Laboratory

Excellent Services, B-34 & 35, Ganpati Paradise, Central Spine,


Vidhyadhar Nagar, Jaipur, Rajasthan

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

20.03.2014

Certificate Number

C-0703

Valid Until

19.03.2016

Last Amended on

Page

1 of 2

Quantity Measured/
Instrument

1.

I. TEMPERATURE
Temperature
Transmitter, RTDs,
Thermocouples with &
without / indicator, Glass
Thermometer,
Temperature Gauge,
Digital Thermometer#

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

-30C to 50C
50C to 250C

0.14C
0.27C

Using PRT 4 Wire,


Precision calibrator &
liquid oil bath
By Comparison Method

2.

Temperature
Transmitter, RTDs ,
Thermocouple with &
without indicator,
Temperature Gauge,
Digital Thermometer#

250C to 900C
900C to 1200C

2.00 C
2.40 C

Using S-Type
Thermocouple,
Precision calibrator &
Dry Block Furnaces
Calsys 650 &
Calsys 1200
By Comparison Method

3.

Infrared Thermometer/
Pyrometer/ optical
Thermometer#

50C to 500C
500C to1300C

2.5C
4.4C to 6.4C

Using Infrared
Thermometer & Black
Body Source
By Comparison Method

4.

Temperature Indicator of
Cold Chamber,
Conditioning Chamber/
BOD Incubator / Liquid/
Water Bath / Dry Block
Furnace/ Oven$

-30C to 250C

0.23C

Using RTD (Pt-100) 4


wire & Precision
calibrator for single
position Calibration
By Direct Method

5.

Temperature Indicator of
Furnace/ Oven/ Muffle
Furnace#

50C to1200C

2.4C

Using S-Type
Thermocouple with
Precision Calibrator for
single position
Calibration
By Direct Method

Laboratory

Excellent Services, B-34 & 35, Ganpati Paradise, Central Spine,


Vidhyadhar Nagar, Jaipur, Rajasthan

Accreditation Standard ISO/IEC 17025:2005


Discipline

Thermal Calibration

Issue Date

20.03.2014

Certificate Number

C-0703

Valid Until

19.03.2016

Last Amended on

Page

2 of 2

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

6.

Freezers, Cold Chamber,


Oven, Environmental
Chamber, Incubator,
BOD Incubator/ Water
Bath$

-30C to 50C
50C to 250C

0.5C
0.22C

Using Multi Point Data


Logger & 3 Wire RTD
(PT100) Sensor & PRT
4 Wire & Universal
Calibrator for Multi
position Calibration
By Direct Method

7.

Temperature Indicator of
Conditioning Chamber/
BOD Incubator / Liquid/
Water Bath / Dry Block
Furnace/ Oven/ Deep
Freezer

-80C to 250C

0.31C

Using RTD (Pt-100) 4


wire with indicator for
single position
Calibration
By Direct Method

II. HUMIDITY
1.

Humidity Chamber

20%RH to 95%RH @25C

1.2% RH

Using Standard RH
Sensor with Indicator
RH Calibrator
By Direct Method

2.

Digital & Analog


Hygrometer$

20%RH to 95%RH @25C

1.1%RH

Using Standard
Rotronics RH Sensor
with Indicator &
RH Calibrator
By Comparison Method

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$

Only in Permanent Laboratory


Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.