Sunteți pe pagina 1din 8

Field Service Report

CHANDLER
AUSTRALIA
Customer
Shipping
Address

Schlumberger
Aksai, Kazakhstan

Date

19th Sep 2015

S.C.#

SC5848

Service Requested By:

Tursynay Urynova

Phone number

+7 777 500 32 65

Mobile Number
E-mail Address
Service Technician

Pavel Lebedev

Serial # 391

Customer ID N/A

Model #

4262 Twin UCA

Service Type

Calibration

Service
Performed

1. Evaluated instrument against the requirements of API 10B


2. Reviewed operating procedures with laboratory staff

Description

Laboratory audit for API testing

Summary
Approved By

Tursynay Urynova
+7 777 500 32 65

Service Technician

Customers
Comments

Replacement Parts Installed


Item

1
2
3
4
5
6
7
8
9
10

Qty

Part #

Description

Part Charged Out of


Inventory Via

Part Billed
To/Charged Against

Melbourne, Australia
+61 3 9761 7194

Calibration Certificate

Chandler
AUSTRALIA

Schlumberger Aksai, Kazakhstan


CUSTOMER:
MANUFACTURER: AMETEK Chandler Engineering
Twin Cell UCA
DESCRIPTION:
CUSTOMER INSTRUMENT NO: N/A

MODEL: 4262 Left Cell


SERIAL NO: 391

CALIBRATION STANDARDS
CALIBRATOR

I.D. Number

Cal Cert No

Serial No

Calibration Date

Calibration Due

Fluke

714

CK0071595

2938088

19-08-15

18-08-16

JHW40KGZ

CK0066425

1449739

03-09-15

02-09-16

9100S

CK0071594

B51460

19-08-15

18-08-16

TRANSDUCER

Honeywell
CALIBRATOR

Hart Scientific

Controller
Ref. Pressure
(psi)
0
500
1000
1500
2000
3000
4000
Controller Type

Initial
(kpsi)

PRESSURE CALIBRATION
Instrument Gauge

Calibrated
(kpsi)
N/A

Initial
(psi)
0
400
900
1400
1900
3000
4000

Data Acquisition

Calibrated
(psi)

Initial
(psi)

Calibrated
(psi)
N/A

Tolerance as per API 10B-2 ( +/- 1% at minimum of 25%, 50% and 75% of user-defined range )

Ref. Temperature
(C)
0
50
100
150
200
Controller Type

TEMPERATURE
Data Acquisition
Controller
Initial
Calibrated Initial
Calibrated
4
53
102
150
199

0
50
100
150
200

0
50
100
150
200

Thermocouple
L
50
100
150
200

7051

Tolerance as per API 10B-2 ( +/- 3F / 2C ); Tolerance for thermocouples ( +/- 2F / 1C )

UCA TRANSIT TIME MEASUREMENT


Measured
Transit Time (micro-sec/in) :
7.925
Tolerance as per manufacturers specification ( 7.925 0.025) with Steel Calibration Bar
The accuracy of the Calibrating Instruments is traceable to National Standards. Evidence of traceability is maintained on file.

CALIBRATED BY:

Pavel Lebedev

DATE:

19-Sep-2015

SIGNATURE:

SV-4001
Revision 7.0

Page 1 of 1

Issue Date:05/14/03
Revision Date:10/06/14

Melbourne, Australia
+61 3 9761 7194

Calibration Certificate

Chandler
AUSTRALIA

CUSTOMER:
Schlumberger Aksai, Kazakhstan
MANUFACTURER: AMETEK Chandler Engineering
DESCRIPTION:
Twin Cell UCA
CUSTOMER INSTRUMENT NO: N/A

MODEL: 4262 Right Cell


SERIAL NO: 391

CALIBRATION STANDARDS
CALIBRATOR

I.D. Number

Cal Cert No

Serial No

Calibration Date

Calibration Due

Fluke

714

CK0071595

2938088

19-08-15

18-08-16

JHW40KGZ

CK0066425

1449739

03-09-15

02-09-16

9100S

CK0071594

B51460

19-08-15

18-08-16

TRANSDUCER

Honeywell
CALIBRATOR

Hart Scientific

Controller
Ref. Pressure
(psi)
0
500
1000
1500
2000
3000
4000
Controller Type

Initial
(kpsi)

PRESSURE CALIBRATION
Instrument Gauge

Calibrated
(kpsi)
N/A

Initial
(psi)
0
350
900
1400
1900
3000
4000

Data Acquisition

Calibrated
(psi)

Initial
(psi)

Calibrated
(psi)
N/A

Tolerance as per API 10B-2 ( +/- 1% at minimum of 25%, 50% and 75% of user-defined range )

Ref. Temperature
(C)
0
50
100
150
200
Controller Type

TEMPERATURE
Controller
Data Acquisition
Initial
Calibrated Initial
Calibrated
0
50
100
150
200

0
50
100
150
200

Thermocouple
R
50
100
148
198

7051

Tolerance as per API 10B-2 ( +/- 3F / 2C ); Tolerance for thermocouples ( +/- 2F / 1C )

UCA TRANSIT TIME MEASUREMENT


Measured
7.925
Transit Time (micro-sec/in) :
Tolerance as per manufacturers specification ( 7.925 0.025) with Steel Calibration Bar
The accuracy of the Calibrating Instruments is traceable to National Standards. Evidence of traceability is maintained on file.

CALIBRATED BY:

Pavel Lebedev

DATE:

19-Sep-2015

SIGNATURE:

SV-4001
Revision 7.0

Page 1 of 1

Issue Date:05/14/03
Revision Date:10/06/14

CS-1102
Page 1 of 2

CHANDLER ENGINEERING
TITLE: PROC,IN-SERVICE PRESSURE TEST AND INSPECTION

Revision

Date

06/03/2015

Description

Revised By:

ECN xxxx; updated content

Checked By:

AKD

Instrument and Pressure Vessel Check List


Instrument Inspection
1. Date:

19th September 2015

2. Instrument Model Number: 4262 Left Cell

Tested by:

Pavel Lebedev

Serial Number:

391

3. Date of Manufacture: April 2004


4. Original Maximum Pressure of Instrument:

5,000psi

5. Original Maximum Temperature of Instrument: 400F


6. Is the instrument equipped with a rupture disk or relief valve?
If yes, what is the pressure relief value?

5,000 6,000psi

Date of last rupture disc replacement:

N/A

Overall Condition of the Instrument:

Relief Valve (84-0108)

(Must be less than 2 yrs. Old)


(1 = Poor, 5 = Excellent)

Details:

Pressure Vessel(s) Inspection


1. Record the heat numbers or other identifying marks from the vessel and plugs, if accessible.
Vessel: Heat number: C14747-6

Serial Number: 391L

Plug: 71594, 391L

Date: 04-04

Base: 71594, 391L

Other:
5

Condition of the high pressure sealing surfaces:

(1 = Poor, 5 = Excellent)

Details:
2. Condition of the API tool joint threads:

(1 = Poor, 5 = Excellent)

Details: Was the heater element removed to inspect the vessel ID:
3. Corrosion on the inside surfaces of the vessel:
4. Estimated Depth of corrosion pits:
5. Vessel Assembly Age:

Not applicable

(1 = Poor, 5 = no significant corrosion)

None present

15 years

6. Vessel de-rating factor: 1.00


7. Recommended maximum testing parameters:

5,000psi,

400F

CS-1102
Page 2 of 2

CHANDLER ENGINEERING
TITLE: PROC,IN-SERVICE PRESSURE TEST AND INSPECTION

Revision

Date

06/03/2015

Description
ECN xxxx; updated content

Revised By:

Checked By:

AKD

Pressure Vessel Inspection Notes

Ensure inside corners have a


radius and not a sharp corner.
Check for cracking.

1. No visible cracking inside cell

CS-1102
Page 1 of 2

CHANDLER ENGINEERING
TITLE: PROC,IN-SERVICE PRESSURE TEST AND INSPECTION

Revision

Date

06/03/2015

Description

Revised By:

ECN xxxx; updated content

Checked By:

AKD

Instrument and Pressure Vessel Check List


Instrument Inspection
1. Date:

19th September 2015

2. Instrument Model Number: 4262 Left Cell

Tested by:

Pavel Lebedev

Serial Number:

391

3. Date of Manufacture: April 2004


4. Original Maximum Pressure of Instrument:

5,000psi

5. Original Maximum Temperature of Instrument: 400F


6. Is the instrument equipped with a rupture disk or relief valve?
If yes, what is the pressure relief value?

5,000 6,000psi

Date of last rupture disc replacement:

N/A

Overall Condition of the Instrument:

Relief Valve (84-0108)

(Must be less than 2 yrs. Old)


(1 = Poor, 5 = Excellent)

Details:

Pressure Vessel(s) Inspection


1. Record the heat numbers or other identifying marks from the vessel and plugs, if accessible.
Vessel: Heat number: C133764

Serial Number: 391R

Plug: 71594, 391R

Date: 04-04

Base: 71594, 391R

Other:
5

Condition of the high pressure sealing surfaces:

(1 = Poor, 5 = Excellent)

Details:
2. Condition of the API tool joint threads:

(1 = Poor, 5 = Excellent)

Details: Was the heater element removed to inspect the vessel ID:
3. Corrosion on the inside surfaces of the vessel:
4. Estimated Depth of corrosion pits:
5. Vessel Assembly Age:

Not applicable

(1 = Poor, 5 = no significant corrosion)

None present

15 years

6. Vessel de-rating factor: 1.00


7. Recommended maximum testing parameters:

5,000psi,

400F

CS-1102
Page 2 of 2

CHANDLER ENGINEERING
TITLE: PROC,IN-SERVICE PRESSURE TEST AND INSPECTION

Revision

Date

06/03/2015

Description
ECN xxxx; updated content

Revised By:

Checked By:

AKD

Pressure Vessel Inspection Notes

Ensure inside corners have a


radius and not a sharp corner.
Check for cracking.

1. No visible cracking inside cell