Service Test Form

System S/N: __________________Service Ser. ____________

New

Customer

Loaner

Resale

Marketing

Final Test /
Upon Arrival Test
Date Originated: __________________
System Damage Upon arrival
Yes /
No
If Yes, Add Damage Description & Cause: _________________________________________________________________________
Function Test - User Mode
Applicator Display
Ensure the OLED screen on applicators panel is illuminated.
Check the default vacuum level is 1. Press the trigger; the vacuum pump will start to operate, and the pressure
valve will operate intermittently. Release the trigger. Repeat the test with the Vacuum set to level 2, and then
Vacuum Levels
to level 3, ensuring the pressure valve operation changes accordingly. Verify all three levels on applicator
mirror the levels on the GUI when changed. Repeat this entire test using the applicator keypad.
Press the IR icon on system display to set IR to level 1 and press the trigger. Verify IR lamp turns on by pressing
IR Lamp
the trigger. Repeat test with IR set to level 2, and then to level 3. Verify all three levels on applicator mirror the
levels on the GUI when changed. Repeat this entire test using the applicator keypad
Press the RF icon on system display and ensure that the RF is set to level 1. Press the trigger. Repeat the test
RF Display
with the RF set to level 2, and then to level 3. Verify all 3 levels on applicator mirror the levels on the GUI when
changed. Repeat this entire test using the applicator keypad.
Footswitch
With an applicator connected, press and hold the footswitch, ensure vacuum is pulsing.
With an applicator and disposable cover connected, turn on system; ensure no error messages appear upon
Initialization Test
start-up. Confirm that the disposable cover is recognized by applicator.
IR Sensor Verification [User Mode, System and Hot Plate must be turned ON for 20 minutes at least]
Place the applicator without the cover over the hot plate (tool P/N TL74260)
VC / VS
_________/________ C
and check the IR sensor temperature results on the screen: 43C 2C
RF Testing [Service Mode, press the RF test button and the footswitch/trigger]
RF out w/Load
VC / VS: _______/_______ Vpp
RF Fans
Pump Fans
Power Supply Fan
Vacuum Testing [Service Mode ]

Measured value should be : 150V<Vpp<175V


Service Mode Turn on the fans, ensure proper operation.

System

Above (0.50) [Bar]

Above (6.8) [PSI]

Umbilical

(0.45) to (0.85) [Bar]

(-6.53) to (12.35) [PSI]

VC Applicator

(0.45) to (0.85) [Bar]

(-6.53) to (12.35) [PSI]

VS Applicator

(0.45) to (0.85) [Bar]

(-6.53) to (12.35) [PSI]

Vacuum Leakage Test [Service Mode]


(Vacuum level drop within 60 second < 60 [mBar])
System
Umbilical
VC S/N: _______________
VS S/N: _______________

Fans Test [Service Mode]


Pump Fans
RF Fans

HC SW Version: ______________
HC PLD Version: ______________
GUI Version: _________________

PS Fan

Notes

Date:_________________________ Call ID:_____________________ Checked By:___________________________________________

2016, Syneron, Inc. All rights reserved.

PB81503LW-TF