Documente Academic
Documente Profesional
Documente Cultură
1.
2.
3.
4.
5.
6.
7.
8.
9.
Page 1 of 58
Paragon-X
o
Option 207 or 217 Jitter generation & measurement for 100M and 1GbE
Option 208 or 218 Jitter generation & measurement for 10GbE (optional)
Page 2 of 58
The front panel of the Paragon-X provides the following interfaces for testing:
1. 100MbE Electrical or Optical (SGMII SFP)
2. 1GbE Electrical or Optical (SFP) with option 110 fitted
3. 10GbE Optical (XFP or SFP+) with option 111 fitted
10GbE Port 1
10GbE Port 2
Status Display
100MbE/1GbE
Port 1
100MbE/1GbE
Port 2
Page 3 of 58
PC Controller Port
The Paragon-X accepts the following reference clocks which should be applied to one of the
Reference Inputs on the back panel:
1. 2.048MHz
2. 10MHz
3. E1 (2.048Mb/s)
4. DS1 (T1) (1.544Mb/s)
Page 4 of 58
button.
e) Click
f)
Select Tx + Rx mode
Page 5 of 58
j)
Select the Close button to close the Physical Settings window, and then select Close again to
close the setup interface window.
Note: once you have selected Sync-E mode, the Paragon-X GUI will display SYNCE LOCK as
grey (This verifies that the SyncE Rx > Tx box is unchecked, meaning that the Paragon-x is using
its internal or external clock. If the SyncE Rx>Tx box were checked in the required Tx+Rx mode,
there would be no clock).
Page 6 of 58
ESMC Generation
Ethernet Synchronisation Message Channel (ESMC) is a Point to Point Protocol in which each
EEC generates and terminates an ESMC PDU. This ESMC PDU is normally transmitted once per
second on the Slow Protocol Channel and contains the Quality Level (QL) of the associated clock
. The receiving EEC uses this QL to select the best quality clock. The EEC then regenerates the
ESMC PDU with the appropriate QL TLV value to indicate the quality of the clock being
transmitted to the next EEC.
If the EEC supports ESMC, it is possible for the Paragon-X to use this functionality to indicate if
an EEC switches clock reference or goes into holdover.
The Paragon-X can generate ESMC messages on both Ports 1 and 2. For G.8262 conformance
testing, ESMC messages are generated on Port 2 with a user- defined Quality Level (QL) such as
PRC.
open.
b) Check the box Enable TX+Rx Mode. This sets the instrument in terminate mode.
Page 7 of 58
c) Select Port 2 tab (the output from the Paragon and input to EEC)
d) Select SSM code QL-PRC, which tells the EEC that the clock from the Paragon-X is of PRC
quality.
e) Set IPG (inter-packet gap) rate to 1000mS. This makes the instrument send ESMC packets once
per second.
f)
Click
and then
g) The ESMC Generation button will go red to indicate the Paragon-X is generating ESMC packets.
After any setup changes to the Paragon-X, please ensure the EEC has had time to settle
before making any measurements
Page 8 of 58
Input Stimulus
Pass/Fail Criteria
EEC Option 1
Free run
EEC Option 2
Hold over
Notes
Measurement Setup
a) Connect the EEC to Paragon-X as shown above
b) Set up the Paragon-X GUI as per Section 3.
Measurement Process
a) Select the Start Capture
b) To stop the measurement after a pre-defined period for running the test, select Stop Capture
button.
Page 9 of 58
c) Paragon-X will provide the ppm frequency accuracy at the bottom of the graph in red text (as
shown in the screen shot at the top of the next page).
Page 10 of 58
Input Stimulus
EEC
Option 1
and
EEC
Option 2
Pass/Fail Criteria
Notes
Lock can also be
monitored by using
ESMC (if supported)
Input Stimulus
Pass/Fail Criteria
EEC
Option 1
EEC
Option 2
Notes
Not Applicable
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
to +/-4.6ppm
Page 11 of 58
Input Stimulus
EEC
Option 1
Pass/Fail Criteria
EEC
Option 2
Notes
G.8262 states this is for
further study
Not Applicable
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X GUI as per Section 3, including setting up ESMC with QL=PRC if using
ESMC to monitor which clock the EEC is locked to.
Measurement Process
a) select the Start Capture
Page 12 of 58
c) In the Frequency Offset window, add the Frequency Offset required and click the Apply Offset
button
d) To remove the Frequency Offset and return it back to 0ppm click the Remove Offset
button
e) To stop the measurement after a pre-defined period for running the test, select Stop Capture
button.
f) The Paragon-X TIE graph indicates if the EEC is in or out of lock as shown in the screenshot
below
Page 13 of 58
Page 14 of 58
Input Stimulus
Pass/Fail Criteria
EEC Option 1
- Locked Mode
(Constant Temp)
EEC Option 1
- Locked Mode
(Temp effects)
EEC Option 2
- Locked Mode
(Constant Temp)
Notes
(G.8262 masks)
MTIE Table 1/ Figure 1
TDEV Table 3/ Figure 2
MTIE Table 1/ Figure 1
TDEV G.8262 states
for further study
MTIE Table 4/ Figure 3
TDEV Table 5/ Figure 4
G.8262 Section 8.2 also mentions measurements in Non-locked mode and refers to section 11.2
Long-term phase transient response (Holdover) and will not be covered in this section.
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if
using ESMC to monitor which clock the EEC is locked to.
Measurement Process
a) Select the Start Capture
Page 15 of 58
e) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV
graph at the bottom.
Page 16 of 58
f)
The MTIE/ TDEV analysis can be carried out against the G.8262 masks for Wander
Generation which can be selected from the measurement mask pull down selection in the left
hand window.
g) Pass/Fail indication against the masks is shown also in this left hand window and will have a
green background for Mask Pass and a red background for Mask Fail
Page 17 of 58
Page 18 of 58
Input Stimulus
Pass/Fail Criteria
Notes
i. Measurement
filter bandwidths
specified in
G.8262.
All Sync-E
interface
rates
None, unless
device requires
packet stream to
function
a) Operating mode must be set to Sync-E, and Sync-E jitter selected as a measurement.
Page 19 of 58
Interface rate choice will then automatically be restricted to 1Gbit/s (Electrical or optical) or
10Gbit/s (100Mbit/s excluded).
For 10Gbit/s optical, device type is automatically restricted to SFP+ (XFP excluded).
Normally, the operating configuration is set to Tx+Rx mode.
b) Configure capture contains a jitter measurement threshold setting. If the checkbox is ticked, the
capture (result) screen will indicate pass/fail of the peak-peak jitter result currently selected,
relative to this limit. Note that the threshold result is compared to the instantaneous graph result
and so its behaviour is different between long-term and short-term peak-to-peak graph selections.
c) G.8262 specifies a pp jitter measurement time of 60s. The global default capture configuration in
the Paragon-X is manual (indefinite). Selecting Sync-E jitter measurement under operating
mode forces the configure capture timing control to timed mode, 60s. Select other periods or
manual if required, eg for experimentation and observing long-term device behaviour. However,
60s must be selected for making a G.8262 compliance measurement.
Page 20 of 58
d) Initiate jitter measurement using the start capture key. This automatically displays the statistics
window, jitter measurement tab. This tab shows long-term (ie over the total elapsed time) pp and
RMS jitter results. It also shows the measurement bandwidth, which is fixed depending on
interface type selected. Limit checking is not applied to the jitter measurement tab. The tab can
be closed, and re-opened any time via the Statistics and Results key.
e) The main capture screen shows table entries updated once per second for long-term pp, longterm RMS and short-term (one second) pp jitter results.
On the graph, use Graph Display Mode to select a result for graphing. The measurement
bandwidth is also displayed alongside the Y-axis legend on the graph.
Note that sample periods (1-second intervals) are numbered based on the start time of each onesecond period (ie, starting from 0), so for example a 60-second total measuring period will show
as containing sample periods 0-59.
Compare the 60s long-term peak-to-peak value to the G.8262 0.5UI limit.
Page 21 of 58
Page 22 of 58
Input Stimulus
EEC Option 1
Pass/Fail Criteria
The EEC is;
MTIE Wander
Table 7/Figure 5
TDEV Wander
Table 8/Figure 6
Sinusoidal Wander
Table 9/Figure 7
EEC Option 2
TDEV Wander
Table 10/Figure 8
Notes
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if using
ESMC to monitor if the EEC is switching clock references or going into holdover.
Page 23 of 58
Measurement Process
a) Select the Wander and Jitter Generation
c) Select the Wander Tolerance Tab and the following window will open
Page 24 of 58
e)
i.
ii.
Table Sinusoidal Wander Can be used for finding Maximum Tolerable Wander
iii.
i. Select the wander mask required from the drop down list. The mask and the maximum
running time will be shown on the right hand side of the window.
G.8262 Option 1 MTIE - Running Time is 1000s
G.8262 Option 1 TDEV - Running Time is 12000s
G.8262 Option 2 TDEV - Running Time is 12000s
ii. Click Generate Wander to start the test. The elapsed time will be displayed on the
bottom right hand side of the window.
iii. The test will stop after the max running time has elapsed, the test can also be stopped
manually by clicking Stop Wander.
f) Table (Sinusoidal)
Can be used to test Maximum Tolerable Wander
Page 25 of 58
The user can enter up to 10 different wander parameter sets in the table. The Paragon-X will
then automatically add each of the specified wander sets in turn giving full indication of progress
in the Status column. Switching between the different sets is always done at a zero crossing to
prevent phase steps.
i.
Enter the frequency, amplitude, and dwell time (number of cycles the
frequency/amplitude pair will be run) for each wander test point.
The same frequency with different amplitudes can be entered to find maximum
tolerable wander
The Restore Defaults button when checked will restore the values to that defined
in table 9 in G.8262.
Only rows that have the enable check box ticked will be executed in the test. To
skip over a selection, un-tick the enable box for that selection.
ii.
iii.
Click Stop Wander to stop the test. The test will terminate at the next zero crossing
iv.
A pop up box stating how long to the next zero crossing is displayed, click the Stop
Immediate button if it is desired to stop the test instantly.
Page 26 of 58
G.8262 states that with the wander applied that the EEC should
i. Maintain the clock within the prescribed performance limits (the exact performance limits
are for further study)
ii. Not cause any alarms
iii. Not cause the clock to switch references
iv. Not cause the clock to go into holdover
For further insight, use the Paragon-X ESMC generation/capture capability and TIE graph
to check whether the EEC is switching clock reference or going into holdover.
i)
Ensure the Paragon-X has been set up to Generate ESMC messages with QL=PRC as
described in section 3 of this document.
j)
k)
To show both the TIE and ESMC graph at the same time click
and then Graph Display Mode, finally select the graph to display
Page 27 of 58
nd
The screenshot above shows an EEC that has switched references or gone into holdover (top
graph is ESMC, bottom graph shows the difference in Frequency between the reference and the
recovered clock from the EEC)
m) To stop the measurement, select Stop Capture.
Page 28 of 58
Input Stimulus
EEC Option 1
and option 2
Pass/Fail Criteria
Notes
Page 29 of 58
Measurement Setup
a) Open Operating Mode dialog, change Operating Mode to Sync-E
Page 30 of 58
d)
Page 31 of 58
Page 32 of 58
f)
From the TestPackets tab, the user configures the test packet they wish to generate.
g) From the Jitter Tolerance Table tab, the user enters the Frequency, Amplitude, and
Dwell Time for each row.
h) From the Jitter Tolerance Table tab, the user clicks the Start Test button. Test Packet
generation will be automatically controlled as each row in the Jitter table is executed.
i)
The Jitter and Test Packet generation will automatically stop after all the enabled table
rows have been executed.
j)
A PASS or FAIL result for each executed row will appear on the Jitter Tolerance Table
tab. The result will indicate if any Dropped packets were detected during the test.
Page 33 of 58
k) In the Max tolerable jitter mode, operation is similar except a percentage above or below
mask can be set for the generated jitter amplitudes.
Page 34 of 58
Input Stimulus
EEC Option
1
EEC Option
2
Not defined
TDEV Wander
Table 10/Figure 8
Pass/Fail Criteria
Notes
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this
section
Page 35 of 58
b) Set up the Paragon-X as described in section 3, including setting up ESMC with QL=PRC if
using ESMC to monitor which clock the EEC is locked to.
Measurement Process
c) Select the Wander Transfer Tab and the following window will open
Page 36 of 58
Page 37 of 58
b) Enter the frequency, amplitude, and dwell time (number of cycles the frequency/amplitude pair
will be run) for each wander test point.
i. The Restore Defaults button when checked will restore the values to that defined in table 9 in
G.8262.
ii. Only rows that have the enable check box ticked will be executed in the test. To skip over a
selection, un-tick the enable box for that selection.
iii. Be careful to choose a dwell time of 300 cycles for any wander frequency of 1Hz or higher.
Additionally, choose the shortest sample period (configure capture button) available for
calibration and definitive measurement runs.
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click
the
button.
Page 38 of 58
e) Remove the short Ethernet cable between Port 1 and Port 2 and connect the EEC under test as
shown in the setup diagram at the start of this section.
f)
g) The Paragon-X will show the status of the test, the measured Gain (dB) value and also a
Pass/Fail indication on the right hand side of the window.
Page 39 of 58
c) To calibrate the Paragon-X, connect a short Ethernet cable between Port 1 and Port 2 and click
the
button.
Page 40 of 58
e) Remove the short Ethernet cable between Port 1 and Port 2 and replace with the EEC under test.
f)
g) The GUI will show the estimated completion time at the bottom of the screen. The measured Gain
(dB) value and also a Pass/Fail indication (Green/Red background) will be displayed.
Page 41 of 58
c) The amount of the time until completion of the test is shown at the bottom right hand side of the
screen.
d) At any time during the test, it is possible to view an updated output TDEV graph by clicking the
TDEV Results button. The TDEV graph will then be displayed.
Page 42 of 58
e) At the end of the test the Paragon-X will automatically show the final MTIE graph and show
Pass/Fail against the MTIE mask.
Page 43 of 58
Input Stimulus
EEC Option 1
EEC Option 2
Pass/Fail Criteria
Maximum phase
transient at the output
due to reference
switching to meet mask
in G.8262 Figure 12
Notes
To emulate the loss of the
link either
Change ESMC
QL=DNU
Change ESMC
QL=DNU
Measurement Setup
a) Connect the EEC to Paragon-X to EEC as shown in the diagram at the beginning of this section
Page 44 of 58
b) Set up the Paragon-X as described in section 3 (including ESMC generation if using the ESMC
method of switching clock references, ensuring SSM code is set to PRC)
c) Select Configure Capture
Measurement Process
a) There are two methods for determining Phase Transient Response with the Paragon-X;
i. Use ESMC Generation (if supported by the DUT).
ii. Remove the link between Port 2 on the paragon-X and the EEC input port.
c) If using the Ethernet cable removal method please disconnect cable between port 2 and the EEC.
d) If using ESMC method
i. Select
Page 45 of 58
Measuring Results
The switching transient will be seen on the Paragon-X TIE graph.
The Paragon-X graphs can show the TIE and also the ESMC QL for each port
In the screenshot bellow the ESMC graph shows the time and QL change of the ESMC transition
on Port 2 Tx of the Paragon-X. This can then be compared in time to the TIE graph showing the
Transient.
Method of evaluating the capture for phase transient response is dependent upon whether EEC
Option 1 or Option 2 is being evaluated.
EEC Option 1
a) To view the output TIE for EEC Option 1, closely monitor the output TIE graph on Paragon-X for
the duration (15s) of the test. While the slave clock is acquiring the new reference, the output
phase transient should be within the limits of figure 12 of G.8262 (provided below)
Page 46 of 58
EEC Option 2
a) To view the output MTIE mask for EEC Option 2 ONLY, perform the following:
b) To display the MTIE and TDEV graphs click
c) A separate window will open showing the TIE in graphical form at the top, with an MTIE/TDEV
graph at the bottom.
Page 47 of 58
d) The MTIE analysis can be carried out against the G.8262 masks for Wander Transient which
can be selected from the measurement mask pull down selection in the left hand window.
e) Pass/Fail indication against the masks is shown also in this left hand window and will have a
green background for Mask Pass and a red background for Mask Fail
Page 48 of 58
Page 49 of 58
10.
Input Stimulus
Pass/Fail Criteria
Notes
To emulate the loss of the
link either
EEC Option 1
Maximum phase
transient and excursion
at the output due to
reference switching to
meet mask in G.8262
Figure 13
Change ESMC
QL=DNU
EEC Option 2
Maximum phase
transient and excursion
at the output due to
reference switching to
meet mask in G.8262
11.2.2 and table 14,
st
nd
including 1 and 2
derivatives of phase vs
time
Change ESMC
QL=DNU
Page 50 of 58
After allowing the DUT to settle, leave for at least 900s (15 minutes), measure the wander on Port 1.
The resulting graph should be flat and the Offset measurement at the foot of the Wander TIE graph
should indicate 0.000 ppm. The ESMC received from the DUT on Port 1 should also indicate PRC.
Restart the wander measurement and then set the ESMC on Port 2 to DNU (Do Not Use). This
should cause the the DUT to search for an alternative master clock signal to lock to. If there are no
other external signals then the DUT will use its internal clock. The ESMC messages on port 1 should
reflect this change with the appropriate Quality Level e.g. EEC1.
The TIE graph should show the SyncE signal from the DUT start to drift off. Leave the capture running
for a further 10,000 seconds (approx 3 hrs) minimum. Stop the measurement.
The limits of operation are given by this mask reproduced from G.8262 fig 13:
Check also that the TIE measurement shows a calculated frequency offset (shown at bottom of TIE
graph) within +/-4.6ppm.
For EEC-2, proceed as above, except the limit for TIE is as described in G.8262 11.2.2 and table 14,
illustrated here as a mask:
Page 51 of 58
1.00E+06
1.00E+05
1.00E+04
TIE (nS)
1.00E+03
1.00E+02
1.00E+01
10
100
1000
Time S (seconds)
st
The limits for EEC-2 1 derivative TIE (fractional frequency offset) and 2
nd
can be used for all cases, and is the only practical way to compute the 1 and 2
nd
differentials of TIE in
Page 52 of 58
used in the calculation need to be changed to match the original capture sample rate, if it is varied
from its default 30/s.
See Calnex document CX5001e (on our website) as a practical example in Excel of the use of a
Paragon-x exported TIE file, with central difference differentiation examples.
Page 53 of 58
Pass/Fail Criteria
EEC Option 1
Free run
Recommend to test
for an hour, longer if
close to limits
Recommend to test
for an hour, longer if
close to limits
EEC Option 2
Hold over
Notes
Pass/Fail Criteria
Notes
Lock can also be
monitored by using
ESMC (if supported)
Pass/Fail Criteria
EEC
Option 1
EEC
Option 2
Notes
Not Applicable
EEC is locked to the clock
from the Paragon-X. The
Frequency is then offset
to +/-4.6ppm
Page 54 of 58
Input Stimulus
EEC
Option 1
Pass/Fail Criteria
EEC
Option 2
Notes
G.8262 states this is for
further study
Not Applicable
Input Stimulus
EEC Option 1
- Locked Mode
(Constant Temp)
EEC Option 1
- Locked Mode
(Temp effects)
EEC Option 2
- Locked Mode
(Constant Temp)
Pass/Fail Criteria
MTIE & TDEV Pass/Fail
masks shown in G.8262
Section 8.1.1
MTIE Pass/Fail masks
shown in G.8262
Section 8.1.2.
MTIE & TDEV Pass/Fail
masks shown in G.8262
Section 8.1.2
Page 55 of 58
Notes
MTIE Table 1/ Figure 1
TDEV Table 3/ Figure 2
MTIE Table 1/ Figure 1
TDEV G.8262 states
for further study
MTIE Table 4/ Figure 3
TDEV Table 5/ Figure 4
Input Stimulus
EEC Option 1
Pass/Fail Criteria
The EEC is;
MTIE Wander
Table 7/Figure 5
TDEV Wander
Table 8/Figure 6
Sinusoidal Wander
Table 9/Figure 7
EEC Option 2
TDEV Wander
Table 10/Figure 8
Notes
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
To Check whether the
EEC is switching
references or going
into holdover, the
Paragon can measure
the wander and/or
ESMC QL of the EEC
output.
EEC Option 1
EEC Option 2
Not defined
TDEV Wander
Table 10/Figure 8
Pass/Fail Criteria
Page 56 of 58
Notes
There is no definition of the
input stimulus to be used in
G.8262. Without further
guidance from the Standards,
it is suggested that the
amplitude and frequency
values associated with mask
points labelled f1, f2 & f3 on
G.8262 Section 9.1.1, Table 9
& Figure 7 are used.
EEC Option 1
EEC Option 2
Pass/Fail Criteria
Maximum phase
transient at the output
due to reference
switching to meet mask
in G.8262 Figure 12
Notes
To emulate the loss of the
link either
Change ESMC
QL=DNU
Page 57 of 58
Change ESMC
QL=DNU
For more information on the Calnex Paragon-X and to take advantage of Calnexs extensive experience in
sync and packet testing technologies, please contact Calnex Solutions on +44 (0) 1506 671 416 or email:
info@calnexsol.com
www.calnexsol.com
CX5001 v10 Oct 2012
This information is subject to change without notice