Documente Academic
Documente Profesional
Documente Cultură
1 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
THE GALVANOMETER
Digital instruments were not readily available until the 1970's and cathodic protection
measurements were carried out with a variety of electrical meters which were based on the
galvanometer.
These instruments work on the reaction of a magnet in an electric field against the elastic effect of a
hair spring.
They are less rugged than digital instruments and more difficult to manufacture. They also require a
considerable amount of current to move the indicator needle.
25/03/2016 11:40
Module 1
2 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
The amount of current drawn by this type of instrument is sufficient to cause a depression in the
potential of the soil in which the electrode is placed.
It can be said that the current flowing through the measuring circuit causes its own IR drop.
This potential depression can be measured using two electrodes and a digital voltmeter.
This 'volts drop' in the measuring circuit caused substantial inaccuracies in the measurement, and
the readings were shown to be lower than the true voltage. This is the opposite effect to that of the
'IR drop in the soil' which is caused by the passage of the cathodic protection current.
During the period in which these meters were used, the fact that the readings were less than the
criteria, encouraged operators to blame instrument inaccuracy when protected criteria could not be
achieved.
25/03/2016 11:40
Module 1
3 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
All the time that a needle was moved against a hair spring, at any time in the measuring process,
this required sufficient current to be affected by any resistance in the measuring circuit. Instruments
which were extremely sensitive had to have extremely fragile mechanisms which were unsuitable
for rough field use.
25/03/2016 11:40
Module 1
4 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
The error in the measuring system tended to give the impression that sections of pipeline were not
protected and engineers worked to increase the protection on these sections. They found that many
sections of pipeline could not be 'protected' as it proved impossible to increase the output of the
cathodic protection system enough to show the required voltage on the meter.
A formula was recommended to correct readings to an increased value, in proportion to the internal
resistance of the voltmeter. This formula corrected the readings upwards towards the required
criteria, in contrast to the present correction that which is obtained by switching the CP current off
and is a correction downwards.
NUL-BALANCE METERS
Attempts were made to design specialised instruments which balanced out this error with an
opposition EMF supplied from an internal battery. The idea was that if they worked on a balancing
process when the actual measurement was made, then they would draw no current. Some of these
instruments became quite complex but they all drew current during the original balancing
operation, and hence none reached the accuracy which is possible with a digital meter.
The designers seemed to have missed the point that the current drawn, to set up the balance, created
an error which was carried through the whole measuring operation.
The complexity of operation of the balancing meters required more care and understanding by the
operators, and the instruments were not as rugged as the simpler ones of the day.
Instruments appeared with more and more dials, switches and adjustments but this resulted in less
and less personnel who understood how to use them. It also had the adverse effect that the amount
of time to take a single reading with such an instrument, precluded the possibility of taking
'immediate off potentials'.
There was then the problem of measuring the EMF which had balanced the measuring circuit. This
had to come from a dry cell battery which would not have a constant potential as the current drawn,
to balance the circuit, would drain the battery each time of use.
Some instruments were then marketed with a more constant (laboratory tested) battery built in,
with which to balance the, commercially available, dry cell.
The problem that still remained, was to establish a potential value, within the meter, against which
a reliable voltage could be measured.
Sophisticated variable resistors were built into some instruments and it was claimed that an error
free voltage could be measured by a complex procedure of balancing potentials using two
galvanometers, two sources of DC charge and a variety of switches, variable resistances and a
component called a potentiometer.
On examination, the 'potentiometer' turned out to be a variable resistor which had been calibrated
to give a reading on a mechanically driven digital dial.
DIGITAL METERS
25/03/2016 11:40
Module 1
5 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
The availability of digital instruments removed the downward error which had been caused by the
volts drop in the measuring circuit and allowed increased voltages to be obtained, because they
dramatically reduced the amount of current flowing.
The solid state electronics of today's meters only require a tiny amount of current to make a reading
and the resistance in the measuring circuit is greater by an order of magnitude than the resistance in
the circuit to be measured.
Using a digital voltmeter with an impedance of 7 to 10 mega-ohms per volt ensures that the current
flowing through the measuring circuit is not sufficient to cause a potential depression in the earth in
which the electrode is placed.
25/03/2016 11:40
Module 1
6 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
When using galvanometers, the error in measuring had helped to balance out the IR drop in the soil
which, at present, causes optimistic readings.
In fact it was the advent of digital metering which bought the question of the cathodic protection
criteria sharply into focus.
Pipelines began to fail in areas where digital voltmeters had shown that the pipeline had achieved a
voltage difference of -0.850 mv with reference to an electrode placed immediately above the
pipeline on the ground surface. It is doubtful if this value would have shown on a galvanometer
which often required the ground contact to be watered to allow sufficient current to pass through
the metering circuit to activate the magnet against the hairspring.
This inaccuracy was not an advantage, but did have some tendency to balance the inherent error in
the measuring technique.
THE COPPER/COPPER-SULPHATE ELECTRODE
It is now necessary to examine the electrode, which is merely a copper rod immersed in a saturated
solution of copper sulphate.
There are many available on the market as this selection shows.
25/03/2016 11:40
Module 1
7 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
But a half-cell for cathodic protection monitoring can be made by anyone. Students are required to
made a half-cell as part of the practical work of this course. The copper/copper-sulphate electrode
is regarded as a reference electrode as, under laboratory
as a known potential against which to measure a voltage.
25/03/2016 11:40
Module 1
8 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
'Half-cell'
In cathodic protection field work, the electrode is simply part of the measuring circuit, and has no
mystical powers. Current can pass through the electrode in either direction to or from the
electrolyte without a surface reaction which is peculiar to the external electrolyte at that location.
LABORATORY USE
An example of laboratory discipline can be seen in the British Standards Institute, Code of Practice
for testing paints and coatings for their cathodic disbondment potential. This code of practice
specifies that the reference electrode is placed exactly 10mm from the metal to electrolyte
interface, at a coating fault which is deliberately made on a test piece.
The pH, temperature and pressure at the time of the test are specified, as are the period of time and
25/03/2016 11:40
Module 1
9 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
Measuring Circuit
25/03/2016 11:40
Module 1
10 of 10
http://www.rogeralexander1938.webspace.virginmedia.com/cpn/Corre...
if there are metal oxides and damp associated with these connections.
These phenomena will be caused by the micro-reactions taking place between the films of moisture
and the variety of salts and metals.
25/03/2016 11:40