Sunteți pe pagina 1din 38

3M Attest Auto-reader

Service Manual

Neither this book nor any part may be reproduced or transmitted in any form or by any means, electronic or mechanical, including
photocopying, microfilming, recording or by any information storage and retrieval system, without permission from 3M Health Care.

Health Care

Attest Auto-reader

Service Manual

________________________________________________________________________
Page 2 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

Revision History

Revision

Date

Page(s) Changed

10/16/2001

All

Change: Birth Notice, Manual Released


B

7/16/2003

Deleted Pages 5-13


Updated New Pages 5-14
Updated New Pages 16-21

Change: Eliminated Operators Manual Section


Change: Updated Acceptance Test Procedure to match C-05-005002 (MCU-0002) Version 7
Change: Updated BOMs to designate spare parts availability
C

7/18/2005

All

Change: Added drawings and parts list for new 290 case design
Change: Added drawings and parts list for 293/293G Auto-reader
Change: Updated Acceptance Test Procedure to include 293/293G Auto-reader
D

4/21/2009

All

Change: Performed an 11-digit stock number review, update some per ECOs.
Change: Added new round cord clip for power supply, 26-1017-3696-0 and deleted the flat
power supply cord clip and screw, 26-1017-2681-3, 26-1014-7689-8, Reference, ECO-05-003550.
Change: Changed the 10 inch ribbon cable (connects the NMOM and display PCB) from 26-10147284-8 to the 12 inch ribbon cable 26-1017-6304-8 for the 293/293G, reference ECO-05-003152.
Change: Added Connectivity Kit stock numbers to the 290 and 290G per ECO-05-003862.
Change: Added stand-offs, Dual Lock, etc. to Service Manual (for shipping purposes for the
293/293G), reference ECO-05-003345.

________________________________________________________________________
3M2009 78-8078-9168-0
Page 3 of 38

Attest Auto-reader

Service Manual

TABLE OF CONTENTS

TROUBLESHOOTING GUIDE
Troubleshooting Guide

ACCEPTANCE TEST PROCEDURE


Special Instructions
Assembly/Test Procedure
Acceptance Test Procedure
Calibration Procedure
Preshipment checks/Packout Procedure
290/290G Acceptance Test Sheet
293/293G Acceptance Test Sheet

7
7
8
10
11
13
15

ILLUSTRATED PARTS LIST


290/290G Assembly
293/293G Assembly
Frame Assembly
Sled Assembly
Sled Holder

18
22
24
26
27

THEORY OF OPERATION
Detailed Discussion

28

SCHEMATIC DIAGRAM
Wiring Diagram

30

________________________________________________________________________
Page 4 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

TROUBLESHOOTING GUIDE
Please refer to Operators Manuals for Normal Operation:
290 Operators Manual
78-8078-8977-5
Literature Code 70-2009-4089-1
290G Operators Manual
78-8083-0136-6
Literature Code 70-2009-4687-2
In the event of an error code, unplug the Attest 290/290G Auto-reader and plug it back in to initiate the
self-diagnostic test. If error code re-occurs, customer should call for repair or replacement.
The display panel will indicate malfunctions detected by a specific error code. Refer to the following:
Code

Problem

Solution

C1

Unit Warm-up Caution


Incubation block not at temperature
RRBI readings are disabled until
The incubator block reaches:
290 60C +/-2C
290G 37C +/-2C

Allow a 30 minute warm-up after


Attest 290/290G Auto-reader is plugged in.

C2

RRBI has been removed before incubation


Replace RRBI in proper well within 10 seconds.
completed. All three lights for affected RRBI
will light up and alarm will sound if active.

E1

Temperature Control Error


This error will occur if system is no
longer able to control the temperature of
the incubator block.
290 60C +/-2C
290G 37C +/-2C

Check heater blanket connection

E3

Memory Error.

Replace NMOM PCB

E4

Analog-Digital Conversion Error.

Replace Sled PCB

E7

Communications Error
This error will occur if the internal
communications have failed.

Check ribbon cable connection between


NMOM and Sled PCB. In addition, check for
power interruptions which could cause
communications error, E7

E8

Bulb Error
This error occurs if the system detects
that the bulb is too dim to properly read
RRBI results.

Check for NMOM Software V1.2


Replace Sled PCB / old bulb

E9

Transport Error
This error will occur if the system detects
a motor failure or positioning problem.

Check for loose set screw on sled


Check for loose set screw on motor drive pulley
Check for loose drive belt
Check bearings for smooth operation
Check for power interruptions which could
cause communications error which could
cause E9

________________________________________________________________________
3M2009 78-8078-9168-0
Page 5 of 38

Attest Auto-reader

Service Manual

TROUBLESHOOTING GUIDE
Trouble Shooting Guide (Continued)
Other Problem

Solution

Red, yellow, or green lights do not


illuminate when expected such as
following insertion of a RRBI or when
time remaining button is pressed.

Specific light burned out replace Display PCB


Check ribbon cable connection between
NMOM and Display PCB

LED flash test does not function

Check alignment and connection of


lid switch

Unwanted alarm sounds when


positive detected.

Turn alarm off, refer to features


Section.

All three lights blink and buzzer sounds.


RRBI has been removed from
incubation/reader well prematurely.

Replace RRBI in incubation/reader


well within 10 seconds to prevent
loss of data collected.

Negative Control Vial


Vial not crushed.

Crush vial before inserting into well.

Media did not wet the spore strip.

Tap vial on table until media wets spore strip.

Control vial was sterilized.

Use a non-sterilized vial as a control.


Check chemical indicator on label.

Unexplained positive
Adhesive residue on vial.

Do not place tape or labels on vial.

Test vial is absorbing fluorescent


residue from a chemical indicator or
tape.

Place the vial so it does not come in


direct contact with chemical
indicators or tape.

RRBI placed into a well immediately


following removal of previous BI.

Allow time for previous BI result to clear.

Inconsistent Read Time


RRBI not inserted all the way.

Ensure RRBI is inserted all the way into the


reader well.

RRBI placed into a well immediately


following the removal of previous BI.

Allow time for previous BI result to clear.

Intermittent Power Connection

Spread split pin on NMOM power jack to


ensure snug fit with plug.

________________________________________________________________________
Page 6 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

ACCEPTANCE TEST PROCEDURE

1.0 DESCRIPTION: Attest Auto-reader Final Acceptance Specification and Packout


1.1

This procedure covers the following Auto-readers:

1.1.1
1.1.2
1.1.3
1.1.4
1.1.5
1.1.6
1.1.7
1.1.8
1.1.9
1.1.10
1.1.11
1.1.12
1.1.13
1.1.14
1.1.15
1.1.16

Finished Goods #

Description

70-2007-2980-7
70-2007-2981-5
70-2007-2982-3
70-2007-2983-1
70-2007-3170-4
70-2007-3171-2
70-2007-3172-0
70-2007-3173-8
70-2007-4567-0
70-2007-4569-6
70-2007-4571-2
70-2007-4573-8
70-2007-4568-8
70-2007-4570-4
70-2007-4572-0
70-2007-4574-6

290 Attest Auto-reader 100VAC


290 Attest Auto-reader 120VAC
290 Attest Auto-reader 230VAC
290 Attest Auto-reader 240VAC
290G Attest Auto-reader 100VAC
290G Attest Auto-reader 120VAC
290G Attest Auto-reader 230VAC
290G Attest Auto-reader 240VAC
293 Attest Auto-reader 100VAC
293 Attest Auto-reader 120VAC
293 Attest Auto-reader 230VAC
293 Attest Auto-reader 240VAC
293G Attest Auto-reader 100VAC
293G Attest Auto-reader 120VAC
293G Attest Auto-reader 230VAC
293G Attest Auto-reader 240VAC

2.0 SPECIAL INSTRUCTIONS:


2.1

SAFETY PRECAUTIONS: Follow facility general safety guidelines for electronic


assembly and testing.

2.2

Standards and specifications given on assembly and component


drawings must be met.

2.3

Hardware fasteners must be fully seated and tight.

2.4

Exterior surfaces must be free from scratches, sharp edges and burrs. Any paint touch-ups
must be neat and blend well with surrounding surface.

2.5

Wiring should be held neatly in place. Wiring should not rub against sharp edges such as
sheet metal ends or screw threads.

2.6

Switches must operate without binding.

2.7

The person performing assembly operations is responsible for meeting the print
requirements. The checksheet is to be filled out by the assembler for each machine.

2.8

The location of the labels is defined on the prints. The labels are to be mounted straight,
without bubbles or lifted edges.

2.9

If a machine fails a test, indicate on the checklist the failure and repair. The unit must be
tested again after the repair for that function. The results of the test after the repair must
be recorded.

2.10

Any/all unexpected error codes displayed during testing must be resolved before
continuing.

________________________________________________________________________
3M2009 78-8078-9168-0
Page 7 of 38

Attest Auto-reader

Service Manual

3.0 ASSEMBLY/TEST PROCEDURE:


3.1

In-Process tests
3.1.1 General Workmanship
3.1.1.1 Screw tightness, solder joint quality, wire connections etc
3.1.2 Ribbon Cable Assembly
3.1.2.1 Ensure ribbon cables are inserted fully and routed and secured properly.
3.1.3 Heater Blanket Application
3.1.3.1 Clean outside surface of the incubator block with isopropyl alcohol
prior to attaching the blanket.
3.1.3.2 The blanket must be laid flat with no lifting of the blanket away from
the incubator block. (Inspection to be done prior to placing insulation
on unit.)
3.1.4 Home Sensor Adjustment
3.1.4.1 Remove outer case.
3.1.4.2 Measure and mark the center of the aluminum optics housing with a
pencil or felt-tip pen.
3.1.4.3 Apply power to unit and allow sled to move to home position.
3.1.4.4 View from top of incubator block and ensure that the center of the
optical housing is directly in line with the center of Well #12.
3.1.4.5 If optics is not centered, adjust position of home sensor vane as
necessary and cycle power to re-home sled position.
3.1.5 Sled Spacing Across Incubator Block
3.1.5.1 Unplug motor from NMOM board and slide sled to home position.
3.1.5.2 Adjust spacing between sled optics and edge of incubator block to be
0.040 (+ 0.00, - 0.025). Tighten screw to secure incubator block to
frame.
3.1.5.3 Slide sled in front of well #1. Adjust spacing between sled optics and
edge of incubator block to be 0.040 (+ 0.00, - 0.025). Tighten screw
to secure incubator block to frame.
3.1.5.4 Slide sled slowly back to home position, ensuring that optics do not
touch incubator block at any point along the full range of travel. At
home position, verify spacing between sled optics and edge of
incubator block to be 0.040 (+ 0.00, - 0.025). If not within
specification, go to 7.2.4.2 and repeat process.
3.1.5.5 Plug motor cable back in to connector.

________________________________________________________________________
Page 8 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

STEPS TO REMOVE COVER OF 293/293G AUTO-READER:


1.

Remove 4 screws from back cover

2.

Remove 8 screws used to mount the top cover

________________________________________________________________________
3M2009 78-8078-9168-0
Page 9 of 38

Attest Auto-reader

3.
4.

Service Manual

Lift RIGHT side of top cover. Display Ribbon cables are connected on LEFT side of top cover.
Un-plug or Re-plug Display Ribbon cables as needed.

LIFT COVER

RIBBON CABLES

ROTARY SWITCHES

________________________________________________________________________
Page 10 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

5.
6.

Service Manual

Set rotary switches for all 3 tiers to setting 0 for functional testing.
Unplug USB-to-RS232 adapter at the RS232 end and connect computer serial cable for functional
testing of each tier (top, center, bottom).

USB to RS232 ADAPTER

7.

After testing:
- Re-plug USB-toRS232 adapters for each tier
- Re-set all three rotary switches based on labeling (1292 or 1294)
Ensure all Display Ribbon Cables are securely connected

Proper Repairs Prior to Shipping the 293/293G


293/293G Frame Top View
Drill holes on ends as shown for mounting stand-offs.
-

________________________________________________________________________
3M2009 78-8078-9168-0
Page 11 of 38

Attest Auto-reader

Service Manual

Proper Repairs Prior to Shipping the 293/293G Continued:


Auto-reader Display Ribbon Cable Disconnection Resolution.
Attach from ribbon cable to ribbon cable connector.

Proper Repairs Prior to Shipping the 293/293G includes:


Add Dual Locks, Stand-offs, lock nuts and screws as shown below. Reference 293/293G
assembly drawing for full blow-up of stand-offs, etc.

________________________________________________________________________
Page 12 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

293/293G ASSEMBLY
ITEM
NUM.

PART
NUMBER

DESCRIPTION

QUANTITY
REQUIRED

1
2
9
39
40
41
42

78-8083-1005-2
78-8083-1004-5
26-1017-1620-2
70-0705-3010-3
70-0707-4015-7
26-1017-6006-9
26-1017-3845-5

3M 8422 Polyester Ribbon Connector Tape


Alignment pins, for shipping
Screw, 8-32 x 0.5, Pan, Phil, Blk Zn
Dual Lock SJ3263 Round 13/16
Dual Lock SJ3550 Square 1 by 1
Standoff, 8-32, M/F Hex, 1.25
8-32 KEPS Lock-nuts, Zinc, Hex
1

3
4
6
6
6
6
6

For Shipping 293/293G:


Note: Alignment Pins to be placed in well #1 and #12 for each heater block.

4.0 ACCEPTANCE TEST PROCEDURE:


Note: This procedure to be done for 100% of units. For 293/293G Auto-readers, perform each test for each
of the three tiers of incubators (top, center, bottom).
4.1 Error Code Test:
4.1.1
4.1.2
4.1.3

4.1.4

4.1.5

Disconnect power and remove cover.


Set rotary switch to position 1.
Generate a failure condition by one of the following 2 methods:
4.1.3.1. Disconnect heater blanket.
4.1.3.2. Disconnect motor wire connector and move sled off home position
toward Well #1.
Apply power to the unit and verify the error code:
4.1.4.1. "E1" for temperature failure.
4.1.4.2. "E9" for motor movement failure.
Disconnect power, repair failure, apply power, and verify no error code is
displayed.

4.2 Display Verification:


4.2.1
4.2.2
4.2.3
4.2.4
4.2.5
4.2.6
4.2.7
4.2.8
4.2.9
4.2.10
4.2.11
4.2.12

Set rotary switch to position 1.


Apply power to the unit.
Verify 7-segment display shows "8.8." on initial power-up as a LED check.
Verify 7-segment display shows "F1" (1291 BIs) as a set-up mode indication.
7-segment display will show "C1", the warm-up caution code, if not up to
temperature.
Open cover on unit and ensure all red, yellow, and green LEDs light.
Insert a BI into any well of the test unit.
Press "time remaining" button and verify that "1H" is shown for the BI.
Wait for the unit to heat up and the first reading to be taken.
Remove the BI and verify that "C2" is displayed and the audible signal sounds.
Re-insert BI within 10 seconds and verify that caution code clears and audible
signal ends.
Press both buttons and verify that period is displayed on the 7-segment display
to signify the audible signal is disabled.

________________________________________________________________________
3M2009 78-8078-9168-0
Page 13 of 38

Attest Auto-reader
4.2.13
4.2.14
4.2.15

4.2.16
4.2.17

Service Manual
Remove the BI and verify that "C2" is displayed and the audible signal does not
sound.
Re-insert BI within 10 seconds and verify that caution code clears.
For 293/293G Auto-reader, press both buttons for EACH tier and verify that
period is toggled on/off on the 7-segment display of ALL THREE TIERS
SIMULTANEOUSLY.
Remove BIs.
Remove Power.

4.3 Automated Test System Setup, Manufacturing Test System (MTS)


4.3.1
4.3.2
4.3.3
4.3.4
4.3.5
Note:

Set rotary switch to position 0.


Plug serial cable into connector on unit.
Cycle power to unit and start Automated Test System Software or select File,
New if test system is already running.
Enter serial number of unit being tested.
Perform the following tests:

If any tests fail, re-test once before re-working unit.

4.4 Serial Communications


4.4.1
4.4.2

Select "Serial Communication" test and press "Test"


Verify Successful Result in status display box.

4.5 Lid Test (290/290G Only)


4.5.1
4.5.2
4.5.3
4.5.4

Select "Lid Test" test and press "Test" or press "Next Test".
Open lid and leave open when prompted and press "OK".
Close lid when prompted and press "OK".
Verify Successful Result in status display box.

4.6 BI Hole Test


4.6.1
4.6.2
4.6.3
4.6.4
4.6.5

Select "BI Hole Test" test and press "Test" or press "Next Test".
Make sure all wells are empty when prompted, close lid, and press "OK".
Insert BIs into all wells when prompted, close lid, and press "OK".
Verify Successful Result in status display box.
Remove all BIs from wells.

4.7 Config Switch Test


4.7.1
4.7.2

Select "Config Switch" test and press "Test" or press "Next Test".
Verify Successful Result in status display box.

4.8 Home Sensor Test


4.8.1
4.8.2

Select "Home Sensor" test and press "Test" or press "Next Test".
Verify Successful Result in status display box.

4.9 E2ROM Read/Write Test


4.9.1
4.9.2

Select " E2ROM Read/Write" test and press "Test" or press "Next Test".
Verify Successful Result in status display box.

4.10 Servo Reading Test


Note: This test uses "Fluorescent Standards, Model 290", part number 78-8078-8421-4

________________________________________________________________________
Page 14 of 38
3M2009 78-8078-9168-0

Attest Auto-reader
4.10.1
4.10.2
4.10.3
4.10.4
4.10.5

Service Manual
Select "Servo Reading" test and press "Test" or press "Next Test".
Insert Low Standard into Well #12 when prompted, close lid, and Press "OK".
Insert High Standard into Well #12 when prompted, close lid, and Press "OK".
Verify completion of test in status display box.
Remove fluorescent standard from unit.

4.11 Motor Movement Test


4.11.1
4.11.2

Select "Motor Movement" test and press "Test" or press "Next Test".
Verify Successful Result in status display box.

4.12 Motor Slip Test


4.12.1
4.12.2

Select "Motor Slip" test and press "Test" or press "Next Test".
Verify Successful Result in status display box.

4.13 Fluorescent Standards Verification


4.13.1
4.13.2
4.13.3
4.13.4
4.13.5
4.13.6

Press "View Report" icon on toolbar.


Verify Low Fluorescent Standard Std Dev is less than 2.
Verify High Fluorescent Standard Std Dev is less than 4.
Verify High Fluorescent Standard Average is less than 250.
Verify difference between High and Low averages is greater than 110.
Close Report.

5.0 CALIBRATION PROCEDURE:

5.1 Temperature Set Point Verification


Use a traceable temperature measurement system and appropriate temperature probe that
makes good contact with incubator block walls to verify temperature in Wells #6 and #12
are within the range including:
Model 290/293
59.0C 61.0C
Model 290G/293G
36.0C 38.0C
Measure temperatures for an appropriate amount of time after temperature has stabilized
to ensure temperature remains within specification.
NOTE: Recommended measurement system or, similar: Fluke 77 Multi-meter and
Fluke 80K Thermocouple Module.
Temperature probe assembly, 3M stock number 78-8078-8982-5
5.2 Fluorescent Standards Verification
Fluorescent standards (78-8078-8421-4) intensity should be verified every 2 year (24
months) according to the procedure outlined on "Calibration Procedure for Fluorescent
Standards", drawing number 78-8078-9589-7.

6.0 PRESHIPMENT CHECKS/PACKOUT PROCEDURE:


6.1 Filled out and signed Machine Check-sheet
6.2 Labeling
All labeling (serial plate, warning, instructions etc.) to be legible. All labeling to be installed
per print requirements. (Adhered to unit or placed inside of box etc.)
6.3 Preparation for Shipping

________________________________________________________________________
3M2009 78-8078-9168-0
Page 15 of 38

Attest Auto-reader

6.3.1

Service Manual

Place rotary switch to appropriate position based on the BI label present on the
unit according to the following table. Apply Power and verify the correct setup
mode is displayed on 7-segment display according to the following table.
Setup Mode

1291 BIs, Blue Label


1292 BIs, Brown Label
1291/1292 BIs, Blue/Brown Label
1294 BIs, Green Label

Rotary Switch
Position
1
3
7
d

7-Segment Power-up
Display
F1
F3
F7
Fd

6.3.2

With unit unplugged, insert a small flat blade screwdriver tip (Apex 00 or
equivalent) into the center split pin of the power jack on the NMOM PCB.
Spreading this split pin will ensure a snug fit of the power plug upon receipt by
customer.

6.3.3

Clean the machine of all smudges, fingerprints or water.

________________________________________________________________________
Page 16 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual
290/290G Acceptance Check Sheet

Approvals:
_______________________
Service

page 1 of 2

_______
Date

Serial Number _______________________


Model Info (Check Appropriate)
Finished Goods #

Description

70-2007-2980-7
70-2007-2981-5
70-2007-2982-3
70-2007-2983-1
70-2007-3170-4
70-2007-3171-2
70-2007-3172-0
70-2007-3173-8

290 Attest Auto-reader 100VAC


290 Attest Auto-reader 120VAC
290 Attest Auto-reader 230VAC
290 Attest Auto-reader 240VAC
290G Attest Auto-reader 100VAC
290G Attest Auto-reader 120VAC
290G Attest Auto-reader 230VAC
290G Attest Auto-reader 240VAC

In process test
characteristics

Test Results
pass/fail

1) General
Workmanship
2) Ribbon Cable
3) Heater Blanket
4) Home Sensor
Adjustment
5) Sled Spacing
Across Bock
6) Error Code Test
7) Display Verification

________________________________________________________________________
3M2009 78-8078-9168-0
Page 17 of 38

Attest Auto-reader

Service Manual
290/290G Acceptance Check Sheet

In process test
characteristics

page 2 of 2

Test Results
pass/fail

8) Fluorescent
Standards
Low Standard StdDev <2
High Standard StdDev <4
High Standard Avg. <250
Avgerage Difference >110?
Low to High Std
9) Temperature
Verification
10) Documentation
11) Labeling
12) Workmanship
13) Rotary Switch
14) No Smudges,
Fingerprints, etc.
15) Final Packed per
Standard

Comments/Repair Information/New Test Results:


______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________

________________________________________________________________________
Page 18 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual
293/293G Acceptance Check Sheet

Approvals:
_______________________
Service

page 1 of 2

_______
Date

Serial Number _______________________


Model Info (Check Appropriate)
Finished Goods #

Description

70-2007-4567-0
70-2007-4569-6
70-2007-4571-2
70-2007-4573-8
70-2007-4568-8
70-2007-4570-4
70-2007-4572-0
70-2007-4574-6

293 Attest Auto-reader 100VAC


293 Attest Auto-reader 120VAC
293 Attest Auto-reader 230VAC
293 Attest Auto-reader 240VAC
293G Attest Auto-reader 100VAC
293G Attest Auto-reader 120VAC
293G Attest Auto-reader 230VAC
293G Attest Auto-reader 240VAC

In process test
characteristics

Test Results pass/fail


TOP TIER

Test Results pass/fail


CENTER TIER

Test Results pass/fail


BOTTOM TIER

1) General
Workmanship
2) Ribbon Cable
3) Heater Blanket
4) Home Sensor
Adjustment
5) Sled Spacing
Across Bock
6) Error Code Test
7) Display Verification

________________________________________________________________________
3M2009 78-8078-9168-0
Page 19 of 38

Attest Auto-reader

Service Manual
293/293G Acceptance Check Sheet

In process test
characteristics

Test Results pass/fail


TOP TIER

Test Results pass/fail


CENTER TIER

page 2 of 2

Test Results pass/fail


BOTTOM TIER

8) Fluorescent
Standards
Low Standard StdDev <2
High Standard StdDev <4
High Standard Avg. <250
Avgerage Difference >110?
Low to High Std
9) Temperature
Verification
10) Documentation
11) Labeling
12) Workmanship
13) Rotary Switch
14) No Smudges,
Fingerprints, etc.
15) Final Packed per
Standard

Comments/Repair Information/New Test Results:


______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________

________________________________________________________________________
Page 20 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

Illustrated Parts List

________________________________________________________________________
3M2009 78-8078-9168-0
Page 21 of 38

Attest Auto-reader

Service Manual

290/290G ASSEMBLY
ITEM
NUM.

PART
NUMBER

DESCRIPTION

1
2
3
4
5
7
8
9
10
11
12
13
14
15
16
16A
16B

78-8083-0098-8
78-8078-8675-5
26-1002-5214-2
78-8083-0207-5
78-8078-8668-0
78-8083-0086-3
26-1004-2652-2
70-0700-7979-6
26-1002-5195-3
78-8078-8756-3
26-1000-4179-2
78-8083-0195-2
78-8078-0402-2
78-8083-0087-7
26-1012-1254-1
26-1014-7689-8
26-1017-3696-0
78-8078-8764-3
78-8078-8773-8

MEMBRANE PANEL OVERLAY


DISPLAY ASSY
SCREW, MACH,6-32 X.250, PAN, PHIL
ENCLOSURE ASSY
FRAME ASSY
BOTTOM PLATE
WASHER, FLAT #6
BUMPER,3M
SCR 2-56 X .188, MACH PAN ZN
LID SWITCH ASSY
NUT, MACH, 2-56, ZN
LID SENSOR BRACKET
LABEL-SERVICE CENTER
COVER, BACK
CLAMP,RICHO, CFCC-3-05A-RT
CLIP for FLAT CORD, WHITE
CLIP for ROUNDED CORD
DECAL, CAUTION
SERIAL PLATE, AUTO READER

17

N/A
N/A

N/A
N/A
N/A
N/A
N/A
N/A
N/A

QUANTITY
REQUIRED

1
1
12
1
1
1
3
6
2
1
2
1
1
1
1
1
1
1
1

________________________________________________________________________
Page 22 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

290/290G ASSEMBLY
ITEM
NUM.

PART
NUMBER

18
19
20
21
22
23
24
26

78-8078-8878-5
26-1004-3301-5
26-1017-0657-5
26-1014-7284-8
78-8078-9170-6
26-1014-7450-5
26-1012-1254-1
26-1017-2681-3

DESCRIPTION

N/A

N/A
N/A

QUANTITY
REQUIRED

LABEL, SHOCK/EXCLAMATION
SCREW, 2-56 X .125
SCREW, 6-19 X .250, PAN, PHIL
CABLE, ELEK PARALEX 1.00 MM
CARD,INSERT
ENVELOPE, CLEAR, 2.5 X 4.0
CLAMP,RICHO, CFCC-3-05A-RT
SCREW, MACH, 4-40 X .188, PHIL

1
2
1
1
1
1
1
1

N/A INDICATES PART NOT AVAILABLE

CONNECTIVITY KIT, Auto-reader, New Style: Connectivity Kit is necessary to link


the 290 Auto-readers to a PC with Merlin SW (Microsystems SPM).
1
2
3

78-8083-1166-2
78-8083-1167-0
78-8083-1211-6

Connectivity Kit
Cover-plate
Instruction Sheet

________________________________________________________________________
3M2009 78-8078-9168-0
Page 23 of 38

Attest Auto-reader

Service Manual

290/290G ASSEMBLY (OLD STYLE CASE)


ITEM
NUM.

PART
NUMBER

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
18
19

78-8078-8772-0
78-8078-8675-5
26-1002-5214-2
78-8078-8869-4
78-8078-8759-7
78-8078-8668-0
78-8078-8761-3
26-1002-5216-7
70-0700-7979-6
26-1002-5206-8
78-8078-8756-3
26-1003-4052-5
78-8078-8760-5
78-8078-0402-2
78-8078-8880-1
78-8078-8764-3
78-8078-8773-8
78-8078-8878-5

DESCRIPTION

N/A

N/A
N/A
N/A
N/A

N/A
N/A
N/A
N/A
N/A
N/A

QUANTITY
REQUIRED

MEMBRANE PANEL OVERLAY


DISPLAY ASSY
SCREW, MACH,6-32 X.250, PAN, PHIL
MASK,DISPLAY BOARD
ENCLOSURE
FRAME ASSY
BOTTOM PLATE
SCREW, MACH, 6-32 X .375, PAN PHIL
BUMPER,3M
SCR 4-40 X .375, MACH PAN ZNC
LID SWITCH ASSY
NUT, MACH, 4-40, KEPS
LID SENSOR BRACKET
LABEL-SERVICE CENTER
COVER, BACK
DECAL, CAUTION
SERIAL PLATE, AUTO READER
LABEL, SHOCK/EXCLAMATION

1
1
5
2
1
1
1
12
6
12
1
4
1
1
1
1
1
1

________________________________________________________________________
Page 24 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

290/290G ASSEMBLY (OLD STYLE CASE)


ITEM
NUM.

PART
NUMBER

DESCRIPTION

20
21
22
23
24
25
26

26-1002-5204-3
26-1014-7284-8
78-8078-9170-6
26-1014-7450-5
26-1014-7495-0
26-1012-1254-1
78-8125-7843-9

27

78-8078-8978-3

SCREW, MACH, 4-40 X .250, PAN, PHIL


CABLE, ELEK PARALEX 1.00 MM
CARD,INSERT
ENVELOPE, CLEAR, 2.5 X 4.0
WASHER,ASM CO.,
CLAMP,RICHO, CFCC-3-05A-RT
BI ID LABEL KIT 1291, 1292,
& 1291/1292 (NOT SHOWN)
BI ID LABEL 1294 (NOT SHOWN)

N/A
N/A

QUANTITY
REQUIRED

4
1
1
1
2
1
1
1

N/A INDICATES PART NOT AVAILABLE

CONNECTIVITY KIT, Auto-reader, Old Style: Connectivity Kit is necessary to link the
290 Auto-readers to a PC with Merlin SW (Microsystems SPM).
1
2
3

78-8083-1168-8
78-8083-1169-6
78-8083-1211-6

Connectivity Kit
Cover-plate
Instruction Sheet

________________________________________________________________________
3M2009 78-8078-9168-0
Page 25 of 38

Attest Auto-reader

Service Manual

293/293G ASSEMBLY
ITEM
NUM.

PART
NUMBER

1
2
2A
3
4
6
7
8
9
10
11
12
13
14
15
19
20
21
22

78-8078-8668-0
78-8083-0380-0
78-8083-0381-8
78-8078-8772-0
26-1003-6373-3
78-8083-0306-5
78-8083-0307-3
26-1002-5226-6
26-1017-1620-2
78-8083-0303-2
78-8083-0396-6
78-8083-0305-7
78-8083-0304-0
78-8083-0392-5
70-0700-7979-6
78-8083-0395-8
78-8083-0394-1
26-1007-4847-9
78-8083-0391-7

DESCRIPTION

N/A

QUANTITY
REQUIRED

FRAME ASSY
3
LABELS, BI WELLS 1294
3
LABELS, BI WELLS 1292
3
MEMBRANE PANEL OVERLAY
3
SCREW, MACH,6-32 X.375, PAN, PHIL 12
END CAPS, PAIR
1
PLATE, BACK
1
SCREW,MACH,8-32 X.375, PAN, PHIL 10
SCREW,MACH,8-32 X.5,PAN, PHIL, BLK14
CASE
1
DECAL, FRONT
1
BRACKETS, PAIR
1
BASE PLATE
1
POWER CABLE ASSY
1
BUMPER,3M
6
CONVERTER, SERIAL TO USB
3
USB HUB;
1
TAPE, DBL COATED, 1W 3M VHB
AR
CABLE, FLAT, RS232
3

________________________________________________________________________
Page 26 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

293/293G ASSEMBLY
ITEM
NUM.

PART
NUMBER

23
24
25
26
27
29
30
31
32
33
34

78-8078-8869-4
26-1002-5214-2
78-8078-8675-5
78-8083-0418-8
26-1017-6304-8
78-8083-0437-8
26-1017-2068-3
26-1017-2787-8
26-1002-5204-3
26-1003-4052-5
26-1009-7475-2

DESCRIPTION

N/A

QUANTITY
REQUIRED

MASK,DISPLAY BOARD
SCREW, MACH, 6-32 X .25, PAN PHIL
DISPLAY ASSY
SPACER, THERMAL INSUL
CABLE, ELEK PARALEX 1.00 MM
WIRE HARNESS, DISPLAY
USB PANEL MOUNT JACK
DUST COVER, USB
SCREW, MACH, 4-40 X .25, PAN PHIL
NUT, 4-40, KEPS
TIE WRAP, #10 MT HOLE

6
12
3
3
3
1
1
1
2
2
1

N/A INDICATES PART NOT AVAILABLE

________________________________________________________________________
3M2009 78-8078-9168-0
Page 27 of 38

Attest Auto-reader

Service Manual

FRAME ASSEMBLY
ITEM
NUM.

PART
NUMBER

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16

78-8078-8389-3
26-1002-5218-3
26-1014-7498-4
26-1002-5216-7
78-8078-8386-0
26-1011-8385-8
78-8078-8758-9
26-1002-5204-3
26-100508084-9
78-8078-8717-5
26-1004-7298-8
26-1014-7496-6
78-8078-8725-8
78-8078-8766-2
26-1014-7300-2
78-8078-8667-2

DESCRIPTION

N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A

QUANTITY
REQUIRED

NMOM ASSYY
SCREW, MACH, 6-32 X .500, PAN, PHIL
SPACER,ASM CO, 3031R-F-.275,.250
SCREW,MACH,6-32 X .375, PAN, PHIL
FRAME
SCR, M3 X .5 X 5.0,SKT,HEX SKT
SLED LOCATOR
SCREW,MACH,4-40 X .250,PAN, PHIL,
NUT, MACH,4-40, KEPS
STEPPER MOTOR
PULLEY, BERG MFG, TP7A5MW2-15,
WASHER, ASM CO., 99B-.062-D, FLAT,
SPACER, INCUBATOR BLOCK
INCUBATOR BLOCK ASSY
SPACER,ASM CO., 3031R-D-.312, .250
SLED HOLDER ASSY

1
5
3
2
1
4
1
2
4
1
1
2
1
1
2
1

________________________________________________________________________
Page 28 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual
290 FRAME ASSEMBLY

ITEM
NUM.

PART
NUMBER

17
18
19
20
21
22
23
24

78-8078-8714-2
78-8078-8716-7
26-1014-7125-3
78-8078-8715-9
78-8656-4008-6
78-8078-8703-5
26-1014-7301-0
78-8078-8905-6

DESCRIPTION

N/A
N/A
N/A
N/A
N/A
N/A
N/A

QUANTITY
REQUIRED

IDLER PULLY SPRIG


IDLER PULLEY
RETAINER RING, WALDES TRUARC,
SHAFT, IDLER PULLY
RETAINING RING, TRUARC,5133-18-H
GUIDE ROD
BELT, TIMING, B.C. OF A.,MXL270T80
PLATE, CAP

1
1
2
1
4
2
1
1

N/A INDICATES PART NOT AVAILABLE

________________________________________________________________________
3M2009 78-8078-9168-0
Page 29 of 38

Attest Auto-reader

Service Manual

SLED ASSEMBLY
ITEM
NUM.

PART
NUMBER

1
2
3
4
5
6
7
8
9
10
11
12
12A
13

78-8078-8211-9
78-8000-7655-2
26-1014-7302-8
78-8078-8670-6
78-8078-8671-4
78-8078-8674-8
78-8078-8312-5
78-8078-8669-8
26-1000-4179-2
78-8078-8673-0
78-8078-8672-2
26-1017-2781-1
26-1014-7272-3
26-1004-3301-5

DESCRIPTION

N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A

SLED ASSEMBLY
SCR, 2-56 X 1.0
O-RING
SIDE COVER, LEFT
GLASS EXCITATION FILTER
MASK
REFLECTOR
SIDE COVER, RIGHT
NUT, 2-56 X
GLASS EMISSION FILTER
GELATIN EMISSION FILTER
LAMP
LAMP, ALTERNATE
SCR.2-56 X .125

QUANTITY
REQUIRED

1
1
2
1
2
1
1
1
1
1
1
1
0
2

N/A INDICATES PART NOT AVAILABLE

________________________________________________________________________
Page 30 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

SLED HOLDER ASSEMBLY


ITEM
NUM.

PART
NUMBER

1
2
3
4
5
6
7
8
9
10
11
12
13

78-8078-8704-3
26-1014-7294-7
26-1014-7295-4
26-1014-9893-4
78-8078-8707-6
78-8078-8706-8
26-1011-5909-8
78-8078-8705-0
78-8078-8710-0
78-8078-8666-4
26-1002-5206-8
26-1014-7293-9
26-1014-7296-2

DESCRIPTION

N/A
N/A
N/A
N/A
N/A
N/A
N/A

N/A

QUANTITY
REQUIRED

GUIDE PLATE
BEARING, OILITE,AA-309-4
SPACER, ASM CO., 14030-H-A-.875-3
CABLE, FLEX
BELT TENSIONER PIN
BELT LOCKING PIN
SCR, SET, 2-56 X .125
PCB CLAMP
STRAIN RELIEF
SLED ASSEMBLY
SCR, 4-40 X 3/8 MACH PAN PHIL ZNC
SCR, SET, 4-40 X .375
SPACER, ASM CO., 14030-H-A-1.125-3

1
4
2
1
1
1
1
1
1
1
4
1
2

N/A INDICATES PART NOT AVAILABLE

________________________________________________________________________
3M2009 78-8078-9168-0
Page 31 of 38

Attest Auto-reader

Service Manual

3M Attest Model 290/290G Theory of Operation


Theory of Operation
This section will describe mechanical, electrical and software component aspects of the 3M Attest
290/290G Auto-reader/Incubator.
A. Electro-Optics Apparatus
The sled board contains the electro-optics apparatus for the 3M Attest 290/290G Auto-reader. Inside
the enclosed extrusion are two photo sensors and a bulb, or excitation source. Light from the excitation
source is shown on and reflected off the Biological Indicator, filtered to the desired wavelength, and a
proprietary embedded control system accurately measures the fluorescent activity. The embedded
control system also monitors the bulb intensity. This apparatus should not be disassembled unless the
manufacturing guidelines are followed.
B. Sweep Measurement Timing
The measurements of relative fluorescence for each biological indicator are internally timed by a
micro-controller to take place at the same on time for all readings. The fluorescent measurement
time and the amount of time to transport the elecro-optics to another biological indicator well is 2
seconds. See Table 1 below.
Reader/
incubator Well #

Order
Read

12
11
10
9
8
7
6
5
4
3
2
1

1st
2nd
3rd
4th
5th
6th
7th
8th
9th
10th
11th
12th

Time after bulb turns


on to fluorescence
measured
T0
T0+2
T0+4
T0+6
T0+8
T0+10
T0+12
T0+14
T0+16
T0+18
T0+20
T0+22

Table 1
Since the excitation sources output intensity varies with respect to time on and the proprietary
embedded control systems ability to minimize the output variations, this timing technique assures that
each biological indicator placed into a reader incubator well will be read in the same excitation source
on time with a variation of less than +/- 1% of the total specified readout time.
C. Measuring Fluorescent Intensity Increases.
Design considerations for the Electro-Optics apparatus were made on the assumption that the intensity
increase of fluorescence would be measured. Since background fluorescence can vary between BI
vials, this measurement of fluorescent intensity increase eliminates the requirement for BIs to have
similar background fluorescent measurements. Furthermore, since the detection of positive and
negative BIs will be based on the change in fluorescence during incubation instead of fixed threshold
values, this measurement method eliminates the need for daily calibration.
D. Excitation Intensity
Each time a biological indicator is read by the electro-optics apparatus, the excitation source will
remain on for a fixed length of time, regardless of the number or location of the biological indicators.

________________________________________________________________________
Page 32 of 38
3M2009 78-8078-9168-0

Attest Auto-reader

Service Manual

After this on time, the excitation source intensity will be measured. This technique will allow the
ability to measure the output intensity of the excitation source after it has been operated, and stabilized.
If this output intensity is less than a chosen set-point, then the excitation source failure determination
can be ascertained. Error message (E8) will be displayed for the user if the excitation source is too dim
to accurately measure fluorescent intensity.
E. Sled Positioning.
The Electro-Optics Apparatus is mounted on a moveable platform called a Sled. The Sled position is
referenced to a home position via a sensor mounted on the sled and a vane mounted on the frame. This
establishes the home position of the sled. The home position is determined automatically at power-up
as well as when a series of BI readings are completed. Positioning is accomplished with a stepper
motor as an open loop system. The motor is given a certain number of steps to move the Sled from
one well position to the next. The number of steps moved away from the home position is
accumulated, and when the Sled returns home, the number of steps needed to move home is subtracted
from the accumulated value. A difference of more than 3 steps between the two values 3 consecutive
times will result in the system preventing normal operation, and an error message (E9) will be
displayed for the user. If the difference is more than 3 steps, the readings will be ignored and retaken.
F. BI Insertion Detection.
BI presence in each Incubator Block Well is determined using a photoelectric sensor mounted on the
NMOM board and placed next to each well. When the unit detects the cover open, the BI detection
sensors are read and the A/D value is compared to a setpoint. A value above the setpoint indicates the
presence of a BI. A value below the setpoint indicates an empty well.
G. Incubator Block Temperature Control.
Incubator Block temperature is maintained by a heater blanket mounted on the Incubator Block. A
thermistor on the heater blanket is measured with an A/D converter and the value is compared to a
setpoint. If the value is at least one count above the setpoint, the heater is turned off. If the value is at
least one count below the setpoint, the heater is turned on. The setpoint is saved in the EEPROM. The
value is automatically changed via the rotary setup switch for 60C or 37C operation. When the unit
is first turned on, a warm-up caution code (C1) is displayed and the unit is non-functional until the
setpoint is reached. During normal operation, a high or low temperature condition will result in the
system preventing normal operation, and an error message (E1) will be displayed for the user.
H. Display.
The display consists of red, yellow, and green LEDs for each Incubator Well and dual 7-segment
displays. When a BI is inserted into a well, the yellow LED will light, indicating a BI in process.
After the full incubation time, if the BI result is negative, the green LED will light and remain on until
the BI is removed. If at any time during incubation a positive result is determined, the red LED will
light and remain on until the BI is removed. The buzzer will also sound if enabled, until the "buzzer"
button is pressed or the BI is removed. If a positive BI is removed then red LED will remain lit for
approximately 10 seconds, allowing the BI to be replaced for growth readout purposes. The 7-segment
display will display caution and error codes as well as time remaining for each BI if the "time
remaining" button is pressed. The LEDs all light in a test mode when the cover is opened. If a BI is
removed before incubation is complete, all LEDs for that well will light and the buzzer will sound for
10 seconds, with caution code C2 displayed, indicating the need to replace the BI. After 10 seconds,
the BI data is lost.

________________________________________________________________________
3M2009 78-8078-9168-0
Page 33 of 38

Attest Auto-reader

POWER
POWER.SCH

JUMPER INSTALLED FOR TEST PU

V+

-10V

-10V

EMRST*
EMINT*
EXRST*
EXINT*

VDD

EMRST*
EMINT*
EXRST*
EXINT*

+12V
+12V
+8V
-10V

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18

HOME
RST*
SLEDCS*

VEM
VEX

VEM
VEX

SDO
SDI
SCK

EXCITE
EXCITE.SCH

PINGMOM*
LAMP
INC*
U/D*
PTCS*
LAMPV
+12V

+12V

LAMP
INC*
U/D*
PTCS*
LAMPV

J4
FPC18

J3

DGND
PGND
JUMPER2X1

EC
EC.SCH

SENSOR
SENSOR.SCH

SLED

SLED INTERFACE

V-

NCOM
NCOM.SCH

NMOM

J1
1

VV-

J12

+8V

+8V

V+
V+

Service Manual

FPC

DGND

18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1

+12V

LO
CONY

VDD

+8V
-10V

RST*

TX
RX
CONY

SLEDCS*
HI
SDO

+8V
-10V

SCK

+8V
-10V

HEATH
HEATL
COOL
HOMELED*

VDD

HRTLO

SW1*
SW2*
SW3*
SW4*

SDI

MEMBRANE INTERFACE

SCK
PINGMOMD*

PGND
BILED
BILED.SCH
LEDEN1*
LEDEN2*
LEDEN3*
LEDEN4*
LEDEN5*
LSCK
LSDO
STORE
VLED

LEDEN1*
LEDEN2*
LEDEN3*
LEDEN4*
LEDEN5*
LSCK
LSDO
STORE

+12V
J1

SDO

DGND

+12V

ASEG
BSEG
CSEG
DSEG
ESEG
FSEG
GSEG
DPSEG
MSD
LSD
BUZ
VLEDEN

FFC12
DGND
DPOWER
DPOWER.SCH
ASEG
BSEG
CSEG
DSEG
ESEG
FSEG
GSEG
DPSEG
MSD
LSD
BUZ
VLEDEN

VDD

NTC

J5

1
2
3
4
5
6
7
8
9
10
11
12

DICS*

FPC CABLE

1
2
3
4
5
6
7
8
9
10
11
12

GLOBAL COMPATIBLE SWITCH MODE P


INPUT 90 < VAC < 264 , 47 < HZ < 62
OUTPUT 12VDC , 30W REGULATED

1
2
3
4

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

FPC8

DISPLAY INTERFACE

FFC5

VDD
RST*

1
2
3
4
5

EXTERNAL POWER SUPPLY

LINE
NEUTRAL
EARTH

MELCOR CP1.0-63-08L

MTE4
J7
INCUBATION HEATER/THERMISTOR

DISPLAY

J2

J2

OPTIONAL INCUBATION TEC

1
2
3
4

HEATH
HEATL
COOL
HRTHI

COUNTRY SPECIFIC PIN AD

PGND PC722A
DRC TRACE

J1
TECA
TECB

PC HOST

SUBD9PCST

DGND
JUMPER

NT
+12V

DEC
DEC.SCH

ELDR-35

1
6
2
7
3
8
4
9
5

J8
1

PINGMOMS*

HOMELED*

L1

STAR GROUND POINT

SDI

DGND
PGND

RXD

DGND
CONY NPOWER
NPOWER.SCH
LO
LO
+PS
HI

1
6
2
7
3
8
4
9
5

TXD

ASYNC SERIAL INTERFACE EIA-232D

NEC
NEC.SCH
SLEDHOME

FPC CABLE

PGND

TX
RX
LO
CONY

HEATER HI
T
THERMISTOR
HEATER LO
COPPER ETCH ON KAPTON

NT

RST*
DICS*

NBI
NBI.SCH

SDI
SCK
SDO
PUR*
SCK
OEXPCS*
LEDEN*
PINGMOMD* PUR*
BIHOMEA
BIHOMEB
BIHOMEC
BIHOMED
SDO

FPC12
DGND
PGND

SCK
SDO
PUR*
OEXPCS*
LEDEN*
PUR*
BIHOMEA
BIHOMEB
BIHOMEC
BIHOMED

I0
I1

M1
PHASE A1

NMOTOR
NMOTOR.SCH
VR
PHA
PHB

VR
PHA
PHB

I0
I1
MPHA1
MPHA2
MPHB1
MPHB2

J6
1
2
3
4
5
6
MTE6

J6
BLACK
BROWN
YELLOW
ORANGE

1
2
3
4
5
6

PHASE A2
PHASE B1

MTR STEP 2PH 4W 17FRAME

PHASE B2
JAPAN SERVO KH42HM2-801

AMP MTE

VLED

________________________________________________________________________
Page 34 of 38
3M2009 78-8078-9168-0

Certificate of Calibration
Serial Number: _____________________
Model Number: ____________________ 290 Auto-reader

____________________ 290G Auto-reader

The serial number listed above has been tested and found to comply with 3Ms manufacturing
specifications applicable to subject device at time of manufacture. It has been tested with NIST
traceable instrumentation to verify it meets the following specifications.
3M Test Procedure Number: C-05-005002
1)

Temperature Specification:
290
290G

2)

60C +/-2C
37C +/-2C

Well #6 Measurement

Well #12 Measurement

Relative Fluorescence Measurement Increase Specification:


Greater Than 110 Relative Fluorescence Units
Units

Measurement
3)

Fluorescence Measurement Standard Deviation Specification:


Less than 4 Relative Fluorescence Units
Units

Measurement

Test Equipment:
Equipment Used (Temperature):
Equipment ID:
Calibration Date:
Calibration Due:
Equipment Used (Fluorescence):
Equipment ID:
Calibration Date:
Calibration Due:

Fluorescent Standards
78-8078-8421-4

For factory calibration and re-certification of this equipment contact:


3M Health Care Service Center
Suite 200, Bldg. 502
3350 Granada Avenue North
Oakdale, MN 55128
1-800-292-6298
Fax: 1-800-770-8016
Outside of the USA, contact your local 3M Subsidiary.

Calibrated By: _________________________________

Calibration Date:

_________________________________

QC:

Date:

_________________________________

_________________________________

Certificate of Calibration
Serial Number: _____________________
Model Number: ____________________ 293 Auto-reader

____________________ 293G Auto-reader

The serial number listed above has been tested and found to comply with 3Ms manufacturing
specifications applicable to subject device at time of manufacture. It has been tested with NIST
traceable instrumentation to verify it meets the following specifications.
3M Test Procedure Number: C-05-005002
1)

Temperature Specification: 293 60C +/-2C, 293G 37C +/-2C


C
C
C
C
C
C

Well #6 Measurement
Well #12 Measurement
Well #18 Measurement
Well #24 Measurement
Well #30 Measurement
Well #36 Measurement
2)

Relative Fluorescence Measurement Increase Specification: Greater Than 110 Relative Fluorescence Units
Units
Units
Units

Top Tier Measurement


Center Tier Measurement
Bottom Tier Measurement

3)

Fluorescence Measurement Standard Deviation Specification: Less than 4 Relative Fluorescence Units
Units
Units
Units

Top Tier Measurement


Center Tier Measurement
Bottom Tier Measurement

Test Equipment:
_______________________
_______________________
_______________________
_______________________

Equipment Used (Temperature):


Equipment ID:
Calibration Date:
Calibration Due:
Equipment Used (Fluorescence):
Equipment ID:
Calibration Date:
Calibration Due:

Fluorescent Standards
78-8078-8421-4

For factory calibration and re-certification of this equipment contact:


3M Health Care Service Center
Suite 200, Bldg. 502
3350 Granada Avenue North
Oakdale, MN 55128
1-800-292-6298
Fax: 1-800-770-8016
Outside of the USA, contact your local 3M Subsidiary.

Calibrated By: _________________________________

Calibration Date:

_________________________________

QC:

Date:

_________________________________

_________________________________

3M Health Care
3M Center
Building 275-4E-01
St. Paul, MN 55144-1000
1-800-228-3957

3M Health Care
D-41453 Neuss, Germany

Recycled paper
40% pre-consumer
10% post-consumer
Litho in U.S.A. with 3M film,
proofing systems and offset
plates.
Attest is a registered trademark
of 3M.
3M 2009
78-8078-9168-0. Rev. D

S-ar putea să vă placă și