MANUAL
SINGLE POLE TIMER
MAKE : ANANDHA JOHI
MODEL: JOTHI 545

SINGLE POLE TIMER


CONTENT
CHAPTER

DESCRIPTION

INTRUDUCTION AND

SPECIFICATION

PAGE
NO.
1
4

PANEL DESCRIPTION AND


3

OPERATING

INSTRUSTRION
CIRCUIT DESCRIPTION
APPLICATIONS

5
6

CHAPTER 1
INTRODUCTION AND SPECIFICATION
1.1

INTRODUCTION

This manual covers the specification and operating instructions for


ANANDHA JOTHI Digital Time interval meter.
ANANDHA JOTHI Time Interval meter JOHI 545 is small,
lightweight digital instrument designed for measurement of elapsed
time between any two events. It measures time interval from 0.1ms
to 99.999sec.
The Silent features of ANANDHA JOTHI Time Interval Meter
are:1. Compact and light weight.
2. Input circuit will draw a maximum of 100mA
3. Possibility of 2 different modes of input signals for time
interval measurement
4. Use of latest IC techniques for persistently excellent
performance.
5. Time base accuracy of better then +/- 0.1%
6. Free from mains frequency variation & 10% of the mains
voltage fluctuations.
7. Highly reliable instrument size LED readout.
8. Maximum voltage selectable by switch.

1.2.TECHNICAL SPECIFICATION :
1. DISPLAY
2. RANGE

: 5 Digits LED
: 0.1msec to 99999 seconds with
resolution of o.1ms.

3. TIME SELECTION

: Auto range 00.000 mSec.to 99999

Example milli Seconds : 1. 00.025 m Sec.


Seconds.

2. 09.564 Sec

Seconds.

3. 1034.7 Sec

Seconds.

4. 57631 Sec.

3. ACCURAC

(Dot. Different)

: +/- 1Count +/- Time base accuracy.

4. TIME BASE ACCURACY : +/- 0.1%


5. SUPPLY VOLTAGE

230 V AC +/- 10% at 50Hz.

START / STOP ARRANGEMENT :


START/ STOP TERMINALS : Timer starts by N/O and N/C
(Normally Open and Normally Closed potential free contacts)
ENVIRONMENT : Up to 50 Degree Celsius

CHAPTER II
2.1.GENERAL :
As only most required controls are provided on ANANDHA
JOTHI Time Interval Meter, time Interval between two signals viz,
former to start the counting and latter to stop the counting, can be
measured by merely connecting those signals to the terminals
provided on the front panel of the instrument and selecting the
proper start/stop modes.
2.2. PANEL DISCRIPTION : FRONT PANEL
1. Mains Power Switch
When set to ON Position the instrument is ready for Operation.
2. Reset Switch
By pressing this switch (Push Button) display is reset to Zero
3. Start /Stop mode selection switch
This is a two position selector switch and is used to select the
desired start / stop modes. ( NO / NC)
4. Time Base AUTO RANGE
Selection : 99.999 Sec. 99.999Sec, 999.99sec, 1Sec.
00.000 TO 99.999 SEC.

CHAPTER III
TIMER TEST TERMINALS : RED & BLOCK
In connection Test terminals start / stop signal should be
connected to this terminal.
2.3. OPERATING INSTRUCTIONS:
Set Power switch ON position. All the digits of LED display
should read zero, if not then press reset push button. Connect the
start signal between terminals T1 & T2 Select the proper mode
(NO/NC) before applying the signal. Same way connect the stop
signal.
The instrument is now ready for receiving the stipulate star / stop
signal. Give supply to the set up of which you want to measure
the time.

CHAPTER IV
Control circuit consists of start and stop circuits with concerned
logic circuitry for starting
Display systems consists of counting and stop circuits with
concerned logic circuitry for
Display system consists of counting and displaying of the pulses
coming from the control circuit.

CHATER 5
APPLICATIONS:
ANANDHA JOTHI TIME INTERVAL METER can be used for a
large number of applications where elapse to time between any two
events is to be measured. A few representative applications are
explained in detail in the following explanatory notes:RELAY / CIRCUIT BREAKER OPERATE TIME :
Set the start mode switch NO/NC position. Connect the start
terminals T1 & T2 in parallel with the circuit. As soon as voltage
across relay coil appears start signal is applied to the meter and
instruments starts counting. N/O Position and connect normally
open contact of the relay will be closes, meter stops counting. The
recorded figure on the LED read out is the operate of relay/circuit
breaker.