Documente Academic
Documente Profesional
Documente Cultură
A rotating stage: This allows the position of the specimen to be set. Rotate ability of the
polarizing stage and center ability are fundamental 360 degree angle. A universal stage
with multiple rotating axes may also be used to enable the observation of specimen from
many directions.
Polarizing objective (strain-free objective) for polarized light: polarizing objective
differs from ordinary objectives in a respect that it possesses a high light polarizing
capability.
A center able revolving nosepiece: This enables optical axis adjustment for the
objective.
Analyzer: The analyzer can rotate 90 degree or 360 degree.
Bertrand lens: It is used for observing the objective. It is located between the analyzer
and eyepiece for easy in and out of the light path.
A test plate: The test plate is a phase plate used for verifying the double refractivity of
specimens, determining the vibration direction of pieces, and for retardation
measurement.
the analyzer. The most common compensators are the quarter wave, full wave, and
quartz wedge plates.
the stage. The stage can be centered with minute adjustments to the rear pivot pins by
turning the centering screws. Start by locating a small circular feature on a relatively simple
bright field specimen in an area devoid of other distracting structures. Next, relocate the
chosen feature to the center of the view field in the eyepiece crosshairs (see Figure 2(a)).
Rotate the stage through 180 degrees until the centered specimen feature is overlapped by
one of the vertical eyepiece crosshairs (Figure 2(b)). Using the pivot screws, relocate the
specimen feature to a distance approximately half way back to the crosshair center (Figure
2(c)). Finally, move the specimen feature back to the center of the crosshairs (Figure 2(a))
and repeat the alignment sequence. After a few cycles, the specimen feature should be
centered in the view field as the stage is rotated.
After the stage has been centered with respect to the microscope optical axis with the 10x
objective installed in the optical train, the other objectives (4x, 20x, 40x, and 100x) should
be sequentially centered to the stage. Microscopes having center able objectives contain a
pair of Allen (or a similar drive) setscrews in the nosepiece that translate the objective
laterally within its seat. Each objective can be centered using the procedure described above
for the circular stage by using the objective centering tools provided by the manufacturer.
Accurate centering of the stage and all objectives is paramount to quantitative analysis in
polarized light, and avoids the operator losing sight of specific features in the view field
during stage rotation. After completing the centering procedure, review the steps to ensure
all of the objectives on the microscope are accurately centered before continuing.
Establishing Khler illumination and centering the microscope stage and objectives is easily
accomplished in bright field mode (but can also be undertaken with a birefringent specimen
in polarized light mode, if necessary). To continue with alignment of the polarizing
components, install the fixed or rotating polarizing element in the condenser or a housing on
the light source exit port in the microscope base. Many polarized light microscopes are
equipped with a fixed polarizer that has a transmission vibration axis oriented East-West, or
left-to-right when standing in front of the microscope. Other microscopes have polarizers
that can be rotated through 180 or 360 degrees, and are often mounted into frames with
graduated increments ranging from 1 to 90 degrees. Ensure that the polarizer is correctly
secured into position and either permanently fixed into its holder or placed in the zero
position. The polarizer transmission vibration axis must be set to the East-West orientation
in this step.
Insert the analyzer into the microscope nosepiece or intermediate tube. Virtually all
research-level polarized light microscopes have a rotating analyzer mounted in a rectangular
frame that enables rotation of the element through 180 or 360 degrees. Student microscopes
often have a fixed analyzer (as well as a fixed polarizer) to ease the burden of crossed
polarizer alignment. Rotate the analyzer while examining a birefringent specimen on the
stage with the 10x objective. When the transmission axes of the analyzer and polarizer are
crossed at a 90-degree angle, maximum specimen birefringence (equivalent to maximum
specimen brightness) should be observed on a very dark background (a condition that is
termed maximum extinction). Under crossed polarized illumination, the vibration axis of
the analyzer should be oriented North-South or front-to-back while standing in front of the
microscope. A critical assessment for maximum extinction can only be accomplished while
observing the background intensity through the microscope eyepieces. Keep in mind that at
this point, maximum extinction only indicates that the polarizers are crossed, but does not
guarantee that their transmission axes are perfectly oriented East-West and North-South. The
latter aspect is addressed in the next step.
Calibration of the polarizer and analyzer axis alignment can be accomplished by one of two
methods. Many manufacturers provide a polarized light microscope orientation plate with
their research-level microscopes to assist in alignment of the crossed polarizers. These plates
contain a perfectly cut rectangular thin birefringent crystal having an optical axis that is
parallel to the long edge of the crystal (see Figure 3). The crystal is mounted flat on a
microscope slide and designed specifically for use on circular polarized light microscope
stages. When the long axis crystal is oriented at a 45-degree angle to the polarizer and
analyzer transmission azimuths (Figure 3(a)), maximum birefringence is observed in
polarized light. Rotating the crystal towards the polarizer axis progressively reduces the
level of birefringence (Figure 3(b)) until it is eliminated when the long axis is exactly
parallel to the polarizer transmission azimuth (and the horizontal eyepiece crosshair; see
Figure 3(c)).
At this point, the crystal birefringence is completely extinguished and should be
indistinguishable from the background. Note that the outline of the crystal is artificial in
Figure 3(c), and is included only to demonstrate that the long axis is parallel to the polarizer
azimuth. If the polarizer and analyzer are not positioned exactly in a East-West and NorthSouth orientation (respectively), then the long axis of the crystal will not be parallel to the
eyepiece crosshairs when the birefringence is completely extinguished. In this case, rotate
the long axis of the crystal until it is parallel to the horizontal microscope crosshair, and then
rotate the polarizer until minimum birefringence is observed in the eyepieces. Next, rotate
the analyzer until the crystal birefringence is completely extinguished.
If a Bertrand lens or focusing telescope is available with the polarized light microscope,
alignment of the polarizer and analyzer transmission azimuths can be performed using one
of these optical components. Without a specimen on the stage, cross the polarizers and insert
the eyepiece telescope into the observation tubes or swing the Bertrand lens into the optical
pathway. Focus the telescope or Bertrand lens on the objective rear aperture. A dark
polarization cross can be observed at maximum extinction (Figure 4(a)) with brighter
regions flanking the arms of the cross, which should be oriented vertically and horizontally.
In most cases, the horizontal and vertical components of the polarization cross are wider in
the center of the field and narrower at the periphery.
The polarization cross geometry originates from the depolarizing effects of spherical lens
surfaces in the condenser and objective, and may vary in size and shape depending upon the
quality (strain reduction or elimination) of the lens systems. However, the presence of a
polarization cross is normal in most microscopes and can not generally be considered an
artifact. The cross should be centered in the view field (when superimposed on the eyepiece
crosshairs) and appear perfectly aligned in a North-South and East-West direction. If this is
not the case (as illustrated in Figure 4(c)), the polarizer and analyzer should be rotated in
combination until the proper orientation is achieved. As discussed above, perfect alignment
of the polarization cross (thus, the polarizers) with a transmission axis at zero degrees is
essential for performing quantitative measurements of birefringence azimuthal angles with
or without a retardation plate or compensator in the optical pathway. The polarization cross
reference calibration is also useful if a fixed polarizer or analyzer becomes misaligned in the
mount.
The final step in polarized light microscope alignment is to adjust the condenser aperture
diaphragm so the bright outer regions of the polarization cross that are visible at the edge of
the objective aperture are blocked. Figure 4(b) illustrates a properly adjusted condenser
diaphragm. This step dramatically improves the extinction factor of the microscope optical
system, and the field will appear maximally dark when viewed in orthoscopic mode
(without a Bertrand lens or focusing telescope). After the microscope has been aligned and
the polarization axes calibrated, it is ready for quantitative analysis of birefringent
specimens.