Documente Academic
Documente Profesional
Documente Cultură
MAC 400
Resting ECG Analysis System
Version - Product Code SCT
Service Manual
2036039-001 Revision D
NOTE
The information in this manual applies to MAC 400 resting ECG analysis systems with product code SCT.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
MUSE, CASE, MAC, MARS, and 12SL are trademarks owned by GE Medical Systems Information Technologies,
a General Electric Company going to market as GE Healthcare. All other trademarks contained herein are the property
of their respective owners.
2007, 2008. General Electric Company. All Rights reserved.
T-2
MAC 400
2036039-001D
21 August 2008
Contents
1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Manual information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Revision history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Manual purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Responsibility of the manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Equipment symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Service information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Equipment identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Serial number format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Serial number label (non-China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Serial number label (China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Product label (non-China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Product label (China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
CE marking information (non-Chinese only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
2036039-001D
MAC 400
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Recommended maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Required tools and supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Inspections and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Leakage current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Earth leakage (AC line) current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Enclosure (chassis) leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Patient leakage current tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Patient leakage current to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Patient leakage current, mains on applied part . . . . . . . . . . . . . . . . . . . . . . . .3-8
Replacing components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Replacing the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Replacing the primary fuses in the power inlet module . . . . . . . . . . . . . . . . . . . .3-11
Opening the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Replacing the thermal printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Replacing the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Replacing the PCB assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Replacing the RTC battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
Replacing the OTPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Replacing the 12V power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Replacing the power inlet module along with the toroid assembly . . . . . . . . . . . .3-19
Adjustment instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20
Functional checkout procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Tools needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-21
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Functional checkout procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Operational checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Diagnostic tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
Electrical safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Device connected to mains but does not turn on . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Green LED does not light up and device does not turn on . . . . . . . . . . . . . . . . . . .4-2
Green LED lights up and device does not turn on . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Battery only connection and device does not turn on . . . . . . . . . . . . . . . . . . . . . . .4-4
Blank display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Errors detected during self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
ii
MAC 400
2036039-001D
2036039-001D
MAC 400
iii
iv
MAC 400
2036039-001D
2036039-001D
Introduction
MAC 400
1-1
Introduction
Manual information
Revision history
The document part number and revision letter are on the bottom of each page. The
revision letter identifies the documents update level. The revision history of this
document is summarized in the following table.
Table 1-1: Revision history, PN 2036039-001
Revision
Date
Comment
2 July 2007
5 November 2007
2 January 2008
21 August 2008
Manual purpose
This manual provides technical information for service representatives and technical
personnel so they can maintain the equipment to the assembly level. It serves as a
guide for maintenance and electrical repairs that are considered field-repairable.
Where necessary, the manual identifies additional sources of relevant information
and/or technical assistance. See the operator guides for the instructions necessary to
operate the equipment safely in accordance with its function and intended use.
Intended audience
This manual is intended for the person who maintains or troubleshoots this
equipment.
Intended use
The MAC 400 device is intended for use under the direct supervision of a licensed
healthcare practitioner. The system is intended to acquire, measure, and record
information from adult and pediatric populations. The basic system delivers 3channel ECG recordings in automatic and arrhythmia modes and 1- or 3-channel
ECG recordings in manual mode. The arrhythmia detection provides the
convenience of automatic documentation. It is not designed to provide alarms for
arrhythmia detection.
1-2
MAC 400
2036039-001D
Introduction
Contraindications
This system is not intended for use as a vital signs physiological monitor or for use
during patient transport.
This device is not intended for use with high-frequency surgical units. Disconnect
the patient from the device before using the high-frequency surgical units.
This device is not intended for use with direct cardiac applications.
Conventions
Table 1-2: Conventions
Bold text
Italicized text
Safety information
Definitions
The terms Danger, Warning and Caution are used throughout this manual to point
out hazards and to designate a degree or level of seriousness. Familiarize yourself
with their definitions and significance.
Hazard is defined as a source of potential injury to a person.
Table 1-3: Hazard terms
Term
2036039-001D
Definition
DANGER
WARNING
CAUTION
NOTE
MAC 400
1-3
Introduction
Safety messages
Additional safety messages that provide appropriate operation information may be
found throughout this manual.
DANGER
EXPLOSION HAZARD Use near flammables may cause
explosion.
Do not use in the presence of flammable anesthetics, vapors or
liquids.
WARNING
CONNECTION TO MAINS The mains plug must be
connected to an appropriate power supply.
Operate the unit from its battery if the integrity of the protective
earth conductor is in doubt.
The electrical installation of the relevant room complies with the requirements
of the appropriate regulations.
General
Failure on the part of the responsible individual, hospital, or institution using this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failure and possible health hazards.
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE.
Parts and accessories used must meet the requirements of the applicable IEC 60601
series safety standards and essential performance standards.
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety. Consideration
relating to the choice should include:
1-4
Evidence that the safety certification of the ACCESSORY has been performed
in accordance to the appropriate IEC 60601-1 harmonized national standard.
MAC 400
2036039-001D
Introduction
Equipment symbols
See the MAC 400 Resting ECG Analysis System Operators Guides for
information about the symbols used on this product and its packaging.
Service information
Refer equipment servicing to GE authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that warranty.
It is the users responsibility to report the need for service to GE or to one of their
authorized agents.
Equipment identification
025
2036039-001D
MAC 400
1-5
Introduction
026
Description
Year of manufacture
Miscellaneous characteristic
027
1-6
MAC 400
2036039-001D
Introduction
Description
E
D
C
B
A
F
G
079
Description
Production license
Serial number
2036039-001D
MAC 400
1-7
Introduction
028
Description
Site of manufacture
A
B
C
D
E
080
1-8
MAC 400
2036039-001D
Introduction
Description
Product name
Company name
Company address
Rated current
Fuse
Type
2036039-001D
Description
2033547-001
MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement feature (non-Chinese only)
2033547-002
MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement and interpretation feature (non-Chinese only)
MAC 400
1-9
Introduction
1-10
Description
2033547-003
MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement feature (China only )
2033547-004
MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement and interpretation feature (China only )
MAC 400
2036039-001D
2036039-001D
Equipment Overview
MAC 400
2-1
Equipment Overview
General description
The MAC 400 resting ECG analysis system is a 12-lead, 3-channel system with a
2-line x 16 character display, passive patient cable and battery operation.
Top view
029
2-2
Description
Keyboard
Display
MAC 400
2036039-001D
Equipment Overview
Side view
030
2036039-001D
MAC 400
Description
Connect the patient cable.
2-3
Equipment Overview
Rear view
031
Description
Product label
AC mains connector
2-4
Power block
MAC 400
2036039-001D
Equipment Overview
Keyboard interface
Display interface
Clocks
PCB assembly
CPU
VCC monitor
Memory block
The available functions and operation of the MAC 400 system are described in the
operators guides.
Mechanical structure
The major mechanical components of the MAC 400 device are the top and bottom
covers. The bottom cover is the basic element housing the following sub-assemblies:
Aluminum plate with a power supply module and toroid assembly on the
bottom and the PCB assembly on the top. The display, the keyboard connector,
printer connector and patient cable connector are mounted to the PCB assembly
The top cover houses the display window, paper cutter and keyboard, which is
linked to the main PCB assembly via a flexible cable.
2036039-001D
MAC 400
2-5
Equipment Overview
The circuitry on the non-isolated section processes, analyses and interprets digitized
ECG data. The non-isolated section interfaces with the keyboard, display and
thermal printer.
032
2-6
MAC 400
2036039-001D
Equipment Overview
Power inlet
The power inlet includes a three-pin IEC plug and fuse holder for two 2-pole
5x20mm fuse links. This system inlet has a snap fitting.
100-240V AC
Frequency
50-60Hz
1.67A
Output voltage
12V
Line regulation
12V +/-0.5%
Load regulation
12V +/-2%
Battery
A customized, rechargeable Lithium-Ion battery with built-in 2S protection is
mounted to the battery pocket in the bottom cover. A dedicated battery charger is
used to charge the battery. The battery specifications are listed in Table 2-5: Battery
specifications on page 2-7.
Table 2-5: Battery specifications
Rated voltage
7.2V
Battery voltage
5.4-8.4V
The MAC 400 system can print only when the battery is connected to the device.
With a fully charged battery (8.3-8.4V), approximately 100 ECGs can be recorded in
auto mode.
NOTE
If the device cannot produce at least fifty ten-second ECG reports sequentially
on a fully charged battery, we recommend replacing the battery. For instructions
on replacing the battery, see Replacing the battery on page 3-10.
2036039-001D
MAC 400
2-7
Equipment Overview
A yellow LED on the keyboard indicates when it falls below 6.8V +/- 2%.
(non-Chinese systems)
If the voltage falls below 6.3V +/- 2%, the unit turns off automatically. (all
systems)
070
2-8
DS2
DS1
Charge in progress
ON
OFF
Charge complete
OFF
ON
OFF
OFF
MAC 400
2036039-001D
Equipment Overview
Memory
Memory on the MAC 400 system includes SRAM, PROM, and Serial EEPROM.
PROM stores the code for the CPU and serial EEPROM stores configuration data.
The beep sounds once every five seconds when the leads are not connected.
The beep sounds once every five seconds to indicate the printer door open,
battery and paper error conditions.
The beep sounds once for every QRS peak detected when the heart rate is
within the specified limits.
NOTE
Pressing the QRS beep key disables the audible tone. Turning off the QRS beep
in configuration settings also disables the audible tone for QRS complex.
Keyboard
The keyboard contains the following control keys: up cursor, down cursor, right
cursor, left cursor, enter, config, lead selection, copy, QRS beep, print start/stop.
The on/off key is not part of the matrix.
There are two LEDs on the keyboard one green and one yellow.
2036039-001D
The green LED indicates that the system is connected to active mains.
(all systems)
On non-Chinese systems, the yellow LED indicates that the battery voltage is
below 6.8V +/- 2%.
MAC 400
2-9
Equipment Overview
Display
The alphanumeric interface displays on a 16X2 character liquid-crystal display
powered by +5V supply. The character format is 5X7 dots plus the cursor, and the
viewing angle is six oclock. The display type is a reflective green STN.
FPGA
The CPU-thermal printer interface, CPU-display enable logic, CPU-motor driver IC
interface are all implemented in FPGA. The CLKOUT signal is the clock signal for
logic implemented and has an output frequency of 16.7MHz.
2-10
Probe ID circuit
Instrumentation amplifiers
Multiplexer
ADC
Voltage reference
DSP
MAC 400
2036039-001D
Equipment Overview
Serial EEPROM
Clocks
Opto-couplers
Transformer
+5V_ISO
+3.3V_ISO
+1.8V
A step-up transformer is used to transfer power from the non-isolated side to the
isolated side. The voltages generated on the acquisition side are +5V_ISO,
+3.3V_ISO and +1.8V. Communication between DSP and CPU follows SPI
protocol by transmitting signals optically using Opto-couplers.
Connector J1 connects the ECG leads to the MAC 400 system. The system supports
10 leadwire ECG cables. Diodes connected in series offer electrostatic discharge
protection. A probe ID circuit detects whether the cable is connected. A lead-fail
bias circuit detects when a particular lead is not connected. The right leg drive
circuit generates the reference signal for all the leads. Instrumentation amplifiers
provide the required gain. An 8:1 multiplexer selects the eight channels one by one.
The ADC samples each sample and sends the digitized data to CPU. The signal
chain in DSP includes notch filter, DC filter, low pass filter, down sampling and line
frequency filter. Serial EEPROM is used to store DSP code. A 20MHz external
crystal provides the reference clock signal to DSP.
Internal interfaces
This section describes the interface signals between the PCB assembly and external
modules.
2036039-001D
Connector
Signal Name
I/O
J10-1
GND
J10-2
GND
J10-3
+12V
J10-4
+12V
MAC 400
2-11
Equipment Overview
Signal Name
I/O
J11-1
VBAT +
J11-2
VBAT +
J11-3
VBAT -
J11-4
VBAT -
2-12
Connector Pin
Function
Description
J8-1
GND
0V
J8-2
+5V
VCC of display
5V
J8-3
CONTRAST
0-2V
J8-4
ADR [0]
TTL/active high
J8-5
R/W*
TTL/active high
J8-6
DISPLAY
3.3V/active high
J8-7
DATA [8]
J8-8
DATA [9]
J8-9
DATA [10]
J8-10
DATA [11]
MAC 400
Level/Polarity
2036039-001D
Equipment Overview
Function
Description
Level/Polarity
J8-11
DATA [12]
J8-12
DATA [13]
J8-13
DATA [14]
J8-14
DATA [15]
J8-15
Unused
J8-16
Unused
Function
I/O
Description
Level
J7-1
PHK
0-3.8V
J7-2
VSEN
5V
J7-3
VHE
800mV-5V
J7-4
SW
J7-5
SW
J7-6
VH
6-8.4V
J7-7
VH
6-8.4V
J7-8
DI
TTL
J7-9
CLK
TTL, L->H
J7-10
GND
Ground
0V
J7-11
GND
Ground
0V
J7-12
STB 5
2036039-001D
MAC 400
Strobe signal
TTL/active high
0V
2-13
Equipment Overview
Function
I/O
Description
Level
J7-13
STB 4
Strobe signal
J7-14
STB 3
Strobe signal
J7-15
VDD
5V
J7-16
TM
Thermistor
J7-17
STB 2
Strobe signal
J7-18
STB 1
Strobe signal
J7-19
AEO2
J7-20
AEO1
J7-21
GND
Ground
0V
J7-22
GND
Ground
0V
J7-23
LAT
Latch signal
J7-24
DO
TTL
J7-25
VH
6-8.4V
J7-26
VH
6-8.4V
J7-27
MT A
6-8.4V
J7-28
MT A*
6-8.4V
J7-29
MT B
6-8.4V
J7-30
MT B*
6-8.4V
2-14
Connector Pin
Function
J9-1
GND
J9-2
Not connected
J9-3
Not connected
J9-4
GND
I/O
MAC 400
Description
Level
Ground connection
0V
Ground connection
0V
2036039-001D
Equipment Overview
Function
I/O
Description
J9-5
On/Off switch
J9-6
ROW1
Row 1 connection
J9-7
ROW2
Row 2 connection
J9-8
ROW3
Row 3 connection
J9-9
ROW5
Row 5 connection
J9-10
COLUMN5
Column 5 connection
J9-11
COLUMN1
Column 1 connection
J9-12
COLUMN2
Column 2 connection
J9-13
ROW4
Row 4 connection
J9-14
COLUMN3
Column 3 connection
J9-15
COLUMN4
Column 4 connection
J9-16
GND
J9-17
J9-18
Battery indication
LED
Non-Chinese systems:
Connected to the yellow LED which
lights up when the battery voltage
goes below 6.8V +/- 2%
Ground connection
Level
0V
Chinese systems:
Connected to yellow LED which will
illuminate continuously when battery
is fully charged or not charging or it
will flash when battery voltage falls
below 6.8V +/- 2%.
2036039-001D
MAC 400
2-15
Equipment Overview
Function
J1-1
V2
J1-2
V3
J1-3
V4
J1-4
V5
J1-5
V6
J1-6
Cable shield
J1-7
Cable shield
J1-8
GND
J1-9
RA
J1-10
LA
J1-11
LL
J1-12
V1
J1-13
Not connected
J1-14
RL
2-16
1.
Press the
configuration key to access the setup menu. The first menu
item is Language.
2.
Use the
3.
Use the
MAC 400
2036039-001D
Equipment Overview
4.
Press
5.
Press the
Setting options
Operating parameters can be customized using the setup menu. To change or set
options, follow the steps for each menu item. Following are several of the most
common settings. For more information about setting operating parameters, see the
MAC 400 Resting ECG Analysis System Operators Guide.
To access the options menu and initiate any of the following setting selections, press
the
configuration key.
Selecting a language
You can select the language for the display text and printed ECG reports.
1.
Use the
2.
Use the
3.
Press
NOTE
When you choose Chinese or Russian, changes only affect printed reports. The
display remains in English.
Use the
2.
Use the
3.
Press
2036039-001D
Use the
MAC 400
2-17
Equipment Overview
2.
Use the
3.
Press
Use the
2.
Use the
3.
Press
Use the
2.
Use the
right/left cursor to change the low limit value in increments of 5
BPM, between 30 and 120 BPM.
3.
Press
4.
Use the
240 BPM).
5.
Press
NOTE
After setting these options, you must press
Printing settings
It is recommended that you print the current settings of the system before
performing any maintenance.
2-18
1.
Press
2.
Use the
3.
MAC 400
2036039-001D
Equipment Overview
Restoring options
When the main PCB assembly is replaced, you need to restore the original options.
2.
3.
Use the
entire 13 digits, the display briefly goes blank and then displays Ser. No. Stored.
2.
3.
Use the
left/right cursor keys to enter the 12-digit numeric option code.
After you enter the entire code, the display briefly goes blank and then displays
Option Enabled.
To confirm that you have restored options, print the configuration list as described in
Printing settings on page 2-18.
Diagnostic tests
All messages displayed during self tests are in English. The system menu guides you
through the functions and testing.
2036039-001D
ROM test
MAC 400
2-19
Equipment Overview
RAM test
DSP test
If any of these tests fail, the corresponding error code displays the message: Test
failed...4 ROM (2 RAM, 5 DSP) error.
Device tests
The device test function tests various modules of the system and is conducted
through the keyboard interface. To initiate device tests, press and hold the right
cursor and configuration key together until Device test displays. Press the right/
left cursor keys to display the previous/next test.
1.
2.
Display test
3.
Motor test
4.
Recording test
5.
Results test
Key test
Device test KEY TEST displays. Press the enter key to initiate the key test. During
the test mode, when each key is pressed the corresponding name of that key
displays. The names are identified in the following table.
2-20
Display message
Enter
KEY_ENTER
Up cursor
CUR_UP
Down cursor
CUR_DOWN
Right cursor
CUR_RIGHT
Left cursor
CUR_LEFT
Lead selection
KEY_LEAD
Copy
KEY_COPY
QRS Beep
Config
KEY_CONFIG
Start/Stop
Exits key test and displays Device test and moves to next
test
MAC 400
2036039-001D
Equipment Overview
Display test
Device test DISPLAY TEST displays. To test the display:
1.
2.
Press the right cursor key. Two rows of 16 black squares display.
3.
4.
Press the enter key to exit and move to the next test.
Motor test
Device test MOTOR TEST displays. To test the speed of the motor:
1.
2.
Press the up/down cursor keys to select the speed: 5, 25, or 50 mm/s.
3.
Press the start/stop key to begin printing short lines on the edges of the paper.
The distance between consecutive short lines indicates the selected speed.
4.
5.
Recording test
Device test RECORDING TEST displays. To test the recording:
1.
2.
Press the up/down and right/left cursor keys to select the speed 5, 25, or 50
mm/s and the gain 5, 10, or 20mm/mV.
3.
Press the enter key to print the two channel wave form.
4.
5.
Results test
Device test RESULTS displays. Press the start/stop key to print one page results.
The printed output supplies data on the software contained in the unit (reference
number, version number, and date of preparation of the firmware), the results of the
tests carried out during power-up, and details on the connected patient lead.
To exit the device diagnostic tests, press the config key and the left cursor together.
2036039-001D
MAC 400
2-21
Equipment Overview
2-22
MAC 400
2036039-001D
2036039-001D
Maintenance
MAC 400
3-1
Maintenance
Recommended maintenance
A regular equipment maintenance program helps prevent unnecessary equipment
and power failures and reduces possible health hazards. This chapter contains
instructions for the following recommended maintenance:
Electrical safety
Leakage tests to verify the equipment does not propose any health hazard.
WARNING
PROPER MAINTENANCE REQUIRED Failure to
implement the recommended maintenance schedule may cause
equipment failure and possible health hazards.
All responsible individuals employing the use of this device
assume the responsibility for performing the recommended
maintenance unless an Equipment Maintenance Agreement exists.
Multimeter
Check the device and accessories for mechanical defects, which can impair
their functioning.
Ensure that the labels and inscriptions on the device relating to safety are clearly
legible.
Check that the patient cable and lead wires are not defective or broken.
Inspect the keys and keyboard to make sure they have correct markings on them
and to ensure proper operation.
Visual inspection
3-2
MAC 400
2036039-001D
Maintenance
Confirm the Green LED is glowing when the power cord is connected to the AC
line and the AC switch is On.
Exterior cleaning
Clean the exterior surfaces once per month, and more frequently if needed.
Use a clean, soft cloth and water with a mild detergent. Wring excess water
from the cloth and wipe down the exterior surfaces of the device.
Interior cleaning
Check for dust on the interior components and clean as needed with commercially
available compressed air.
Clean the thermal printhead periodically to ensure good print quality and long
printer life. To clean the thermal printhead:
1.
2.
Open the printer door and separate the platen from the printhead.
3.
Wipe the glazed surface of the thermal printhead gently with cotton swabs and
ethyl alcohol.
4.
When the surfaces are dry, fix the platen in place and close the printer door.
NOTE
Do not hit the thermal printhead with anything hard.
Do not use sand paper on the thermal printhead.
Do not add any unnecessary force on the thermal printhead.
2036039-001D
MAC 400
3-3
Maintenance
071
3-4
This test measures the ground resistance from the device power inlet ground
terminal to any exposed conductive part of the device.
Per IEC 60601-1, test the devices power inlet ground terminal. Any resistance
greater than 100 mOhm results in failure.
MAC 400
2036039-001D
Maintenance
033
078
This test measures leakage current of the devices internal power supply. The device
has to be turned on and off, and connected to your safety-testing equipment.
Measurements should be taken under the following conditions (refer to the safety
tester diagram):
2036039-001D
Device in ON state
MAC 400
3-5
Maintenance
The test has failed if the measured values are greater than:
N.C.
S.F.C.
S1 (neutral): closed
S1 (neutral): open
GND: open
GND: open
Total tests/combinations: 2
Total tests/combinations: 2
500 uA (IEC)
300 uA (UL)
034
This test measures leakage current from the chassis to ground during normal
condition (N.C.) and single fault conditions (S.F.C.). In all cases, the leakage current
is measured from any exposed conductive parts to ground; the device has to be
turned on and off, and connected to your safety-testing instrument.
During N.C., measurements should be taken under the following conditions (refer to
the safety tester diagram):
Device in ON state
S1 (neutral) closed
During S.F.C., measurements should be taken under the following conditions (refer
to the following safety tester diagram):
Device in ON state
The test has failed if the measured values are greater than:
3-6
MAC 400
2036039-001D
Maintenance
N.C.
S.F.C.
S1 (neutral): closed
GND: closed
Total tests/combinations: 2
Total tests/combinations: 2
500 uA (IEC)
300 uA (UL)
077
This test measures leakage current from the floating input (patient connection) to
ground during normal condition (N.C.) and single fault condition (S.F.C.). In all
cases, the leakage current is measured from the floating input to ground: The device
has to be turned on and off, and connected to your safety testing instrument.
During N.C., measurements should be taken under the following conditions (refer to
the safety tester diagram):
Device in ON state
S1 (neutral) closed
During S.F.C., measurements should be taken under the following conditions (refer
to the safety tester diagram):
2036039-001D
Device in ON state
MAC 400
3-7
Maintenance
The test has failed if the measured values are greater than:
N.C.
S.F.C.
S1 (neutral): closed
GND: closed
Total tests/combinations: 2
Total tests/combinations: 4
035
WARNING
PERSONAL SAFETY MAINS VOLTAGE is applied to
floating input (patient connection) and patient cables/test plug
conductors and safety tester PATIN JACK.
For the protection of the person performing these tests, the
following values of resistor R may be used:
This test performs leakage current tests during single fault conditions (S.F.C.) with
line voltage applied to the floating (patient connection) inputs.
3-8
MAC 400
2036039-001D
Maintenance
In all cases, the leakage current is measured from the floating (patient) inputs jack of
the device to ground. To setup this leakage current test:
1.
2.
3.
4.
5.
Ensure that the person/tester in not touching the patient cable/test plug or
floating (patient) input or device conductive surfaces
6.
During S.F.C, measurements should be taken under the following conditions (refer
to safety tester diagram):
Device in ON state
S1 (neutral) open
The test has failed if the measured values are greater than:
N.C.
N/A
S.F.C.
Polarity: NORM & RVS
S1 (neutral): open
GND: closed
Total tests/combinations: 0
Total tests/combinations: 2
N/A
After completing this test, disconnect the safety tester from line voltage before
disconnecting the patient cable/test plug and/or the device.
Replacing components
When it is determined that a component is faulty, refer to the following instructions
for replacement. Follow the safety instructions when replacing components.
WARNING
OPERATOR SAFETY Repairing device when powered on
may cause injury.
Turn off the device before replacing components.
2036039-001D
MAC 400
3-9
Maintenance
Safety instructions
When repairing the MAC 400 device, the following considerations apply:
Before opening the device, turn off the power and disconnect the mains plug
from the power inlet module. Never connect the mains plug when the device is
open.
Before opening the device or completing any servicing, remove the battery.
Before replacing the primary fuses in the power inlet module, turn off the power
and disconnect the mains plug.
Only the original GE Healthcare components identified in the FRU list may be
used.
036
3-10
MAC 400
2036039-001D
Maintenance
1.
Remove the screw fastening the battery cover to the bottom cover.
2.
Slide and raise the battery cover and remove the battery from its compartment.
3.
Place the new battery in to the battery compartment with the battery label facing
downwards, and the battery terminals facing the battery contacts on the PCB.
Reverse voltage protection is implemented mechanically.
037
1.
Depress the latch and slide the fuse holder from the power inlet module.
2.
3.
4.
2036039-001D
MAC 400
3-11
Maintenance
038
1.
2.
Turn the device upside down and remove the four fastening screws on the
bottom.
3.
With the keyboard facing up, open the printer door along with the platen.
4.
Raise the top cover of the housing and disconnect the keyboard cable from the
J9 connector.
The display remains fixed on the PCB.
NOTE
If you need to replace the top cover assembly, dispose of the faulty (old)
one and reassemble with the new top cover assembly.
3-12
MAC 400
2036039-001D
Maintenance
039
1.
Open the device and printer door as described in Opening the device on
page 3-12.
2.
3.
4.
Remove the two screws fastening the base of the printer to the printer bracket.
5.
Align the mounting holes of the new printer with those on the printer bracket.
6.
Use the screws to fasten the printer onto the printer enclosure.
7.
Connect the flat ribbon cable to the printer connector on the PCB.
8.
Align the platen on the printer door and press it into position.
NOTE
Ensure that the platen gear aligns with the gears of the thermal printer.
2036039-001D
MAC 400
3-13
Maintenance
040
3-14
1.
2.
Disconnect the display cable connector from the J8 connector on the main PCB.
3.
Remove the four screws fastening the display to the display mounting bracket.
4.
Align the new display mounting holes to those on the display mounting bracket
and fasten the display with the screws.
5.
Connect the display cable connector to the J8 connector on the main PCB.
MAC 400
2036039-001D
Maintenance
041
NOTE
If possible, print out the current settings before replacing the PCB assembly. For
information about the configuration settings, see Printing settings on page 218.
1.
2.
3.
4.
Remove the two screws fastening the display bracket assembly and remove it.
5.
6.
Remove the four screws fastening the PCB assembly to the aluminum plate.
NOTE
Return the PCB assembly to the factory in ESD packaging.
2036039-001D
7.
Orient the new main PCB assembly, align the mounting holes with those in the
aluminum plate, and fasten the new PCB assembly to the plate with the screws.
8.
Re-fasten the display and display bracket to the PCB and plate with the screws.
MAC 400
3-15
Maintenance
9.
Re-connect the printer cable to J7, the power supply cable to J10, the keyboard
cable to J9 and the display cable to J8.
10. Re-attach the upper section of the device to the lower section with the four
screws.
11. Replace the battery in the battery pocket, slide on the cover and re-fasten with
the screw.
12. Turn on the device and enter the settings as described in Setting options on
page 2-17.
13. Enter the serial number and option code (optional) as described in Restoring
options on page 2-19.
14. Adjust the display contrast as described in Adjustment instructions on page 320.
15. Adjust the time and date.
16. If possible, set the configuration to the previous user settings. If not, select the
default settings.
042
1.
Remove the PCB assembly as instructed in steps 1-6 of Replacing the PCB
assembly on page 3-15.
2.
With the display connector at the top and facing upwards, locate the RTC
battery BT1 on the left side of the PCB.
3.
Turn the PCB assembly over and de-solder the battery contacts to remove the
battery.
NOTE
Dispose of battery according to local municipal waste guidelines.
3-16
MAC 400
2036039-001D
Maintenance
4.
5.
6.
7.
Turn on the device and when you see the main menu, press the configuration
key.
8.
075
2036039-001D
1.
Remove the PCB assembly as instructed in steps 1-6 of Replacing the PCB
assembly on page 3-15.
2.
With the display connector at the top and facing upwards, locate the OTPROM
in the upper center of the PCB.
3.
4.
MAC 400
3-17
Maintenance
0432
1.
Remove the main PCB as instructed in steps 1-6 of Replacing the PCB
assembly on page 3-15.
2.
Remove the five screws fastening the aluminum plate to the bottom cover.
3.
4.
Remove the screw fastening the ground wire to the printer bracket.
5.
Turn over the aluminum plate and disconnect the yellow/green wire from the
yellow/green wire of the toroid.
6.
Disconnect the white and black wires from the power inlet module.
7.
Remove the four screws to unfasten the power supply from the mounting plate.
8.
Align the new power supply mounting holes to the power supply mounting
plate and replace the four screws.
9.
Connect the yellow/green wire of the power supply to the yellow/green wire of
the toroid.
10. Connect the black wire of the power supply to the terminal marked N of the
power inlet module, and connect the white wire to the terminal marked L.
11. Complete steps 11-15 ofReplacing the power inlet module along with the
toroid assembly on page 3-19.
3-18
MAC 400
2036039-001D
Maintenance
Replacing the power inlet module along with the toroid assembly
044
1.
Remove the main PCB assembly as instructed in steps 1-6 of Replacing the
PCB assembly on page 3-15.
2.
Remove the power supply shield as described in steps 1-6 of Replacing the
12V power supply on page 3-18.
3.
4.
To remove the power inlet from the mounting bracket, depress the locking tabs
on the top and bottom and slide the inlet out.
5.
Insert the new power inlet into the bracket and snap into place.
6.
b.
Place the new toroid into position on the shield, remove the plastic from the
adhesive-backed holder and place as shown in drawing.
NOTE
Ensure that the toroid is centered between the mounting holes.
2036039-001D
MAC 400
3-19
Maintenance
7.
Connect the yellow/green wire of the toroid to the yellow/green wire of the
power supply.
8.
Connect the yellow/green wire terminal to the power inlet terminal marked
9.
Connect the white wire to the terminal marked L and the black wire to the
terminal marked N of the power supply to the power inlet module.
NOTE
Use care to ensure that the power inlet connections are properly assembled
as indicated in drawing.
10. Connect the earthing cable of the toroid supply to the thermal printer base
mounting hole.
11. Replace the shield and power supply assembly by turning it over and aligning
the screw holes.
NOTE
Before tightening the shield screws, confirm that the power supply cable
and grounding cable are properly positioned.
12. Replace the screws fastening the power inlet bracket to the lower housing.
13. Replace the screw holding the ground cable to the printer bracket.
14. Replace the screws on the shield.
15. Replace the main PCB assembly as described in Replacing the PCB assembly
on page 3-15.
Adjustment instructions
No adjustment of components is required for the MAC 400 device.
When the PCB assembly or display is exchanged, the contrast of the display must be
adjusted to optimum contrast. To adjust the contrast:
1.
2.
3.
4.
When you set the desired contrast level, press the configuration key to exit the
setup menu.
3-20
MAC 400
2036039-001D
Maintenance
FRU Replacement
Visual Inspection
Electrical Safety
Test
FRU Description
Tools
Checkout Procedure
1 to 6 and 8
1,3,5
1, 2
Battery assembly
1, 3, 9
N/A
1, 5
1 to 6 and 8
1, 2
1 and 8 to 12
1, 3
N/A
1 to 6 and 8
1, 2
1 to 12
1, 2, 3, 8
N/A
1, 11
1. 2. 3
N/A
1, 4, 9
1, 3, 8
N/A
1, 10
1, 3, 8
N/A
1, 10
1, 3, 8
3, 4, 5
N/A
1, 8, 11
1 to 6 and 8
1, 2
1 to 12
Hardware kit
1, 3
N/A
Fuse kit
1, 3
N/A
1, 2, 3, 7, 8
N/A
1, 6
N/A
N/A
Tools
Visual Inspection
Electrical Safety
Test
Checkout Procedure
No parts replaced
3,5
N/A
N/A
N/A
2, 3, 4, 5, 6, 8
2, 3
1, 2
1, 11
2, 3, 4, 5, 6, 8
1, 2
1, 11
Tools needed
2036039-001D
1.
Phillips screwdriver
2.
3.
MAC 400
3-21
Maintenance
4.
5.
6.
Multi-meter
7.
Soldering iron
8.
ECG simulator
9.
MAC 400 Resting ECG Analysis System Operators Guides for reference
Visual inspection
Verify the following items have passed inspection.
Operational checks
3-22
1.
Verify the power-up self-test passed. See Power on self test on page 2-19.
2.
Verify the serial number restored successfully. See Restoring the serial number
on page 2-19.
3.
For units with the interpretation option enabled, verify the interpretation option
restored successfully. See Restoring the interpretation option on page 2-19.
4.
5.
Verify that the new battery was charged successfully. See the MAC 400 Resting
ECG Analysis System Operators Guides.
6.
Verify the time and date were set successfully. See Setting the date and time for
printed ECG reports on page 2-17.
7.
Verify that the last ECG was copied successfully. See the section Automatic Mode
in the MAC 400 Resting ECG Analysis System Operators Guides.
MAC 400
2036039-001D
Maintenance
Diagnostic tests
8.
Verify the keyboard test was successful. See Key test on page 2-20 for more
information.
9.
Verify the display test was successful. See Display test on page 2-21 for more
information.
10.
Verify the printer motor test was successful. See Motor test on page 2-21 for
more information.
11.
Verify the recording test was successful. See Recording test on page 2-21 for
more information.
12.
Verify the results test was successful. See Results test on page 2-21 for more
information.
Verify the equipment does not propose any health hazard. See Earth leakage
(AC line) current test on page 3-5,Patient leakage current tests on page 3-7,
and Enclosure (chassis) leakage current test on page 3-6 for more
information.
2.
Verify all exposed metal is properly grounded. See Ground resistance test on
page 3-4 for more information.
Condition1
UUT - ON2
Result
Forward Polarity
NC
______
Pass/Fail
300
SFC
______
Pass/Fail
1,000
SFC
______
Pass/Fail
1,000
Reverse Polarity
NC
______
Pass/Fail
300
Forward Polarity
NC
______
Pass/Fail
100
SFC
______
Pass/Fail
300
SFC
______
Pass/Fail
300
2036039-001D
MAC 400
3-23
Maintenance
Condition1
UUT - ON2
Result
SFC
______
Pass/Fail
300
SFC
______
Pass/Fail
300
Reverse Polarity
NC
______
Pass/Fail
100
Forward Polarity
NC
______
Pass/Fail
10
SFC
______
Pass/Fail
50
SFC
______
Pass/Fail
50
SFC
______
Pass/Fail
50
SFC
______
Pass/Fail
50
Reverse Polarity
NC
______
Pass/Fail
10
Ground Resistance
1
3-24
Resistance
N/A
______
MAC 400
Pass/Fail
2036039-001D
2036039-001D
Troubleshooting
MAC 400
4-1
Troubleshooting
Green LED does not light up and device does not turn on
When valid mains is connected to the MAC 400 device, the green LED on the
keyboard lights up. When the device is connected to mains and it does not turn on,
follow the steps in the following flowchart to pinpoint the defective component. To
replace components, see Opening the device on page 3-12.
045
4-2
MAC 400
2036039-001D
Troubleshooting
046
2036039-001D
MAC 400
4-3
Troubleshooting
047
4-4
MAC 400
2036039-001D
Troubleshooting
Blank display
If the device is on and the audible tone (beep) sounds but the display is blank, follow
the steps in the following flowchart to pinpoint the faulty component.
048
2036039-001D
MAC 400
4-5
Troubleshooting
Battery error
The MAC 400 device can only print when it is powered by the battery. When
Battery Error displays, follow the steps identified in the following flowchart.
049
4-6
MAC 400
2036039-001D
Troubleshooting
050
2036039-001D
MAC 400
4-7
Troubleshooting
Paper problem
When a problem occurs with the thermal paper, Paper Problem displays. Follow the
steps outlined in the following flowchart to diagnose the problem.
051
4-8
MAC 400
2036039-001D
Troubleshooting
052
2036039-001D
MAC 400
4-9
Troubleshooting
053
4-10
MAC 400
2036039-001D
Troubleshooting
054
2036039-001D
MAC 400
4-11
Troubleshooting
4-12
MAC 400
2036039-001D
2036039-001D
MAC 400
5-1
Ordering parts
General information
This chapter identifies the FRU parts and FRU kits with part numbers and images.
You can order parts for the assemblies, stand-alone FRUs, and FRU kits that are
considered field-serviceable. Only items, assemblies, and kits which have part
numbers provided in this chapter are available as FRUs. To order parts, contact GE
Service or a GE approved vendor.
Accessory kits
Table 5-1. Accessory kits
Part number
Description
2035819-001
IEC starter kit (includes one ECG cable with lead wires,
four clamp electrodes, six bulb electrodes, one pack of zfold paper & gel).
2035819-002
AHA starter kit (includes one ECG cable with lead wires,
four clamp electrodes, six bulb electrodes, one pack of zfold paper & gel).
Patient Cables
Table 5-2. Patient Cables
Part number
Description
2029890-001
2029893-001
Electrodes
Table 5-3. Electrodes
Part number
5-2
Description
2029891-001
2029894-001
2029892-001
MAC 400
2036039-001D
Consumables
Table 5-4. Consumables
Part number
217 083 18
Description
Electrode cream, 250-ml refill bottle
Writer paper
Table 5-5. Writer paper
Part number
Description
2030887-001
2030888-001
Service manual
Table 5-6. Service manual
2036039-001D
Part number
Description
Language
2036039-001
English
2036039-002
English
MAC 400
5-3
Operators guides
Table 5-7. Operators guides
5-4
Part Number
Description
Language
2032589-001
English (ENG)
2032589-002
German (GER)
2032589-003
French (FRE)
2032589-004
Dutch (DUT)
2032589-005
Italian (ITA)
2032589-006
Spanish (SPA)
2032589-007
2032589-008
Russian (RUS)
2032589-009
Hungarian (HUN)
2032589-010
Polish (POL)
2032589-011
Czech (CZE)
2032589-012
Chinese (CHS)
2032589-013
Slovakian (SLO)
2032589-014
Greek (GRE)
2032589-015
Korean (KOR)
2032589-016
Bulgarian (BUL)
2032589-017
Turkish (TUR)
2032589-018
Lithuanian (LIT)
2032589-019
Latvian (LAT)
2032589-020
Estonian (EST)
2032589-023
2032589-024
MAC 400
2036039-001D
Power cords
Table 5-4: Power cords
Part Number
Description
Country
405535-006
North America/
Central America/
Japan
401855-018
China
401855-101
Europe
401855-102
Britain
401855-103
Italy
401855-104
Israel
401855-107
Switzerland
2035953-001
India
401855-109
Denmark
40185-110
Australia
2036039-001D
Part Number
Description
2037511-001
MAC 400
5-5
055
Item
5-6
Item description
Fuse holder
Toroid assembly
MAC 400
2036039-001D
056
057
2036039-001D
MAC 400
5-7
058
Item
5-8
Item description
MAC 400
2036039-001D
059
_____
2036039-001D
MAC 400
5-9
060
Item
Item description
060
Item
Item description
061
5-10
MAC 400
2036039-001D
062
063
Item
2036039-001D
Item description
MAC 400
5-11
064
065
Item
5-12
Item description
Bottom cover
Battery cover
MAC 400
2036039-001D
066
Item
Item description
067
2036039-001D
MAC 400
5-13
068
069
5-14
MAC 400
2036039-001D
2036039-001D
Technical Specifications
MAC 400
A-1
Technical Specifications
Instrument type
Microprocessor augmented automatic electrocardiograph: 10leadwire acquisition with programmable lead configuration.
Processing
Input impedance
> 10M
@ 10 Hz,
defibrillator protected
Acquisition and analysis
12 leadssimultaneous
Patient leakage
Signal input
Operating modes
Digital sampling
>1000 samples/second/channel
500 samples/second/channel
Report formats
Dynamic range
30 to 300 BPM 2% or 1
BPM, whichever is greater.
Heart rates outside this range
do not display.
Resolution
Baseline adjustment
Frequency response
0.01 to 150 Hz
Automatic compensation of
baseline
Lead-fail indication
A-2
MAC 400
< 10
2036039-001D
Technical Specifications
Software options
Paper size
Width 80 mm
Typical +/- 1
Maximum +/- 5%
Type
Writer amplitude
accuracy
Displayed data
Keyboard
Writer
Type
+/- 5%
Writer technology
Power supply
Resolution
Vertical 8 dots/mm
Ratings
Frequency
50/60 Hz +/-10%
Battery type
Writer speeds
5, 25, 50 mm/s
5mm/s is not available in auto mode
Writer sensitivity
(gain)
5, 10, 20 mm/mV
Approximately 3 hours
Paper type
Z-fold, roll
Battery charging
Battery capacity
Physical specifications
Relative humidity
Transport/storage
25-95%
25-95%
Width
263 mm
Operating
Height
78 mm
Pressure
Depth
208 mm
Transport/storage
2036039-001D
MAC 400
A-3
Technical Specifications
Weight
Operating
Environmental specifications
Temperature (with battery)
Transport/storage
-15 C to + 50 C
Certifications:
Operating
+10 C to 40 C
China-specific certifications
a) SFDA Certificate
b) CCC Certificate
c) CMC Certificate
Certifications for all other countries
a) Complies to UL 60601-1 and CAN/CSA C22.2 No. 601.1
c) CB certified for IEC 60601-1, IEC 60601-2-25, IEC 60601-2-51
d) CE marking for Council l DIrective 93/42/EEC concerning
medical devices
e) Meets applicable AAMI EC-11 requirements
Warranty
Standard warranty is 1 (one) year.
Ordering information
Visit www.gehealthcare.com or contact your local GE representative
A-4
MAC 400
2036039-001D
2036039-001D
Electromagnetic
compatibility
(not applicable
for China)
MAC 400
B-1
CAUTION
EQUIPMENT INTERFERENCE If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in that configuration.
The equipment or system should not be used adjacent to, or stacked
with, other equipment.
Compliance
Group 1
Class A
EN 55011
RF emissions
EN 55011
Harmonic Emissions
Electromagnetic environment-guidance
Class A
EN 61000-3-2
Voltage fluctuations/Flicker emissions
Complies
EN 61000-3-3
B-2
MAC 400
2036039-001D
Compliance level
Electromagnetic environment-guidance
6 kV contact
6 kV contact
8 kV air
8 kV air
EN 61000-4-4
1 kV for input/output
lines
Surge
1 kV differential mode
1 kV differential mode
EN 61000-4-5
2 kV common mode
2 kV common mode
3 A/m
3 A/m
Electrostatic discharge
(ESD)
EN 61000-4-2
EN 61000-4-11
Power frequency(50/60
Hz) magnetic field
EN 61000-4-8
2036039-001D
MAC 400
B-3
Immunity test
Compliance level
Electromagnetic environment-guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the MAC 400, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
EN 61000-4-6
d = 1.2
d = 1.2
Radiated RF
3V/m
EN 61000-4-3
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the equipment.
B-4
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
MAC 400
2036039-001D
150KHz to 80MHz
outside ISM bands
d = 1.2
150KHz to 80MHz
outside ISM bands
d = 1.2
80MHz to 800MHz
d = 1.2
800MHz to 2.5GHz
d = 2.3
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
At 80MHz and 800MHz the separation distance of high frequency range applies.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (w) according to the transmitter
manufacturer.
2036039-001D
MAC 400
B-5
B-6
MAC 400
2036039-001D
Index
Index
C
CE marking information 9
E
equipment identification 5
equipment service information 5
equipment symbols 5
F
FRUs
bottom cover kit 12
clock battery 14
complete printer module 11
fuse kit 13
hardware kit 13
LCD assembly with cable and connector 10
lithium-ion battery 7
main PCB assembly 9
paper roll axle 14
power bracket kit 8
power inlet module assembly 6
power supply AC/DC assembly 7
printer door assembly 11
software OTP ROM 10
top cover assembly with keyboard 12
functionality, description 5
M
maintenance
monthly inspections 2
mechanical structure 5
P
parts
ordering 2
PCB assembly 8
power supply 7
R
replacing components 9
2036039-001D
MAC 400
Index-1
Index
Index-2
MAC 400
2036039-001D
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel: + 1 414 355 5000
www.gehealthcare.com
GE Medical Systems
Information Technologies GmbH
Munzinger Strae 3-5
D-79111 Freiburg
Germany
Tel: + 49 761 45 43 - 0
Fax: + 49 761 45 43 - 233
Asia Headquarters
GE Medical Systems
Information Technologies Asia; GE (China) Co., Ltd.
11th Floor, Shanghai MAXDO Center,
8 Xing Yi Road, Hong Qiao Development Zone
Shanghai 200336, Peoples Republic of China
Tel: + 86 21 5257 4650
Fax: + 86 21 5208 2008