Documente Academic
Documente Profesional
Documente Cultură
DOC0899034, Rev:1
DOC0899034
Revision
Type
Controlled Document
Title
Originator
212030967_william_r_phelan
Name
iVent201 MRI Image Quality Test doc with doc number and
page numbers added per Shlomi Deler
188928
Route
Signer
Status
R-4426975
100035647_shlomi__deler
Approve
R-4426975
212064784_rebecca_n_zimmer
Approve
Approved
Comments
Completion Date
DOC0899034, Rev:1
iVent201
Guidance on MRI Requirements and Image Quality Test
Table of Contents
1
Introduction ..............................................................................................................................................2
Purpose .....................................................................................................................................................2
Approved
3.1
3.2
3.3
MR Preparation Area.......................................................................................................................... 4
3.4
4.2
DOC0899034
Page 1
DOC0899034, Rev:1
1 Introduction
The special conditions of the Magnetic Resonance (MR) Environment (as defined below) increase the
requirements both on the equipment used and on the operators. The strong magnetic field requires that only
non-magnetic devices or devices with a low content of magnetic material are allowed in this environment. The
use of Radio Frequency (RF) fields puts extreme demands on the Electromagnetic compatibility (EMC) of
devices used during the imaging sequences of the MR scanner.
The IVent201 ventilator is classified as MR Conditional. MR Conditional means that the ventilator has been
demonstrated to pose no known hazards in a specified MR Environment with specified MR conditions for use.
The conditions required for use of the iVent201 in the MR Environment are described in this iVent201
Guidance on MRI Requirements and Image Quality Test (Guidance). If these conditions are met, a successful
Image Quality Test is performed, and the MR Image Quality Validation Record attached hereto at Section 5 is
signed, the use of the iVent201 is indicated for use in the specific MR Suite where the testing occurred.
2 Purpose
This Guidance outlines the requirements necessary to use the iVent201 ventilator in the MR Environment, and
the measures that must be undertaken for the use of the iVent201 with Magnetic Resonance Imaging (MRI).
Note: MR Environment is defined as the volume within the 0.50 millitesla (mT) (5 Gauss) line of an MR system,
which includes the entire three-dimensional space surrounding the MR scanner. For cases where the 0.50 mT
line is contained within the Faraday shielded volume, the entire room shall be considered the MR
environment1.
The iVent201 ventilator cannot be subjected to field strength greater than 100 Gauss and must
be kept outside of 100 Gauss perimeter, typically about 3.0 meters from the magnets
ASTM standard F2503-5, Marking Medical Devices and other items for safety in the magnetic resonance environment ,
ASTM International, West Conshohocken PA, 2006, www.astm.org.
Approved
A Responsible Organization is an entity that is accountable for the use and maintenance of the medical equipment or
system that includes the MR Environment.
DOC0899034
Page 2
DOC0899034, Rev:1
isocenter. For maximum safety, each facility must measure the Gauss field around the imaging
device and map the field force levels to ensure that the ventilator is not placed inside the
mapped Gauss lines.
NOTE:
All staff operating the ventilator in the MR suite must be advised that the static magnetic field
is always present whether or not a scan is occurring. Passing the ventilator along the side of
the scanner, even though a scan is not taking place, will damage the ventilator or may bring it
within the field that causes physical attraction between the ventilator and the MR device. Do
not bring the ventilator within the 100 Gauss boundary at any time.
Upon completion of mapping, the 100 Gauss strength lines should be permanently marked on the floor with a
durable tape, or with GE Healthcare Gauss line warning floor label (GE P/N 350C0323-01), shown below. It is
recommended that the clinical staff put tape marks on the floor indicating where specific equipment can be
safely placed within the MR suite.
To ensure that the ventilator remains outside the 100 Gauss boundary while the patient is within the range of
the imaging magnet, a patient circuit extension may be required. In such cases, GE Healthcare offers an MRI
Patient Circuit extension (GEHC P/N M1173617).
It is very likely that the magnetic shielding of some MRI units may not be 100% effective. Therefore, the Gauss
levels at a given distance may differ in different areas within the same radius from the center of the magnet
core. For this reason, it is important to map correctly the Gauss levels for each individual MRI suite. Further,
stationary objects that are ferrous, such as steel I-beams in the building wall or floor, affect the field lines of
the magnet, and therefore the 100 Gauss line may not be symmetrical, and should be mapped out on both
sides of the MR device.
Approved
Use of the ventilator other than as instructed may cause the ventilator to
malfunction, and may cause permanent damage to the ventilator.
CAUTION!
Connect the potential equalization pin on the back of the ventilator to the ground
during battery (DC power) operation.
Before using the ventilator in the MR suite, remove the green cover and the
DOC0899034
Page 3
DOC0899034, Rev:1
NOTE
Oxygen Cylinder Holder GE P/N M1206572 in the USA, and GE P/N M1162046
outside the USA) and Roll Stand (GE P/N M1206571 in the USA, and GE P/N
M1162044 outside the USA) are the only metal GEHC accessories that are
acceptable for use within the MRI Suite.
Remove gas cylinder (if used). Connect IVent201 to the gas supply of the MR preparation area.
Make sure the IVent201 is labeled MR Conditional and that a laminated MRI Card (GEHC P/N
M1161864) is attached to the units handle.
Transfer the IVent201 to an MR-approved mounting solution.
Make sure that the patient tubing is sufficiently long, attaching an extension if needed.
Approved
MRI Card
DOC0899034
Page 4
DOC0899034, Rev:1
The ventilator must be prevented from crossing the 100 Gauss line, whether the scanner is
active or not. There is a magnetic moment to the ventilator and if it is brought closer than the
100 Gauss line, it can be pulled toward the magnet. If the ventilator is running on AC and it is
brought closer than the 100 Gauss line, the power supply will be damaged and will require
repair. It is recommended that either a physical barrier be placed on the floor, or a tethering
system be deployed to prevent the ventilator from crossing the 100 Gauss line.
Approved
Execute a gradient echo scan with a small flip angle of 10 degrees with the ventilator turned off in the MRI
room using a standard phantom placed in the center of a head coil. This scan will serve as a reference
against which scans with the ventilator running shall be compared.
2.
Now using identical parameters and the identical setup, execute the same scan as in 1 above, but this
time, set the iVent201 ventilator to the 50Kg settings running with an Rp20 and attach a balloon test lung
to the patient circuit. Make sure the ventilator is oriented so that the screen is facing in the direction of
DOC0899034
Page 5
DOC0899034, Rev:1
the control room window and that the ventilator roll stand is on the far side of the 100 Gauss line from
the scanner magnet.
3.
Next, perform a scan with the ventilator running in the MR suite, but remove the head coil and set the
scan parameters for a body scan. Place a phantom target in the scanner.
4.
Lastly, the technician should perform a scan that is known to be sensitive to RF interference in order to
determine the sensitivity of the system to interference.
5.
If RF interference is noted, the hospital staff may elect to run the test again with the ventilator running
under battery power only to ascertain whether or not this reduces interference. The ventilator may be
run under battery power under the close supervision of respiratory therapy staff that are trained to
understand the battery charge status of the ventilator.
The Radiologist and MR technician must keep the scan files acquired in Scan Sequences 1-4 above and the site
Radiologist should qualify the scan results as acceptable or not. If the image quality is deemed not acceptable,
then the iVent201 is not to be used in this particular MR suite.
NOTE If the interference is deemed acceptable, the Radiologist and MR technician must be advised that any
addition of equipment, or change in MR scanner configuration (software or hardware) may result in an
increase in interference which is beyond the scope of the MR Environment Guidance.
2.
Approved
Yes
Yes
No
No
(tick as appropriate)
(tick as appropriate)
Page 6
DOC0899034, Rev:1
The purpose of this Validation Record and Acknowledgement is to verify that the requirements set forth in the
foregoing Guidance document and Image Quality Test, have been implemented at the time of installation. The
Guidance discloses the limitations of use of the iVent201 in the MR Environment due to possible image
artifacts. The customer must operate the ventilator in a manner consistent with this Guidance.
By signing below, the Responsible Organization hereby confirms and acknowledges the following:
1.
Implementation of all applicable Responsible organization actions according to this Guidance. This
applies to all iVent201 units used in specific MR environments.
2.
Successful completion of the MR Image Quality Test according to Section 4 of this Guidance.
3.
Authorized representatives of the Responsible Organization have read and understood the limitations of
use of the iVent201 in the MR Environment and the possibility that an increase in image artifact may
result if the MR Environment or MR scanner configuration are changed following the completion of the
MR Image Quality Test.
4.
The Responsible Organization shall retain full responsibility for MR image quality when using the iVent201
in the MR Environment, including but not limited to an increase in image artifact due to changes in the
MR hardware, software, or MR Environment.
RESPONSIBLE ORGANIZATION:
Place and Date
Signature
Print Name
Title or
Position
WITNESSED BY [GE Healthcare representative]:
Signature
Print Name
Title or Position
A copy of this Guidance, together with copies of the completed Tables 1 and 2 of section 4.2, should be
returned to _____________________________at GE Healthcare, [add relevant address].
Approved
DOC0899034
Page 7