Documente Academic
Documente Profesional
Documente Cultură
242249
Advance Access publication 18 January 2010
doi:10.1093/rpd/ncp303
INTRODUCTION
In recent years, high-energy medical accelerators
have been the main tools in radiation therapy
centres. Higher beam penetration, low scattering of
beam to normal tissues and skin sparing effect are
the advantages of high-energy photon beams. But,
the use of high-energy photons may lead to the production of secondary photoneutrons which require
extra radiation protection measures in terms of
room shielding and possible patient whole body
dose(1 4). When photon energy becomes higher than
the threshold energy required for photonuclear ( g,n)
interactions, photoneutrons are produced in
materials which are in the path of photons, such as
target, primary collimator, flattening filter, jaws and
lead shielding of head as well as materials found in
the walls of treatment room(5 8). Shielding against
these highly penetrating radiations has been a challenging concern for radiation protection scientists(9,10). Otherwise, although room concrete is used
as a shielding for photons and photoneutrons, concrete walls in high-energy X-ray rooms also act as
photoneutron sources(11).
According to International Atomic Energy
Agency (IAEA)(12), total photoneutron fluence,
wA, at the inner maze per gray X-ray at the isocentre
(n m22 Gy21) is given by the following equation:
wA
QN
5:4QN 1:26QN
2
4pd
2pS
2pS
exp
dEn
T2
T
0:1071 LnEmax =En 7:34
Emax 7:34
LnEmax =En 7:34 dEn
0
# The Author 2010. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
243
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
A. MESBAHI ET AL.
Rooms layouts
To study the effect of room geometry on photoneutron fluence and photoneutron dose, four different
room layouts with ordinary concrete with the density
of 2.35 g cm23 and the dimensions of 673.5 m3
were simulated. The height and width of maze for
all layouts was 3.65 and 2 m, respectively, except
layout 4. The width of the maze was considered 4 m
for layout 4. Figure 2 shows the layout of rooms and
locations for which calculations were performed.
Three spherical cells with 10 cm radius and the
material of water were simulated so that one of them
located at the maze entrance door (cell 3) and the
others was positioned at the distances of 350 cm (cell
2) and 700 cm (cell 1) from the maze entrance door
(Figure 2).
Analytical methods
We used two analytical methods, which were proposed by two studies(15,30) and also recommended
by IAEA 47 for photoneutron dose calculations in
four layouts.
244
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
Dn 2:4 1015 wA
r
Ar
1:64
S1
Dn H1 103
245
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
Figure 3. Photoneutron spectra at the distance of 1 m from the target derived from full simulated linac without room.
A. MESBAHI ET AL.
Table 1. Photoneutron dose equivalent in different layouts for three points including a point at the maze entrance door,
a point in the middle of the maze and a point at the inner maze entrance.
Location
MCNPX
Difference (%)
2.45E202
1.69E203
2.54E204
2.20E202+2.71E204
1.26E203+5.34E205
2.20E204+1.55E206
10
34
13
2.45E202
1.69E203
2.54E204
2.23E202+2.71E204
1.01E203+9.16E205
2.20E204+1.55E206
7
40
13
9.78E203
1.69E203
1.79E207
8.16E203+1.03E205
5.51E204+4.96E206
1.34E207+3.57E209
16
67
25
9.78E203
1.69E203
1.79E207
8.64E203+2.38E205
7.86E204+1.02E205
1.02E207+1.02E208
11
53
43
Room 1
1
2
3
Room 2
1
2
3
Room 3
1
2
3
Room 4
1
2
3
MCNPX
Difference
(%)
2.78E203 2.01E203+2.3E205
4.40E204 4.24E204+5.01E205
1.17E204 9.81E205+9.95E207
27
22
16
2.78E203 1.81E203+2.01E205
4.40E204 4.34E204+5.21E205
1.17E204 9.98E205+9.95E207
34
13
17
2.78E203 2.01E203+2.3E205
4.40E204 3.59E204+5.61E206
1.17E204 8.59E206+8.90E208
27
18
86
2.78E203 2.01E203+2.3E205
4.40E204 3.02E204+3.85E206
1.17E204 5.62E206+5.96E208
27
31
95
246
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
Room 1
1
2
3
Room 2
1
2
3
Room 3
1
2
3
Room 4
1
2
3
McGinley
CONCLUSION
In the current study the photoneutron and capture
gamma dose were calculated by MC and recommended analytical methods in four different
room layouts. Our results showed that the analytical
methods either for single- or double-bend mazes
overestimate the dose at the maze entrance door significantly for both photoneutrons and capture
gamma rays. Moreover, additional bend caused a
significant reduction in both photoneutron and
capture gamma dose equivalents. It can be concluded that room and maze layouts have significant
effects on maze received photoneutrons and capture
gamma rays. This should be considered in
247
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
A. MESBAHI ET AL.
14.
FUNDING
16.
17.
15.
REFERENCES
1. Allen, P. D. and Chaudhri, M. A. The dose contribution due to photonuclear reactions during radiotherapy. Med. Phys. 9(6), 904 906 (1982).
2. Vanhavere, F., Huyskens, D. and Struelens, L.
Peripheral neutron and gamma doses in radiotherapy
with an 18 MV linear accelerator. Radiat. Prot. Dosim.
110(1 4), 607612 (2004).
3. Zanini, A. et al. Monte Carlo simulation of the photoneutron field in linac radiotherapy treatments with different collimation systems. Phys. Med. Biol. 49(4),
571 582 (2004).
4. Attix, F. H., Rank, E. X., August, L. S., Miller, G. E.
and Shapiro, P. A shielding maze at a neutron radiotherapy facility. Health Phys. 31(1), 78 80 (1976).
5. Facure, A., da Silva, A. X. and Falcao, R. C. Monte
Carlo simulation of scattered and thermal photoneutron
fluences inside a radiotherapy room. Radiat. Prot.
Dosim. 123(1), 5661 (2007).
6. Mesbahi, A. A Monte Carlo study on neutron and electron contamination of an unflattened 18-MV photon
beam. Appl. Radiat. Isot. 67(1), 5560 (2009).
7. Waller, E. J., Jamieson, T. J., Cole, D., Cousins, T. and
Jammal, R. B. Experimental and computational determination of neutron dose equivalent around radiotherapy
accelerators. Radiat. Prot. Dosim. 107(4), 225232
(2003).
8. Zanini, A., Durisi, E., Fasolo, F., Visca, L., Ongaro,
C., Nastasi, U., Burn, K. W. and Annand, J. R.
Neutron spectra in a tissue equivalent phantom during
photon radiotherapy treatment by LINACS. Radiat.
Prot. Dosim. 110(1 4), 157160 (2004).
9. Facure, A. and Silva, A. X. The use of high-density
concretes in radiotherapy treatment room design. Appl.
Radiat. Isot. (2007).
10. Waller, E. J., Jamieson, T. J., Cole, D., Cousins, T. and
Jammal, R. B. Effectiveness of customised neutron
shielding in the maze of radiotherapy accelerators.
Radiat. Prot. Dosim. 107(4), 233 238 (2003).
11. McCall, R. C. Shielding for thermal neutrons. Med
Phys. 24(1), 135136 (1997).
12. International Atomic Energy Agency (IAEA).
Radiation protection in the design of radiotherapy facilities. Safety Reports series No. 47 (IAEA, Vienna)
2006.
13. Followill, D. S., Stovall, M. S., Kry, S. F. and Ibbott,
G. S. Neutron source strength measurements for
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
248
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010
249
Downloaded from rpd.oxfordjournals.org at TABRIZ University of Medical Sciences on September 17, 2010