REPAIR MANUAL

ISSUED............... OCT. 2005


REVISED ............ JAN. 2006

SPECULAR MICROSCOPE

SP-3000P

TOPCON CORPORATION
R-SP3000P-1-0601-01

CONTENTS
Page

1. INTRODUCTION ..................................................................................................................................1
1-1
1-2
1-3
1-4
1-5

Features of Product ..................................................................................................................................1


Specifications and Performance of Product.............................................................................................1
Target User of This Manual.....................................................................................................................1
How to Use This Manual .........................................................................................................................2
Cautions ...................................................................................................................................................2

2. PRODUCT OUTLINE .........................................................................................................................3


2-1
2-2
2-3
2-4
2-5

Nomenclature...........................................................................................................................................3
Optical System (Arrangement) ................................................................................................................5
Electric System ........................................................................................................................................6
Principles .................................................................................................................................................8
PCB Arrangement..................................................................................................................................10

3. REPAIR WORK ..................................................................................................................................16


3-1
3-2
3-3
3-4

FACTORY MENU Program .................................................................................................................16


Disassembly and Assembly Procedures ................................................................................................24
Troubleshooting .....................................................................................................................................34
Assembly and Adjustment Procedures ..................................................................................................44

4. TOOL LIST............................................................................................................................................72
4-1 Special Repair Tools..............................................................................................................................72
4-2 General Repair Tools .............................................................................................................................73

R-SP3000P-1-0601-01

1. INTRODUCTION
1-1 Features of Product
(1) Photography of the corneal endothelium and the measurement of the corneal thickness can be performed
at the same time.
(2) The auto-alignment function ensures quick and easy photography and measurement.
(3) Simplified cell analysis enables the user to easily understand the rough value of the cell density.

1-2 Specifications and Performance of Product


Field of coverage

0.25 0.5mm

Photography/measuring point

Corneal center and 4 places in the peripheral area

Photography magnification

165 on the instrument monitor

Corneal thickness measurement repeatability

0.01mm

Alignment

To be displayed on the monitor screen.

Monitor

5.6 type (color character display)

Photography mode

AUTO/MANUAL/SEMI-AUTO switching

Flash level

HIGH/LOW switching

Photography/measurement result display

To be displayed on the monitor screen.

Image memory

5 displays for each of right and left eyes

Working distance

25mm

Power saving function

Power save method

External output terminal

VIDEO OUT, CONTROL, RS-232C OUT and USB2.0

External input terminal

RS-232C IN

Operating temperature

10 ~ 40C

Instrument longitudinal mobility

40mm

Instrument lateral mobility

92mm

Measuring head vertical mobility

28mm

Chinrest height adjustment quantity

60mm

Source voltage

AC100-240V, 50Hz/60Hz

Power input

190VA

Dimensions

274mm (W) 485mm (D) 410mm (H)

Weight

22kg

* Subject to change in specifications and appearance without advance notice for future improvement.

1-3 Target User of This Manual


SP-3000P is an optical precision inspection instrument to be used for ophthalmology and internal medicine.
Only a specialist for assembly and adjustment of medical machines can repair this instrument.
For others except the specialist, it is recommended to take the TOPCON repair technique training course.

R-SP3000P-1-0601-01

1-4 How to Use This Manual


When inspecting, adjusting or repairing the SP-3000P microscope for any problem, use this manual according
to the flow chart below.
Manual
Operation procedure

Initial inspection

Applicable item

Page

FACTORY MENU program

P. 16 ~ 23

Disassembly and assembly


procedures

P. 24 ~ 33

Troubleshooting

P. 34 ~ 43

Trouble

Inspection

OK

NO
Trouble location
check

Searching for the


cause

Repair operations

Inspection

NO

Disassembly and assembly


procedures
Assembly and adjustment
procedures
FACTORY MENU program

P. 24 ~ 33
P. 44 ~ 71
P. 16 ~ 23

Normal

YES
End

Photography principles

SP-3000P basic materials

P. 8 ~ 9

1-5 Cautions

Refer to the illustrations of the parts list for the disassembly and assembly of the unit which is not
described in this manual. Refer to the instruction manual for replacing the consumables.

If the optical system is soiled when repairing, photographing is adversely affected. Be careful not to touch
or soil the optical system when disassembling, assembling or adjusting.

Order the repair parts by using the service parts list.

R-SP3000P-1-0601-01

2. PRODUCT OUTLINE
2-1 Nomenclature
1

4
6

No.

Name

Function

Lamp house cover

Remove it when replacing the xenon lamp.

Color video monitor

On this monitor, the aligned anterior eye segment image, the aligned
corneal endothelium image and the photography result are displayed.

Photography switch

Press this switch after alignment while observing the corneal endothelium, and it is possible to take a picture. Press this switch when the
memory pictures are displayed, and the memory pictures will be
switched by turns. This switch is also available for resetting when the
power saving function is working.

Joystick

Incline this, and alignment in the longitudinal and lateral directions is


possible. Turn this, and alignment in the vertical direction is possible.

Locking knob

This is used to fix the instrument in transportation.

RS-232C input terminal

This is used to connect the instrument to a bar code reader, etc.

VIDEO OUT output terminal

This is used to connect the instrument to an image processing unit


adopting the EIA (NTSC) signal system.

USB output terminal

This is used to connect the instrument to a personal computer (PC), etc.

RS-232C output terminal

This is used to connect the instrument to a personal computer (PC), etc.

R-SP3000P-1-0601-01

10

11

12
13
16
17

15

No.

14

Name

10

Forehead rest

11

Photography window

12

Vertical positioning mark

13

Chinrest

14

Power cord

15

Fuse holder

16

Mouse

17

Power switch

Function

This is standard for adjusting the height of the patient's eye.


.
This is used for the simplified cell analysis, the fixation target switching, etc.

R-SP3000P-1-0601-01

2-2 Optical System (Arrangement)

No.

Name

Function

XY LED

Confirms the alignment in the XY direction on the sensor.

Center fixation LED

Fixes the patient's eye as a target.

XY alignment sensor

Detects the alignment range in the XY direction.

Z LED

Used as a light source for observation.

Observation slit

Fixes the width of the observed image.

Photographing slit

Fixes the width of the photographed image.

Xenon

Used as a light source for photography.

Photography system condenser lens

Forms the endothelium image on the mask position.

Z alignment sensor

Detects the alignment range in the Z direction.

10

Mask lens

The mask is displayed on the monitor.

11

Knife edge

Shields the clear reflected image.

12

Photography system relay lens

Forms the endothelium image on the CCD camera.

13

CCD camera

Displays an image on the monitor.

14

Anterior eye segment illumination LED

Illuminates the patient's eye.

15

Peripheral fixation LED

Fixes the patient's eye.

R-SP3000P-1-0601-01

POWER CORD IN JAPAN


DC POWER CABLE UNIT

SWITCHING POWER SUPPLY


CHINREST
CABLE UNIT

CONTROL SW
PCB UNIT

CHINREST CONTROL
PCB UNIT

CPU-I/O
CABLE 2 UNIT

DISCHARGE RESISTANCE UNIT

CPU PCB UNIT

CONDENSER UNIT

LCD YC CABLE
STEPPING MOTOR
FOR Z AXIS

XENON UNIT

Z AXIS ZERO DETECTION SENSOR UNIT

Y AXIS LOWER LIMIT DETECTION


SENSOR UNIT

Y AXIS UPPER LIMIT DETECTION


SENSOR UNIT

X AXIS ZERO DETECTION SENSOR UNIT

Z AXIS MOTOR
CABLE UNIT

Y AXIS MOTOR UNIT


FOR Y AXIS

STEPPING MOTOR
FOR X AXIS

XY ANALOG PCB UNIT

PD SENSOR UNIT

XY SENSOR CABLE 1 UNIT


X AXIS MOTOR
CABLE UNIT

XY SENSOR CABLE 2 UNIT

CHARGING CONTROL CABLE UNIT

XE POWER CABLE UNIT

USB I/F
CABLE UNIT

USB MOUSE
CABLE UNIT

RS-232C OUT
CABLE UNIT

RS-232C IN
CABLE UNIT

EXTERNAL SYSTEM
CABLE UNIT

VIDEO OUT
CABLE UNIT

R/L SENSOR UNIT

LAMP HOUSE COVER


SENSOR UNIT

SOLENOID UNIT

ANTERIOR EYE SEGMENT


ILLUMINATION LED PCB UNIT

LATERAL FIXATION LED


PCB UNIT

VERTICAL FIXATION LED


PCB UNIT

CENTER FIXATION LED


PCB UNIT

Z LED
PCB UNIT

XY LED
PCB UNIT

2-3 Electric System

2-3-1 Wiring diagram

CAMERA DATA CABLE UNIT

CAMERA POWER CABLE UNIT

XY SENSOR
PCB UNIT

XENON CABLE UNIT

Z SENSOR CABLE 2 UNIT

Z SENSOR
PCB UNIT
LCD POWER CABLE

Z SENSOR CABLE 1 UNIT

I/O PCB UNIT

CONTROL SW CABLE UNIT

CPU-I/O
CABLE 1 UNIT

POWER SUPPLY UNIT

EARTH CABLE UNIT

3P SW CABLE UNIT

6
R-SP3000P-1-0601-01

2-3-2 Block diagram

Photography head unit


LED control unit

Xenon tube

Sensor unit
CCD camera

Xenon flashing control unit

Solenoid

Base unit

Automatic drive motor


LCD monitor
Interface control unit

Chinrest unit
Chinrest
motor

Power unit
Video OUT connector
Commercial
power supply

Switching power supply

RS-232C IN connector

USB I/F connector

RS-232C OUT connector

USB mouse connector

R-SP3000P-1-0601-01

2-4 Principles
2-4-1 Photography principles
Slit light which enters the cornea gets through it and reaches the aqueous humor. Part of the light is reflected
on the front of the cornea in the boundary whose refraction index is different.
When the light reflected on the backside of the cornea is magnified, it can be observed as an image, because
the cells on the backside of the cornea are uneven.

Cornea
Photography
system (P)
22.5
22.5

R'
R

Illumination
system (I)

S'
S

Arrange the illumination system (I) and the photography system (P) so that the normal line which is perpendicular to the cornea may divide into I and P equally. Consequently the light is reflected on the backside of
the cornea. Move the instrument while keeping the angles of I and P uniform so that the normal line may
divide into I and P equally. So the front reflection of the cornea (S'-S) and the backside reflection of the cornea (endothelium cells R'-R) can be observed.

2-4-2 Corneal thickness measurement principles


Slit light is slantingly emitted for the patient's cornea and the light reflected by the front and backside of the
cornea is brought in by the line sensor.
The corneal thickness is calculated by using the interval (L) between the front reflection image and backside
reflection image on the line sensor.
Backside of cornea

Front of cornea

Backside reflection

L
Reflected
light
Front reflection
Line sensor
Slit light
Corneal
thickness

R-SP3000P-1-0601-01

2-4-3 Photographed position detection function


XY detection function: This function focuses the spot XY LED ray reflected on the cornea on XY sensor to
detect the XY shift and direction.
Z detection function : Slit light as illumination light is brought into the cornea. As mentioned above, when
the spot XY LED ray is brought into the XY sensor, the light is turned on and the
front reflection (S'-S) and the backside reflection (endothelium cells R'-R) are done.
Electrically the Z sensor detects the position of the front reflection.
Automatic alignment is done according to the detection results of the XY detection
and Z detection. When the XY alignment and Z alignment are in the OK range, photographing is done automatically.

R-SP3000P-1-0601-01

2-5 PCB Arrangement

Z SENSOR PCB
J502

Z SENSOR PCB
J501

XY ANALOG PCB
J401

CCD CAMERA DATA

2-5-1 4190371000 CPU PCB

VIDEO OUT

EXTERNAL
SYSTEM
D-sub
CONTROL SW
PCB
J702

IO PCB
J204

IO PCB
J203

USB MOUSE

10

R/L SENSOR

RS-232C IN
Din

RS-232C OUT
D-sub

USB I/F

R-SP3000P-1-0601-01

X AXIS MOTOR

Z AXIS MOTOR

Y AXIS MOTOR

2-5-2 4190372000 I/O PCB

X AXIS ZERO
DETECTION
SENSOR
Z AXIS ZERO
DETECTION
SENSOR
CCD CAMERA
POWER SUPPLY
LCD POWER
SUPPLY
PD SENSOR

DETECTION

CHINREST
CNTROL PCB
J601

Y AXIS
LOWER LIMIT
DETECTION
SENSOR

SWITCHING
POWER SUPPLY
CN6

Y AXIS
UPPER LIMIT
SENSOR

SWITCHING
POWER SUPPLY
CN3
CPU PCB
J102

SWITCHING
POWER SUPPLY
CN4

11

CONDENSER UNIT

XENON UNIT

CHARGE RESISTANCE UNIT

CPU PCB
J101

R-SP3000P-1-0601-01

XY SENSOR PCB
J701

2-5-3 4190374000 XY ANALOG PCB

SOLENOID UNIT
XY LED PCB
Z LED PCB
CENTER FIXATION LED
PCB
VERTICAL FIXATION LED
PCB
LATERAL FIXATION LED
PCB
ANTERIOR EYE SEGMENT
ILLUMINATION LED PCB
LAMP HOUSE COVER
SENSOR
CPU PCB
J105

12

R-SP3000P-1-0601-01

2-5-4 4190375100 XY SENSOR PCB

XY ANALOG PCB
J402

CPU PCB
J107

CPU PCB
J106

2-5-5 4190375500 Z SENSOR PCB

13

R-SP3000P-1-0601-01

2-5-6 4190366000 SWITCHING POWER SUPPLY

I/O PCB
J202

I/O PCB
J218
3P SWITCH

I/O PCB
J201

CHINREST DC
MOTOR

I/O PCB
J216

2-5-7 4190368000 CHINREST CONTROL PCB

14

R-SP3000P-1-0601-01

CPU PCB
J110

2-5-8 4190381000 CONTROL SW PCB

JOYSTICK SW UNIT

15

R-SP3000P-1-0601-01

3. REPAIR WORK
3-1 FACTORY MENU Program
Carry out the following check to find a trouble early and to adjust the instrument.
Program name

Contents

FACTORY MENU GROUP-A


1. E2PROM INITIALIZE

Writes the initial values for the correction values in EEPORM.

2. CCD CENTER ADJUST

Stores the reference position to perform XY alignment with CCD in


EEPROM.

3. ANTERIOR LED SETTING

-----

4. FLASH ROM CHECK

Performs the memory check for FLASH ROM. (Sum check)

5. FIXATION LED TEST

Checks if the lighting position displayed on the screen is fit to the


LED mounting position.

6. XY LED ADJUST

Adjusts the XY LED luminance level.


Within the luminance level value range: OK is displayed.
Out of the luminance level value range: NG is displayed.

7. IMAGEnet SLIT MEMORY

Sets a test eye and stores the mask slit inside dimension.
Initial value: 0.1985

FACTORY MENU GROUP-B


1. XY SENSOR ADJUST

Sets a test eye and checks the XY alignment circuit.


Within the range: OK is displayed. Circle is displayed.
Out of the range : NG is displayed. Circle is displayed.
Nothing
: NOTHING is displayed. Circle disappears.

2. Z SENSOR ADJUSTMENT

Sets a test and checks the Z alignment circuit.


Within the range: OK is displayed. Bar is displayed.
Out of the range : NG is displayed. Bar is displayed.
Nothing
: NOTHING is displayed. Bar disappears.

3. PACHOMETRY ADJUST

Sets the correction value of the corneal thickness.


Setting the correction value is successful : OK is displayed.
Setting the correction value is failure
: NG is displayed.

4. OPTICAL ADJUST

The displayed data used for optical adjustment

5. OPTICAL CHECK

The displayed data used for optical function check

FACTORY MENU GROUP-C


1. E2PROM CHECK

Performs the memory check for E2PROM.

2. SWITCH CHECK

Checks the switches.


ON or OFF is displayed.

3. LED CHECK

Checks each LED by blinking.


Each LED blinks every 2 seconds.

4. XENON CHECK

Checks the flashing of the xenon lamp.


The xenon lamp flashes every 4 seconds.

5. POWER OFF CHECK

Checks the automatic power off function.


Power ON: 2 seconds/Power OFF: 2 seconds

16

R-SP3000P-1-0601-01

6. MOTOR P.I. CHECK

Checks the operation of the X, Y and Z motors and the photo interrupter.
X drive - Reciprocation : OK or NG is displayed.
Y drive - Reciprocation : OK or NG" is displayed.
Z drive - Reciprocation : OK or "NG" is displayed.

7. SOLENOID CHECK

Checks the operation of solenoid.


Solenoid is ON/OFF repeatedly at intervals of 2 seconds.

8. XENON COVER CHECK

Checks the charge and discharge of the condenser under the cover
IN/OUT condition.
Checks the operation of the lamp house cover photo interrupter.
The following data are displayed by opening/closing the lamp house
cover.
COVER
: IN/OUT OK or NG is displayed.
CHARGE ON : OK or NG is displayed.
CHARGE OFF : OK or NG is displayed.

9. AGING TEST

Used for the aging test.


Xenon lamp flashes. Motors operate. Solenoid operates.
Each LED is ON. This sequence is ON/OFF repeatedly in a cycle
of 3 minutes.

10. RS232C CHECK

Checks data sending/receiving of RS-232C.


OK or FAIL is displayed.
Cable connection is not correct: UNCONNECTED is displayed.

11. USB CHECK

Checks data sending of USB.

FACTORY MENU GROUP-D


1. PROGRAM DOWNLOAD

Performs download of the SP-3000P software and writes it on the


flash ROM installed in the CPU PCB.

2. ID INPUT

Data cannot be transferred when the ID input is not done.


ON : The ID input is required.
OFF : Data can be transferred without the ID input.

3. ADVANCED MORPHOLOGY

It is OFF in normal condition. This is a factory mode (changing the


cell analysis mode).

4. MACHINE PARAM

This is a factory mode. (This is the function to write the correction


values for a specified instrument forcedly.)

17

R-SP3000P-1-0601-01

[FACTORY MENU program start method]


(1) Turn on the power switch while pressing the [Delete] and [Clear] switches at the same time.

Clear switch

Delete switch

(2) The following display appears on the screen. Press the [Image transfer] switch.

Image transfer switch

(3) Press the [Photography] switch. Move the cursor to the item to be checked and press the [Image transfer]
switch.

The item list is divided into four displays. Use the [Image selector] switch to move to the next display.
(4) On each checking display, press the [Clear] switch to return to the item list display.
(5) Turn off the power switch to end FACTORY MENU.

18

R-SP3000P-1-0601-01

Caution:If the programs of "FACTORY MENU GROUP-A/B" are executed without installing the special tools, the memory data may be changed. Be careful.
This chapter describes the programs 1 ~ 12 of "FACTORY MENU GROUP-C" which are used to check the
functions. Refer to the items of "3-4 Assembly and Adjustment Procedures" for the "FACTORY MENU" programs which are used for adjustment.

Operation of "FACTORY MENU GROUP-B [OPTICAL ADJUST]"

Clear switch

Delete switch

Menu switch
Image transfer switch

Delete switch
Menu switch

Image selector switch

: Turns on/off the XY LED and Z LED.


: Changes the displayed mask image.

Mask #1
The alignment frame is displayed at the center.

Mask #2

Mask #3
The perpendicular line is displayed at the center.

Mask #4
This is the display for correcting the slit inside dimension.

19

R-SP3000P-1-0601-01

Clear switch
: Turns on/off the mask image display.
Image transfer switch : Lights the xenon and takes a picture only when observing the endothelium.
Image selector switch: Opens and closes the solenoid.
The following table shows the relation between the light sources in this case.
Anterior eye segment observation

Endothelium observation

XY LED

ON

OFF

Z LED

OFF

ON

20

R-SP3000P-1-0601-01

Check of FACTORY MENU GROUP-C


Check program

Display on monitor

Check for trouble

E2PROM CHECK

When "NG" is displayed:

Make sure that "OK" is displayed.

Replace the CPU PCB.

SWITCH CHECK

Check the control switch

Make sure that "" is displayed for the pressed switch.

PCB.

Check the joystick switch.


Check the cable between
the control switch PCB and
the I/O PCB.

LED CHECK

XY LED

Check the wiring of each

(1) Align the tool test eye.

LED.
(2) Make sure that each LED
blinks every 2 seconds on
the monitor.

Check the XY analog PCB.


Check the CPU PCB.

(3) Observe the fixation LED


from the patient side.

Anterior eye segment LED

21

R-SP3000P-1-0601-01

Check program

Display on monitor

Check for trouble

XENON CHECK

The xenon lamp has burned

Make sure that the xenon


lamp lights up every 4 seconds.

out.

POWER OFF CHECK

Replace the I/O PCB.

Make sure that "power ON"


(2 seconds) and "power OFF"
(2 seconds) are repeated.

MOTOR P.I. CHECK

Check each of the photo

Make sure that "OK" is displayed.

interrupters and motors.

Check the I/O PCB.

SOLENOID CHECK

Check the solenoid unit.

Make sure that the solenoid


repeats its operation every 2
seconds.

Check the XY analog PCB.


Check the CPU PCB.

22

R-SP3000P-1-0601-01

Check program

Display on monitor

Check for trouble

XENON COVER CHECK

When "IN" and "OUT" are

(1) When there is nothing


near the photo interrupter,
make sure that the following messages are displayed on the monitor
screen:
XENON COVER : OUT
CHARGE ON : NG

not switched:
Check the photo interrupter.
Check the CPU PCB.
Check the I/O PCB.

When "NG" is displayed:


Check the I/O PCB.

(2) When there is a white


paper near the photo interrupter, make sure that the
following messages are
displayed on the monitor
screen:
XENON COVER : IN
CHARGE ON : OK

RS232C CHECK

Check the CPU PCB.

(1) Connect the RS-232C


cable to the output terminal and the input terminal.
(2) Make sure that "OK" is
displayed.

RS-232C output
terminal

RS-232C cable

RS-232C input
terminal

23

R-SP3000P-1-0601-01

3-2 Disassembly and Assembly Procedures


3-2-1 Removing the external covers

3
9

7
6

10
12

Illustration

11

13
9

Proc.
1

10 14
14

Disassembly item
Removing
cover 2
Removing
right cover

the

the

front

head

12

Removal method

Points for assembly

Remove the two screws


Remove the front cover

.
.

Remove the cap 4 .


Remove the two screws

.
.

Removing the head left


cover 6

Remove the four screws

Removing the monitor


cover 8

Remove the cap 9 .


Remove the two screws

10 .

Removing the frame


right cover 11

Remove the two screws

12 .

Removing the frame


left cover 13

Remove the two screws

14 .

24

Perform the reversed procedure of disassembly.

R-SP3000P-1-0601-01

3-2-2 Removing the color video monitor

Illustration

Proc.
1

Disassembly item
Removing the
video monitor 2

color

Removal method

Points for assembly

Loosen the four screws 1 .


Slide the color video monitor 2
upward and remove it toward you.
Remove all the cables from the color
video monitor.

Perform the reversed procedure of disassembly.

25

R-SP3000P-1-0601-01

3-2-3 Removing the PCB's

5
4

Illustration

3
Proc.
1

Disassembly item
Removing
PCB 1

the

CPU

Removing the I/O PCB


4

Removal method
Remove the spacer 2 .
Remove the two screws

Remove the spacer 5 .


Remove the two screws

26

Points for assembly


Perform the reversed procedure of disassembly.

R-SP3000P-1-0601-01

Illustration

3-2-4 Removing the control switch PCB

1
3

4
Proc.

Disassembly item

Removal method

Points for assembly


Perform the reversed procedure of disassembly.

Removing
cover 1

base

Remove the two screws 2 .


Lift up the base cover 1 and remove
it.

Removing the control


switch PCB 3

Remove the two screws 4 .


Remove the connector and then
remove the control switch PCB 3 .

the

27

R-SP3000P-1-0601-01

3-2-5 Removing the measurement unit PCB

Illustration

10
9

Proc.

Disassembly item

Removal method

Removing the frames


1 , 2 and 3

Remove the two screws


frame.
Remove each frame.

Removing the XY analog PCB 5

Remove the connectors at 10 places.


Remove the three screws 6 .

Removing the XY sensor PCB 7

Remove the connector at 1 place.


Remove the four screws 8 .

Removing the Z sensor


PCB 9

Remove the connectors at 2 places.


Remove the four screws 10 .

28

Points for assembly


4

for each

Perform the reversed procedure of disassembly.

It is necessary to adjust
each PCB.

R-SP3000P-1-0601-01

3-2-6 Removing the CCD camera unit

Illustration

Proc.

Disassembly item

Removing the frame

Removing
camera 3

the

Removal method
1

CCD

Points for assembly

Remove the two screws 2 .


Remove the frame 1 toward you.

Perform the reversed procedure of disassembly.

Remove the two screws 4 .


Remove the connector and then
remove the CCD camera unit 3
toward you.

It is necessary to adjust
the CCD camera position.

29

R-SP3000P-1-0601-01

3-2-7 Removing the X/Z original point sensor

5
Illustration

Proc.

Disassembly item

Removal method

Points for assembly

Remove the four screws

Removing the X original point sensor 3

Remove the two screws

Removing the Z original point sensor 5

Remove the two screws


Remove the two screws

.
.

Removing the plate

30

Perform the reversed procedure of disassembly.

R-SP3000P-1-0601-01

3-2-8 Removing each LED PCB unit 1

1
9

10

9
Illustration

8
7
10

Proc.

Disassembly item

Removal method

Removing the center


fixation LED PCB 1

Remove the connector J405.


Remove the two screws 2 .

Perform the reversed procedure of disassembly.

Removing the XY LED


PCB 3

Remove the connector J403.


Remove the two screws 4 .

It is necessary to adjust the


position and electricity.

Removing the Z LED


PCB 5

Remove the connector J404.


Remove the two screws 2 .

It is necessary to adjust the


position and electricity.

Removing the anterior


eye segment illumination LED PCB 7

Remove the connector J408.


Remove the two screws 8 for each
fixing plate. Then, remove each fixing
plate 9 .
Remove the two screws 10 at each
place.

31

Points for assembly

R-SP3000P-1-0601-01

3-2-8 Removing each LED PCB unit 2

Illustration

1
2

Proc.

Disassembly item

Removing the vertical


fixation LED PCB 1

Removing the horizontal fixation LED PCB

Remove the connector J407.


Remove the two screws 2 at each
place.

Removal method
Remove the connector J406.
Remove the two screws 4 .
Remove the two screws 6 .
Remove the two screws 2 at each
place.

32

Points for assembly


Perform the reversed procedure of disassembly.

R-SP3000P-1-0601-01

3-2-9 Removing the switching power supply

Illustration

7
4

6
3

1
2

Proc.

Disassembly item

Removal method

Points for assembly

Removing the adjusters

Turn the two adjusters 1 at the LCD


side counterclockwise and remove
them.

Perform the reversed procedure of disassembly.

Removing the locking


knob unit 2

Remove the four screws 3 .


Remove the locking knob unit.

Removing the bottom


plate 6

Loosen the four screws 4 and move


each unit inward.
Remove the six screws 5 .
Pull out the bottom plate 6 slowly.
Note: Dont pull out the bottom plate
6 forcedly because the cables
are connected to it.

Removing the switching power supply 7

Remove the connectors at 3 places.


Remove the four screws 8 .

33

Be careful not to catch


the cables.

R-SP3000P-1-0601-01

3-3 Troubleshooting
3-3-1 Troubles related with the color video monitor
(1) No image is displayed.
(2) No character is displayed.
3-3-2 Error messages are displayed on the screen.
(1) XENON OVER
(2) XENON POWER PCB ERROR
(3) CABLE UNCONNECTED
(4) RS232C FAIL
(5) USB FAIL
(6) CHANGE BATTERY
(7) COVER OUT
3-3-3 Troubles at alignment
(1) Automatic alignment does not operate.
(2) Automatic alignment operates but the endothelium is not photographed. (The image cannot be changed to
the endothelium one or it is difficult to change.)
(3) The image is stored but flash is not done. (The photographed image is very dark.)
(4) The image is not stored and flash is not done. (Alignment still continues.)
3-3-4 After photographing
(1) The image cannot be transferred for IMAGEnet.
(2) Data cannot be inputted or outputted by RS-232C.
(3) Data cannot be outputted by USB.
(4) Zooming and simplified cell analysis cannot be done.

34

R-SP3000P-1-0601-01

3-3-1 Troubles related with the color video monitor

J103

CPU PCB

LCD

J101102

J104

VOLUME
PCB

J203J204

DATA

I/O PCB
J206

+12V
J205

+12V

CCD camera

Image signal

CCD camera
A/D conversion

DATA

CPU PCB
A/D conversion

35

Image
signal

R-SP3000P-1-0601-01

(1) No image is displayed.


[CHECK POINT]
1 Adjust "BRIGHT" of LCD.
2 Check the power supply (+12V).
Power supply for LCD (+12V)
I/O PCB: J205 No. 1, 2 +12V
No. 3, 4 GND
Power supply for CCD camera (+12V) I/O PCB: J206 No. 1
+12V
No. 2 GND
3 Check the cable between the CPU PCB and the CCD camera and the cable between the CPU PCB and
the LCD.
4 When the instrument operates:
The probable cause is the LCD, CCD camera or CPU PCB.
It is highly possible that the LCD malfunctions.
5 When the photographed image is not displayed:
The probable cause is the CPU PCB or CCD camera.
(2) No character is displayed.
[CHECK POINT]
1 Adjust the brightness of characters.
2 The CPU PCB malfunctions.

36

R-SP3000P-1-0601-01

3-3-2 Error messages are displayed on the screen.


(1) The following message is displayed.
XENON OVER
[CHECK POINT]
1 Conditions of the message:
Charge is stopped and discharge is forcedly done because the charge voltage is excessive.
2 Replace the I/O PCB if the message is still displayed though the charge voltage adjusting VR (I/O
PCB: VR201) is readjusted.
(2) The following message is displayed.
XENON POWER PCB ERROR
[CHECK POINT]
1 Conditions of the message: The xenon charging condenser is not charged.
2 Check if the connectors J202 and J218 are plugged in.
3 If the above 2 is OK, replace the I/O PCB.
(3) The following message is displayed.
CABLE UNCONNECTED
[CHECK POINT]
1 Conditions of the message:
When data is outputted in the RS-232C or USB data output allowable mode and when the cable is not
connected.
2 Check if the RS-232C output terminal or the USB output terminal is properly connected to a personal
computer (PC).
3 Check if the connectors at the DATA OUT side (J115 and J117) are plugged in.
(4) RS232C FAIL is displayed.
[CHECK POINT]
1 Conditions of the message:
When data is inputted or outputted in the RS-232C data input or output allowable mode, the communication protocol is not good.
2 Check if the communication specification of PC or the data input device is set correctly.
3 Check if the data input/output terminals are connected correctly.
(5) The following message is displayed.
USB FAIL
[CHECK POINT]
1 Conditions of the messages:
When data is not outputted well in the USB data output allowable mode.
2 Start the application program at the PC side. Check if communication is possible by restarting.
3 If the communication is still NG, restart the instrument and PC to check the operation.
(6) The following message is displayed.
CHANGE BATTERY
[CHECK POINT]
1 Conditions of the message: The clock IC battery is dead.
2 Replace the battery of the clock IC on the I/O PCB.

37

R-SP3000P-1-0601-01

(7) The following message is displayed.


COVER OUT
[CHECK POINT]
1 Conditions of the message: The lamp house cover is removed with the power ON.
2 Check if the lamp house cover is set correctly.
(Set a white paper near the photo interrupter and check if it is detected.)
3 Check the connectors and cables.
4 Check the J410 3-pin signal.
COVER is set...........L (0V)
COVER is not set.....H (5V)
5 Check the connector J401 and cable on the XY analog PCB and the connector J105 and cable on the
CPU PCB.
6 If the above 4 and 5 are OK, replace the CPU PCB.
* When this error message is displayed, the xenon condenser is being discharged.

38

R-SP3000P-1-0601-01

3-3-3 Troubles at alignment


(1) Automatic alignment does not operate.
[CHECK POINT]
1 Start FACTORY MENU Program and check the operation of the X, Y and Z motors in MOTOR
P.I. CHECK.
When NG is displayed, check the connection of the relevant motor and the wiring of the relevant
original point sensor.

Check if the mechanism of the relevant motor unit operates smoothly.

Connect the synchroscope to the following test point of the I/O PCB, which corresponds to the relevant
motor. Start MOTOR P.I. CHECK and check if the signal is changed (LOW level HIGH level).
X motor: TP214
LOW level (near 0V) in normal status
HIGH level (5V) when the original point is detected
Z motor: TP217
LOW level (near 0V) in normal status
HIGH level (5V) when the original point is detected
Y motor: TP215
LOW level (near 0V) in normal status
HIGH level (5V) when the upper limit is detected
Y motor: TP216
LOW level (near 0V) in normal status
HIGH level (5V) when the lower limit is detected

If the signal of the test point is changed, it is highly possible that the I/O PCB malfunctions.
If the signal is not changed, it is judged that the relevant original point sensor malfunctions.
Replace the relevant original point sensor.

If the instrument is still abnormal though the above-mentioned check is done, it is judged that the relevant motor malfunctions.
Replace the relevant motor.

When OK is displayed, check if XY alignment or Z alignment is properly adjusted.

If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or XY analog PCB.
Z alignment : Replace the Z sensor PCB.

39

R-SP3000P-1-0601-01

(2) Automatic alignment operates but the endothelium is not photographed. (The image cannot be changed to
the endothelium one or it is difficult to change.)
[CHECK POINT]
1 Start FACTORY MENU Program and perform SOLENOID CHECK to check if the image is
changed.
If the solenoid is not turned ON/OFF, check the connection of J410 on the XY analog PCB.

Check the light path selector mechanism.

If there is no problem in the light path selector mechanism, it is judged that the solenoid unit malfunctions. Replace the solenoid unit.

If the solenoid is turned ON/OFF, check if the XY alignment or Z alignment is properly adjusted.

If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or the XY analog PCB.
Z alignment : Replace the Z sensor PCB.

40

R-SP3000P-1-0601-01

(3) The image is stored but flash is not done. (The photographed image is very dark.)
[CHECK POINT]
1 Start FACTORY MENU Program and perform XE CHECK to check if flashing is done.
If the xenon lamp does not flash, check the connection of J215 on the I/O PCB.

If flashing is not done though the xenon lamp is replaced, it is judged that the I/O PCB malfunctions.
Replace the I/O PCB.

If the xenon lamp flashes, replace the xenon lamp or check the optical system light path and check if the
xenon lamp is stained.

(4) The image is not stored and flash is not done. (Alignment still continues.)
[CHECK POINT]
1 Check if XY alignment or Z alignment is properly adjusted.
If the above-mentioned check or adjustment cannot be done, it is judged that the relevant PCB malfunctions.
XY alignment : Replace the XY sensor PCB or the XY analog PCB.
Z alignment : Replace the Z sensor PCB.

41

R-SP3000P-1-0601-01

3-3-4 After photographing


(1) The image cannot be transferred for IMAGEnet.

I/O PCB

J112
1
2

VIDEO OUT

DG

1
2
3
4
5
6
7
8
9

J113
1
2
3
4
5

SIGNAL

D-SUB 9PIN
CONTROL

DG

[CHECK POINT]
1 Is the VIDEO signal outputted?
2 Is D-SUB9PIN connected from J113 of the I/O PCB?

(2) Data cannot be inputted or outputted by RS-232C.


[CHECK POINT]
1 Is the initial setting correct for inputting or outputting data?
2 Connect the cross cable (commercial product) for D-SUB9PINs and start FACTORY MENU Program. Then, perform RS232C CHECK and check if OK is displayed.
When OK is displayed Check the setting for RS-232C on the partner side.
When FAIL is displayed The CPU PCB malfunctions. Replace the CPU PCB.

(3) Data cannot be outputted by USB.


[CHECK POINT]
1 Check if RS-232C is initially set for outputting data.

42

R-SP3000P-1-0601-01

(4) Zooming and simplified cell analysis cannot be done.


[CHECK POINT]
1 Does the mouse operate normally?
Check the connection of the mouse connector.
Check if the mouse operates normally by inserting/removing the connector.

Check the connection of J116 on the CPU PCB.

Replace the mouse with a new one and check if it operates normally.
If the mouse does not operate normally, it is judged that the CPU PCB malfunctions.
Replace the CPU PCB.

43

R-SP3000P-1-0601-01

3-4 Assembly and Adjustment Procedures


3-4-1 Adjustments and cautions when replacing each unit
When replacing the units of the electric system as the result of troubleshooting and self-check, the necessary
adjustments are shown below.
Unit name
Switching power supply

Unit number
41903 66300

Adjustment and caution

Page

Checking of the switching power supply output

45

voltage
I/O PCB

CPU PCB

41903 72000

41903 71000

Replacement of EEPROM

(The old EEPROM before replacement The


EEPROM to be newly installed)
Adjustment of the xenon light intensity

57

Storing of the XY alignment circuit offset


Storing of the Z alignment circuit offset
Storing of the corneal thickness correction value
Storing of the IMAGEnet mask slit inside dimen-

52
56
58
59

sion
XY analog PCB

41903 74000

XY sensor PCB

41903 75100

Z sensor PCB

41903 75500

XY LED PCB

41903 76100

Z LED PCB

41903 76200

Adjustment of the XY alignment circuit frequency


Adjustment of the XY alignment circuit gain
Storing of the XY alignment circuit offset
Adjustment of the XY alignment sensor position
Adjustment of the XY alignment circuit gain
Storing of the XY alignment circuit offset
Adjustment of the Z LED luminance
Adjustment of the Z alignment sensor position
Adjustment of the Z alignment sensor circuit gain
Storing of the Z alignment circuit offset
Adjustment of the XY LED position
Adjustment of the Z LED position

49
51
52
50
51
52
53
54
55
56
71
69

When replacing or adjusting the PCB's mentioned above, be sure to carry out the checks or adjustments in the
following order.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17

Checking of the switching power supply output voltage


EEPROM initial setting
Adjustment of the XY LED luminance
Storing of the XY alignment offset on CCD
Adjustment of the XY alignment circuit frequency
Adjustment of the XY alignment sensor position
Adjustment of the XY alignment circuit gain
Storing of the XY alignment circuit offset
Adjustment of the Z LED luminance
Adjustment of the Z alignment sensor position
Adjustment of the Z alignment sensor circuit gain
Storing of the Z alignment circuit offset
Adjustment of the xenon light intensity
Storing of the corneal thickness correction value
Storing of the IMAGEnet mask slit inside dimension
Date setting
Time setting
44

R-SP3000P-1-0601-01

3-4-2 Checking of the switching power supply output voltage


[Purpose]

[Procedure]

Check the switching power supply output voltage so that the voltage at each unit can meet the
standard of the element being used during operation.
Make sure that the power supply voltage at each PCB is within the standard.
1

Make sure that the output voltage of the switching power supply is within the following
standard.
Output voltage

Standard

DC+24V

Reference voltage 10%

DC+285V

Reference voltage 10%

Make sure that all the cables of the electric parts are correctly connected. Then, make sure
that the output voltage at each PCB is within the standard.
Output voltage

Standard

PCB to be measured

Pin No.

DC+18V output

+18V10%

I/O PCB (4190372000)

TP204

DC+15V output

+15V5%

I/O PCB (4190372000)

TP207

DC+12V output

+12V0.5V

I/O PCB (4190372000)

TP208

DC-15V output

-15V5%

I/O PCB (4190372000)

TP210

DC+5.0V output

+5.0V0.5V

I/O PCB (4190372000)

TP205

DC+3.3V output

+3.3V0.3V

I/O PCB (4190372000)

TP209

DC+1.5V output

+1.5V0.1V

CPU PCB (4190371000)

45

R-SP3000P-1-0601-01

3-4-3 EEPROM initial setting


[Purpose]

[Procedure]

Store the initial value of each correction value before adjusting the electricity of SP-3000P.
Unless this setting is done, all the adjustments are adversely affected hereafter (for example, the
character display position is not proper on the monitor screen).
1
2
3
4
5
6
7

Start the "FACTORY MENU" program and make sure that the selection display appears on
the screen.
Set SW201 to the WP-OFF side on the I/O PCB to cancel WRITE-PROTECT. (LED208 is
turned on.)
Press the [Photography] switch. Select "A-1. EEPROM INITIALIZE" and press the [Image
transfer] switch.
The display as shown below appears on the monitor screen. Press the [Image transfer]
switch, and EEPROM is initialized.
When the initialization is finished normally, "EEPROM INITIALIZATION DONE" is displayed on the monitor screen.
When the initialization fails, "EEPROM INITIALIZATION FAILED" is displayed on the
monitor screen.
After the initialization is finished, press the [Image transfer] switch to return to the selection
display.

46

R-SP3000P-1-0601-01

3-4-4 Adjustment of the XY LED luminance


[Purpose]
[Procedure]

Make the light intensity of the XY alignment LED uniform for each instrument.
1
2
3

Set the tool base and the XY detection system tool.


Select "GROUP-A XY LED ADJUST" in the FACTORY MENU program.
The luminance level of the XY LED and the OK/NG for judgment are displayed. Adjust
VR9 (XY analog PCB) so that the luminance may be within the standard.
Standard

Adjustment place

The value on the monitor should be


1055 and "OK" should be displayed.

VR9 (XY analog PCB)

47

R-SP3000P-1-0601-01

3-4-5 Storing of the XY alignment offset on CCD


[Purpose]
[Procedure]

Store the reference position to perform the XY alignment on CCD in EEPROM.


1
2
3

Set the tool base and the XY detection system tool.


Select "GROUP-A CCD CENTER ADJUST" in the "FACTORY MENU" program.
The "CCD CENTER ADJUST" display appears on the monitor screen as shown below.
Make sure that the numerical values are displayed and then exit from this program with the
[Clear] switch.

48

R-SP3000P-1-0601-01

3-4-6 Adjustment of the XY alignment circuit frequency


[Purpose]

[Procedure]

Tune the synchronized frequency of the band-pass filter circuit in the XY alignment (PSD sensor) circuit in the LED modulated frequency (3.6kHz). Unless this adjustment is done correctly,
the filter characteristics are deteriorated by temperature change.
1
2
3
4

Set the tool base and the XY detection system tool.


Select "GROUP-B XY SENSOR ADJUST" in the "FACTORY MENU" program.
The display as shown below appears on the monitor screen. The XY LED is turned on and
adjustment is possible.
Connect the probes of the oscilloscope as shown below.
Observation probe : TP5 (XY analog PCB)
Ground
: TP7 (XY analog PCB)
Turn VR6 clockwise to maximize the signal amplitude. At this time, set the perpendicular
axis sensitivity of the oscilloscope at 2V/div and run electricity for the XY LED not to cause
saturation (by VR9 on the XY analog PCB).
Connect the observation probe of the oscilloscope to TP4, TP2 and TP1 in turn and adjust
VR5, VR1 and VR2 so that each signal amplitude may be maximum. Adjust each VR by
turning it in the same direction.
Standard
Maximum amplitude
Conditions:
The perpendicular axis sensitivity: 2V/div
Power up the XY LED not
to cause saturation.

Adjustment place
X1 axis

VR6

TP5

X2 axis

VR5

TP4

Y1 axis

VR1

TP2

Y2 axis

VR2

TP1

Condition: Adjust each VR by turning it in


the same direction (clockwise).

49

R-SP3000P-1-0601-01

3-4-7 Adjustment of the XY alignment sensor position


[Purpose]

[Procedure]

Adjust the XY alignment light to be approximately at the center of the XY sensor. Unless this
adjustment is done correctly, the alignment detection range is not well-balanced.
1
2

3
4

5
6

Carry out this adjustment just after "3-4-6 Adjustment of the XY alignment circuit frequency" without changing the condition.
Connect the probes of the oscilloscope as shown below.
Observation probe (CH1) : TP5 (XY analog PCB)
(CH2) : TP4 (XY analog PCB)
Ground
: TP7 (XY analog PCB)
Set "CHOP MODE" for the oscilloscope. Move the XY sensor PCB in the X direction to
adjust its position so that the level of TP5 may be the same as TP4.
Connect the probes (CH1) and (CH2) of the oscilloscope to TP2 and TP1. Move the XY
sensor PCB in the Y direction to adjust its position so that the level of TP2 may be the same
as TP1.
Repeat the above 3 and 4 several times. Adjust the XY sensor PCB position until each
level for TP5 and TP4 and for TP2 and TP1 is within the standard.
Press the [Image transfer] switch to return to the selection display.
Standard

Adjustment place

Level difference V = Within 0.4V


Conditions:
Perpendicular axis sensitivity: 2V/div
Power up the XY LED not to cause
saturation.

XY sensor PCB

50

R-SP3000P-1-0601-01

3-4-8 Adjustment of the XY alignment circuit gain


[Purpose]

[Procedure]

Adjust the circuit sensitivity so that the XY alignment can be detected correctly when the
patient's eye is aligned.
1
2
3
4

Carry out this adjustment just after "3-4-7 Adjustment of the XY alignment sensor position"
without changing the condition.
Select "GROUP-B XY SENSOR ADJUST" in the "FACTORY MENU" program.
The gain level values of the XY alignment circuit are displayed. Adjust VR8, VR7, VR4
and VR3 (XY analog PCB) so that the level values may be within the standard.
After adjustment, press the [Image transfer] switch to return to the selection display.
Standard
The displayed value on the
monitor:
1075

Adjustment place
X1 axis

VR8

X2 axis

VR7

Y1 axis

VR4

Y2 axis

VR3

51

R-SP3000P-1-0601-01

3-4-9 Storing of the XY alignment circuit offset


[Purpose]

[Procedure]

Store the condition of the XY alignment circuit system in EEPROM so that the adjustment error
of the circuit system may be finally canceled. Unless this storing procedure is done correctly,
the position detection accuracy of the XY alignment cannot be obtained.
1
2
3

Set the tool base and the XY detection system tool.


Select "GROUP-B XY SENSOR ADJUST" in the "FACTORY MENU" program.
Press the [Image transfer] switch, and measurement starts.
The offset values (Level of X1, Level of X2, Level of Y1 and Level of Y2) on the sensor are
displayed. By pressing the [Image transfer] switch, measurement is automatically repeated
until "100" is shown on "COUNT". After measurement, these values are automatically
stored in EEPROM.
If "COUNT" does not reach "100", the tools are not set properly or the XY alignment circuit
system is not adjusted correctly.
If you want to store the values again, press the [Clear] switch to return to the selection display and select "GROUP-B XY SENSOR ADJUST" again.
After storing, press any switch to return to the selection display.

52

R-SP3000P-1-0601-01

3-4-10 Adjustment of the Z LED luminance


[Purpose]
[Procedure]

Make the light intensity of the Z alignment LED uniform for each instrument.
1
2
3

Set the tool base and the Z detection system tool B.


Select "GROUP-B Z SENSOR ADJUST" in the "FACTORY MENU" program.
Connect the probes of the oscilloscope as shown below.
Observation probe: Z VIDEO (Z sensor PCB)
Ground
: AG (Z sensor PCB)
Set the instrument so that you can observe one perpendicular synchronization section.
Adjust VR10 (XY analog PCB) so that the voltage level of the endothelium signal against
the blanking level may be within standard.
Standard: 0.3V0.2V

Voltage: 0.3V0.2V

53

R-SP3000P-1-0601-01

3-4-11 Adjustment of the Z alignment sensor position


[Purpose]

[Procedure]

Adjust the Z alignment slit light to be approximately at the center of the Z sensor. Unless this
adjustment is done correctly, the Z alignment detection range is not well-balanced.
1
2
3

Carry out this adjustment just after "3-4-10 Adjustment of the Z LED luminance" without
changing the condition.
Remove the Z detection system tool B and set the Z slit tool.
Connect the probes of the oscilloscope as shown below.
Observation probe : Z VIDEO (Z sensor PCB)
Ground
: AG (Z sensor PCB)
The different-level reflected slit light is displayed on the monitor screen as illustrated. As
watching the oscilloscope, adjust the Z sensor PCB position in the Y direction so that the
light on the center of the shadowed section A may be received. The Z sensor PCB position
in the X direction should be the position where it touches the metal part.
When the light of the shadowed section A is received, the waveform on the oscilloscope
shifts rightward as shown below. At the middle position of this shift, fix the PCB.

Different-level slit light

54

R-SP3000P-1-0601-01

3-4-12 Adjustment of the Z alignment sensor circuit gain


[Purpose]

[Procedure]

Adjust the circuit sensitivity so that Z alignment can be detected correctly when the patient's eye
is aligned.
1
2
3

Carry out this adjustment just after "3-4-11 Adjustment of the Z alignment sensor position"
without changing the condition.
Remove the Z slit tool and set the Z projection system tool B.
Adjust VR1 (Z sensor PCB) so that the amplitude of the endothelium signal may be within
the standard on the oscilloscope.
Standard: 3.0V0.1V

3.0V0.1V

55

R-SP3000P-1-0601-01

3-4-13 Storing of the Z alignment circuit offset


[Purpose]

[Procedure]

Store the condition of the Z alignment circuit system in EEPROM so that the adjustment error of
the circuit system may be finally canceled. Unless this storing procedure is done correctly, the
position detection accuracy of the Z alignment cannot be obtained.
1
2

3
4

Carry out this storing procedure just after "3-4-12 Adjustment of the Z alignment sensor circuit gain" without changing the condition.
Press the [Image transfer] switch, and the value for "COUNT" is increased.
Then, the endothelium reflection position, "N", and the optical magnification "M" on the
sensor are displayed as the numerical values.
If the value for "COUNT" is not increased, the tools are not set properly or the Z alignment
circuit system is not adjusted correctly.
After storing, press the [Image transfer] switch to return to the selection display.

COUNT UP

Display while storing the offset

Display when the offset storing procedure is


finished

56

R-SP3000P-1-0601-01

3-4-14 Adjustment of the xenon light intensity


[Purpose]

[Procedure]

Adjust the charging voltage of the xenon tube lighting condenser. Unless this adjustment is done
correctly, the brightness of the photographed endothelium image is different between the instruments.
1
2
3
4

While the power is OFF, set VR201 of the I/O PCB to the medium position.
Set the tool base and the Z projection system tool B.
Connect the BNC video cable from the VIDEO OUT terminal to the oscilloscope. Set the
instrument so that you can observe one horizontal synchronization section.
Turn on the power. When the photography result display appears, observe the luminance
level of one horizontal synchronization section on the sample chart for 2000 through oscilloscope.
If the luminance level is out of the standard, adjust VR201 again, perform photography and
check the luminance level.
Repeat this procedure until the luminance level is within the standard. The VIDEO signal
level in "FLASH HIGH" should be a reference value.
The light intensity is increased by turning the volume clockwise.

Voltage
V

Screen on oscilloscope

Standard:

In "FLASH LOW" : V = 550mV ~ 600mV


In "FLASH HIGH": (V = 825mV ~ 900mV)

57

R-SP3000P-1-0601-01

3-4-15 Storing of the corneal thickness correction value


[Purpose]

[Procedure]

Store the correction value which is to be used for the final correction calculation in EEPROM so
that the absolute accuracy of the corneal thickness can be ensured.
1
2
3

Set the corneal thickness test eyes (three test eyes with different thickness) onto the chinrest
unit.
Select "GROUP-B PACHOMETRY ADJUST" in the "FACTORY MENU" program.
The corneal thickness of the test eye as a reference is displayed on the monitor screen as
shown below. Change the value with the [Chinrest up/down] switches and change the digit
with the selector switch to adjust finely in order to get the corneal thickness of the test eye to
be measured.

By using the [Image selector] switch, move to the approximate position of the test eye,
which is the target displayed on the screen. Measurement is performed ten times according
to automatic alignment.
Repeat the procedure of above 3 and 4 by using the three test eyes. The correction values
are displayed and they are automatically written in EEPROM. Then, press any switch to
return to the MENU display.

58

R-SP3000P-1-0601-01

3-4-16 Storing of the IMAGEnet mask slit inside dimension


[Purpose]

[Procedure]

Store the mask slit inside dimension in EEPROM in consideration of the optical magnification
error so that the scale calibration, which is necessary for IMAGEnet cell analysis, can be done
correctly.
1
2

Set the tool base and the Z detection system tool A.


Select "GROUP-A IMAGEnet SLIT MEMORY" in the "FACTORY MENU" program and
set the test eye so that you can see the following image.

UPPER

LOWER

4
5

Select the "UPPER" and "LOWER" lines with the [Image selector] switch. By using the
[Chinrest up/down] switches, align "UPPER" to the upper third square and "LOWER" to the
lower third square with the black point of the test eye chart as the center. Click the [Image
selector] switch to store the interval of the six squares on the test eye chart.
Press the [Photography] switch, and the mask slit inside dimension, which is calculated in
consideration of the optical magnification error, is stored.
Press any switch on the control panel to return to the "FACTORY MENU" display.

59

R-SP3000P-1-0601-01

3-4-17 Date setting


[Procedure]

1
2

Turn on the power switch.


Press the [MENU] switch on the control panel.

[MENU] switch
Line B

The "ID INPUT" dialog box is displayed as shown below.

Fit the mouse pointer to the "DATE SET" tab and click it.
The "DATE SET" dialog box is displayed. Fit the mouse pointer to the "Y/M/D" icon,
which is the date order icon, and click it.

6
7
8

Fit the mouse pointer to the desired numeral on the "Ten key" icons and click it.
The entered numeral is displayed on the date display area.
Enter the numerals continuously until the desired date is displayed. Then, fit the mouse
pointer to the "Enter key" icon and click it. The date is set.
60

R-SP3000P-1-0601-01

3-4-18 Time setting


[Procedure]

1
2

3
4

Fit the mouse pointer to the "TIME SET" tab and click it.
The "TIME SET" dialog box is displayed. Fit the mouse pointer to the "AM/PM key" icon
and click it.

Fit the mouse pointer to the desired numeral on the "Ten key" icons and click it.
Enter the numerals continuously until the desired time is displayed. Then, fit the mouse
pointer to the "Enter key" icon and click it. The time is set.

61

R-SP3000P-1-0601-01

3-4-19 Adjustment of the anterior eye segment rear lens group


[Procedure]

1
2
3
4
5
6
7

Remove the diaphragm (41902 1122) and set the tool cross scale.
Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [MENU] switch on the control panel to erase "! scale" on the screen.
Illuminate from the patient's eye side.
Adjust the CCD camera so that the tool cross scale may be displayed at the center of the
monitor.
By moving the rear lens group cell unit back and forth, fit the focus to the first long scale
line from the center and then fix the unit.
Remove the tool cross scale and set the diaphragm (41902 1122).

Diaphragm
(41902 1122)
Tool cross scale

Adjust the focus on the


monitor screen.
Rear lens group cell unit

Illumination

CCD camera
surface

62

R-SP3000P-1-0601-01

3-4-20 Adjustment of the anterior eye segment front lens group


[Procedure]

1
2
3

Carry out this adjustment just after "3-4-19 Adjustment of the anterior eye segment rear lens
group" without changing the condition.
Set the tool base and the XY detection system tool.
By moving the front lens group cell unit (41902 1150) back and forth, make adjustment so
that XY bright spot may be circular and then fix the unit.

The XY bright spot should be


circular.

XY detection
system tool

Front lens group cell unit

63

R-SP3000P-1-0601-01

3-4-21 Adjustment of the CCD camera position


[Procedure]

1
2
3

Carry out this adjustment just after "3-4-20 Adjustment of the anterior eye segment front
lens group" without changing the condition.
Press the [MENU] switch on the control panel to display "! scale".
By moving the CCD camera, make adjustment so that the XY bright spot may be within the
"!" (the displayed mark for adjustment) and then fix the camera.
* Be careful not to set the CCD camera slantingly. (If it is set slantingly, the observation/
photography slits are tilted.)
Press the [Photography] switch to return to the "FACTORY MENU" program display.

Monitor screen
The displayed mark
for adjustment

XY bright spot image

XY detection
system tool

64

R-SP3000P-1-0601-01

3-4-22 Adjustment of the photography system relay lens


[Procedure]

1
2
3
4
5
6

7
8
9

10
11
12

Remove the slit holder (41902 1307) from the mask unit sub.
Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
Illuminate from the patient's eye side.
By moving the relay lens unit (41902 1380) back and forth, adjust the focus temporarily so
that the mask image can be seen.
Move the relay lens unit by using the pulling screw and pushing screw on the side of the
endothelium reflection mirror unit so that the mask may be displayed at the center of the
instrument monitor.
Turn the mask lens unit (41902 1350) so that the mask slit may be perpendicular on the
instrument monitor screen.
Press the [MENU] switch on the control panel, and the memory image appears on the instrument monitor screen.
Move the relay lens unit a little by using the pulling screw and pushing screw on the side of
the endothelium reflection mirror unit to make adjustment so that the memory image may be
positioned at the center.
By moving the relay lens unit back and forth, make adjustment to set the mask slit focus
equally in right and left.
Repeat the above 9 to 10 to adjust the position and focus of the mask.
Press the [Photography] switch to return to the "FACTORY MENU" program display.

Slit holder

[Mask unit sub]

If the mask is positioned between the


memory images, the point is within "!3mm".
Relay lens unit
(Move this unit back and forth to adjust focus.)

Endothelium reflection
mirror unit

65

R-SP3000P-1-0601-01

3-4-23 Adjustment of the photography system condenser lens


[Procedure]

1
2
3
4
5
6
7
8

9
10
11

Set the tool base and the Z detection system tool B.


Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
Press the [DELETE] switch on the control panel to turn off the Z LED.
Illuminate from the patient's eye side.
By moving the condenser lens cell unit (41902 1320) back and forth with the screws, carry
out focusing so that the chart can be seen.
By turning the eccentric adjustment lens sub eccentrically, make adjustment so that the chart
center may be fit to the mask center.
By moving the condenser lens cell unit (41902 1320) back and forth with the screws, make
adjustment to bring the center longitudinal line of the chart into focus.
* Adjust focus so that the blur may be uniform from the center longitudinal line to the right
and left longitudinal lines on the chart.
Repeat the above 7 to 8 to adjust the position and focus.
Press the [MENU] switch on the control panel to make sure that the black point is within the
"!" of the memory image.
Press the [Photography] switch to return to the "FACTORY MENU" program display.

Focus position

Eccentric adjustment lens sub


Approx. center of mask

Z detection system
tool B

66

R-SP3000P-1-0601-01

3-4-24 Adjustment of the observation slit


[Procedure]

1
2
3
4
5
6
7

Set the tool base and the Z detection system tool A.


Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
Move the observation slit with the adjustment screws (pushing screw and pulling screw) to
be at the center of the instrument monitor screen.
By moving the mirror holder unit (41902 1220) back and forth, adjust the focus of the observation slit correctly.
Press the [MENU] switch on the control panel, and the memory image appears on the instrument monitor screen.
Move the observation slit a little with the adjustment screws (pushing screw and pulling
screw) to make adjustment so that the memory image may be positioned at the center of the
slit center section.

Slit image

Memory image

Z detection system
tool A
Adjustment screw
Mirror holder unit (41902 1220)

67

R-SP3000P-1-0601-01

3-4-25 Adjustment of the photography slit


[Procedure]

1
2
3
4
5
6
7
8
9

Carry out this adjustment just after "3-4-24 Adjustment of the observation slit" without
changing the condition.
Remove the reflection mirror unit and the xenon lamp unit. Illuminate with the halogen
light source device equipped with light guide from the xenon lamp position.
Press the [DELETE] switch on the control panel to turn off the Z LED.
Move the photography slit right and left to be positioned at the center of the instrument monitor screen.
By moving the slider back and forth, adjust the slit focus correctly.
Press the [MENU] switch on the control panel, and the memory image appears on the instrument monitor screen.
Make adjustment so that the photography slit may be positioned at the center of the memory
image.
Press the [Photography] switch to return to the "FACTORY MENU" program display.
Set the reflection mirror unit and the xenon lamp unit.

Memory image

Photography slit
(By moving back and forth, adjust focus.
By moving right and left, adjust position.)

Slit image
LED holder

Tool
(Z test eye A)

Visible-light
illumination

Slider

68

Xenon lamp unit

Reflection mirror unit

R-SP3000P-1-0601-01

3-4-26 Adjustment of the Z LED position


[Procedure]

1
2
3
4
5

6
7

Set the tool base and the Z projection system tool A.


Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [MENU] switch on the control panel to erase "! scale" on the screen.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
By moving the LED holder longitudinally and laterally, make adjustment so that the center
of the Z LED luminous flux may be positioned at the center of the observation slit on the
instrument monitor screen.
Remove the tool base and the Z projection system tool A. By viewing from the patient side,
make sure that the luminous flux center is at the diaphragm center of the lens unit (#1210).
Press the [Photography] switch to return to the "FACTORY MENU" program display.

Lens unit (#1210)


LED holder

Z projection
system tool A

69

R-SP3000P-1-0601-01

3-4-27 Adjustment of the knife edge position


[Procedure]

1
2
3
4
5
6
7
8

9
10

Slit holder

Set the tool base and the Z detection system tool B.


Set the slit holder.
Remove the reflection mirror unit and the xenon lamp unit. Illuminate with the halogen
light source device equipped with light guide from the xenon lamp position.
Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [MENU] switch on the control panel to erase "! scale" on the screen.
Press the [Image selector] switch on the control panel, raise the solenoid and change to the
photography system.
Press the [DELETE] switch on the control panel to turn off the Z LED.
By turning the bracket and by using the pulling screw and pushing screw on the side of the
slit holder, make adjustment so that the analyzed image of the knife edge may be positioned
perpendicularly to the left fifth line from the chart center longitudinal line.
Press the [Photography] switch to return to the "FACTORY MENU" program display.
Set the reflection mirror unit and the xenon lamp unit.

Bracket
Eccentric adjustment lens sub
Analyzed image

Z detection
system tool B

70

R-SP3000P-1-0601-01

3-4-28 Adjustment of the XY LED position


[Procedure]

1
2
3
4

Set the tool base and the XY detection system tool.


Select "GROUP-B OPTICAL ADJUST" in the "FACTORY MENU" program.
Press the [MENU] switch on the control panel to erase "! scale" on the screen.
Adjust the position of the XY LED PCB so that the XY bright spot may be brightest. Then,
fix the XY LED PCB.

XY LED PCB
XY bright spot

71

R-SP3000P-1-0601-01

4. TOOL LIST
4-1 Special Repair Tools

Tool No.

Usage
rank

Name

Illustration

Application

SP-2000P
-01

Tool base

Optical and electrical adjustment

-02

XY detection system
tool

Adjustment of the XY alignment


Adjustment of the anterior eye
segment front lens group and CCD
camera
Adjustment of the XY LED position

-03

Z projection system
B tool

Adjustment of the Z alignment

-04

Z slit tool

Adjustment of the Z alignment


sensor position

-05

Adjustment of the anterior eye

Cross scale tool

segment rear lens group

-06

-07

-08

Z detection system
A tool

Adjustment of the projection sys-

Z projection system
A tool

Adjustment of the observation/

Z detection system
B tool

Adjustment of the knife edge

tem objective lens focus

projection slit

72

R-SP3000P-1-0601-01

Tool No.

Usage
rank

SP-3000P
-01

-02

Name

Illustration

Application

Corneal thickness
test eye
0.4, 0.5, 0.6, 0.7,
0.78mm

Checking of the corneal thickness

Corneal thickness
test eye and
attachment
0.4, 0.6, 0.78mm

Checking of the corneal thickness

accuracy and corneal thickness


correction

accuracy and corneal thickness


correction
By combining with the corneal
thickness test eye (Tool No. SP2000P-09), the configuration is the
same as the above corneal thickness test eye.

4-2 General Repair Tools


Synchroscope
Digital multimeter
Halogen lamp illumination device equipped with light guide (50W or higher)
Phillips screwdriver set
Slotted screwdriver set
Hexagonal wrench set

73

R-SP3000P-1-0601-01

TOPCON MEDICAL SYSTEMS, INC


37 West Century Road,Paramus,New Jersey 07652,U.S.A. Phone:201-261-9450 Fax:201-387-2710 www.topcon.com

TOPCON CANADA INC.


110 Provencher Avenue, Boisbriand, QC J7G 1N1 CANADA Phone:450-430-7771 Fax:450-430-6457 www.topcon.ca

TOPCON EUROPE B.V.


(European Representative)
Essebaan 11, 2908 LJ Capelle a/d IJssel,THE NETHERLANDS Phone:010-4585077 Fax:010-4585045 www.topconeurope.com

TOPCON EUROPE MEDICAL B.V.


(European Sole Sales Company)
Essebaan 11, 2908 LJ Capelle a/d IJssel,THE NETHERLANDS Phone:010-4585077 Fax:010-2844940 www.topconeurope.com
ITALY OFFICE:Via Dell' Industria n.60, 20037 Paderno Dugnano, (Milano), ITALY Phone:02-61-25-583 E-mail:info.topconitaly@tiscali.it www.topcon.it

TOPCON DEUTSCHLAND G.m.b.H.


Giesserallee 31-33 D-47877 Willich GERMANY Phone:02154-8850 Fax:02154-885111 www.topcon.de Med@topcon.de

TOPCON ESPAA S.A.


HEAD OFFICE:Frederic Mompou 5, ED. Euro 3, 08960,Sant Just Desvern Barcelona,SPAIN Phone:93-4734057 Fax:93-4733932 www.topconesp.com
MADRID OFFICE:Avenida Burgos, 16E,1 28036,Madrid,SPAIN Phone:91-302-4129 Fax:91-383-3890

TOPCON S.A.R.L.
89, rue de Paris 92585 Clichy, Cedex,FRANCE Phone:01-4106-9494 Fax:01-4739-0251

TOPCON SCANDINAVIA A.B.


Neongatan 2 S-43151 Mlndal, SWEDEN Phone:031-7109200 Fax:031-7109249 info@topcon.se

TOPCON (GREAT BRITAIN) LTD.


Topcon House,Kennet Side,Bone Lane,Newbury,Berkshire RG14 5PX United Kingdom Phone:01635-551120 Fax:01635-551170

TOPCON SOUTH ASIA PTE.LTD.


Blk 192 Pandan Loop, #07-01 Pantech Industrial Complex, SINGAPORE 128381 Phone:62780222 Fax:62733540 www.topcon.com.sg

TOPCON INSTRUMENTS (MALAYSIA) SDN.BHD.


Excella Business Park Block C,1st Floor,Jalan Ampang Putra,Taman Ampang Hillir, 55100 Kuala Lumpur,MALAYSIA Phone:03-42701192 Fax:03-42704508

TOPCON INSTRUMENTS (THAILAND) CO.,LTD.


77/162 Sinn Sathorn Tower, 37th Fl.,Krungdhonburi Rd.,Klongtonsai, Klongsarn, Bangkok 10600,THAILAND Phone:440-1152~7 Fax:440-1158

TOPCON AUSTRALIA PTY.LTD.


Unit 18,4 Avenue of Americas Newington NSW 2127 AUSTRALIA Phone:02-8748-8777 Fax:02-9647-2926 www.topcon.com.au

TOPCON KOREA CORPORATION


2F Yooseoung Bldg., 1595-3, Seocho-Dong, Seocho-Gu, Seoul, 137-876 KOREA Phone:02-2055-0321 Fax:02-2055-0319 www.topcon.co.kr

TOPCON OPTICAL (H.K.) LTD.


2/F.,Meeco Industrial Bldg.,No.53-55 Au Pui Wan Street,Fo Tan Road,Shatin,N.T.,Hong Kong Phone:2690-1328 Fax:2690-2221 E-mail:sales@topcon.com.hk

TOPCON CORPORATION BEIJING OFFICE


1070 Poly Plaza Building,14 Dongzhimen Nandajie Dongcheng District,Beijing,100027,CHINA Phone:10-6501-4191 Fax:10-6501-4190

TOPCON CORPORATION BEIRUT OFFICE


P.O.Box 70-1002 Antelias,BEIRUT-LEBANON Phone:961-4-523525/523526 Fax:961-4-521119

TOPCON CORPORATION DUBAI OFFICE


C/O Atlas Medical FZCO., P.O.Box 54304 C-25, Dubai Airport Free Zone, UAE Phone:971-4-2995900 Fax:971-4-2995901

75-1 Hasunuma-cho,Itabashi-ku,Tokyo,174-8580 Japan.


Phone:3-3558-2520 Fax:3-3960-4214 www.topcon.co.jp