Sunteți pe pagina 1din 31

Technical specification

centrifuge No.:
user manual - Cat. No.:

MPW-54
20054/.EN

2014-12-09

USER MANUAL

LABORATORY CENTRIFUGE

MPW-54
READ IT BEFORE USE!

MPW MED. INSTRUMENTS, 46 Boremlowska Street, 04-347 Warsaw/Poland


tel. +48 22 610 81 07 (service), fax +48 22 610 55 36
mpw@mpw.pl, www.mpw.pl

Technical specification

WARNING SIGNS AND HAZARD ICONS

WARNING!
Warning of potential injury or health risk.
DANGER!
Risk of electric shock with potential for severe injury or death as a consequence.
DANGER!
Biohazard with potential for risk to health or death as a consequence.
DANGER!
Risk of explosion with potential for severe injury or death as a consequence.

This manual was prepared with special care. MPW MED. INSTRUMENTS may change the manual at any
time and without notice because of improvements and inaccuracies of current information.
2

Technical specification

Content
1 TECHNICAL SPECIFICATION .................................................................................................................... 4
2 APPLICATION ......................................................................................................................................... 5
3 SAFETY NOTE ......................................................................................................................................... 6
3.1 PERSONNEL .................................................................................................................................................. 6
3.2 GUARANTEE.................................................................................................................................................. 6
3.3 LOADING THE ROTOR ...................................................................................................................................... 7
3.4 CURRENT PROTECTION .................................................................................................................................... 8
3.5 SAFETY HINTS ................................................................................................................................................ 8
3.6 MAINTENANCE CONDITIONS............................................................................................................................. 9
3.7 SAFETY PRECAUTIONS ..................................................................................................................................... 9
4 INSTALATION ....................................................................................................................................... 11
4.1 THE CONTENTS OF THE PACKAGE ..................................................................................................................... 11
4.2 LOCATION .................................................................................................................................................. 11
5. OPERATING AND DESIGN ...................................................................................................................... 12
5.1 DESCRIPTION OF CENTRIFUGE ......................................................................................................................... 12
5.2 CONTROL ELEMENTS ..................................................................................................................................... 12
5.3 ROTOR AND ACCESSORIES INSTALLATION........................................................................................................... 13
5.4 CONTROL DEVICE ......................................................................................................................................... 13
5.5 SET UP PARAMETERS..................................................................................................................................... 13
5.6 SAFETY FEATURES......................................................................................................................................... 14
6. CENTRIFUGING ...................................................................................................................................... 15
6.1 CONTROL PANEL .......................................................................................................................................... 15
6.2 HOLD MODE .............................................................................................................................................. 16
6.3 CENTRIFUGING FINISH ................................................................................................................................... 16
6.4 SIGNALLING STATES OF CENTRIFUGE ................................................................................................................. 17
6.5 MAINTENANCE OF CENTRIFUGING ELEMENTS ..................................................................................................... 18
6.6 STERILIZATION ............................................................................................................................................. 19
6.7 CHEMICAL RESISTANCE .................................................................................................................................. 21
7. GUARANTEE, SERVICE.......................................................................................................................... 22
8. DISPOSAL ............................................................................................................................................ 23
9. MANUFACTURERS INFO ..................................................................................................................... 24
10. ANNEXES ............................................................................................................................................. 25

ADDITIONAL ACCESSORIES

DECLARATION OF CONFORMITY

DECLARATION OF DECONTAMINATION (REPAIR)

DECLARATION OF DECONTAMINATION (RETURN)

Technical specification

Technical specification

manufacturer
type
mains voltage L1+N+PE, 10%
frequency, 10%
connected load (max)
overcurrent protection
capacity (max)
speed RPM
force RCF
running time

MPW MED. INSTRUMENTS SPDZIELNIA PRACY,


Boremlowska 46 Street, 04-347 Warsaw
MPW - 54
MPW 54s
230V
115V
230V
115V
50/60Hz
60W
fuse WTA-T 2A 250V
120ml
3500, 5800,
1000, 3500,
3122
1137
01:00 30:00 [min., s] (step 1min.)

short-time operation mode SHORT

no

continuous operation mode HOLD


electromagnetic compatibility
ambient conditions
set-up site

yes
according to PN-EN 55011
PN-EN 61010-1 (pkt.1.4.1)
indors only

ambient temperature

2 40C

humidity (maximum relative humidity)

< 80%

excess-voltage category

II

pollution degree

PN-EN 61010-1
PN-EN 61010-1

safety area
dimensions
height (H)

180 mm

180 mm

width (W)

220 mm

220 mm

depth (D)

270 mm

270 mm

367 mm
56 dB
4 kg
4 kg

367 mm
56 dB

height with open lid(Hoc)


noise level
weight of centrifuge 230V
weight of centrifuge 230V

300 mm

4kg
4 kg

4 kg
4 kg

There is a possibility of using power inverter 12DC/230AC (look: 10.ANNEXES ADDITIONAL


ACCESSORIES OTHER)

Application

Application

The MPW-54 is a table top laboratory centrifuge for in vitro diagnostic (IVD). Device is used for
separation samples taken from people's, animals and plants components of different densities,
under the influence of the centrifugal force, to provide information about their biological state. Its
construction ensures easy operation, safe work and wide range of applications at laboratories
engaged in routine medical analyses, biochemical research works etc. This centrifuge is not biotight
and therefore during centrifugation of preparations requiring biotightness one has to use closed and
sealed containers and rotors. In the centrifuge, it is prohibited to centrifuge caustic, inflammable and
explosive preparations.

Safety notes

Safety notes

3.1 Personnel

Laboratory centrifuge can be operated by laboratory personnel after getting


acquainted with user manual.

User manual shall be always held near the centrifuge.

The centrifuge can not be misused.

Guarantee period amounts to minimum 24 months (unless otherwise specified in


the purchase documents).

The service life of the centrifuge specified by the manufacturer amounts to 10


years.

After termination of guarantee period it is necessary to carry out yearly technical


inspections of the centrifuge.

Manufacturer reserves the right to make technical changes in manufactured


products.

Maximum period of storage of not used centrifuge amounts to 1 year. After this
period, a service authorized by manufacturer should carry out technical
inspection of the centrifuge.

3.2 Guarantee

Safety notes

3.3 Loading the rotor

Fix the rotor on the motor axis firmly.

Avoid unbalance.

Load opposite buckets with the same accessories.

Centrifugation of the test tubes of different sizes:


There is a possibility to centrifuge test tubes of different sizes; however, it is
absolutely necessary in such cases that opposite buckets and round carriers be the
same.

CORRECT

INCORRECT

It is necessary to insert test tubes symmetrically on the opposite sides


It is necessary to insert buckets symmetrically on the opposite sides

UWAGA

If the centrifuge will run with a large unbalance it will occur dangerous vibrations.
Buckets 13080 and 13081 are part of common balanced kit and marked by one number.
One should load rotor with buckets from the same kit.

FILLING TUBES

Fill test tubes outside the centrifuge.

Please pay special attention to the quality and proper thickness of the glass test
tubes walls. Those shall be test tubes for centrifuges.

Safety notes

3.4 Current protection


The centrifuge is equipped with thermal current protection. Fuse is
situated in the plug-in socket unit at back wall of the centrifuge

Supply voltage given on the rating plate has to be consistent


with local supply voltage. MPW MED. INSTRUMENTS laboratory
centrifuges are 1st safety class devices and they are provided
with the three-core cable with the plug resistant to dynamic
loadings.

Mains socket shall be provided with the safety pin. It is


recommended to install emergency cut-out that shall be located
far from the centrifuge, near the exit or beyond the room.

Before switching on, check whether the centrifuge is


connected to power supply correctly.

3.5 Safety hints


ROTORS MAINTENANCE

Use only accessories in good condition.

Protect equipment against


maintenance.
HAZARDOUS
MATERIALS

corrosion

using

accurate

preventive

Infectious materials could be processed in closed buckets only.

It is not allowed to subject to centrifugation toxic or infectious materials


with damaged leak proof seals of the rotor or test-tube. Proper
disinfection procedures have to be carried out when dangerous
substances contaminated the centrifuge or its accessories.

EXPLOSIVE AND COMBUSTIBLE MATERIALS

It is not allowed to centrifuge explosive and inflammable materials.

It is not allowed to centrifuge substances prone to reacting in result of


supplying high energy during centrifugation. The centrifuge can not be
operated in explosion-endangered areas

It is not allowed to centrifuge materials capable of generating inflammable


or explosive mixtures when subjected to air.

Safety notes

3.6 Maintenance conditions


START-UP
Prior to switching the centrifuge on, one shall read all sections of this instruction
carefully in order to ensure smooth operation and avoid damages of this device or
its accessories.

In order to protect the centrifuge against unbalance, fill in the test tubes up to the
same weight.

TRANSPORTATION

Centrifuge must not be transported with the rotor mounted on the shaft.

GENERAL HINTS

One must use original rotors, test-tubes and spare parts only.

In case of faulty operation of the centrifuge one shall ask for assistance of service
of MPW MED. INSTRUMENTS company or its authorized representatives.

It is not allowed to switch the centrifuge on if it is not installed properly or rotor is


not fitted correctly.

CENTRIFUGES SUBSTANCES

It is not allowed to exceed load limit set by the manufacturer. Rotors are intended
for fluids of average homogeneous density equal to 1,2 g/cm3 or smaller when
centrifugation is carried out at maximum or submaximum speed.

In order to avoid overloading of the rotor, please observe the maximum load, which is
stored on each rotor. The maximum permissible load is achieved when all the tubes are
filled with a liquid of 1,2 g/cm3 density.
If the density of the centrifuging liquid is greater than 1,2 g/cm3, a tube can be filled only
partially, or limit the speed of a centrifuge, which shall be determined from the formula:
n admissible = nmax x

1,2

= density 3
cm

nmax =max speed [rpm]

3.7 Safety precautions


For safety reasons, inspections of the centrifuge carried out by the authorized service at least
once a year after the period of warranty. The reason for more frequent inspections could be
corrosion inducing environment. Examinations should end with issuing report of validation that
checks on the technical state of the laboratory centrifuge. It is being recommended to establish
document where every repairs and reviews are being registered. Both these documents should be
stored in the place of use of the centrifuge.

Safety notes

INSPECTION PROCEDURES CARRIED OUT BY THE OPERATOR


Operator has to pay special attention to the fact that the centrifuge parts of key
importance due to safety reasons are not damaged. This remark is specifically important
as for:
Centrifuge accessories and especially structural changes, corrosion, preliminary
cracks, abrasion of metal parts.

Screw joints.

Inspection of the rotor assembly.

Inspection of bioseals of the buckets if such are used.

Control of execution of the guarantee yearly technical inspection of the centrifuge.


Only the manufacturer-specified holders, included in the equipment list, as well as
centrifuge capillaries, which diameter, length and durability are suitable, should be used
for spinning in this centrifuge. The use of equipment made by other manufacturers should
be consulted with the manufacturer of the centrifuge.

It is not allowed to lift or shift the centrifuge during operation, and rest on it.

It is nor allowed to stay in the safety zone within 30 cm distance around the
centrifuge neither leave within this zone some things, e.g. glass vessels.

It is not allowed to put any objects on the centrifuge.

COVER OPENING

It isn't allowed to open the cover manually in emergency procedure when rotor is
still turning.

ROTORS

It is not allowed to use the rotors and round carriers with signs of corrosion or
other mechanical defects.

It is not allowed to centrifuge highly corrosive substances which may cause


material impairment and lower mechanical properties of rotor and round carriers.

It is not allowed to use rotors and accessories not admitted by the manufacturer.
Let to use commercial glass and plastic test tubes, which are destined to
centrifuging in this laboratory centrifuge. One should absolutely not use poor
quality elements. Cracking of glass vessels and test tubes could result in dangerous
vibration of the centrifuge.

It is not allowed to carry out centrifugation with the rotor caps taken off or not
driven tight.

10

Instalation

Instalation

Open the package. Remove the box containing the accessories. Take out centrifuge from the
container. Keep the box and packing materials in case of service shipping

4.1 The contents of the package


Quantity
(pcs)
1
1
1
1
1
2
1
1

Name
Centrifuge MPW-54
Complete clamp
Spanner for the rotor
Key for emergency cover release
Power cord 230V/115V
Fuse WTA T2A 250V
Petroleum jelly 20ml
User manual

Cat No.
See name plate
17168
17099L
17162
17866/17867
17859
17201
20054.EN

4.2 Location

The centrifuge shall not be located near source of heat and shall not
be subjected to direct sunlight.

The table for the centrifuge shall be stable and shall have flat-levelled
table top.

It is necessary to ensure a safety zone of the minimum 30cm round the


centrifuge from every direction.

At the change of the place from cold to warm one, condensation of


water will occur inside the centrifuge.

It is important then that sufficient time be provided for drying the


centrifuge prior to starting the centrifuge again (min. 4 hours).

11

5. Operating and design

5. Operating and design


5.1 Description of centrifuge
New generation of MPW MED. INSTRUMENTS laboratory centrifuges is provided with state-ofthe-art microprocessor control systems, very durable and quiet asynchronous brushless motors and
accessories consistent with requirements of the present-day user.
The centrifuge has rigid self-supporting structure. Housing was made of ABS. Cover is fixed on
steel axles of hinges and from the front it is locked with electromagnetic lock blocking opening during
centrifugation. Rotation chamber was made from plastic. Rotors and containers are from aluminum,
reductive inserts from the polypropylene.

5.2 Control elements

5
4

1. Control panel

3. Hole for emergency lid

2. Power switch

iiiopening
4. Cover

5. Glass inspection

1. General view

1. Motor axle
3

2. Rotor
3. Rotor lid

4. Complete clamp

1. Mains socket

2. Fuse socket

3. Back of the centrifuge


2. Angle rotor assembly

12

5. Operating and design

5.3 Rotor and accessories installation

Connect the centrifuge to the mains (master switch on left side of the centrifuge).

Open the cover of the centrifuge by pressing the COVER key. Prior to putting the rotor in,
one has to check if the rotating chamber is free of impurities, e.g. such as dust, glass
splinters, residues of fluids that must be taken away.

One shall fit the rotor on the motor shaft driving it home on the cone.
Fitting the rotor too shallow will result in lack of identification of the rotor after start of
the centrifuge, displaying the error message and stopping the centrifuge.

Screw-in the bolt for fixing the rotor (clockwise) and screw it tightly home with the supplied
spanner for the rotor.

In case of rotors designed with the cover they must not be used without it. Rotor covers
must be closed exactly. Rotor covers ensure smaller drags of the rotors, proper setting of the
test-tubes and airtight sealing.

Fill test tubes outside the centrifuge.

Put on or screw the caps on vessels and rotors (if applicable).

In case of centrifuging in an angle rotor, test tubes (buckets) have to be filled properly in
order to prevent from pouring fluids during centrifuging.
Centrifuge will tolerate small weight differences occurring during loading of rotors.
However it is recommended to equalize vessels loads as much as possible in order to
ensure minimal vibrations during operation. When the centrifuge is started with large
imbalance, the unbalance control system will switch-off the drive system and error signal
will be transmitted. On the monitoring panel, error message will be displayed.

Threaded parts shall be lubricated with the petroleum jelly.

For replacement of the rotor one shall unscrew clamping and then grab the rotor with both
hands at opposite sides, taking it away from drive shaft by pulling it up.

5.4 Control device


The microprocessor control unit of the centrifuge ensures broad possibilities of providing,
realisation and reading of work parameters.

5.5 Set up parameters


Data setting and read-out system forms hermetically closed keyboard with distinctly accessible
operation points. Easily readable display signaling time to end of centrifuging and set mode of speed.
It make easy for the operator to the registration status of the device.
Control system provides possibility of:
set up speed mode (H) 5800 or (L) 3500 rpm (MPW-54), (H) 3500 or (L) 1000 rpm (MPW-54s),
set up time of centrifuging in the range 130 min or hold time mode.

13

5. Operating and design

5.6 Safety features


Cover lock
The centrifuge can be started only with properly closed cover. While, the cover can be
opened only after stopping the rotor. In case of emergency opening of the cover during operation,
the centrifuge will be immediately switched-off and the rotor will brake till complete stopping.
Rest state inspection
Opening of the centrifuges cover is possible only with the rotor in the state of rest.

EMERGENCY COVER RELEASE


In case of e.g. mains failure it is possible to open cover manually. At first, one must be sure
that rotor is not in the move (use inspection glass). On the right-hand side of the casing
there is a hole. Insert emergency opening key (17162) into the hole.
It is not allowed to use emergency cover release when the rotor is running!

14

6. Centrifuging

6. Centrifuging
Power switching ON/OFF is carried out with master switch situated on the side wall of the centrifuge.
All settings on the centrifuge are done by means of the control panel.
6.1 Control panel
The control panel (equipped with membrane keyboard) placed on the front casing serves the
purpose of controlling centrifuge operation. STATUS/SPEED display show information about state of
centrifuge (left side) and speed mode (right side). TIME display show information about time remain
to the end of centrifuging.

4. Control panel

Buttons application:

START

Start of centrifuging

STOP

Cancel of centrifuging

COVER

Open the lid

TIME

Set up runtime (in 1 minute step)

L/H

SPEED

Change speed mode

15

6. Centrifuging

6.2 HOLD mode


continuous mode (centrifuging until
pressing STOP)

HOLD mode

In order to run the HOLD mode, set the


time value to - -

6.3 Centrifuging finish


WAYS OF FINISH THE CENTRIFUGING

After reaching the preset working time centrifuging is stopped.

Before the expiry of the set time, you can interrupt the cycle by pressing
the STOP

x1

16

6. Centrifuging

6.4 Signalling states of centrifuge

17

Cover is locked

Cover is open

Rotor is running (speed is increasing or


constant) mark turn counter-clockwise

Rotor is running (breaking) mark turn


clockwise

Rotor is running, time is counting dot


next to time indication is blinking

Cycle cancelled or end of centrifuging a


dot next to the time indication stops
flashing, after braking the rotor, four
short beeps are issued

6. Centrifuging

Centrifuging in high rotational speed

Centrifuging in low rotational speed

Centrifuge has memory for remembering the last settings even after turn off and turn on the
centrifuge

6.5 Maintenance of centrifuging elements

In order to increase durability of gaskets and threaded places, it shall be


lubricated with petroleum jelly (cat. No 17201).

Cleaning of the accessories

In order to ensure safe operation one shall carry out in regular way periodical
maintenance of the accessories.

Rotors, buckets and round carriers have to withstand steady high stresses
originating from the centrifugal force. Chemical reactions as well as corrosion
(combination of variable pressure and chemical reactions) can cause
corrosion or destruction of metals. Hard to observe surface cracks increase
gradually and weaken material without visible symptoms.

In case of observation of surface damage, crevice or other change, as well as


the corrosion, the given part (rotor, bucket, etc.) shall be immediately
replaced.

In order to prevent corrosion one has to clean regularly the rotor with the
fastening bolt, buckets and round carriers.

Cleaning of the accessories shall be carried out outside of the centrifuge once
every week or still better after each use.

Then, those parts shall be dried using soft fabric or in the chamber drier at ca.
50C.

Especially prone to the corrosion are parts made of aluminium. For cleaning
them one should use neutral agent of pH value 68.

It is forbidden to use alkaline agent of pH > 8.

In this way, the useful service life of the device is substantially increased and

18

6. Centrifuging

susceptibility to corrosion is diminished.

Accurate maintenance increases the service life as well and protects against
premature rotor failures.

Corrosion and damages resulting from insufficient maintenance could not be


subject of claims lodged against the manufacturer.

One can use all standard disinfectants. The centrifuges and accessories are constructed from
various materials and one should to take into account possible variety of materials. During
sterilization by means of steam one should to consider temperature resistance of individual
materials.

6.6 Sterilization

sterilization *

PS
SAN
PMMA
PC
PVC
POM
PE-LD
PE-HD
PP
PMP
ECTFE
ETFE
PTFE
FEP/PFA
FKM
EPDM
NR
SI

radiation
radiation

C2H4O
(ethylene oxide)

formalin,
ethanol

25 kGy

121C,
20 min

1)
2)
1)

may be used

cannot be used

Laboratory vessels have to be exactly cleaned and rinsed with the distilled water before the sterilization in the
autoclave. It is always necessary to remove closures from containers!

1)

The frequent steam sterilization reduces mechanical durability! PC test tubes may become useless.

2)

Except PVC hoses which are resistant to the steam sterilization in the temperature 121C

19

6. Centrifuging

very good

Permanent action of the substance does not cause damage through 30 days. The
material is able to be resistant through years.

good to limited

Continuous action of the substance causes insignificant and partly reversible damage
through the period of 7-30 days (e.g. puffing up, softening, reduced mechanical
durability, discolouring).

limited

The material should not have the continuous contact with the substance. The
immediate occurrence of damage is possible (e.g. the loss of mechanical durability,
deformation, discolouring, bursting, dissolving).

Plastics
polystyrene

ECTFE

ethylene/chlorotrifluoroethylene

styrene-acrylonitrile

ETFE

ethylene/tetrafluoroethylene

polymethyl methacrylate

PTFE

polytetrafluoroethylene

polycarbonate

FEP

tetrafluoroethylene/perfluoropropylene

PVC

polyvinyl chloride

PFA

tetrafluoroethylene/perfluoroalkylvinylether

POM

acetal polyoxymethylenel

FKM

fluorcarbon rubber

PE-LD

low density polyethylene

EPDM

PE-HD

high density polyethylene

NR

natural rubber

polypropylene

SI

silicon rubber

PS
SAN
PMMA
PC

PP
PMP

alkalis

ahs

ether

esters

haloid
hydrocarbons

cyclic
hydrocarbons

weak or diluted
acids

aldehydes

PS
SAN
PMMA
PC
PVC
POM
PE-LD
PE-HD
PP
PMP
ECTFE
ETFE
PTFE
FEP
PFA
FKM
EPDM
NR
SI

oxidizing
substances

strong or
concentrated acids

cyclic alcohols

ketones

Chemical resistance of plastic

polymethylpentene

20

ethylene propylene diene

6. Centrifuging

DANGER!
For centrifuging infectious materials it is necessary to use hermetically closed buckets, in
order to prevent they migration into the centrifuge.

Rotors, buckets and round carriers can be sterilized in autoclave with temperature 121 124C and
pressure 215 kPa during 20 min. In the centrifuge, disinfectants and cleaning agents generally used in
medical care should be used (e.g. Aerodesina-2000, Lysoformin 3000, Melseptol, Melsept SF,
Sanepidex, Cutasept F).
Additional accessories can be sterilized using autoclave depending on material that they are made of.
See table STERILIZATION.
User is responsible for proper disinfections of the centrifuge, if some dangerous material
was spilled inside or outside of the centrifuge. During the above mentioned works one
must wear safety gloves.

6.7 Chemical resistance


One can use all standard disinfectants. The centrifuges and accessories are constructed from various
materials and one should to take into account possible variety of materials. During sterilization by
means of steam one should to consider temperature resistance of individual materials.

DANGER!
For centrifuging infectious materials it is necessary to use hermetically closed buckets,
in order to prevent they migration into the centrifuge.
User is responsible for proper disinfections of the centrifuge, if some dangerous
material was spilled inside or outside of the centrifuge. During the above mentioned
works one must wear safety gloves.

21

Guarantee, service

7. Guarantee, service
Manufacturer grants to the Buyer the guarantee on conditions specified in the Guarantee
Certificate. Buyer forfeits the right to guarantee repair when using the device inconsistently with the
User manual provisions, when damage results from the Users fault.
Repairs should be carried out in authorized service workshops, granted with the MPW
Certificate.
The centrifuge shall be sent to repair after decontaminating disinfections. Information about
authorized service workshops could be obtained from the Manufacturer.

22

Disposal

8. Disposal

When you are disposing the device, the respective statutory rules must be
observed.

Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August


13, 2005, may not be disposed as part of domestic waste.

The device belongs to 8th group (medical devices) and is categorized in


business to business field.

The icon of the crossed-out trash can shows that the device may not be
disposed as part of domestic waste. The waste disposal guidelines of the
individual EC countries might vary. If necessary, contact your supplier.

23

Manufacturers info

9. Manufacturers info

MPW MED. INSTRUMENTS

Boremlowska 46 Street
04-347 Warsaw

+48

22 610 56 67
22 610 81 07
22 610 55 36

www.mpw.pl
e-mail: mpw@mpw.pl
http://

DISTRIBUTOR:

24

sales department
service
fax

ANNEXES

10. ANNEXES
Capacity [ml]
Tube
x L [mm]
Catalogue number

15

15

10

10

glass

Nalgene
PP

conical
bottom
PP

with cap
PP

glass

12x100
15119

16x113
15048

17x120
15050

16x108
15053

12x75
15118

Angle rotor (30) 11101 with


buckets 13080
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
14083
13,3x30
13,3x30
lack of bottom
6
3120
5800
8,3
15
29

Capacity [ml]

10

Tube

with cap PP

glass

with cap PP

with cap
PP

x L [mm]
Catalogue number

11,7x95
15054

12x75
15120

12x85
15419

16x108
15053

Angle rotor (30) 11101 with


buckets 13081
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
13,3x30

14083
13,3x30

14082
14083
13,3x30
13,3x30
lack of bottom
6
3120
5800
8,3
15
29

25

14082
13,3x30

14083
13,3x30

ANNEXES
Capacity [ml]

Tube
x L [mm]
Catalogue number

15

15

10

10

glass

Nalgene
PP

conical
bottom
PP

with cap
PP

glass

12x100
15119

16x113
15048

17x120
15050

16x108
15053

12x75
15118

Angle rotor (30) 11144 with


buckets 13080
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
14083
13,3x30
13,3x30
lack of bottom
4
3120
5800
8,3
15
29

Capacity [ml]

10

Tube

with cap PP

glass

with cap PP

with cap
PP

x L [mm]
Catalogue number

11,7x95
15054

12x75
15120

12x85
15419

16x108
15053

Angle rotor (30) 11144 with


buckets 13081
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
13,3x30

14083
13,3x30

14082
14083
13,3x30
13,3x30
Brak dna
4
3120
5800
8,3
15
29

26

14082
13,3x30

14083
13,3x30

ANNEXES
Capacity [ml]
Tube
x L [mm]
Catalogue number

15

10

glass

Nalgene
PP

with cap
PP

10
glass

12x100
15119

16x113
15048

16x108
15053

12x75
15118

Angle rotor (30) 11145 with


buckets 13080
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
14083
13,3x30
13,3x30
lack of bottom
8
3120
5800
8,3
15
29

Capacity [ml]

10

Tube

with cap PP

glass

with cap PP

with cap
PP

x L [mm]
Catalogue number

11,7x95
15054

12x75
15120

12x85
15419

16x108
15053

Angle rotor (30) 11145 with


buckets 13081
Catalogue number
Hole x L [mm]
Shape of round carrier bottom
Quantity of tubes per rotor
Max RCF [g]
Max speed [rpm]
Centrifuging radius [mm]
Time of acceleration [s]
Time of decelaration [s]

14082
13,3x30

14083
13,3x30

14082
14083
13,3x30
13,3x30
lack of bottom
8
3120
5800
8,3
15
29

27

14082
13,3x30

14083
13,3x30

ANNEXES

ADDITIONAL ACCESSORIES:
INNE/OTHER
nr kat./cat no

opis

description

16696

Przetwornica
napicia
300W
umoliwiajca zasilanie prdem staym
o napiciu 12 V (m.in. z zapalniczki
samochodowej)

Power inverter 300W (12DC 230 AC) (e.g.


possibility of supply from the cigarette
lighter socket)

Only original MPW buckets can be used! One should use tubes which
dimensions and durability is proper!Using of tubes from other
companies should be consulted with manufacturer of centrifuge. In
the centrifuge, disinfectants and cleaning agents generally used in
medical care should be used (e.g. Aerodesina-2000, Lysoformin 3000,
Melseptol, Melsept SF, Sanepidex, Cutasept F).

28

DECLARATION OF CONFORMITY
Product

Laboratory centrifuge

Model

MPW-54

Product classification on the basis of


the Directive 98/79/EC

Non classified to list A or B and not for


self-testing

Product complies with the requirements:


Directive 98/79/EC (IVD), including the requirements of harmonised
standards:
PN-EN ISO 13485:2012

PN-EN ISO 18113-3:2011

PN-EN ISO 13485:2012/AC:2013-03

PN-EN 61010-2-101:2005

PN-EN 13612:2006

PN-EN 61326-2-6:2013-08

PN-EN ISO 14971:2012

PN-EN ISO 62366:2008

selected harmonized standards of Directive 2006/95/EC (LVD):


PN-EN 61010-1:2011

PN-EN 61010-2-020:2008

Directive 2004/108/WE (EMC)


standard PN-EN ISO 15223-1:2012

MPW MED. INSTRUMENTS


SPDZIELNIA PRACY
Warsaw, 46 Boremlowska Street
Quality policy in line with ISO 9001:2008
Certifying authority

Warsaw, 13.11.2014

nr 10.054.03

DECLARATION OF DECONTAMINATION

In order to protect our employees please fill out the declaration of decontamination
completely before sending centrifuge to the manufacturer (repair).

1.

2.

Device
type:

serial No.:

Description of decontamination
(see user manual)

3.

4.

Decontamination carried out by:


name:

Date and signature

DECLARATION OF DECONTAMINATION

In order to protect our employees please fill out the declaration of decontamination
completely before sending centrifuge to the manufacturer (return).

5.

6.

Device
type:

serial No.:

Description of decontamination
(see user manual)

7.

8.

Decontamination carried out by:


name:

Date and signature

S-ar putea să vă placă și