51
R-SP3000P-1-0601-01
[Procedure]
Store the condition of the XY alignment circuit system in EEPROM so that the adjustment error
of the circuit system may be finally canceled. Unless this storing procedure is done correctly,
the position detection accuracy of the XY alignment cannot be obtained.
1
2
3
52
R-SP3000P-1-0601-01
Make the light intensity of the Z alignment LED uniform for each instrument.
1
2
3
Voltage: 0.3V0.2V
53
R-SP3000P-1-0601-01
[Procedure]
Adjust the Z alignment slit light to be approximately at the center of the Z sensor. Unless this
adjustment is done correctly, the Z alignment detection range is not well-balanced.
1
2
3
Carry out this adjustment just after "3-4-10 Adjustment of the Z LED luminance" without
changing the condition.
Remove the Z detection system tool B and set the Z slit tool.
Connect the probes of the oscilloscope as shown below.
Observation probe : Z VIDEO (Z sensor PCB)
Ground
: AG (Z sensor PCB)
The different-level reflected slit light is displayed on the monitor screen as illustrated. As
watching the oscilloscope, adjust the Z sensor PCB position in the Y direction so that the
light on the center of the shadowed section A may be received. The Z sensor PCB position
in the X direction should be the position where it touches the metal part.
When the light of the shadowed section A is received, the waveform on the oscilloscope
shifts rightward as shown below. At the middle position of this shift, fix the PCB.
54
R-SP3000P-1-0601-01
[Procedure]
Adjust the circuit sensitivity so that Z alignment can be detected correctly when the patient's eye
is aligned.
1
2
3
Carry out this adjustment just after "3-4-11 Adjustment of the Z alignment sensor position"
without changing the condition.
Remove the Z slit tool and set the Z projection system tool B.
Adjust VR1 (Z sensor PCB) so that the amplitude of the endothelium signal may be within
the standard on the oscilloscope.
Standard: 3.0V0.1V
3.0V0.1V
55
R-SP3000P-1-0601-01
[Procedure]
Store the condition of the Z alignment circuit system in EEPROM so that the adjustment error of
the circuit system may be finally canceled. Unless this storing procedure is done correctly, the
position detection accuracy of the Z alignment cannot be obtained.
1
2
3
4
Carry out this storing procedure just after "3-4-12 Adjustment of the Z alignment sensor circuit gain" without changing the condition.
Press the [Image transfer] switch, and the value for "COUNT" is increased.
Then, the endothelium reflection position, "N", and the optical magnification "M" on the
sensor are displayed as the numerical values.
If the value for "COUNT" is not increased, the tools are not set properly or the Z alignment
circuit system is not adjusted correctly.
After storing, press the [Image transfer] switch to return to the selection display.
COUNT UP
56
R-SP3000P-1-0601-01
[Procedure]
Adjust the charging voltage of the xenon tube lighting condenser. Unless this adjustment is done
correctly, the brightness of the photographed endothelium image is different between the instruments.
1
2
3
4
While the power is OFF, set VR201 of the I/O PCB to the medium position.
Set the tool base and the Z projection system tool B.
Connect the BNC video cable from the VIDEO OUT terminal to the oscilloscope. Set the
instrument so that you can observe one horizontal synchronization section.
Turn on the power. When the photography result display appears, observe the luminance
level of one horizontal synchronization section on the sample chart for 2000 through oscilloscope.
If the luminance level is out of the standard, adjust VR201 again, perform photography and
check the luminance level.
Repeat this procedure until the luminance level is within the standard. The VIDEO signal
level in "FLASH HIGH" should be a reference value.
The light intensity is increased by turning the volume clockwise.
Voltage
V
Screen on oscilloscope
Standard:
57
R-SP3000P-1-0601-01
[Procedure]
Store the correction value which is to be used for the final correction calculation in EEPROM so
that the absolute accuracy of the corneal thickness can be ensured.
1
2
3
Set the corneal thickness test eyes (three test eyes with different thickness) onto the chinrest
unit.
Select "GROUP-B PACHOMETRY ADJUST" in the "FACTORY MENU" program.
The corneal thickness of the test eye as a reference is displayed on the monitor screen as
shown below. Change the value with the [Chinrest up/down] switches and change the digit
with the selector switch to adjust finely in order to get the corneal thickness of the test eye to
be measured.
By using the [Image selector] switch, move to the approximate position of the test eye,
which is the target displayed on the screen. Measurement is performed ten times according
to automatic alignment.
Repeat the procedure of above 3 and 4 by using the three test eyes. The correction values
are displayed and they are automatically written in EEPROM. Then, press any switch to
return to the MENU display.
58
R-SP3000P-1-0601-01
[Procedure]
Store the mask slit inside dimension in EEPROM in consideration of the optical magnification
error so that the scale calibration, which is necessary for IMAGEnet cell analysis, can be done
correctly.
1
2
UPPER
LOWER
4
5
Select the "UPPER" and "LOWER" lines with the [Image selector] switch. By using the
[Chinrest up/down] switches, align "UPPER" to the upper third square and "LOWER" to the
lower third square with the black point of the test eye chart as the center. Click the [Image
selector] switch to store the interval of the six squares on the test eye chart.
Press the [Photography] switch, and the mask slit inside dimension, which is calculated in
consideration of the optical magnification error, is stored.
Press any switch on the control panel to return to the "FACTORY MENU" display.
59
R-SP3000P-1-0601-01
1
2
[MENU] switch
Line B
Fit the mouse pointer to the "DATE SET" tab and click it.
The "DATE SET" dialog box is displayed. Fit the mouse pointer to the "Y/M/D" icon,
which is the date order icon, and click it.
6
7
8
Fit the mouse pointer to the desired numeral on the "Ten key" icons and click it.
The entered numeral is displayed on the date display area.
Enter the numerals continuously until the desired date is displayed. Then, fit the mouse
pointer to the "Enter key" icon and click it. The date is set.
60
R-SP3000P-1-0601-01
1
2
3
4
Fit the mouse pointer to the "TIME SET" tab and click it.
The "TIME SET" dialog box is displayed. Fit the mouse pointer to the "AM/PM key" icon
and click it.
Fit the mouse pointer to the desired numeral on the "Ten key" icons and click it.
Enter the numerals continuously until the desired time is displayed. Then, fit the mouse
pointer to the "Enter key" icon and click it. The time is set.
61
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
Remove the diaphragm (41902 1122) and set the tool cross scale.
Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [MENU] switch on the control panel to erase "! scale" on the screen.
Illuminate from the patient's eye side.
Adjust the CCD camera so that the tool cross scale may be displayed at the center of the
monitor.
By moving the rear lens group cell unit back and forth, fit the focus to the first long scale
line from the center and then fix the unit.
Remove the tool cross scale and set the diaphragm (41902 1122).
Diaphragm
(41902 1122)
Tool cross scale
Illumination
CCD camera
surface
62
R-SP3000P-1-0601-01
1
2
3
Carry out this adjustment just after "3-4-19 Adjustment of the anterior eye segment rear lens
group" without changing the condition.
Set the tool base and the XY detection system tool.
By moving the front lens group cell unit (41902 1150) back and forth, make adjustment so
that XY bright spot may be circular and then fix the unit.
XY detection
system tool
63
R-SP3000P-1-0601-01
1
2
3
Carry out this adjustment just after "3-4-20 Adjustment of the anterior eye segment front
lens group" without changing the condition.
Press the [MENU] switch on the control panel to display "! scale".
By moving the CCD camera, make adjustment so that the XY bright spot may be within the
"!" (the displayed mark for adjustment) and then fix the camera.
* Be careful not to set the CCD camera slantingly. (If it is set slantingly, the observation/
photography slits are tilted.)
Press the [Photography] switch to return to the "FACTORY MENU" program display.
Monitor screen
The displayed mark
for adjustment
XY detection
system tool
64
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
8
9
10
11
12
Remove the slit holder (41902 1307) from the mask unit sub.
Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
Illuminate from the patient's eye side.
By moving the relay lens unit (41902 1380) back and forth, adjust the focus temporarily so
that the mask image can be seen.
Move the relay lens unit by using the pulling screw and pushing screw on the side of the
endothelium reflection mirror unit so that the mask may be displayed at the center of the
instrument monitor.
Turn the mask lens unit (41902 1350) so that the mask slit may be perpendicular on the
instrument monitor screen.
Press the [MENU] switch on the control panel, and the memory image appears on the instrument monitor screen.
Move the relay lens unit a little by using the pulling screw and pushing screw on the side of
the endothelium reflection mirror unit to make adjustment so that the memory image may be
positioned at the center.
By moving the relay lens unit back and forth, make adjustment to set the mask slit focus
equally in right and left.
Repeat the above 9 to 10 to adjust the position and focus of the mask.
Press the [Photography] switch to return to the "FACTORY MENU" program display.
Slit holder
Endothelium reflection
mirror unit
65
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
8
9
10
11
Focus position
Z detection system
tool B
66
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
Slit image
Memory image
Z detection system
tool A
Adjustment screw
Mirror holder unit (41902 1220)
67
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
8
9
Carry out this adjustment just after "3-4-24 Adjustment of the observation slit" without
changing the condition.
Remove the reflection mirror unit and the xenon lamp unit. Illuminate with the halogen
light source device equipped with light guide from the xenon lamp position.
Press the [DELETE] switch on the control panel to turn off the Z LED.
Move the photography slit right and left to be positioned at the center of the instrument monitor screen.
By moving the slider back and forth, adjust the slit focus correctly.
Press the [MENU] switch on the control panel, and the memory image appears on the instrument monitor screen.
Make adjustment so that the photography slit may be positioned at the center of the memory
image.
Press the [Photography] switch to return to the "FACTORY MENU" program display.
Set the reflection mirror unit and the xenon lamp unit.
Memory image
Photography slit
(By moving back and forth, adjust focus.
By moving right and left, adjust position.)
Slit image
LED holder
Tool
(Z test eye A)
Visible-light
illumination
Slider
68
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
Z projection
system tool A
69
R-SP3000P-1-0601-01
1
2
3
4
5
6
7
8
9
10
Slit holder
Bracket
Eccentric adjustment lens sub
Analyzed image
Z detection
system tool B
70
R-SP3000P-1-0601-01
1
2
3
4
XY LED PCB
XY bright spot
71
R-SP3000P-1-0601-01
4. TOOL LIST
4-1 Special Repair Tools
Tool No.
Usage
rank
Name
Illustration
Application
SP-2000P
-01
Tool base
-02
XY detection system
tool
-03
Z projection system
B tool
-04
Z slit tool
-05
-06
-07
-08
Z detection system
A tool
Z projection system
A tool
Z detection system
B tool
projection slit
72
R-SP3000P-1-0601-01
Tool No.
Usage
rank
SP-3000P
-01
-02
Name
Illustration
Application
Corneal thickness
test eye
0.4, 0.5, 0.6, 0.7,
0.78mm
Corneal thickness
test eye and
attachment
0.4, 0.6, 0.78mm
73
R-SP3000P-1-0601-01