Documente Academic
Documente Profesional
Documente Cultură
DOI 10.1007/s00330-005-2678-0
Fritz Schick
F. Schick (*)
Section of Experimental Radiology,
Department of Diagnostic Radiology,
Eberhard Karls University of Tbingen,
Tbingen, Germany
e-mail: fritz.schick@med.unituebingen.de
Introduction
In the past 2 decades technical development of clinical
magnetic resonance (MR) units was extremely fast. Meanwhile, modern standard whole-body MR systems operating at 1.5 T provide high image quality in morphological
and functional clinical examinations of all body regions in
reasonable measuring times. However, further increased
spatial resolution and shorter examination times remain
desirable in order to improve diagnostic sensitivity and
specificity of MR examinations.
In the early days of whole-body MR systems the construction of supraconducting magnets with high spatial and
temporal homogeneity played an important role, but as
947
to reposition the patient in the scanner manually. Additionally, the presence of multiple receiver coils was found
to be useful for improvement of spatial resolution or
shortening of measuring times. So-called parallel imaging
strategies [13] can clearly reduce the number of phaseencoding steps at the expense of a slightly decreased signal-to-noise ratio. There are still possibilities to further
improve MR examinations in the future using multiple coil
systems and parallel imaging applications; however, the
size of the single coils must not be chosen too small, since
sensitivity to regions deep in the body gets worse for
smaller coils [4]. This phenomenon restricts the possible
number of single receiver coils and the corresponding
saving of examination time by parallel imaging. An extreme increase in signal yield or noise suppression from
new receiver systems cannot be expected in the near future
for units operating at magnetic fields up to 1.5 T.
Parallel imaging techniques would allow the recording of
even large 3D data sets in relatively short measuring times,
but the signal-to-noise ratios in the resulting images are
often insufficient if high spatial resolution is selected. For
this reason the main limitation in high-resolution MRI is
now the signal yield per volume compared with the noise
level. Parallel imaging has also paved the way to highly
resolved imaging of the body trunk in very short measuring
times, allowing even non-ECG-triggered examinations of
the heart. However, those applications are also limited by
the relatively low signal-to-noise ratio achievable on common MR units operating at 1.5 T.
For a long period of nearly 2 decades progress in MRI
took place at a nearly constant magnetic field strength of
1.5 T. Since developments of gradient and receiver systems
are now considered to be close to the borders of feasibility,
the problem of low signal yield can only be solved by
increasing the field strength. In addition, high-field animal
scanners have shown excellent results with extremely high
spatial resolution [5] and are supportive for further developments of human scanners with higher fields.
Human neuroscience groups have already demonstrated
clear advantages of higher field strength in blood oxygen
level dependent (BOLD) studies of brain activity using
high-field systems mostly operating at field strengths of
34 T [6, 7]. Prototype MR head scanners (or small-bore
whole-body units) with extremely high magnetic field
strengths of 79.4 T showed further improved functional
MRI (fMRI) and spectroscopic results from human examinations with small receiver coils from brain areas near the
scull [8, 9], but no high-quality images covering the entire
human head were reported from such ultra-high MR
systems.
Only a few reports on preclinical or clinical whole-body
applications at 3 T or more have been published [1032].
By this review the author intends to provide a comprehensive overview of relevant technical aspects gaining
more and more importance for increasing field strength,
their implications for various MRI techniques and spec-
948
1.5
H
C
23
Na
31
P
1
H
13
C
23
Na
31
P
1
H
13
C
23
Na
31
P
13
3.0
7.0
470
1,870
1,780
1,160
235
940
890
580
100
400
380
250
Wavelength in water
(cm)
52
210
200
129
26
105
100
64
11
45
42
28
Relaxation times
The longitudinal relaxation times T1 of most tissues are
significantly longer at higher field strength; however, this
prolongation is tissue-dependent and influences the contrast behaviour in T1-weighted imaging. Several publications comparing relaxation times between 1.5 and 3.0 T
[3436] have reported changes between 5% for skeletal
musculature and nearly 40% for white matter in brain and
liver.
Transverse relaxation times T2 have been reported to be
less dependent on field strength and mostly slightly reduced. Only changes in the range 510% have been found
for most organs.
Sequences with T2* weighting are markedly influenced
by the field strength, since those sequences are sensitive to
microscopic field inhomogeneities due to tissue susceptibility. However, this is not really a relaxation effect and is
therefore reported in the Susceptibility effects section of
this article.
Susceptibility effects
All paramagnetic and diamagnetic tissue compartments as
well as implanted metallic structures lead to magnetic field
distortions inside those materials and in their surroundings. The induced field distortions are proportional to the
949
1.0 Tesla:
1.5 Tesla:
3.0 Tesla:
7.0 Tesla:
H2O
H2O
H2O
H2O
= 79 cm
= 52 cm
= 26 cm
= 11 cm
Fig. 1 RF wavelength (according to the 1H Larmor frequency) in the medium water for several common field strengths. The wavelengths
are compared with body dimensions.
applied outer magnetic field strength, as long as no ferromagnetic materials are involved. This statement is valid
on a microscopic and on a macroscopic scale. Microscopic
magnetic field inhomogeneities inside picture elements
increase linearly with field strength, leading to accordingly
faster dephasing of the contributing frequency components.
This increasingly rapid dephasing for higher field strength
is especially important for tissue types with marked inherent structural inhomogeneities, such as lung tissue or
spongy bone marrow. It should be mentioned that susceptibility-related dephasing of magnetisation due to substances such as deoxyhaemoglobin in the blood (relevant
950
EMCL
a
TMA
Cr2
100
IMCL
Cr3
-100
-200
Chem. Shift / Hz
-300
EMCL
b
Cr2
100
-100
IMCL
TMA Cr
3
-200
-300
Chem. Shift / Hz
-400
-500
-600
951
Fat-suppressed imaging
Frequency-selective or inversion-recovery fat-suppression
techniques are often applied in clinical routines using spinecho or gradient-echo sequences at 1.5 T. Since both
macroscopic field distortions and the chemical shift difference between water and fat signals are increased in the
same manner at higher fields, shimming works similarly
efficiently at higher field strength in most body regions as
known from examinations at 1.5 T. One important effect of
fat suppression at high field is the lack of chemical shift
artifacts allowing application of low readout bandwidth.
For this reason one can gain full advantage from the
increased signal yield at higher field strength. Improved
spatial resolution or shorter measuring times, especially in
combination with parallel imaging strategies such as
SENSE and SMASH [13], can be easily obtained at 3 T
using fat-suppressed imaging techniques.
Echoplanar and BOLD contrast imaging
Magnetic field inhomogeneities induced by diamagnetic or
paramagnetic material in the body linearly increase with
field strength. These field inhomogeneities lead to more
pronounced distortions in echoplanar images at higher field
strength if sequence parameters are not adapted. Adaptation
of sequence parameters means increasing the readout bandwidth in order to reduce echo spacing and echo train length.
Doing so enables one to record more echoplanar images per
measuring time or even multiecho echoplanar imaging [39],
but the signal-to-noise ratio in every image is only slightly improved at higher field strength owing to the higher
readout bandwidth. However, combination of higher field
strength with parallel imaging techniques and multishot
echoplanar techniques provides more flexibility and increased image quality of echoplanar images at higher field.
The BOLD effect, i.e., signal alteration owing to tissue
oxygenation, is clearly more pronounced at higher field for
a given echo time in T2*-weighted sequences. For this
reason it is possible to reduce the echo time to achieve
more total signal and a shorter measuring time. Using
adapted echoplanar techniques for BOLD imaging leads
finally to more than linear gain in accuracy or effectiveness for higher field strengths.
MR angiography and perfusion imaging with and
without contrast media
At higher field strength longitudinal relaxation times T1 of
tissues tend to get longer, whereas the relaxivity of most
paramagnetic contrast media (e.g., Gd-DTPA) in blood is
only slightly decreased [40, 41]. Altogether, conditions for
perfusion studies with contrast media are improved and
952
xx
Dielectric Pad
953
Spectroscopy
954
Inhomogeneities of RF fields
Safety issues
A higher RF leads to altered conditions for electromagnetic
resonance of conductive material at higher field strength.
955
with increasing field strength [6, 7, 52]. In addition, spectroscopic studies of the brain show clearly higher quality at
3 T, with better spectral separation of relevant metabolites
such as glutamine and glutamate [5355].
Anatomical imaging of the brain at 3 T also profits from a
higher signal-to-noise ratio (nearly doubled as expected)
compared with imaging at 1.5 T, but reduced contrast
between grey and white matter is obvious in T1-weighted
images (Fig. 7). The contrast behaviour in T2-weighted
imaging is very similar for both field strengths as demonstrated in Fig. 7. More detail on specific sequences for brain
imaging is beyond the scope of this article and is reported in
the literature [5660].
Soft tissue in the neck contains layers of fat tissue. For
this reason head and neck imaging requires an increased
receiver bandwidth at higher field strength, since otherwise the chemical shift artifacts get too prominent. This
leads to higher noise levels in the images and the overall
quality in standard T1- or T2-weighted spin-echo or fast
spin-echo imaging is only slightly increased at 3 T.
956
Fig. 10 Three-dimensional
knee imaging at 3 T with an
isotropic resolution of 0.5 mm in
a measuring time of 8 min 20 s
using a knee coil of the manufacturer with eight elements. A
double-echo steady-state sequence with water-selective
excitation and TR=22 ms,
TE=6.9 ms and fl=25 was
applied, resulting in high contrast between bone, cartilage and
synovial fluid. Parallel imaging
with factor 2 provided an acceptable measuring time.
957
field inhomogeneities in each pixel combined with diffusion make lung imaging at 3 T very difficult, and extremely
short echo times are necessary to achieve sufficient signal
intensity from lung structures.
Morphological and functional cardiac imaging using
spin-echo or spoiled gradient-echo sequences, respectively,
shows clearly improved results at 3 T [12, 21, 25, 26]. The
increased signal-to-noise ratio allows one to take advantage
of parallel imaging strategies [13], leading to shorter measuring times or a higher number of recorded slices. It should
be mentioned that higher field strengths unfortunately limit
applications of modern SSFP (e.g., True-FISP) sequences.
Artificial dark lines occur with higher spatial frequency
owing to the more pronounced macroscopic static field
inhomogeneities, but after shimming artifact-free images
of the entire human heart can usually be obtained (Fig. 8).
The maximal flip angles in such sequences are restricted
owing to the SAR limits. However, experimental comparison showed improved image quality at 3 T even for TrueFISP sequences.
Abdomen and pelvis
Abdominal and sometimes pelvic imaging is affected by the
mentioned dielectric effects. Distinct signal voids most
often occur in areas behind the ventral abdominal wall
(Fig. 5a). The relaxation times of abdominal and pelvic
organs are slightly different at 3 T [36], but T1- and T2weighted images at higher field strength show contrast behaviour similar to the findings at 1.5 T to which radiologists
are used to. An example is shown in Fig. 9. The current
quality of abdominal and pelvic MR images recorded at
3 T is not always superior to examinations with optimised
array coil systems on modern MR units operating at 1.5 T.
Further improvements in RF transmission and receiver coil
systems are expected to lead to superior results on 3 T
systems within the next few years.
Knee and wrist/hand
Musculoskeletal imaging of knee and wrist clearly benefits
from a higher field strength for several reasons. The crosssectional dimensions of the extremities allowed development of adapted transmit/receive coils or even multichannel
coil systems with high sensitivity to the increased MR
signals. For this reason extremely high spatial resolution
can be obtained in a moderate examination time. Parallel
acquisition strategies are clearly supportive for imaging at 3
T and allow more flexible optimisation regarding spatial
resolution and examination times. Multiplanar images
reconstructed from a 3D data set of a knee examination
are shown in Fig. 10. High resolution and good contrast
between cartilage, bone and synovial fluid were achieved
Conclusions
Altogether whole-body MR imaging at 3 T is feasible now,
but the quality of the results is particularly dependent on the
body region examined and on the sequence technique
applied. In the present stage, fMRI of the brain, anatomical
imaging of the extremities and spectroscopic examinations
(brain, prostate, and musculature) show clearly improved
results at 3 T compared with 1.5 T. Most other examinations
provide at least similar image quality at higher field
strength. Only examinations of the abdomen and pelvis are
in some cases inferior owing to problems with dielectric
effects. In addition, fast spin-echo sequences and True-FISP
sequences are often hampered by the SAR limits, especially
when the body coil must be used for RF transmission.
Undesired heating hazards in the presence of conductive
material have to be considered very carefully at higher field
strengths.
The course of hardware and software development for
optimisation of MRI at 1.5 T has a history of approximately
20 years, and a very high level has already been reached. In
contrast, clearly less effort could be made for whole-body
applications at 3 T, and there are probably more opportunities for further developments with high impact in the
future. It is expected that current problems, especially for
imaging of the body trunk, will be solved within the next
few years. Thereafter, complete whole-body examinations
could be possibly performed with clearly higher quality as
on modern 1.5 T systems at present.
Acknowledgements The author wishes to thank all contributors
from several research groups working on the 3-T whole-body system in Tbingen, which was financed by the Deutsche Forschungsgemeinschaft (Th 812/1-1). The images and spectra presented in
this review article were recorded and prepared with the help of
Michael Fenchel, Hansjrg Graf, Sabine Lenk, Matthias Lichy,
Jrgen Machann, Petros Martirosian, Stephan Miller, Heinz-Peter
Schlemmer, Gnter Steidle and Beate Wietek. The Chairman of the
Department of Diagnostic Radiology of the University Clinic of
Tbingen, Claus D. Claussen, is acknowledged for continuous support. Many experimental and clinical studies at 3 T were supported
by members of Siemens Medical Solutions.
958
References
1. Pruessmann KP, Weiger M,
Scheidegger MB, Boesiger P (1999)
SENSE: sensitivity encoding for fast
MRI. Magn Reson Med 42:952962
2. Wang Y (2000) Description of parallel
imaging in MRI using multiple coils.
Magn Reson Med 44:495499
3. Paschal CB, Morris HD (2004) K-space
in the clinic. J Magn Reson Imaging
19:145159
4. De Zwart JA, Ledden PJ, van Gelderen
P, Bodurka J, Chu R, Duyn JH (2004)
Signal-to-noise ratio and parallel imaging performance of a 16-channel
receive-only brain coil array at 3.0
Tesla. Magn Reson Med 51:2226
5. Marzola P, Osculati F, Sbarbati A
(2003) High field MRI in preclinical
research. Eur J Radiol 48:165170
6. Chen W, Ugurbil K (1999) High spatial
resolution functional magnetic resonance imaging at very-high-magnetic
field. Top Magn Reson Imaging 10:63
78
7. Silvennoinen MJ, Clingman CS, Golay
X, Kauppinen RA, van Zijl PC (2003)
Comparison of the dependence of
blood R2 and R2* on oxygen saturation
at 1.5 and 4.7 Tesla. Magn Reson Med
49:4760
8. Tkac I, Andersen P, Adriany G, Merkle
H, Ugurbil K, Gruetter R (2001) In
vivo 1H NMR spectroscopy of the
human brain at 7 T. Magn Reson Med
46:451456
9. Ugurbil K, Adriany G, Andersen P,
Chen W, Garwood M, Gruetter R,
Henry PG, Kim SG, Lieu H,
Tkac I, Vaughan T, Van De Moortele
PF, Yacoub E, Zhu XH (2003) Ultrahigh field magnetic resonance imaging
and spectroscopy. Magn Reson Imaging 21:12631281
10. Rotter M, Berg A, Langenberger H,
Grampp S, Imhof H, Moser E (2001)
Autocorrelation analysis of bone structure. J Magn Reson Imaging 14(1):87
93
11. Uematsu H, Dougherty L, Takahashi
M, Ohno Y, Nakatsu M, Song HK,
Ferrari VA, Gefter WB, Schnall MD,
Hatabu H (2001) Pulmonary MR angiography with contrast agent at 4
Tesla: a preliminary result. Magn Reson
Med 46:10281030
12. Dougherty L, Connick TJ, Mizsei G
(2001) Cardiac imaging at 4 Tesla.
Magn Reson Med 45:176178
13. Klarhofer M, Csapo B, Balassy C,
Szeles JC, Moser E (2001) High-resolution blood flow velocity measurements in the human finger. Magn
Reson Med 45:716719
959