This guide details installation process for WT210 WiHART sensors. It assumes the user is familiar with
the operation of the IK200 installation kit.
Refer to User guide IK200 installation kit for more detailed usage instructions.
Refer to Overview WT210 WiHART system deployment for the full list of system documentation.
Table of contents
Safety notices .......................................................................................................................................................................... 2
Compatibility .......................................................................................................................................................................... 2
Technical background............................................................................................................................................................ 2
Installation requirements ...................................................................................................................................................... 4
Task 1. Mounting the sensor ................................................................................................................................................ 5
Task 2. Mechanical installation and validation procedure ............................................................................................... 7
Task 3. Completing sensor installation .............................................................................................................................. 9
Re-installing a sensor (if required) .................................................................................................................................... 10
Re-provisioning a sensor (if required) .............................................................................................................................. 10
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
Safety notices
Installation of this sensor in an explosive environment must be in accordance with the
standards and practices appropriate to the site.
Review the Regulatory Compliance section in Datasheet WT210 WiHART sensor for
restrictions for safe installation.
Only fit approved Permasense BP10E or BP20E battery packs.
Ensure the device is installed with the antenna at least 20cm (8) from all persons.
Compatibility
Permasense WT210 WiHART sensors are only compatible with Emerson 1420 gateways with the EIMP
order option.
Permasense BP10 and BP20 battery packs can be used but note that they have a narrower ambient
temperature range than the BP10E and BP20E.
Antenna
Technical background
Permasense systems are permanently installed
ultrasonic corrosion monitoring solutions for use in
oil and gas and other industrial facilities.
At the heart of the systems are permanently
attached dual-probe ultrasonic wall thickness
measurement sensors that use breakthrough
waveguide technology to enable them to operate in
high-temperature environments.
These unique waveguides provide separation
between the sensor head, rated at -50C to +75C
(-58F to +167F) and the potentially hot mounting
location which can be up to 600C (1112F).
Battery pack
Head
(Contrining
electronics)
Waveguides
Stabiliser
Foot
Thermocouple
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
Each sensor is dry-coupled (meaning no couplant is required) to the surface of the pipe using a pressure
contact, obtained with force applied from mounting the sensor on studs and securing with nuts.
Note: Installing the sensor correctly, to ensure proper coupling between the sensor tip and the pipe
surface, is key to ensuring correct functionality of each sensor.
Once correctly installed, the ultrasonic signal produced by the sensor travels down one waveguide. Where
the waveguide contacts the pipe, ultrasound waves travel across the outer surface which is seen in the
received signal as a surface wave (S). Ultrasound waves also travel through and reverberate in the pipe wall
which is seen in the received signal as multiple back wall reflections (B1, B2, etc). The thickness is
calculated from these signals.
Once a sensor is installed, it will communicate with the gateway specified during installation, and any other
sensors already installed that are assigned to the same gateway. These sensors form a wireless mesh
network.
The gateway manages this mesh, selecting the best path to transmit data from each sensor to the gateway.
The gateway then delivers this data to the Permasense database.
Data from all sensors can be viewed by the end user via the Permasense Data Manager browser interface.
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
Installation requirements
Before sensor installation:
Pairs of M8 (5/16") studs must be welded to the pipe work at the desired sensor locations. See
Installation Guide Stud welding.
Required hardware
(supplied with Permasense WT210 sensors)
Consumables
(supplied in Permasense IK200 Installation Kit)
Required tooling
(supplied in Permasense IK200 Installation Kit)
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
6. Locate and place the WT210 sensor over the studs, ensuring the waveguides are positioned centrally
(5mm, 0.2) .
7. Place Nord-Lock washers over the studs.
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
8. Thread nuts onto the studs, and run them down several threads.
9. Observe the sensor contact from the side. As shown in the following illustrations, check that:
10. Carefully secure the sensor in this position by finger tightening each nut.
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
6) Enter
network ID and
Join Key
7) Click
Provision
8) Check sensor
can hear the
network
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
be completed, but the sensor is unlikely to join the network without additional sensors to extend the
mesh.
9. Check that:
As the sensor is tightened and becomes coupled to the pipe, the coupling amplitude will rise in line
with each 1/4 turn tightening of the nuts.
At 10 seconds after the start button is pressed, an ultrasonic waveform will appear in the bottom
window. This waveform is continuously updated every 10 seconds.
The ultrasonic waveform will comprise several peaks. It is important that the first TWO peaks are
clean, and well defined, as in the example illustrated.
12. Continue to tighten each nut in 1/4 turn increments, alternating between nuts until torque is reached
on each nut. Continue to monitor the install tool software during this process.
Overtightening the nuts can damage the waveguides
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
The coupling amplitude history plot shows an increasing trend, in-line with nut tightening.
The coupling amplitude is above 40.
The last displayed ultrasonic waveform shows the first two peaks are clean and well defined.
The measured thickness displayed in the bottom right of the window is comparable with expectations
for the measurement location.
14. When all of the above criteria are satisfactorily met, press the Complete button.
15. Ensure all required sensor information is accurately logged (e.g. sensor ID and location).
16. Remove the CC21 and fit the battery pack.
When the battery pack is fitted, the sensor will restart and try to join the WiHART gateway. In a large
network of 100 sensors this can often take 2 hours, and sometimes 6 hours.
Refer to Installation guide BP series battery pack.
Sensor installation is now complete.
17. To protect the sensor from high levels of radiated heat, install the circular stainless steel heat shield. At
insulated locations the heat shield can be fixed to the insulation jacket.
See the document Guidelines - Re-insulation of lines following sensor installation.Task 4. Fitting the
lanyard
Install the lanyard around the pipe and sensor to
provide a fall protection restraint.
1. Wrap the lanyard around the circumference of the
pipe. The 2m (7ft) length will accommodate a
maximum diameter of 24.
2. Thread the bare end of the wire through the loop
in the lanyard to secure it to the pipe.
3. Feed the bare end of the lanyard into the gripple
and push the gripple 15cm (6) up from the bare
end.
4. Feed the bare end between the waveguides near
the sensor head, loop back around a waveguide
without trapping the thermocouple. Feed into the
return hole of the gripple. One of the waveguides
should now be in the loop.
Note. The wire can be released from the gripple
using the release key.
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
10
Permasense Ltd, Century House, 100 Station Road, Horsham, RH13 5UZ, UK
www.permasense.com
support@permasense.com
+44 20 3002 0922
Revision 1, 20/05/2014 Permasense Ltd. Permasense is a registered trademark of Permasense Ltd.