CLB 21103
PROCESS INSTRUMENTATION
BACHELOR OF CHEMICAL ENGINEERING
TECHNOLOGY
EXPERIMENT 2
Pressure Measurement &
Calibration
SEMESTER
January 2013
EXPERIMENT 2
PRESSURE MEASUREMENT & CALIBRATION SYSTEM
1.0
EXPERIMENT PROCEDURES
Check the readings at the Manometer and Pressure Gauge (PG). Once the
pressures readings in both the instruments are stabilized shut the vent (V)
valve and record the data in Table 1.
Open the Vent (V) valve to release the pressure to the next suggested
pressure.
Continue measure the pressure from 25 psig until 0 psig.
Plant Shut-down
1. Make sure that the remaining air in pressure tank, T1 is released with vent (V)
valve fully open.
2. Shut the manual valve MV6 next to AFR and all other manual valves fully.
3. Reconnect the original tubing of both PT1 and PT2.
4. Switch OFF the main power supply to the plant at the front of the panel/ cubical.
Fig. 1.1: Connection of instrument to Pressure Transmitter, Hand Held & Manometer
Fig. 1.2: Connection between Multimeter & the Pressure Transmitter terminal
2.0 RESULTS
TABLE 1: Pressure Measurement
RUN I
SET 1
SET 2
Average
30.00
30.00
30.00
29.47
29.46
29.74
30.00
30.00
30.00
28.50
29.50
29.00
44.20
44.30
44.25
F=B+ 14.70
44.17
44.16
44.44
0.53
0.54
0.63
0.97
0.04
0.38
0.03
0.14
0.19
SET 1
SET 2
Average
G1 = | B C |
G2 = | B D |
H=|FE|
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
RUN II
A
25.00
25.00
25.00
24.74
24.63
24.69
25.00
25.00
25.00
24.80
24.90
24.85
39.50
39.50
39.50
F=B+ 14.70
39.44
39.33
39.39
At Pressure Gauge
0.26
0.27
0.32
At Pressure Indicator,
P12
0.06
0.27
0.17
0.06
0.17
0.12
G1 = lB Cl
Gauge Pressure
Deviation, psig
G2 = lB Dl
H = IF El
RUN III
SET 1
SET 2
Average
20.00
20.00
20.00
19.53
19.45
19.49
20.00
20.00
20.00
20.00
20.00
20.00
34.50
34.50
34.50
F=B+ 14.70
34.23
34.15
34.19
0.47
0.55
0.51
0.47
0.55
0.51
0.27
0.35
0.31
SET 1
SET 2
Average
G1 = lB Cl
G2 = lB Dl
H = IF El
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
RUN IV
A
15.00
15.00
15.00
14.38
14.34
14.36
15.00
15.00
15.00
14.50
15.10
14.8
29.20
29.60
29.40
F=B+ 14.70
29.08
29.04
29.06
0.62
0.66
0.64
0.12
0.76
0.44
0.12
0.28
0.40
SET 1
SET 2
Average
G1 = lB Cl
G2 = lB Dl
H = IF El
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
RUN V
A
10.00
10.00
10.00
9.56
9.60
9.58
10.00
10.00
10.00
9.90
10.10
10.00
24.50
24.50
24.50
F=B+ 14.70
24.26
24.30
24.28
0.44
0.40
0.42
0.34
0.50
0.42
0.24
0.20
0.22
SET 1
SET 2
Average
G1 = lB Cl
G2 = lB Dl
H = IF El
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
RUN VII
A
0.00
0.00
0.00
0.03
0.01
0.02
0.00
0.00
0.00
0.50
0.10
0.30
15.10
14.80
14.95
F=B+ 14.70
14.73
14.71
14.72
0.03
0.01
0.02
0.47
0.09
0.19
0.37
0.09
0.23
SET 1
SET 2
Average
G1 = lB Cl
G2 = lB Dl
H = IF El
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
RUN VI
A
5.00
5.00
5.00
4.64
4.66
4.65
5.00
5.00
5.00
5.00
5.10
5.05
19.70
19.70
19.70
F=B+ 14.70
19.34
19.36
19.35
0.36
0.34
0.35
0.36
0.44
0.04
0.36
0.34
0.35
G1 = lB Cl
G2 = lB Dl
H = IF El
Gauge Pressure
Deviation, psig
At Pressure Gauge
At Pressure Indicator,
P12
Remarks:
B: Read from the Manometer (AMETEK JOFRA CPC-200 INDG).
C: Read from the Pressure Gauge (PG).
D: Read from the Pressure Indicator (PI2) on the front panel.
E: Read from the Pressure Indicator (PI1) on the front panel.
Input
Output
Measured
Theoretical
% Accuracy
% Deviation
Dev.
Remarks
Pressure Applied,
Pressure,
Output,
Output, mA
Limit
psig/psia
psia
mA
C D mA
D mA
x 100
C D mA x 100
20 - 4 mA
Fail : If Otherwise
at% of Calibration
Range
0.30/ 15.00 at 25.0%
15.10
7.98
8.00
0.25
0.13
0.075
Fail
15.10
7.97
12.00
33.58
25.19
0.075
Fail
15.10
7.98
16.00
50.13
50.13
0.075
Fail
15.10
7.98
20.00
60.10
75.13
0.075
Fail
Remarks:
B: Read from the Pressure Indicator (PI1) on the front panel.
C: Read from the Multimeter (GW MODEL GDM-391A).
Product Code:
Serial Number:
Maximum deviation : 0.075 (Zero Ini. Err = 0.03)
Tag Name:
Calibration Range:
A
Input
Output
Measured
Theoretical
% Accuracy
% Deviation
Dev.
Remarks
Pressure Applied,
Pressure,
Output,
Output, mA
Limit
psig
psig
mA
C D mA
D mA
x 100
C D mA x 100
20 - 4 mA
Fail : If Otherwise
at% of Calibration
Range
0.00 at 0.0%
0.00
3.98
4.00
0.50
0.13
0.03
Fail
7.50 at 25.0%
6.50
7.49
8.00
6.38
3.19
0.075
Fail
15.00 at 50.0%
14.40
11.68
12.00
2.67
2.00
0.075
Fail
22.50 at 22.5%
22.00
15.78
16.00
1.38
1.38
0.075
Fail
30.00 at 100.0%
28.20
19.96
20.00
0.20
0.25
0.075
Fail
Remarks:
B: Read from the Pressure Indicator (PI2) on the front panel.
C: Read from the Multimeter (GW MODEL GDM-391A).
3.0 QUESTIONS
1. State whether the pressure is higher/lower/equal at the bottom of a tank filled with
a gas than at the top.
2. Differentiate absolute pressure and gauge pressure.
3. If the contents of a tank holding 20m3 of gas are transferred to a tank with twice
as much volume and all other factors remain unchanged, what will happen to gas
pressure?
4. If the gas in a tank heated and all other factors remain unchanged, what will
happen to the gas pressure?
5. What is the pressure measuring instrument used in this plant?
6. Discuss briefly the possible errors affecting the accuracy in pressure measurement
as observed in the experiment.
4.0 REFERENCES
1. Experiment Manual PMC 200, QDR Marketing Sdn Bhd