Documente Academic
Documente Profesional
Documente Cultură
PROCEDURE
OBI Advanced Imaging version 1.4
Cone Beam CT version 2.1
4D ITC version 8.2
Clinac version 7.5
Revision 2.0.0
Objective:
After completing this module the student will have performed the following tasks
using the OBI Operations Manual as a resource:
Setup the patient
Perform a Cone Beam CT
Acquire a Topogram
Export the Images to OBI
Analyze images
Shift the Exact Couch into treatment position
Treat the patient
A. Exercise Key
1. Right mouse click denoted as RMC.
2. Left mouse click denoted as LMC.
3. Bracket () denotes additional options.
4.
B. Procedure Content:
1. Load Session on 4D ITC
2. On Board Imager Graphical User Interface
3. Acquire Images
4. Imaging and Verification Tools
Varian Medical Systems
FOR TRAINING PURPOSES ONLY!
rh, March 09
1/21
C. 4D ITC
1. Check-in the patient
2. Select the patient from the treatment queue and click OK (Figure 1).
3. Select the CBCT Setup Field and click Mode Up (Figure 2).
rh, March 09
2/21
D. OBI workstation
1. If not already positioned, extend the arms from outside the room by clicking
Download Axes.
Press Motion Enable and Auto buttons on the OBI Console to position the
arms.
WARNING: Always verify clearance prior to extending the OBI arms from the
OBI console.
2. On OBI workstation click 3D/3D Match to start acquisition (Figure 3). The
CBCT application will open automatically. The original planning CT will be
downloaded on to the OBI for matching.
3. The following warning message may appear: Make sure Couch Rails are
removed if fitted.
Remove the couch rails and confirm the warning message by clicking OK
(Figure 4). This message can be disabled in CBCT Administration.
4. If this is the patients first Cone Beam CT the following options will be available.
rh, March 09
3/21
rh, March 09
4/21
6. Select the appropriate predefined CBCT mode from the CBCT Mode Menu and
if necessary select the appropriate patient orientation. (Figure 7)
Diameter and
Acquisition Mode will
vary based on the
selected CBCT mode.
rh, March 09
5/21
10. Press Start Scan (Figure 10). The following message will appear: Acquiring
Dark Field Please Wait (Figure 11).
The system performs a Dark Field to account for and eliminate background
noise.
Figure 10
11. Once the dark field has been acquired the gantry needs to be moved to the start
position. This will be 185 or 175 whichever is closer to the actual gantry
position. The couch may also move to a safe position for the Cone Beam CT.
Press and hold MEB and >> on the Clinac dedicated keyboard to move gantry
and/or couch into position (Figure 12).
The following message will appear if the couch movement is aborted. Couch
Position error the couch is outside of the allowed area (Figure 13).
13. Press and hold X-ray footswitch or the Hand Pendant. Then press and hold
the MEB and >> on the Clinacs dedicated keyboard. This will start CT
acquisition.
Varian Medical Systems
FOR TRAINING PURPOSES ONLY!
rh, March 09
6/21
Hold all three depressed until the acquisition has finished (audio signal will
stop).
Once the scan is complete release the MEB and >> first and then release the
X-ray footswitch.
14. Acquisition and reconstruction progress will be displayed on the screen (Figure
15).
The Acquisition Progress is the data being acquired and sent from the CBCT
cache to the Reconstructor computer.
The Serialization Progress is the process of sending the data back from the
Reconstructor computer to the OBI computer.
15. If the scan is interrupted due to a purposeful action (ex. Breath-hold CBCT) or
an accident (release of dedicated keyboard or foot/handswitch) the scan can be
resumed. Select Resume Acquisition(Figure 16a). The system will request
the gantry be repositioned approximately 10 degrees prior to the interruption
position (Figure 16b). The system will then request that the acquisition be
resumed(Figure 16c)
rh, March 09
7/21
16. If the couch was moved to a safe position for scanning it will need to be returned
to the original set up position prior to beginning the match process (Figure 17).
Press and hold MEB and >> on the Clinac dedicated keyboard to move the
couch into position (Figure 18).
rh, March 09
8/21
17. Once the scan is complete the reconstructed images will be displayed on the
screen. Review them if needed (Figure 20).
If the patient moved during the scanning process, select Decline Scan and restart the process.
18. Once the Accept and Export Scan button has been selected, the CBCT image is
stored on the OBI workstation. The CBCT image will then be imported into the
3D/3D Match workspace of OBI. The entire process may take a few seconds to
complete. Upon completion, the Acquisition workspace closes and the OBI
Analyze workspace opens automatically.
E. OBI 3D/3D Workspace
1. At this point there are two options for matching: Automatic Match (see step 3)
and Manual Match (see step 5). They can be used individually or in
combination.
Varian Medical Systems
FOR TRAINING PURPOSES ONLY!
rh, March 09
9/21
OBI will perform an automatic match using the Mutual Information Algorithm
(see Appendix E for more information on the Algorithm)
3. The Auto Match Control dialog box opens and the matching process starts
(Figure 22).
4. The system shifts the two superimposed images relative to each other during
the match and calculates the couch correction values (Figure 23 and 24).
rh, March 09
10/21
is active.
Adjust the image using the arrow keys on the keyboard. This allows for small
adjustments in the match alignment. Make large adjustments by clicking and
dragging the image with the mouse or hold the Alt key and press the
appropriate arrow key.
Rotate the image by holding the left mouse button and then click and drag
outside of the dotted red circle or hold the Ctrl key and use the right or left
arrow keys. Rotate the image in large steps by holding the Alt+Ctrl keys and
using the right or left arrow key.
To rotate the main viewing window between Transversal, Sagittal, and Coronal
views press Ctrl+R or this icon
11/21
Verify the match using the match verification tools (see Step 6).
Right Click on the Window and Level bar to select Range which automatically
adjusts to a specific anatomical selection and can be used on both the CBCT
and the reference CT (Figure 26b). (Can also be used in the CBCT window
(Figure 20).)
rh, March 09
12/21
From the Dynamic Window check or uncheck contours as needed during the
verification process (Figure 27a).
Contours can be turned off and on as a group using the Contour button
(Figure 27b)
rh, March 09
13/21
rh, March 09
14/21
Tool
Description
You can scroll through the sagittal, coronal, and transversal views of CBCT or CT
images by pressing the Page Up/Page Down keys.
Flicker Display
rh, March 09
15/21
7. Once you have verified the match result and accepted the couch correction
values, you can apply the couch shift. If for some reason you do not want to
send a specific shift it can be deselected by un-checking the include box.
Reset Shift will zero out any shifts and realign the images to there initial
positions.
The Save Match button saves the match back to the database for immediate
review in Offline Review.
Click Apply Shifts to send the Target positions to the clinac console. (Figure
31).
Use the MEB and the >> key on the Clinac dedicated keyboard to move the
couch into the new position.
9. Prior to treating, retract the OBI arms to reduce damage to the kV detector from
scattered radiation.
NOTE: After acquiring a CBCT the OBI is unable to communicate
correctly with the generator causing a problem with radiographic
acquisitions. Switch the generator from radiographic to fluoro mode and
back to radiographic mode. This will reset generator communication and
solve the problem.
F. 4D ITC
Varian Medical Systems
FOR TRAINING PURPOSES ONLY!
rh, March 09
16/21
Select Apply for Session and then click Apply (Figure 33).
Select Apply Permanently only if its desired to apply the couch values
permanently (Figure 34).
3. Select the first field to treat and click Mode Up (Figure 36).
rh, March 09
17/21
4. If the couch values are out of tolerance the following message will appear
asking for confirmation of table parameters (Figure 37).
rh, March 09
18/21
Reconstruction Filter:
Standard: 50/50 smooth and
sharp
Smooth: Best at defining soft
tissue
Sharp: Emphasis on bone
Ultra Sharp: Max emphasis on
bone
Ring Artifact Suppression:
You can control the strength of the
image correction by selecting one of
the following levels of ring artifact
suppression:
None: allows you to switch ring
artifact suppression off
Weak: might miss some rings
Medium: is recommended
Strong: might remove essential
information
Figure 39: Scan Parameters
H. Quick Scan
Varian Medical Systems
FOR TRAINING PURPOSES ONLY!
rh, March 09
19/21
1. When Quick Scan is selected the system will apply the same parameters that
were selected in the previous CBCT acquisition for that patient. (Only settings in
the parameter window are saved, dashboard changes will not be saved.)
Continue from step D-10.
5. The CBCT application will reopen. Select Reconstruct Existing Scan and
select the most current (just performed) scan (Figure 42a and 42b). Click Next.
rh, March 09
20/21
7. When reconstruction has finished select Accept and Export Scan (Step D17)
and continue with section E.
rh, March 09
21/21