Documente Academic
Documente Profesional
Documente Cultură
OPTIONAL
Antenna Extension : 6 or 8 Lengths.
- Use only with Teflon Rod Antenna !
Refer to catalogue numbering Page for
ordering info.
High Temperature Unit :
- Teflon De-coupler threads on to
1 1/2NPT Mtg. Adaptor with a
2 NPT mtg. Process Connection.
APPLICATIONS
Food and Beverages
Water / Wastewater
Chemicals with vapors
MECHANICAL
Conduit Entry
Enclosure
Ingress Protection
: 1/2 NPT x 2
: Aluminum or S.S. - 94V0
: NEMA 4 (IP65) Type 4 / 4X
ENVIRONMENTAL
Antenna Extension
OPERATIONAL
: Pulse Radar
: +/- 0.1% of max. range in lab using
4-20mA current output
: +/-0.25% of max. range (typically in field )
Response Time : Std. Unit 2 - 3 echos per sec.
PROCESS
: Std. with less damping 6 echos / sec.
Temperature PP Rod : - 40 to 176F (- 40 to 80C)
: Fast Protocol Unit 10 - 30 echos / sec.
Frequency
: 5.8 GHz or 6.3 GHz.
De-coupler & PTFE Rod - 40 to 350F (- 40 to 177C)
Loss of Echo
: Hold 1 min., 22mA or 2 mA. output
Material Dielectric
: Er >2
Transmit Power : 50 uW average
Max. Pressure
: 5 bar (without De-coupler)
Calibration
: Via communications port (required)
Diagnostics
: (Echo Profile) via communications port
Catalogue # - On the Web return to Home Page
: Dielectric rod PP, & Optional PTFE
& Refer to Catalogue Number Structure for Ordering Antenna
Horn 316 S.S. for very low dielectric
information. In Product Documentation refer to Page 4.
constant materials and short blanking
Temperature
Approvals
: - 40 to 140F (- 40 to 60C)
: FCC Part 15 - Low Power
Communication Device
Installation Category : Class II
Operation
Accuracy
TECHNICAL SPECIFICATIONS
Radar Range
Code
017
033
050
100
240
RANGE
RESOLUTION
MOUNTING
In Liquids
* - 17 ft.
* - 5m
0.08
* - 33 ft.
* - 10 m
0.15
* - 50 ft.
* - 15 m
0.22
* - 100 ft.
* - 30 m
0.44
11 mm
* - 240 ft.
* - 73 m
1.06
26 mm
2.0 mm
3.9 mm
5.7 mm
ELECTRICAL SPECIFICATIONS
**
2.0,1 1/2 NPT
**
2.0,1 1/2 NPT
**
**
Power
ABM400 AC
**
Inter-Connection Diagram
FUSE
- F1/8 A 250V
8
L1
AC/DC
IN
L2/N 7
DC Power Input
Note - TB #7 is
Connected to TB #4
4-20 mA
Control Panel
Fastening Screw
RS232/485 A/TX 2
6
+5
4
B/RX 1
Terminal Block
STATUS
Status LED
CALIBRATION
SWITCH
ABM
SENSOR TECH. INC
Calibration Pushbutton
NOTE Use only 1/2 NPT Conduit
Typical Installation
1) Direct mounting ultrasonic sensor - Simply thread sensor directly into metal or
plastic nozzle.
2) Radar unit must be installed into metal fitting with the antenna pointing downward.
3) Do not mount in the center of a domed tank.
12 - 30 Vdc
115 / 230 Vac
60 / 50 Hz
CALIBRATION PUSHBUTTON
/ Led Indicator For 4 - 20 mA (750 ohm Max.)
OPTIONAL Programmable Calibration &
Diagnostics, Data Logging , RS232 or RS485
Communications.
Temperature Compensation for the speed of
sound for ultrasonic.
Deadzone
FULL
20 or 4 mA
LED
Red
+/0.25%
S
P
A
N
R
E
S
O
L
U
T
I
O
N
Per
Chart
Yellow
A
C
C
U
R
A
C
Y
Material
Level
EMPTY
4 or 20mA
Operation - An ultrasonic/electromagnetic pulse is transmitted from the
ABM sensor . The pulse travels to the surface being monitored and is
reflected off this surface back to the sensor . The time of flight is divided
by 2 ,corrected with temp. and converted to an output signal directly
proportional to the material level .
Wiring Information
- Ground shield at one end only.
- All terminal block wiring must be rated for 250V.
- Power input wiring must be protected by a 15A double pole
circuit breaker .
- Terminal is for use only with equipment which has no live
parts which are accessible .
- Terminal is for use with equipment which maintains basic
insulation from hazardous voltage under normal and single
fault conditions .
- Connection used at the remote end of external circuit .
Recommended Wiring
For AC Sensor
Power
3 Wire unshielded 22 AWG , 300 V
Current Output 1 Pair shielded 24 AWG , 300 V
Communication 1 Pair shielded 24 AWG , 300 V
For DC Sensor
Power & Current output 3 Wire shielded 24 AWG , 300 V
Communication
1 Pair shielded 24 AWG , 300 V
1
2
3
4
5
6
7
8
9
-- 4
3
DB9
A/TX 2
Red
B/RX 1
WHT. or BLK.
L1
8
L2/N 7
6
+ 5
SHIELD
E-mail: info@abmsensor.com
L1
8
L2 / N 7
RS485 to
RS232
Converter
To PC
Serial Port
+ 5
-- 4
3
A/TX 2
B/RX 1
Red
WHT. or BLK.
Connect Shield to
Ground at one End
To Power Source
FUSE
+ 8
24VDC
-- 7
6
+ 5
4-20 mA
-- 4
3
!
RS232/485 A/Tx 2
B/RX 1
STATUS
Calibration
Sensor Calibration
Switch & Status Led
ABM
ULTRASONIC
RADAR
Model#
ABMXXX XXXRXCX-XXYY-AEX
POWER RATING : 12-30 Vdc , 0.07 A
: max. @ 24Vdc
AMB. TEMP. : - 40 TO 60 Deg. C.
OUTPUT: 4-20mA , RS
ENCLOSURE: NEMA 4x , IP65
De-energize circuit before removing
access cover.
SERIAL#
ID#
Nameplate Recess
Cover Tool Access Hole
7.3
Electronic Enclosure
Aluminum or S.S.
3.5 O.D.
1
1/2 NPT Conduit Entry
18.6
Mounting Thread 2or 1 1/2 NPT
(Aluminum or S.S.)
1.5 DIA.
Antenna Polypropylene (Std,)
Teflon (optional)
10.3
Radar ABM300-XXXR
Complies with FCC Part 15
FCC I.D. # QVK-ABMXXX
I.C. 4509A-ABMXXX Made in Canada
SERIAL# _______________Manufactured :_______
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference and
(2) this device must accept any interference received,
including interference that may cause undesired operation