Data Sheet

D184S073U02

Level 0: Information

Function
Electromagnetic flowmeters can be used to accurately
measure the flowrate of liquids, slurries, pastes and
sludges with an electrical conductivity of at least 20 S/cm
(opt. 0.5 S/cm).

Applications
The SM4000 is specially designed to measure liquids in the
pulp & paper and food & beverage industry. It can be used
to measure pulp stocksflows, fast changing process, two
phase liquids, mining slurries and pulsating flows (pist on
pump application).

Instrument Advantages
Digital Signal Processing (DSP) assures long term stability
and accuracy in both flow directions with absolute zero
stability.
Integrated diagnostic tools including system monitoring
with the error analysis displayed and signalled over a
contact output for increased operational reliability.
Easy to use menudriven operating structure for
application the configuration. Easy to read the fullgraphic
display is backlighted.

Important Instrument Features

flowmeter primary designed per PEDGuideline 97/23 EC


diameter range from DN 1 to DN 1000 [1/25 to 40]
accuracy < 0.5 of rate
fluid temperatures 40 C to 130 C (opt. 180 C)
Hygienic Certification 3A, FML, EHEDG
liners, PFA, PTFE, hard rubber
standardized installation lengths for DIN or ANSI flanges
Power supply voltage AC/DC

FieldIT
Electromagnetic Flowmeter
FSM4000

Electromagnetic Flowmeter FSM4000


D184S073U02

Accuracy, Reference Conditions and Operating


Principle
Reference Conditions per EN 29104
Fluid Temperature
20 C 2 K
Ambient Temperature
20 C 2 K
Supply Power
Nominal voltage per name plate UN 1 %
Reference installation conditions
Upstream > 10 x D,
Downstream > 5 x D,
D = nominal size of the flowmeter primary

Operating Principle
Faradays Laws of Induction form the basis for the electromagnetic
flowmeters. It states that a voltage is induced in a conductor as it
moves through a magnetic field.
This principle is applied to a conductive fluid which flows through a
magnetic field generated perpendicular to the flow direction (see
Schematic).
The voltage induced in the fluid is maesured at two electordes,
installed diametrically opposed. This signal voltage UE is proportional
to the magnetic induction B, the electrode spacing D and the average
flow velocity v.
Noting that the magnetic induction B and the electrode spacing D are
constants, a proportionality exists between the signal voltage UE and
the average flow velocity v. The equation for the volume flow shows
that the signal voltage UE is linear and proportional to the volume
flowrate.
The induced signal voltage is processed in the converter into scaled,
analog and digital signals.

Warm-up Time
30 min

Magnet Coil

Analog Output Effects


Same as pulse output plus 0.1 % of rate 0.01 mA

z
Meter Tube in
Electrode Plane

Accuracy

Pulse Output
Q> 0.05 Rangemax 0.5 % of rate
Q< 0.05 Rangemax 0.00025 Rangemax

9
8
7
6

B
D
Signal electrode

5
4
3
2
1
0.5

E
x

UE
Signal voltage

10

20

90

100 %

0.5

10

Q
Range max
v

m
s

UE
B
D
v
qv

= Signal voltage
= Magnetic induction
= Electrode spacing
= Average flow velocity
= Volume flowrate

UE ~ B D v
2

D
qv = 
v
4

UE ~ qv

Fig. 1

Accuracy

Fig. 2

Electromagnetic Flowmeter Schematic

Electromagnetic Flowmeter FSM4000


D184S073U02

Overview, Flowmeter Primary and Converter Design


Variable Connections

Connection Types

Rigid Flanges

Wafer Design

Rigid Flanges

Weld stubs

DN 3DN 40
1/1011/2

Pipe fitting
Food industry fitting

DN 3DN 40
1/1011/2

Tri-Clamp

External threads

DN 50DN 100
2 4

Additional, see Dimensions,


Others upon request

DN 50DN 100
24

0.5 % of rate
Complete stainless steel housing Series 2000

Accuracy
Flowmeter primary hsg. material

Alum. Series 4000

Flowmeter Primary
Wafer Design

SE21*
PN
Inch
1/104

Flanged DIN 2501


Flanged ASME B16.5

3100
3100

1/104
1/104

Food Industry Fitting DIN 11851

1/101 1/2 40 S
340
16 S
50, 80 2, 3
10 S
65, 100 2 1/2, 4
1/101 1/2 40 R
340
16 R
50, 80 2, 3
10 R
65, 100 2 1/2, 4
1/101 1/2 40 Q
340
16 Q
50, 80 2, 3
10 Q
65, 100 2 1/2, 4
40 P
2540 11 1/2
16 P
50, 80 2, 3
10 P
65, 100 2 1/2, 4
16 T
1/102
350
10 T
65, 100 2 1/24
10 E
1/101
325
1/251/12 10 B
12
PEEK, Torlon (<DN 3[1/10])
PFA (>DN 2 [1/12])

Weld stubs DIN 11850

Weld stubs DIN 2463

Weld stubs ISO 2037

TriClamp DIN 32676


External threads ISO 228
1/8 Sanitary connectors
Liner
Conductivity
Electrodes
Process connection material
Protection Class per EN 60529
Fluid temperature

SE41F
SE21W
PN
PN
Inch
DN

350 1/102 1040

65100 21/24 1016


1040
DN3DN1000

3100 1040
CL150
1/104150300 CL/JIS 350 1/102 CL300/JIS 1/1040
65100 21/24 CL150/JIS 1/1012
CL 300

SE21F

DN
3100

Model Number

Process Connections

*
W
F
F

PN

DN

PFA

PFA

Hard rubber, soft rubber,


PTFE, PFA, others

20 S/cm Option 0.5 S/cm


SST No. 1.4571 [316Ti], 1.4539, Hastelloy B2/C4, PlatinumIridium, Tantalum, Titanium
Steel,SSTNo.1.4571[316Ti]

SST No. 1.4571 [316Ti]


SST1.4404[316L]
IP 67 / IP 68
IP 67/ IP 68
IP 67 / IP 68
IP 67/ IP 68
25 to 130 C / 180 C
25 to 130 C
40 to 130 C
25 to 130 C

Approvals
Hygienic & sterile requirements

CIP/SIPcapable FML, 3A, EHEDG

CIPcapable

Converter
Model Number
Supply power
Current output standard
Pulse output, 1channel std.
Ext. Zero return
Ext. Totalizer reset
Forward/reverse metering
Communication
Fluid monitor std.
Detection empty pipe
Local indicator/totalizer
Auto. density correction
Protection Class / EN 60529
Housing

S4
85 253 V AC, 24 V AC/DC
0/210 mA, 0/420 mA
active (24 V), optocoupler (220 mA)
yes
yes
yes
HARTProtocol (PROFIBUS PA, FF in preparation)
yes, from DN 10 [3/8] and 20 S/cm
yes
yes
yes, manual entry (totalize and indicate in mass units)
IP 67, NEMA 4X
Field mount housing

Electromagnetic Flowmeter FSM4000


D184S073U02

Specifications: Flowmeter Primary

Flange Designs and Pressure Ratings


Meter Size Flanges

Material

Press. Rating

PED

3 25
1/10 1

DIN
ASME
JIS

SS1.4571[316Ti]
or Steel

40 bar
CL 150, CL 300
10 bar

SEP
Art.3
Par. 3

32
11/4

DIN
ASME
JIS

SS1.4571[316Ti]
or Steel

40 bar
CL 150, CL 300
10 bar

40
11/2

DIN
ASME
JIS

SS1.4571[316Ti]
or Steel

40 bar
CL 150, CL 300
10 bar

50
2

DIN
ASME
JIS

SS1.4571[316Ti]
or Steel

40 bar
CL 150, CL 300
10 bar

65
21/2

DIN
ASME
JIS

SS1.4571[316Ti]
or Steel

16, 40 bar
CL 150, CL 300
10 bar

Meter
Std. Pressure
Size
Rating
Dn Inch
PN

Min. Flow Range


Flow Velocity
0 to 0.5 m/s

Max. Flow Range


Flow Velocity
0 to 10 m/s

1
1/25
1.5 1/17
2
1/12

10
10
10

0 to
0 to
0 to

0.03 l/min
0.06 l/min
0.1 l/min

0 to
0 to
0 to

0.6 l/min
1.2 l/min
2 l/min

3
4
6

1/10
5/32
1/4

40
40
40

0 to
0 to
0 to

0.2 l/min
0.4 l/min
1 l/min

0 to
0 to
0 to

4 l/min
8 l/min
20 l/min

8
10
15

5/16
3/8
1/2

40
40
40

0 to
0 to
0 to

1.5 l/min
2.25 l/min
5 l/min

0 to
0 to
0 to

30 l/min
45 l/min
100 l/min

20
25

3/4
1

40
40

0 to
0 to

7.5 l/min
10 l/min

0 to
0 to

150 l/min
200 l/min

32
40

11/4
11/2

40
40

0 to
0 to

20 l/min
30 l/min

0 to
0 to

400 l/min
600 l/min

80
3

SS1.4571[316Ti]
or Steel

50
65
80

2
21/2
3

40
40
40

0 to
0 to
0 to

3 m3/h
6 m3/h
9 m3/h

0 to
0 to
0 to

60 m3/h
120 m3/h
180 m3/h

DIN
ASME
JIS

40 bar
CL 150, CL 300
10 bar

100
4

SS1.4571[316Ti]
or Steel

100 4
125 5
150 6

16
16
16

0 to
0 to
0 to

12 m3/h
21 m3/h
30 m3/h

0 to
0 to
0 to

240 m3/h
420 m3/h
600 m3/h

DIN
ASME
JIS

16, 40 bar
CL 150, CL 300
10 bar

125
5

DIN
ASME

SS1.4571[316Ti]
or Steel

16, 40 bar
CL 150, CL 300

200 8
250 10
300 12

10/16
10/16
10/16

0 to
0 to
0 to

54 m3/h
90 m3/h
120 m3/h

0 to
0 to
0 to

1080 m3/h
1800 m3/h
2400 m3/h

150
6

DIN
ASME

SS1.4571[316Ti]
or Steel

16, 40 bar
CL 150, CL 300

350
400
450
500

10/16
10/16
Cl150
10

0 to
0 to
0 to
0 to

165
225
300
330

m3/h
m3/h
m3/h
m3/h

0 to
0 to
0 to
0 to

3300
4500
6000
6600

200
8

DIN
ASME

SS1.4571[316Ti] 10, 16, 25, 40 bar


or Steel
CL 150, CL 300

250
10

DIN
ASME

SS1.4571[316Ti] 10, 16, 25, 40 bar


or Steel
CL 150, CL 300

600 24
700 28
800 32

10
10
10

0 to
0 to
0 to

480 m3/h
660 m3/h
900 m3/h

0 to 9600 m3/h
0 to 13200 m3/h
0 to 18000 m3/h

300
12

DIN
ASME

SS1.4571[316Ti] 10, 16, 25, 40 bar


or Steel
CL 150, CL 300
SS1.4571[316Ti]
or Steel

10
10

0 to
0 to

0 to 24000 m3/h
0 to 27000 m3/h

DIN
ASME

900 36
100040

1200 m3/h
1350 m3/h

350 1000
14 40

14
16
18
20

m3/h
m3/h
m3/h
m3/h

Conformity Evaluation per Category III, Module B1 + D, Fluid Group 1

Flow Ranges, Meter Sizes and Pressure Ratings

10, 16, 25, bar


CL 150

Other meter sizes, press. ratings, temperature classes upon request.

Designs
Information:
When a preamplifier is installed for low conditivity
applications and the fluid has a high r (e.g. demineralized
water r = 78) the fluid velocity is limited < 1 m/s.

Meter tube
Stainless steel No. 1.4301 [304]
DN 3 to DN 300 [1/10 to 12]
Two piece housing: Cast Alum., painted1)
Flanges
Steel, zinc plated, standard
Stn. stl 1.4571[316Ti] ( DN15[1/2] standard)
DN 350 to DN 1000 [14 to 40]
Steel, welded design, painted1)
Flanges
Steel, painted, standard,
stn. stl. 1.4571[316Ti]
Connection Box
Body: RAL 70122)
Cover: RAL 90021)
1) Paint coat 80 m thick, RAL 9002 (white grey), painting see page 8
2) Paint coat 80 m thick, RAL 7012 (Basalt grey)

Electromagnetic Flowmeter FSM4000


D184S073U02

Material strength for Model SE41F

Installation Length, Flanged (Short Design)


Meter sizes DN 10 DN 400 [3/816] correspond to DIN/ANSI
Flanged design VDE/VDI 2641 and the installation lengths defined in
DVGW Working Paper, W420 f (Water Totalizers, Design WP ISO
4064 Short as well as ISO 13359).
ANSI CL 150/CL 300

Limits for the allowable fluid temperature (TS) and allowable


pressure (PS) are a function of the liner and flange materials of
the flowmeter (see instrument Name Plate).
Material Strength Curves
Liner: PTFE, hard/soft rubber (limited to 90 C, 194 F)
45

Pipeline Vibrations
Max. allow. 15 m/s2 (10 150 Hz)

Temperature Diagram
Ambient Temperature C
60

High temperature range

40

Druck / Pressure (PS) [bar]

With PFA, installation length, flanged to ISO 13359


with PTFE, Hard/soft rubber
Installation length, flanged to ISO 13359 or ABB (Installation length
Series 1000)

Attention:

PN 40

35

30
25

PN 25

20

15

PN 16

10

PN 10

50

0
-30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110 120 130140150160170180190

Temperatur / Temperature (TS) [C]

10

Fig. 4

DINFlanges SST No. 1.4571[316Ti] to DN 600 [24]

0
180

25

Fluid Temp.
C

Liner: PTFE, hard/soft rubber (limited to 90 C, 194 F)


55

797.5

Standard Flanges

High Temperature Design

Stn. Stl. may be used for the entire range

Note: Max. measuring temperature for hard/soft rubber 90 C

Fig. 3

Fluid Temperature as a Function of the


Ambient Temperature

Druck / Pressure (PS) [bar]

50

725

High temperature range

300lb

45

652.5

40

580

35

507.5

30

435

25

362.5

20

290
150lb

15

217.5

10

145

72.5

0
-30

Max. Allowable Cleaning Temperatures


PTFE, PFA Designs

psi

130

25

-10

10

30

50

70

90

110

130

150

170

190

Temperatur / Temperature (TS) [C]


-22 F 14 F

50 F

86 F

122 F 158 F 194 F 230 F 266 F 302 F 338 F 374 F

0.01 bar = 1 kPa

Tmax C

tmax
Minutes

TAmb. C

Steam cleaning

PTFE, PFA,
PEEK, Torlon

150

60

25

Liquid cleaning

PTFE, PFA,
PEEK, Torlon

140

60

25

If the ambient temperature >25 C, the difference must be subtracted


from the max. cleaning temperature. Tmax C.
C = (TAmb. 25 C [13 F])
Weight
See Dimensions

Min. allowable Pressure and Fluid Temperature


Liner material
Hard rubber
Soft rubber
PTFE
PFA

Meter size
DN
15 1000
50 1000
10 600

inch
1/240
240
3/824

3 100

1/104

TOperation
mbar abs
0
0
270
500
0

at

TOperation
C
< 90
< 90
< 20
< 130
< 130

F
< 194
< 194
< 68
< 266
< 266

Fig. 5

ANSIFlanges SST No. 1.4571[316Ti] to DN 600 [24]

Liner: PTFE, hard/soft rubber (limited to 90 C, 194 F)


45

Druck / Pressure (PS) [bar]

Liner
Primary

CIPCleaning

High temperature range

40
PN 40

35
30
25
PN 25

20
15

PN 16

10

PN 10

5
0
-30

-10

10

30

50

70

90

110

130

150

170

190

Temperatur / Temperature (TS) [C]

Fig. 6

DINFlanges Steel to DN 600 [24]

Other meter sizes, pressure ratings and Temperature Classes upon re
quest

Electromagnetic Flowmeter FSM4000


D184S073U02

Specifications: Stainless Steel Flowmeter Primaries,


Model SE21_, DN 1 to DN 100 [1/25 to 4]

Liner: PTFE, hard/soft rubber (limited to 90 C, 194 F)


P/T-Ratings
797.5

Connection Overview Series 2000

725

362.5

20
15

217.5

10

145

72.5

0
-30

-10

10

30

50

70

90

110

130

150

0
190

170

Temperatur / Temperature (TS) [C]


50 F

-22 F 14 F

86 F

122 F 158 F 194 F 230 F 266 F 302 F 338 F 374 F

Meter Size

290

150lb

Wafer Design

435

25

Flange

30

Food Industry Fitting


DIN 11851

507.5

SE21
B

Weld Stubs
ISO 2037

35

Model:

Weld Stubs
DIN 2463

580

Weld Stubs
DIN 11850

40

TriClamp DIN 32676

652.5

External Threads

45

psi

Druck / Pressure (PS) [bar]

300lb

1/8Sanitary Connectors

Hochtemperaturbereich
High temperature range

50

Process Connections

55

0.01 bar = 1 kPa

ANSIFlanges Steel to DN 600 [24]

JIS 10K-B2210 Flanges SS No. 1.4571 [316Ti] or Steel


Meter Size
DN
Inch

Material

PN

TS [C]

PS [bar]

32100 11/4  4 SST 1.4571[316Ti]

10

25 to +180

10

32100 11/4  4

10

10 to +180

10

Steel

Liner: PTFE, hard/soft rubber


(limited to 90 C)
Liner: Hard/soft rubber 90 C (194 F)
17

DN 700 PN 16

15

1/25 x

PED

DN 2

1/12 x

DN 3

1/10

DN 4

5/32

DN 6

1/4

DN 8

5/16

DN 10

3/8

DN 15

1/2

DN 20

3/4

DN 25

DN 32 11/4

DN 40 11/2

DN 50

DN 65 21/2

DN 80

DN 100

14
13
12

DN 900 PN 16
DN 800 PN 16

11

DN 1000 PN 16

10

DN 900 PN10
DN 800 PN 10
DN 700 PN 10

DN 1000 PN 10

-30 -20 -10

10

20

30

40

50

60

70

80

90

Temperatur / Temperature (TS) [C]

Fig. 8

DINFlange SST No. 1.4571 [316Ti]


DN 700 [28]DN 1000 [40]

Material Strength Curves for Flanged Flowmeters Model SE21F


Liner: PFA

Liner: Hard/soft rubber 90 C (194 F)

45

17

Druck / Pressure (PS) [bar]

15
14
DN 700 PN 16

13
12
11

DN 900 PN 16
DN 800 PN 16

10

DN 1000 PN 16

9
8

Druck / Pressure (PS) [bar]

40

16

DN 900 PN 10
DN 800 PN 10
DN 700 PN 10

7
6
-10

10

20

30

40

50

60

70

80

PN 40

35

30
25

PN 25

20

15
10

PN 16

PN 10

5
0
30 20 10 0 10 20 30 40 50 60 70 80 90 100 110 120 130 *)

DN 1000 PN 10
90

Temperatur / Temperature (TS) [C]

Temperatur / Temperature (TS) [C]

Fig. 9

Conformity.
Evaluation per Category III
Module B1+D, Fluid Group 1

Druck / Pressure (PS) [bar]

16

DN 1

SEP
Art. 3,
Par. 3

Fig. 7

DINFlanges Steel DN 700 [28]DN 1000 [40]

Fig. 10

DINFlanges SS 1.4571 [316Ti] to DN 100, 4

Electromagnetic Flowmeter FSM4000


D184S073U02

Ambient Conditions

Liner: PFA
Druck / Pressure (PS) [bar]

55

Ambient Temperature

50

300lb

25 C to + 60 C [13 F to +140 F]

45

Fluid Temperature DN 3100

40
35

40 C to +130 C, CIPcapable, see Temperature Diagram and max.


allowable cleaning temperature.
The maximum allowable ambient temperature as a function of the
fluid temperature for stainless steel process connections and the
Wafer Design

30
25
20
150lb

15
10

Ambient Temperature C

60

0
30

10

10

30

50

70

90

110

130 *)

Temperatur / Temperature (TS) [C]

Fig. 11

ANSIFlanges SS 1.4571 [316Ti] to DN 100, 4

Material Strength Curve for Wafer Design Model SE21W

40

Liner: PFA

50

130 *)
Fluid
Temp. C

25

652.5

40

580
PN 40/CL 300

35

435

25

362.5
PN 25

10 C to +120 C [14 F to 248 F], CIPcapable to max. 120 C.

290

15

Temperature Diagram

Fluid Temperature DN 12 [1/251/12]

507.5

30

20

Fig. 13

psi

Druck / Pressure (PS) [bar]

P/TRatings
45

Storage Temperature
 25 C to +70 C [13 F to +158 F]

217.5
PN 16/CL 150

10

145
PN 10

0
60 50 40 30 20 10 0 10 20 30 40 50 60 70 80 90 100 110120130*)

72.5

Liner

Temperatur / Temperature (TS) [C]


40 22 4

Minimum Allowable Absolute Pressure

14 32 50 68 86 104 122 140 151 176 194 212 230 248 266 284 F

PFA
Peek/Torlon

Meter Size
DN
Inch
3100 1/104
12 1/251/12

POperate
mbar abs
0
0

at

TOperate
C
130*)
120

Maximal Allowable Fluid Temperature and Pressure


Fig. 12

Process Connections
Liner PFA
Wafer Design

JIS 10K-B2210 Wafer Design


Meter Size
DN

Material

PN

TS [C]

PS [bar]

32100
11/4  4

1.4404[316L]
1.4435[316L]
1.4301[304]

10
145 psi

40 to +130
40 to +266 F

10
145 psi

Weld stubs

TriClamp
Food Industry fittings

External threads
1/8Sanitary connectors
Liner
Peek/Torlon

Meter Size
DN
Inch
350
1/102
65100 21/24
340
1/101 1/2
50, 80 2, 3
65, 100 2 1/2, 4
350
1/102
65100 2 1/24
340
1/101 1/2
50, 80 2, 3
65, 100 2 1/2, 4
325
1/101

PS
bar
40
16
40
16
10
16
10
40
16
10
10

at

TS
C
130*)
130*)
130*)
130*)
130*)
121
121
130*)
130*)
130*)
130*)

12

10

120

1/251/12

Process Connections
DN 12 [1/251/12]
1/8Sanitary connectors with 2 grounding electrodes in the same
material as the signal electrodes, standard
DN 3100 [1/104]
*) Higher temperature are allowed for CIP/SIP cleaning for limited time
periods, see Table Maximum Allowable Cleaning Temperature

Wafer Design, TriClamp, Food Industry Fittings, External threads.


Others upon request

Electromagnetic Flowmeter FSM4000


D184S073U02

Maximal Allowable Cleaning Temperature


CIPCleaning
Steam
Liquid cleaning

Liner
PFA/Peek
PFA/Peek/Torlon

Tmax
C
150
140

Tmax
Minutes
60
60

Specification for painting


TAmb
C
25
25

If the ambient temperature >25 C, the difference must be subtracted


from the max. cleaning temperature.
Tmax  C, C = (TAmb. 25 C) [13 F].

Maximal Allowable Temperature Shock

Type of painting
Epoxy powder coating
Priming coat
No priming coating
Preparing of the surface
Cleaning in a high alkaline bath, Alodine procedure  conversion layer
for a better adherion of the painting
Thickness of painting

Liner
PFA, Peek, Torlon

Temp.Shock
max. Temp.Diff. C
arbitrary

Temp.Gradient
C/min
arbitrary

RAL 9002 (White grey)


RAL 7012 (Basalt grey)

Specifications Flowmeter Primary

Flowmeter Primary Materials


Liner
Materials
PFA,
Peek,
Torlon

Electrode Materials
Standard
Others
Hast.B2
Hast.C4
(1.4539 for SST 1.4539
SST 1.4571[316Ti]
Food Ind.
fittings and Tantalum, Titanium,
PlatinumIridium
TriClamp)

Process Connection
Materials
Flange
Wafer Design
Weld stubs
Food Ind. fittings
TriClamp
External threads
1/8Sanitary connectors
Connection Box
Meter tube
Cable connector
Primary housing

Standard

Flat housing gaskets

Electrode Design
Standard Others
Flat head Conical
head
( DN10
[3/8])
SST No.
1.4539
Option

SST 1.4571 [316Ti] upon req.


none
SST 1.4404 [316L] upon req.
SST 1.4404 [316L] upon req.
SST 1.4404 [316L] upon req.
SST 1.4404 [316L] upon req.
SST 1.4571 [316Ti] POM, Brass, PVC
Stainless Steel

SST 1.4301 [304]

Polyamide
PVDF
Deep drawn housing SST No. 1.4301[304]

Process Connection Material


Wafer Design,
Weld stubs,
Food Ind. fitting
TriClamp
External threads

Gasket Materials
none
EPDM (EthylenePropylene) std.
with FDAApproval (CIP capable, no oil
or grease)
Silicone with FDAApproval (option,
suitable for oil and grease)
Silicone (suitable for oil and grease)

Protection Class
IP 67 Standard
IP 68 with connector Pg 13.5 and PU resin compound
Pipeline Vibration
Maximum allowable 15 m/s2 (10150 Hz)

80 m
Color

Electromagnetic Flowmeter FSM4000


D184S073U02

Specifications: Flowmeter Primary, Flowrate Nomographs


Flowrate Nomograph
l/min m3/h

6
5
4
3
2

10-1
8
6
5
4
3
2

10-2
8
6
5
4

103
8
4

10
8
6
5
4

102
8

103
8
6
5
4

3
2

3
10-3
8

6
5
4

10-3

4 5 6
m/s

6
5
4
3

-2

0.5 0.6 0.8 1

Fig. 14

2
0

50
DN
40
DN
32
DN
8
DN
6
DN

DN

6
5
4

6
5
4

10-1
8

1
8

10
8

6
5
4

DN

10-1
8

6
5
4

6
5
4

DN

6
5
4

6
5
4

10
8

104
8

10

6
5
4

8
6
5
4
3
2

10

8
6
5
4

1
8

DN DN DN
15 20 25

6
5
4

10

1
8

1.

DN

6
5
4

6
5
4

10
8

102
8

DN

10
8

DN

Example

6
5
4

DN DN DN DN DN DN DN DN DN DN DN
25 30 35 40 45 50 60 70 80 90 10
0 0 0 0 0 0 0 0 0
0
0 0

10
8

20

6
5
4

DN

6
5
4

3
4

15

DN

102
8

l/s

12

6
5
4

6
5
4

l/min m3/h

DN

103
8

DN DN
80 10
0

65

DN

5
4

l/s

10
8

3
2

6
5
4

10

0.5 0.6 0.8 1

4 5 6
m/s

8 10

8 10

Flowrate Nomograph DN 1 to DN 100 [1/25 to 4]

Flowrate Nomographs
The volume flowrate is a function of the flow velocity and the
flowmeter size. The Flowrate Nomographs, Fig. 14 and Fig. 15
indicate the flowrate range for a specific flowmeter size and which
flowmeter sizes are suitable for a specific flowrate.

Fig. 15

Flowrate Nomograph DN 125 to DN 1000 [5 to 40]

Example:
Flowrate = 7 m3/h (Maximum value = range end value). Suitable are
flowmeter sizes DN 20 to DN 65 [3/4 to 2/12] for a flow velocity
between 0.5 and 10 m/s.

Electromagnetic Flowmeter FSM4000


D184S073U02

Up- and Downstream Straight Sections


Flowmeter installation is the key to an accurata flow measurement
system. As a rule of thump, generally 3 pipe diameters uystream and
2 pipe diameters downstream are sufficient. Other upstream
obstructions however, such as double elbows out of plane, control
valves and nonconcentric pipe reducers etc. may require longer
lengths.

Installations in Larger Size Pipelines


The flowmeter can be installed in larger size pipelines without
problems using reducer sections. The pressure drop when reducers
are used can be determined from the Nomograph Fig. 17. The
procedure for determining the pressure drop is as follows:
1. Determine the diameter ratio d/D.
2. Calculate the flow velocity as a function of the flowmeter size
and the flowrate:
Q (flowrrate)
v = --------------------------------------------Primary Constant

The flow velocity can also be determined from the Flowrate


Nomographs, Fig. 14 and Fig. 15.
In Fig. 17 read the Pressure drop value, yaxis, at the
intersection of the Diameter Ratio, xaxis, and the Flow
Velocity curve.

Grounding
3.

Flanged Reducer
8
D

Grounding the flowmeter primary is necesary not only for safety


reasons but also to assure proper opration of the electromagnetic
flowmeter. The grouding screws on the flowmeter primary, which
correspond to VDE 0100, Part 540, are to be connected to earth
potential. For measurement reasons this potential should be identical
to the fluid potential if at all possible.
In applications using plastic or insulating lined pipelines the ground
connections are made utilizing grouding plates or grounding
electrodes. If stray currents are present in the pipeline, grounding
plates up and downstream of the flowmeter primary are
recommended.

Specifications: Flowmeter Primary, Installation


Requirements and Grounding

Electrode Axis
Neither electrode should be positioned at the highest pont in the
pipeline in horizontal installations. Gas bubbles, which may be present
in the fluid, could interrupt the electrical connection between the
electrodes and the fluid. An ideal installation is shown in Fig. 16.
Assure that the meter tube is always completely filled with fluid.
Instrument designs with removable electrodes for cleaning are
available upon request.

d
D
v
p

= EMF inside diameter


= Pipe inside diameter
= Flow velocity in m/s
= Pressure drop in mbar

Pressure Drop Nomograph for EMF


Flanged reducers with /2 = 8
100

Electrode Axis

v=8m/s
7m/s

Pressure Drop p [mbar]

6m/s
5m/s

10
4m/s

3m/s

2m/s

Fig. 16

Electrode Axis

0.5

0.6

0.7

Diameter Ratio d/D


Fig. 17

10

Pressure Drop Nomograph

0.8

0.9

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Flowmeter Primary DN 3 to DN 300 [1/1012], DIN Flanges


Model SE41F
Cable Connector M20 x 1.5

Cable Connector
M20 x 1,5

88

Number of Holes N

Number of
Holes N

d4
D

d2

Tolerances L*
b

k
H

L*

Flange Dimensions

DN 3200
1/108

+0

DN 250300
1012

+0

3
5

Instrument Dimensions

Weight

DN

PN1)

D/H

d4

d2

L*

L2)

L3)

E4)

G4)

kg

38
10
15
20

1040
1040
1040
1040

90
90
95
105

40
40
45
58

18
18
18
20

60
60
65
75

14
14
14
14

4
4
4
4

67
67
67
87

130
200
200
200

133
203
203
203

136
206
206
206

101
101
101
112

62
62
62
73

129
129
129
140

2.5
3
3
3.5

5.51
6.61
6.61
7.72

25
32
40

1040
1040
1040

115
140
150

68
78
88

20
20
20

85
100
110

14
18
18

4
4
4

87
95
100

200
200
200

203
203
203

206
206
206

112
117
121

73
78
82

140
145
149

4
5
6

8.82
11.02
13.23

50

1040

165

102

21

125

18

116

200

203

206

129

90

157

17.64

65

1016
2540

185
185

122
122

21
25

145
145

18
18

4
8

100
100

200
200

203
203

206
206

153
153

104
104

171
171

12
12

26.45
26.45

lb (pound)

80

1040

200

138

27

160

18

100

200

203

206

159

110

177

16

35.27

100

1016
2540

220
235

158
162

23
27

180
190

18
22

8
8

130
130

250
250

253
253

256
256

179
179

130
130

197
197

15
15

33.07
33.07

125

1016
2540

250
270

188
188

25
29

210
220

18
26

8
8

124
124

250
250

255
255

260
260

170
170

127
127

195
195

23
30

50.71
66.14

150

1016
2540

285
300

212
218

25
31

240
250

22
26

8
8

170
170

300
300

305
305

310
310

190
190

148
148

216
216

28
34

61.73
74.96

200

10
16

340
340

268
268

28
28

295
295

22
22

8
12

195
195

350
350

355
355

360
360

222
222

179
179

248
248

54
54

119.05
119.05

250

10
16

395
405

320
320

30
30

350
355

22
26

12
12

250
250

450
450

455
455

460
460

250
250

207
207

276
275

80
80

176.37
176.37

300

10
445 370 31 400
22
12
250
500
505
510
293
250
318
83
182.98
16
460 378 33 410
26
12
250
500
505
510
293
250
318
92
202.82
1) Other pressure ratings upon request.
2) Standard with one grounding plate mounted on one flange SST No. 1.4571 [316Ti]. Other materials upon request.
See also Information Grounding. DN 3100 L + 3 mm, DN 125300 L + 5 mm.
3) With protection flanges. Protection flanges replace grounding plates. DN 3100 L + 6 mm, DN 125300 L + 10 mm.
Series 1000 (for replacement only) also available with protection flanges, Dimensions see page 16.
4) For high temperature design + 120 mm.

All dims in mm
Fig. 18

Projektion nach ISO Methode E

Flowmeter Primaries DN 3 to DN 300 [1/10: to 12]

11

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Flowmeter Primary DN 3 to DN 300 [1/1012], ANSI Flanges


Model SE41F
Cable Connector M20 x 1.5

88

88
35

Cable Connector
M20 x 1.5

Number of Holes N
d2

d4
D

d2
Number of
Holes N

k
H

L*

Tolerances L*
3

DN 250300
1012

+0

Weight

Flange Dimensions
ANSI CL 300

Flange Dimensions
ANSI CL 150

Meter Instrument Dimensions ANSI CL 150/CL 300


Size

+0

DN 3200
1/108

L1)2) *
DN

Inch

38 1/105/16
10, 15 3/8, 1/2
3/4
20
1
25

ABB
ISO
13359 (Ser.1000)

E3)

D/H

129 89
129 89
140 99
140 108

67
67
87
87

130
200
200
200

130
270
270
270

101
101
112
112

62
62
73
73

78 145 117
82 149 127
90 157 152

d4

d2 N D/H

kg

lb

CL150/300

(pound)

d2 N

35
35
43
51

16
16
19
19

4
4
4
4

2.5
3
3.5
4

5.51
6.61
7.72
8.82

64 18 89 16
73 20 99 16
92 21 121 19

4 134
4 156
4 165

64 21 99 19
73 23 115 23
92 24 127 19

4
4
8

5
6
8

11.02
13.23
17.64

4 191 105 27 150 23


4 210 127 30 168 23
8 254 157 34 200 22

8
8
8

12
16
15

26.45
35.27
33.07

36/38
38
66
98

79/84
83.77
145.50
216.05

124/181

273/399

60
60
70
80

95
100
116

200
200
200

280
280
280

117
121
129

65
80
100

2 1/2
3
4

100
100
130

200
200
250

330
340
400

153 104 171 178 105 25 140 19


159 110 177 191 127 26 153 19
179 130 197 229 157 26 191 19

125
150
200
250

5
6
8
10

124
170
195
250

250
300
350
450

450
450
500
550

169
190
222
250

254
279
343
406

4 95
4 95
4 117
4 124

13
13
14
16

1 1/4
1 1/2
2

195
216
248
276

16
16
16
16

35
35
43
51

32
40
50

127
148
179
207

d4

186
216
270
324

28
30
34
35

216
241
298
362

22 8
22 8
22 8
25 12

280
318
381
445

186
216
270
324

16
16
18
20

39
41
47
53

67
67
83
89

235
270
330
387

22 8
22 12
25 12
28 16

293 250 319 483 381 37 432 26 12 521 381 56 451 32 16


620
250 500
12
300
1) Standard with one grounding plate mounted on one flange SST No. 1.4571[316Ti]. Other materials upon request.
See also Information Grounding. DN 3100 L + 3 mm, DN 125300 L + 5 mm
2) With protection flanges. Protection flanges replace grounding plates. DN 3100 L + 6 mm, DN 125300 L + 10 mm
Series 1000 ( for replacement only) also available with protection flanges, Dimensions see page 16.
3) For high temperature design + 120 mm.

All dims in mm
Fig. 19

12

Flowmeter Primaries DN 3 to DN 300 [1/10: to 12], ANSI Flanges

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Flowmeter Primary DN 350 to DN 1000 [1440], DIN Flanges


Model SE41F

Cable Connector M20 x 1.5

Number of
Holes N

d4

d2

d4
k
D

A
L*
Tolerances L*

DIN Flange Dimensions

DN 350500
1420

DN 6001000
2440

10

Instrument Dimensions

Weight

DN

PN1)

d4

d2

L*

L2)

L3)

350
350

10
16

505
520

460
470

430
438

22
26

16
16

31
35

322
322

550
550

555
555

560
560

341
341

313
313

250
250

126
140

278
309

400
400

10
16

565
580

515
525

482
490

26
30

16
16

31
37

370
370

600
600

605
605

610
610

367
367

339
339

275
275

155
175

342
386

500
500

10
16

670
715

620
650

585
610

26
33

20
20

33
39

407
407

650
650

655
655

660
660

403
403

375
375

310
310

191
235

421
518

600
600

10
16

780
840

725
770

685
725

30
36

20
20

33
41

469
469

780
780

785
785

790
790

454
454

426
426

361
361

243
302

536
666

700
700

10
16

895
910

840
840

800
795

30
36

24
24

35
41

537
537

910
910

915
915

920
920

495
495

469
469

405
405

315
340

694
750

800
800

10
16

1015
1025

950
950

905
900

33
39

24
24

37
43

605
605

1040
1040

1045
1045

1050
1050

545
545

519
519

455
455

405
467

893
1030

900
900

10
16

1115
1125

1050
1050

1005
1000

33
39

28
28

39
45

671
671

1170
1170

1175
1175

1180
1180

595
595

569
569

505
505

483
567

1065
1250

619
619
619

555
555
555

475
575
764

1047
1268
1684

645
739 1300 1305 1310
31
1175 1120 1080 30 28
6
1000
645
739 1300 1305 1310
39
1230 1160 1110 36 28
10
1000
645
739 1300 1305 1310
47
1255 1170 1115 42 28
16
1000
1) Other pressure ratings upon request.
2) One grounding plate mounted on one flange SST No. 1.4571[316Ti].
Other materials upon request. L + 5 mm.
3) With protection plates mounted on both flanges, grounding plate not required L + 10 mm.
Series 1000 (for replacement only) also available with protection flanges, Dimensions see page 16.

appr. kg lb (pound)

All dims in mm
Fig. 20

Projektion nach ISO Methode E

Flowmeter Primaries DN 350 to DN 1000 [14 to 40], DIN Flanges

13

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Flowmeter Primary DN 350 to DN 1000 [1440], ANSI Flanges


Model SE41F

Cable Connector M20 x 1.5

Number of
Holes N

d4

d2

d4
k
D

A
L*
Tolerances L*

Meter Size

DN 350 500
1420

DN 600 1000
2440

10

Instrument Dimensions

0
0

Flange Dimensions
ANSI CL 150 1)

Weight

L2)3) *
E

d4

d2

appr. kg
CL150

appr. lb
(pound)

650
700

780

313
329
375
375

250
275
310
310

341
367
403
403

534
597
635
699

476
540
578
635

413
470
533
584

28.6
28.6
31.7
31.7

12
16
16
20

40
42
45
48

178
225
230
263

392
496
507
580

850
910

423
469

361
405

454
495

813
837

749
749

692
762

34.9
34.9

20
20

53
50

327
385

721
849

48
22.2
864
900
942
545
455
519
1040

605
800 32
44
25.4
972
595 1057 1010
505
569
1170

671
900 36
28
56
1140
645 1380 1210
555
619
1300

739
1000 40
1) Other pressure ratings upon request.
2) One grounding plate mounted on one flange SST No. 1.4571[316Ti].
Other materials upon request. L + 5 mm.
3) With protection plates mounted on both flanges, grounding plate not required L + 10 mm.
Series 1000 (for replacement only) also available with protection flanges, Dimensions see page 16.

51
57
63

513
662
1132

1131
1459
2496

DN

Inch

ISO
ABB
13359 (ser.1000)

350
400
450
500

14
16
18
20

322
370
407
416

550
600
686
762

600
700

24
28

469
537

914

All dims in mm

Fig. 21

14

Flowmeter Primaries DN 350 to DN 1000[14 to 40], ANSI Flanges

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering information: Flanged Instruments, Model SE41F


In addition to the Ordering Number please supply the following information: Fluid, fluid temperature, operating pressure, flow range, pipeline type
(grounding plate, grounding electrodes)1)
Ordering Number
SE41F
Liner Material
H
Hard rubber (DN15DN1000 [1/240]) KTW approved
S
Soft rubber (DN50DN1000 [240])
T
PTFE
(DN10DN600 [3/824]) FDA approved
P
PFA
(DN3DN 8 [1/105/16]) FDA approved
Z
Others
Meter Size Inch
03
DN
3
1/10
04
DN
4
5/32
06
DN
6
1/4
08
DN
8
5/16
10
DN
10
3/8
15
DN
15
1/2
20
DN
20
3/4
25
DN
25
1
32
DN
32
11/4
40
DN
40
11/2
50
DN
50
2
65
DN
65
21/2
80
DN
80
3
1H
DN 100
4
1Q
DN 125
5
1F
DN 150
6
2H
DN 200
8
2F
DN 250
10
3H
DN 300
12
3F
DN 350
14
DN 400
16
4H
DN 450
18 (only ANSI Cl150/300 ISO 13359 Installation length)
4F
5H
DN 500
20
6H
DN 600
24
7H
DN 700
28
8H
DN 800
32
9H
DN 900
36
1T
DN 1000
40
Measuring
Grounding Electrode Material 2)
SST No. 1.4571[316Ti] none
S
Hastelloy B2
none
B
Hastelloy C4
none standard
H
Titanium
none
M
Tantalum
none
T
SST No. 1.45393)
F
none
P
PlatinumIridium
none
SST No. 1.4571[316Ti] with
E
Hastelloy B2
with
N
Hastelloy C4
with
O
Titanium
with
I
Tantalum
with
Q
SST No. 1.45393)
R
with
G
PlatinumIridium
with
Z
Others
Pressure Rating
C
DIN PN 10
D
DIN PN 16
E
DIN PN 25
F
DIN PN 40
K
JIS K10
P
ANSI CL 150 (Installation length Series 1000 only for replacement)
Q
ANSI CL 300 (Installation length Series 1000 only for replacement)
R
ANSI CL 150 (ISO 13359 Installation length)
S
ANSI CL 300 (ISO 13359 Installation length)
Z
Others
Continued on next page
1) If a grounding plate is required (see Part No.) The standard grounding plate is made of SST No. 1.4571[316Ti]. Others upon request.
2) From DN 125 [5] a conductive grounding surface is integrated in the hard/soft rubber liners, standard.
3) For Food Industry applications.

15

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Number
Flange Material
Steel (standard from DN 20[3/4])
1
SS No. 1.4571[316Ti] (standard for DN3DN15[1/101/2], option from DN 20[3/4]) 3
Others upon request
9
Flange Accessories
None
Protection plates SST 1.4571[316Ti] (mounted one side)
Only PTFE/PFA
Grounding plate SST 1.4571[316Ti] (mounted one side)
Only PTFE/PFA
Protection flanges SST 1.4571[316Ti] (mounted both sides)4) Only PTFE
Temperature Range
Standard (see Specifications, Liner Materials) 130 C
High temperature design 180 C (only PTFE/PFA)
Certifications
None
Material certified per EN 10204 Par. 3.1B and pressure test per AD2000
Inspection Certificate EN 102043.1B
Pressure test per AD2000
Others
Calibration Certificates
Standard
Others
Protection Class per EN 60529
IP 67 (connection box with M20 x 1.5 cable connector)
IP 68 (connection box with M20 x 1.5 cable connector) and PU resin compound
IP 67 (connection box with NPTthreads)
Others
Design5)
Without preamplifier (Standard DN 10 [3/8] to DN 1000 [4])
Preamplifier Typ A (Standard DN 3 [1/10] to DN 8 [5/16])
Preamplifier Typ B
4) Only with Installation Length Series 1000 Short and ANSI/DINFlanges.
Installation length DN 80[3] L + 20 mm, from DN 100[4] L + 25 mm.
5) Notes on design
Design

A
B
C
D
S
H
A
D
F
G
Z
A
Z
2
3
4
9

Minimum Conductivity
DN 3 [1/10] to DN 8 [5/16]

DN 10 [3/8] to DN 1000 [4]

without preamplifier

not possible

= > 20 S/cm

Preamplifier Typ A

= > 20 S/cm

= > 5 S/cm

Preamplifier Typ B
= > 5 S/cm
Max. cable lenght without preamplifier 50 m
Max. cable lenght with preamplifier 200 m (Empty pipe detection not pos
sible)

Additional Ordering Information may be provided in


writing:
Name Plate (also available in Stainless Steel, option)
German
English
French
Electrode Design
Standard
Conical head (from DN 10 [3/8] material SST No. 1.4539 used for
high grease content fluids)

16

= > 0,5 S/cm

1
2
4

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Stainless Steel Flowmeter Primaries DN 3 to DN 40 [1/10 to 1-1/4], Flanged


Model SE21F
103

83
Cable Connector
Pg 13.5

41

Number of Holes N

d4

d2

k
D

DIN Flange Dimensions


Liner PFA
Connections

DN

1/105/16
3/8, 1/2
3/4
1
11/4
11/2

38
10,15
20
25
32
40

PN

L1)

d4

d2

1040

130
200
200
200
200
200

37
37
42
54
62
67

90
95
105
115
140
150

60
65
75
85
100
110

42
36
41
54
64
74

14
14
14
14
14
14

18
18
20
20
20
20

108
108
120
127
132
136

62
62
66
73
78
82

134
134
136
143
148
152

L1)

130
200
200
200
200
200
130
200
200
200
200
200

37
37
42
54
62
67
37
37
42
54
62
67

Weight
appr. kg
appr. lb
(pound)
3.5
7.72
3.5
7.72
3.5
7.72
4
8.82
5
11.02
5.5
12.13

ANSI Flange Dimensions


Liner PFA
Connections

DN

1/105/16
3/8, 1/2
3/4
1
11/4
11/2
1/105/16
3/8, 1/2
3/4
1
11/4
11/2

38
10,15
20
25
32
40
38
10,15
20
25
32
40

PN

CL 150

CL 300

88.9
60.3
88.9
60.3
98.4
69.8
108
79.2
117.5 88.9
127
98.6
95.2
66.7
95.2
66.7
117.5 82.5
123.8 88.9
133.3 98.4
155.6 114.3

d4

d2

42
34.8
42.9
50.8
63.5
73.0
42
34.8
42.9
50.8
63.5
73.0

15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
19
19
19
22.2

18
12.6
14.2
15.8
17.4
19
18
15.8
17.4
19.0
20.5
22.1

108
108
120
127
132
136
108
108
120
127
132
136

62
62
66
73
78
82
62
62
66
73
78
82

134
134
136
143
148
152
134
134
136
143
148
152

Weight
appr. kg
appr. lb
(pound)
3.5
7.72
3.5
7.72
3.5
7.72
4
8.82
5
11.02
5.5
12.13
3.5
7.72
3.5
7.72
3.5
7.72
4
8.82
5
11.02
5.5
12.13

1) If a grounding plate is required, L + 3 mm, materials upon request.


With 2 grounding plates (protection plates) L + 6 mm, materials upon request.

All dims in mm
Fig. 22

Projektion nach ISO Methode E

Flowmeter Primaries DN 3 to DN 40 [1/10 to 11/2]

17

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Stainless Steel Flowmeter Primaries DN 50 to DN 100 [2 to 4], Flanged


Model SE21F
83
Cable Connector Pg 13.5

41

d4

d2

Number of Holes N

A
b
L

DIN Flange Dimensions


Liner PFA

Weight

DN

PN

L1)

d4

d2

appr. kg

appr. lb (pound)

50

1040

200

100

165

125

104

18

24

135

161

17.64

65

1016
2540

200
200

107
107

185
185

145
145

124
124

18
18

4
8

26
26

149
149

175
175

10
10

22.05
22.05

80

1040

200

107

200

160

139

18

28

155

181

12

26.45

100

1016
2540

250
250

159
159

220
235

180
190

161
167

18
22

8
8

24
28

175
175

201
201

18
18

39.68
39.68

L1)

d4

d2

appr. kg

appr. lb (pound)

200

100

152

121

99

19

23

135

163

17.64

21/2

200

107

178

140

118

19

26

149

177

10

22.05

200

107

191

152

131

19

28

155

183

12

26.45

ANSI Flange Dimensions


Liner PFA
Meter Size

Weight
PN

CL 150

250

159

229

190

171

19

28

175

203

18

39.68

200

100

165

127

102

19

26

135

163

17.64

200

107

191

149

124

22

29

149

177

10

22.05

200

107

210

168

143

22

32

155

183

12

26.45

250

159

254

200

177

22

36

175

203

18

39.68

21/2
3
4

CL 300

1) If a grounding plate is required, L + 3 mm, materials upon request.


With 2 grounding plates (protection plates) L + 6 mm, materials upon request.

All dims in mm
Fig. 23

18

Flowmeter Primaries DN 50 to DN 100

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Information: Stainless Steel Flowmeter Primaries DN 3 to DN 100 [1/10 to 4], Flanged
In addition to the Ordering Number please supply the following information: Fluid, fluid temperature, operating pressure, flow range,
pipeline type (grounding plate)1)
Ordering Number
SE21F
Liner Material: PFA
P
Meter Size
Inch
03
DN
3
1/10
04
DN
4
5/32
06
DN
6
1/4
08
DN
8
5/16
10
DN 10
3/8
15
DN 15
1/2
20
DN 20
3/4
25
DN 25
1
32
DN 32
11/4
40
DN 40
11/2
50
DN 50
2
65
DN 65
21/2
80
DN 80
3
1H
DN 100
4
Signal Electrode
/ Grounding Electrode Material
S
SST No. 1.4571[316Ti] / none
B
Hastelloy B2
/ none
H
Hastelloy C4
/ none
M
Titanium
/ none
T
Tantalum
/ none
F
SST No. 1.4539
/ none (Food Ind. applications)
P
PlatinumIridium
/ none
E
SST No. 1.4571[316Ti] / with
N
Hastelloy Bs
/ with
O
Hastelloy C4
/ with
I
Titanium
/ with
Q
Tantalum
/ with
R
SST No. 1.4539
/ with (Food Ind. applications)
G
PlatinumIridium
/ with
Z
Others
Pressure Rating
PN 16 Standard DN 100 [4]
D
PN 40 Standard DN 380 [1/103]
F
K
JIS K10
P
ANSI CL 150 (ISO 13359 Installation length)
Q
ANSI CL 300 (ISO 13359 Installation length)
Z
Others
Process Connection Materials
SST No. 1.4571[316Ti]
3
Others
9
Flange Accessories1)
A
None
B
Protection Plates SST No. 1.4571 (mounted both sides)
C
Grounding Plate SST No. 1.4571 (mounted one side)
Z
Others
Temperature Range
Standard design 130 C (266 F)
S
Continued on next page
1) Protection plates (2 grounding plates) mounted on both sides and grounding plate mounted on one side to flange,
material SST No. 1.4571 [316Ti].
Others upon request.

19

Electromagnetic Flowmeter FSM4000


D184S073U02
In addition to the Ordering Number please supply the following information: Fluid, fluid temperature, operating pressure, flow range,
pipeline type (grounding plate))1)
Ordering Number
SE21F
Certifications
A
None
D
Material certified per EN 10204 Par. 3.1B and pressure test per AD2000
F
Inspection Certificate EN 10204  3.1B
G
Pressure test per AD2000
Z
Others
Calibration Certificates
Standard
A
Others
Z
Protection Class
IP 67 (connection box with Pg 13.5 cable connector)
2
IP 68 (connection box with Pg 13.5 cable connector) and PU resin compound
3
IP 67 (connection box with NPTAdapter)
4
Design1)
Without preamplifier (Standard DN 10 [3/8] to DN 1000 [4])
1
Preamplifier Typ A (Standard DN 3 [1/10] to DN 8 [5/16])
2
Preamplifier Typ B
4
1) Notes on design
Design

Minimum Conductivity
DN 3 [1/10] to DN 8 [5/16]

DN 10 [3/8] to DN 1000 [4]

without preamplifier

not possible

= > 20 S/cm

Preamplifier Typ A

= > 20 S/cm

= > 5 S/cm

Preamplifier Typ B
= > 5 S/cm
= > 0,5 S/cm
Max. cable lenght without preamplifier 50 m
Max. cable lenght with preamplifier 200 m (Empty pipe detection not possible)

Additional Ordering Information may be provided in


writing:
Name Plate (also available in Stainless Steel, option)
German
English
French
Electrode Design
Standard
Conical head (from DN 10 [3/8] material SST No. 1.4539 used for
high grease content fluids)

20

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Stainless Steel Flowmeter Primaries, DN 3 to DN 100 [1/10 to 4], Wafer Design
Model SE21W

83
41

Cable Connector Pg 13.5

Di

DN 3DN 8 [1/105/16]
A

Di
13.6

C#

B
A*
L

Meter Size
DN
Inch

PN

L1)

A*

3
1/10
4
5/32
68
64
37
28
6
1/4
1040
8
5/16
CL 150/
10
3/8
CL 300
15
1/2
JIS
20
3/4
78
74
42
28
25
1
90
86
54
46
32
11/4
98
94
62
46
40
11/2
103
99
67
46
50
2
117
112

60
103
99

60
65
21/2 1016
CL 150 103
80
3
99

60
JIS
100
4
133
129

60
1) Installation length with 2 grounding plates L + 3 mm
# Mounting bracket optional

50

42

50
70
70
70
110
110
110
110

50
59
69
77
95
111
128
155

Di
3
4
6
8
10
15
20
25
32
40
47
62
74
96

Weight
appr. kg

Weight
appr. lb
(pound)

45

108

62

197

1.5

3.31

54
63
73
82
100
116
133
160

112
119
124
128
136
150
157
176

66
73
78
82
50
58
66.5
80

205
219
229
237
213
235
250
283

2.0
2.0
2.5
3.0
4.0
4.5
6.5
8.5

4.41
4.41
5.51
6.61
8.82
9.92
14.33
18.74

All dims in mm
Fig. 24

Projektion nach ISO Methode E

Flowmeter Primaries DN 3 to DN 100 [1/10 to 4], Wafer Design

21

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Flowmeter Primary, Variable Process Connections, DN 3 to DN 100 [1/10 to 4]


Model SE21_
DN 3DN 40
[1/1011/2]

Cable Connector
Pg 13.5

103

83

83

3
A

L Rohr

DN 50DN 100
[24]

103
Cable Connector Pg 13.5

83

83

Number of Holes N

C#

B
A

2 1/24
DN 65100

Meter Size
DN
Inch

F2)

G2)

Weight
appr. kg1)

Weight
appr. lb1) (pound)

310

1/103/8

37

44

28

50

108

62

197

1.5

3.31

15

1/2

37

44

28

50

108

62

197

1.5

3.31

20

3/4

42

63

28

50

112

66

205

2.0

4.41

25

54

63

46

70

119

73

219

2.0

4.41

32

11/4

62

78

46

70

124

78

229

2.5

5.51

40

11/2

67

78

46

70

128

82

237

3.0

6.61

50

128

100

60

110

136

50

213

4.0

8.82

65

21/2

114

116

60

110

150

58

235

4.5

9.92

80

114

133

60

110

157

66.5

250

6.5

14.33

283

9.0

19.84

100
4
144
160
60
110
176
80
For installation lengths with process connections, see Pages 2324.
1) For additional weight due to process connections, see Pages 2324.
2) With mounting bracket G and F + 10.5 mm
# Mounting bracket optional

Fig. 25

22

2
DN 50

Flowmeter Primaries DN 3 to DN 100 [1/10 to 4], Variable Process Connections

All dims in mm

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Stainless Steel Flowmeter Primaries, Adapters for Various Process Connections
Model SE21_
Weld Stubs

Meter Size
DN

ISO 2037

Inch
Di

Da

DIN 11850
Weight1)

Di

Da

DIN 2463
Weight1)

Di

Da

Weight1)

kg

lb

kg

lb

kg

lb

310

1/103/8

10.0

13.0

127

0.4

0.88

10.3

13.5

127

0.4

0.88

15

1/2

16.0

19.0

127

0.4

0.88

18.1

21.3

127

0.4

0.88

20

3/4

20.0

23.0

132

0.7

1.54

23.7

26.9

132

0.7

1.54

25

22.6

25.0

149

0.7

1.54

26.0

29.0

149

0.7

1.54

25

28

149

0.7

1.54

32

11/4

31.3

33.7

166

1.0

2.20

32.0

34.0

166

1.0

2.20

32

35

166

1.0

2.20

40

11/2

35.6

38.0

171

1.0

2.20

38.0

41.0

171

1.0

2.20

36.8

40

171

1.0

2.20

50

48.6

51.0

173

1.0

2.20

50.0

54.0

173

1.0

2.20

49

52

173

1.0

2.20

65

21/2

60.3

63.5

165

1.4

3.09

66.0

70.0

165

1.4

3.09

66

70

165

1.4

3.09

80

72.9

76.1

169

2.0

4.41

81.0

85.0

169

2.0

4.41

81

85

169

2.0

4.41

100

97.6

101.6

199

2.6

5.73

100.0

104.0

199

2.6

5.73

100

104

227

3.0

6.61

Food Industry Fittings

Meter Size
DN

DIN 11851

Inch
Rd. Thds.

Tri-Clamp

DIN11864-1 (Form B)
Weight1)

kg

lb

Rd. Thds.

DIN 32676

Weight1)
kg

lb

Di

Weight1)
kg

lb

28 x 1/8

169

0.5

1.10

34 x 1/8

161

0.5

1.10

10.0

163

0.5

1.10

15

1/2

34 x 1/8

169

0.5

1.10

44 x 1/6

161

0.5

1.10

16.0

163

0.5

1.10

20

3/4

44 x 1/6

180

0.9

1.98

44 x 1/6

170

0.9

1.98

20.0

168

0.7

1.54

25

52 x 1/6

207

0.9

1.98

52 x 1/6

197

0.9

1.98

26.0

192

0.8

1.76

32

11/4

58 x 1/6

230

1.4

3.09

58 x 1/6

220

1.4

3.09

32.0

209

1.5

3.31

40

11/2

65 x 1/6

237

1.4

3.09

65 x 1/6

227

1.4

3.09

38.0

214

1.4

3.09

50

78 x 1/6

243

1.4

3.09

78 x 1/6

233

1.4

3.09

50.0

216

1.2

2.65

65

21/2

95 x 1/6

245

2.2

4.85

95 x 1/6

233

2.2

4.85

66.0

221

1.6

3.53

80

110 x 1/4

259

3.2

7.05

110 x 1/4

245

3.2

7.05

81.0

225

2.4

5.29

130 x 1/4

307

4.4

9.70

130 x 1/4

291

4.4

9.70

100.0

255

3.1

6.83

Weld Stubs per


DIN 11850 or ISO 2037 or DIN 2463

Di

Di

100
4
1) Weight per pair

RdWgt.

1/103/8

Da

310

TriClamp per
DIN 32676

Food Industry Fitting


per DIN 11851

All dims in mm
Fig. 26

Projektion nach ISO Methode E

Dimensions DN 3 to DN 100 [1/10 to 4], Adapters for Variable Process Connections

23

Electromagnetic Flowmeter FSM4000


D184S073U02

Model DS21

DN
310
15
20
25
32
40
50
65
80
100

Inch
1/103/8
1/2
3/4
1
11/4
11/2
2
21/2
3
4

Weld Stubs
DIN 11850
L
127
127
132
149
166
171
173
165
169
199

Weight1)
kg
lb
0.4
0.4
0.7
0.7
1.0
1.0
1.0
1.4
2.0
2.6

0.88
0.88
1.54
1.54
2.20
2.20
2.20
3.08
4.41
5.73

Food Industry Fittings


Weight1)
DIN 11851
kg
lb
L
169
169
180
207
230
237
243
245
259
307

0.5
0.5
0.9
0.9
1.4
1.4
1.4
2.2
3.2
4.4

Di

Meter Size

1.10
1.10
1.98
1.98
3.08
3.08
3.08
4.85
7.05
9.70

Weld Stubs

L
DIN 11850, ISO 2037

Food
Industry
Fitting

L
DIN 11851

DN
Inch
310 1/103/8
15
1/2
20
3/4
25
1
1) Weight per pair

R
3/8
1/2
3/4
1

External Threads ISO 228 / DIN 2999


Weight 1)
a
L
kg
lb
18
139
0.4
0.88
18
139
0.4
0.88
25
164
0.8
1.76
25
179
0.8
1.76

a
External
Threads

Meter Size

All dims in mm
Fig. 27

24

Dimensions, Model SE21, DN 3 to DN 100 [1/10 to 4], Adapters for Variable Process Connections

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Stainless Steel Flowmeter Primaries, 1/8-Sanitary Connectors, DN 1 to DN 2 [1/25 to 1/12]


Model SE21B

64

83
41

92

118

Cable Connector Pg 13.5

48

10

4 x 5.5

26
60

40

90

25

(135)

69

Note:
1

Connection dimension for G 1/8 coupling, internal threads

Mounting bracket optional

Information:
Grounding electrodes are integrated in the flowmeter primary as standard
and are made of the same material as the signal electrodes.

Process Connections
1/8 Sanitary connection1)
1) For hose 6 x 4 mm

DN

Inch

PN

12

1/25  1/12

10

Weight
kg

lb

1.5

3.31

All dims in mm
Fig. 28

Projektion nach ISO Methode E

Flowmeter Primaries DN 1 to DN 2 [1/25 to 1/12]

25

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Information: Stainless Steel Flowmeter Primaries


In addition to the Ordering Number please supply the following information: Fluid, fluid temperature, operating pressure, flow range,
pipeline type (grounding plate, grounding electrode)1)
Ordering Number
SE21
Process Connections
W
Wafer Design
DN 3100 [1/104]
R
Weld stubs DIN 11850
DN 3100 [1/104]
Q
Weld stubs DIN 2463
DN 3100 [1/104]
P
Weld stubs ISO 2037
DN 25100 [14]
S
Food Industry fitting DIN 11851
DN 3100 [1/104]
T
TriClamp DIN 32676
DN 3100 [1/104]
E
External threads ISO 228/DIN 2999 DN 3 25 [1/01]
V
No adapter
B
1/8Sanitary connectors
DN 12 [1/253/32]
Z
Others
Liner Material
PFA (DN 3  100 [1/104])
P
Peek (DN 12 [1/251/12])
K
Torlon (DN 12 [1/251/12])
N
Meter Size
Inch
01
DN
1
1/25
1S
DN 1.5
1/17
02
DN
2
1/12
03
DN
3
1/10
04
DN
4
5/32
06
DN
6
1/4
08
DN
8
5/16
10
DN 10
3/8
15
DN 15
1/2
20
DN 20
3/4
25
DN 25
1
32
DN 32
11/4
40
DN 40
11/2
50
DN 50
2
65
DN 65
21/2
80
DN 80
3
1H
DN 100
4
Measuring Electrode /Grounding Electrode Material2)
S
SST No. 1.4571[316Ti] /none
B
Hastelloy B2
/none
H
Hastelloy C4
/none standard
M
Titanium
/none
T
Tantalum
/none
F
SST No. 1.4539
/none (Food Industry applications)
P
PlatinumIridium
/none
E
SST No. 1.4571[316Ti] /with
N
Hastelloy B2
/with
O
Hastelloy C4
/with Standard
I
Titanium
/with
Q
Tantalum
/with
R
SST No. 1.4539
/with (Food Industry applications)
G
PlatinumIridium
/with
Z
Others
Pressure Rating
PN 10
TriClamp (DN 65100); external threads, Food Ind. fitting, weld stubs (DN 65, 100)
PN 16
Wafer Design, TriClamp DN (350); Food Ind. fitting, weld stubs DN 50, 80)
PN 40
Wafer Design (DN 350); Food Ind. fitting, weld stubs (DN 340)
JIS K10
Wafer Design
DN 3100 [1/104]
ANSI CL 150
Wafer Design
DN 3100 [1/104]
ANSI CL 300
Wafer Design
DN 350 [1/102]
Others
Continued on next page
1) Only required for Wafer Design and pipelines with insulating liners.
2) DN 12 [1/251/12] always order with same material as signal electrodes, standard.

26

C
D
F
K
P
Q
Z

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Number
SE21
Process Connection Materials
None
(only Wafer Design)
SST No. 1.4571 [316Ti] (only DN 12 [1/251/12])
SST No. 1.4404 [316L] standard
POM
(DN 12 [1/251/12])
Others
Instrument Mounting Bracket
None
With
Others
Temperature Range
Standard design 130 C (DN 12 [1/251/12] max. 120 C)
Certifications
None
Material certified per EN 10204 Par. 3.1B and pressure test per AD2000
Inspection Certificate EN 10204  3.1B
Pressure test per AD2000
Calibration Certificates
Standard
Others
Protection Class
IP 67 (connection box with Pg 13.5 cable connector)
IP 68 (connection box with Pg 13.5 cable connector) and PU resin compound
IP 67 (connection box with NPTAdapter)
Design1)
Without preamplifier (Standard DN 10 [3/8] to DN 1000 [4])
Preamplifier Typ A (Standard DN 3 [1/10] to DN 8 [5/16])
Preamplifier Typ B

0
3
4
8
9
A
C
Z
S
A
D
F
G
A
Z
2
3
4
1
2
4

1) Notes on design
Design

Minimum Conductivity
DN 3 [1/10] to DN 8 [5/16]

DN 10 [3/8] to DN 1000 [4]

without preamplifier

not possible

= > 20 S/cm

Preamplifier Typ A

= > 20 S/cm

= > 5 S/cm

Preamplifier Typ B

= > 5 S/cm

= > 0,5 S/cm

Max. cable lenght without preamplifier 50 m


Max. cable lenght with preamplifier 200 m (Empty pipe detection not possible)

Additional Ordering Information may be provided in


writing:
Name Plate (also available in Stainless Steel, option)
German
English
French
Electrode Design

Standard
Conical head (from DN 10 [3/8] material SST No. 1.4539 used
for high grease content fluids)
Others
Gasket Materials
EPDM with FDAApproval (CIP capable, no oil or grease)
Silicone with FDAApproval (suitable for oil and grease)
PTFE (Teflon)
None

27

Electromagnetic Flowmeter FSM4000


D184S073U02

Installation Accessories

Chemical Applicationd (Grounding)


For plastic pipelines or pipelines with insulating liners, protection/
grounding plates made of conductive PTFE (patent applied for),
replace the expensive special materials (Platinum, Tantalum etc.)
presently used. Succesfully tested in numerous applications.
N Holes at scale

d2

di

R8

R2

16

6.6

l1

Fig. 29

Fig. 31

Protection Plates/Grounding Plates

The loose grounding plates are installed between the flowmeter flange
and the mating flange.
Available for sizes DN 3DN 600 [1/1024]. Mat. SS 1.4571[316Ti]
other materials upon request .
DN
(inch)
PN
T=Thick
380
1/103
1040
1,5
100150
46
2540
1,5
100250
410
1016
1,5
300600
1224
1016
3

40.5

PTFE Protection/Ground Plates

Nennweite
DN 10100
< DN 200

(inch)
3/84

PN
1040
upon request

Thickness
4 mm

Wafer Design Accessories for Model SE21W


The following accessories are available as a function of the meter size
and the pressure rating:
Bolts, nuts, lockwasher, centering elements
Gaskets are not included ion the accessories.

Material: SS
C

R2

Ground
connection 1)

Meter Size
DN 3 to 10
1/10 to 3/8

R2

Ground
connection 1)

R8

10

Grounding Plate, Loose (Centered


by Mounting Bolts)

R8

16

16

20

6.4

6,6

Protection Plate/Grounding Plate


Mounted on Flowmeter 2)

1) Attach ground cable with a M6 screw,

lock washer and nut to the plate.


B = Inside diameter of the lining
2) Grounding plate mounted directly on the instrument
DN 10 to DN 600 [3/8 to 24]
Material: Stn. stl. 1.4571 [316Ti]
Thickness DN 100 [4] = 3 mm, DN 125 [5] = 5 mm
B = Inside diameter of the lining

Fig. 30

Grounding plate

Protection Flanges
Protection flanges protect the liner from abrasive fluids and are
centered in the pipeline using a groove or spring. Protection flanges
are available in 2 thicknesses and are mounted on the intrument.
Added installation length for Series 1000 and DIN/ANSI flanges
DN 80 [3] L + 20 mm, from DN 100 [4] L + 25 mm.
Gaskets for Protection Plates/Grounding Plates
Only one gasket at the mating flange is required for PTFE/PFA lined
instruments installed with metal grounding plates. A gasket should be
placed on either side of the grounding plate for hard/soft rubber lined
instruments. Commercially available gaskets made of a material
(rubber, lt, PTFE, etc) compatible with the fluid and the operating
temperatures should be used.

28

DN 15
1/2
DN 20
3/4
DN 25
1
DN 32
1 1/4
DN 40
1 1/2
DN 50
2
DN 65
2 1/2
DN 80
3
DN 100
4

Pressure Rating
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  40
150 lb
300 lb
PN 10  16
150 lb
PN 10  40
150 lb
PN 10  16
150 lb

Part Number
D614L265U03
D614L265U03
D614L265U04
D614L265U03
D614L266U05
D614L266U06
D614L267U04
D614L267U05
D614L267U06
D614L268U04
D614L268U05
D614L268U06
D614L269U04
D614L269U05
D614L269U06
D614L270U04
D614L270U05
D614L270U06
D614L296U04
D614L296U05
D614L296U06
D614L297U08
D614L297U10
D614L298U08
D614L298U09
D614L299U07
D614L299U09

Electromagnetic Flowmeter FSM4000


D184S073U02

Specifications: Converter

Protection Class per EN 60529


IP 67 for field mount housing and NEMA 4X
Design
Field Mount Housing made of cast Alum. per DIN 1725, painted. Paint
coat 60 m thick. Lower section (RAL 7012), upper section (RAL
9002). Dimensions Page 35. Weight appr. 3.3 kg (7.28 pound).
Electrical Connections
Cable entry M20 x 1.5, NPT, PF with adapter (upon request), screw
terminals
Damping
Selectable from 0.1 to 99.0 s
Zero Return
Selectable from 0 to 10 % of range end value
Signal Cable
Max. cable length between flowmeter primary and converter is:
50 m for the standard design and versions with automatic zero return,
from DN 10 [3/8] and from 20 S/cm.
200 m for designs with preamplifiers. A 10 m long signal cable is
included with flowmeter. If a longer cable is required, see Footnote on
Page 35, Ordering Information, Converter.
Magnet Coil Supply Cable

Fig. 32

Is required for connecting the flowmeter primary to the converter. For


EMCreasons the 2 wire cable is shielded. A 10 m long cable is
included with flowmeter. If a longer cable is required, see Footnote on
Page 35, Ordering Information, Converter.

Converter FSM4000S4

Forward/Reverse Flow Metering

Flow Range
Any flow range whose 100 % value corresponds to a flow velocity
between 0.5 m/s and 10 m/s can be selected.
Minimum Conductivity
20 S/cm standard
5 S/cm with preamplifier
0.5 S/cm with preamplifier

DN 31000 [1/1040]
DN >8 [ 5/16]
DN 101000 [3/840]

Reproducibility
0.2 % of rate
Response Time
1 = 70 ms (066 %) in Operation mode Fast
1 = 200 ms (066 %) in Operation mode (Standard)
Supply Power
Unom = 115230 V 50/60 Hz
Urat = 85253 V 50/60 Hz
50/60 Hz 6 %
20,426,4 V AC,
20,431,2 V DC Ripple 5 %
Power

The flow direction is indicated in the display by direction arrows. It is


also possible to transmit the flow direction information over the
contact output.
Display
4line, lighted, graphic display. To improve the readability, the flow
information can be displayed on two lines with larger characters. The
individual flowrate and output values can be displayed. Nonswitching
dotmatrix display with max. 4 x 16 characters. An clear text error
diagnosis is displayed when the system monitor detects an error
condition. The alarm condition can also be signaled over the alarm
output.
Data Security
All data is stored in a FRAM (Ferroelectric Nonvolatile Random Access
Memory) for a 10 year period without requiring power during a power
outage or when the main supply power is turned off. Additional
security is provided by a serial FRAM installed in the converter and on
the external connection board in which the parameter settings and
process information are stored. The data in the external FRAM can be
uploaded into the converter. Thereby an easy converter exchange is
possible  without reconfiguration  by simply uploading the stored
data from the external FRAM into the new converter.

S 45 VA (flowmeter primary including converter)


Ambient Temperature
20 to +60 C [4 to +140 F]
Storage Temperature
20 to +80 C [4 to +176 F]
Relatively humidity
Acc. to IEC 6006821
Vibration
Acc. to DIN EN 6006826 (1055 Hz: amplitude = 0.15 mm)

29

Electromagnetic Flowmeter FSM4000


D184S073U02
Flow Totalization
The flow is totalized in engineering units. The pulse factor, pulses per
unit, can be set between 0.001 and 1000. The totalizer values for the
forward and reverse flow directions can be displayed simultaneously.
Display selections are made in the submenu "Display".
Parameter Settings
Entries are made using menus displayed in a number of languages
(German, English, French, Finnish, Swedish, Italian, Spanish and
Dutch) using 3 keys in dialog with the display or by communication
over a data link, using a PC, HARTProtocol or Profibus (in
preparation).
A Magnet Stick can be used for configuration without opening the
front door. Position the Magnet Stick over the magnet symbols on the
cover over the display.
Alarm Signals
The converter can incorporate a max/minalarm. The alarm limits can
be set anywhere within the range of 0105 % of the flow range end
value. When the flowrate is outside of the alarm limits set, an error
message is displayed in the upper line. It is also possible to transmit
a signal over the contact output.
Upgrade
The converter with all its functions can also be used with older model
electromagnetic flowmeters. The appropriate model number must be
selected in the converter. It is also possible to use the converter with
flowmeters Model 10D1422. A special converter version is required
when updating an older model (see Ordering Information, Converter)
and is applicable to the meter size range DN 3 to DN 1000 [1/10" to
40"] only in the low voltage design.
Isolation In-/Output
The current output, pulse output and the contact input and contact
output are galvanically isolated from the input circuit and from each
other.

In-/Output Signals
Current Output (terminals 31, 32)
The following functions are available in the software:
0 to 20 mA load 560 Ohm
4 to 20 mA load 560 Ohm
0 to 10 mA load 1120 Ohm
2 to 10 mA load 1120 Ohm
Contact Output (terminals 41,42)
The following functions are available in the software:
Forward/reverse flow direction signal
Flowrate alarm (max. or min.)
General alarm
Empty pipe
The output contact can be configured as normally open or normally
closed.
Optocoupler specifications
16 V UCEH 30 V, 2 mA ICEL 220 mA
Scaled Pulse Output (terminals 51, 52)
Maximum count frequency 5 kHz. Pulse factor per engineering unit
from 0.001 to 1000/unit selectable in the software. The pulse width
can be set from 0.100 ms to 2000 ms.
The design (active, passive) can be selected using jumpers.
Active
Voltage pulse 24 V rectangular,
Load 150 Ohm,
Pulse width 50 ms, count frequency 3 Hz,
Load 500 Ohm,
Pulse width 0.1 ms, max. count frequency: 5 kHz.
Passive
Optocoupler
"closed ":
0V UCEL 2 V, 2 mA ICEL 220 mA
"open":
16 V UCEH 30 V, 0 mA ICEH 2 mA

Contact Input (terminals 81/82)


Passive contact (normally open). The following functions are available
in the software:
External Zero Return
When the pipeline empties all output signals can be turned off. The
current output can be set to 0 mA or 2/4 mA dependent on the
current output range selection.
External Totalizer Reset
Resets the internal totalizer value in the display and the pulse counter
for both flow directions.
Design Passive (optocoupler)
ON:
OFF:
Ri:

30

16 V UKL 30 V
0 V UKL 2 V
2 k

Electromagnetic Flowmeter FSM4000


D184S073U02

Converter

4-20 mA

Rb min = 250 Ohm

HART
Bell 202
Modem Box

PC
Fig. 33

Flowmeter
Primary
Hand Terminal

HARTProtocol

HART-Protocol Version 5.x

The HART Protocol provides for communication between a


process control system, Hand Terminal and the EMF field instrument.
The communication signals, which are superimposed as an AC
voltage on the current output, do not affect any other instruments
connected to the current output. This option is only available with a 4
20 mA current output.
Transmission Type
FSKModulation on the current output 420 mA per
Bell 202 Standard.
Baudrate
1200 Baud

System Connections
In conjunction with the DTM (Device Type Manager) available of the
instrument (Software Revision from B.10) the communication
(Configuration, Parameter Settings) can be activated using the
appropriate applications per FDT 0.98 or 1.2 (SMART VISION 4.01
R2).
AMS
in preparation
PROFIBUS PA
in preparation
FOUNDATION Fieldbus FF
in preparation

Format
Logic 1: 1200 Hz
Logic 0: 2200 Hz
Cable
Max. 1500 m AWG 24 twisted and shielded
Max. Signal Amplitude
1.2 mApp
Load Current Output
min.: 250 ,
max.: 560
The digital communication on the current output does not affect any
other connected instruments.
To configure using SmartVision a DTM (Device Type Manager) is
required.

31

Electromagnetic Flowmeter FSM4000


D184S073U02

Connection Examples for Peripherals

Scaled Pulse Output active


internal
+24 V DC

Scaled Pulse Output passive, optocoupler

external

internal

external

51

51

52

52

RB*
+1630 V DC


Current Output active with/without HART-Protocol (420) mA


internal

external

0/420 mA

+31

0/210 mA

32

Contact output for system monitor, max.min.alarm, empty


pipe or forward/reverse flow direction signal. Function soft
ware selectable

internal

internal

external
41

+U

Contact input for external totalizer reset and


external zero return
Function software selectable

RB*

42

0V
+24 V DC

external
81

RI
82

+1630 V DC
0V

CE
* RB 

I

Fig. 34

32

Connection Examples for Peripherals

CE

Electromagnetic Flowmeter FSM4000


D184S073U02

Interconnection Diagram: Flowmeter Primary DN 1 to DN 1000 [1/2540], Converter Field Mount


Housing

Converter S4
Field Mount Housing

U+

2S

U
2

1S

Functional Earth

24 V

8)

51 52 81 82 41 42 31 32

16

7)

6)

5)

4)

M3 M1

3)

2

1+

2)

1)

Attention!
SE

2S

1S

16

M3 M1 SE

SE

U+

U

16

M3 M1 SE

Flowmeter Primary

1)

Supply power
High voltage:
Low voltage:
Frequency:

With preamplifier
(always at DN1[1/25]
 DN8 [5/16])

Functional Earth

Is the flowmeter primary includes a preampli


fier either for low conductivity applications or
for flowmeter sizes DN 1 to DN 8 [1/25 to 5/
16], the supply voltage 12 V DC must be
connected to terminals U+ and U at both
the flowmeter primary and at the converter.

Urat = 85 V AC to 253 V AC / Unom = 115 to 230 V AC


Terminals L, N,
20.4 to 26. V AC, 20.4 to 31.2 V DC
Terminals 1+, 2,
47 Hz f 53 Hz; 50 Hz Supply power
56 Hz f 64 Hz, 60 Hz Supply power

2)

Magnet coil supply:

Shielded 2 x 1 mm CE Type 227 TEC 74


ABB Part No. D173D147U01
10 m included with shipment standard

3)

Current output:
Function:

Software selectable
Active
0/420 mA, 0/210 mA,0 RB 560
Terminals: 31, 32

4)

Contact output:
Function:

Software selectable
Passive
closed: 0V UCEL 2 V, 2 mA ICEL 220 mA
open: 16 V UCEH 30 V, 0 mA ICEH 2 mA
Terminals: 41, 42

5)

Contact input:
Function:

Software selectable
Passive
On:
16 V UKL 30 V
Off:
0 V UKL 2 V
Ri = 2 k
Terminals: 81, 82

6)

Pulse output:
Function:

Selectable
Active/Passive
fmax:
5 kHz, pulse width 0.1 ms2000 ms
Setting range:0.0011000 pulses/units
closed: 16 V UCEH 30 V, 2 mA ICEH 220 mA
open: 0 V UCEL 2 V, 0 mA ICEL 2 mA
Terminals: 51, 52

7)

Shielded signal cable:

ABB Part No. D173D018U02, 10 m included with shipment.

8)

Shielded signal cable:

Voltage supply for flowmeter primary with preamplifier


Terminals +U, U instead of on 2S and 1S in the standard instrument.

Comment:
We recommend that the output cables be shielded with the shield at one end connected to functional earth.
Fig. 35

Interconnection Converter S4

33

Electromagnetic Flowmeter FSM4000


D184S073U02

Interconnection Diagram: Upgrade Flowmeter Primary DN 3 to DN 1000 [1/1040], Converter


Field Mount Housing, Model 10D1422 (10DI1425, 10DS3111A-C DN 5001000 [2040])
Terminal board: D685A1020U03
Switch S903 must be closed

Converter S4
Field Mount Housing

3 2S

1 1S 3

16 7

Functional Earth

24 V

8)

51 52 81 82 41 42 31 32 M3 M1

7)

6)

5)

4)

3)

1)

2)

Attention!
SE

Supply power see name plate primary

Flowmeter Primary

1)

Supply Power
High voltage:
Frequency:

2)
3)

Magnet coil supply:


Current output:
Function:

4)

Contact output:
Function:

5)

Contact input:
Function:

6)

Pulse output:
Function:

7)
8)

Shielded signal cable:


Reference cable:

Functional Earth

Urat = 85 V AC to 253 V AC / Unom = 115 to 230 V AC


Terminals L, N,
47 Hz f 53 Hz; 50 Hz Supply power
56 Hz f 64 Hz, 60 Hz Supply power
Magnet coil supply from Supply power
Software selectable
Active
0/420 mA, 0/210 mA, 0 RB 560
Terminals: 31, 32
Software selectable
Passive
closed: 0V UCEL 2 V, 2 mA ICEL 220 mA
open: 16 V UCEH 30 V, 0 mA ICEH 2 mA
Terminals: 41, 42
Software selectable
Passive
On:
16 V UKL 30 V
Off:
0 V UKL 2 V
Ri = 2 k
Terminals: 81, 82
Selectable
Active/Passive
fmax:
5 kHz, pulse width 0.1 ms2000 ms
Setting range: 0.0011000 pulses/units
closed:
16 V UCEH 30 V, 2 mA ICEH 220 mA
open:
0 V UCEL 2 V, 0 mA ICEL 2 mA
Terminals: 51, 52
ABB Part No. D173D018U02 using this signal cable or use the available installation
Only for connecting to reference cable terminals: Terminals: 6, 7 in Model 10D1422

Comment:
We recommend that the output cables be shielded with the shield at one end connected to functional earth.

Fig. 36

34

Interconnection Upgrade Converter S4

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Information: Converter S4


In addition to the Ordering Number please supply the following information:
Flow range, if applicable current output, pulse output active/passive, pulse factor and pulse width. These selections can be factory preset.
Ordering Number
External Converter

S4
S4 0.5 % v. M.

Housing
Field mount housing (with cable connector M20x1.5), standard
Field mount housing (with adapter for cable connector NPT 1/2)
Supply Power
High voltage 85253 V AC
Low voltage 20.426.4 V AC, 20.431.2 V DC
Display
Magnet Stick operation and graphic display, lighted

A
B
G
K
A

In-/Output Options
1)
Current output + Pulse output active + Contact input + Contact output
Current output + Pulse output active + Contact input + Contact output + HART
1)
Current output + Pulse output passive + Contact input + Contact output
Current output + Pulse output passive + Contact input + Contact output + HART
Contact output + PROFIBUS PA
Contact output + FOUNDATION Fieldbus FF
Contact output + PROFIBUS PA with M12 plug
Application
Standard
For Flowmeter Primary
Flowmeter Primary Type SE2_, SE4_, standard
Flowmeter Primary Type DS2_, DS4_, 10DS3111, 10DI1425
Flowmeter Primary Type 10D1422, 10DI1425 ( DN 500)2)
Others
Approvals
None
Name Plate
German
English
French
Stainless Steel German
Stainless Steel English
Stainless Steel French
Design Level
Software Level

01
02
03
04
07
08
10

in preparation
in preparation
in preparation
0
1
2
3
9
0
G
E
F
GN
EN
FN
*
*

1) Option not available for Supply power Low voltage


Shielded signal cable, 10 m included with each flowmeter. For longer cables order Part No. D173D018U02.
Shielded coil supply cable, 10 m included with each flowmeter. For longer cables order Part No. D173D147U01.
2) The connection board for the reference cable includes a voltage splitter, magnet coil supply for the flowmeter primary
from a separate external power supply. FRAMVariant Cs = 100 % and Cz = 0 % required.

35

Electromagnetic Flowmeter FSM4000


D184S073U02

Dimensions: Field Mount Housing


198

139.7

167

10

1)

1)

1)

1)

14.5

249

265

38

66

83.5

132

Field Mount Housing with Window

Cable Connector M20 x 1.5

Mounting Dimensions
1) Mounting holes for
2 pipe mounting kit
Mounting kit upon request

Opening the Housing Cover


Loosen the cover screws.
Remove cover or flip open the cover with a
window.

Press the cover with window on its closure


side against the gasket surface and install
and tighten the cover screws.
Set the small cover in the frame and install
and tighten the cover screws.

Min.175

Closing the Housing Cover

min. 62

All dims in mm
Fig. 37

36

Dimensions, Converter Housing and Installation proposal

Projektion nach ISO Methode E

Electromagnetic Flowmeter FSM4000


D184S073U02

Ordering Number: Test simulator


Flowmeter Primary Simulator
Setting the Flow Signal
None (adapter only)
3position Digiswitch with 1000 steps
Supply power 1)
None (Adapter orders only)
110 V240 V (50/60 Hz) with Schuko plug
24 V/48 V (AC/DC) with 4 mm plug
110 V240 V (50/60 Hz) with USA plug
Others

55XC4
0
1
0
1
2
3
9

Accessories
None
Adapter for converter 50XM2000/50XE4000/E4/50XF4000
Adapter for simulator 55XC2000 2)
Adapter plate for converter S4
Others
Design Level
(Specified by ABB)
Name Plate
German
English
French
1) Supply power must correspond to value listed on the converter Name Plate.
2) The adapter listed above for Converters 50XM2000/50XE4000/E4 can also be used for existing simulators 55XC2000.
If two adapters are required, then the Ordering Number for each adapter type is to be used.

0
1
2
5
9
*
1
2
3

Software
Meter parameter documentation software for PC upon request.
SMART VISION Communication tool for HART
No charge for 90 day evaluation version.
DTMs are included in SMART VISION Tool.

37

Electromagnetic Flowmeter FSM4000


D184S073U02

38

Electromagnetic Flowmeter FSM4000


D184S073U02

39

The IndustrialIT wordmark and all mentioned


product names in the form XXXXXXIT are registered
or pending trademarks of ABB.

The Companys policy is one of continuous product


improvement and the right is reserved to modify the
information contained herein without notice.

ABB has Sales & Customer Support


expertise in over 100 countries worldwide.

Printed in the Fed. Rep. of Germany (10.03)

D184S073U02 Rev. 03

ABB 2003

www.abb.com

ABB Ltd.
Oldends Lane, Stonehouse
Gloucestershire, GL 10 3TA
UK
Phone: +44(0)1453 826661
Fax: +44(0)1453 827856

ABB Inc.
125 E. County Line Road
Warminster, PA 18974
USA
Phone: +1 215 674 6000
Fax: +1 215 674 7183

ABB Automation Products GmbH


Dransfelder Str. 2
37079 Gttingen
GERMANY
Phone: +49 551 905534
Fax:
+49 551 905555
EMail Customer Care Center:
CCC-support.deapr@de.abb.com

S-ar putea să vă placă și