Sunteți pe pagina 1din 7

Procedure No.

436-20081015-01
Acceptance

Major

Minor

Anesthesia Vaporizers
Used For
Anesthesia Unit Vaporizers [10-144]

Also Called
By trade names (e.g., Fluotec 5, Vapor 2000, Tec 6 Plua), which are registered trademarks and
should be used only when referring to the specific devices

Commonly Used In
Operating rooms, emergency departments, delivery rooms, trauma departments, ambulatory
surgical centers, and any areas requiring the administration of an inhalation agent (with
anesthesia units)

Scope
Applies to the various anesthesia vaporizers used to deliver a known concentration of vaporized
liquid anesthetic

Risk Level:
High

Type

Interval

Time Required

Major
Minor

6 Months
0 NA

0 Hours
0 Hours

Overview
An anesthesia unit vaporizer is used to vaporize a liquid anesthetic agent and deliver a controlled
amount to the patient.
According to the American Society for Testing and Materials (ASTM) standard ASTM F1850,
anesthetic agent vaporizers are required to be concentration calibrated (i.e., a calibrated knob
controls the output concentration). Older vaporizers that do not have a single control for
selecting anesthetic vapor concentration should be removed from service. there are two types of
concentration-calibrated vaporizers: variable bypass and heated blender.
Conventional (variable-bypass) vaporizers.In a variable-bypass vaporizer, the total background
gas flow that enters the unit is split into two streams. The smaller stream, which acts as the
carrier gas, passes through the vaporizing chamber containing the anesthetic agent and
becomes saturated with agent vapor; the remainder of the gas bypasses this chamber. A wick
may be used in the vaporizing chamber to provide increased surface area for efficient
evaporation of the drug and saturation of the carrier gas. The saturated carrier gas leaves the
chamber and mixes with the bypass gas. One adjustment is made to set the desired
concentration. This adjustment simultaneously balances the carrier and bypass flows to produce
the blend required for the set concentration. The mixture exits the vaporizer and is delivered
from the anesthesia machine as the fresh gas to be inspired by the patient.
Evaporation of the liquid agent contained in the chamber is driven by heat absorbed from the
walls of the vaporizer; consequently, when evaporation is occurring, the vaporizer and its
contents cool. Because the equilibrium vapor pressure of an agent changes with temperature, a
temperature-sensitive mechanism is used to automatically adjust the carrier and bypass flows to
compensate for temperature changes. Figure 1 presents a schematic of a variable-bypass
vaporizer.

Figure1. Schematic illustrating the basic elements of a variable-bypass vaporizer


Desflurane (heated-blender) vaporizers.Desflurane, a volatile inhalation anesthetic marketed by
Baxter Pharmaceutical Products Division under the trade name Suprane, has characteristics that
allow rapid induction of and emergence from anesthesia.
The boiling point of desflurane22.9C at 760 mm Hgis just above room temperature;
therefore, small increases in ambient temperature or decreases in atmospheric pressure can
cause it to boil. Also, because of desflurane's high minimum alveolar concentration (i.e., its low
potency), evaporation of sufficient agent to achieve a given anesthetic effect would require much
more heat absorption from the vaporizer than occurs with other agents. Furthermore, the
change in vapor pressure of desflurane per change in temperature is as much as three times
that for the other volatile agents at sea-level atmospheric pressure. These profound effects of
temperature and ambient pressure on the vapor pressure of desflurane make stabilizing the
delivered concentration at a set point extremely difficult in a passive mechanical system, such as
a variable-bypass vaporizer. As a result, the variable-bypass design was abandoned for
desflurane, and Datex-Ohmeda developed the Tec 6 vaporizer, based on a heated-blender
design. Figure 2 shows a schematic of this vaporizer. Datex-Ohmeda has since released the Tec 6
Plus, which is similar in design.

Figure 2. Schematic illustrating the basic elements of the Datex-Ohmeda Tec 6 vaporizer
Datascope, Drager, Medical, and Penlon have not developed an equivalent desflurane vaporizer. A
version of the Tec 6 has been adapted for Drager machines and is compatible with Drager's
triple-exclusion interlock system.

A desflurane vaporizer requires electrical power to heat the agent to a thermostatically controlled
39C, producing a stable, saturated vapor pressure of 1,500 mm Hg. No wick is used, and no
carrier gas enters the sump chamber. Instead, a stream of vapor under pressure flows out of the
sump; this stream blends with the background gas stream, which originates from the anesthesia
machine's flowmeters, to achieve the desired concentration.
The background gas stream passes through a fixed-flow resistor, producing a back pressure
upstream of this resistor that is proportional to the background gas flow. The desired desflurane
concentration is set on the dial of the adjustable metering valve in the vapor stream; this setting
produces a predetermined aperture. The pressure in the vapor upstream of the aperture and the
back pressure in the background gas stream are continually sensed by a differential pressure
transducer. The transducer controls a pressure-regulating valve in the vapor stream between the
sump and the adjustable metering valve. The pressure-regulating valve permits only that flow
from the sump necessary to cause the pressure upstream of the adjustable metering valve to
equal the back pressure in the background gas stream. In this way, the ratio of the adjustable
metering valve's resistance to the resistance of the fixed-flow resistor determines the ratio of the
flows in each stream, and therefore, the concentration of vapor in the blended output. If the flow
from the anesthesia machine's flowmeters through the vaporizer is altered, the flow of vapor
from the sump is automatically adjusted so that the pressures at the two monitored points
remain equal, the flow ratio does not change, and the output concentration continues to match
its setting.
The control circuits and heating elements in the vaporizer are turned on by the act of connecting
the vaporizer to electrical power. The unit then heats to and remains at operating temperature as
long as it receives power, whether it is delivering agent or is in the standby mode. Consequently,
it is warm to the touch while plugged into a live socket.

Test Apparatus and Supplies, Parts


Electrical safety analyzer (line powered vaporizers)
Halogenated anesthetics analyzer

Special Precautions
Do not fill a vaporizer with an inhalation agent unless you are qualified to do so. Always use a
scavenging system or appropriate ventilation when inspecting vaporizers. For personal safety,
when inspecting vaporizers alone, notify other personnel of your location. Be sure that filler ports
are tightly capped before passing gas through the vaporizer.
As a general precaution, older vaporizers containing an anesthetic agent should not be tipped. If
such tipping occurs, notify the user and follow the manufacturer's recommended procedures for
airing or drying the vaporizer.

Procedure
Be sure that you understand how to operate the equipment, the significance of each control and
indicator, and the alarm capabilities. Before beginning an inspection, carefully read this
procedure, the operators manual, and the inspection and preventive maintenance procedures
recommended by the manufacturer (typically included in the service manual). Use
the BiomedicalBenchmark Support Assessment Form to document a maintenance decision that
reflects past experience with this type of equipment and the environment where it is used. Then
use the IPM Procedure Customization Tool to modify this procedure as needed; the program will
generate a documentation form with the corresponding changes.
Note:This procedure should be performed simultaneously with Anesthesia Units Procedure 400,
where leak testing of the vaporizer has been included with the anesthesia unit.

Qualitative Tasks
Chassis/Housing ( Acceptance )
Check for shipping damage; report any damage to the manufacturer, shipper, or service
organization, and arrange for repair or replacement.
Check that the vaporizer is suitably constructed to withstand normal hospital use and abuse. For
instance, a unit with venting on the top of the housing or poorly protected or sealed controls and
indicators may be prone to fluid entry.
Examine the exterior of the vaporizer for cleanliness and general physical condition. Ensure that
plastic housings are intact, that all assembly hardware (e.g., screws, fasteners) is present and
tight.
Chassis/Housing ( Major )
Examine the exterior of the vaporizer for cleanliness and general physical condition. Be sure that
housings are intact, that all assembly hardware is present and tight, and that there are no signs
of spilled liquids or other serious abuse.
Mount/Fasteners ( Acceptance )
Check security of mounts or support mechanisms. Verify that the vaporizer is firmly mounted on
the anesthesia unit and the presence of o-rings, if applicable.
AC Plug ( Acceptance )
If the vaporizer has an AC plug, it should be a solidly constructed, good quality plug with
adequate strain relief, but the use of a Hospital Grade plug (identifiable by a green dot and/or
labeling) will eliminate guesswork and ensure a plug of acceptable construction quality. Rightangle plugs are unacceptable for devices that are moved frequently. A good quality two-prong
plug is acceptable for double-insulated devices. Replace the plug or have the supplier replace it if
it is not Hospital Grade or otherwise suitable. Hospital Grade molded plugs are acceptable.
Examine the AC power plug for damage. Attempt to wiggle the blades to determine if they are
secure. Shake nonmolded plugs and listen for rattles that could indicate loose screws.
AC Plug ( Major )
If the vaporizer has an AC plug, examine it for damage. Attempt to wiggle the blades to
determine that they are secure. Shake the plug and listen for rattles that could indicate loose
screws. If any damage is suspected, open the plug and inspect it.
Line Cord ( Acceptance )
Ensure that the line cord, if so equipped, is long enough for the vaporizer's intended application;
an extension cord should not be required.
The cord should be of suitable quality and current-carrying capacity. Hard Service (SO, ST, or
STO), Junior Hard Service (SJO, SJT, or SJTO), or an equivalent-quality cord should be used.
Verify that the vaporizer has adequate protection against power loss (e.g., from accidental
disconnection of a detachable power cord, disconnection of the power cord from the wall, or
depleted battery if a battery-powered device is not plugged in). Equipment having a detachable
power cord should also have adequate capture devices, cleats, or channels to hold the cord in
place. If these are absent, request that the supplier provide suitable means of securing the cord.
Verify that the vaporizer has adequate alarms or indicators for line-power loss and battery
depletion and an adequate battery-charging indicator.
Line Cord ( Major )
Inspect the cord, if so equipped, for signs of damage. If damaged, replace the entire cord, or if
the damage is near one end, cut out the defective portion.
Strain Reliefs ( Acceptance, Major )
Examine the strain reliefs at both ends of the line cord, if so equipped. Be sure that they hold
the cord securely.

Fittings/Connectors ( Acceptance )
Verify appropriate connectors are supplied if connection to other hospital equipment or systems
is required. Verify that suitable connectors are supplied with the vaporizer so that adapters are
not required.
Examine all gas and liquid fittings and connectors, as well as all electrical cable connectors, for
general condition. Electrical contacts should be straight, clean, and bright. Gas and liquid fittings
should be tight and should not leak. If keyed connectors are used (e.g., pin-indexed gas
connectors), ensure that no pins are missing and that the keying is correct. Keying pins should
be securely seated in "blind" holes so that they cannot be forced in farther.
Fittings/Connectors ( Major )
Examine all gas and liquid fittings and connectors for general condition. Be sure all fittings are
tight.
Controls ( Acceptance )
Examine all controls and switches for physical condition, secure mounting, and correct motion. If
a control has fixed-limit stops, check for proper alignment, as well as positive stopping.
Controls ( Major )
Examine all controls for physical condition, secure mounting, and correct motion. Where a
control should operate against fixed-limit stops, check for proper alignment, as well as positive
stopping. During the course of the inspection, be sure to check that each control performs its
proper function. Return all controls to the off position following the test.
Fluid Levels ( Acceptance )
Check all fluid levels. If the fluid level is low, a qualified user should fill the vaporizer with
anesthetic agent to continue the inspection.
Battery ( Acceptance )
If so equipped, determine the replacement interval for the battery and document the interval in
an equipment control system or a hard copy file for the vaporizer. Be sure to include
batteries/cells for clocks and/or memory logs, if so equipped. It may be desirable to disconnect
the battery and determine if the vaporizer still operates on line power.
Operate the vaporizer on battery power for several minutes to verify that the battery is charged
and can hold a charge. Activate the battery test function, if so equipped. Provide users with
instructions and/or checklist procedure to ensure adequate battery charging and performance.
Battery ( Major )
If so equipped and readily accessible inspect the physical condition of the battery and battery
connectors. Operate the battery-powered functions of the vaporizer for several minutes to check
that the battery has an adequate charge. Check remaining battery capacity by activating the
battery test function or measuring the output voltage. If it is necessary to replace a battery,
label it with the date.
Indicators/Displays ( Acceptance, Major )
Confirm the operation of all indicators and visual displays on the vaporizer, if so equipped.
Self-Test ( Acceptance, Major )
For units with a self-test mode, activate it and determine if the expected response is produced.
Time/Date Settings ( Acceptance, Major )
Verify that the time and date settings on the unit are correct.
Network/Wireless Interfaces ( Acceptance )
Assess the potential risks and vulnerabilities to the confidentiality, integrity, and availability of
electronic information stored or transmitted by the device or system and take appropriate

preventive measures. Make sure that necessary information, including device identification,
operating system, IP (internet protocol) address, is documented in the appropriate log. As
appropriate, verify that confidentiality (e.g., password protection) and malicious software
protective (including mechanisms for future OS and virus protection patches and upgrades)
measures are implemented. Determine if a VPN (virtual private network) is needed. For wireless
devices, ensure that appropriate wireless security measures have been implemented and that
requirements of the healthcare organization's wireless management policies have been met.
Verify that data backup processes are activated and verified.
Network/Wireless Interfaces ( Major )
Review measures taken to ensure protection against the potential risks and vulnerabilities to the
confidentiality, integrity, and availability of electronic information stored or transmitted by the
device or system and verify that preventive measures are still active. For example, are
passwords being applied correctly, have operating system and virus protection patches and
upgrades been installed, is the device still operating on a VPN (virtual private network) and are
wireless security measures still in place. Verify that data backup processes are activated and that
data can be retrieved from backups.
Alarms ( Acceptance )
If the vaporizer has alarms, it should not be possible for them to be turned off, silenced, or
defeated without adequate warning to the operator or automatic alarm reactivation after a short
delay.
Verify that alarms are loud, distinctive, and/or bright enough to be noticed in the environment in
which the vaporizer will normally be used. Audible alarm-volume controls should not allow the
alarm to be turned off or lowered to an indiscernible volume. Operate the vaporizer in such a
way as to activate each audible and visual alarm, if so equipped. If the vaporizer has an alarmsilence feature, check the method of reset (i.e., manual or automatic) against the
manufacturer's specifications.
Alarms ( Major )
If the vaporizer has alarms, operated it in such a way as to activate each audible and visual
alarm. If the vaporizer has an alarm-silence feature, check the method of reset (i.e., manual or
automatic) against the manufacturer's specifications.
Audible Signals ( Acceptance, Major )
If the vaporizer has audible signals, operate it in such a way as to activate the signals. Confirm
appropriate volume, as well as the operation of a volume control, if so equipped.
Interlocks ( Acceptance, Major )
Check that the vaporizer interlock allows activation of only one vaporizer at a time.
Labeling ( Acceptance, Major )
Check that all necessary placards, labels, conversion charts, and instruction cards are present
and legible.
Site Glass, O-Rings, Keyed Filler Mechanism ( Acceptance, Major )
Examine the physical condition of the site glass, O-rings, and keyed filler mechanism, if so
equipped.
Accessories ( Acceptance, Major )
Verify that all necessary features and accessories have been supplied with the vaporizer. A copy
of the operators and service manuals (electronic or hard copy), including schematics, should be
shipped with the equipment. Manuals should be filed in the central equipment file and clinical
instructions should be kept in the patient care area for easy access by clinicians.

Quantitative Tasks
Grounding Resistance ( Acceptance, Major ) 0.5
Measure the resistance between the grounding pin of the power cord (if so equipped) and
exposed (unpainted and not anodized) metal on the chassis. Grounding resistance should not
exceed 0.5 . If the unit is double insulated, grounding resistance need not be measured.
Chassis Leakage Current ( Acceptance ) 500 A
Leakage current must be measured with the device powered by a conventional (grounded)
power system, even if it is normally used in an area with isolated power. ECRI Institute does not
recommend chassis leakage current tests of double-insulated devices.
With the polarity of the power line normal and the equipment ground wire disconnected,
measure chassis leakage current with the device operating in all normal modes, including on,
standby, and off. Maximum leakage current should not exceed 500 A.
Inspect AC adapters used to power (or recharge) certain devices for CE mark or UL (or other
testing laboratory) listing and to verify that it is labeled to identify the device with which it is to
be used. ECRI Institute recommends testing of adapters, particularly those that are not listed,
by measuring the leakage current from each secondary (low voltage) connection to ground. The
leakage current should not exceed the limits for the device chassis leakage current to ground.
Concentration Check ( Acceptance, Major ) 0.3% vapor or 10% of the selected value,
whichever is greater
Record the type and control number of each vaporizer. Because there are various types of
halogenated anesthetic analyzers, follow the manufacturer's procedure for setup and use of the
analyzer.
Vaporizers should usually be tested with an oxygen flow of 4 to 5 L/min (nitrous oxide may
affect the readings of some vapor analyzers). Test the vaporizers at low, medium, and high
concentration settings in the normal clinical use range (e.g., 0.5%, 1.0%, and 3.0% for
halothane). At one concentration setting (e.g., 1.0% for halothane, 10% for desflurane), test the
vaporizer at another flow (e.g., 1 L/min). The concentration should be 0.3% vapor or 10% of
the selected value, whichever is greater. [If errors in concentration are observed, allow the
vaporizer to operate for a minute or two and recheck the unit. Some units may require a short
stabilization period.]

Preventive Maintenance
Clean ( Major )
Clean the exterior.
Replace ( Major )
Replace the battery, if indicated.
Change ( Major )
Change CO2 absorbent if the procedure or investigation results in excessive gas flow through the
carbon dioxide absorber.

Notes
System Components
0

S-ar putea să vă placă și