Sunteți pe pagina 1din 2

Keyboard

Type

Membrane keyboard with


tactile feedback

Software Standard
Resting ECG mode Records and prints 12-lead
resting ECGs with 10 seconds
duration as a standard feature.
Language support English
Electrical
Power Supply

External ACDC adaptor or


battery operation (system
acquires and prints only on
battery)

External ACDC adaptor specifications


Input voltage
100 to 240 VAC 10%
Input current
Maximum 0.5A @ 90 VAC,
0.3A @ 240 VAC
Operating frequency 50 to 60 Hz 3Hz
Output voltage
12V 5%
Battery Specification
Battery type
Replaceable and rechargeable,
Lithium Ion
Battery capacity

7.2V typical, 2.25 AH 10%


250 continuous ECGs in
Simultaneous 1 lead report
format at 25 mm/s and
10mm/mV
500 continuous ECGs in
Simultaneous 3 lead report
format at 25 mm/s and
10mm/mV
240 minutes of continuous
rhythm print at 25 mm/s and
10 mm/mV

Physical Specification
Height
Approximately 85 mm
Width
Approximately 275 mm
Depth
Approximately 215 mm
Weight
Approximately 1.2 Kg including
battery and paper roll
Environmental Specification
Temperature
Operating
Transport/storage

5 to 40C
-15C to 50 C

Humidity
Operating
transport/storage

25% to 95% RH non-condensing


25% to 95% RH non-condensing

Pressure
Operating
Transport/storage

700 to 1060 hPA


500 to 1060 hPA

MAC i
More ECG Per ECG

Key International Safety & Performance Standards


IEC 60601-1: 1988 +Amd-1: 1991, +Amd-2: 1995
General Requirements far Safety
IEC 60601-2-25: 1993 +Amd-1: 1999
Particular Requirements for the Safety of
Electrocardiographs
IEC 60601-2-51: 2003
Particular requirements for safety, including essential
performance, of recording and analyzing single
channel and multi channel electrocardiographs.
ANSI/AAMI EC-11: 1991 / 2001 / 2007
American National Standard Diagnostic
Electrocardiography Devices

Since 1965, GE has introduced technology innovation, which has profoundly impacted clinical guidelines and ECG
standards. Through its collaboration with world-leading cardiologists, GE continues to develop ECG products to

Battery charge time Approximately 3 hours from


total discharge

higher standards of quality. With over 40 years' heritage in highly reliable ECG analysis and interpretation, we offer
you the Marquette 12SL program to support your clinical decisions.
MAC i is GE Healthcare's new compact electrocardiograph, specially designed to meet the needs unique to

GE Healthcare
Plot No. 4, Kadugodi Industrial Area
Sadaramangala
Whitefield, Bangalore 560 067.
www.gehealthcare.com

2011 General Electric Company - All rights


reserved.
GE Medical Systems Information Technologies, a
General Electric company, doing business as GE
Healthcare.
General Electric Company reserves the right to make
changes in specifications and features shown herein,
or discontinue the product described at any time
without notice or obligation.
Contact your GE representative for the most current
information.
GE, GE Monogram, MAC, Marquette and 12SL are
trademarks of General Electric Company.

2036097-002-2007.10-V2.0

Indian environment. Being a 'GE healthymagination' product, MAC i offers Accuracy, Convenience and Value for
Money.
Light weight & portable MAC i is easy to operate, aids in faster, more confident cardiac assessment ensuring
better patient care.

Better...
Accuracy!

Instrument Type

Convenience!

Value for Money!

Dual ECG Modes


Auto and Manual modes allow you to adapt to
your specific patient needs.

Ultra Portable
ECG device (with battery) weighs less than 1200
Gms. for easy transport.

Approved as per International Standards


MACi complies to all the applicable requirements
of international safety and performance standards
namely IEC standards and AAMI EC 11.

Long-lasting Battery
Uses Li-Ion platform, lasts for 500 ECGs in three
channel mode & 250 ECGs in single channel mode.

Reliable Decision Support


Marquette 12SL Measurement and Interpretation
aids in faster decision making.
Wide Paper Size
58 mm paper is used for clear visualization of ECG
waveforms and reports.
One Touch Operation
For acquisition, analysis and printing, minimizes
training needs.
Alphanumeric Display
For easy visualization of ECG setup information.

Microprocessor augmented automatic electrocardiograph; 10-Ieadwire, 12 lead


simultaneous acquisition with programmable lead configuration.

Processing

Lower Recharge Time


Battery recharges in less than 3 hours.
Smart Options
Three channel print supports paper savings.
Decision support tool for auto measurement and
interpretation using GE Marquette 12SL are
available to meet specific customer needs.
MAC i Supplies & Accessories
GE Healthcare offers a wide range of tested and
approved quality accessories and consumables
that are specifically designed to maintain the
performance of the MAC i.

MAC i available as

Single Channel with Measurement


Single Channel without Measurement
Three Channel with Measurement
Single Channel with Measurement and Interpretation
Three Channel with Measurement and Interpretation

Acquisition

12 leads simultaneous ECG acquisition

Digital sampling rate:

2,000 samples/second/channel

Acquisition mode

10 seconds of simultaneous and Sequential


ECG recording

Computerized Measurements

Supports Measurement with Marquette


12SL ECG Analysis Program (Optional feature)

Interpretation

Supports Adult Interpretation with Marquette


12SL ECG Analysis Program (Optional Feature)

ECG analysis frequency

500 samples/second/channel

Dynamic range

AC Differential 5 mV, DC offset 300 mV

Resolution

4.88 V/LSB @ 500 sps

Frequency response

-3 dB @ 0.01 and 150 Hz

Low cut-off frequency

0.01 Hz, 0.02 Hz, 0.16 Hz or 0.32 Hz

High cut-off frequency

150 Hz, 40 Hz or 20 Hz

Operating modes

Automatic ECG (12 lead data), Rhythm/Manual


(real-time data)

Common mode rejection

>100 dB (with AC filter enabled)

Input impedance

>10 M @ 10 Hz, defibrillator protected,

Patient leakage

<10 A in Normal Condition and <50 A in


Single Fault Condition

Lead-fail indication

Disconnected lead detection.

Heart rate meter

30 to 300 BPM 10% or 5 BPM, whichever is


greater.

Start up time

7 seconds

Display
Display type

2x16 alphanumeric display with adjustable


contrast

Display data

Operating Mode, Heart rate, print formats/Lead


display, speed, gain, muscle filter, AC filter and
Configuration menu

Writer

Example of single channel ECG report with Marquette 12SL Measurement and Interpretation.

Writer technology

Thermal dot array print head

Writer speed

5, 25, & 50 mm/s

Number of traces

1 channel rhythm, 1 or 3 channel (option) Auto


ECG user selectable

Writer sensitivity/gain

2.5, 5, 10, 20mm/mV

Writer speed accuracy

5%

Writer amplitude accuracy

5%

Writer resolution

Horizontal 40 dots/mm @ 25 mm/s, 8 dots/mm


vertical

Paper type

Thermal Roll paper 20 m.

S-ar putea să vă placă și