Documente Academic
Documente Profesional
Documente Cultură
HELP
Labview library
Version Content
Pressure&force.llb
Transducer_calibration.llb
3.37
4.32
isel_ui4.x.llb
1.21
PointEditor.llb
ARA_Joystick.llb
ARA_Utils.llb,
0.19
0.1
1.06
2/37
Introduction .................................................................................................................. 23
Use of the program ....................................................................................................... 24
The panel on the left ..................................................................................................... 24
Lower control buttons ................................................................................................... 25
Settings tab ................................................................................................................... 25
Voltages tab .................................................................................................................. 25
Calculate matrix tab ...................................................................................................... 26
Loading recalculated tab ............................................................................................... 27
Graph, 3D Graph tab .................................................................................................... 27
Pressure&Force ........................................................................................................ 3
Introduction ............................................................................................................ 3
Channel settings...................................................................................................... 8
Program control .................................................................................................... 10
Single measurement...................................................................................................... 10
Show actual channel data ............................................................................................. 10
Automatic measurement panel ..................................................................................... 10
Use ........................................................................................................................ 28
Main dialog panel ......................................................................................................... 28
Profile editor window ................................................................................................... 29
Use ........................................................................................................................ 32
Initial settings - Settings tab ......................................................................................... 32
Initialization, reference movement - Init tab................................................................. 33
Coordinate systems - Coordinate system tab ................................................................ 34
Movement - Move tab .................................................................................................. 34
Automatic movement.................................................................................................... 35
Log file - Log tab .......................................................................................................... 36
3/37
Introduction
Pressure&Force
a measurement software for multi-channeled, pressure, force and hotwire measurements, synchronized with a probe positioning traverse
Introduction
o Starting the program
o Data structure
Main window
o Initial settings - Settings tab
o Traverse program tab
o Measurement file tab
Shifting between measurement points
o Actual channel data tab
o Time resolved data tab
FFT tab
Channel settings
Program control
o Single measurement
o Show actual channel data
o Automatic measurement panel
Features accessible from the menu
o Setting the atmospheric pressure and temperature
o Calibration
o Traverse settings
o Update time-resolved data paths...
o Convert time-resolved data...
Name
of
Labview
library: Pressure&force.llb
Version descriptions: pressure&force_versions.txt
Required for running the program:
ARA_Utils.llb, Transducer_calibration.llb
isel_ui4.x.llb,
PointEditor.llb,
Note: The current version's name always includes the version number (e.g.
Pressure&force_v3.21.llb)
Data structure
The main measurement file with a .txt extension consists of two main parts:
The channel settings (.chs) file gives the number of channels we are
going to use for data acquisition and how the measured voltage values
should be converted to physical quantities. The .chs file refers to the
calibration file (.txt), which contains the calibration constants. We can
also set if the measured voltage and deviation values should be written
in the measurement file during the measurement.
body: each line stores the measured mean and deviation values for
every channel, the atmospheric pressure and temperature values. At
the end of the lines the name of the datasetfile can be seen.
The dataset file (arranged in columns) contains the voltage
values measured on each channel and the time-resolved data of
the physical quantities calculated from them.
4/37
Main window
At the top of the main window the measurement control buttons are located,
beyond them the measurement data and the settings can be found on the tabs
of a Tab Control. Proceeding from left to right we begin with the settings and
end up with the results.
Note: after the first click on the Apply settings button the program
would like to save the file on the hard disk drive, and will ask where to
save it.
The measurement settings stored in the file:
5/37
In the bottom right corner of the tab the current position can be seen.
The settings located on the lower part of the tab are not being stored in the
measurement file. These are:
The measurement program (traverse program) that has been given on the
Measurement settings tab can be seen in the list on the left side. On right side
the points can be displayed in a 3D diagram, the point selected from the list is
marked with a different color. The diagram can be refreshed using the
Redraw button.
With the Point file editor the measurement program can be edited. The
detailed instruction manual can be found in the help of the Pointeditor.llb.
Note: The traverse program saved in the Point file editor has to be
chosen with the Measurement program...button on the
Measurement settings tab in order to be loaded into the main
program!
6/37
After selecting a line, the Name and Comment fields can be modified with
the Add comment & name button. The lines can be deleted using the Delete
line button. Multiple lines can be selected by pressing the Shift button.
Note: If the Save file after each measurement option (at the bottom of
the Measurement settings tab) is switched on, then the program will add
(save to the hard drive disk) the new line to the measurement file after
each measurement so that in case of a crash the important measurement
data will not get lost. The changes made on the header of the
measurement file or on the Name and Comment fields of the
measurement lines are only stored in the memory. Therefore at the end
of the measurement series it is recommended to save the main
measurement file one more time, for example by overwriting the original
measurement file.
In the table of this tab the measurement results can be seen line by line. The
columns are:
7/37
For the continuous and immediate display of the measured physical quantities
the Show actual channel data button can be used. On the left side of the
Actual channel tab the current values of the measured quantities (calculated
pressure, force, etc.) can be seen.
From the Show actual channel drop-down menu we can choose the channel
whose graph we would like to display in the diagram to the right. By
activating the Large display in separate window setting the current value of
the channel selected from the list will appear in a large window.
Note: We have to select the channels which should be displayed in the
large window before starting the measurement. The large display
window can be closed or opened during the measurement by unticking
or ticking the box of the Large display in separate window setting.
FFT tab
On this tab the Fourier transform of the measured data can be generated, with
various window functions, sample size, number of averages. The diagram can
be customized as shown in the case of the Time resolved data tab.
With the Recalculate FFT button the FFT can be recalculated with
the modified parameters.
With the Save FFT... button the FFT graph can be saved to a text file,
separated with Tabs.
Channel settings
The channel settings editor window can be reached by clicking on the
Channel settings... button on the Measurement settings tab. Here:
The lines of the table at the top can be modified by selecting (one or
multiple, by pressing the Shift button) them. The settings of the
selected channels appear in the cluster under the table, the
modifications can be made here.
With the Load and Save buttons the channel settings can be loaded
and saved to a file with a .chs extension. Saving the file is necessary,
because the main measurement file refers to the .chs file.
8/37
The dialog box can be closed with the OK and Cancel buttons.
We can select the channels from which we want to acquire data. (Edit
measurement channels... button). From the drop-down menu on the
left side of the dialog box the DAQmx global channels that have
already been defined can be selected or new ones can be created. The
order of the channel list can also be changed.
9/37
After closing the channel list editor window the measurement channels will
appear in the table in the bottom left corner.
After this we have to create the data processing channels, i.e. we have to tell
what shall be calculated from the measured voltage. We can provide this
information in the Add data processing channel... dialog box.
10/37
starts after pressing the button and lasts until the button is unpressed. The
display can be seen on the Actual channel data tab.
Note: do not make menu or file operations, and do not modify the
comments of the measurement lines during the measurements.
Single measurement
Pause button: stops the sequence after the end of the current measurement.
This way, for instance, the modification of the settings (e.g. sampling
frequency) can be made, and then the measurement can be continued. The
sequence continues after unpressing the button.
Program control
Go to program line> button: if pressed, the program jumps to the line given.
Thus a measurement program can be started at a different point instead of the
beginning, or if the running program is paused by the Pause button then
measurement positions can be skipped/repeated.
11/37
Barometer: dstp://152.66.21.31/setra470
Big
windtunnel
thermometer
(yellow):
dstp://152.66.21.31/GMH3710
Laboratory thermometer (blue): dstp://152.66.21.31/GMH3530
Calibration
The calibration of the linear transducer can be made in the Linear
transducer calibration... item of the Setting menu. The one component
hotwire calibration can be found in the 1-D hotwire calibration... item of the
Settings menu. The matrix calibration of multicomponent force measurement
systems can be made in the Multi-component balance calibration... menu
item. The detailed description of these can be found in the help of the
Transducer_calibration.llb. (It can be openedwith the Help button of each
dialog box.)
Traverse settings
The settings can be given:
manually
by retrieving the data from a thermometer or a digital barometer
connected to another computer which is accessible through the
network. The data can be retrieved by giving the so-called DataSocket
link of the instrument and pressing the Test channel button. After 1-2
seconds the measured value appears, or an error message if the data
could not be retrieved.
1 to 3-axis, ISEL type motion controllers can be used for the automatic
traverse movement performed by the Pressure & Force program. The settings
of the controllers (initialization, change of the coordinate system, manual
motion etc.) can be made in the Traverse settings item of the menu. The
detailed description of the motion controller can be found in the help of the
isel_UI4.x.llb.
12/37
Note: To save the modification of the main measurement file to the hard
disk drive, the file has to be saved with the Save measurement...!
The original files can be overwritten or the new files can be copied to another
directory. In the latter case the references of the main measurement file can
be redirected with the update measurement file links option.
With the update measurement file header option the general settings of the
time-resolved data can be set to a new format or to a new position in the file's
header.
Note: To save the modification of the main measurement file to the hard
disk drive, the file has to be saved with the Save measurement...!
13/37
Transducer calibration
Data
Introduction
Starting the program
o Independent run
o In case of a version built into another measurement program
Linear transducer calibration
o Calibration mode
o Measurement mode
o Diagram and error bar settings
Hotwire calibration
o Introduction
o Temperature compensation
o Use of the program
o Calibration mode
o Hotwire calibration with measurement in single points
o Curve fitting
o Hotwire calibration with blow down method
o Advanced tab
Matrix calibration for multiple load cell balances
o Introduction
o Use of the program
o The panel on the left
o Lower control buttons
o Settings tab
o Voltages tab
o Calculate matrix tab
o Loading recalculated tab
o Graph, 3D Graph tab
Other VI-s, LabVIEW technical information
14/37
Data
Name
of
Labview
library: Transducer_calibration.llb
Version descriptions: Transducer_calibration_versions.txt
Required for running the program: ARA_Utils.llb
Note: The current version's name always includes the version number (e.g.
Transducer_calibration_v4.15.llb). To use it in other programs (ARALDA.llb, Pressure&Force.llb) the version number has to be removed from the
file name (Transducer_calibration.llb) and the file has to be copied to the
subdirectory where the program is.
Introduction
For fluid mechanical measurements numerous measurement devices are used
which have analog voltage signal output, e.g. pressure transducer, load cell,
hotwire probe. Most of these have linear characteristics, i.e. there is a linear
relation between the measured physical quantity and the voltage, e.g. in the
case of pressure:
p[Pa] = slope * U[V] + zero
where slope and zero are the constants of the calibration curve.
15/37
2. Hotwire_calibration.vi: subprogram for the calibration of non-linear,
one-component hotwire probes (CTA)
3. Multiple_load_cell_calibration.vi: subprogram for the calibration of
multi-component force measurement systems.
By double-clicking on these , and pressing the Run button in the toolbar, the
selected program starts.
Calibration mode
The calibration mode can be opened with the New calibration button and can
be exited using the End Calibration button.
16/37
added to the chosen line with the Add reference value command. The wrong
measurement can be deleted with the Delete line command.
The curve fitting happens automatically after each measurement. The
calibration curve of a specific channel can be selected in the Show channel
graph drop-down menu under the diagram on the right. If required the curve
fitting and drawing can be repeated with the Redraw button.
With the Show error graph option the difference from the calibration curve
can be shown, here the X axis coincides with the calibration curve, the error
bars are horizontal lines. With this diagram any outstanding differences can
be identified, and in measurement mode it can also be seen if the verifying
measurements are within the error bar.
In the Channel list editor the measurement channels can be selected for the
measurement. After closing the panel the physical quantities measured with
the reference device can be seen in the first column of the table on the left, in
the following columns the measured voltage values (in [V]) of each channel
can be found.
The data of the channel selected in the Show channel graph drop-down
menu can be seen in the blue box under the diagram. The date of the
calibration and the ID of the transducer is set here manually! If required, the
constants can also be edited.
Before finishing the calibration the calibration files have to be saved with the
Save to file... command. The program offers the channelname.txt file name.
Measurement mode
Initially the program is in measurement mode. If we have not done it already,
the channels through which we want to measure should be selected in the
Channel list editor. Using the Open file... button the calibration files can be
opened. Here, a calibration file can be assigned to each channel by selecting
the line and pressing the Add button.
In calibration mode after setting the channels, the voltage measurements can
be carried out by pressing the Measure button. The sampling time and the
sampling frequency can be modified. The results of the measurement appear
in the table. The physical quantity measured with the reference device can be
17/37
The loaded calibration curves appear in the diagram on the right, and in the
blue bordered box under it. If it is needed the Show channel graph dropdown menu and the Redraw button can be used.
Zooming and shifting is possible with the 3 gray buttons under the
diagram.
By right-clicking on the axes the autoscale can be turned on/off and
the axis can be formated.
The appearance of the points and curves can be changed by rightclicking on the legend.
With the Show error graph option the error of the curve fitting can
be shown.
In the Error bar settings box we can set the error type which should be
displayed:
The mean error of the curve fitting
The maximum error of the curve fitting
An absolute error value (fix error value, can be set manually, for
example it can be the measurement error given in the specification of
the device, e.g. +/- 5 Pa)
Absolute error value given in the percentage of a measurement value
(e.g. the measuring range of the device)
Hotwire calibration
Introduction
Control measurements can be made with the Measure button. The difference
between the calibration curve and the control measurement can be zeroed
with the Zero channels with selected line button. Prior to this a line has to be
selected from the table to which we want to fit the calibration curve.
The modified calibration can be saved with the Save to file... command. It is
advisable to simply overwrite the existing calibration file.
18/37
Calibration mode
Knowing the constants - similarly to the sensors with linear characteristics the flow velocity can be determined from the voltage signal.
The calibration mode can be opened with the New calibration button and can
be exited using the End Calibration button. With the help of the Open file...
/ Save file... the calibration files with .htwcal extension can be loaded and
stored.
The following settings have to be made on the Settings tab before starting the
calibration:
Temperature compensation
In contrast to most of the pressure transducers and load cells, the CTA
systems are not temperature compensated, i.e. the temperature change of the
flowing fluid causes error in the measurement. Therefore the temperature of
the fluid has to be measured during the calibration and the measurement, and
a corrected voltage signal (referring to the calibration temperature) can be
calculated from the current voltage signal.
19/37
o
dstp://152.66.21.31/GMH3710
Laboratory
(blue):
Note: the thermometer can be connected to the computer with which the
measurement is carried out, e.g. if we want to measure the temperature of the
air flowing out from the calibrator. In this case the NI DataSocket Server and
the GMH3710.llb driver have to be started. The program gives the current
DataSocket URL, which can be given in the dialog box seen above.
The p0 and T0 values can be refreshed with the Update button.
Wire settings: The data of the hotwire sensor can be set here:
o Wire channel: the data acquisition channel measuring the
output voltage of the hotwire bridge
o Wire ID: the ID/serial number of the sensor. It is practical to
set it because, for example when comparing with former
calibrations, it can turn out that the hotwire is damaged if its
cold resistance changes significantly.
o Overheat ratio (a): the ratio of the hot (during the
measurement) and cold (before the calibration, measured at
ambient temperature) resistance of the hotwire. This is
important because the feedback control of the CTA measuring
bridge will keep the hot resistance of the wire at a constant
value.
manually
thermometer
dstp://152.66.21.31/GMH3530
It can be set:
o
a = R_hot / R_ambient
Cold resistance [ohm]: the wire resistance value determined
before the calibration with the help of the bridge.
20/37
The calibration measurements appear in the table on the right, i.e. the voltage
measured on the pressure transducer, the pressure and velocity calculated
from it, the voltage measured on the hotwire and the current temperature. The
lines can be deleted with the Delete line(s) button. (Multiple selection:
shift+click)
Curve fitting
The measurement values shown in the table of the Calibration tab also
appear in the diagram of the Curve fitting tab. The diagram can be refreshed
with the Redraw button.
21/37
In the Iteration constants frame the type of the fitted curve can be selected
(Curve fit type), it also contains the results of the curve fitting iteration, the
constants and the difference between the points and the fitted curve in m/s
(Mean error, max error). The fitting is done by the Fit curve button.
In the Error bar settings frame the displayed error bars can be chosen. If the
Show error graph is checked then the digram shows the difference of the
points around the fitted curve, and not the fitted curve. If the Show both
curve fits is checked then both curves (according to King's law and the
polynomial) are displayed. (The 'selected' curve is the one chosen in the
Curve fit type. The curve fitting for both types has to be made beforehand!)
1. The data acquisition is started with the Measure blowdown button.
2. The blow down begins after pressing the Scan button on the hotwire
calibrator, after some time the velocity decrease stops. If the
3.
4.
5.
6.
With the help of the zoom and Number of data points the blowdown points
can be adjusted arbitrarily. If we do not like the added points, then they can
be deleted with the Delete selected line(s) button on the Calibration tab.
It is worth repeating the blow down at different states of the calibrator's
FLOW control, thus the points cover a wider velocity range.
Advanced tab
Iteration settings
o Termination: the parameters of the curve fitting iteration
according to King's law can be set here.
o Extrapolation limits: the fitted curve can be extended beyond
the calibration limits (min. and max. voltage).
o Error out: error message of the iteration routine
Velocity indicator moving average: the actual flow velocity value
shown on the Calibration tab is formed from the moving average of
the current measurements. By increasing the length of the moving
average, the fluctuation of the displayed value can be decreased.
22/37
23/37
The relation is nonlinear, therefore even a forth or fifth order fitting can be
necessary. However the cells' effect on each other is not always considered in
higher order products.
aij E j
j =1
1
424
3
lineris klcsnhats
b112
b122
c11
b 211
b 212
b 222
c 21
F = K E
b111
Fi = f (E j )
Fi =
F1 a11 a12
F = a
2 21 a 22
E1
E
22
E
c12 1
E1E 2
c 22 2
E2
3
E1
E3
2
+ bijk E j Ek + cij E j
j =1 k = j
j =1
1
4
4244
3 1
424
3
msodfok ( szorzat )
klcsnhats
harmadfok
klcsnhats
For example when measuring two force components with two cells, the third
order relation is:
where K is the calibration matrix, F is the force and torque vector, E is the
vector of cell voltages' products with different orders. The number of
constants in one line of the K matrix is: 2 n + ( n 1 ) n 2 . From the table
below it can be seen that in case of 6 force and torque components there can
be about 200 constants.
Number of force and torque The size of the calibration matrix
components
linear 2nd order 3rd order
2
2x2
2x5
2x7
3
3x3
3x9
3x12
4
4x4
4x14
4x18
5
5x5
5x20
5x25
6
6x6
6x27
6x33
Accordingly during the calibration the number of the recorded Load1,
Load2,...load combinations have to exceed significantly the number of
constants.
From the F and E vectors recorded during the calibration the following
equation system can be formed for every force component:
F1, Load1 E1 E 2
F
1, Load 2 = E1 E 2
... .
.
24/37
E1E 2
E2
E1
E1E 2
.
E2
.
E1
.
E1
E1
.
a11
a
12
3
E 2 Load1 b111
3
E 2 Load 2 b112
... b122
c11
c
12
The values of the F vector contain the different loads for the same force
component, the lines of the E... matrix contain the measured voltages for a
particular load value.
The relation above can be solved as a linear equation system, the vector of the
unknown constants [ a 11 . . . . c 12 ] can be determined. (Labview Solve
Linear Equations VI) If there are more calibration measurement results than
needed for the solution of the equation system, then the routine returns the
approximate value calculated with the method of least squares. By
performing the operation for every component, the K calibration matrix can
be obtained. During the solution it is supposed that at 0 load the voltage of
the cells is also 0. As this is not true in practice, the cell voltages measured
before the calibration at 0 force load (Zero Vector) are to be extracted from
the measured voltage.
The Dimension variable gives the number of measured force and torque
components. The former data is deleted when the Dimension is adjusted. The
Order sets the order of the calibration relation. On the "Settings" tab of the
tab control on the right a table shows the calibration matrix (Parameter
names) belonging to the set dimension and order.
25/37
Settings tab
With the help of Open file... / Save to file... the calibration files with .mcal
extension can be loaded and saved. The files are in ASCII format. The arrays
recorded during the calibration can be saved to a text format separated by
Tabs, which can be easily loaded to Excel, using the Save Arrays button.
On the upper part of the tab the constants of the calibration matrix
belonging to the set dimension and order can be seen.(Parameter
names)
In the middle the data of the calibration can be set: Balance ID,
Personnel, Date, Reference unit, Calibration comment
At the bottom the Sampling frequency and the sampling time of the
calibration measurement can be set, and the set channels can also be
seen.
26/37
The Zero Vector stores the cell voltages measured in the unloaded state of
the balance. The Tare vector cannot be set in this program, later it is used for
subtracting the constant non-aerodynamic loads (e.g. the weight of the
vehicle model) during wind tunnel measurements. It is NOT allowed to use
the Zero Vector for this purpose.
27/37
On the tab the load recalculated with the calibration matrix and the difference
between the recalculated and measured values, i.e. the error of the calibration
can be seen. By examining the latter the calibration are very much out of
place can be detected. points which are very much out of place can be
detected. (The calibration can be corrected by deleting or remeasuring these
points.)
On the two tabs the recorded calibration points and the calibration relation
can be displayed in 2D and 3D. The content of the axes can be modified
(Graph axes). For example the impact of one force component on the voltage
of another cell (cross impact) can also be displayed.
The Redraw button refreshes the diagram.
The root-mean-square error of the recalculation (Recalculation RMSE) and
the relative value of it, applied to the maximum load (Recalculation relative
error) can be found under the tables.
28/37
library: PointEditor.llb
Required for running the program: Note: The current version's name always includes the version number (e.g.
PointEditor_0.17.llb). To use it in other programs (ARA-LDA.llb,
Pressure&Force.llb) the version number has to be removed from the file
name (PointEditor.llb) and the file has to be copied to the subdirectory where
the program is.
Introduction
By using the program, measurement point meshes, tables consisting of X, Y,
Z point triplets can be generated, and saved to text files separated by Tabs.
All these could be done by spreadsheet softwares, however the PointEditor
has multiple advantages:
Introduction
Starting the program
o Independent run
o In case of a version built into another measurement program
Use
o Main dialog panel
o Profile editor window
LabVIEW technical information
Name
of
Labview
Version descriptions: pointeditor_versions.txt
Use
Main dialog panel
29/37
The finished point mesh can be displayed in two types of diagrams depending
on whether the 3D diagram box is selected or not:
The definitions of the angles can be seen in the figures below, in case of
directions x+, y+ and z+ :
With the Add profile to points command the profile is added to the given
reference point. With the Undo command the profile additions can be
withdrawn, with the Reset button the whole point mesh can be deleted.
Measurement point files with .pts extension (text files separated by Tabs) can
be loaded with the Load file... command and saved with the Save points....
One dimensional point series can be created here. After giving the minimum
and maximum values, linear profiles with given distance between the points
or with given number of points, exponential profiles or profiles with given
30/37
31/37
Introduction
Starting the program
o Independent run
o In case of a version built into another measurement program
Use
o Initial settings - Settings tab
o Initialization, reference movement - Init tab
o Coordinate systems - Coordinate system tab
o Movement - Move tab
o Automatic movement
o Log file - Log tab
Other VI-s, LabVIEW technical information
Name
of
Labview
Version descriptions: isel_ui4.x_versions.txt
graphical interface and allow easy control of the linear units from other
Labview
programs.
The stepper motor controller is able to perform the following functions:
library: Isel_UI4.x.llb
Independent run
Note: The current version's name always includes the version number (e.g.
isel_ui4.x_v1.18.llb). To use it in other programs (ARA-LDA.llb,
Pressure&Force.llb) the version number has to be removed from the file
name (isel_ui4.x.llb) and the file has to be copied to the subdirectory where
the program is.
After clicking on the Isel_ui4.x.llb a Labview window opens and the contents
of the library appear. In the front, there is the options.vi, by double-clicking
on it and pressing the Run button in the toolbar, the main program of the
positioning traverse starts.
Introduction
The linear units applicable for moving the probe are usually driven by stepper
motors. The limit switches indicating the end points of the stepper motor and
the shafts are managed by the stepper motor controllers of the ISEL
Automation company. The motion controller is connected to the computer
with a serial port, and commands can be sent and feedback can be received
(e.g. the current position) using the Isel UI 4.x or the UI 5.x serial protocol.
The Labview subroutines (VI-s) introduced herein provide a user friendly
32/37
The properties of the axes are indicated on the linear units, these have to be
written into the Traverse Properties frame:
Use
The functions of the program are located on the tabs of a Tab Control, the
settings have to made from left to right. At the bottom of the window the
error feedback can be seen on the execution of the operation, a green tick and
the error code 0 indicates the error free execution of the command. In case of
position request, the returned position can be seen in the source field in
hexadecimal numbers.
Note: After the first start of the program only the first tab of the panel
can be seen, the other tabs are blocked so that the order of the starting is
kept.
33/37
on the side of the motor, after reaching the limit switch they change direction
and they leave the limit switch at slow pace, and they stop. The reference
movement has to be made after every initialization, otherwise the program
does not know the position of the slides.
Note: Before the reference movement, it should be checked that during
the motion the axes will not hit the stationary parts of the measuring
equipment or of the wind tunnel. If it is necessary the Reference STOP
button or the emergency switch can be used.
The motion controller is able to move 2 axes synchronously, in this case the
probe will move along a diagonal. After selecting the two axes, this can be set
by pressing the Set interpolation plane button. In order to avoid collisions it
is easier not to use this feature.
If - for example after the restart of the software, if the motion controller itself
has already been initialized - the initialization and the reference movement
are not needed, then with the Skip start procedure button the block can be
removed from the tabs, which perform motional operations.
The motion controller has to be initialized after it has been turned on. Among
the communication settings the serial port (COM1, COM2 etc.) of the motion
controller and the speed of the communication have to be given. Generally
this is 9600 baud, which is set by jumpers in the motion controller. The
initialization means that the internal motion controller program is restarted
and the internal position memory is cleared. The number of axes to be used
has to be given, and the Initialize button has to be pressed.
Note: The initialization also has to be made after the Reset button,
situated on the front panel of the motion controller is pressed.
The next step is the movement to the zero point of the axes, which is called
the reference movement (Find reference). The slides move to the endpoint
34/37
The functions of the new coordinate system can be modified manually on the
right side of the tab. The conversion is done by the Set new coordinate
system button.
Note: the change can only be made from hardware to software or from
software to hardware coordinate system, it is not possible to switch from
a software to another software. First the switch has to be made to the
hardware and then to the new software.
In the upper right corner of the tab the current limits of the motion are shown,
the endpoints of the axes in the applied coordinate system (Coordinate system
limits).
35/37
Automatic movement
Ask position : retrieving the position, the last known position of the
traverse is shown in the fields.
Move (abs): move to the given point of the coordinate system. The 0
and A buttons serve for zeroing and for taking the last requested
position to the target coordinate fields.
Move (rel) : move to a given distance from the current position. The 0
button serves for the zeroing of the fields.
During every motion operation a small dialog box appears with a big STOP
button. Therefore the motion can also be aborted from the software. During
the motion, the reading of the position does not work, thus the position
feedback in this panel reports an error.
36/37
The moving software records every completed operation in a log file, which
can be useful e.g. for troubleshooting. Its location can be given on the
Settings tab, its content can be seen on the Log tab.
The isel_ui4x_3.vi does the serial communication with the motion controller.
The Isel_UI4.x.llb contains further SubVI-s for the insertion of the
positioning traverse into other measurement programs. I can provide further
information on these personally.
37/37